COMPUTED RADIOGRAPHY CR: USES PHOTOSTIMULABLE PLATE IMAGE PLATE INSTEAD OF FILM, AS THE IMAGE RECEPTOR.. CR CASSETTES ARE NOT LIGHT-TIGHT IMAGE PLATE IS NOT SENSITIVE TO LIGHT... CROSS
Trang 1DIGITAL PROJECTION
RADIOGRAPHY
COMPUTED RADIOGRAPHY
Trang 2DIGITAL PROJECTION RADIOGRAPHY DEPENDS ON COMPUTER TECHNOLOGY TO
PRODUCE DIGITAL RADIOGRAPHIC IMAGE
Trang 3DIGITAL PROJECTION RADIOGRAPHY
Trang 4DIGITAL RADIOGRAPHY
vsCOMPUTED RADIOGRAPHY
Trang 7COMPUTED RADIOGRAPHY
(CR):
USES PHOTOSTIMULABLE PLATE (IMAGE
PLATE) INSTEAD OF FILM, AS THE IMAGE
RECEPTOR.
CR USES CASSETTE THAT LOOKS VERY
SIMILAR TO CONVENTIONAL RADIOGRAPHIC
CASSETTE
Trang 8CR PLATE !!!!!
DOES NOT USE SCREENS OR FILM ( CONVENTIONAL FILM)
Trang 10CR PLATE
Trang 12CR CASSETTES ARE NOT
LIGHT-TIGHT
IMAGE PLATE IS NOT
SENSITIVE TO LIGHT
Trang 13CR CASSETTE IS PROTECTING IMAGE PLATE
FROM DAMAGE
+
IS THE STABLE VEHICLE FOR
TRANSPORT AND PLACEMENT OF THE CASSETTE UNDER THE
PATIENT
Trang 14BACK PANEL CONTAINS LEAD FOIL TO PROTECT THE PLATE
FROM BACKSCATTER
Trang 15BACK PANEL ALSO CONTAINS
CHIP TO RECORD PATIENT
DEMOGRAPHIC INFORMATION
Trang 16REMOTE OPERATOR PANEL
Trang 20IMAGE PLATE
APPROX 1MM
Trang 21CROSS-SECTION OF CR
PLATE
PROTECTIVE LAYER PHOSPHOR LAYER ANTI-HALO & REFLECTIVE LAYER
BASE
BACKING LAYER
Trang 22► PET- POLYETHYLENE TERAPHTALATE
Trang 23PHOSPHOR LAYER
► Ba FX: Eu +2
Trang 24PROTECTIVE LAYER
► FLUORINATED POLYMER MATERIAL
PROTECTS PHOSPHOR LAYER
Trang 25ANTI-HALO LAYER + REFLECTIVE
LAYER
► PREVENTS LASER FROM PASSING THROUGH.
► REFLECTED LIGHT FROM PHOSPHOR IS ALLOWED
TO PASS.
Trang 26BACKING LAYER
► PROTECTS THE BASE FROM DAMAGE.
Trang 27LEAD COATING
SCATTER
Trang 28The simplest explanation for luminescence is that impurities in the
crystal lattice are responsible for
luminescence As the concentration of impurity ions increase the greater the
intensity of the luminescence CR
screens use barium fluorohalides doped with europium (europium is the impurity
in the crystal).
Trang 29When phosphors are stimulated with ray photon energy electron pair holes are created In effect, europium is raised
x-to an excited state and upon luminescence it is returned to its ground Eu2+ state This mechanism holds for both spontaneous luminescence and
photostimulated luminescence
Trang 30The shifting of europium from its excited state back to its ground state for both spontaneous and photostimulated
luminescence is about 0.6 - 0.8 microseconds With screen-film imaging
these crystals spontaneously luminescence to expose a film, but with
CR imaging the luminescence occurs, then there is also photoluminescence
that occurs when the screen is stimulated by a narrow beam of infrared
light.
Trang 31The holes or vacancies in the lattice are portions of the lattice normally occupied
by halogens (fluoride, bromide, or
iodine) These vacancies will trap free electrons when irradiated and are called
Farbzentren centers or F-centers.
Trang 32When the photostimulable plate is
exposed to high frequency light, usually from a helium laser, the electrons in these F-centers are liberated and cause
luminescence at readout.
Trang 33THE LATENT IMAGE WILL REMAIN STORED FOR 24
HOURS
IT WILL FADE THROUGH
PHOSPHORESCENCE
Trang 34► 25% OF STORED
ENERGY
► WITHIN 8 HOURS
Trang 35CR AFTER EXPOSURE IS
INSERTED INTO CR SCANNER
Trang 36CR PLATE
CR SCANNER/READER
Trang 37THE SCANNER SCANS THE IMAGE PLATE WITH LASER LIGHT AND READS THE
LATENT IMAGE FROM THE
PHOSPHORREADOUT
Trang 38LASER LIGHT
HELIUM-NEON (633 nm) LASER BEAM SWEEPS THE PLATE IN RASTER FASHION
Trang 40RASTER
Trang 41PM TUBE
AMPLIFIER ADC
LASER
LIGHT GATE
Trang 42SWEEPING OF THE LASER
ACROSS THE PLATE
► FAST SCAN DIRECTION
► SCAN DIRECTION
► LASER SCAN DIRECTION
Trang 43PLATE MOVEMENT IN THE
Trang 44THE SPEED OF THE SCANNING LASER BEAM IS
ADJUSTED TO THE LUMINESCENT SIGNAL
Trang 45PLATE & F-CENTERS
LASER 633 nm
PHOTOSTIMULABLE LUMINESCENCE
390-400 nm
Trang 46LASER LASER SPOT SIZE -
100 MICROMETERSPIXEL SIZE - 100 MICROMETERS
Trang 47P.M TUBE DETECTION
SENSITIVITY MATCHED TO BLUE-PURPLE PSL (390-400
nm)
Trang 48AFTER SCANNING THE CR
PLATE NEED TO BE CLEARED
FROM RESIDUAL SIGNAL
PLATE EXPOSED TO HIGH
INTENSITY SODIUM VAPOR
OR FLUORESCENT LIGHT
Trang 50Typical image plate can be reused thousands of times
Trang 51IMAGE DISPLAY
Trang 52IMAGE CHECK-INDEX???
Trang 53IMAGE
MANIPULATION-WINDOWING
Trang 54PICTURE DESTINATION -PACS
Trang 55PACS
Trang 56CR CHARACTERISTIC CURVE
► VERY WIDE LATITUDE
Trang 57CR SYSTEMS
► LESS RADIOGRAPH REPEATS
Trang 58CHARACTERISTIC CURVE
LATITTUDE OF CR LATITTUDE OF FILM
Trang 59IMAGE CAN BE SALVAGED WITH
CR
► 500% OVEREXPOSURE
► 80% UNDEREXPOSURE
USE ADEQUATE KVP
Trang 60HARD COPY GENERATED IN
LASER PRINTER
Trang 61CR SYSTEM EFFICIENCY
CHARACTERISTIC
► QDE- QUANTUM DETECTION EFFICIENCY
NO NEED FOR SCREENS
Trang 62THE MEASURE OF THE EFFICIENCY OF
CR SYSTEM TO CONVERT REMNANT
X-RAYS TO USEFUL IMAGE SIGNAL
Trang 63PATIENT DOSE
Trang 64DIGITAL RADIOGRAPHY
Trang 65ELECTRONIC DETECTORS
► INDIRECT TECHNIQUE CLASS
► DIRECT TECHNIQUE CLASS
Trang 66INDIRECT TECHNIQUE
► INDIRECT CONVERSION FLAT PANEL DETECTORS
► INDIRECT CONVERSION CHARGED COUPLE DEVICE DETECTORS
Trang 67INDIRECT CONVERSION FLAT
PANEL DETECTORS
► USE THIN LAYERS OF SILICON COMBINED WITH THE ARRAYS OF PHOTODIODES P.D IS COATED WITH CESIUM IODIDE OR RARE EARTH
MATERIAL- WHEN STRUCK BY REMNANT X-RAYS THEY EMIT LIGHT
ADVANTAGE: HIGH QDE DISADVANTAGE: LIGHT DIVERGENCE
Trang 68PHOTODIODE
Trang 70INDIRECT CONVERSION CHARGE COUPLED DEVICE
DETECTORS
► USE CCD ARRAYS TO RECORD THE VISIBLE LIGHT EMITTED BY A SCINTILLATION CRYSTAL THAT HAS ABSORBED THE ENERGY AN OPTICAL COUPLING SYSTEM DEMAGNIFIES THIS LIGHT AND SENDS IT
TO THE CCD ARRAY.
Trang 71OPTICAL COUPLING
Trang 72CCD
Trang 73DIRECT TECHNIQUE
► DIRECT CONVERSION FLAT PANEL DETECTORS
Trang 74DIRECT CONVERSION FLAT
ADVANTAGE: NO LIGHT DIVERGENCE
Trang 75READY - The 17"x17"
amorphous selenium-based
sensor is electrically sensitized
EXPOSE - Incident X-ray photons are
converted into electron-hole pairs A
transverse electric field separates the
electrons and holes and creates a
charge image precisely
corresponding to the X-ray image.
READOUT - The microplasma line
scanner sweeps across the sensor in
less than 2 seconds causing readout
of the charge image and sensor reset.
Trang 78PORTABLE????
Trang 79PORTABLE CT???!!
Trang 814K MONITORS USED IN MAMMOGRAPHY
AND CHEST RADIOGRAPHY
Trang 83PIXEL
Trang 84
The formation