Table 17.3Leaks Section 17.4.1: Symptoms and Sources Pre-pump 17.4.1.1 Fitting Loose fitting Tighten or replace Proportioning manifold low-pressure mixing Loose or damaged valve Tighten v
Trang 1Table 17.3
Leaks (Section 17.4.1): Symptoms and Sources
Pre-pump
(17.4.1.1)
Fitting Loose fitting Tighten or replace Proportioning
manifold (low-pressure mixing)
Loose or damaged valve
Tighten valve or replace manifold
Pump (17.4.1.2) Fittings Loose or damaged Tighten or replace
Cross-threaded Replace; may require new
pump head Seal damaged Replace Pump seal Worn seal Replace Auxiliary component Loose fitting;
damaged component
Tighten or replace
High-pressure
fitting (17.4.1.3)
Stainless steel Loose or
contaminated
Tighten 1/4-turn; clean or replace
contaminated
Turn off pump, reseat tube, retighten; not for use
>6000 psi
Autosampler or
manual injector
(17.4.1.4)
Fittings Loose or damaged See fittings adjustments,
Sections 17.4.1.1, 17.4.1.3 Low-pressure needle
seal
Worn or loose seal Adjust or replace High-pressure needle
seal
Worn or damaged seal
Adjust or replace
Cross-port leakage Damaged rotor seal Clean, replace rotor seal;
may require stator replacement
External leaks not at fittings
Worn rotor seal Clean, replace rotor seal Insufficient
rotor-to-seal pressure
Adjust
Seal-pack failure (Waters only)
Worn or damaged seals
Service or replace
Trang 2Table 17.3
(Continued)
Column (17.4.1.5) Tube connections Loose or damaged See fittings adjustments,
Section 17.4.1.3 End-fitting Loose or damaged Tighten; may require new
column End-fitting on
cartridge column
Loose or damaged fitting; damaged seal
Tighten; replace seal; may require new assembly Detector (17.4.1.6) Connecting fittings Loose or damaged See fittings adjustments,
Sections 17.4.1.1, 17.4.1.3 Cell leaks Loose fitting Tighten
Over-pressure damage
Remove source of excessive pressure
Failed window seal Adjust or replace Non-UV detector See Chapter 4 and
operator’s manuals
Table 17.4
Pressure Problems (Section 17.4.2): Symptoms and Sources
Pressure too high
(17.4.2.1)
General Wrong mobile phase;
temperature too low; flow rate too high; wrong column; wrong particle size
Use correct conditions
Blocked tubing Buffer precipitation;
sample particulates
Replace tubing; remove problem source Blocked or partially
blocked frit
In-line filter Replace frit; consider
additional sample cleanup
Guard column Replace guard column Analytical column Reverse column (if
permitted); replace column; consider using in-line filter
Upper pressure limit mid-gradient
Normal gradient pressure changes
Adjust upper pressure limit to accommodate
(continued overleaf)
Trang 3Table 17.4
(Continued)
Pressure too low
(17.4.2.2)
General Wrong mobile phase;
temperature too high; flow rate too low; wrong column
Use correct conditions
Lower pressure limit Insufficient mobile
phase
Refill reservoir Overly sensitive limit
switch
Disable lower pressure limit
Massive leak Isolate and fix
(Section 17.3.1) High-pressure fitting Loose or
contaminated fitting
Tighten, clean, or replace Low-pressure fitting Loose or
contaminated fitting
Tighten, clean, or replace Low-pressure mixing
manifold
Air leak from unused solvent line; leaky proportioning valve
Prime unused lines with organic solvent; replace faulty proportioning valve
Inlet check valve Sticking with ACN
mobile phases
Sonicate in MeOH; replace
Pump starvation (failed siphon test)
Blocked inlet-line frit Replace Faulty proportioning
valve
Replace Pinched or blocked
tubing
Clear or replace Bubbles in pump Inadequate degassing
or purging; degasser failure
Degas mobile phase and purge pump; replace degasser
Pump seal Worn pump seal Replace Pump piston Broken or scratched
piston
Replace Pressure too
variable
(17.4.2.3)
Bubbles in pump Inadequate degassing
or purging
Degas mobile phase and purge pump
Inlet check valve Sticking with ACN
mobile phases
Sonicate in MeOH; replace
High-pressure fitting Loose or
contaminated fitting
Tighten, clean, or replace Low-pressure fitting Loose or
contaminated fitting
Tighten, clean, or replace
Trang 4Table 17.4
(Continued)
Low-pressure mixing manifold
Air leak from unused solvent line; leaky proportioning valve
Prime unused lines with organic solvent; replace faulty proportioning valve
Pump starvation (failed siphon test)
Blocked inlet-line frit Replace Faulty proportioning
valve
Replace Pinched or blocked
tubing
Clear or replace Pump seal Worn pump seal Replace Pump piston Broken or scratched
piston
Replace
Table 17.5
Interpreting Retention-Factor k and Retention-Time t RChanges (Section 17.4.3)
k Changes t RChanges Possible Cause More Information
No Yes Flow rate (settings, leaks, bubbles, pump problems) Section 17.4.3.1
Yes Yes Mobile-phase error (%B, pH, additives) Section 17.4.3.3 Yes Yes Stationary-phase problem (aging, dirty samples) Section 17.4.3.4 Yes Yes Temperature problem (poor control) Section 17.4.3.5
Table 17.6
Retention-Time Problems (Section 17.4.3.6): Symptoms and Sources
Abrupt change
in t R
When column changed Wrong column; poor
column-to-column reproducibility (especially type-A columns)
Use correct column; use columns from same batch; adjust method for more robust separation; use more reproducible columns (e.g., type-B); Section 2.5.4.6, Table 17.5 When mobile phase
changed
Improperly formulated mobile phase
Make new batch of mobile phase; Table 17.5
(continued overleaf)
Trang 5Table 17.6
(Continued)
In gradient elution, with change in instrument
Different dwell volume Adjust or compensate dwell
volume; Section 9.3.8.2
No apparent change in conditions
Equipment, bubble, or temperature problem
Check for leaks (Table 17.3, Section 17.4.1); check for pressure problems (Table 17.4,
Section 17.4.2); check k
changes (Table 17.5, Section 17.4.3); Section 2.5.4.6
Retention drifts During first few
injections only
Slow column equilibration;
column ‘‘loading’’
for sample
Normal for some columns and samples; allow longer equilibration; ignore first few injections (Section 3.10.2.2); Sections 2.7.1 (general), 7.4.3.2 (ion-pair), 8.5 (normal-phase), 9.3.7 (gradient elution) Over entire sample
batch
Unstable mobile phase; mobile-phase evaporation;
excessive helium sparging
Use stable mobile phase; cover reservoir to reduce evaporation; reduce helium sparging time or change to different degassing technique (Section 3.3) Retention too
small
Only for large-mass injections for some
or all peaks, with right-triangle peak shape
Mass overload Reduce mass of sample
injected; Table 17.11, Sections 17.4.5.3, 2.6.2
All peaks Mobile phase error;
flow rate, column,
or temperature change
See Table 17.5
Some peaks pH problem or other
mobile-phase error;
column aging
Make appropriate adjustments; Section 7.3 (pH); replace column Only for polar solutes Ionic samples Change mobile-phase pH
(Section 7.3); use ion pairing (Section 7.4)
Trang 6Table 17.6
(Continued)
Neutral or non-ionic samples
Try EPG or AQ column; use normal phase (Chapter 8)
or HILIC (Section 8.6) Retention too
large
All peaks Mobile-phase error;
flow rate, column,
or temperature change; leak
See Tables 17.5, 17.3
Some peaks pH problem or other
mobile-phase error;
column aging
Make appropriate adjustments; Section 7.3 (pH); replace column
Table 17.7
Peak-Area Problems (Section 17.4.4): Symptoms and Sources
Peak area to large
(17.4.4.1)
All peaks larger by same proportion
Injection volume too large
Adjust Sample preparation errors
Use correct procedure Detector settings
wrong
Adjust Different peaks in
sample larger by different amounts
Detector settings wrong
Adjust
Peak(s) present in blank
Late elution from prior injection
Extend run; add strong-solvent flush; modify sample pretreatment Carryover Add autosampler wash steps;
change wash solvent; adjust connections; change sample loop material; rearrange injection order Peak area too small
(17.4.4.2)
All peaks smaller by same proportion
Injection volume too small
Adjust Sample preparation errors
Use correct procedure
(continued overleaf)
Trang 7Table 17.7
(Continued)
Detector settings wrong
Adjust Different peaks in
sample smaller by different amounts
Detector settings wrong
Adjust
Peak area too
variable
(17.4.4.3)
Replicate injections give constant areas
Problem prior to injection
Check sampling and sample pretreatment steps Replicate injections
give inconsistent areas
Problem from injector onward
Check autosampler, pump, detector, and integration
Table 17.8
Baseline-Drift Problems (Section 17.4.5.1): Symptoms and Sources
Periodic drift 1 cycle/run Normal with
gradient elution
Ignore; use higher wavelength; change solvent; add UV absorber Several hours or 1
cycle/day
Laboratory temperature cycle
Thermostat oven; move system from drafts; adjust lab HVAC
Other drift Baseline drifts, then
stable
Normal equilibration;
detector warm-up
Wait until baseline stabilizes
Table 17.9
Baseline-Noise Problems (Section 17.4.5.2): Symptoms and Sources
Short-term noise High-frequency (50 or
60 Hz)
‘‘Dirty’’ electrical supply; too small
of detection time constant; too high of data rate
Use UPS to cleanup supply; use larger time constant or
RC filter; use slower data collection rate
50–60 Hz< noise <
peak width
Inadequate mobile-phase degassing
Use better degassing technique
Trang 8Table 17.9
(Continued)
Long-term noise Cyclic Bubble in pump;
piston seal or piston damage;
sticking or leaking check valve
Degas mobile phase; service pump; clean or replace check valve
Appears as chromatographic peak
Late-eluted solute Extend run; add
strong-solvent flush to method; change sample cleanup
Random, irregular baseline disturbances
Accumulated late-eluted non-polar materials from sample
Add strong-solvent flush to method; replace column; change sample cleanup
Peaks appear in blank gradients (‘‘ghost’’
peaks, Figs 17.11, 17.12)
Contaminated mobile phase
Replace mobile-phase reagents with fresh and/or higher purity reagents
Table 17.10
Peak-Shape Problems (Section 17.4.5.3): Symptoms and Sources
Source
Solution
Peak tailing TF ≤ 1.2 Normal None required
1.2 < TF ≤ 2 If increased over time:
column aging or contamination;
mobile phase error
Flush or replace column; prepare new batch of mobile phase
If constant, may be normal
No action or explore
solutions for TF > 2
TF > 2 For all peaks Also see split or distorted
peaks Early peaks tail more
than later peaks (Fig 17.13):
extra-column effects
Reduce extra-column volume (Sections 2.4.1, 3.9)
(continued overleaf)
Trang 9Table 17.10
(Continued)
Source
Solution
Worse for bases and ionic compounds
(Fig 5.9a): silanol
effects; trace metal contamination
Adjust pH; use higher purity or less-active column; see Sections 5.4.4.1, 7.3.4.2 Inadequate buffering
or mobile-phase additive concentration
Increase buffer or additive concentration; see Section 7.2.1.1 Right-triangle peak
tailing accompanied
by earlier retention
(Fig 17.15a)
Mass overload of the column
Reduce injection volume
or dilute sample for smaller mass on column (Section 2.6.2) Peak fronting TF ≥ 0.9 Normal None required
TF < 0.9 For ion pairing,
especially with type-A columns
Change temperature
±5−10◦C may help; change to type-B column
Column void or bed collapse
Replace column; operate column below high pH-limit
Broad peaks N > 75% of
manufacturer’s test
May be normal No action required;
compare to new column test with sample solutes
High-molecular-weight compounds (proteins, polymers, etc.)
Some broadening normal
No action required
Gradual broadening over>500
injections
Normal column aging Flush column with strong
solvent; replace column
In the presence of narrower peaks (Fig 17.5)
Late elution Extend run; adjust run
length; use gradient flush; improve sample cleanup
Accompanied by longer retention times
Column temperature too low
Increase temperature
Trang 10Table 17.10
(Continued)
Source
Solution
Narrower peaks broadened more than normally wider peaks
Detector time constant too large; integrator data-rate too low;
excessive smoothing (especially MS detection)
Use time constant≤1/10 peak width; collect a minimum of 20 points across a peak; use less smoothing
early peaks broad or flat-topped
(Fig 17.16b)
Too large an injection volume
Reduce injection volume; dilute injection solvent (Sections 2.6.1, 3.6.3) Split or distorted
peaks
All peaks distorted in same manner (Figs
17.17, 17.19)
Partially blocked column inlet-frit;
column void
Reverse-flush column; use in-line filter or better cleanup; replace column (Section 5.8) Broad or distorted
peaks, especially at end of run
(Fig 17.20b),
accompanied by change in retention time
Mismatch of mobile phase and column temperature
Use better temperature control; use mobile phase pre-heater (Section 3.7.1)
Early peaks distorted
(Fig 17.16c),
usually accompanied by earlier retention
Too large a volume of too strong a sample-injection solvent
Use more dilute injection solvent; reduce injection volume (Sections 2.6.1, 3.6.2.2)
Table 17.11
Failed System Performance Tests (Section 17.4.6): Symptoms and Sources
Source
Solution
Gradient linearity Steps in gradient Blocked reservoir
frit; bad proportioning valve
Check/replace frit; replace proportioning manifold
Segmented appearance
in linear gradient
Software control error
Adjust control parameters (see service manual)
(continued overleaf)