In the remainder of Section 5.4, we will first discuss the basis of column selectivity, which can be attributed to different interactions between solute molecules and the column.. 5.4.1 B
Trang 1C ≡N
CH3
(a) Alkylsilica columns
X
Phenyl columns Cyano column “Other”columns
Embedded-polar-group R-NH-(C=O)-O- Carbamate
R-NH-(C=O)-NH- Urea R-(C=O)-NH- Amide Fluoro columns
Perfluorophenyl (PFP)
-CF2CF2CF3 Fluoroalkyl
F
Figure5.19 RPC columns classified according to the ligand (figures omit the connecting silane group [–Si(CH3)2–])
simplified cartoons of Figure 5.19 (the—Si[CH3] group is omitted in Fig 5.19a–d).
The ligand of a RPC column is often an alkyl group, for example, C3, C8, C18
(Fig 5.19a) Alternatively, the ligand may consist of phenylpropyl or phenylhexyl, called phenyl columns (Fig 5.19b) If the ligand is –C3–C≡N (Fig 5.19c), we have a
cyano column The alkyl group may also be substituted by other functional groups
X (Fig 5.19d), and this gives rise to the additional column types listed at the bottom
of Figure 5.19 So-called embedded-polar-group (EPG) phases have been growing
in popularity, because of their compatibility with low %B mobile phases, their reduced silanol interactions, and unique selectivity (Section 5.4.1); peak shape for basic solutes is usually quite good with these columns The ligands in these phases contain amide, carbamate, urea (all of which are strong hydrogen-bond bases), or other polar functional groups embedded within the ligand structure Some EPG
Trang 2packings tend to be less stable than comparable alkyl or aryl columns The nature
of the ligand mainly determines column selectivity, which is the subject of following Section 5.4
Column selectivity can be important for different reasons During method devel-opment a change of column may be necessary to improve selectivity and increase resolution (Sections 2.5.2, 5.4.3) For the latter application we must be able to
iden-tify a second column with quite different selectivity When a routine RPC procedure
is used at different times and places, a replacement column from the same source may not be immediately available locally, or too costly, or impractical for other reasons (Sections 5.4.2, 6.3.6.1) In this case we must identify a column of equivalent (or
at least similar) selectivity For either situation, we require a quantitative procedure that allows us to compare column selectivity Column selectivity is also related to certain problems that can arise during either method development or the routine use
of an RPC procedure: peak tailing, the deterioration of a column during use, and
‘‘de-wetting’’ of the column when used with mobile phases that are predominantly aqueous (Section 5.3.2.3) Finally, knowledge concerning column selectivity helps
us understand sample retention as a function of the column and solute molecular structure, in turn preparing us to better deal with various separation challenges
In the remainder of Section 5.4, we will first discuss the basis of column
selectivity, which can be attributed to different interactions between solute molecules and the column This will lead to quantitative values of those properties of a
column that determine its selectivity Finally, we will discuss the use of these
column-selectivity properties for both method development and the routine use of
an RPC procedure
5.4.1 Basis of RPC Column Selectivity
As discussed in Section 2.3.2.1, solute retention is determined by various interactions among the solute, mobile phase, and stationary phase (column) The relative impor-tance of different solute–column interactions—and column selectivity—depends on the composition of the stationary phase and the molecular structure of the solute Figure 5.20 illustrates eight different interactions that can affect column selectivity:
(a) hydrophobic interaction
(b) steric exclusion of larger solute molecules from the stationary phase
(here referred to as ‘‘steric interaction’’)
(c) hydrogen bonding of an acceptor (basic) solute group by a donor (acidic)
group within the stationary phase (usually a silanol –SiOH)
(d) hydrogen bonding of a donor (acidic) solute group by an acceptor (basic)
group within the stationary phase (represented here by a group ‘‘X’’)
(e) cation-exchange or electrostatic interaction between a cationic solute and
an ionized silanol (–SiO−) within the stationary phase; also repulsion of
an ionized acid (e.g., R–COO−)
Trang 3(f) dipole–dipole interaction between a dipolar solute group (a nitro group
in this example) and a dipolar group in the stationary phase (a nitrile group for a cyano column)
(g, h) π –π interaction between an aromatic solute and either a phenyl group
(phenyl column) (g), or a nitrile group (cyano column) (h)
(a)
Hydrophobic interaction
COCH3
Hydrogen bonding (acidic solute)
(d)
O
C
OH
(e)
Cation exchange
NH +
(g)
π-π interaction
(phenyl column)
O
O2N
π-π interaction (cyano column) O
C=N
O2N
NO2
(c)
Hydrogen bonding
(basic solute)
N
(b)
Steric exclusion
(f)
Dipole-dipole interaction
O
C= N
+ −
O 2 N
−
Figure5.20 Solute-column interactions that determine column selectivity (figures omit the connecting silane group [–Si(CH ) –])
Trang 4O M ++ O
N N
Complexation of chelating solutes
(i )
Figure5.20 (Continued)
(i) complexation between a chelating solute and metal contaminants on the
particle surface
Interactions (a–e) can be significant for every column; dipole interactions (f ) are
only important in the case of cyano columns, andπ –π interactions (g, h) occur only
for phenyl and cyano columns [57] Both dipole andπ –π interactions are inhibited
by the use of acetonitrile as B-solvent, which further minimizes their importance
for separations with acetonitrile Complexation with surface metals (i) can result
from the use of a less pure, type-A silica, leading to broad, tailing peaks (very undesirable); the chelating soluteα,α-bipyridyl has been used to test columns for
metal complexation Because phenyl and cyano columns are used less often, and
type-A columns are not recommended, we will emphasize interactions (a–e) in this
chapter (but see [57, 58])
5.4.1.1 Solute–Column Interactions
The various solute–column interactions of Figure 5.20, which determine column selectivity, have been understood in general terms since the 1980s; see [59] for a good discussion of recent attempts at characterizing column selectivity However, only after 2000 did it become possible to reliably characterize RPC column selectivity
in terms of these interactions This was accomplished by the development and
application of the hydrophobic-subtraction model [60–62], which recognizes that
hydrophobic interactions are by far the most important contribution to RPC retention If only hydrophobic interactions were significant, a plot of values of log
k for one column against another would give a straight line with no scatter of
data around the line As seen in Figure 5.21, this is approximately the case for these two C18 columns (Inertsil ODS-3 and Stablebond C18)—however, values
of log k for aliphatic amides () and protonated strong bases () fall below
the best fit to these data These latter deviations are due to interactions of these solute molecules with silanol groups (silanol interactions are more significant for the StableBond C18 column) These and other smaller deviationsδ log k from this
plot (see the expanded inset of Fig 5.21) represent contributions to retention from
nonhydrophobic interactions b–e of Figure 5.20 It is possible to analyze values of
δ log k for the combination of different solutes and columns so as to separately
evaluate the five interactions of Figure 5.20a–e For columns other than phenyl or cyano (i.e., those for which only interactions a–e of Fig 5.20 are significant), values
Trang 5δlog k
2.0
1.5
1.0
0.5
0.0
−0.5
−1.0
log k (StableBond C18)
y = 0.21 + 1.01 x
r 2 = 0.995 amides protonated bases
Figure5.21 Comparison of retention on two different C18columns Data for 90 different organic compounds Conditions: 15× 4.6-mm columns; 50% acetonitrile-water, pH-2.8
phosphate buffer; 2.0 mL/min; 35◦C Adapted from [61]
of k can be related to the interactive properties of the solute and the column:
log
k
k EB
=ηH − σS∗+ βA + αB + κC (5.3)
(i) (ii) (iii) (iv) (v)
Here k and k EB are values of the retention factor for a given solute and the
reference compound ethylbenzene (EB), respectively Terms i–v of Equation (5.3) correspond, respectively, to the interactions of Figure 5.20a–e Quantities η, σ,
β, α, and κ refer to properties of the solute molecule: hydrophobicity (η),
‘‘bulkiness’’ (σ), hydrogen-bond (H-B) basicity (β), H-B acidity (α), and effective ionic charge (κ) Corresponding column parameters are of primary practical interest:
H, hydrophobicity; S*, steric interaction, or resistance by the stationary phase to penetration by bulky solutes; A, H-B acidity; B, H-B basicity; and C, ion-exchange capacity or electrostatic (coulombic) interaction Columns with similar values of H, S*, etc., will possess similar selectivity and provide a similar retention order for peaks within the chromatogram Columns with different values of H, S*, etc., will differ in
selectivity and provide changes in relative retention The hydrophobic-subtraction model and Equation (5.3) best summarize our present understanding of RPC retention and column selectivity [59]
We will next relate terms i–v of Equation (5.3) to the interactions of Figure 5.20a–e Hydrophobic interaction is illustrated in Figure 5.20a, by the
Trang 6interaction of the solute 2-n-octanone (CH3COC6H13) with the ligand groups of
a C8 column Values of column hydrophobicity H increase for longer ligands
(e.g., C18 vs C8), a higher concentration of ligand groups on the silica surface (μmoles/m2), smaller pore diameters (e.g., 8- vs 30-nm pores), and the presence of
column end-capping An increase in H increases the retention of more hydrophobic
molecules, those with larger values ofη
Steric exclusion or ‘‘steric interaction’’ is illustrated in Figure 5.20b by the
retention of two polycyclic aromatic hydrocarbon (PAH) isomers: the narrow, long naphthacene and the more ‘‘bulky’’ triphenylene Naphthacene is better able to squeeze between adjacent ligands, but if the spacing of column ligands is increased (lower ligand concentration), it becomes easier for the bulky triphenylene to enter
the stationary phase The column parameter S* measures the ‘‘tightness’’ of the
stationary phase or the difficulty that bulky solute molecules experience in squeezing
between the ligands; larger values of S* mean a ‘‘tighter’’ stationary phase and relatively less retention of bulky solute molecules Values of S* increase for longer
ligands, a higher concentration of the ligand (ligands closer together), and smaller pore diameters Solute bulkiness is measured by its value ofσ Steric exclusion is a somewhat complex phenomenon; see Section 5.4.1.2 below for further insights
Hydrogen bonding of a non-ionized basic solute (e.g., pyridine) by a column
silanol is illustrated in Figure 5.20c The hydrogen-bond acidity A of the column
is due to the presence of surface silanols, and therefore decreases when the column
is end-capped (due to the removal of some silanols and blocking of others; see the
example of Fig 5.16d) The silanols of type-A columns are usually more acidic than
those present in type-B columns; therefore values of A tend to be larger for type-A
columns The H-B basicity of the solute is measured by its value ofα; unprotonated amines and amides are more basic and have larger values ofα, while nitriles and nitro compounds are much less basic and have smaller values ofα Most other polar compounds have intermediate H-B basicities and values ofα
Hydrogen bonding of a solute that is a H-B acid (e.g., butyric acid) is illustrated
in Figure 5.20d The H-B basic group ‘‘X:’’ in the stationary phase is not specified because the nature of X differs for different kinds of RPC columns For type-B
alkylsilica columns, there is so far no compelling explanation for what groups
‘‘X:’’ consist of It has been suggested that water dissolved into the stationary phase corresponds to these groups ‘‘X:’’, on the basis of an inverse correlation
of values of A and H —as well as other arguments In the case of some type-A alkylsilica columns with large values of B, contaminating metals in the silica appear
to comprise the ‘‘X:’’ groups Alternatively, for embedded-polar-group columns, the polar group (which is usually a H-B base) very likely corresponds to the ‘‘X:’’ group.
An increase in B leads to increased retention of carboxylic acids, which have large
values ofα; the retention of other H-B acids, such as alcohols and phenols, is less
dependent on B (smaller values of α)—although phenols are preferentially retained
on embedded-polar-group columns (i.e., phenols appear as stronger H-B acids when these columns are used)
Column ion-exchange capacity C is a measure of the ionization of the silica
and the accessibility of ionized silanols Silanol ionization and/or accessibility (and
values of C) increase (1) as mobile-phase pH increases, (2) for non–end-capped
columns, and (3) for type-A versus type-B columns While the main effect of larger
values of C is to increase the retention of protonated bases, it also results in a
Trang 7decrease in retention for ionized acids (because of electrostatic repulsion); the value
ofκfor a solute is approximately equal to its molecular charge (e.g., +1 for fully protonated bases,−1 for fully ionized acids) The main difference in selectivity for
type-A versus type-B columns is determined by their low-pH values of C; type-B columns have values of C < 0.25 at pH 2.8, while type-A columns have C > 0.25.
For columns with values of C < 0.00 at low pH, it is believed that these columns
carry a net positive charge [63], presumably the result of protonated amine groups
that are introduced during the manufacturing process for some columns Values of
H, S*, A, and B are assumed not to change with the pH of the mobile phase Values of the column-selectivity parameters H, S*, etc., have been measured
for over 400 different columns; see [64] for a partial listing, or for a current list of values contact one of the authors (or http://www.USP.org/USPNF/columnsDB.html) Average values of these column parameters are summarized at the top of Table 5.8
for several different kinds of RPC column Within a given column type, there is
also a significant variation in values of H, S*, etc., as illustrated at the bottom of
Table 5.8 for several type-B C18columns Consequently not all columns of a given
kind can be regarded as equivalent in terms of selectivity Apart from values of H and S*, for example, average retention as measured by values of k for ethylbenzene
(last column of Table 5.8) increases with the surface area of the particle
5.4.1.2 Shape Selectivity
The following, minor digression examines two distinct forms of steric exclusion; for now, the reader may prefer to skip to Section 5.4.2.
Two separate manifestations of steric exclusion have been described: steric
interaction, as measured by term ii of Equation (5.3), and shape selectivity [65].
Differences between these two phenomena are illustrated in Figure 5.22 for the
separation of two isomeric hydrocarbons on a polymeric column (Fig 5.22a) and
a monomeric column (Fig 5.22b) The basis of shape selectivity is illustrated in Figure 5.22a for a ‘‘narrow’’ molecule i, a ‘‘wide’’ molecule j, and a polymeric alkylsilica column The ‘‘wide’’ molecule j is excluded from part of the stationary
phase because its minimum cross-section (double-headed arrow) exceeds the spacing
between ligands (molecule j cannot ‘‘squeeze’’ between the ligands) In a monomeric column (Fig 5.22b), the ligands are further apart, so as to allow access of both narrow and wide molecules (i and j) to the stationary phase When a molecule
has access to the stationary phase, steric exclusion affects retention in a different way; now the hydrodynamic diameter of the molecule becomes important, rather than its minimum cross-section When the hydrodynamic diameter of the solute
molecule is comparable in size to the spacing between ligands (as for molecule i),
the retained molecule is restricted in its possible orientations within the stationary phase This restriction of the solute molecule reduces its retention, in a similar way as for size-exclusion chromatography (Section 13.8.1) Note that shape selectivity and steric interaction lead to dissimilar effects on retention as a function of molecular shape; thus they clearly represent two different contributions to retention [61] Because of the ‘‘either-or’’ nature of shape selectivity, it can result in relatively large changes in relative retention—whereas steric interaction has a smaller effect on RPC selectivity
Figures 5.22c, d illustrates the potential advantage of shape selectivity for
the separation of polycyclic aromatic hydrocarbons (PAHs) The polymeric column
Trang 8Table 5.8
Characterization of Column Selectivity by means of the Hydrophobic-Subtraction Model
(Eq 5.3)
Different column types
C18(type-B, wide-pore) 0.95 0.01 −0.05 0.01 0.22 0.31 3.2
C18(type-B, monolith) 1.01 0.02 0.12 −0.02 0.11 0.31 3.2
C18(type-B, hybrid) 0.98 0.01 −0.14 −0.01 0.13 0.05 6.3
C18(polar end-capped) 0.90 −0.04 −0.02 0.02 −0.02 0.40 7.4
Different narrow-pore, type-B C18columnsa
Zorbax StableBond C18c 1.00 −0.03 0.26 0.00 0.14 1.04 7.6 Zorbax Eclipse XDB-C18c 1.08 0.02 −0.06 −0.03 0.05 0.09 9.1 Kromasil 100–5C18d 1.05 0.04 −0.07 −0.02 0.04 −0.06 12.5 ProntoSIL 120–5 C18 SHe 1.03 0.02 −0.11 −0.02 0.11 0.40 8.7
Nucleodur C18 Gravityh 1.06 0.04 −0.10 −0.02 −0.08 0.32 11.0
aAll columns 5- μm particles; data of [73].
bAdvanced Materials Technology;cAgilent;dAkzo Nobel;eBischoff;fGL Science;gGrace-Alltech;
hMacherey Nagel;iACT;jMerck;kPhenomenex;lSupelco;mThermo/Hypersil;nWaters;oTosoh Bioscience.
Trang 9(a) (b)
(min)
(c)
(d)
Solute minimum cross-section
(width)
O
OH
Si
OH
Si
Si-OH
i
j
Solute hydrodynamic diameter
i
j
shape selectivity
Polymeric column with
closely spaced ligands
steric interaction
Monomeric column with widely spaced ligands
steric interaction
column
shape selectivity
column
Figure5.22 Different manifestations of steric exclusion Shape selectivity (a) compared with steric interaction (b) (c) Separation of a mixture of 13 polycyclic aromatic hydrocarbons on
a polymeric column (d) Separation of same sample with same conditions on a monomeric column (c) and (d) are adapted from [65].
for the separation of Figure 5.22c exhibits greater shape selectivity and therefore
provides a much greater differentiation (and better separation) of these different isomeric C22 PAHs, versus the corresponding separation in Figure 5.22d with a
monomeric column (where shape selectivity is minimal) Long, narrow molecules (compared to those that are short and wide) are preferentially retained when shape selectivity is more important, while short, wide solute molecules (of similar molecular weight) are more retained when steric interaction is dominant As a rule, we can say that shape selectivity is more important when C30 or polymeric columns are used, and sample molecules are both large and have very different ratios of length to width Most RPC separations are carried out with monomeric columns other than
C30, in which case steric interaction and values of S* largely define the effect of steric
Trang 10exclusion on column selectivity For further details on the practical utility of shape selectivity, see Section 6.3.5.2
5.4.2 Column Reproducibility and ‘‘Equivalent’’ Columns
Column manufacturers try to ensure that each column (e.g., Waters Symmetry C18) has similar properties and will perform satisfactorily and reproducibly in a routine
RPC assay Consequently the plate number N and column pressure drop for each
column usually is measured prior to its sale (Section 5.7); columns whose values of
N fall below some minimum value are discarded Similarly other tests are carried out
by the manufacturer (Section 5.7) to ensure that column selectivity stays the same from one batch to the next of the column packing (similar to the measurements of
values of H, S*, etc.) An example is shown in Figure 5.23 for several successive
batches of ZorbaxR Rx-C18, where the retention times for dimethylaniline and
toluene are plotted against the batch number Values of k for the two solutes vary
by±4% (1 SD), mainly as a result of small, unimportant differences in the surface
areas of the silica particles The ratio of these two k-values ( α) is a more direct
measure of column selectivity (primarily the important column-selectivity parameter
C of Eq 5.3); values of α vary by only ±0.5% Consequently it appears that the
selectivity of different batches of this column packing should be similar, especially for separations that involve protonated basic compounds
During the 1970s and 1980s HPLC column manufacturing had not yet devel-oped to its present advanced state, and column selectivity often varied significantly among different batches of a particular column More recently column repro-ducibility has improved (e.g., see [67]), but it is still possible that one column batch will differ enough from another in terms of selectivity to result in a failed separation—especially for demanding separations Various means exist for dealing with the latter problem (Section 6.3.6.1), one of which is to select an ‘‘equivalent’’
column from a different source Using values of H, S*, etc., as in Table 5.8b, it is possible to identify one or more columns with similar values of H, S*, etc., by means
5
4
3
2
1 α 0
k
Lot number
CH3 N(CH3)2
Figure5.23 Monitoring different batches of column packing for possible changes in selectiv-ity Sample: dimethylaniline and toluene Conditions: 150× 4.6-mm Zorbax Rx-C18columns; 50% acetonitrile-water plus pH-7 phosphate buffer; 1.6 mL/min; 22◦C Adapted from [66]