To develop and validate the determination method of nifedipine in drug - excipient compatibility testing samples by high performance liquid chromatography.
Trang 1DEVELOPMENT AND VALIDATION OF HIGH PERFORMANCE
LIQUID CHROMATOGRAPHY METHOD FOR THE
DETERMINATION OF NIFEDIPINE IN DRUG - EXCIPIENT
COMPATIBILITY TESTING SAMPLES
Tran Quang Trung 1 ; Nguyen Thanh Hai 2 ; Trinh Van Lau 3 ; Truong Ngoc Hien 1
SUMMARY
Objectives: To develop and validate the determination method of nifedipine in drug - excipient compatibility testing samples by high performance liquid chromatography Methods: Chromatographic conditions include: PDA detector, analytical wavelength, column and mobile phase were studied; the evaluation of the system suitability, specificity - selectivity, the linearity range, repeatability, intermediate precision, accuracy, range, limit of detection and limit of quantitation following the guidance of ICH Results: Chromatographic conditions were established, including: PDA detector, analytical wavelength 237 nm, column RP-C18 Phenomenex, mobile phase MeOH:H 2 O (65/35, v/v) The method had been evaluated and achieved the ICH’s regulations Conclusion: The developed HPLC method can be used to determine the nifedipine in drug - excipient compatibility testing samples, which will use in pre-formulation studies of the nifedipine dosage forms
* Keywords: Nifedipine; High performance liquid chromatography; Drug - excipient compatibility
INTRODUCTION
Nifedipine (NIF), a dihydropyridine
calcium channel blocker, is widely used in
the treatment of angina, hypertension and
other vascular disorders [2, 3] However,
nifedipine is relatively unstable [4] and
can interact with excipients in tablets
formulation that reduce its content In
order to determine the interaction between NIF and excipients to select suitable excipients for formulation development of NIF push - pull osmotic pump extended-released tablets, we have developed and validated the HPLC method to quantify NIF in drug - excipient compatibility testing samples created by combining NIF with potential excipients
1 Vietnam Military Medical University
2 Vietnam National University, Hanoi
3 National Institute of Drug Quality Control
Corresponding author: Tran Quang Trung (tqt201316@gmail.com)
Date received: 05/10/2019
Date accepted: 21/11/2019
Trang 2MATERIALS AND METHODS
1 Materials and equipment
* Materials and chemicals:
- Reference standard of NIF (potency
99.82%) was provided by National Institute
of Drug Quality Control
- NIF was supplied by Baoji Guokang
Bio-Technology Co., Ltd (China) and
meet USP 38 standards
- PEO (polyethylene oxide) N10, PEO
N80, PEO N750, PEO 301, PEO 303,
PEO Coagulant provided by Colorcon
Asia Pacific Pte., Ltd
- Lactose monohydrate, sodium chloride,
magnesium stearate, PVP K30, red iron
oxide were sourced from China and all
meet USP 38 standards
- HPLC grade acetonitrile, methanol
were procured from Merck Ltd
- Other used chemicals were of HPLC
or analytical grade
* Equipment:
HPLC system Alliance Waters 2695D;
Detector 2998 dual α Absorbance (USA);
Phenomenex LUNA column, Xbridge
column (RP-C18, 5 µm, 250 x 4.6 mm);
Spectrometer EMC-61PC-UV, Emclab
(Germany); pH meter Mettler Toledo
(Switzerland); analytical balance Mettler
Toledo (precision 0.1 mg), other usually
equipment in laboratory
2 Methods
* Method development:
- Preparation for standard solutions:
+ Standard stock solution: A standard
stock solution containing 500 µg/mL NIF
was prepared by exactly approximately
weighing 50.0 mg NIF standard,
transferring it into a 100 mL volumetric flask and adding about 70 mL mobile phase composed of methanol/H2O (65/35, v/v) The flask was sonicated for
30 mins, allowed to cool at room temperature and the volume was made
up with mobile phase
+ Standard solution: Accurately taken
3 mL of the stock standard solution
500 µg/mL into a 10 mL volumetric flask and made up the volume with mobile phase, mixed well to give a standard solution of about 150 µg/mL of NIF All standard solutions were filtered through a 0.45 µm PTFE membrane filter prior to injection into the HPLC system
- Preparation for sample solutions:
Accurately weighed an amount of powder mixture of NIF and excipients (in
a ratio of 1/1) equivalent to about 50.0 mg
of NIF, transfered into a 100 mL volumetric flask; 70 mL mobile phase (methanol/H2O, 65/35, v/v) was added, and the flask was sonicated for 30 mins, allowed to cool at room temperature and made up the volume with mobile phase From this solution, a volume of about 10 mL of solution was withdrawn and was centrifuged at 4,000 rpm for 20 minutes Taken exactly 3 mL
of centrifuged solution into a 10 mL volumetric flask and add the mobile phase just to the mark, mixed well, filtered through a 0.45 µm PTFE membrane filter prior to injection into the HPLC system
- Chromatographic conditions:
The NIF standard solution of 150 µg/mL concentration was used for investigation
of chromatographic conditions The solutions were scanned under 200 - 500 nm wavelength range to find the maximum
Trang 3absorption wavelength The other
chromatographic conditions such as:
stationary phase (silica gel column C18
Phenomenex LUNA, Xbridge); mobile
phase (mixture of solvents MeOH, ACN,
H2O at different proportions) were also
studied Maintained a flow rate at
1 mL/min, sample injection volume of
25 µL Identified chromatographic
conditions for peaks with good symmetry,
max peak area/concentration ratio and
peak height/concentration ratio
* Method validation:
The proposed HPLC-UV method was validated in accordance with the International Conference on Harmonization (ICH) guidelines and Circular 32/2018 of the Vietnamese Ministry of Health [1,5], by evaluating the validation characteristics such as the system suitability, specificity - selectivity, the linearity range, precision (repeatability, intermediate precision), accuracy, specified range, limit of detection and limit of quantitation
RESULTS AND DISSCUSION
1 Method development results
* Analytical wavelength:
The appropriate wavelength for NIF analysis was determined by recording UV spectrum of a NIF standard sample of 150 µg/mL on the EMC-61PC-UV spectrophotometer, Emclab The solution above was scanned from 200 to 500 nm with quartz cuvettes 1 cm of thick The blank sample was mobile phase composed of methanol/H2O (65/35, v/v)
Figure 1: Ultraviolet absorption spectrum of NIF standard solution of 150 µg/mL
UV-Vis spectrum showed that the absorption maxima of NIF was 237 nm and 370 nm From there, choose the wavelength 237 nm as the wavelength for NIF qualification by HPLC method
Trang 4* Column efficiency validation:
Analyzed samples on two reversed silicagel columns C18 Xbridge and Phenomenex LUNA
Table 1: Column analysis validation
Results
Figure 2: Chromatogram of the NIF standard sample using Xbridge column
Figure 3: Chromatogram of the NIF standard sample using
Phenomenex LUNA column
The results of asymmetry factor, the number of theoretical plates, retention time and
chromatograms showed that both columns had a symetrical peak, the AF was within the permitted limits (0.8 ≤ AF ≤ 1.5); however, the Phenomenex LUNA column with a smaller number of theoretical plates, shorter retention time compared to these ones in the Xbridge column (13.108 mins), therefore, the Phenomenex LUNA column was selected for the quantitative analysis of NIF
Trang 5* Mobile phase:
Based on the reference documents and chemicals, solvents available in laboratory, three different mobile phase systems were studied including:
- System 1: MeOH:H2O (65/35, v/v)
- System 2: ACN:H2O (65/35, v/v)
- System 3: ACN:MeOH:H2O (25/25/50, v/v/v)
Table 2: Mobile phase investigation results
Figure 4: Chromatogram of standard sample in mobile phase system MeOH:H2O
(65/35, v/v)
Figure 5: Chromatogram of standard sample in mobile phase system ACN:H2O
(65/35, v/v)
Trang 6Figure 6: Chromatogram of standard sample in mobile phase system ACN:MeOH:H2O
(25/25/50, v/v)
Chromatogram of standard sample in system 2 included ACN:H2O (65/35, v/v) had unbalanced peak with AF = 1.261 and the baseline was not stable That one in system
3 composed of ACN:MeOH:H2O (25/25/50, v/v/v) had balanced peak but retention time was too short (1.050 minutes) Whereas, this one in system 1 including MeOH:H2O (65/35, v/v) had stable baseline, sharp and proportioned peak with AF = 1.126, indicated good separation ability
* Mobile phase ratio:
The mobile phase system which composed of mixture of MeOH and H2O was investigated at different ratio 50/50, 65/35, 60/40, 70/30, 80/20
Table 3: Results of investigation mobile phase ratio
Figure 7: NIF standard chromatogram with mobile phase MeOH:H2O (50:50, v/v)
Trang 7Figure 8: NIF standard chromatogram with mobile phase MeOH:H2O (65:35, v/v)
Figure 9: NIF standard chromatogram with mobile phase MeOH:H2O (60:40, v/v)
Figure 10: NIF standard chromatogram with mobile phase MeOH:H2O (70:30, v/v)
Figure 11: NIF standard chromatogram with mobile phase MeOH:H2O (80:20, v/v)
Trang 8It was found that the mobile phase with the ratio of MeOH:H2O (65:35, v/v) was
selected for quantitative analysis of NIF, because of good separation with a suitable
retention time, compact and balanced peak
2 Method validation results
* System suitability test:
Replicated the injection of a NIF standard solution at concentration of about
150 µg/mL 6 times on HPLC system Evaluated retention time (RT), peak area and
asymmetry factor and theoretical plate (N) in all chromatograms
Table 4: Results of investigation system suitability
The results showed that %RSD for the retention time, peak area, asymmetry factor
and theoretical plate of NIF standard were less than 2.0% The obtained value
demonstrated the suitability of the system for the analysis of NIF
* Specificity:
Prepared 3 samples: blank sample (mixture of excipients), standard sample, test
sample Sample processing and chromatography carried out under selected conditions
Figure 12: Chromatogram of blank sample
Trang 9Figure 13: Chromatogram of NIF standard solution at concentration of 150 µg/mL
Figure 14: Chromatogram of NIF test sample at concentration of 150 µg/mL
Standard sample peak was balanced, retention time was 8.874 minutes On the
chromatogram of the blank sample, no peak was observed in a period of 8 - 10
minutes The chromatograms of both the test sample and the standard sample had
1 peak at approximately 8.8 minutes From there, we concluded that the method was
highly specific
* Calibration curve and linearity range:
Standard solutions were preparedat concentrations from 30 µg/mL to 240 µg/mL
Each concentration was analyzed 3 times Record chromatograms and peak response
The relationship between the area and concentration of NIF was studied
Table 5: Results of calibration curve (n = 3)
Regression equation: y = 83930x - 284700
Trang 10The correlation coefficient R2 = 1 indicated that a linear correlation between the
concentration of NIF and the peak area was obtained in the range of concentrations
investigated This linear range was suitable for quantifying NIF in drug - excipient
compatibility testing samples
* Precision:
- Repeatability:
6 test samples were prepared according to the selected procedure and injected into
the HPLC system (prepared by tester 1) Determined the precision by calculating the
relative standard deviation between the quantifications
Table 6: Results of method precision
Under the selected chromatographic conditions, RSD of the quantitative results in
each day of analysis wss 0.53%, which was less than 2% As such, the chosen method
guarantees accuracy
- Intermediate precision:
6 test samples were prepared by another tester (tester 2) following the selected
procedure and injected them into the HPLC system Determined the intermediate
precision by calculating the relative standard deviation between the quantifications
Table 7: Results of intermediate precision
Under the selected chromatographic conditions, RSD of the quantitative results was
analyzed by each analyst was 0.24% and 0.53%, all of them were less than 2%
Trang 11By both of analysts were 0.3% which was less than 3% As such, the chosen method
guarantees precision
* Accuracy:
The accuracy of the method expressed in % recovery % recovery was determined
by calculating the amounts of standard added into the placebo powder at three different
concentration levels (80%, 100%, 120%) Each level was made in triplicate (n = 3)
Table 8: Results of accuracy test
Recovery
level
Amount of the standard added (mg)
Total peak area (µV.s)
Amount of the standard recovered (mg)
%
80%
Mean = 99.2 RSD = 0.36%
100%
Mean = 98.84 RSD = 0.38%
120%
Mean = 99.68 RSD = 0.59%
The amount of standard recovery at each concentration level was in the range of
98 - 102% compared to the amount of standard added, with the RSD of % recovery at
each concentration level < 2% indicated that the proposed method was highly
accurate
* Specified range:
The accuracy results showed that at the NIF concentration of about 80% to 120%
compared to the quantitative concentration, the quantitation method achieved good accuracy and repeatability Thus, the specified range of the method was
120 - 180 µg/mL
* Determination of limit of determination and limit of quantification:
The limit of detection and limit of quantification were estimated from signal to noise
ratio Limit of detection and limit of quantification were found to be 0.0375 µg/ml and
0.105 µg/mL, respectively
Trang 12CONCLUSION
A HPLC method has been developed
to quantify NIF in drug - excipient
compatibility testing samples with
chromatographic conditions such as:
HPLC system Alliance Waters 2695D;
detector 2998 dual α Absorbance; column
(RP-C18, 5 µm, 100A0, 250 x 4.6 mm)
Phenomenex LUNA; mobile phase
MeOH:H2O (65/35, v/v); flow rate:
1 mL/min; detector PDA 237 nm; injection
volume 25 µL; column temperature: room
temperature
The proposed HPLC method was
validated by evaluating the validation
characteristics such as system suitability,
specificity - selectivity, the linearity range,
repeatability, intermediate precision,
accuracy, range, limit of detection and
limit of quantitation The analytical method
had met the requirements according to
ICH regulations and Circular 32/2018 of the Vietnam Ministry of Health This method can be used to quantify NIF in
drug - excipient compatibility testing samples
REFERENCES
1 Vietnamese Ministry of Health Circular 32/2018, Regulate the registration of drugs and drug materials
2 Uday Y.A et al Estimation of nifedipine
by reverse phase high performance liquid chromatography tablet dosage form IJPLS
2011, pp.610-612
3 Sweetman S.C et al Martindale
Pharmaceutical Press 2014, pp.1447-1455
4 Klaus Florey et al Analytical profiles of
drug substances Academic Press InC 1989, Vol 18, pp.245-253
5 ICH Harmonised Tripartite Guideline
procedures: Q2 (R1) Text and Methodology 2005.