COMPUTED RADIOGRAPHY CR: USES PHOTOSTIMULABLE PLATE IMAGE PLATE INSTEAD OF FILM, AS THE IMAGE RECEPTOR.. CR CASSETTES ARE NOT LIGHT-TIGHT IMAGE PLATE IS NOT SENSITIVE TO LIGHT... CROSS

Trang 1

DIGITAL PROJECTION

RADIOGRAPHY

COMPUTED RADIOGRAPHY

Trang 2

DIGITAL PROJECTION RADIOGRAPHY DEPENDS ON COMPUTER TECHNOLOGY TO

PRODUCE DIGITAL RADIOGRAPHIC IMAGE

Trang 3

DIGITAL PROJECTION RADIOGRAPHY

Trang 4

DIGITAL RADIOGRAPHY

vsCOMPUTED RADIOGRAPHY

Trang 7

COMPUTED RADIOGRAPHY

(CR):

USES PHOTOSTIMULABLE PLATE (IMAGE

PLATE) INSTEAD OF FILM, AS THE IMAGE

RECEPTOR.

CR USES CASSETTE THAT LOOKS VERY

SIMILAR TO CONVENTIONAL RADIOGRAPHIC

CASSETTE

Trang 8

CR PLATE !!!!!

DOES NOT USE SCREENS OR FILM ( CONVENTIONAL FILM)

Trang 10

CR PLATE

Trang 12

CR CASSETTES ARE NOT

LIGHT-TIGHT

IMAGE PLATE IS NOT

SENSITIVE TO LIGHT

Trang 13

CR CASSETTE IS PROTECTING IMAGE PLATE

FROM DAMAGE

+

IS THE STABLE VEHICLE FOR

TRANSPORT AND PLACEMENT OF THE CASSETTE UNDER THE

PATIENT

Trang 14

BACK PANEL CONTAINS LEAD FOIL TO PROTECT THE PLATE

FROM BACKSCATTER

Trang 15

BACK PANEL ALSO CONTAINS

CHIP TO RECORD PATIENT

DEMOGRAPHIC INFORMATION

Trang 16

REMOTE OPERATOR PANEL

Trang 20

IMAGE PLATE

APPROX 1MM

Trang 21

CROSS-SECTION OF CR

PLATE

PROTECTIVE LAYER PHOSPHOR LAYER ANTI-HALO & REFLECTIVE LAYER

BASE

BACKING LAYER

Trang 22

► PET- POLYETHYLENE TERAPHTALATE

Trang 23

PHOSPHOR LAYER

► Ba FX: Eu +2

Trang 24

PROTECTIVE LAYER

► FLUORINATED POLYMER MATERIAL

PROTECTS PHOSPHOR LAYER

Trang 25

ANTI-HALO LAYER + REFLECTIVE

LAYER

► PREVENTS LASER FROM PASSING THROUGH.

► REFLECTED LIGHT FROM PHOSPHOR IS ALLOWED

TO PASS.

Trang 26

BACKING LAYER

► PROTECTS THE BASE FROM DAMAGE.

Trang 27

LEAD COATING

SCATTER

Trang 28

The simplest explanation for luminescence is that impurities in the

crystal lattice are responsible for

luminescence As the concentration of impurity ions increase the greater the

intensity of the luminescence CR

screens use barium fluorohalides doped with europium (europium is the impurity

in the crystal).

Trang 29

When phosphors are stimulated with ray photon energy electron pair holes are created In effect, europium is raised

x-to an excited state and upon luminescence it is returned to its ground Eu2+ state This mechanism holds for both spontaneous luminescence and

photostimulated luminescence

Trang 30

The shifting of europium from its excited state back to its ground state for both spontaneous and photostimulated

luminescence is about 0.6 - 0.8 microseconds With screen-film imaging

these crystals spontaneously luminescence to expose a film, but with

CR imaging the luminescence occurs, then there is also photoluminescence

that occurs when the screen is stimulated by a narrow beam of infrared

light.

Trang 31

The holes or vacancies in the lattice are portions of the lattice normally occupied

by halogens (fluoride, bromide, or

iodine) These vacancies will trap free electrons when irradiated and are called

Farbzentren centers or F-centers.

Trang 32

When the photostimulable plate is

exposed to high frequency light, usually from a helium laser, the electrons in these F-centers are liberated and cause

luminescence at readout.

Trang 33

THE LATENT IMAGE WILL REMAIN STORED FOR 24

HOURS

IT WILL FADE THROUGH

PHOSPHORESCENCE

Trang 34

► 25% OF STORED

ENERGY

► WITHIN 8 HOURS

Trang 35

CR AFTER EXPOSURE IS

INSERTED INTO CR SCANNER

Trang 36

CR PLATE

CR SCANNER/READER

Trang 37

THE SCANNER SCANS THE IMAGE PLATE WITH LASER LIGHT AND READS THE

LATENT IMAGE FROM THE

PHOSPHORREADOUT

Trang 38

LASER LIGHT

HELIUM-NEON (633 nm) LASER BEAM SWEEPS THE PLATE IN RASTER FASHION

Trang 40

RASTER

Trang 41

PM TUBE

AMPLIFIER ADC

LASER

LIGHT GATE

Trang 42

SWEEPING OF THE LASER

ACROSS THE PLATE

► FAST SCAN DIRECTION

► SCAN DIRECTION

► LASER SCAN DIRECTION

Trang 43

PLATE MOVEMENT IN THE

Trang 44

THE SPEED OF THE SCANNING LASER BEAM IS

ADJUSTED TO THE LUMINESCENT SIGNAL

Trang 45

PLATE & F-CENTERS

LASER 633 nm

PHOTOSTIMULABLE LUMINESCENCE

390-400 nm

Trang 46

LASER LASER SPOT SIZE -

100 MICROMETERSPIXEL SIZE - 100 MICROMETERS

Trang 47

P.M TUBE DETECTION

SENSITIVITY MATCHED TO BLUE-PURPLE PSL (390-400

nm)

Trang 48

AFTER SCANNING THE CR

PLATE NEED TO BE CLEARED

FROM RESIDUAL SIGNAL

PLATE EXPOSED TO HIGH

INTENSITY SODIUM VAPOR

OR FLUORESCENT LIGHT

Trang 50

Typical image plate can be reused thousands of times

Trang 51

IMAGE DISPLAY

Trang 52

IMAGE CHECK-INDEX???

Trang 53

IMAGE

MANIPULATION-WINDOWING

Trang 54

PICTURE DESTINATION -PACS

Trang 55

PACS

Trang 56

CR CHARACTERISTIC CURVE

► VERY WIDE LATITUDE

Trang 57

CR SYSTEMS

► LESS RADIOGRAPH REPEATS

Trang 58

CHARACTERISTIC CURVE

LATITTUDE OF CR LATITTUDE OF FILM

Trang 59

IMAGE CAN BE SALVAGED WITH

CR

► 500% OVEREXPOSURE

► 80% UNDEREXPOSURE

USE ADEQUATE KVP

Trang 60

HARD COPY GENERATED IN

LASER PRINTER

Trang 61

CR SYSTEM EFFICIENCY

CHARACTERISTIC

► QDE- QUANTUM DETECTION EFFICIENCY

NO NEED FOR SCREENS

Trang 62

THE MEASURE OF THE EFFICIENCY OF

CR SYSTEM TO CONVERT REMNANT

X-RAYS TO USEFUL IMAGE SIGNAL

Trang 63

PATIENT DOSE

Trang 64

DIGITAL RADIOGRAPHY

Trang 65

ELECTRONIC DETECTORS

► INDIRECT TECHNIQUE CLASS

► DIRECT TECHNIQUE CLASS

Trang 66

INDIRECT TECHNIQUE

► INDIRECT CONVERSION FLAT PANEL DETECTORS

► INDIRECT CONVERSION CHARGED COUPLE DEVICE DETECTORS

Trang 67

INDIRECT CONVERSION FLAT

PANEL DETECTORS

► USE THIN LAYERS OF SILICON COMBINED WITH THE ARRAYS OF PHOTODIODES P.D IS COATED WITH CESIUM IODIDE OR RARE EARTH

MATERIAL- WHEN STRUCK BY REMNANT X-RAYS THEY EMIT LIGHT

ADVANTAGE: HIGH QDE DISADVANTAGE: LIGHT DIVERGENCE

Trang 68

PHOTODIODE

Trang 70

INDIRECT CONVERSION CHARGE COUPLED DEVICE

DETECTORS

► USE CCD ARRAYS TO RECORD THE VISIBLE LIGHT EMITTED BY A SCINTILLATION CRYSTAL THAT HAS ABSORBED THE ENERGY AN OPTICAL COUPLING SYSTEM DEMAGNIFIES THIS LIGHT AND SENDS IT

TO THE CCD ARRAY.

Trang 71

OPTICAL COUPLING

Trang 72

CCD

Trang 73

DIRECT TECHNIQUE

► DIRECT CONVERSION FLAT PANEL DETECTORS

Trang 74

DIRECT CONVERSION FLAT

ADVANTAGE: NO LIGHT DIVERGENCE

Trang 75

READY - The 17"x17"

amorphous selenium-based

sensor is electrically sensitized

                    

EXPOSE - Incident X-ray photons are

converted into electron-hole pairs A

transverse electric field separates the

electrons and holes and creates a

charge image precisely

corresponding to the X-ray image.

READOUT - The microplasma line

scanner sweeps across the sensor in

less than 2 seconds causing readout

of the charge image and sensor reset.

Trang 78

PORTABLE????

Trang 79

PORTABLE CT???!!

Trang 81

4K MONITORS USED IN MAMMOGRAPHY

AND CHEST RADIOGRAPHY

Trang 83

PIXEL

Trang 84

        

The formation