1. Trang chủ
  2. » Khoa Học Tự Nhiên

Introduction to Modern Liquid Chromatography, Third Edition part 30 pot

10 327 0
Tài liệu đã được kiểm tra trùng lặp

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Định dạng
Số trang 10
Dung lượng 165,21 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

Table 5.10Approximate Plate Number for Well-Packed Columns under Optimized Test Conditions Particle Typea Particle Diameter μm Column Length mma Plate Number N Note: Small, neutral test

Trang 1

Table 5.10

Approximate Plate Number for Well-Packed Columns under Optimized Test Conditions

Particle Typea Particle Diameter ( μm) Column Length (mm)a Plate Number N

Note: Small, neutral test-solute, low viscosity mobile phase, ambient temperature, measured at the plate-height minimum.

aEstimated values for 4.6 mm i.d columns.

bAn average for commercially available, sub-2- μm particles.

new column that met the manufacturer specifications If problems are encountered

with a routine method that might be caused by the column, values of N and either

‘‘bad’’ column can be confirmed as the cause of the problem

Every column used for a routine method has a finite lifetime (number of injected samples before column failure) that depends on separation conditions—especially mobile-phase pH and temperature For ‘‘clean’’ samples, 1000 to 2000 analy-ses should be possible for must silica-based columns, particularly reversed phase columns However, for other samples (with minimal sample preparation), such as extracts of blood, plant or animal tissue, or soil, 200–500 analyses is a more typical column lifetime

5.8 COLUMN HANDLING

The performance and life of the column depend on how it is used and handled During heavy use with ‘‘dirty’’ samples (especially samples from biological sources),

columns can develop severe peak tailing (Fig 5.26a) or double peaks for each component (Fig 5.26b)— usually the result of a partly blocked frit, a contaminated

Trang 2

(a) (b)

peak tailing split peaks

Figure5.26 Examples of peak tailing (a) and split peaks (b).

column, or deterioration of the column packing The following restorative measures are sometimes effective, but the time and effort involved are often not cost-effective Usually it is more economical to simply replace the column

A blocked frit or contaminated column can sometimes be restored by periodi-cally purging the column with a strong solvent A 20-column-volume purge (about

0.1% ammonium hydroxide in 4% methanol is often effective for RPC columns Pure methanol or isopropanol can be used for normal-phase columns Flushing a RPC column (at least) daily with a strong solvent, such as methanol or acetoni-trile, can enhance column performance and lifetime for isocratic separations This approach removes strongly retained sample components that can build up at the

column inlet Back-flushing a column with a strong solvent at 0.2 to 0.5 mL/min

may be more effective, so as to avoid driving the column contaminants into the column However, some manufactures recommend against back-flushing because their columns are fitted with an inlet frit that has larger pores that might allow particles to be swept out (consult the column care-and-use instructions for a specific column) In gradient elution, clearing the column of strongly retained components can be accomplished by using a steep or step gradient (Section 9.2.2.5) The use of a

blockage of the column inlet frit, and is highly recommended

To reduce the possible impact of ‘‘dirty’’ samples on column lifetime, sample

pretreatment is commonly used (Chapter 16) A guard column can also be used to

protect the column, and it is recommended for routine analysis A guard column

is short (e.g., 10–20 mm) and preferably contains a packing that is the same as

or similar to that in the main column (Section 5.4.2) The guard column captures strongly retained sample components (and particulates), and prevents these from fouling the analytical column Guard columns must be replaced at regular intervals, before the column becomes saturated with strongly retained sample components that then pass into the analytical column However, because of their added expense and inconvenience, some users prefer to avoid guard columns and replace the main column more frequently The use of guard columns with low-volume, high-efficiency columns (e.g., sub-2-μm columns) requires special care, because of the greater importance of extra-column peak broadening when small columns are used For columns that are not well packed, a sudden pressure surge (as during sample injection) can cause a void at the column inlet, with a decrease in column performance Fortunately, this problem is no longer common for columns from

Trang 3

silica-based columns, other types of particles (e.g., organic gels, graphitized carbon) are more fragile and less able to withstand sudden changes in flow, pressure, or temperature

Column performance can be reduced significantly by a loss of the bonded phase during use, leading to a short column lifetime The recommendations of the manufacturer, especially with regard to mobile-phase pH and temperature, should be

(Fig 5.16b) provide additional stability at low pH High-pH mobile phases (e.g., pH

> 8) can slowly dissolve silica-based packings, again resulting in a degradation of

column performance Columns of hybrid silica-silane particles (Section 5.3.2.2) and those based on zirconia (Section 5.2.5.1) are especially resistant to degradation by high-pH mobile phases; some other special stationary phases (e.g., bonded bidentate

silanes, Fig 5.14b or c) are also more stable for high-pH applications [44].

Stationary-phase loss from silica-based columns at low pH is accelerated at higher temperatures Therefore higher temperatures as a means of improving col-umn performance or separation selectivity should be used carefully Many workers

Degradation of silica-based columns at higher pH occurs via dissolution of the silica, and its degradation is also accelerated by higher temperatures Figure 5.27

0 5 10 15

500 400 300 200 100 0

Mobile-phase volume (L)

60 °C

40 °C

Figure5.27 Effect of temperature on silica-support dissolution Column: Zorbax Rx-C18,

15× 0.46 cm; continuous nonrecycled 20% acetonitrile/80% sodium phosphate buffer,

0.25 M, pH 7.0 Flow rate: 1.0 mL/min Adapted from [78]

Trang 4

columns should be avoided because of rapid dissolution of the silica support [78].

On the other hand, use of organic buffers (e.g., TRIS, HEPES, citrate) may increase column lifetime over that with phosphate and carbonate buffers when operating

at intermediate and higher pH [79] One speculation is that these basic, partially hydrophobic, organic buffer compounds tend to bind to unreacted silanol groups

on the packing so as to create an additional barrier to the dissolution of the silica support [79] However, the advantage of the organic buffers may be misleading, as the actual pH of the buffer/organic solvent mobile phase may be somewhat lower than that measured for the buffer solution itself (the pH of phosphate and carbonate buffers in organic-containing solvents is somewhat higher that the aqueous buffer [80]); see the further discussion of Section 7.2.3

The performance and lifetime of a column can also be affected by improper

microbial growth if they are stored for more than a day at room temperature

The resulting particulates can in turn block the column inlet, reducing N and

increasing the column pressure Therefore it is good practice to formulate buffers

absence of oxygen due to helium sparging, can inhibit bacterial growth and prolong buffer life

When removed from the system, the column is best stored in a nonprotic solvent such as acetonitrile (100% B) For short-term applications (overnight or

a few days) it is convenient (and acceptable) to leave the mobile phase in place However, prolonged storage with buffered solutions, particularly those with high concentrations of water or alcohols, should be avoided Prior to storage, the column should be flushed with 5 to 10 column volumes of the same aqueous-organic

mobile phase but without buffer before an additional 5-column-volume flush with

100% organic phase (this avoids precipitation of the buffer within the column) Flushing columns with pure water for long periods should be avoided because of

prevent columns from drying out, they should be tightly capped for storage

order to retain these very small particles Therefore both the sample and mobile phase should be passed through 0.2-μm filters to ensure that particulates do not block the frits and degrade column performance Small-particle columns are often

very narrow widths—measured either in time or volume This requires instrumental conditions that minimize potential extra-column effects that artificially broaden peaks:

• short, low-volume tubing that connects the sampling valve, column, and detector

• high data-capturing rate (at least 20 points/sec or 20 Hz)

For further details, see [81, 82]

Trang 5

1 K K Unger, Porous Silica, Elsevier, Amsterdam, 1979, p 294.

2 K Kalghatgi and C Horv ´ath, J Chromatogr., 443 (1988) 343.

3 K K Unger and H Giesche, Ger Pat DE-3543 143.2 (1985).

4 J J Kirkland, F A Truszkowski, C H Dilks, Jr., and G S Engel, J Chromatogr A,

890 (2000) 3

5 J J Kirkland, T J Langlois, and J J DeStefano, Amer Lab., 39 (2007) 18.

6 N B Afeyan, N F Gordon, I Mazsaroff, L Varady, S P Fulton, Y B Yang, and F E

Regnier, J Chromatogr., 519 (1990) 1.

7 D Whitney, M McCoy, N Gordon, and N Afeyan, J Chromatogr A, 807 (1998)

165

8 L R Snyder and M A Stadalius, in High-performance Liquid Chromatography:

Advances and Perspectives, Vol 4, C Horvath, ed., Academic Press, San Diego, 1986.

9 R K Iler, The Chemistry of Silica, Wiley, New York, 1979, p 639.

10 J Billen, P Gzil, F Lynen, P Sandra, P Van der Meeren, and G Desmet, Poster, HPLC

2006, San Francisco, June 2006

11 K K Unger and E Weber, A Guide to Practical HPLC, Git Verlag, Darmstadt, 1999,

p 45

12 U D Neue, HPLC Columns, Wiley-VCH, New York, 1997, p 82.

13 J J DeStefano, T J Langlois and J J Kirkland, J Chromatogr Sci., 46 (2008) 1.

14 J K ¨ohler, D B Chase, R D Farlee, A J Vega, and J J Kirkland, J Chromatogr., 352

(1986) 275

15 J K ¨ohler and J J Kirkland, J Chromatogr., 385 (1987) 125.

16 J Nawrocki, Chromatographia, 31 (1991) 177.

17 J Nawrocki, Chromatographia, 31 (1991) 193.

18 H Engelhardt, H Low, and W G ¨otzinger, J Chromatogr., 544 (1991) 371.

19 J J Kirkland, B E Boyes, and J J DeStefano, Amer Lab., 26 (1994) 36.

20 D W Sindorf and G E Maciel, J Am Chem Soc., 105 (1983) 1487.

21 P G Dietrich, K.-H Lerche, J Reusch, and R Nitzsche, Chromatographia, 44 (1997)

362

22 A M´endez, E Bosch, M Ros´es, and U D Neue, J Chromatogr A, 986 (2003) 33.

23 M Jacoby, Chemical and Engineering News, Dec 11, 2006

24 F Svec and C G Huber, Anal Chem., 78 (2006) 2100.

25 G Guiochon, J Chromatogr A, 1168 (2007) 101.

26 K K Unger, R Skudas, and M M Schulte, J Chromatogr A, 1184 (2008) 393.

27 F Svec, T Tennikova, and Z Deyl (eds.), Monolithic Materials: Preparation, Properties

and Applications, J Chromatogr Library, Vol 67, Elsevier, Amsterdam, 2004.

28 M Kele and G Guiochon, J Chromatogr A, 960 (2002) 19.

29 X Wang, D R Stoll, P W Carr, and P J Schoenmakers, J Chromatogr A, 1125

(2006) 177

30 S Eeltink, G Desmet, G Viv ´o-Truyols, G P Rozing, P J Schoemnmakers, and W Th

Kok, J Chromatogr A, 1104 (2006) 256.

31 T Hara, H Kobayashi, T Ikegami, K Nakanishi, and N Tanaka, Anal Chem., 78

(2006) 7632

32 B Bidlingmeyer, K K Unger, and N von Doehren, J Chromatogr A, 832 (1999) 11.

Trang 6

33 L Geiser, S Eeltink, F Svec, and J M J Fr´echet, J Chromatogr A, 1140 (2007) 140.

34 J Nawrocki, C Dunlap, J Li, J Zhao, C V McNeff, A McCormick, and P W Carr,

J Chromatogr A, 1028 (2004) 31.

35 J Nawrocki, C Dunlap, A McCormick, and P W Carr, J Chromatogr A, 1028 (2004)

1

36 U Bien-Vogelsang, A Deege, H Figge, J K ¨ohler, and G Schomburg, Chromatographia,

19 (1984) 170

37 J J Pesek and M T Matyska, J Chromatogr A, 952 (2001) 1.

38 P Ross, LCGC, 18 (2000) 14.

39 J H Knox, B Kaur, and G R Millward, J Chromatogr., 352 (1986) 3.

40 M J Wirth and H O Fatunmbi, Anal Chem, 65 (1993) 822.

41 J J Kirkland, J L Glajch, and R D Farlee, Anal Chem., 61 (1988) 2.

42 J L Glajch and J J Kirkland, US Pat 4,705,725, 1987

43 J L Glajch and J J Kirkland, US Pat 4,847,159, 1989

44 J J Kirkland, J B Adams, M A Van Straten, and H A Claessens, Anal Chem., 70

(1998) 4344

45 L C Sanders and S A Wise, Anal Chem., 56 (1984) 504.

46 N S Wilson, J Gilroy, J W Dolan, and L R Snyder J Chromatogr A, 1026 (2004)

91

47 L Brown, B, Ciccone, J J Pesek, and M T Matyska, Amer Lab., Dec (2003) 23.

48 K Miyabe and N Orita, Talanta, 36 (1989) 897.

49 M R Buchmeiser, J Chromatogr A, 918 (2001) 233.

50 K K Unger and J Schick-Kalb, US Pat 4,017,528, 1977

51 K D Wyndham, J E O’Gara, T H Walter, K H Glose, N L Lawrence, B A Alden,

G S Izzo, C J Hudalla, and P C Iraneta, Anal Chem., 75 (2003) 6781.

52 R E Majors, LCGC, 20 (2002) 516.

53 B A Bidlingmeyer and A D Broske, J Chromatogr Sci., 42 (2004) 100.

54 T H Walter, P Iranetaund, and M Capparella, J Chromatogr A, 1075 (2005) 177.

55 R Eksteen and J Thoma, in: Chromatography, the State of the Art, Vol 1, Akademia

Kiado, Budapest, 1985

56 L Zhang, L Sun, J I Siepmann, and M R Shure, J Chromatogr A, 1079 (2005) 127.

57 K Croes, A Steffens, D H Marchand, and L R Snyder, J Chromatogr A, 1098 (2005)

123

58 M R Euerby, P Petersson, W Campbell, and W Roe, J Chromatogr A, 1154 (2007)

138

59 U D Neue, J Sep Sci., 30 (2007) 1611.

60 N S Wilson, M D Nelson, J W Dolan, L R Snyder, R G Wolcott, and P W Carr,

J Chromatogr A, 961 (2002) 171.

61 L R Snyder, J W Dolan, and P W Carr, J Chromatogr A, 1060 (2004) 77.

62 L R Snyder, J W Dolan, and P W Carr, Anal Chem., 79 (2007) 3255.

63 D H Marchand and L R Snyder, J Chromatogr A, 1209 (2008) 104.

64 L R Snyder and J W, Dolan, High Performance Gradient Elution, Wiley-Interscience,

Hoboken, NJ, 2007, p 420

65 L C Sander and S A Wise, J Chromatogr A, 656 (1993) 335.

66 J J Kirkland, Amer Lab., 26 (1994) 28K.

Trang 7

849 (1999) 87.

68 L R Snyder, A Maule, A Heebsch, R Cuellar, S Paulson, J Carrano, L Wrisley, C

C Chan, N Pearson, J W Dolan, and J Gilroy, J Chromatogr A, 1057 (2004) 49.

69 J Pellett, P Lukulay, Y Mao, W Bowen, R Reed, M Ma, R C Munger, J W Dolan,

L Wrisley, K Medwid, N P Toltl, C C Chan, M Skibic, K Biswas, K A Wells, and

L R Snyder, J Chromatogr A, 1101 (2006) 122.

70 Y Mao and P W Carr, LCGC, 21 (2003) 150.

71 J W Dolan and L R Snyder, J Chromatogr A, 1216 (2009) 3467.

72 D V McCalley, Adv Chromatogr., 46 (2008) 305.

73 D M Marchand, L R Snyder, and J W Dolan, J Chromatogr A, 1191 (2008) 2.

74 H Colin P Hilaireau, and J De Tournemire, LCGC, 8 (1990) 302.

75 F Couillard, Chromatography Apparatus, US Patent 4,597,866, 1986-07-01

76 J J Kirkland and J J DeStefano, J Chromatogr A, 1126 (2006) 50.

77 L R Snyder and J J Kirkland, Introduction to Modern Liquid Chromatography, 2nd

ed., Wiley, New York, 1979, ch 5

78 H A Claessens, M A Van Straten, and J J Kirkland, J Chromatogr A, 728 (1996)

259

79 J J Kirkland, M A Van Straten, and H A Claessens, J Chromatogr A, 797 (1998)

111

80 G W Tindall and R L Perry, J Chromatogr A, 988 (2003) 309.

81 J J Kirkland, J Chromatogr Sci., 38 (2000) 535.

82 F Gerber, M Krummen, H Potgeter, A Roth, C Siffrin, and C Spoendlin, J

Chro-matogr A, 1036 (2004) 127.

Trang 8

CHAPTER SIX

REVERSED-PHASE

CHROMATOGRAPHY

FOR NEUTRAL SAMPLES

6.1 INTRODUCTION, 254

6.1.1 Abbreviated History of Reversed-Phase Chromatography, 255

6.2 RETENTION, 256

6.2.1 Solvent Strength, 257

6.2.2 Reversed-Phase Retention Process, 259

6.3 SELECTIVITY, 263

6.3.1 Solvent-Strength Selectivity, 263

6.3.2 Solvent-Type Selectivity, 265

6.3.3 Temperature Selectivity, 270

6.3.4 Column Selectivity, 273

6.3.5 Isomer Separations, 276

6.3.6 Other Selectivity Considerations, 278

6.4 METHOD DEVELOPMENT AND STRATEGIES

FOR OPTIMIZING SELECTIVITY, 284

6.4.1 Multiple-Variable Optimization, 286

6.4.2 Optimizing Column Conditions, 295

6.5 NONAQUEOUS REVERSED-PHASE CHROMATOGRAPHY

(NARP), 295

6.6 SPECIAL PROBLEMS, 297

6.6.1 Poor Retention of Very Polar Samples, 297

6.6.2 Peak Tailing, 298

Introduction to Modern Liquid Chromatography, Third Edition, by Lloyd R Snyder,

Joseph J Kirkland, and John W Dolan

Copyright © 2010 John Wiley & Sons, Inc.

253

Trang 9

6.1 INTRODUCTION

This chapter describes the separation of neutral samples by means of reversed-phase chromatography (RPC) By a ‘‘neutral’’ sample, we mean one that contains no molecules that carry a positive or negative charge—usually as the result of the ionization of an acid or a base Although a neutral sample implies an absence

of acidic and basic solutes, this is not necessarily the case Depending on mobile phase pH, any acids or bases in the sample may be present largely (e.g., 90%+)

in the neutral (non-ionized) form—in which case their chromatographic behav-ior is similar to that of non-ionizable compounds The separation of ‘‘ionic’’ samples (which contain one or more ionized compounds) by RPC is covered in Chapter 7

RPC is usually a first choice for the separation of both neutral and ionic samples, using a column packing that contains a less polar bonded phase such

acetonitrile (ACN) or methanol (MeOH); other organic solvents (e.g., isopropanol [IPA], tetrahydrofuran [THF]) are used less often A preferred organic solvent for

an RPC mobile phase will be water-miscible, relatively nonviscous, stable under the conditions of use, transparent at the lowest possible wavelength for UV detection, and readily available at moderate cost Commonly used B-solvents can be ranked in terms of these properties as follows:

In a few countries, ACN is considered sufficiently toxic to limit its general use, but this is not true elsewhere See Appendix I for further information concerning solvent properties and the choice of B-solvent for a given application Samples that contain acids or bases normally require a buffered mobile phase, in order to maintain a constant pH throughout the separation (Chapter 7) Strongly retained, very hydrophobic samples may require a water-free mobile phase (nonaqueous reversed-phase chromatography [NARP], Section 6.5) Normal-phase chromatog-raphy (Chapter 8) can also provide acceptable separations of very hydrophobic samples, as sample hydrophobicity contributes little to retention for this HPLC mode Preferred conditions for the isocratic separation of neutral samples by RPC are listed in Table 6.1

Compared to other forms of HPLC (normal-phase, ion-exchange chromatog-raphy, etc.; Table 2.1), separations by RPC are usually more convenient, robust, and versatile RPC columns also tend to be more efficient and reproducible, and are available in a wider range of choices that include column dimensions, particle size,

used for RPC tend to be less flammable or toxic, and are more compatible with UV detection at wavelengths below 230 nm for increased detection sensitivity (Table I.2

of Appendix I) An additional advantage of RPC is generally fast equilibration of the column after a change in the mobile phase—or between runs when using gradient elution (Section 9.3.7) Finally, because RPC has been the dominant form of HPLC since the late 1970s, a better practical understanding of this technique has evolved This usually means an easier development of better separations All of the foregoing reasons have contributed to the present popularity of RPC

Trang 10

Table 6.1

Preferred Conditions for the Separation of Neutral Samples by Reversed-Phase

Chromatography

Dimensions: 100× 4.6-mm

particle size: 3 μm Pore diameter: 8–12 nm

Weight ≤ 50 μg

aAlternatively, use a 150× 4.6-mm column of 5-μm particles; flow rate, column dimensions, and particle

size can be varied, depending on the anticipated difficulty of the separation and the maximum allowable column pressure (Section 2.4.1)

bInitial values, which may be changed during method development (Section 2.5); a temperature 5–10◦C above ambient is suggested in most cases Also consult the column manufacturer’s recommendations for

a maximum column temperature.

cVaries with the sample; start with 80%B and adjust further as described in Section 2.5.1.

Many organic compounds have limited solubility in either water or the water-organic mobile phases used for RPC, but this is rarely a practical con-cern Thus very small weights (nanograms or low micrograms) of individual solutes are usually injected, so the required sample concentration is usually only

a few micrograms/mL or less In those cases where sample solubility in water or water-organic mixtures is exceptionally poor (very hydrophobic samples), the use

of normal-phase chromatography with nonaqueous mobile phases may be preferred (Section 8.4.1)

Some samples are less well separated by RPC For example, very polar

molecules may be retained weakly in RPC (k

mobile phase; these samples may require a different approach (Section 6.6.1) Similarly enantiomers require separation conditions that exhibit chiral selectivity (Chapter 14) While many achiral isomers can be separated by RPC (Section 6.3.5), these compounds are often better separated by normal-phase chromatography using

an unbonded silica column (Section 8.3.4.1) Finally, normal-phase chromatography

is often a better choice for preparative HPLC (Chapter 15)

6.1.1 Abbreviated History of Reversed-Phase Chromatography

Prior to the invention of RPC in 1950 by A J P Martin [1], the chromatographic separation of neutral samples was carried out with a polar column (or stationary phase) and a less polar mobile phase; such separations are now referred to as

Ngày đăng: 04/07/2014, 01:20

TỪ KHÓA LIÊN QUAN