Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X Tia gamma tia X
Trang 1Gamma and X-Ray Detection
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
Figure 1.1 Efficiency Calibration
The choice of a particular detector type for an application depends
upon the X-ray or gamma energy range of interest and the
applica-tion’s resolution and efficiency requirements Additional
consider-ations include count rate performance, the suitability of the detector
for timing experiments, and of course, price.
DETECTOR EFFICIENCY
The efficiency of a detector is a measure of how many pulses occur
for a given number of gamma rays Various kinds of efficiency
defini-tions are in common use for gamma ray detectors:
a Absolute Efficiency: The ratio of the number of counts
pro-duced by the detector to the number of gamma rays emitted
by the source (in all directions).
b Intrinsic Efficiency: The ratio of the number of pulses
pro-duced by the detector to the number of gamma rays striking
the detector.
c Relative Efficiency: Efficiency of one detector relative to
an-other; commonly that of a germanium detector relative to a
3 in diameter by 3 in long NaI crystal, each at 25 cm from a
point source, and specified at 1.33 MeV only.
d Full-Energy Peak (or Photopeak) Efficiency: The efficiency
for producing full-energy peak pulses only, rather than a
pulse of any size for the gamma ray.
Clearly, to be useful, the detector must be capable of absorbing a
large fraction of the gamma ray energy This is accomplished by
us-ing a detector of suitable size, or by choosus-ing a detector material of
suitable high Z An example of a full-energy peak efficiency curve
for a germanium detector is shown in Figure 1.1.
DETECTOR RESOLUTION
Resolution is a measure of the width (full width half max) of a
single energy peak at a specific energy, either expressed in
ab-solute keV (as with Germanium Detectors), or as a percentage of
the energy at that point (Sodium Iodide Detectors) Better (lower
FWHM value) resolution enables the system to more clearly
sep-arate the peaks within a spectrum Figure 1.2 shows two
spec-tra collected from the same source, one using a sodium iodide
(NaI(TI)) detector and one using germanium (HPGe) Even though
this is a rather simple spectrum, the peaks presented by the
so-dium iodide detector are overlapping to some degree, while those
from the germanium detector are clearly separated In a complex
spectrum, with peaks numbering in the hundreds, the use of a
germanium detector becomes mandatory for analysis.
GAS-FILLED DETECTORS
A gas-filled detector is basically a metal chamber filled with gas and
containing a positively biased anode wire A photon passing through
the gas produces free electrons and positive ions The electrons are
attracted to the anode, producing an electric pulse.
At low anode voltages, the electrons may recombine with the ions
Recombination may also occur for a high density of ions At a
suffi-ciently high voltage nearly all electrons are collected, and the
detec-tor is known as an ionization chamber At higher voltages the trons are accelerated toward the anode at energies high enough to ionize other atoms, thus creating a larger number of electrons This detector is known as a proportional counter At higher voltages the electron multiplication is even greater, and the number of electrons collected is independent of the initial ionization This detector is the Geiger-Mueller counter, in which the large output pulse is the same for all photons At still higher voltages continuous discharge occurs The different voltage regions are indicated schematically in Figure 1.3 The actual voltages can vary widely from one detector to the next, depending upon the detector geometry and the gas type and pressure.
elec-IONIZATION CHAMBER
The very low signal output for the ionization chamber makes this detector difficult to use for detecting individual gamma rays It finds use in high radiation fluxes in which the total current produced can
be very large Many radiation monitoring instruments use ionization chambers Absolute ionization measurements can be made, using
an electrometer for recording the output 1
A detector is typically specified in terms of its physical size, fective window size and gas path length, operating voltage range and resolution for the 5.9 keV X ray from a 55 Fe source (Mn X ray) Typical resolutions are about 16 to 20% full-width at half maximum (FWHM).
Trang 2ef-Operating voltages depend upon the fill gas as well as the
geom-etry For X rays, noble gases are often used, with xenon, krypton,
neon and argon common choices Xenon and krypton are selected
for higher energy X rays or to get higher efficiencies, while neon
is selected when it is desired to detect low energy X rays in the
presence of unwanted higher energy X rays Sometimes gas
mix-tures are used, such as P-10 gas, which is a mixture of 90% argon
and 10% methane Gas pressures are typically one atmosphere
The 2006 preamplifier available for proportional counters is shown
in Figure 1.4.
GEIGER-MUELLER COUNTER
The Geiger-Mueller counter produces a large voltage pulse that is
easily counted without further amplification No energy
measure-ments are possible since the output pulse height is independent
of initial ionization Geiger-Mueller counters are available in a wide
variety of sizes, generally with a thin mica window The operating
voltage is in the plateau region (see Figure 1.3), which can be
rela-Figure 1.2
Figure 1.3 Gas Detector Output vs Anode Voltage
tively flat over a range of bias voltage The plateau is determined by measuring the counting rate as a function of the anode voltage The discharge produced by an ionization must be quenched in or- der for the detector to be returned to a neutral ionization state for the next pulse This is accomplished by using a fill gas that contains
a small amount of halogen in addition to a noble gas The voltage drop across a large resistor between the anode and bias supply will also serve to quench the discharge since the operating voltage will
be reduced below the plateau.
The Geiger-Mueller counter is inactive or “dead” after each pulse until the quenching is complete This dead time can be hundreds
of microseconds long, which limits the counter to low count rate applications.
Figure 1.4 Proportional Counter and Preamplifier
SCINTILLATION DETECTORS
A gamma ray interacting with a scintillator produces a pulse of light, which is converted to an electric pulse by a photomultiplier tube The photomultiplier consists of a photocathode, a focusing electrode and
10 or more dynodes that multiply the number of electrons striking them several times each The anode and dynodes are biased by a chain of resistors typically located in a plug-on tube base assembly Complete assemblies including scintillator and photomultiplier tube are commercially available from CANBERRA.
The properties of scintillation material required for good detectors are transparency, availability in large size, and large light output proportional to gamma ray energy Relatively few materials have good properties for detectors Thallium activated NaI and CsI crys- tals are commonly used, as well as a wide variety of plastics LaBr 3 (Ce) crystals are a newer type of scintillation detector material of- fering better resolution, but otherwise, similar characteristics to
Trang 3Figure 1.5 57 Co Spectrum from Counter
SEMICONDUCTOR DETECTORS
A semiconductor is a material that can act as an insulator or as a conductor In electronics the term “solid state” is often used inter- changeably with semiconductor, but in the detector field the term can obviously be applied to solid scintillators Therefore, semicon- ductor is the preferred term for those detectors which are fabricated from either elemental or compound single crystal materials having
a band gap in the range of approximately 1 to 5 eV The group IV elements silicon and germanium are by far the most widely-used semiconductors, although some compound semiconductor materi- als are finding use in special applications as development work on them continues.
Table 1.1 shows some of the key characteristics of various ductors as detector materials:
semicon-Table 1.1 Element vs Band Gap
pair (eV)
Si Ge CdTe HgI 2 GaAs
14 32 48-52 80-53 31-33
1.12 0.74 1.47 2.13 1.43
3.61 2.98 4.43 6.5 5.2 Semiconductor detectors have a p-i-n diode structure in which the intrinsic (i) region is created by depletion of charge carriers when
a reverse bias is applied across the diode When photons interact within the depletion region, charge carriers (holes and electrons) are freed and are swept to their respective collecting electrode by the electric field The resultant charge is integrated by a charge sen- sitive preamplifier and converted to a voltage pulse with an ampli- tude proportional to the original photon energy.
Since the depletion depth is inversely proportional to net electrical impurity concentration, and since counting efficiency is also depen- dent on the purity of the material, large volumes of very pure mate- rial are needed to ensure high counting efficiency for high energy photons.
NaI detector crystals NaI is still the dominant material for gamma
detection because it provides good gamma ray resolution and is
economical However, plastics have much faster pulse light decay
and find use in timing applications, even though they often offer little
or no energy resolution.
NaI(Tl) SCINTILLATION DETECTORS
The high Z of iodine in NaI gives good efficiency for gamma ray
detection A small amount of Tl is added in order to activate the
crystal, so that the designation is usually NaI(Tl) for the crystal
The best resolution achievable ranges from 7.5%-8.5% for the 662
keV gamma ray from 137 Cs for 3 in diameter by 3 in long crystal,
and is slightly worse for smaller and larger sizes Figure 1.7 shows,
respectively, the absorption efficiencies of various thicknesses
of NaI crystals and the transmission coefficient through the most
commonly used entrance windows Many configurations of NaI
de-tectors are commercially available, ranging from crystals for X-ray
measurements in which the detector is relatively thin (to optimize
resolution at the expense of efficiency at higher energies), to large
crystals with multiple phototubes Crystals built with a well to allow
nearly spherical 4π geometry counting of weak samples are also a
widely-used configuration A typical preamplifier and amplifier
com-bination is shown in Figure 1.6.
Figure 1.6 NaI(Tl) Detector Electronics
The light decay time constant in NaI is about 0.25 microseconds,
and typical charge sensitive preamplifiers translate this into an
output pulse rise time of about 0.5 microseconds For this reason,
NaI detectors are not as well-suited as plastic detectors for fast
coincidence measurements, where very short resolving times are
required LaBr3 (Ce) detectors have a light decay time constant of
0.03 microseconds making them another possible solution for
coin-cidence measurements.
Trang 4Prior to the mid-1970’s the required purity levels of Si and Ge could
be achieved only by counter-doping p-type crystals with the n-type
impurity, lithium, in a process known as lithium-ion drifting Although
this process is still widely used in the production of Si(Li) X-ray
detectors, it is no longer required for germanium detectors since
sufficiently pure crystals have been available since 1976.
The band gap figures in Table 1.1 signify the temperature
sensitiv-ity of the materials and the practical ways in which these materials
can be used as detectors Just as Ge transistors have much lower
maximum operating temperatures than Si devices, so do Ge
detec-tors As a practical matter both Ge and Si photon detectors must
be cooled in order to reduce the thermal charge carrier generation
(noise) to an acceptable level This requirement is quite aside from
the lithium precipitation problem which made the old Ge(Li), and to
some degree Si(Li) detectors, perishable at room temperature.
The most common medium for detector cooling is liquid nitrogen,
however, recent advances in electrical cooling systems have made
electrically refrigerated cryostats a viable alternative for many
detector applications.
In liquid nitrogen (LN 2 ) cooled detectors, the detector element (and
in some cases preamplifier components), are housed in a clean
vacuum chamber which is attached to or inserted in a LN 2 Dewar
The detector is in thermal contact with the liquid nitrogen which
cools it to around 77 °K or –200 °C At these temperatures, reverse
leakage currents are in the range of 10 -9 to 10 -12 amperes.
Figure 1.7
In electrically refrigerated detectors, both closed-cycle mixed frigerant and helium refrigeration systems have been developed to eliminate the need for liquid nitrogen Besides the obvious advan- tage of being able to operate where liquid nitrogen is unavailable or supply is uncertain, refrigerated detectors are ideal for applications requiring long-term unattended operation, or applications such as undersea operation, where it is impractical to vent LN 2 gas from a conventional cryostat to its surroundings.
re-A cross-sectional view of a typical liquid nitrogen cryostat is shown
in Figure 1.8.
DETECTOR STRUCTURE
The first semiconductor photon detectors had a simple planar ture similar to their predecessor, the Silicon Surface Barrier (SSB) detector Soon the grooved planar Si(Li) detector evolved from attempts to reduce leakage currents and thus improve resolution The coaxial Ge(Li) detector was developed in order to increase overall detector volume, and thus detection efficiency, while keep- ing depletion (drift) depths reasonable and minimizing capacitance Other variations on these structures have come, and some have gone away, but there are several currently in use These are il- lustrated in Figure 1.9 with their salient features and approximate energy ranges.
struc-For more information on specific detector types refer to the Detector Product Section of this catalog.
Trang 5DETECTOR PERFORMANCE
Semiconductor detectors provide greatly improved energy
resolu-tion over other types of radiaresolu-tion detectors for many reasons
Fun-damentally, the resolution advantage can be attributed to the small
amount of energy required to produce a charge carrier and the
con-sequent large “output signal” relative to other detector types for the
same incident photon energy At 3 eV/e-h pair (see Table 1.1) the
number of charge carriers produced in Ge is about one and two
or-ders of magnitude higher than in gas and scintillation detectors
re-spectively The charge multiplication that takes place in proportional
counters and in the electron multipliers associated with scintillation
detectors, resulting in large output signals, does nothing to improve
the fundamental statistics of charge production.
The resultant energy reduction in keV (FWHM) vs energy for ous detector types is illustrated in Table 1.2.
vari-Table 1.2 Energy Resolution (keV FWHM)
vs Detector Type
Proportional Counter X-ray NaI(Tl)
3 x 3 NaI(Tl) Si(Li) Low Energy Ge Coaxial Ge
1.2 3.0
— 0.16 0.14
—
— 12.0 12.0
— 0.5 0.8
—
— 60
—
— 1.8
At low energies, detector efficiency is a function of cross-sectional area and window thickness while at high energies total active detec- tor volume more or less determines counting efficiency Detectors having thin contacts, e.g Si(Li), Low-Energy Ge and Reverse Elec- trode Ge detectors, are usually equipped with a Be or composite carbon cryostat window to take full advantage of their intrinsic energy response.
Coaxial Ge detectors are specified in terms of their relative full- energy peak efficiency compared to that of a 3 in x 3 in NaI(Tl) Scintillation detector at a detector to source distance of 25 cm De- tectors of greater than 100% relative efficiency have been fabricated from germanium crystals ranging up to about 75 mm in diameter About two kg of germanium is required for such a detector.
Curves of detector efficiency vs energy for various types of Ge detectors can be found in the Detector Product Section of this catalog.
Figure 1.8 Model 7500SL Vertical Dipstick Cryostat
Figure 1.9 Detector Structures and Energy Ranges
1 A.C Melissinos, Experiments in Modern Physics, Academic Press, New York (1966), p 178.
Trang 6Charged Particle Detection
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
SILICON CHARGED PARTICLE DETECTORS
Silicon Charged Particle detectors have a P-I-N structure in which
a depletion region is formed by applying reverse bias, with the
re-sultant electric field collecting the electron-hole pairs produced by
an incident charged particle The resistivity of the silicon must be
high enough to allow a large enough depletion region at moderate
bias voltages A traditional example of this type of detector is the
Silicon Surface Barrier (SSB) detector In this detector, the n-type
silicon has a gold surface-barrier contact as the positive contact,
and deposited aluminum is used at the back of the detector as the
ohmic contact.
A modern version of the charged particle detector is the CANBERRA
PIPS ® detector (Passivated Implanted Planar Silicon) This
detec-tor employs implanted rather than surface barrier contacts and is
therefore more rugged and reliable than the Silicon Surface Barrier
(SSB) detector it replaces.
At the junction there is a repulsion of majority carriers (electrons in
the n-type and holes in p-type) so that a depleted region exists An
applied reverse bias widens this depleted region which is the
sensi-tive detector volume, and can be extended to the limit of breakdown
voltage Detectors are generally available with depletion depths of
100 to 700 µm.
Detectors are specified in terms of surface area and alpha or beta
particle resolution as well as depletion depth The resolution
de-pends largely upon detector size, being best for small area
detec-tors Alpha resolution of 12 to 35 keV and beta resolutions of 6 to
30 keV are typical Areas of 25 to 5000 mm 2 are available as
stan-dard, with larger detectors available in various geometries for
cus-tom applications Additionally, PIPS detectors are available fully
depleted, so that a dE/dx energy loss measurement can be made
by stacking detectors on axis Detectors for this application are
sup-plied in a transmission mount, (i.e with the bias connector on the
side of the detector).
A chart of the energies of various particles measured at several
depletion depths is shown in Table 1.3 Note that even the thinnest
detector is adequate for alpha particles from radioactive sources,
but that only very low energy electrons are fully absorbed However,
for a detector viewing a source of electron lines, such as conversion
electron lines, sharp peaks will be observed since some electron
path lengths will lie fully in depleted region Figure 1.10 shows
rang-es of particlrang-es commonly occurring in nuclear reactions.
Table 1.3 Particle Ranges and PIPS Depletion Depth
Maximum Particle Energy Depletion
7
15
21
27 33
15
55
85
105 130 Since charge collected from the particle ionization is so small that it
is impractical to use the resultant pulses without intermediate
am-plification, a charge-sensitive preamplifier is used to initially prepare
the signal.
Figure 1.11 illustrates the electronics used in single-input alpha spectroscopy application Note that the sample and detector are lo- cated inside a vacuum chamber so that the energy loss in air is not involved.
LIQUID SCINTILLATORS
Two very important beta-emitting isotopes, tritium and 14 C, have very low energy beta rays These are at 19 and 156 keV respec- tively, too low to detect reliably with solid scintillators The liquid scintillation technique involves mixing a liquid scintillator with the sample, and then observing the light pulses with one or more pho- tomultiplier tubes The efficiency of such a counter is virtually 100% – essentially 4π geometry with no attenuation between source and detector Pulse processing of the resultant Photomultiplier outputs allows the rejection of cosmic events, and the separation, if desired,
of alpha and beta events The increased sensitivity of the Liquid Scintillation counter, coupled with advances in sample preparation techniques, has led to its increasing use for low-level alpha and beta measurements.
Figure 1.10 Range-Energy Curves in Silicon
Figure 1.11
©2006 Canberra Industries, Inc All rights reserved.
Trang 7Basic Counting Systems
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
PULSE ELECTRONICS
The nuclear electronics industry has standardized the signal
defi-nitions, power supply voltages and physical dimensions of basic
nuclear instrumentation modules using the Nuclear Instrumentation
Methods (NIM) standard initiated in the 1960s This
standardiza-tion provides users with the ability to interchange modules, and the
flexibility to reconfigure or expand nuclear counting systems, as
their counting applications change or grow CANBERRA is a
lead-ing supplier of Nuclear Instrumentation Modules (also called NIM),
which are presented in greater detail in Section 1 of this catalog In
the past several years, the industry trend has been to offer modular
detector electronics with the multichannel analyzer (MCA) and all
supporting instrumentation for spectroscopy with a single
detec-tor combined in a compact, stand-alone enclosure These modular
MCAs are smaller, lighter and use less power than the NIM-based
counting systems that preceded them However, their performance
is equal to, or greater than, comparable NIM-based systems
CANBERRA is also a leading supplier of these modular detector
electronics which are described in the Multichannel Analyzers
Sec-tion of this catalog Depending on the applicaSec-tion and budget, NIM
or modular electronics may be the best counting equipment solution
for the user, and CANBERRA supports both of these form factors
with a wide variety of products.
Basic electronic principals, components and configurations which
are fundamental in solving common nuclear applications are
discussed below.
PREAMPLIFIERS AND AMPLIFIERS
Most detectors can be represented as a capacitor into which a charge is deposited, (as shown in Figure 1.12) By applying detec- tor bias, an electric field is created which causes the charge carriers
to migrate and be collected During the charge collection a small current flows, and the voltage drop across the bias resistor is the pulse voltage.
The preamplifier is isolated from the high voltage by a capacitor The rise time of the preamplifier’s output pulse is related to the collection time of the charge, while the decay time of the preamplifier’s output pulse is the RC time constant characteristic of the preamplifier itself Rise times range from a few nanoseconds to a few microseconds, while decay times are usually set at about 50 microseconds Charge-sensitive preamplifiers are commonly used for most solid state detectors In charge-sensitive preamplifiers, an output voltage pulse is produced that is proportional to the input charge The output voltage is essentially independent of detector capacitance, which is especially important in silicon charged particle detection (i.e PIPS ® detectors), since the detector capacitance depends strongly upon the bias voltage However, noise is also affected by the capaci- tance.
To maximize performance, the preamplifier should be located at the detector to reduce capacitance of the leads, which can degrade the rise time as well as lower the effective signal size Additionally, the preamplifier also serves to provide a match between the high im- pedance of the detector and the low impedance of coaxial cables
to the amplifier, which may be located at great distances from the preamplifier.
The amplifier serves to shape the pulse as well as further amplify it The long delay time of the preamplifier pulse may not be returned
to zero voltage before another pulse occurs, so it is important to shorten it and only preserve the detector information in the pulse rise time The RC clipping technique can be used in which the pulse
is differentiated to remove the slowly varying decay time, and then integrated somewhat to reduce the noise The unipolar pulse that results is much shorter The actual circuitry used is an active filter for selected frequencies A near-Gaussian pulse shape is produced, yielding optimum signal-to-noise characteristics and count rate performance.
Figure 1.12 Basic Detector and Amplification
Figure 1.13 Standard Pulse Waveforms
Trang 8A second differentiation produces a bipolar pulse This bipolar pulse
has the advantage of nearly equal amounts of positive and negative
area, so that the net voltage is zero When a bipolar pulse passes
from one stage of a circuit to another through a capacitor, no charge
is left on the capacitor between pulses With a unipolar pulse, the
charge must leak off through associated resistance, or be reset to
zero with a baseline restorer.
High performance gamma spectrometers are often designed today
using Digital Signal Processing (DSP) techniques rather than
ana-log shaping amplifiers The shaping functions are then performed in
the digital domain rather than with analog circuitry This is discussed
later in this section.
Typical preamplifier and amplifier pulses are shown in Figure 1.13
The dashed line in the unipolar pulse indicates undershoot which can
occur when, at medium to high count rates, a substantial amount of
the amplifier’s output pulses begin to ride on the undershoot of the
previous pulse If left uncorrected, the measured pulse amplitudes
for these pulses would be too low, and when added to pulses whose
amplitudes are correct, would lead to spectrum broadening of peaks
in acquired spectra To compensate for this effect, pole/zero
cancel-lation quickly returns the pulse to the zero baseline voltage.
The bipolar pulse has the further advantage over unipolar in that
the zero crossing point is nearly independent of time (relative to the
start of the pulse) for a wide range of amplitudes This is very useful
in timing applications such as the ones discussed below However,
the unipolar pulse has lower noise, and constant fraction
discrimina-tors have been developed for timing with unipolar pulses.
For further discussions on preamplifier and amplifier characteristics,
please refer to each applicable module’s subsection.
Figure 1.14 Multichannel Analyzer Components with Analog Signal Processing
PULSE HEIGHT ANALYSIS AND COUNTING TECHNIQUES
Pulse Height Analysis may consist of a simple discriminator that can be set above noise level and which produces a standard log-
ic pulse (see Figure 1.13) for use in a pulse counter or as gating signal However, most data consists of a range of pulse heights of which only a small portion is of interest One can employ either of the following:
1 Single Channel Analyzer and Counter
2 Multichannel Analyzer The single channel analyzer (SCA) has a lower and an upper level discriminator, and produces an output logic pulse whenever an in- put pulse falls between the discriminator levels With this device, all voltage pulses in a specific range can be selected and counted If additional voltage ranges are of interest, additional SCAs and coun- ters (i.e scalers) can be added as required, but the expense and complexity of multiple SCAs and counters usually make this con- figuration impractical.
If a full voltage (i.e energy) spectrum is desired, the SCA’s nators can be set to a narrow range (i.e window) and then stepped through a range of voltages If the counts are recorded and plotted for each step, an energy spectrum will result In a typical example
discrimi-of the use discrimi-of the Model 2030 SCA, the lower level discriminator (LLD) and window can be manually or externally (for instance, by
a computer) incremented, and the counts recorded for each step However, the preferred method of collecting a full energy spectrum
is with a multichannel analyzer.
The multichannel analyzer (MCA), which can be considered as a series of SCAs with incrementing narrow windows, basically con- sists of an analog-to-digital converter (ADC), control logic, memory and display The multichannel analyzer collects pulses in all voltage ranges at once and displays this information in real time, providing
a major improvement over SCA spectrum analysis.
Trang 9Figure 1.14 illustrates a typical MCA block diagram An input energy
pulse is checked to see if it is within the selected SCA range, and
then passed to the ADC The ADC converts the pulse to a number
proportional to the energy of the event This number is taken to be
the address of a memory location, and one count is added to the
contents of that memory location After collecting data for some
pe-riod of time, the memory contains a list of numbers corresponding
to the number of pulses at each discrete voltage The memory is
accessed by a host computer which is responsible for spectrum
dis-play and analysis as well as control of the MCA Depending on the
specific model MCA, the host computer may be either a dedicated,
embedded processor or a standard off-the-shelf computer.
PULSE HEIGHT ANALYSIS WITH DIGITAL SIGNAL
PROCESSORS
Today’s high performance Multichannel Analyzer systems are
de-signed using Digital Signal Processing (DSP) techniques rather
than the traditional analog methods DSP filters and processes the
signals using high speed digital calculations rather than
manipula-tion of the time varying voltage signals in the analog domain The
preamplifier signal first passes through an analog differentiator,
then is delivered to a high speed digitizing ADC (Figure 1.15) The
output of the ADC is a series of digital values that represent the
dif-ferentiated pulse Those signals are then filtered using high-speed
digital calculations within the Digital Signal Processor.
For optimal speed and accuracy in signal processing, a trapezoidal
filter algorithm is deployed in the DSP implementation Trapezoidal
filtering has been shown to allow for the highest throughput
perfor-mance with the least degradation of spectral resolution
Addition-ally, the DSP based design is intrinsically more stable, resulting in
better performance over a range of environmental conditions.
COUNTERS AND RATEMETERS
Counters and ratemeters are used to record the number of logic
pulses, either on an individual basis as in a counter, or as an
aver-age count rate as in a ratemeter Counters and ratemeters are built
with very high count rate capabilities so that dead times are
mini-mized Counters are usually used in combination with a timer (either
Figure 1.15 Multichannel Analyzer Components with Digital Signal Processing
Figure 1.16 NaI Detector and Counter/Timer
with Alarm Ratemeter
built-in, or external), so that the number of pulses per unit of time are recorded Ratemeters feature a built-in timer, so that the count rate per unit of time is automatically displayed Whereas counters have
an LED or LCD for the number of logic pulses, ratemeters have a mechanical meter for real-time display of the count rate Typically, most counters are designed with 8-decade count capacity and offer
an optional external control/output interface, while ratemeters are designed with linear or log count rate scales, recorder outputs and optional alarm level presets/outputs Additional information may be found in the Counters and Ratemeters Introduction.
SIMPLE COUNTING SYSTEMS
As related above, pulse height analysis can consist of a simple gle channel analyzer and counter, or a multichannel analyzer Gen- erally, low resolution/high efficiency detectors (such as proportional counters and NaI(Tl) detectors) are used in X ray or low-energy gamma ray applications where only a few peaks occur An example
sin-of a simple NaI(Tl) detector-based counting system sin-of this type is illustrated in Figure 1.16.
Trang 10In this configuration, a Model 2015A Amplifier/SCA is used to
gen-erate a logic pulse for every amplified (detector) pulse that falls
within the SCA’s “energy window” The logic pulse is then used as
an input to the Model 512 Counter/Timer which provides the user
with a choice of either preset time or preset count operation The
Model 512 is equipped with an RS-232 interface, which enables
it to be controlled and read out to a computer for data storage or
further analysis.
Alternatively, Model 1481LA Linear/Log Ratemeter is used as the
counter, with an alarm relay that will trigger if the count rate exceeds
a user preset value.
Although counters are still used in some applications, most of
today’s counting systems include a multichannel analyzer (MCA)
Besides being more cost effective than multiple SCA-based
sys-tems, a MCA-based system can provide complete pulse height
analysis such that all nuclides, (i.e., even those not expected), can
be easily viewed and/or analyzed.
NaI(Tl) DETECTORS AND MULTICHANNEL ANALYZERS
The need for a single-input Pulse Height Analysis system for use with
a Sodium Iodide detector is served most simply by a photomultiplier
tube (PMT) base MCA such as the uniSpec (Figure 1.17) The
uni-Spec MCA includes a high voltage power supply, preamplifier,
am-plifier, spectrum stabilizer and ADC in addition to its MCA functions,
and thus, there is no need for any NIM modules or a NIM Bin All of
Figure 1.18 HPGe Detector and Analog MCA Configuration
this capability is provided in an enclosure no larger than a standard tube base preamplifier, and the computer interface is via a USB port
on the host computer or a USB hub Further technical discussions concerning multichannel analyzers and multichannel analysis sys- tems (including spectroscopy software) may be found in the Multi- channel Analyzers and Counting Room Software sections.
GERMANIUM DETECTORS AND MULTICHANNEL ANALYZERS
A typical analog HPGe detector-based gamma spectroscopy tem consists of a HPGe detector, high voltage power supply, pream- plifier (which is usually sold as part of the detector), amplifier, ADC and multichannel analyzer As will be discussed in more detail later, DSP configurations replace the amplifier and ADC with digital signal processing electronics.
sys-The analog system components are available in several different types, allowing the system to be tailored to the precise needs of the application and the available budget For example, low-end ampli- fiers such as the Model 2022 offer basic capabilities, but users with higher count rate or resolution requirements may consider the Mod-
el 2026 or 2025 with Pileup Rejection/Live Time Correction (PUR/ LTC) feature and both Gaussian and triangular shaping Similarly, the ADC chosen for a system including a 556A NIM MCA could be either an economical Wilkinson ADC like the Model 8701 or a faster Fixed Dead Time (FDT) ADC like Model and 8715 For more Figure 1.17 NaI Detector and MCA Configuration
Trang 11information about selecting specific modules, refer to the
introduc-tion secintroduc-tions for those specific components.
For applications requiring security of the signal processing,
CANBERRA offers a variety of computer controlled electronics
which require access via a host computer, rather than unprotected
front panel for adjustment For example, the AIM/ICB NIM family is
a network based, computer controlled signal processing line that
can be controlled remotely by a Genie 2000 or Genie-ESP
spec-troscopy workstation.
Spectroscopy systems based on Digital Signal Processing (DSP)
have been widely accepted as the state of the art In a DSP based
system, the amplifier and ADC are replaced by a set of digital
cir-cuits which implement the filtering functions in high speed digital
calculations CANBERRA offers several DSP based products, all
of which offer superior environmental stability, higher count rate
throughput performance and better resolution over a range of count
rate conditions Models 2060, 9660, DSA-1000, DSA-2000 and the
InSpector 2000 all employ this advanced DSP technology.
Figures 1.18, 1.19 and 1.20 show several of the available
Germanium Detectors/MCA configurations Optional LN2 Monitors,
Level Alarms, and Control Systems are available for most types of
be obtained with the Model 2025 AFT Research Amplifier Besides allowing the user to select a long shaping time constant, the Model
2025 features an enhanced baseline restorer which is ideal for set preamplifiers Any of the CANBERRA Digital Signal Processing MCAs or components can be used with these detectors and provide even better throughput and resolution performance.
re-MULTIPLE INPUT SYSTEMS
CANBERRA offers two solutions for multiple input counting systems which process the amplified signals from a number of detectors A multiple input scenario would typically be considered six or more detector inputs – or the point at which multiple independent MCA systems become cost prohibitive for a given counting application The Multiport II (Figure 1.18) is the first solution and, also, the more robust of the two It offers the capability for up to six totally indepen- dent MCAs and ADCs housed in one double-wide NIM Because the MCAs and ADCs are separate from each other, any combina- tion of detectors and channel number settings may be used for each input.
Figure 1.20 HPGe Detector with DSA-2000 Digital Signal Processor (DSP) Figure 1.19 HPGe Detector with DSA-1000 or InSpector 2000 Digital Signal Processor (DSP)
Trang 12The second solution employs a Model 8224 Multiplexer (or Mixer/
Router) to route the signals from multiple detectors to a single ADC
for digitizing and on to a 556A MCA for processing as shown in
Figure 1.21 Since this configuration shares the MCA and ADC
among the detectors, it has a lower cost per input than the
Multi-port II – particularly for large numbers of detectors However, the
Multiplexer configuration has a major drawback due to the single
ADC; the count rate of the individual detectors must be relatively
low to avoid excessive signal pileup Additionally, a Multiplexer must
allocate the memory of the MCA to its various inputs (same amount
for each input), which decreases the number of channels available
for each individual detector Within these constraints, Multiplexers
can be quite efficient for applications such as low-level
environmen-tal alpha spectroscopy in which multiple low-intensity inputs are
col-lected in MCA memory segments of 512 channels or less Low-level
gamma counting with NaI detectors, which typically don’t need more
than 1024 channels, is another application that can make use of a
Multiplexer An example configuration is depicted in Figure 1.21.
It should be noted that the Multiport II and the 8224 Multiplexer do
not include spectroscopy amplifiers or detector bias supplies These
components must be supplied by other parts of the signal chain
Also, these two solutions do not include the benefits of Digital
Sig-nal Processing
Advances in electronics technology have dramatically lowered the
cost of MCAs, so that today, it is frequently more effective to use
multiple complete MCA systems (or the Multiport II) in place of a
Multiplexer.
LOW LEVEL GAMMA RAY COUNTING
Large volume HPGe detectors have become dominant over other
detector types for low level gamma ray spectroscopy because of
their inherently good resolution and linearity It is only in the
analy-sis of single radionuclides that NaI(Tl) detectors can compare in
sensitivity with HPGe detectors Since the majority of all gamma
spectroscopy applications require the analysis of more complex,
multi-radionuclide samples, the following discussion will be limited
to the application of HPGe detectors to low level counting.
The sensitivity of a HPGe spectrometer system depends on
sev-eral factors, including detector efficiency, detector resolution,
background radiation, sample constituency, sample geometry and
counting time The following paragraphs discuss the role these
fac-tors play in low level gamma ray counting.
1 EFFICIENCY: Generally, the sensitivity of a HPGe system will be in direct proportion to the detector efficiency HPGe detectors are almost universally specified for efficiency rela- tive to a 3 in NaI(Tl) at 25 cm detector-to-source distance at 1.33 MeV, and from this benchmark one may roughly com- pute the efficiency at lower energies However, for the cus- tomer who is counting weak samples with lower gamma energies, for instance 100-800 keV, the following subtle con- siderations to the detector design are important to system performance:
a The detector should have an adequate diameter This sures that the efficiency at medium and low energies will
as-be high relative to the efficiency at 1.33 MeV, where it is bought and paid for.
b The detector-to-end-cap distance should be minimal – five millimeters or less The inverse square law is real and will affect sensitivity.
c The detector should be of closed end coaxial geometry, to assure that the entire front face is active.
2 RESOLUTION: Generally, the superior resolution of a HPGe detector is sufficient enough to avoid the problem of peak convolution, (i.e., all peaks are separate and distinct) The sensitivity of a system improves as the resolution improves because higher resolution means that spectral line widths are smaller, and fewer background counts are therefore in- volved in calculating peak integrals.
Since the sensitivity is inversely related to the square root of the background, that is,
improvements in resolution will not improve sensitivity as dramatically as increased efficiency.
Figure 1.21 Multiple Input NaI Detector System
Sensitivity = 1
Bkg
√
Trang 133 BACKGROUND RADIATION AND SAMPLE
CONSTITU-ENCY: Interfering background in gamma spectra originates
either from within the sample being counted
(Compton-pro-duced) or from the environment If the sample being
ana-lyzed has a high content of high-energy gamma emitting
radioisotopes, the Compton-produced background will
eas-ily outweigh the environmental background For extremely
weak samples, the environmental background becomes
more significant Obviously, massive shielding will do little to
improve system sensitivity for low energy gamma rays in the
presence of relatively intense higher energy radiation
How-ever, Compton-suppression can be very effective in reducing
this background.
4 SAMPLE GEOMETRY: An often overlooked aspect of HPGe
detector sensitivity is the sample geometry For a given
sam-ple size (and the samsam-ple size should be as a large as
prac-ticable for maximum sensitivity), the sample should be
dis-tributed so as to minimize the distance between the sample
volume and the detector itself.
This rules out analyzing “test tube” samples with non-well
type detectors, or “large area flat samples” with standard
de-tectors It does rule in favor of using re-entrant or
Marinelli-beaker-type sample containers, which distribute part of the
sample around the circumference of the detector.
GERMANIUM DETECTORS WITH INERT SHIELDS
There are many different types of shield designs that are available,
and because of the difficulty in determining the background
contri-bution of the materials used in a given shield, it is difficult to assign
performance levels to various types of shields However, some
crite-ria for shield designs have evolved over the years, such as:
1 The shield should not be designed to contain unnecessary
components like the Dewar It will only contribute to increased
background if it is within the walls of the shield, as well as
unnecessarily increase the shield’s size, weight and cost.
2 The detector should be readily installed and removable from
the shield.
Pity the person who has to move lead bricks (at 12 kg each)
to disengage a HPGe detector A HPGe detector and shield
system should have a liquid nitrogen transfer system to avoid
removing the detector for the weekly filling.
3 Sample entry should be convenient to the operator.
4 The shield should accommodate a variety of sample sizes
and configurations.
The HPGe detector should be located in the center of the
shield so as to minimize scatter from the walls In this
posi-tion, the shield must accommodate the largest sample that
is anticipated Also, sample placement should be accurately
repeatable and easily verified by the operator.
The shield design that has all these features and is moderately
priced is the CANBERRA Model 747 Lead Shield illustrated in
Trang 14The performance of the shield using a CANBERRA HPGe detector
Resolution 12%1.95 keV FWHM at 1.33 MeV
0.90 keV FWHM at 1.22 keV Background
Count: 2.25 counts/second in the 50 keV–2.7 MeV range
Sensitivity: Table 1.4 lists the sensitivities of several single
radioisotopes, assuming a counting time of
50 000 seconds, a 50% error and a
2
3
6 10
LOW BACKGROUND CRYOSTATS
The design or configuration of the cryostat is another factor in
sys-tem performance Some cryostat/shield designs do not prevent
streaming from the outside environment, nor do they provide
self-shielding from their own relatively hot components Through an
improper choice of material types and/or thicknesses, the cryostat
may appreciably contribute to the background CANBERRA has
developed sources for select, low-background, materials, and has
invested in the design and fabrication of low-background cryostats,
as described in the Introduction to the Cryostats and Cryostat
Options Section.
HPGe COMPTON SUPPRESSION SPECTROMETER
When the ultimate in low level counting is required, a Compton Suppression Spectrometer, in conjunction with an appropriate low- background shield/cryostat design, is the answer In this configura- tion, the HPGe detector is surrounded by an active NaI(Tl) or plastic scintillation guard detector (also known as an annulus detector), with the electronics configured in an anticoincidence counting mode The Compton continuum, which is primarily caused by gamma rays which sustain one or more inelastic collisions and escape (i.e scat- ter out of) the germanium detector material without imparting their full energy, can lead to concealment of low activity peaks Since this is undesirable in low level counting applications, a Compton Suppression Spectrometer can be used to gate (i.e turn off) data acquisition whenever one of the incompletely absorbed photons es- capes the germanium detector and is “seen” by the annulus detec- tor When acquisition is complete, the resultant spectrum contains only peaks attributed to gamma rays which have imparted their full energy within the detector material.
It should be pointed out that some radioisotopes (those having incident gamma rays) such as 60 Co, will not be analyzed properly
co-by the anticoincidence spectrum from a Compton Suppression tem Therefore, two spectra are usually obtained from such a spec- trometer – one in the anticoincidence mode, and the other in the normal (ungated) mode.
Sys-Figure 1.24 illustrates a typical example of a Compton Suppression System.
One type of annulus has six (6) 5.08 cm (2-inch) diameter multiplier tubes (PMTs) on one end, and a 7.62 cm (3-inch) diam- eter NaI(Tl) plug with one PMT (which is operated in parallel with the other PMT) on the other end A simpler type of annulus detector uses a 15.24 cm (6-inch) diameter NaI(Tl) well detector on a single PMT In either configuration, the annulus must be large enough to allow the insertion of the HPGe detector’s endcap along with the sample.
photo-Figure 1.24 Compton Suppression System
Trang 15While some endorse the use of a fairly complex Timing Chain to
derive the anti-Compton gate signal, CANBERRA has found that
the simplified circuit shown in Figure 1.24 yields equivalent results 2
The “Incoming Count Rate” signals from the Spectroscopy
Ampli-fiers are checked for coincidence, and, if it exists, the 2040
Co-incidence Analyzer’s output is used as an anti-coCo-incidence input
to the ADC’s Gate When coincidence occurs, this gate “turns off”
the delayed unipolar signal from the Spectroscopy Amplifier Typical
Compton Suppression Spectrometer results are illustrated in Figure
1.25 It can be seen that the ‘figure of merit’ – the value of the 137 Cs
peak at 662 keV divided by the average contents of the Compton
continuum (the energy range 358-382 keV) – is on the order of
1000:1.
HIGH COUNT RATE GAMMA RAY SYSTEMS
High count rate applications require special techniques to assure
good resolution and/or good throughput In general, “high count
rate” is used to refer to incoming count rate, that is, the number
of events seen by the detector The term “throughput rate” may be
of more interest to the researcher, being a measure of the rate at
which the system can accurately process these incoming counts
In high count rate HPGe detector applications, problems such as
the loss of resolution, excessively long counting times, erroneous
peak to background ratios, inaccurate counting statistics or system
shutdown due to overload and saturation begin to appear In some
experiments, the solution to these problems merely lies in reducing
the incoming count rate to the detector, or by employing
electron-ics which inhibit the processing of pulses through the electronelectron-ics
when events are occurring so fast that they are overlapping (pulse
pileup) In this latter solution, system throughput will of course be
reduced, but parameters such as resolution will be enhanced Table
1.5 indicates the throughput limitations of the major components of
a spectroscopy chain Note that the term “energy rate limited” refers
to the fact that the component’s performance is not only affected
by the incoming count rate, but by the relative energy (amplitude)
of the incoming counts as well Each element in the chain can be
2 Compton Suppression Made Easy, Application Note
Figure 1.25 Ge Spectra with Compton Suppression
optimized for high count rate performance.
or more sophisticated peak shapes.
Some ways to address high count rate in the detector include ing the detector farther away from the source, or collimating the de- tector – which in both cases reduces the number of events seen by the detector – or using a detector of lesser efficiency The detector
mov-in the latter case will ‘see’ fewer events, and furthermore will have a lower charge collection time.
THE PREAMPLIFIER
Most Germanium detectors in use today are equipped with feedback, charge sensitive preamplifiers In the RC-feedback pre- amplifier, a feedback resistor discharges the integrator, typically in one or two milliseconds If the incoming energy rate (count rate X energy/count) produces a current that exceeds the capability of the resistor to bleed it off, the output will increase until, in the extreme, the preamplifier saturates and ceases to operate This limit occurs
RC-at approximRC-ately 200k MeV/s The sRC-aturRC-ated condition remains til the count rate is reduced The saturation limit is dependent on both energy and count rate and is usually specified in terms of the
un-“energy/rate limit” The energy/rate limit can be increased by ing the value of the feedback resistor, but this in turn allows more noise to pass through the preamplifier, resulting in a degradation in
lower-Table 1.5 Major System Components and their Throughput Limitations
Trang 16When a Coaxial Germanium detector is used in applications
requir-ing high throughput, the Model 2101 Transistor Reset Preamplifier
(TRP) is favored over traditional RC feedback Preamplifiers The
higher cost of the TRP is justified by its much higher energy rate
ca-pacity, an enhancement obtained by replacing the Feedback
Resis-tor of a typical RC feedback preamplifier with a special reset circuit
This circuit monitors the dc level of the preamplifier and discharges
the feedback capacitor whenever the output reaches a
predeter-mined reset threshold At moderate to high count rates (i.e above
20 000 cps), the absence of the feedback resistor and its attendant
noise and secondary time constant contributions lead to: 1) lower
preamplifier noise contributions, 2) inherently better resolution and
reduced spectrum broadening vs count rate, 3) elimination of the
need for pole/zero cancellation, and 4) elimination of ‘lock-up’ due
to saturation Figure 1.26 illustrates the throughout performance of
the two preamplifier styles.
Although the Model 2101 TRP virtually never shuts down due to
saturation, its reset process and the amplifier overload which it
causes does induce intervals of dead time into the counting system
The Model 2101 has been designed with a small Charge Gain (50
mV/MeV) and a wide Dynamic Range (4 V) to significantly reduce
the dead time due to resets in comparison to competitive units.
DIGITAL SIGNAL PROCESSOR
As we described in an earlier section, Digital Signal Processors
(DSP) have come to replace the analog shaping amplifier and ADC
in most high performance gamma spectroscopy systems It is in
applications involving high count rate performance where the
advantages of DSP become most pronounced.
In gamma spectroscopy systems, the DSP replaces the
functional-ity of both the shaping amplifier and the ADC The DSP first filters
the signal for optimum signal to noise ratio and to provide gain It
then detects the peak amplitude of the filtered pulse to calculate the
memory address of the MCA channel into which the event is to be
stored.
In the DSP, the analog signal from the preamplifier is first
differenti-ated in the analog domain to provide a rapid return to baseline This
is depicted in Figure 1.27 The resulting time varying voltage signal
is sampled by a high speed sampling analog to digital converter This
results in a digitized profile of the differentiated preamplifier signal
represented in internal memory of the DSP From this point on, the
signal is processed in the digital domain by the DSP – essentially
a high speed digital computer executing calculations as opposed
to analog circuits manipulating time varying voltage signals.
Figure 1.26 Throughput vs Count Rate:
Throughput Optimization
Figure 1.27 Typical Amplifier Pulses
Figure 1.28 Trapezoidal Pulse Waveform as processed in DSP
Processing the signals digitally allows more sophisticated filtering functions to be applied to the signal It also allows greater flexibility
to the user in terms of adjusting filtering parameters – more ble settings are available because they are handled as digital com- mands, not the selection of discrete analog components Finally, the use of high speed digital electronics allows the signals to be processed more rapidly, thus contributing further to the count rate performance of the system.
possi-CANBERRA’s DSP products deploy a trapezoidal filtering algorithm
as shown in Figure 1.28 Two parameters are available for user adjustment – the rise/fall time of the trapezoid (hereafter referred to
as rise time) and the flat top time
Adjusting the rise time changes the filter characteristics to optimize for noise characteristics The larger the rise time, the better the sig- nal to noise ratio Shorter rise times will adversely affect signal to noise ratio and degrade the resolution of the system Flat top ad- justments are made to accommodate the variations in pulse rise time which in turn is proportional to the charge collection time in the detector Larger detectors tend to have a larger number of long rise time (large charger collection time) events, thus requiring a lon- ger flat top time Failure to set the DSP rise time long enough to accommodate the longest charge collection time events results in degraded resolution, an effect known as ballistic deficit Note that for some types of smaller detectors, the flat top time can be set near or very close to zero, resulting in a triangular shape
Trang 17Figure 1.29 A comparison of the system throughput as a function
of input count rate for a DSP and an analog system optimized for
high throughput for a small detector (11%)
Figure 1.30 A comparison of the system resolution as a function
of input count rate for a DSP and an analog system optimized for maximum throughput for a small detector (11%)
Figure 1.31 A comparison of the system throughput as a function
of input count rate for an analog system optimized for maximum throughput with a DSP system set for a similar throughput
These two parameters together control the total event processing
time The total processing time for an event processed with the DSP
trapezoidal algorithm is defined by the equation:
T p = (2T r ) + T flat top
We see that the settings for both parameters effect the total
pro-cessing time, which in turn effects the count rate throughput of the
system As we noted earlier, setting either parameter too fast can
result in lost resolution Increasing the settings improve resolution,
but lengthen processing time and sacrifice throughput A tradeoff
exists (as it did in analog systems) between count rate throughput
and resolution Higher throughput can be attained with a loss of
resolution and better resolution can be attained at a loss of
through-put – up to the limits imposed by the performance of the detector
and preamplifier components.
These tradeoffs also existed in traditional analog systems, but the
tradeoffs can now be made at a higher level – the DSP provides
both improved throughput and improved resolution as compared to
analog This is due to a number of factors First, the trapezoidal
algorithm is simply more efficient and can process the signals more
accurately and rapidly than analog electronics.
Secondly, the user has much more flexibility to vary the components
of the processing time In analog systems, the processing was
con-trolled by a single parameter – the shaping time Now with DSP, two
parameters are available – one to accommodate noise level and
one to accommodate detector pulse rise time By adjusting these
two separately, optimum settings can more readily be attained
re-sulting, generally, in shorter total processing time to reach the same
resolution result Additionally, the analog amplifiers typically were
limited to six or fewer shaping time selections If, say, 2 µs shaping
was too short, the next available selection was usually 4 µs – twice
the processing time With the CANBERRA DSP products, the user
can typically select from 35 to 40 rise times and 21 flat top times
Again, this greater granularity of adjustment makes it possible to
more closely optimize the performance.
Note that the CANBERRA DSP products also implement Pile Up
Rejection/Live Time Correction (PUR/LTC) Earlier products
imple-mented this feature with analog circuitry, but in the DSP this is
in-corporated into the digital domain functions Pulse pileup occurs
when a new pulse from the preamplifier reaches the input stages
of the DSP before the previous pulse is fully processed In such
cases, the PUR/LTC function a) inhibits the processing of any
inval-id, composite pulses and b) turns off the live time clock during the
time pulse processing is gated off In this manner, piled up events
– which would serve only to distort the spectrum – are rejected
before storage by the MCA and the actual live counting time of the
MCA remains correct.
The improved performance of the DSP as compared to analog
systems is shown in Figures 1.29 to 1.34 Figures 1.29 and 1.30 show
real performance data collected with a DSP and an analog gated
integrator and fast ADC (the fastest available using analog
technol-ogy) For this experiment, a Model 2060 DSP was set for rise time
of 0.72 µs and flat top time of 0.68 µs The analog gated integrator
amplifier (Model 2024) was set for shaping of 0.25 µs and paired with a
800 ns Fixed Dead Time ADC These settings were chosen for
op-timal throughput with a relatively small (≈11% efficient) germanium
detector.
Trang 18As we can see from Figure 1.29, the DSP based system provides
higher throughput by approximately 50% Figure 1.30 shows the
resolution comparison for the same experiment and demonstrates
that the DSP also provides significantly better resolution once the
input count rate exceeds approximately 150 kcps Note that the
shape of the resolution curve in Figure 1.30 is also much flatter,
indicating that widely varying count rates can be accommodated at
a relatively constant resolution.
Note that with these settings chosen for highest throughput, the
res-olution performance at lower count rates is actually slightly worse
with the DSP However, in an application involving those count rates,
it is unlikely those settings would be used Figures 1.31 and 1.32
show the same analog data compared to the DSP system with the
rise time extended to 1.24 µs This reduces the throughput of the
DSP system although it is still superior to that of the analog Further,
we see now that with these settings, the resolution of the DSP is
su-perior to the analog across the full range of incoming count rates.
Figures 1.33 and 1.34 compare a Model 2060 DSP to a
Gauss-ian analog system consisting of a Model 2025 amplifier and Model
8715 ADC In this case, the settings of both systems were chosen
to provide optimal resolution under the high incoming count rates
Analog systems were set for 2 µs and 4 µs Gaussian shaping times
while the DSP settings were 5.6 µs rise time and 0.8 µs flat top
Fig-ure 1.33 shows that, with these settings, the throughput of the DSP
system is approximately equal to that of the 2 µs Gaussian system
Yet Figure 1.34 shows the resolution of the DSP system is superior
to the 4 µs Gaussian system Again, the DSP allows the
spectros-copist to achieve a significantly better tradeoff between throughput
and resolution.
LOSS FREE COUNTING APPLICATIONS
The correction of the Live Time Clock as described above,
effec-tively extending the counting time to account for those periods when
the system could not accept an input, is adequate for most samples,
in particular those for which the count rate is relatively constant
However, for short half-lived samples, or samples whose
constitu-ents change (as in a flow monitoring application), this method will
not be accurate In addition, even if the “counts per unit time” are
accurate using the traditional method for dead time correction, the
“real” counting time will have been extended by an amount equal to
the dead time, which may in fact increase the actual collection time
to an undesirable length.
The principal goal of Loss Free Counting (LFC) is to insure that at
the end of any data acquisition interval, the electronics have
accu-mulated all of the events that occurred regardless of any dead time
that may have been present in the system LFC is based on the
concept that by adding “n” counts per event to an MCA’s channel
register, rather than digitizing and storing a single count at a time,
a “zero dead time” energy spectrum can be accumulated that
as-sures all counts are included in the spectrum Assuming that “n” is
correctly derived, (“n” should equal “1” plus a “weighting factor”
rep-resenting the number of events that were lost since the last event
was stored), and the data is truly random in nature, the concept is
statistically valid The factor “n” is derived on a continuous basis by
examining the state of the Amplifier and ADC every 200 ns The
pro-portion of time during which the Amplifier and ADC are processing a
pulse provides a measure for the magnitude of the weighting factor
“n”, which is updated every 20 µs Loss free counting requires that
the MCA support “add-n” or multiple “add-one” data transfer; consult
the factory for details.
Figure 1.32 A comparison of the resolution between an analog system optimized for maximum throughput and a DSP system
set for a similar throughput
Figure 1.33 A comparison of the system throughput as
a function of input count rate for a DSP and two analog systems optimized for resolution
Figure 1.34 A comparison of the system resolution as
a function of input count rate for a DSP and two analog systems optimized for resolution
Trang 19Unfortunately, counting statistics in a Loss Free Counting system
cannot be calculated from the corrected spectrum One basic
as-sumption used by all peak fitting algorithms is that of Poisson
count-ing statistics That is, the uncertainty of the counts is proportional
to the square root of the number of counts While this assumption
is true for traditional “add-1” front-ends, it is not true of the “add-n”
Loss Free Counting front-end This assumption is especially poor
as the weighting factor becomes large To properly quantify the
un-certainty in each channel’s contents, the peak fitting program must
have access to both the corrected “add-n” and the uncorrected
“add-1” spectra Therefore, CANBERRA also offers a “Dual-LFC”
hardware option for the Model 599 Loss Free Counting Module
which allows the collection of both of these spectra so that
statisti-cally correct peak filling can occur Note that the correction software
for the “Dual-LFC” system is only available for VMS-based Genie
Systems.
PIPS DETECTORS AND MULTICHANNEL ANALYZERS
Alpha spectroscopy measurements of low-level samples require
long counting times A large area PIPS detector, when configured
Figure 1.35 Example Large Scale Alpha Spectroscopy System based on the Alpha Analyst
with a CANBERRA alpha spectrometer and multichannel analyzer, provides a high resolution, low background, counting system that will satisfy a multitude of alpha spectroscopy applications.
An example of a single chamber alpha spectroscopy system (that can easily be upgraded) is illustrated in Figure 1.11 Note that the Model 7401 Alpha Spectrometer is a complete, self-contained, dou- ble-width NIM module that contains a vacuum chamber, vacuum gage, detector bias supply, preamplifier/amplifier, SCA, counter/tim-
er and pulser for setup and test Multiple Model 7401 Alpha trometers can be configured with a vacuum system that allows indi- vidual vacuum chambers to be opened and loaded without affecting the vacuum or data acquisition of the other spectrometers.
Spec-However, where numerous samples are counted simultaneously, it
is more cost effective and user efficient to select a system based on the Alpha Analyst (Figure 1.35) This turn-key system supports mul- tiple detectors in a comprehensive software environment featuring full computer control of all vacuum elements and acquisition elec- tronics To learn more about CANBERRA’s Alpha Analyst, please refer to Section 1 of this catalog.
Trang 20Timing and Coincidence Counting Systems
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
COINCIDENCE TECHNIQUES
There are many applications that require the measurement of events
that occur in two separate detectors within a given time interval, or
the measurement of the time delay between the two events These
two approaches are used in gamma-gamma or particle-gamma
co-incidence measurements, positron lifetime studies, decay scheme
studies and similar applications, and are titled coincidence or timing
measurements.
A coincidence system determines when two events occur within
a certain fixed time period However, in practice it’s not possible
to analyze coincidence events with 100% confidence due to the
uncertainties associated with the statistical nature of the process
Statistical timing errors may occur from the detection process and
uncertainties in the electronics resulting from timing jitter, amplitude
walk and noise, which lead to statistically variable time delays
be-tween processed events A simple coincidence circuit solves this
problem by essentially summing the two input pulses, passing the
resultant sum pulse through a discriminator level, and generating
an output pulse when the two input pulses overlap Figure 1.36
illustrates this process Note that the period of time in which the two
input pulses can be accepted is defined as the resolving time, which
is determined by the width of the pulses, τ, such that the resolving
time is equal to 2τ.
The 2040 Coincidence Analyzer uses a more sophisticated scheme
allowing analysis of several input signals It produces a logic pulse
output when the input pulses, on the active inputs, occur within the
resolving time window selected by the front panel control.
Since detector events occur at random times, accidental
coinci-dences can occur between two pulses which produce background
in the coincidence counting The rate of accidental or random
coin-cidences is given by:
N acc =N 1 N 2 (2τ)
Where:
N 1 = Count rate in detector number 1
N 2 = Count rate in detector number 2 2τ = The resolving time of the coincidence circuit The number of counts in the detectors depends upon the experiment and the detectors, so the best way to reduce accidental coincidences
is to make the resolving time as small as possible However, the resolving time cannot be reduced below the amount of time jitter in the detector pulses without losing true coincidences, so the type of detector determines the minimum resolving time usable.
A coincidence setup with NaI detectors is shown in Figure 1.37 The unipolar pulse from the 2022 Amplifier is processed by a Model 2037A Constant Fraction Timing SCA to produce a standard NIM logic pulse for the 2040 Coincidence unit The 814FP Pulser is used
to set up delays and test operation.
Figure 1.36 Coincidence Pulses
Figure 1.37 Coincidence Electronics
Trang 21In order to properly operate the system, a delay curve is obtained
in which coincidences are measured as a function of relative delay
between the two detectors In the ideal case of no time jitter in either
detector, the solid curve in Figure 1.38 is obtained However, real
detectors will produce the dashed curve, and the minimum
resolv-ing time settresolv-ing is where there is a flat region (indicatresolv-ing all true
coincidences are collected) Thus, the proper relative delay is the
value for the center of the flat region.
Typical resolving times are 10 nanoseconds or better, for an energy
range of 0.1 to 1 MeV Shorter resolving times are possible for
plas-tic scintillators and silicon charged parplas-ticle detectors, even down to
less than 1 nanosecond In general, the shorter the rise time of the
preamplifier pulse, the smaller the resolving time This is discussed
further with fast discriminators.
Another coincidence technique involves the direct measurement of
the time delay between two pulses A time-to-amplitude converter
(TAC) converts a time difference between two input pulses to a 0 to
10 volt pulse This analog pulse can then be used as an input to an
SCA, or ADC and MCA The Model 2145 TAC provides an integral
SCA capability as well as an output pulse, gate delays and other
features.
Figure 1.39 Coincidence Electronics with TAC Unit
The use of a 2145 TAC with NaI detectors in exact analogy with the
2040 Coincidence unit is shown in Figure 1.39 There are several advantages to the use of a TAC First, no delay curve needs to be taken since all relative decays occurring are recorded, and second,
no resolving time setting is involved.
The natural time spectrum of the two detectors can be stored directly in an MCA, and if a window is set very narrow, as in Figure 1.40, then there are a minimum number of accidental coincidences recorded The 2058 Nanosecond Delay is required to delay one de- tector pulse from the other so that a measurable time difference occurs.
TIMING DISCRIMINATORS
A crucial part of any coincidence system is the timing tor used to determine when a pulse occurs There are two general categories:
discrimina-• Slow (or energy) Timing
• Fast Timing The timing single channel analyzers used in Figures 1.37 and 1.39 are examples of slow timing The single channel analyzer operates with shaped pulses to select the range of energies involved in the coincidence, and produces an output logic pulse that is, ideally, in- dependent of input pulse amplitude Fast timing uses pulses directly from the detector, without regard to specific energy.
Three basic techniques are used in both fast and slow modes for acquiring timing information:
• Leading Edge
• Crossover
• Constant Fraction CANBERRA provides electronic modules for performing any of these techniques, and the proper choice depends upon the detec- tor and application, as discussed below.
Figure 1.38 Resolving Time Curve
Trang 22The most fundamental timing circuit generates a logic signal when
the leading edge of an input pulse crosses through a discriminator
level as shown in Figure 1.41.
The main problem is that the time of the output pulse varies
mark-edly with amplitude, as can be seen by comparing the two signals
shown This effect can be reduced by setting the discriminator at a
very low level, such as just above noise The Model 2037A Timing
SCA sets the discriminator up to a maximum of 200 millivolts.
Crossover timing relies on the fact that the zero-crossing point in a
bipolar pulse is very nearly independent of pulse amplitude (See
Figure 1.42) The Model 2037A Timing Single Channel Analyzer
offers a crossover mode of operation for bipolar input pulses The
crossover technique has some limitations in that there is still time
dependence or “walk” for different amplitudes, and that signals with
varying rise times from the same detector (such as occurs with
germanium detectors), will produce walk.
Figure 1.40 TAC Spectrum
Figure 1.41 Leading Edge Timing
The constant fraction technique will eliminate most of the
short-comings of the two former methods of timing The constant fraction
timing technique is similar to a discriminator, but with a threshold
that is a constant fraction of the signal amplitude A discussion of
the constant fraction technique, as implemented in the Model 2126
Constant Fraction Discriminator, is given in the Timing Section.
This module can be connected to NaI or fast plastic detectors with negative amplitude output signals ranging from –5 mV to –5 V, and rise times down to 1 nanosecond The 2126 performs no signal pro- cessing on the input and is very often attached directly to the anode
of a photomultiplier tube As mentioned above, constant fraction discrimination is a method that offers a timing output signal when
a constant ratio of the pulse height is reached This ratio, once set,
is consistent from pulse to pulse, thus removing the amplitude and rise time errors that arise The main problem with this method is that it is still sensitive to pulse shape distortion, exhibiting poor time resolution for energies less than 200 keV and for poorly shaped or noisy pulses.
Whenever noise or low amplitude is a significant characteristic of a detector, a filter network is required between the signal source and amplifier to alleviate the noise distortion The Model 2111 Timing Fil- ter Amplifier has a built-in filter that attenuates the noise component before amplification of a low signal When used with a constant frac- tion discriminator such as the 2126, a stable time reference can be derived The 2111/2126 combination has widespread application in gamma-gamma coincidence and lifetime studies, offering excellent time resolution, which in connection with the high resolution of large germanium detectors, increases the rate of useful data collection.
The 2111 Timing Filter Amplifier is applicable for both surface rier and germanium timing applications Both of these detectors produce signals of low amplitude, distorted with noise, and in the case of germanium detectors, poorly shaped rise times A gamma- gamma coincidence system with NaI and germanium detectors is shown in Figure 1.43 This is an example of a “fast-slow” coincidence system in which Model 2111 and 2126 constant fraction discrimina- tors are used to indicate the presence of a pulse, and an energy range on the NaI detector is selected The energy spectrum of the germanium detector is stored in the MCA if the ADC gate is opened
bar-by a coincidence pulse representing a combination of proper germanium timing and the selected NaI energy A Model 2145 TAC
NaI-is used to set the true coincidence range because of the set-up convenience offered, as described above The TAC SCA output and the Timing SCA output of a Model 2015A Amplifier/Timing SCA are placed in coincidence with Model 2040 Coincidence unit If desired,
an energy requirement could be placed on the germanium tor by adding a Model 2037A Timing SCA on the 2025 amplifier’s bipolar output, and feeding the 2037A’s output to the Reset/Inhibit input on the 2145 TAC.
detec-Anticoincidence systems are occasionally required, as mentioned above for Compton suppression in germanium detectors (see Figure 1.24) or cosmic ray suppression in alpha/beta counting.
Figure 1.42 Bipolar Pulses and Crossover Timing
Trang 23Figure 1.43 NaI-Ge Fast-Slow Coincidence Electronics
Trang 24Spectrum Analysis
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
CANBERRA offers a variety of nuclear systems which perform
data analysis as well as data acquisition These systems range
from small stand alone systems to more sophisticated
configura-tions involving a variety of computer platforms Typical applicaconfigura-tions
include Environmental Monitoring, Body Burden Analysis, Nuclear
Waste Assay, Safeguards and other applications Details of these
systems are provided later in this catalog, or in various brochures
that are available from CANBERRA The following section presents
some of the typical procedures and calculations involved in nuclear
applications.
COUNTING STATISTICS
Radioactive decay occurs randomly in time, so the measurement of
the number of events detected in a given time period is never
ex-act, but represents an average value with some uncertainty Better
average values can be obtained by acquiring data over longer time
periods But, since this is not always possible, it is necessary to be
able to estimate the accuracy of any given average.
Nuclear events follow a Poisson distribution which is the limiting
case of a binomial distribution for an infinite number of time
inter-vals, and closely resembles a Gaussian distribution when the
num-ber of observed events is large The Poisson distribution for
observ-ing N events when the average is N, is given by:
PN =
and has standard deviation σ (sigma) equal to √N A graph of P N for N equal to 3 and to 10 is shown in Figure 1.44 The curves are asymmetric and have the property that N is not exactly the most probable value but is close to it However, as N increases the curve becomes more symmetric, and approaches the Gaussian distribu- tion:
PN = • e –x 2 / 2N = • e –x / 2σ 2
Where: x = N – N The integral of the area under the Gaussian curve is often used to report errors in terms of a confidence level in percent For example,
in the value reported as 64 ± 8, 8 is equal to σ and represents 68%
of the area under the appropriate Gaussian curve for N=64 It may
be stated as the value one is 68% confident of obtaining if the surement is repeated Traditionally, many of CANBERRA’s MCAs have used 1.65 σ, which corresponds to a 90% confidence level Probable error is often used, which corresponds to a 50% confi- dence level These can be user-set to other values, such as:
0.68 1.0 1.15 1.65 1.96
50 68.3 75 90 95
Figure 1.44 Poisson Distributions for N=3 and 10
NNe –N
N!
1 √2πN
1 √2πα
Trang 25Since the uncertainty depends upon the square root of the counts,
improvements in accuracy by counting longer, or by using a more
efficient detector, only increase as the square root For example, if
564 counts are obtained in an hour for σ ≈ √564 ≈ 24 for a 24/564
= 4.3% accuracy, counting for two hours to get 1133 counts with σ
≈ 34 only gives an improvement to 3.0% In other words, counting
twice as long gives an improvement of √2 = 1.4, or 40%.
Examples of data in which counting statistics apply include: the
counts in a counter, the counts in a single channel of an MCA
trum, or the sum of counts in a group of channels in an MCA
spec-trum The situation becomes even more complicated when
subtract-ing a background as shown in the followsubtract-ing separate, but frequent,
cases.
• Subtracting background counts, as in one counter’s value
from another, or for each channel (when subtracting one
spectrum from another).
• Subtracting a straight line background from a peak on top of
the background in a spectrum, such as a HPGe peak on top
of Compton pulses from higher energy gamma rays.
The error in adding or subtracting two Poisson distributed numbers
with errors, as in:
Ntotal = (N1 ± √N 1 ) ± (N2 ± √N 2 )
is given by:
σ N total = √(√N 1 ) 2 + (√N 2 ) 2
Consider a low level counting situation in which 56 counts are
ob-tained in 10 minutes, and a background of 38 counts in 10 minutes
was measured without the sample The result is 56–38 = 18 counts,
with an error of √56 + 38 = √94 = or approximately 9.7, a σ value
of 54%.
A better procedure is to measure the background over a longer
period of time to obtain a small percentage error and factor to the
appropriate time for each sample analyzed Using the same
exam-ple as above, but with a 100 minute background of 380 counts, the
result would be 56–(380/10) = 18 counts, with an error of
√ 56 + ([ 10 / 100 ] 2 x 380) = √56 + 38 = √59.8
or approximately 7.7, a σ value of 43%.
NET AREA CALCULATION
For the case in which a peak lies on a background that cannot be
subtracted by a background spectrum, such as shown in Figure
1.45 for an MCA spectrum from a HPGe detector:
The area above the background represents the total counts between
the vertical lines minus the trapezoidal area below the horizontal
line If the total counts are P and the end-points of the horizontal line
are B 1 and B 2 , then the net area is given by:
A = P – (B1 + B2)
Where: n = The number of channels between B 1 and B 2
It is tempting to calculate the uncertainty by just using the formula
for subtracting two numbers, with standard deviations of:
σ N = √P + (B1 + B2)
However, this is incorrect because the trapezoidal area is not son distributed and its error is not just the square root of the counts, but depends upon how the errors in B 1 and B 2 affect the horizontal line across the entire region The proper procedure, which is imple- mented in CANBERRA MCAs and in analysis of peak areas in vari- ous HPGe software packages, is derived as follows:
Pois-The standard deviation in a function A is given by:
A = f (N1 N2 Nn) where Nn is the counts in channel N.
Figure 1.45 Net Area Determination
n 2
n 2
n 2
= (P1 + Pn) +
2 (√B 1 )2 +
2 (√B 2 )21/2
= P +
2 (B 1 + B 2 )
A = P – +
and the standard deviation is:
σ(A) = P +
2 +
1/2
B2
n2
Trang 26Most CANBERRA MCAs and analysis software packages perform
end-point averaging with a user-selectable number of end-points.
There are many ways of calculating the net counts under a peak
The method described above is a valid, common method, provided
that there are no interferences from photopeaks adjacent to the
peak of interest, and assuming that the background continuum
var-ies linearly from one side of the peak to the other.
However, if interferences exist, other methods of calculating the
net area of a peak must be employed which can include, (but are
not limited to), the use of parabolic or step background algorithms,
as well as non-linear fitting algorithms, etc For further discussions
concerning these techniques and others, the reader is referred to
more detailed texts and formal spectroscopy training courses.
ENERGY CALIBRATION
Many nuclear applications require a means for determining the
en-ergy at a particular channel location of a spectrum To meet this
need, CANBERRA has implemented various techniques which are
briefly discussed.
In some MCAs, a simple two-point energy calibration is used to
determine both the offset and slope by the equation:
E = A (ch) + B
Where: ch = channel number
Thus, the energy vs channel number can be directly read out
How-ever, the more advanced MCA Systems, such as those based on
Genie 2000 or Genie-ESP, allow users to choose between
first-order (i.e linear) or second-first-order (i.e quadratic) equations that use
a least squares fit to multiple data points.
Most preamplifier, amplifier and ADC systems are very linear and
first-order energy calibration can properly describe the data For
ex-ample, a CANBERRA germanium detector with 2002 Preamplifier,
2025 Amplifier and 8701 ADC has nonlinearities less than 0.05%
for the preamplifier and amplifier, and 0.025% for the ADC The
combined nonlinearity is then:
± √(0.05%) 2 + (0.05%) 2 + (0.025%) 2 = ±0.075%
This is still a very small number, but for a spectrum of 4000
chan-nels, the low and high energy channels may be correct and leave
a 0.00075 x 2000 = 1.5 channel uncertainty at channel 2000 A
second-order term in the energy calibration can remove this in order
to provide very accurate energy-channel calibration over the entire
range, according to the equation:
E = A(ch) 2 + B(ch) + C
Where: ch = channel number
A further refinement is provided by using least-squares techniques
to determine the equation that best fits the data, when more than
the minimum number of points is available, (2 for first-order, 3 for
second-order) The Genie 2000 and Genie-ESP MCA Systems use
this technique.
NUCLIDE IDENTIFICATION AND QUANTITATIVE ANALYSIS
Many applications with high purity germanium (HPGe) detector
spectra involve identifying particular gamma rays with specific
nu-clides The sharp peaks in the HPGe spectra, coupled with a
care-ful precise energy calibration, can be used for generally unique
determinations of the nuclides in a sample If an automatic peak
search capability is provided then a complete sample analysis can
be accomplished without operator intervention, which is ideal for analyzing large numbers of samples All CANBERRA HPGe/com- puter-based gamma spectroscopy systems provide nuclide identifi- cation through peak searches of spectra and scans of standard and user-generated nuclide libraries A sample printout of a Genie 2000 nuclide identification report is shown in Figure 1.46.
In the Genie software platforms, the peak search locates peak centroids and then enters a region of interest about each peak This is especially useful for observing the quality of data obtained CANBERRA analysis software provides the additional capability of resolving overlapping peaks into individual components.
Figure 1.46 Isotope ID
Figure 1.47 88 Y Decay Scheme
Trang 27The final step in nuclide analysis is to determine the intensity of
the radioactivity corresponding to each isotope The net area of the
peak is directly related to the intensity, but it is also necessary to
correct for the efficiency of the detector, the branching ratio of the
source, and the half life (if it is desired to relate the activity to an
ear-lier or later time) The efficiency was discussed earear-lier and has an
energy dependence such as shown in Figure 1.1 The branching
ra-tio (or yield) is used to correct the number of gamma rays observed
to obtain the number of disintegrations of the source Figure 1.47
shows the decay scheme for 88 Y and the percent of disintegrations
leading to the various gamma rays.
The activity of a particular isotope is given in microcuries as:
A(µCi) =
Where yield is the branching ratio fraction and live time is the actual
ADC live time in seconds Half-life corrections are made by
multiply-ing the activity by an exponential factor.
A(at time t o ) = Ae
Where decay time and half-life must be in the same units (seconds,
minutes, hours, or years).
Further specific data analysis is highly dependent upon the
appli-cation, detector and electronics configuration, and is beyond the
scope of this brief presentation.
EFFICIENCY CALIBRATION
In the equation for activity cited above, the value for efficiency is
dependent on the geometry of the sample – size, density, and
dis-tance from detector For the detectors used in gamma analysis,
ef-ficiency varies significantly with energy (see Figure 1.1) Therefore,
each counting geometry requires an efficiency calibration, using a
known standard in the same geometry which includes multiple
en-ergies A series of data pairs of efficiency vs energy are generated
from the relationship:
Efficiency =
Where Activity is the strength (in Bq) of the standard source (at
col-lection time) at the given energy, yield is the branching ratio fraction
and live time is the actual ADC live time in seconds.
In the Genie software system, the calibration data from the standard
are entered into a “Certificate File”, with these data being used for
subsequent efficiency calibrations The software will automatically
correct for source decay by the formula:
A(at count time) = A(at certificate time)e –
Where decay time and half-life are in the same units (seconds,
min-utes, hours, or years).
In(2)xDecay Time Half Life
Net Area (Live Time)(Activity)(Yield)
In(2)xDecay Time Half Life
CANBERRA has recently introduced Mathematical Efficiency Calibration products (S573, S574) that do not require radioactive sources for efficiency calibrations These new products (ISOCS™, LabSOCS™) rely on fundamental physical measurements and nu- clear constants to accurately determine the energy-efficiency pairs From the series of data pairs, a curve of efficiency versus energy is generated, with the user having a choice of fitting paradigms Thus, the software can calculate efficiency at any energy in the calibrated energy range when analyzing an unknown spectrum.
MINIMUM DETECTABLE ACTIVITY
The calculation of Minimum Detectable Activity for a given nuclide,
at the 95% confidence level, is usually based on Currie’s
deriva-tion (Currie, L.A (1968) Anal Chem 40:586.), with one simplified
T over the energy range of interest
T is the collect time (sec) EFF is the efficiency at the energy of interest
Y is the Branching Ratio
wt is sample weight This formulation takes into account both kinds of errors – false positive and false negative, and yields the smallest level of activity which can be detected with 95% confidence, while also having 95% confidence that “activity” is not detected falsely from a null sample When the measurement is made on a ‘blank’, with no activity, but with the same form and density as an actual sample, the calculated MDA
is an a priori estimate of the best sensitivity that can be expected
from sample measurements When the calculation is applied to a spectrum collected from an actual sample, the background at the energy of interest will in most cases be higher, due to interference and Compton scattering from other nuclides in the sample Thus, the
MDA for an actual sample, computed a posteriori, will be somewhat higher than the a priori estimate.
The MDA – also referred to as Lower Limit of Detection (LLD) – can
be improved by increasing the efficiency of detection, decreasing the background, or, for a given experimental setup, by increasing the collect time or the sample size It is frequently necessary to select the appropriate collect time to ensure that the measured MDA will be below the action level mandated by the count-room procedures.
The above formula for MDA, generally accepted in the United States and many other countries, is implemented in a more complete form
in CANBERRA Analytical software Some CANBERRA software packages, such as Genie 2000, offer the user a choice of additional formulas required in other countries.
2.71 + 4.66(σ)
T • EFF • Y • wt Net Area
(Live Time)(Efficiency)(Yield)(3.7x10 4 )
Trang 28Neutron Detection and Counting
Phone contact information
Benelux/Denmark (32) 2 481 85 30 • Canada 905-660-5373 • Central Europe +43 (0)2230 37000 • France (33) 1 39 48 52 00 • Germany (49) 6142 73820 • Japan 81-3-5844-2681 •
Russia (7-495) 429-6577 • United Kingdom (44) 1235 838333 • United States (1) 203-238-2351
SOURCES OF NEUTRONS
There are several methods by which neutrons may be produced in
the fuel cycle principle amongst these are:
1 Alpha Particle Induced Reactions
Plutonium and uranium isotopes decay by alpha particle
emission The alpha particle is absorbed by the nuclei of the
low atomic number elements (Li, B, Be, O, F, C, Si, etc.) and
a neutron is produced The yield depends upon the chemical
composition of the matrix and the alpha production rate for
plutonium and uranium Neutrons from (α,n) reactions are
produced randomly (not time-correlated) and they exhibit a
broad energy spectrum Other α-emitting nuclides can also
make important contributions, for example 241 Am.
2 Spontaneous Fission
The even-numbered isotopes of plutonium ( 238 Pu, 240 Pu,
and 242 Pu) spontaneously fission (SF) at a rate of 1100,
471, and 800 SF/gram-second respectively Like (α,n)
trons, SF neutrons have a broad energy spectrum SF
neu-trons are time-correlated (several neuneu-trons are produced at
the same time), with the average number of neutrons per
fission being between 2.16 and 2.26 Uranium isotopes and
odd-numbered plutonium isotopes spontaneously fission
at a much lower rate (0.0003 to 0.006 SF/gram-second) In
spent fuel Cm and Cf isotopes may be significant.
3 Induced Fission
Fissions can be induced in 239 Pu, 235 U, and 238 U by neutron
interrogation of the sample with an external neutron source
Like SF neutrons, they have a broad energy spectrum and
are time-correlated.
NEUTRON DETECTION
Neutrons have mass but no electrical charge Because of this they
cannot directly produce ionization in a detector, and therefore
can-not be directly detected This means that neutron detectors must rely
upon a conversion process where an incident neutron interacts with
a nucleus to produce a secondary charged particle These charged
particles are then directly detected and from them the presence of
neutrons is deduced.
The most common reaction used for high efficiency thermal neutron
detection today is:
n + 3 He → p + 3 H + 765 keV
where both the proton and the triton are detected by a gas filled
proportional counters using 3 He fill gas Quench gas is also added
to control the ionization process.
Another common method uses BF 3 filled detectors that utilize the
fission of the 10 B atom to provide the charged particle.
n + B →
A typical instrument configuration that can be used with either of
these detectors can be seen in Figure 1.48.
Gas filled proportional counters offer low noise amplification of the ionization event producing a charge pulse processed by the attached nucleonics chain They offer high detection efficiency with excellent gamma discrimination They also provide a cost effec- tive and stable means of constructing detectors for a wide range of applications.
Gas proportional detectors such as these are efficient only for mal (low energy) neutrons; for high energy neutrons their capture cross sections are very small, making it very unlikely that a neutron will interact with the fill gas and cause the necessary detection reac- tion Because of this it is necessary to slow the neutrons down to increase the probability of interaction.
ther-This is usually done by surrounding the detector and/or the sample being counted by a hydrogen-rich material (moderator) such as high density polyethylene Typically 10 cm (4 in.) of polyethylene surround the detector with a few cm being between the proportional counters and the neutron source.
The pulse height spectrum from the interaction of a thermal neutron
in a typical 3 He neutron detector will look as shown in Figure 1.49 Perhaps the most important point to note is there is no information about the primary neutron spectrum Because all of the neutrons which are detected have been moderated to reduce their energy
to the thermal level, all neutron energy information is lost All of the events of interest fall into one peak, which is the reaction energy (765 keV) Once a discriminator has been set to eliminate the gam-
ma interference and those events produced by interactions with the walls of the detector tube, simple gross counting is all that is required.
NEUTRON COINCIDENCE COUNTING
One of the more common applications of neutron detection and counting is the assay of fissile materials via the SF signature of the fertile nuclides When the fissile material is contained in a high den- sity matrix or a matrix which includes fission products, the interfer- ing gammas from those fission products may very well overwhelm the weak gammas emitted from the fissile material, making neutron counting the only viable method available for performing the assay
In general neutron and gamma techniques are complementary HRGS may provide relative isotopic information, for example, and the neutron assay bulk non destructive quantification.
Figure 1.48 Neutron Counter Electronics
Trang 29sample matrix For many materials of interest, the vast majority of
the neutrons that are detected will be from the (α,n) reaction and not
from the spontaneous fission of the fissile material.
It is not possible to use energy discrimination to distinguish
neu-trons from different sources, therefore, traditional spectroscopy
techniques cannot be used However, there is a characteristic time
distribution difference between (α,n) neutrons and those neutrons
produced as the result of a fission event that can be exploited
Specifically, fission events will produce multiple neutrons –
usu-ally two and sometimes three – simultaneously; (α,n) neutrons,
on the other hand, are produced individually and randomly This
allows coincidence counting techniques to be used to distinguish
the prompt fission neutrons from the random (α,n) neutrons.
A neutron coincidence system is shown schematically in Figure
1.50 The outputs from the 3 He tubes are first processed by fast
pre-amplifier/amplifier/discriminator (PAD) circuits, and logically ORed
to the input of the neutron coincidence analyzer for processing.
The coincidence logic identifies those neutron counts that occur within
a short time of each other: fission neutrons, closely spaced (α,n)
neu-trons, and counts due to “accidental coincidences” Once one neutron
has been detected, the probability of detecting another neutron from
the same fission decreases approximately exponentially with time
according to the following equation:
P(t) = exp (–t/t d )
where
P(t) = Probability of detecting coincidence neutrons in time t
t d = die-away time of the moderated detector assembly
The die-away time is the characteristic time a neutron will
survive before it is absorbed in the 3 He tubes or escapes the
counter The neutron die-away time normally ranges from 10 to
128 µs depending upon the counter geometry.
Figure 1.49 Thermal Neutron Induced Pulse Height Spectrum
from a Moderated 3 He Detector
The probability for detecting random neutrons from an item is stant with time To determine whether the neutron events which are detected are time-correlated, two equal time periods are sampled
con-by the coincidence logic for each neutron that is detected The first gate or counting window is opened for a time period equal to about 1.267 t d after a neutron is detected Other counts within this time period are due to multiple fission neutrons from the triggering event, other fissions, and (α,n) reactions After a delay of approximately
4000 µs, the second gate is opened and random neutron events are counted The 4000 µs delay assures no time correlation with the neutron which triggered the count The difference in the two time- gated counts (Reals+Accidentals and Accidentals, respectively) is the net Real coincidence count, or Reals The net Reals count is related to the fissile material in the sample by a calibration con- stant Modern CANBERRA neutron analysis instruments are based
on shift register counting The shift register preserves the pulse quence allowing coincidence data to be evaluated for each neutron event.
se-MULTIPLICITY COUNTING
Neutron coincidence counting provides two measured values (Reals and Totals) while in some cases there are three (or more) unknown variables which need to be determined: mass of
240 Pu-effective, (α,n)-to-(SF,n) ratio, and multiplication factor This condition arises, for example, when impurities in the material under analysis preclude the estimation of alpha ratio from the isotopics
In Multiplicity Counting, a third measured parameter – the tion of multiple counts – is derived, and thus the three unknowns may be calculated A special version of the shift register neutron coincidence analyzer performs the necessary data collection and software reduces the histogram of multiplicity events recorded to singles (Totals or Gross), Doubles (or Reals) and Triples event rates These are used in conjunction with interpretational models to ex- tract the unknown variables for product material m eff , α and M L are normally extracted as discussed already for low level waste where
distribu-ML ~1 then m eff , α and ε, the detection efficiency, can be extracted Alternatively, special algorithms can be invoked to reduced de- tection limits and low level bias caused by cosmic-ray induced spallation events in the waste.
Figure 1.50 Schematic Arrangement of a Thermal Well Coincidence Counter
Trang 30System Selection Guide
Gamma
Spectroscopy Multi-Station/Multi-Input DSA-2000AIM/ICB NIM Genie-ESP orGenie 2000 Section 1
DSA-1000, Multiport II Genie 2000 Standalone/Single Input DSA-2000
DSA-1000 Multiport II Eagle Plus
Genie 2000 Section 1
Sample Changer Based Gamma Analyst Genie-ESP
or Apex (Gamma)
Section 1
(HPGe) Portable/Battery Powered InSpector 2000 Genie 2000 Section 1(NaI) required
Portable/No AC Power uniSpecInSpector 1000 Genie 2000Genie 2000 optional Section 1Alpha
Spectroscopy Multi-Station/Multi-Input or
Single Station/Multi-Input
Alpha Analyst Genie-ESP/Apex-Alpha
Client/Server or
Apex-Alpha Desktop
Section 1
Single Station/Limited Number of Inputs Alpha Analyst Apex-Alpha Desktop Section 1Alpha/Beta
Counters Short Count Times Computer Based: Series 5 XLB
Embedded: Series 5 E
iMatic
Eclipse SW LB-Link optional
iLink optional
Section 1
Transportable iSolo
Series 5 APC iLink optionalLB-Link optional Section 1Neutron
Counting Neutron Coincidence Counting JSR-12, JSR-14, 2150, Genie 2000/NDA-2000 Section 3
Alpha/Beta and Gamma
Distributed Alpha Sentry Sampling Heads ASM-1000
Gamma only G64 Area Gamma Monitor RADACS Section 2
PASSIVE NEUTRON COUNTERS
When a neutron coincidence counter is used for the assay of 238 Pu,
240 Pu, and 242 Pu, the neutrons from the spontaneous fission of
these isotopes are detected and counted Since no external neutron
source is required to induce fission, assay systems of this type are
known as Passive Neutron Counters.
ACTIVE NEUTRON COUNTERS
235 U, 238 U, and 239 Pu do not spontaneously fission at a high enough rate to allow passive assay techniques to be used For this reason, uranium assays utilize an external neutron source to induce fission
in the sample Assay systems using this technique are known as Active Neutron Counters Various kinds exist; e.g Active Well Co- incidence Counter which makes use of Am Li sources; 252 Cf shuf- fler; and the Differential Die Away techniques which uses a pulsed neutron generator CANBERRA offers all of these methods and also undertakes special commissions.
Trang 31Introduction
Standard Canberra cryostats and
shields are built with materials which
are screened for extraordinary levels
of natural and man-made radioactive
impurities By this means we are
assured that our standard systems
have relatively low background
levels and are suitable for routine use
in normal applications such as
Radiochemistry, Internal Dosimetry,
Activation Analysis, Waste Assay
and Environmental Counting
Ordinary construction and shielding
materials, however, do contain trace
amounts of naturally occurring and
man-made radionuclides which result
in prominent peaks and an elevated
background continuum, with a
resultant compromise in Minimum
Detectable Activity (MDA)
The design or configuration of
cryostat-shield systems is another
factor in system performance Some
cryostat and shield designs do not
adequately prevent streaming from
the outside environment nor do they
provide self-shielding from their own
relatively “hot” components Worst
of all, through an inappropriate choice
of material types and/or thicknesses
they may actually contribute to the
background or compromise system
performance and/or reliability
Canberra has many years of
experience in materials selection and
in the design and fabrication of low
level counting systems We have
found and developed reliable sources
for select materials and we have
invested in the laboratory facilities
necessary to screen materials and to
test complete systems This effort has
led to the development of standard
Ultra Low-Background Detector
Systems which are described herein
While these standard systems do not
entirely eliminate the need for customdesign systems, they provide predict-able high performance at modest costand can save users the immenseexpense and struggle associated withdesigning, specifying, and testing ofcustom systems
Apologies are due those who havepushed detector system backgrounds
to the utmost limits These systemsbetter deserve the “Ultra” superlativebut they are far beyond practicallimits for commercial applications
Cryostat Design
Five factors are of paramount concern
in the design of a low backgroundcryostat:
1) Background from materials
in close proximity todetector element must beminimized
2) There must be self-shieldingfrom “hotter” materials used
in construction
3) Streaming from the outsideenvironment must bereduced or eliminated.4) Materials having highcross sections for cosmicneutrons with attendantgamma emission should
be kept away from thedetector element
5) The design should notcompromise performanceand reliability
Canberra has two cryostat designswhich satisfy these five concernsadequately They are the Model7500SL-RDC and the 7915-30.Refer to Figure 1 and consider thefollowing explanation of the benefits
of these designs
A The detector chamber is simple
and of low mass The variety ofmaterials is kept to a minimum
so that background contributioncan be controlled
Application
NoteULTRA LOW-BACKGROUND DETECTOR SYSTEMS
Trang 32Figure 1
Cryostat designs suitable for low-background detector systems
B The preamplifier body which
contains fiberglass
printed-circuit boards, aluminum
hardware, and a wide variety
of components, is located
remotely and is shielded from
the detector by the full shield
thickness – not by some token
internal lead shield of
inadequate thickness and
compromising location
C Offset between the detector
element and the outside world
prevents streaming and reduced
background from the hotter
cryostat materials such as
molecular sieves or activated
charcoal This offset is common
to U-style cryostats like the
7915 Among dipstick cryostats,
only the Canberra 7500SL is
built with this offset Internal
lead shields are sometimes
proposed to overcome this
fault but they are usually of
inadequate thickness, they can
produce unwanted X rays in
close proximity to the detector
element, and they can
compro-mise reliability of the detector
system
Cryostat Materials
The difficulty in building low
background detectors is nowhere
more evident than in the choice of
materials offered by some detector
manufacturers for various detectortypes Canberra too, has struggled
to find materials which provide lowbackgrounds without compromisingnormal detector performance andwithout compromising long-termreliability Our choices, andcomments on the alternatives,are illustrated in Figure 2
content of less than one part perbillion This material is much moreexpensive than the high purity magne-sium which has been used for endcaps.Corrosion problems with magnesium,however, are endemic and use of thismaterial in thin cross-section forvacuum systems is pure folly Inaddition, magnesium with guaranteedlimits of primordial radionuclides isnot available
B Vacuum Chamber
The vacuum chamber is fabricatedfrom selected stainless steel or fromhigh purity copper
Post World War II steel is nated by 60Co which was widely used
contami-in blast furnace crucible lcontami-iners tomonitor wear or breakthrough Sincestainless steel is a heavily recycledmaterial there is virtually no such thing
as virgin stainless By batch testing,however, we can select stainless withdiminishingly small 60Co content.The alternative to stainless steel is highpurity copper Canberra uses copperwhich is 99.99% pure Standard OFHC(Oxygen-Free, High-Conductivity)copper, which is often mentioned as amaterial of choice for low-backgrounddetectors, can have up to five times theimpurity concentration of the Canberracopper
It is Canberra’s choice to use stainlesssteel for the detector chamber when-ever the material supply allows and
to substitute high purity copper onlywhen a supply problem exists Theadvantage of stainless steel is that nopassivating coatings are necessary toprevent corrosion
C Detector Holder
The detector holder has two tions: It must physically contain thedetector element in good thermalcontact with the cold finger and it mustprovide an infrared radiation shield forthe detector element to reduce IRgenerated leakage current and noise.For the latter function, the detectorholder or a separate IR shield mustsurround the detector element
func-High purity copper is a good choice of
Trang 33holder material for Coaxial, XtRa, and
Well detectors because these detectors
have a substantial Ge dead layer on
the cylindrical surface which limits
low-energy response from the side
These Canberra detectors generally
have a dead layer of less than 0.5 mm
which stops 60% of 50 keV
photons Much thicker dead layers
have been found on detectors made by
other manufacturers
The additional attenuation by the
copper holder, while not insignificant
(75% at 50 keV) is not prohibitive
in the energy range where these
detectors are most often used
Canberra provides an appropriately
thin IR shield at the normal entrance
window in any case
For reverse-electrode (REGe)
detectors, however, copper is not the
best choice of holder material These
detectors are most often purchased
because of their thin dead layer for
low-energy response and/or for
Compton-Suppressed Spectrometers
which require efficient capture of
escaping scattered photons Low
energy efficiency for the Marinelli
(re-entrant) beaker counting geometry
is compromised by the use of
a copper holder
Low-Energy Ge (LEGe) detectors,
optimized for very low energies, are
designed to detect radiation entering
the window (front) only
For these reasons Canberra uses
high-purity aluminum for REGe detectors
and the less expensive high-purity
copper for other types If your
application warrants the use of
high-purity aluminum with other detector
types it can be supplied as
an option
D Entrance Window
For Coaxial detectors the aluminum
end-cap, which is made seamless and
thus demonstrates excellent vacuum
integrity, is the obvious choice
The aluminum window, which can be
made as thin as 0.5 mm on request,
transmits 60% of 20 keV photons but
only 2% of 6 keV photons Our
low-energy detectors, the XtRa, REGe andLEGe types, are normally equippedwith a beryllium window which, at0.5 mm thickness, transmits 65% of 6keV photons Beryllium, however,even in the purest grade available, isnot low background
The best commercially available highpurity beryllium foil is 99.8% pureand contains significant level ofprimordial radionuclides (severalparts per million) While it is farbetter than commercial grades ofberyllium foil which may have up
to 100 PPM uranium content, it isnot a good choice of Ultra Low-Background systems
Faced with this dilemma, Canberrahas mastered the use of space-agecomposite materials in this applica-tion The resultant window is acarbon-fiber composite which is lightyet strong It is conductive andtherefore provides EMI resistance It
is helium leak-tight and it is virtuallyfree of radioactive contaminants
it can be supplied as an option
Shields
As in the case of cryostats, shielddesign and material choices have adrastic effect on system performancewhen it comes to low-level counting.Shield design also has a big impact onease of detector installation and use.And, as the most massive objectsfound in most counting systems,shields affect the overall appearance
of the laboratory where they arelocated
Canberra shields provide a balancedcombination of performance,ergonomics, and appearance TheUltra Low-Background versions
of the standard shields differ only
in the materials used in construction
so both standard and low-levelshields are compatible with awide variety of standard andlow-background cryostats
Figure 3
Window transmission characteristics
In the standard 0.5 mm thickness ittransmits 80% of 10 keV photons
Finally, this window does not corrode,
is virtually unbreakable, and unlikeberyllium, is non-toxic
For all these reasons Canberra’scarbon-composite window should
be used whenever possible in lowenergy, low background systems
If an alternative window is necessary,
Materials
Bulk - The best choice for the bulkshielding for low-level counting islead Canberra has a source of 60Cofree steel and this material is suitablefor many applications, but for ultralow-level counting it is a poor choicebecause of the increased Comptonscatter and resultant continuum ofbackground counts in the range of
Trang 34is designed to stop the K-shell X raysand everything below them in energy,there is no huge advantage to old lead
in applications where backgroundcontinuum from cosmic interactionsdominates, as is the case for mostabove-ground systems having inertshielding only For systems operatingunderground or with active cosmicguard detectors, the beta bremsstrahl-ung contribution to background maybecome significant
If the effects of 210Pb are stood, the solutions proposed for leadbackground range from the sublime
misunder-to the ridiculous “Virgin” lead issuggested but in this case the 210Pbwill have had little or no chance todecay “Doe Run Mine” lead issometimes specified but this mineclosed around the turn of the century.Indeed some lead deposits may beless active in primordial radionuclides,but no “clean” virgin lead is known
to exist so selection is the rule for thelead bulk
If the operating conditions warrantthe use of lead with certified low
210Pb content, we can supply shieldsmade from selected lead Depending
on the 210Pb content, this can drasticallyincrease the cost
Figure 4
Background continuum vs shield material
100-300 keV This effect is shown in
Figure 4 which compares the
back-ground continuum from a detector
operating in a 15 cm thick steel shield
to that of the same detector operating
in a 10 cm thick lead shield
This continuum can be reduced by
adding a lead liner to the steel shield
Our experiments have shown that
such a liner, even as great as 25 mm
in thickness, does not reduce the
continuum to the level obtained with
lead bulk
Lead contains 210Pb in varying
concentrations The refining process
does not separate 210Pb from the stable
isotope, but since 210Pb has a half life
of 22 years, “old” lead can be notably
lower in 210Pb than “new” lead The
source of the lead ore
is also a big factor in 210Pb
concentration
Now the 46.5 keV gamma ray from
210Pb is readily stopped by the graded
liner used to suppress lead K-shell
X rays However, the 210Bi daughter
of 210Pb is a beta emitter with an
end-point energy of 1161 keV It has been
suggested that bremsstrahlung from
this beta leads to a significant increase
in the background continuum up to
several hundred keV The normal
graded liner would be ineffective in
this energy range To check this
theory we tested samples of lead
with varying 210Pb content on a
low-background LEGe detector with
210Pb levels (the beta obviously excitesthe K-shell X rays) However, thecontinuum differences are fairly small,even with the roughly 3:1 difference
in 210Pb content Above 500 keV, there
is no difference in backgrounds Sincethe graded liner
Figure 5
Background spectra from lead samples Top - 210 Pb content ≈ 20 mBq/gm Bottom - 210 Pb content ≈ 60 mBq/gm
Trang 35Liner
Again there are trade-offs between
background continuum and lead
X-ray peaks The graded liners
typically used to suppress the lead
X rays (75-85 keV) consist of 0.5 to
1.5 mm thick layers of cadmium and
copper The cadmium is an effective
filter for lead X rays while the copper
attenuates the cadmium X rays and
prevents personnel exposure to the
toxic cadmium
This graded liner has the undesirable
effect of increasing the background
continuum however This effect is
illustrated in Figure 6 which shows
that copper alone in the thickness
necessary to stop lead X rays (5 mm
for 98%) will almost double the
background continuum in the 200
keV range
If the lead X-rays were of no concern
in the application, and if toxicity and
decontamination were of no concern,
shields would perform better without a
liner Generally this is not the case,
however, so Canberra shields are
equipped with graded liners – but
with a difference Canberra shields
are built with a tin and copper liner
In the interest of the environment
and in the safety of our workers
and customers we have eliminated
cadmium entirely
Many shields are equipped with
0.5 mm of cadmium but this will stop
only about 70% of the lead X rays
One mm of tin will stop about 95%
of the lead X rays With an additional
1.5 mm of copper, the total attenuation
of lead X rays in the Canberra shields
is about 98.5% Another disadvantage
of cadmium is high cross-section for
neutrons from cosmic radiation For
example, the 113Cd (η, γ) 114Cd reaction
results in a prominent background
peak at 558.2 keV and a lesser peak at
651 keV
Important features of the detector/
shield system are illustrated in Figure 7
A Door closes tightly against
shield body to prevent
streaming and to allow
shielding against radon
Figure 6
Background continuum as a function of copper liner thickness
B Shielding materials chosen
for attenuation, backgroundcontribution, and scatteringproperties
C Shield penetration for
detector entry held to aminimum
D Preamplifier and Dewar
Trang 36Canberra Industries Inc., Nuclear Products Group, 800 Research Parkway, Meriden, CT 06450 U.S.A.
Tel: (203) 238-2351 Toll Free 1-800-243-4422 FAX: (203) 235-1347 With Offices In: Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Russia, United Kingdom.
System Performance
These overlapped spectra (Figure 8)
show backgrounds from a 40% coaxial
detector (1) unshielded (2) with standard
cryostat in 4 inch thick shield and
(3) with Ultra Low-Background
7500SL-RDC cryostat in 4 inch thick
shield Virtually no peaks due to
cryostat materials appear in spectrum (3)
Further reduction in the continuum
and in the 40K and 208Tl (Nat Th) high
energy lines require thicker shielding,
active shielding (cosmic guard detectors)
or subterranean operation
Cryostat Options
Table 1 lists the Ultra Low-Background
cryostat hardware options and model
numbers The base cryostat and
hardware options are the same for all
detector types The materials options
are different for different detector
types Consult the factory for a
specific proposal for detectors,
cryostats and shields
CAN0012 8/95 Printed in U.S.A.
For example a coaxial detector in a vertical dipstick cryostat requires the following items:
Cryostat: 7500SL, Remote Detector Chamber: RDC, Low-Background materials: ULB - GC
Table 1 - Model List For Ultra Low-Background Cryostat Options
Cryostat Type
Base Cryostat
Hardware Option
Ultra Low-Background Material Vertical Dipstick
U-Style
7500SL 7915-30
RDC SL
ULB-GC (Coaxial) ULB-GR (REGe) ULB-GL (LEGe) ULB-GW (Well)
Trang 37Germanium Detectors
Trang 38The Company
Canberra Industries is the world's largest manufacturer of nuclear radiation detection and analysis systems In order to better serve our customers worldwide, Canberra operates detector manufacturing facilities in the U.S and in Europe Canberra Semiconduc- tor, N.V., which is located in Olen, Belgium, provides Ge detectors to the European market In addition, C.S.N.V manufactures a semiconductor detector for charged particles known as the PIPS (Passivated Implanted Planar Silicon) Detector Detector Products Division, which shares the home plant in Meriden, Connecticut, provides Ge detectors to the world market outside Europe, and manufactures Si(Li) X-ray Spectrometers as well as a comprehensive line of shields and accessories for detectors and detector systems.
The cover shows a few of the many types of detectors, shields and accessories manufactured by Canberra Our facility in Olen, Belgium and our home plant in Meriden, Connecticut are shown on the back.
Consult the catalog or your Canberra sales office for information on other Canberra products, including Si(Li) detectors, PIPS tors, NaI (TI) detectors, shields, and accessories.
detec-Germanium Detectors
Germanium detectors are semiconductor diodes having a P-I-N structure in which the Intrinsic (I) region is sensitive to ionizing radiation, particularly X-rays and gamma rays Under reverse bias, an electric field extends across the intrinsic or depleted region When photons interact with the material within the depleted volume of a detector, charge carriers (holes and electrons) are produced and are swept by the electric field to the P and N electrodes This charge, which is in proportion to the energy deposited in the detector
by the incoming photon, is converted into a voltage pulse by an integral charge sensitive preamplifier.
Because germanium has a relatively low band gap, these detectors must be cooled in order to reduce the thermal generation of charge carriers (thus reverse leakage current) to an acceptable level Otherwise, leakage current induced noise destroys the energy resolution
of the detector Liquid nitrogen, which has a temperature of 77°K is the common cooling medium for such detectors The detector is mounted in a vacuum chamber which is attached to or inserted into an LN2 dewar or an electrically powered cooler The sensitive detector surfaces are thus protected from moisture and condensable contaminants.
The Best Detector for Your Application
Canberra offers the widest choice of detector types in the industry Employing the appropriate technology in both materials and processing techniques, Canberra can offer the optimum detector for a wide range of applications We use both p-type and n-type germanium and we use diffused, implanted, and barrier contacts to achieve this product variety.
The illustrations and charts below depict the various detector geometries that are available from Canberra, the energy range they cover, and their salient performance characteristics Consult the individual specification sheets for detailed descriptions, performance ranges, and model availability of each type.
Trang 40Your Choice of Cryostat Type
Canberra Cryostats are manufactured in flanged and slimline
styles, with configurations to suit virtually any application Salient
features of both types of cryostats are shown in the illustrations
below Slimline Cryostats are available in the Canberra patented
convertible version.
Consult the individual cryostat spec sheets for more details on the complete product line Special cryostats are our forte Consult your Canberra salesman for information on:
• Low Background Cryostats and Systems
• Multi-Element Detector Telescopes
• Remote Detector Chamber Cryostats for Compton Suppression and Low Background Counting
• Cryostat Arrays for Pu and U Lung Burden Systems
• Electrically Cooled Detectors
• X-ray and γ -ray Array Detectors
Standard Cryostat Configurations
• Vertical Dipstick
• Horizontal Dipstick
• Horizontal Integral
• Vertical (Down-looking) Integral
• Portable Multi-Attitude (MAC and Big MAC)
• U-Type Integral (Fixed and Swivel-head)