Off-line calibration of the zero and span of measurement was the topic of the previous section. In this section, the on-line methods of response time determination and calibration verification will be described for sensors that have already been installed in operating processes. As in Section 1.8, the discussion here will also focus on temperature and pressure sensors. FUNDAMENTALS OF RESPONSE TIME TESTING The response time of an instrument is measured by applying a dynamic input to it and recording the resulting output. The recording is then analyzed to measure the response time of the instrument. The type of analysis is a function of both the type of instrument under test and on the type of dynamic input applied, which can be a step, a ramp, a sine wave, or even just random noise. The terminology used in connection with time response to a step change was defined in Figure 1.3z. The time constant (T ) of a first-order system was defined as the time required for the output to complete 63.2% of the total rise (or decay) resulting from a step change in the input. Figures 1.9a and 1.9b show the responses of instruments to both step changes and ramps in their inputs and identify the time constant (T) and response times (τ) of these instruments. As shown in Figure 1.9a, the time constant of an instrument that responds as a first-order system equals its response time and it is determined by measuring, after a step change in the input, the time it takes for the output to reach 63.2% of its final value. The response of a first-order system is mathematically described by a first-order differential equation, c(t) = K(1 – e–t/τ) 1.9(1) where c = output t = time K = gain τ = time constant of the instrument The 63.2% mentioned earlier is obtained from this equation by calculating the output when the time equaling the time constant (t = τ) has passed. c(τ ) = K(1 − e−1) = 0.632 K 1.9(2) Although most instruments are not first-order systems, their response time is often determined as if they were, and as if their response time were synonymous with their time constant. However, if the system is of higher than first order, there is a time constant for each first-order component in the system. In spite of this, in the field, the definition of the firstorder time constant is often also used in connection with higher-order systems. The ramp response time is the time interval by which the output lags the input when both are changing at a constant rate. For a ramp input, the response time (τ) is defined as the delay shown in Figure 1.9b. This is also referred to as ramp time delay and can be measured after the initial transient, when the output response has become parallel with the input ramp signal. For a first-order system, the ramp time delay, response time, and time constant are synonymous. The ramp time delay can be mathematically described as c(t) = C(t − τ) 1.9(3) FIG. 1.9a Illustration of step response and calculation of time constant. Input Output Sensor Response τ Time 63.2% of A
Trang 11.9 Response Time and Drift Testing
H M HASHEMIAN (2003)
Off-line calibration of the zero and span of measurement was
the topic of the previous section In this section, the on-line
methods of response time determination and calibration
ver-ification will be described for sensors that have already been
installed in operating processes As in Section 1.8, the
discus-sion here will also focus on temperature and pressure sensors
FUNDAMENTALS OF RESPONSE TIME TESTING
The response time of an instrument is measured by applying
a dynamic input to it and recording the resulting output The
recording is then analyzed to measure the response time of
the instrument The type of analysis is a function of both the
type of instrument under test and on the type of dynamic
input applied, which can be a step, a ramp, a sine wave, or
even just random noise
The terminology used in connection with time response
to a step change was defined in Figure 1.3z The time constant
(T) of a first-order system was defined as the time required
for the output to complete 63.2% of the total rise (or decay)
resulting from a step change in the input Figures 1.9a and
1.9b show the responses of instruments to both step changes
and ramps in their inputs and identify the time constant (T)
and response times (τ) of these instruments
As shown in Figure 1.9a, the time constant of an instru-ment that responds as a first-order system equals its response
time and it is determined by measuring, after a step change in
the input, the time it takes for the output to reach 63.2% of its
final value The response of a first-order system is
mathemat-ically described by a first-order differential equation,
where
c= output
t= time
K= gain
τ= time constant of the instrument The 63.2% mentioned earlier is obtained from this equa-tion by calculating the output when the time equaling the
time constant (t=τ) has passed
c(τ)=K(1 −e−1) = 0.632 K 1.9(2)
Although most instruments are not first-order systems, their response time is often determined as if they were, and as if their response time were synonymous with their time con-stant However, if the system is of higher than first order, there is a time constant for each first-order component in the system In spite of this, in the field, the definition of the first-order time constant is often also used in connection with higher-order systems
The ramp response time is the time interval by which the output lags the input when both are changing at a constant rate For a ramp input, the response time (τ) is defined as the delay shown in Figure 1.9b This is also referred to as ramp time delay and can be measured after the initial transient, when the output response has become parallel with the input ramp signal For a first-order system, the ramp time delay, response time, and time constant are synonymous The ramp time delay can be mathematically described as
c(t) =C(t −τ) 1.9(3)
FIG 1.9a
Illustration of step response and calculation of time constant.
Sensor
63.2% of A
A
Trang 21.9 Response Time and Drift Testing 115
where C is the ramp rate of the input signal The derivations
of Equations 1.9(1) through 1.9(3) and the topic of Laplace
transformation is covered in the second volume of the
Instru-ment Engineers’ Handbook and also in Reference 1
LABORATORY TESTING
The response time of temperature sensors is measured by using
a step input, whereas the response time of pressure sensors is
usually detected by using ramp input signals This is because
obtaining a step change in temperature is easier and more
repeatable than obtaining a step change in pressure Ramp
inputs are also preferred for the testing of pressure sensors,
because a step input can cause oscillation of the pressure
trans-mitter output, which may complicate the measurement
Testing of Temperature Sensors
Figure 1.9c illustrates the equipment used in determining the
response time of a temperature sensor This experiment is
called the plunge test At the beginning of the test, the sensor
is held by a hydraulic plunger, and its output is connected to
a recorder The heated sensor is then plunged into a tank of
water at near-ambient temperature This step change in
tem-perature determines the type of transient in its output, as was
illustrated in Figure 1.9a To identify the response time of
the temperature sensor, the time corresponding to 63.2% of
the full response is measured
Because the response time of a temperature sensor is a function of the type, flow rate, and temperature of the media
in which the test is performed, the American Society for Testing and Material (ASTM) has developed Standard E644 (Reference 2), which specifies a standard plunge test This document specifies that a plunge test should be performed in water that is at near room temperature and is flowing at a velocity of 3 ft/sec (1 m/sec) A plunge test can therefore be performed by heating the sensor and then plunging it into a rotating tank that contains water at room temperature By controlling the speed and the radial position of the sensor, the desired water velocity can be obtained for the plunge test There can be other ways for performing the plunge test For example, the sensor can be at room temperature and plunged into warm water Although the actual temperatures have an effect on response time, this effect is usually small; therefore, the response time is not significantly different if the water is at a few degrees above or below room temperature
Testing of Pressure Sensors
The response time of pressure sensors is usually determined
by using hydraulic ramp generators, which produce the ramp test input signals A photograph of a hydraulic ramp generator
is provided in Figure 1.9d This equipment consists of two pressure bottles, one bottle filled with gas or air and the other
FIG 1.9b
Illustration of ramp response and calculation of ramp time delay.
Sensor
Time
Response Time
Input Output
τ
FIG 1.9c
Plunge test setup.
Channel 2
Channel 1
Response Time
Timing Signal
Test Transient
Trigger
Signal Conditioning Sensor Hot Air Blower
Rotating Tank Timing Probe
Multimeter
Data Recorder
ω
0.632 × AΑ Water
r
Trang 3116 General Considerations
with water, as shown in Figure 1.9e In the outlet from the
gas bottle, an on–off and a throttling valve is provided The
setting of the adjustable valve determines the flow rate of the
gas into the water bottle Therefore, the desired ramp pressure
rate can be generated by adjusting the throttling valve
The water pressure is detected simultaneously by two
sensors, a high-speed reference sensor and the sensor under
test, as shown in Figure 1.9f The outputs of the two sensors
are recorded on a two-pen recorder, and the time difference
(delay) between the two outputs is measured as the response
time of the sensor being tested This delay time measurement
is taken after the pressure in the water bottle has reached a
predetermined setpoint or after the input and output curves
have become parallel
The pressure setpoint is based on the requirements of the
process where the sensor is going to be used For example, if
a full process shutdown is initiated, and if the pressure exceeds
a certain upper limit, then this pressure is likely to be used as
the setpoint pressure at which the response time of the
pres-sure sensor is meapres-sured When testing differential-prespres-sure sensors (serving the measurement of level or flow), the set-point pressure can be selected to correspond to the low level
or flow alarm setpoint of the process In such cases, a decreas-ing ramp input signal is used durdecreas-ing the response time test and the setpoint that initiates the test reading corresponds to the low d/p pressure setting at which the alarm or shutdown
is triggered in the process
These response time measurements can be important to overall process safety if the instrument delay time is signif-icant relative to the total time available to take corrective action after the process pressure has exceeded safe limits
The laboratory testing methods described earlier are useful for testing of sensors if they can be removed from the process and brought to a laboratory for testing, but this is often not the case For testing installed sensors, a number of new tech-niques have been developed as described below They are referred to as in situ, on-line, or in-place testing techniques
To measure the in-service response time of a temperature
response time of a temperature sensor always is a function
of the particular process temperature, process pressure, and process flow rate The most critical effect is process flow rate, followed by the effect of process temperature and then pres-sure The reason why the response time is affected by the process pressure and flow rate is because they affect the heat transfer of the film of the temperature-sensing surface of the detector In contrast, the process temperature affects not only the heat transfer of the film but also the properties of the sensor internals and sensor geometry
Consequently, it is not normally possible to accurately predict or model the effect of process temperature on the response time of temperature sensors; predicting the effects
FIG 1.9d
Photograph of pressure ramp generator for response time testing of
pressure sensors.
FIG 1.9e
Simplified diagram of pressure ramp generator.
Gas Supply
Gas
Signal Rate Adjust
Signal Initiate Solenoid
Reference Sensor
Sensor Under Test
Data Recorder Water
Trang 41.9 Response Time and Drift Testing 117
of process pressure and flow rate are easier This is because
we know that, as the process pressure or flow rate increases,
the heat transfer coefficient on the sensor surface also
increases and causes a decrease in the response time, and
vice versa In contrast, an increase in process temperature
can cause either an increase, or a decrease in the response
time of a sensor This is because, on the one hand, an increase
in process temperature can result in an increase in the heat
transfer coefficient, which reduces sensor response time On
the other hand, an increase in process temperature can also
expand or contract the various air gaps in the internals of the
temperature sensor, causing dimensional changes or altering
material properties, which can increase or decrease the
response times of the various sensors
In the case of pressure sensors, the response time is
nor-mally not changed by variations in process conditions Thus,
for pressure sensors, the choice of in situ response time testing
is based on considering the convenience of in situ testing and
less on the basis of the accuracy of the test results Therefore,
one can measure the response time of an installed pressure
sensor without removing it from the process by taking the ramp
test generator (Figure 1.9d) to the installed sensor (if this can
be done efficiently and safely) In fact, this operation is often
tedious, time consuming, and expensive, especially in
hazard-ous locations or in processes such as exist in nuclear power
plants Still, if one can afford it, using an in situ technique to
measure the response time of a pressure sensor is preferred
Testing of Temperature Sensors
The in situ response time testing of temperature sensors is
referred to as the loop current step response (LCSR) test LCSR
is performed by electrically heating the temperature sensor by
sending electric current through the sensor extension leads
This causes the temperature of the sensor to rise above the
ambient temperature Depending on the sensor involved, the
amount of current and the amount of temperature rise used in
the LCSR test can be adjusted When testing resistance
tem-perature detectors (RTDs), the use of 30 to 50 mA of DC
current is normally sufficient This amount of current raises
the internal temperature of the RTD sensor by about 5 to 10°C
(8 to 18°F) above the ambient temperature, depending on the
RTD and the process fluid surrounding it
For thermocouples, a higher current (e.g., 500 mA) is typically required This is because the electrical resistance of
a thermocouple is distributed along the length of the thermo-couple leads, but the resistance of an RTD is concentrated at the tip of the sensing element In the case of thermocouples, the LCSR current heats the entire length of the thermocouple wire, not only the measuring junction Because, in testing thermocouples, we are interested only in heat transfer at the measuring junction, it is preferred to heat up the thermocou-ple first and measure its output only after the heating current has been turned off Also, for LCSR testing of thermocouples,
AC current is used instead of DC to avoid Peltier heating or cooling, which can occur at the thermocouple junction if DC current is used The direction of the DC current determines whether the measuring junction is cooled or heated
is used in the LCSR testing of RTDs The RTD is connected
to one arm of the bridge, and the bridge is balanced while the electrical current in the circuit is low (switch is open) Under these conditions, the bridge output is recorded, and the current
is then switched to high (switch closed) to produce the bridge output for the LCSR test shown in Figure 1.9h In preparing for the LCSR test, the power supply is adjusted to provide a low current within the range of 1 to 2 mA and a high current
in the range of 30 to 50 mA The actual values depend on the RTD and on the environment in which the RTD is operating
In addition, the amplifier gain is adjusted to give an output in the range of 5 to 10 V for the bridge
Figure 1.9i shows a typical LCSR transient for a 200-Ω RTD that was tested with about 40 mA of current in an operating power plant In some plants, because of process
FIG 1.9f
Ramp test setup.
Pressure
Test Signal
Time
Test Sensor Reference Sensor
Output
Time τ
FIG 1.9g
Wheatstone bridge for LCSR test of RTDs.
RRTD
R1
RS
LCSR Transient
Variable Resistor Fixed
Resistors
Switch
DC Power Supply
Amplifier
Trang 5118 General Considerations
temperature fluctuations, the LCSR transient is not as smooth
as shown in Figure 1.9i In such cases, the LCSR test is repeated several times on the same RTD, and the results are averaged to obtain a smooth LCSR transient as in Figure 1.9i The LCSR test duration is typically 30 sec for RTDs mounted
in fast-response thermowells and tested in flowing water The LCSR test duration, when the sensor is detecting the temper-ature of liquids, typically ranges from 20 to 60 sec but is much longer for air or gas applications
ther-mocouples includes an AC power supply and circuitry shown
in the schematic in Figure 1.9j The test is performed by first applying the AC current for a few seconds while the thermo-couple is heated above the ambient temperature After that, the current flow is terminated, and the thermocouple is con-nected to a millivolt meter to record its temperature as it cools down to the ambient temperature (Figure 1.9k) The millivolt output records a transient representing the cooling of the thermocouple junction alone The rate of cooling is a function
of the dynamic response of the thermocouple
Figure 1.9l shows an LCSR transient of a thermocouple that was tested in flowing air As in the case of RTDs, LCSR transients for thermocouples can also be noisy as a result of fluctuations in process temperature and other factors To over-come noise, the LCSR test can be repeated a few times, and the resulting transients can be averaged to produce smooth
FIG 1.9h
Principle of LCSR test.
FIG 1.9i
In-plant LCSR transients for RTDs.
Time
0.0
0.2
0.4
0.6
0.8
1.0
Single Transient
0.0
0.2
0.4
0.6
0.8
1.0
1.2
Time (sec)
Averaged Transient
FIG 1.9j
Simplified schematic of LCSR test equipment for thermocouples.
V
Sensor Output
Power Supply
Test Medium
Thermocouple
Trang 61.9 Response Time and Drift Testing 119
LCSR results In the case of thermocouples, extraneous
high-frequency noise superimposed on the LCSR transient can be
removed by electronic or digital filtering
test cannot be interpreted easily into a response time reading
This is because the test data is the result of step change in
temperature inside the sensor, whereas the response time of
interest should be based on a step change in temperature
out-side the sensor Fortunately, the heat from inout-side the sensor to
the ambient fluid is transferred through the same materials as
the heat that is transferred from the process fluid to the sensor
(Figure 1.9m) Therefore, the sensor response due to internal
temperature step (LCSR test) and external temperature step
(plunge test) are related if the heat transfer is unidirectional (radial) and the heat capacity of the sensing element is insig-nificant These two conditions are usually satisfied for indus-trial temperature sensors Nevertheless, to prove that the LCSR test is valid for an RTD or a thermocouple, laboratory tests using both plunge and LCSR methods should be performed
on each sensor design to ensure that the two tests produce the same results
Because, for most sensors, the heat transfer path during LCSR and plunge tests is usually the same, one can use LCSR test data to estimate the sensor response of a plunge test where the step change in temperature occurs outside the sensor The equivalence between the two tests has been shown math-ematically (theoretically) as well as in numerous laboratory tests (see References 3 through 5) Therefore, it can be con-cluded that the test results gained from internal heating of a sensor (LCSR) can be analyzed to yield the response time of
a sensor to a step change in temperature that occurred in the medium outside the sensor
One can mathematically prove the similarities between the transient outputs, which are generated by the same temperature sensor, when evaluated by the plunge and the LCSR tests For the plunge test, the sensor output response T(t) to a step change
in the temperature of the surrounding fluid is given by
1.9(4)
Each of the three (or more) elements in Equation 1.9(4) is referred to as a mode, while the terms τ1 and τ2 are called the modal time constants, and the terms (A0, A1, A2,…) are called the modal coefficients. For the LCSR test, the sensor output response T′(t) to a step change in the temperature inside the sensor is given by
1.9(5)
Note that the exponential terms in the above two equations are identical; only their modal coefficients are different The response time (τ) of a temperature sensor is defined by Equation
FIG 1.9k
Illustration of LCSR test principle for a thermocouple.
FIG 1.9l
LCSR transient from a laboratory test of a sheathed thermocouple.
Time
Time Ambient Temperature
0
1
Time (sec)
1/16" Dia K-Type
FIG 1.9m
Heat transfer process in plunge and LCSR tests.
Heat Transfer:
Surrounding Fluid
to Sensor
Heat Transfer: Sensor
to Surrounding Fluid
LCSR
Plunge
T t( )=A +A e−t/ +A e−t/ +
1
T t( ) B0 B e1 t/τ 1 B e2 t/τ 2 L
Trang 7120 General Considerations
1.9(5), and it is therefore independent of the modal coefficients,
although it does depend on the modal time constants
1.9(6)
The other terms in the equation are the natural logarithm, ln,
and the response time (τ) of the sensor Therefore, one might
list the steps required in the LCSR test to obtain the response
time of a temperature sensor as follows:
1 Perform the LCSR test and generate the raw data
2 Fit the LCSR data to Equation 1.9(5) and identify the
modal time constants (τ1, τ2,…)
3 Use the results of Step 2 in Equation 1.9(6) to obtain
the sensor response time
The above procedure has been successfully used for
determining the response times of both RTDs and
thermo-couples, both in laboratory and in situ applications As a
result, it has been demonstrated that the LCSR test can
deter-mine the response time of a temperature sensor within about
10% of the conclusions of a plunge test if both were
per-formed under the same conditions
applica-tions of the LCSR test include the response time
determina-tion of reactor coolant temperature sensors The LCSR
tech-nique has been approved by the U.S Nuclear Regulatory
Commission (NRC) for in situ measurement of the response
also been used in aerospace applications to correct transient
temperature data and in solid rocket motors to determine the
quality of the bonding of thermocouples with the solid
mate-rials such as the nozzle liners.7
In addition to response time measurements, the LCSR
test has been used for sensor diagnostics such as
1 The in situ determination of discontinuities or
nonho-mogeneities in thermocouples.8 In this case, the
pur-pose of running the LCSR test on the thermocouple is
to check if the resulting LCSR signal is normal This
test is especially useful if a reference set of baseline
LCSR data is available, representing the test results
on normal thermocouples so that gross
nonhomoge-neities can be easily noted
2 Determining if “strap-on” RTDs are properly bonded
to pipes or tubes In case of the Space Shuttle main
engine,7 in an experiment, the LCSR test was used to
verify the quality of “strap-on” RTD bonding within
the fuel lines In this application, the RTD-based
tem-perature measurement is used to detect fuel leakages
3 Verifying the bonding of strain gauges to solid
sur-faces Figure 1.9n illustrates how the transients resulting
from LCSR tests change as a function of the strength
of RTD bonding to the pipe Therefore, the LCSR test
can determine the degree of bonding between the solid surface and RTDs or strain gauges (Figure 1.9o)
Figures 1.9p and 1.9q illustrate the commercial equipment used in LCSR testing of RTDs and thermocouples In Figure 1.9p, an LCSR test system includes six channels for RTD response time measurements so that six RTDs can be simultaneously tested This system automatically performs the LCSR test, obtains and analyzes the LCSR data, and
τ
τ τ
− −
1
2 1
3 1
FIG 1.9n
LCSR test to verify the attachment of a temperature sensor to a solid surface.
FIG 1.9o
LCSR test to verify the attachment of a strain gauge to a solid surface.
FIG 1.9p
RTD response time test equipment.
0 1
Time (sec)
LCSR Response (Normalized) Good Bond
75% Bonded
25% Bonded
Unbonded
0 1
Time (sec)
Good Bond Medium Bond
Partial Bond
Trang 81.9 Response Time and Drift Testing 121
determines the response times for each RTD The system can
send the data to a printer and print a table of RTD response
times A response time test transient display of a thermocouple
is illustrated in Figure 1.9q
In Situ Testing of Pressure Sensors
The response time of installed pressure sensors can be
mea-sured remotely while the plant is in operation This technique
is called noise analysis and is based on the monitoring of the
normally present fluctuations of the pressure transmitter output
signals In Figure 1.9r, such an output signal is shown at a
steady state that corresponds to the normal process pressure
This steady-state value is referred to as the DC reading When
magnified, it displays some small fluctuations This magnified
signal is called the noise or the AC component of the signal
sources The first source is the fluctuation of the process
pres-sure caused by turbulence, random heat transfer, vibration, and
other effects Second, there is electrical noise superimposed
on the pressure transmitter output signal Fortunately, these
two phenomena occur at widely different frequencies and thus
can be separated by filtering This is necessary, because only the process pressure fluctuations are of interest
Figure 1.9s illustrates how the noise can be extracted from
a raw signal that includes both the DC and the AC components The first step to remove the DC component is by adding a
negative bias or by highpass electronic filtering Next, the
signal is amplified and passed through a lowpass filter, which
removes the extraneous noise and provides for anti-aliasing.
Next, the signal is sent through an analog-to-digital (A/D) converter and subsequently to a data acquisition computer The computer samples the data and stores it for analysis
The raw noise data from a pressure transmitter (Figure 1.9t) represents the natural process pressure fluctuations and includes the information required to determine the response time of the pressure sensor that generated the steady-state (DC) signal The raw noise data is a small portion of a noise record, which
is normally about 30 to 60 min
For noise data analysis, the two techniques available are the frequency-domain analysis and the time-domain analysis The first uses the power spectral density (PSD) technique involving fast Fourier transform (FFT) The PSD is obtained
by bandpass filtering the raw signal in a narrow frequency band and calculating the variance of the result This variance
is divided by the width of the frequency band, and the results are plotted as a function of the center frequency of the band pass This procedure is repeated from the lowest to the highest expected frequencies of the raw signal to obtain the PSD In
Figure 1.9u, the frequency spectrum of the noise signal from
a pressure transmitter in an operating power plant is shown against PSD If the pressure transmitter is a first-order system, its response time can be determined on the basis of measuring the break frequency of the PSD as shown in Figure 1.9v
FIG 1.9q
Thermocouple response time test analyzer.
FIG 1.9r
Principle of noise analysis technique.
Noise
DC Signal
Time
FIG 1.9s
Block diagram of the noise data acquisition equipment.
FIG 1.9t
Raw noise data from a flow sensor in a power plant.
Data Sampling and Storage Device Low-Pass
Filter
High-Pass Filter
or Bias
Isolated Plant Signal
Amplifier
−4.0
−2.0 0.0 2.0 4.0
Time (sec)
Trang 9122 General Considerations
However, pressure sensors are not necessarily first order,
and PSD plots for actual process signals are not smooth
enough to allow the accurate measurement of the break
fre-quency In addition, PSDs often also contain resonance and
other disturbances that further complicate the response-time
analysis Therefore, both experience and a validated dynamic
model of the sensor are needed to obtain the sensor response
time by analyzing a PSD plot The model, which usually is
a frequency-domain equation, is fit to the PSD to yield the
model parameters, which are then used to calculate the
response time of the pressure sensor A PSD for a flow sensor
in an operating power plant and its model fit are shown in
Figure 1.9w
Autoregressive (AR) modeling is used for noise data
analysis in the time domain An AR model is a time series
equation to which the noise data is fit and the model param-eters are calculated These paramparam-eters are then used to cal-culate the response time of the sensor.9 Time-domain analysis
is generally simpler to code in a computer and therefore
is preferred for automated analysis However, in time-domain analysis, it is often difficult to remove noise data components that are unrelated to the sensor response time For example, if the noise data contains very low-frequency process fluctuations, the AR model will take them into account In such a case, it gives an erroneously large response time value In contrast, in frequency-domain analysis, it is easier to ignore low-frequency process fluctuations and to fit the PSD to that portion of the data that most accurately represents the sensor
Commercial, off-the-shelf equipment is available for both the frequency-domain and the time-domain analysis
of noise data A number of companies provide spectrum ana-lyzers (also called FFT anaana-lyzers), which take the raw noise data from the output of a sensor and provide the necessary conditioning and filtering to analyze it and calculate the sen-sor response time However, because of resonance and other influences, simple FFT analysis does not always yield the correct response time reading This is not a shortcoming of the FFT equipment but a consequence of the inherent nature
of the input signal with which they must work
ON-LINE VERIFICATION OF CALIBRATION
The calibration of installed instruments such as industrial pressure sensors involves (1) the decision whether calibration
is needed at all and (2) the actual calibration, when necessary The first step can be automated by implementing an on-line drift monitoring system This system samples the steady-state output of operating process instruments and, if it is found to have drifted, it calls for it to be calibrated Conversely, if there is no (or very little) drift, the instrument is not calibrated
at all (or calibrated less frequently) The accuracy require-ments of the sensor involved determines the amount of allow-able drift
Drift Evaluation Using Multiple Sensors
In drift evaluations, it is necessary to distinguish the drift that occurs in the process from instrument drift before a reference limit of “allowable drift” can be defined For example, if redundant sensors are used to measure the same process parameter, their average reading can be assumed to closely represent the process and used as the reference This is done
by first sampling and storing the normal operating outputs of the redundant instruments and then averaging these readings for each instant of time These average values are then sub-tracted from the corresponding individual readings of the redundant instruments to identify the deviation of each from the average
FIG 1.9u
Pressure sensor PSD
FIG 1.9v
First-order system PSD.
FIG 1.9w
Flow sensor PSD and its model fit
1.0E −06
1.0E −05
1.0E−04
1.0E −03
1.0E −02
1.0E−01
1.0E+00
Frequency (Hz)
1
Frequency (Hz)
b
Fb= Break Frequency
Fb
1.0E−06
1.0E −04
1.0E −02
1.0E+00
1.0E+02
Frequency (Hz)
Trang 101.9 Response Time and Drift Testing 123
In Figure 1.9x, the results of on-line monitoring of four
steam-generator level transmitters in a nuclear power plant
are shown The difference between the average of the four
transmitters and the individual readings are shown on the
y axis as a function of time in months The data are shown
for a period of about 30 months of operation, and the four
signals show no significant drift during this period
Conse-quently, one can conclude that the calibration of these
trans-mitters did not change and, therefore, they do not need to be
recalibrated If it is suspected that all four transmitters are
drifting in an identical manner (drifting together in one
direc-tion), the data for deviation from the average would not reveal
the drift Therefore, to rule out any systematic or common
drift, one of the four transmitters can be recalibrated
detecting systematic drift is to obtain an independent estimate
of the monitored process and track that estimate along with
the indication of the redundant sensors Both empirical and
physical modeling techniques are used to estimate systematic
drift They each monitor various related process variables
and, based on their values, evaluate the drift in the monitored
parameter For example, in a process involving the boiling
of water (without superheating of the steam), temperature
and pressure are related Thus, if temperature is measured,
the corresponding saturated steam pressure can be easily
determined, tracked, and compared with the measured
pres-sure as a reference to identify systematic drift The use of this
method of drift detection does not require the use of multiple
sensors, and individual sensors can also be tracked and their
calibration drift evaluated on line
The relationship between most process variables is much
more complex than the temperature–pressure relationship of
saturated steam Therefore, most process parameters cannot
be evaluated from measurement of another variable In
addi-tion, an in-depth knowledge of the process is needed to
pro-vide even an estimate of a parameter on the basis of physical
models Therefore, for the verification of on-line calibration, empirical models are often preferred Such empirical models use empirical equations, neural networks, pattern recognition, and sometimes a combination of these, including fuzzy logic for data clustering, are used to generate the model’s output(s) based on its multiple inputs.10–14
Before using the empirical model, it is first trained under
a variety of operating conditions As shown in Figure 1.9y,
if the output parameter (y) is to be estimated on the basis of measuring the input parameters x1, x2, and x3, then, during the training period, weighting factors are applied to the input variables These factors are gradually adjusted until the dif-ference between measured output and the output of the neural network is minimized Such training can continue while the neural network learns the relationship between the three inputs and the single output, or while additional input and output signals are provided to minimize the error in the empirical model Training of the model is completed when the measured output is nearly identical to the estimate gen-erated by the neural network Once the training is completed, the output of the model can be used for drift evaluation or control purposes
An on-line calibration monitoring system might use a combination of averaging of redundant signals (averaging can be both straight and weighted), empirical modeling, physical modeling, and calibrated reference sensor(s) in a configuration similar to the one shown in Figure 1.9z In such
a system, the raw data is first screened by a data-qualification algorithm and then analyzed to provide an estimate of the process parameter being monitored In the case of averaging analysis, a consistency algorithm is used to make sure that a reasonable agreement exists among the redundant signals and that unreasonable readings are either excluded or weighted less that the others before the signals are averaged Such systems as the one illustrated in Figure 9.1z can be consid-ered for both power plants and chemical industry applications for the on-line verification of the calibration requirements of process sensors
The data for on-line monitoring can be obtained from the plant computer or from a dedicated data acquisition system power plant for on-line calibration monitoring purposes The computer applies the on-line calibration algorithms and, based on the sampled data from a variety of process instru-ments, provides such information as plots of deviation for each instrument from a process estimate and a listing of instruments that have drifted The data acquisition system
FIG 1.9x
On-line monitoring data for steam generator level transmitters.
−3.0
−1.5
0.0
1.5
3.0
Time (Month)
SG D Level
15
FIG 1.9y
Illustration of training of a neural network
X1
Y output
X2
X3
Figure 1.9aa illustrates a data acquisition system used in a