Point level control (using contact and proximity sensors) and continuous level transmission for liquids, granular solids, and liquid–liquid interface Design Pressure Routinely to 1000 PSI (7 MPa), others to 5000 PSI (35 MPa), specialized applications to 20,000 PSI (140 MPa) Design Temperature 500°F (260°C) maximum with insulated sensors; 1000°F (540°C) bare metal, sealed to 200 PSI (30 kPa); 2000°F (1100°C) bare metal at atmospheric pressure Excitation Less than 10 V @ 10 kHz to 1 MHz Wetted Materials Type 316 SS and TFE for common models, with options for CPVC, FEP, PE, PEEK, PFA, PP, PVDF, urethane, Hastelloy, Inconel, Monel, nickel, titanium Span 2 to 3 in. (50 to 75 mm) of insulating liquid to 1000 ft. (300 m) for immersion probes and 0.1 in. (2.5 mm) to 10 in. (250 mm) with proximity sensors
Trang 13.3 Capacitance and Radio Frequency (RF)
Admittance
D S KAYSER (1982) B G LIPTÁK (1969, 1995) J B ROEDE (2003)
Service Point level control (using contact and proximity sensors) and continuous level
trans-mission for liquids, granular solids, and liquid–liquid interface
Design Pressure Routinely to 1000 PSI (7 MPa), others to 5000 PSI (35 MPa), specialized applications
to 20,000 PSI (140 MPa)
Design Temperature 500 ° F (260 ° C) maximum with insulated sensors; 1000 ° F (540 ° C) bare metal, sealed
to 200 PSI (30 kPa); 2000 ° F (1100 ° C) bare metal at atmospheric pressure
Excitation Less than 10 V @ 10 kHz to 1 MHz
Wetted Materials Type 316 SS and TFE for common models, with options for CPVC, FEP, PE, PEEK,
PFA, PP, PVDF, urethane, Hastelloy, Inconel, Monel, nickel, titanium
and 0.1 in (2.5 mm) to 10 in (250 mm) with proximity sensors
Inaccuracy Horizontal, less than the diameter of the probe rod
Vertical, less than 0.1 in (2.5 mm) for bare single points in conducting material, roughly 1% of maximum active length for all insulated probes in conducting or interface service, 3% in insulating liquids (or 0.5% with dielectric compensation), roughly 5% for granular insulating solids with constant density and composition, and
2 to 5% for conducting granular solids
Dead Band Unmeasurable with analog instruments and horizontal probes; dependent on A/D
resolution with digital continuous instruments; small and application dependent on vertical single points, but optional dead band adjustment is available
Temperature Coefficient Extremely variable depending on (a) probe insulation and degree of probe-to-sheath
bonding in conducting materials, and (b) composition and density variation in insu-lating liquids
Damping and Time Delay Adjustable time delay of 0 to 30 sec is included on most single-point controls;
adjustable time constants up to 30 sec are available on most analog transmitters, and digital instruments offer zero to several minutes
Cost $200 to $800 for single-point controls; $500 to 1500 for two-wire level transmitters;
all with type 316 SS and TFE wetted parts; increased cost with exotic metals, hermetic seals, flange mounting, longer insertion length, digital output, dielectric compensa-tion, extended press and temp, and longer inactives
Vendors (partial list) ABB Process Automation Instrumentation Div ( www.abb.com/us )
Arjay Engineering Ltd ( www.arjayeng.com ) Babbitt International Inc ( www.babbittlevel.com ) Bindicator ( www.bindicator.com )
LT
CA To Continuous Receiver
To On-off Receiver CA
Flow Sheet Symbol
LS
Trang 23.3 Capacitance and Radio Frequency (RF) Admittance 431
BinMaster ( www.binmaster.com ) Delavan Inc.
Delta Controls Corp ( www.deltacnt.com ) Endress + Hauser Inc ( www.us.endress.com ) FMC Invalco ( www.fmcinvalco.com ) GLI International ( www.gliint.com ) HiTech Technologies Inc ( www.hitechtech.com ) K-Tek Corp ( www.ktekcorp.com )
Lumenite Control Technology Inc ( www.lumenite.com ) Magnetrol International ( www.magnetrol.com )
Monitor Technologies LLC ( www.monitortech.com ) Monitrol Manufacturing Co ( www.monitrolmfg.com ) Omega Engineering Inc ( www.omega.com )
Penberthy ( www.penberthy-online.com ) Princo Instruments Inc ( www.princoinstruments.com ) Robertshaw Industrial Products Div., an Invensis Co ( www.robertshawindustrial.com ) Scientific Technology Inc ( www.automationsensors.com )
Systematic Controls ( www.systematiccontrols.com ) Vega Messtechnik AG ( www.vega-g.de )
INTRODUCTION
Characteristic of probe-type sensors, the RF probes operate
by applying a constant voltage to a metallic rod and
moni-toring the current that flows This current is proportional to
the admittance or capacitance (if conductivity is absent)
from the metallic rod to a second electrode Because the
tank wall is the most convenient second electrode, most
instruments monitor current to ground (which is usually
connected through the probe mounting) The obvious
differ-ence between conductance and RF probes is the frequency
of that constant voltage Whereas conductance types use DC
or low-frequency AC, the RF items usually operate in the
range of 0.1 to 1.0 MHz (although special applications can
operate at 15 kHz or even lower frequencies) RF probes are
connected to their associated electronic units with coaxial
cable except when the electronics are integrally mounted on
the head of the probe
The classic shortcoming of capacitance probes is false
HI level indications caused by conductive coatings that
con-nect the above-level sensing element to ground (or the actual
process level) Since 1970, solutions to this problem have
existed in all but the heaviest, high-conductance process
sit-uations In the case of single-point level switches, the answer
is electrogeometric In the continuous level transmitters, the
approach is purely electrical
Because of the solid, no-moving-parts construction, there
is very little to deteriorate or fail once an RF probe is installed
Compatibility with process liquids is the most obvious obstacle
to a satisfactory life span This is no problem with common,
activity and accelerated permeation experienced by
poly-mers can produce unexpected results Abrasion of metals
and insulators is another cause of shortened life that must
be anticipated Baffles to protect the sensor from high-velocity
solids, combined with judicious location, can minimize this danger
Probe failure in heavily agitated tanks can be avoided by attention to structural considerations In the case of rigid probes, the most likely cause of breakage is fatigue failure caused by eddies rapidly pushing the probe in one direction and then in the opposite A support, with an insulated bush-ing, near the tip of the probe greatly reduces the possibility
of such a failure Flexible cable types can wrap around an agitator and fail in minutes if not adequately anchored to the tank structure If they are anchored without removing slack, they can whip back and forth, causing insulation failure and eventual breakage Intermediate supports are possible in highly agitated service using insulated bushings Beware of thick, conductive coatings that can cause substantial errors
at each support point These “shortcuts” to ground can defeat the electrical coating rejection in the worst cases Correct structural design is the responsibility of the system designer, not the probe supplier Most suppliers can give rules of thumb for their probes and provide structural details of the construc-tion It may require a consultant who is skilled in fluid dynam-ics and mechandynam-ics to arrive at a sure configuration in a highly agitated vessel
Process instruments that depend on electrical character-istics of the process material are at a disadvantage, given that the electrical character is of little interest to most instrument users Fortunately, exact values of conductivity (g) are never required, and changes in relative dielectric constant (K) are more important than the precise value In most cases, classi-fication as conducting or insulating goes a long way toward successful application Within the conducting category, it is sufficient to know that, except for completely deionized water, aqueous solutions will be conductive On the insulating side, it is generally sufficient to understand that most liquids will have a K of 2 or greater The main exceptions are liquids that would be gases at room temperature (with the notable
Trang 3432 Level Measurement
except for ammonia, for which K> 15) and atmospheric
pres-sure Not only do these fluids tend to have K less than 2, they
tend to have much higher temperature coefficients of K than
insulators that are normally liquids
This section is divided into single-point and continuous
transmitter categories to reduce confusion The RF sensors
are somewhat unique in that single-point switches have a
substantial performance advantage over the continuous type
in terms of accuracy, temperature capability, coating
rejec-tion, self-checking, and reliability It seems that, in many
cases, the transmitters are thought to be the superior approach
to control, and they obviously are necessary to obtain a
pro-portional band In many cases, strategically placed
single-point units are a superior route to precise and reliable process
control Considering that any instrument can fail on occasion,
single-point probes provide highly reliable backup, with
self-test capability, and are superior to any transmitter
TYPES OF PROBES
The most basic probe configuration is a metal rod The rod
is insulated from a metal mounting element that connects it
to the process vessel via threads into a half coupling or a flange
that mates with one on a tank nozzle The insulated junction
of rod and mounting includes whatever seal is required
between the process and outside world The next step in
com-plexity includes an insulating coating on the rod, (Figure 3.3a)
that isolates it chemically as well as electrically from the
process Insulated probes should have the insulation securely
bonded to the metal rod over the entire range of service
tem-perature This bonding ensures that process pressure changes
will not compress air space and change the calibration Bond-ing is also important to minimize permeation, which is present to some slight degree with all polymers The addition
of a tight-fitting metal tube over part of the insulation, welded
to the mounting element (Figure 3.3b), will make the covered section of the probe “inactive,” because it will always see the same impedance to ground, regardless of the process material
on the outside A more sophisticated way to “inactivate” a
guard principal A tight-fitting metal tube, insulated both from the rod and the mounting element (Figure 3.3c), with a voltage identical to that on the rod, will not only deactivate that section of the probe but will also negate the rod to mounting capacitance These are sometimes referred to as
three-terminal probes, because there is now a rod connection,
a ground connection, and a guard (sometimes called a shield)
style connections An additional variation is the probe that carries its own intrinsic ground reference This can be a larger
3.3d), with bleed holes at the top to avoid compressing gas
as the liquid rises in a closed chamber Perforated tubes, insulated and bare ground rods, as well as structural cages (Figure 3.3e) are also available for various conditions of agitation, temperature, and chemical compatibility The sen-sors, which use a ground wire wrapped in a helix directly
on the probe insulation, are unstable, unreliable, and facilitate coating
A three-terminal probe, with a plate welded on for greater
sensor The field from the guard can even be used to direct
or focus the field from the sensing element and determine the region of sensitivity In proximity measurements, the
FIG 3.3a
Insulated two-terminal probe.
Center Rod Connection Process Seal
Mounting Flange
Insertion
Length
Metal Rod
Polymer Insulation
Polymer Plug
FIG 3.3b
An insulated, two-terminal probe with grounded inactive section.
Center Rod Connection Process Seal
Mounting Flange Inactive
Length
Insertion Length
Metal Rod Polymer Insulation
Grounded Inactive Section
Polymer Plug Weld
Trang 43.3 Capacitance and Radio Frequency (RF) Admittance 433
capacitance change (proportional to sensor area) is typically
very small but stable, because it is effectively looking at the
air capacitance between sensor and process material Even
minute capacitance changes due to temperature or stress in
the mounting area could be catastrophic with an unguarded
probe The distance between the probe and process material
should be as short as feasible The capacitance produced is
inversely proportional to this distance This means that beyond
10 in (250 mm), the capacitance change becomes extremely
hard to detect The error produced by splashing and
conden-sation buildup on the plate is fairly benign—never more than
the actual coating thickness, and usually less Spans of 2 or
3 picofarads (pF) have been used successfully Such small ranges are the result of level change in shallow pans, with limited space for proximity plates
The electronic guard principle can be employed in an infinite variety of geometries A typical example is the flat plate probe (Figure 3.3g), which mounts flush with the tank wall This is very effective in reducing abrasion, eliminating bending of low-level sensors under heavy mechanical loads, and eliminating sparks from static discharge The monitored current flows from the center plate, and the guard surrounds
FIG 3.3c
A three-terminal probe that employs the “electrical guard.”
FIG 3.3d
A two-terminal probe with intrinsic, concentric ground reference.
Center Rod Connection Guard Connection Process Seal
Mounting Flange
Insertion
Length
Guard Element Polymer Insulation
Polymer Insulation
Metal Rod
Guard Length
Center Rod Connection Process Seal
Mounting Flange
Insertion
Length
Metal Rod Polymer Insulation
Concentric Ground Tube
Weld Bleed
Holes
FIG 3.3e
A two-terminal probe with a cage, for grounding in viscous liquids.
FIG 3.3f
A three-terminal probe with plate for proximity sensing.
Center Rod Connection Process Seal Mounting Flange
Insertion Length
Polymer Insulation
Grounding Cage Welds
Polymer Plug
Metal Rod
Center Rod Connection Guard Connection Process Seal
Mounting Flange
Guard Element
Proximity Plate
Polymer Insulation
Polymer Insulation
Trang 5434 Level Measurement
it, interrupting a path to ground through any coating on the
face of the probe
For the ultimate in seal reliability and hermeticity, plastic
welding offers unique benefits By plastic welding a polymer
flange facing to the probe insulation (Figure 3.3h), a
high-pressure seal is formed (and has been used to 5000 PSI [35
MPa] with the correct flange rating), with the normal process
seal acting as a backup This type of weld is not possible
using TFE, which does not melt, but most other polymers
are candidates It also precludes the use of thin probe
insu-lation The flange facing should be used as the only gasket,
because it must be pinched between the flange faces Only
raised face flanges provide the correct sealing, so RTJ and
flat-faced flanges are not applicable
MOUNTING AND TANK ENTRY
One of the most frustrating barriers to good RF probe
appli-cation is inadequate or misplaced tank access on existing
tanks Often, adding a nozzle or even a half-coupling is
impossible, because the tank is pressure coded, requires ardu-ous inerting, or cannot be taken out of service One very simple-minded and geometrically demanding approach is to use a bent probe inserted through a side entry so as to get its active area to the desired measurement location A flange mounting
is usually mandatory to avoid “screwing,” which would cause the bent probe to rotate inside the tank like an airplane pro-peller Its angular location, once the thread is tightened, would also be a question If a suitable nozzle is available, it must be short enough to get the bend in the probe “around the corner” before it binds If the measuring leg of the probe
is relatively long, it must not hit the opposite wall of the vessel before the bend allows it to pivot Another drawback
to the bent probe is the requirement that any insulation be flexible enough to tolerate the bending process without tearing or splitting This limits the selection to soft, thick insulation and therefore relatively low capacitance (60 to
100 pF/ft)
Many strategies allow us to avoid the bent-probe trap by using a little bit of imagination Every tank or bin should have a vent or pressure relief pipe It is usually possible to tee that pipe so that the vent or relief goes off to the side and
a probe has straight access into the process (Figure 3.3i) This does not affect the pressure code of the vessel, since it is external plumbing Other ways to avoid the “pretzel probe” include the following:
• Use a grounded inactive section to get through various impediments
• “Dog-leg” a pressure gauge to give the probe a straight shot
• Cut a hole in the building’s roof or ceiling in lieu of headroom
FIG 3.3g
A three-terminal flush mounting probe.
FIG 3.3h
Probe with plastic-welded process seal (Courtesy
AMETEK-Drexelbrook.)
Sensing Plate
Polymer Insulator Mounting Holes
Guard Element Gasket Surface
Connection Enclosure
Hermetic
Facing
Probe Insulation
FIG 3.3i
Vent pipe serving dual function for tank access.
Electronics or Connection Enclosure
Vent Added Tee
Probe Inactive
Section
Trang 63.3 Capacitance and Radio Frequency (RF) Admittance 435
col-lector or a side of the tank
• Add a stand pipe (sidearm) parallel to the tank (Figure
3.3j)
• Tee into a fill pipe (this needs detailed application
anal-ysis)
Ingenuity will always trump bent probes for performance,
cost, and efficiency
ELECTRONIC UNITS
Electronic units for single-point instruments are available in
line-powered configurations for 24 VDC; 120 V, 50 to 60 Hz
AC; and 230 V, 50 to 60 Hz AC They also can be obtained
with “universal power” capability that allows them to use any
of these plus 130 VDC The output from these units is
gen-erally a set of double-pole, double-throw relay contacts Some
of the DC-powered instruments use an NPN or PNP output
transistor to effect the switching There are also two-wire,
loop-powered versions that offer intrinsic safety and
auto-matic self-checking The signal from these instruments is a
high or low current within the 4- to 20-mA range Various
types of automatic calibration are available for these
instru-ments, but they all encounter certain conditions that prevent
them from being calibrated properly under every possible
condition Regardless, the calibration of single-point
instru-ments is hardly rocket science
Level transmitters are primarily loop powered, although line-powered items are available from some suppliers The traditional analog 4- to 20-mA instruments have been giving way to those with microprocessors on board Most of the digital instruments are capable of communication using one
of the HART, Honeywell, or fieldbus protocols This allows interrogation and modification of the instrument by means
of a digital communicator One of the most popular features
of these instruments is their ability to calibrate on any two points in the range It is possible to enter the correct output (in milliamps or level units) at an existing low level and enter
a high point days later The instrument locks the input/output curve into those two points The signal from the instrument
is either analog 4- to 20-mA or digital The digital mode allows more than one transmitter to use a single loop The limited power available at the transmitter, combined with the multi-tude of calculations being executed, causes digital instru-ments to be considerably slower than the analog ones Response times of 3 to 4 sec are possible, whereas analog instruments are capable of responses within 100 to 300 ms For small tanks with high fill and drain rates, the digital instru-ments might not be an option
A hybrid instrument is the “multipoint” control It is essentially an analog instrument with internal, adjustable pick-offs and multiple relay outputs It is useful for sump control where several pumps might be involved, but the absence of an analog output makes calibration lengthy Each pick-off must be adjusted with the level at the desired point
on the probe This type of instrument is specified for HI, HI-HI,
FIG 3.3j
Side arm (cage) mounting, when there’s no other way (Courtesy of Robertshaw Controls Co.)
Probe
Cage Isolation Valves
Trang 7436 Level Measurement
and HI-HI-HI points at times This doesn’t make much sense,
because one failure kills the whole operation
SINGLE-POINT SWITCHES
The typical capacitance switch employs a vertical or
hori-zontal probe, either polymer insulated or bare metal,
project-ing some distance into the process vessel with the center rod
insulated from the mounting (Figure 3.3k) With any gas or
current flowing from the center rod to the metallic tank wall
When a liquid or granular solid covers the center rod, the
current from there to ground will increase This change in
current is detected by the electronic unit, which switches the
output When the process material drops below the probe,
the RF current decreases and, hopefully, returns to its
previ-ous low-level state If the process material is thick and
con-ductive, any coating remaining on the insulator between
probe and mounting will maintain a higher RF current level
and can cause the switch to continue signaling a high level
The answer to false high-level indications is based on the
classic electronic guard principle By interposing a third
metal-lic element (shield) with an identical RF voltage between the
flow from center rod to mounting The shield element
sup-plies whatever RF current the coating demands, but this
cur-rent is not included in the measurement “Length is strength”
in capacitance probes Longer active lengths translate to more
substantial capacitance changes in insulating processes, and
longer interelement insulators allow the instrument to reject
higher conductance coatings This means using horizontal
probes for switch points close to the bottom and top of a tank rather than a vertical 2 or 3 in (50 or 75 mm)
Conducting Process Materials
Conducting materials (aqueous, metals, and most forms of carbon) carry the ground potential of the tank walls right to the probe If bare metal, the probe will signal “high level” the instant it contacts the process If it is insulated, the switch-ing point will depend on the capacitance settswitch-ing of the instru-ment With a vertical, insulated probe, it is possible to adjust the level at which the switching takes place by varying the capacitance setting of the instrument A horizontal probe of either type allows no level adjustment other than by relocat-ing the mountrelocat-ing
Coating is a serious consideration in all conducting mate-rials, and the guard-type construction is usually required With bare metallic probes mounted vertically, the plain capacitance probe will do as long as the process never reaches the mount-ing area and no crystallization or heavy condensation occurs When using the guarded probes, the guard should be long enough to project well into the vessel, beyond nozzles and potential wall buildup
Insulating Process Materials
The insulating materials (oils, solvents, resins) are subtler in their effect on the switching circuit A horizontal probe pro-duces a sharp capacitance change over the thickness of the center rod Any portion that is in a nozzle should be inactive, either with the guard or a grounded inactive With a vertical probe, capacitance increases gradually as more of the active length is covered The switching point is adjustable even if the probe is bare metal Good practice requires that the probe
be at least 2 in (50 mm) into the process material at the desired switching point Attempting to get switching at the tip of the probe will lead to unreliable performance, because the slightest change in probe mounting or electronics can cause a constant high-level signal
Dielectric constant (K) is relative to the absolute dielec-tric of a vacuum (K of gases barely differs) This means that the capacitance (and the proportional RF current) in air will
be doubled in gasoline (K = 2) It will be multiplied by 20
in ethanol (K= 20) It will only be raised by 60% in liquid carbon dioxide (K= 1.6) Insulating coatings are a very minor problem Just avoid bridging to a grounded part
Plastic, Concrete, or Fiberglass Tanks and Lined Metal
The absence of metallic contact for a ground reference is seldom a problem using single-point RF probes in conducting process media The probe will indicate high level whenever
it touches the process Unlike metallic tanks (or those with metal pipes, pumps, or grounding rods), it must be tuned to
a value less than the capacitance-to-ground value of the tank Most tanks have at least 10 pF to ground, which is a perfectly adequate level for a reliable measurement Probes with a
FIG 3.3k
Bare metal, two-terminal probe.
Center Rod Connection Process Seal
Mounting Flange
Insertion
Length
Metal Rod Polymer Insulation
Trang 83.3 Capacitance and Radio Frequency (RF) Admittance 437
guard element should not be mounted horizontally in
ungrounded tanks with conductive contents The guard can
drive the entire process at the same voltage as the center rod,
hence there is no current flow and no high-level signal
In the case of an insulating medium in a fiberglass tank,
even metal pipes offer little help to the miniscule
capaci-tances The only sure answer is a vertically mounted probe
with its own metal ground rods or concentric tube This is
also a good precaution against RFI from walkie-talkies and
other sources, which can cause false high-level signals
Metal tanks that are rubber or plastic lined and
steel-reinforced concrete vessels, on the other hand, represent
excellent RF grounds The capacitance from process to metal
structure is very high as compared with the capacitance
pro-duced by an insulating medium This makes it look like a
short circuit to the RF current It is excellent for conducting
media, too, but it still requires the avoidance of horizontal
probes with driven guards Underground concrete or
fiber-glass tanks (except in desert conditions) present a similar
ground configuration
Concrete structures of cinder block with no vertical steel
reinforcing bars are completely ungrounded and perform
exactly the same as fiberglass tanks
Interface
Electrical sensing is the premier method of detecting the
interface between an insulating and a conducting process
medium The typical margin between the conductivity of
organic and aqueous phases is greater than 1000:1 The
mea-surement is completely independent of temperature and
den-sity variation Probes may be mounted vertically or
horizon-tally Vertical probes should be inactive down to about 6 in
above the desired interface control point This can be
accom-plished by use of the electronic guard, a grounded inactive,
or a short probe mounted from the rear (rear mount) on the
end of a suitable pipe The instrument will indicate high level
as soon as conductive material contacts the tip of the bare
probe In the case of heavy oil separators, it may be desirable
to use a probe with sharp edges machined into the tip
(Figure 3.3l) This will ensure good contact between water
and steel in spite of oil coating An interface detector with
bare metal rod should be tuned to the maximum capacitance
level of which it is capable
Horizontal probes should be bare metal with relatively
long insulators between probe and mounting (ground) Typical
proportions, to maximize the margin between insulating and
conducting phases, would be 12 in (300 mm) overall length
with a 10-in (250-mm) long insulator The use of sharp edges
machined on the tip of the probe is also advantageous in this
orientation This is one situation in which the electrical guard
is of no advantage (coatings are usually insulating rather than
conducting) and can actually be a drawback if the guard drives
the entire conducting phase at the same potential as the probe
This phenomenon is usually a product of insufficient
ground-ing, but it has been observed in metal tanks
GRANULAR SOLIDS
The best approach for establishing the high level in large silos is usually via a flexible cable probe with a long, thin weight at the bottom It can be mounted close to the fill point (Figure 3.3m) and negate most of the angle of repose that side-mounted sensors encounter The flexible aspect allows
it to swing out of the way when struck by incoming solids The actual switching should take place on the weight that will be least vulnerable to abrasion In insulating materials,
at least 6 in of weight should be covered at the desired switching point Conducting materials, of course, will switch
as soon as they touch the tip of the probe Incoming material will not cause false high-level indication as long as it is in free fall This is because there is a very small air capacitor between each grain and all the others All this series air capacitance means that any deviation from normal air capac-itance is a negligible quantity If the incoming material is compressed (especially conducting materials), it IS possible that it could cause a trip, so the probe should be located accordingly Rigid probes, located near the fill, are suitable for smaller bins and lighter duty
FIG 3.3l
Oil shedding tip for interface with viscous organic phase.
FIG 3.3m
High-level probe for heavy granulars in large silos.
Metal Rod
Sharp Edges
Process Granular
Electronics or Connection Enclosure Half Coupling
Fill
Top of Silo
Insertion Length
Cable
Weight
Trang 9438 Level Measurement
Horizontal probes mounted in the bin wall should be of
the “guarded” variety to avoid false high-level alarms—
especially for low-level service The primary concern is
bend-ing or abrasion, which becomes more acute as the depth and
density of the material increases Short, fat probes have much
better lifespan, and it is possible to use a flat-plate probe
(Figure 3.3g) mounted flush with the wall for optimal life
CONTINUOUS TRANSMITTERS
Probes are most commonly mounted vertically from the top
of the tank Angle mounting through the side of a tank is
possible, as is mounting up from the bottom When tank
penetrations are at a premium, it is often possible to “tee”
into a vent or drain line for probe mounting There are even
cases in which probes have been successfully “teed” into fill
nozzles, but this requires considerable application analysis
to predict the effect of incoming material The insertion
length of the probe should not extend beyond the desired
measuring range by more than 5% An active probe that is
not producing signal can still produce errors
Conducting Liquids
The main concept required to understand this class of
appli-cations is called saturation Assume that an insulated probe
is immersed in deionized water with minimal conductivity
(Figure 3.3n) The addition of drops of hydrochloric acid to the vessel will gradually increase the conductivity The output will also rise as more RF current flows from probe to ground Eventually, the conductivity will be high enough that the resistance (R) from probe to tank wall is negligible compared
to the capacitive impedance of the probe insulation (C5) Further increases in conductivity will make no observable difference in the RF current and, therefore, the output (i.e., the RF current has reached its saturation point) This is the concept that makes the RF transmitter an instrument rather than a lab curiosity A probe that is not saturated will have
a calibration that is a function of two variables (conductivity and level), just as a d/p transmitter calibration is a function
of density and level It is possible to adjust the threshold of saturation by adjusting insulation thickness (hence capaci-tance) and excitation frequency Raising the impedance of the insulation lowers the threshold, and vice versa
accom-plished A reasonable question might be, “Why not just use the 0.1µS/cm threshold at all times?” The answer is “coating rejection.”
The higher impedance that correlates with a low saturation threshold is less able to ignore high-conductivity coatings
In fact, a large mismatch may allow the coating to saturate the probe so that the output reflects the actual level plus coating length In general, the path to best conducting coating rejection lies in the highest probe capacitance and excitation frequency that will allow saturation by the process material
FIG 3.3n
Electrical representation of an insulated probe in conductive liquid (Courtesy of The Foxboro Co.)
L
I
A1
A2 B
Ce
Ce
C2 C4
C2+ C 4
C3 C5
C3+ C 5
C1
C4
C1
C2
C2
C4
C3
C5
Ka
Kp
R
R = Any Value
Trang 103.3 Capacitance and Radio Frequency (RF) Admittance 439
being measured A special case for coating rejection is one
in which the process liquid is conductive but can dry on the
probe, forming an impervious, insulating coating An
exam-ple is latex paint The liquid is quite conductive, but the dried
paint adds to the insulation thickness of the probe, decreasing
its capacitance and changing the calibration The solution is
to use a thick (low-capacitance) probe insulation so that thin
additions will have negligible effect on calibration By using
the highest possible frequency, conductive coating rejection
will also be maximized, and the resulting accuracy will be
the best possible compromise
mea-surement allows the electronic unit to measure two variables:
the capacitive component of RF current and the resistive
component of RF current The actual liquid level that
satu-rates the probe produces a pure capacitive current The
con-ductive coating, because it is relatively thin, has a much
higher resistance than the bulk liquid, so it will produce both
capacitive and resistive phase current By subtracting the
resistive component from the capacitive, the effect of
con-ductive coating on the output will be decreased In fact, once
the coating is longer than a nominal value, the two RF current
phases equate, and the effect of coating is precisely zero
Maximum coating error occurs when the coating length is
relatively short and its resistance therefore is low The
max-imum error is a fixed, predictable number of inches for a
given probe capacitance, excitation frequency, coating
thick-ness, and conductivity This means that percent inaccuracy
due to coating will be greater on short ranges than on long
ones In other words, length is strength
Insulating Liquids
Probes for measuring insulating liquids should generally
include their own parallel ground reference to guarantee a
uniform distance to ground through the process liquid
Lin-earity, sensitivity, and immunity to RFI are enhanced by a
concentric ground tube (Figure 3.3d) or other construction
In some cases, it is efficacious to use a metallic tank wall,
baffle, or ladder to furnish the required parallel ground
ref-erence This is frequently the case in pharmaceutical or
bev-erage applications where a concentric tube interferes with
any clean-in-place function It is also common in tall tanks
that require flexible cable sensors and with slurries, which
can accumulate solids and plug the ground reference
Changes in dielectric constant will cause a change in
calibration If composition is constant, temperature will be
the only concern for variation of K Because K is proportional
to the number of molecules between probe and ground,
out-put (as with a d/p transmitter) will be proportional to density
and, hence, the weight of process material For cases in which
constant composition and a reasonably narrow temperature
range are not possible, a transmitter with dielectric
compen-sation is available By using such an instrument, it is possible
to obtain a level signal that is independent of dielectric, and
even conductivity, when tanks are not dedicated This tech-nology demands two conditions:
1 A short inactive section at the bottom of the probe, approximately 6 in (150 mm)
2 Homogeneity of the process liquid (no stratification) The compensation is accomplished by making two inde-pendent measurements The first measurement is made with
a short sensor (composition probe) below the tip of the level probe (Figure 3.3o) This segment is assumed always to be covered by the process liquid Its capacitance is proportional
to the electrical character of the process material The second measurement uses the level probe Its output is proportional
to the level times the electrical character of the process fluid
By dividing the output of the level probe by the output of the composition probe, the transmitter output becomes inde-pendent of the electrical properties Please note that conduc-tive coatings can cause large inaccuracies in these particular transmitters Even so, conducting liquids that do not coat will
be measured with accuracy equal to the insulating ones For the sake of utility, both probe segments are usually provided, coaxially, in the same assembly It is also possible to use two completely separate probes, with the composition probe mounted horizontally below the tip of the level probe Inac-curacy can be limited to 0.5% over a wide range of electrical values and a wide temperature range
Continuous Liquid–Liquid Interface
Two elements are key to successful interface level measure-ment: immunity to total level variation and maximum margin between the signal contribution of the organic and aqueous
FIG 3.3o
A probe with a second element for dielectric compensation (Cour-tesy of AMETEK-Drexelbrook.)
Concentric Shield (Ground Reference)
Reference (Composition) Sensor
Insertion Length (I.L.)
5.5"
Measured Level Level Sensor