Trang 1HƯỚNG DẪN SỬ DỤNG
MÁY XÉT NGHIỆM NƯỚC TIỂU
NUA-50
v.160703
Trang 2I
A (Stand-by) MENU
NO code
STRIP test
START : Start test
B
(Stand-by MENU
↔ ↕ SELECT 1 System SELECT System :
Mode (Continuous) (Once) EDIT
↔ OK Speed Normal 60s Fast 30s EDIT
↔ OK Sensit Default Langauge ENG
EDIT OK Detecter ON OFF
EDIT OK
CENCEL System 2 Contrast LCD SELECT Menu m -] OK 3 Initialize SELECT Menu
DATA
OK
4 DataBank 5000 test test
5 RS232 ON
6 Info Xem thông tin
7 Printer Menu Printer SELECT
↔ OK 8 Date & Time Menu Date & Time
SELECT OK 9 GreyMeas
Trang 31
Urine Analyzer NUA- 500 Service Manual
NEOMEDICA d.o.o.
Trang 42
TABLE OF CONTENTS
1 Overview 5
2 Common malfunction and remedy 5
2.1 Malfunctions due to improper installation environment 5
2.2 Common malfunction and remedy 6
3 Malfunction and Remedy 8
3.1 There is no response when turning on the power 8
3.2 The instrument can not get to the ready screen when turning on the power (Self-checking not pass) 9
3.3 There is no display on the screen 10
3.4 Push bar doesn’t move right when a strip is put on the platform 12
3.5 Push bar moves when there is no strip on the platform 13
3.6 The results could not be printed by inner printer 14
3.7 The keys don’t work well 15
3.8 Transfer electrical motor doesn’t work well 17
3.9 Detecting electrical motor doesn’t work well 18
3.10 Strip transfer mechanism works abnormal 19
3.11 LED lights don’t work when detecting (The light value is weak) 20
3.12 Serial port output/input doesn’t work 21
3.13 Parallel interface output doesn’t work 21
3.14 USB output/input doesn’t work 22
3.15 Bar code input doesn’t work 22
3.16 “T-1”is displayed on the screen when turning on the power (Self-checking error of the detecting motor.) 23
3.17 “T-2”is displayed on the screen when turning on the power (The fixed platform doesn’t be put in position or the limit circuit of fixed platform is damaged.) 23
Trang 53
3.18“T-3”is displayed on the screen when testing (The strip doesn’t be put in the right position.) 24 3.19 “T-4”is displayed on the screen when testing (No strip is sensed or light value
is too weak.) 25 3.20 “T-4”is displayed on the screen when turning on the power (Light value is abnormal.) 25 3.21“T-5”is displayed on the screen when turning on the power (Initialize error of inner communication or bad contact of 6-core cable for inner communication.) 26
Version:16.07
Trang 64
Product information
Product model: NUA-500
Product name: Urine Analyzer
Manufacturing site: NeoMedica DOO Bul Cara Konstantina 82-86, 18000 Niš, Serbia
(1) Operating this device following official NeoMedica Operating Manual
(2) Assembling, upgrading, resetting, and repairing are performed by NeoMedica’s authorized personnel
(3) Product storing environment, operating environment, and electrical environment are
as described in this manual
(4) Product serial number and labels are intact to verify the product identity as manufactured by NeoMedica
(5) Damages not by miss-use, or accidental dropping
Free services apply to those products with applicable items within warranty period For those beyond the description of warranty conditions, NeoMedica will charge for service Any returned goods to NeoMedica for service, customers should pay for the transportation and any applicable customs fees
Return Procedures
If return is necessary, take the following steps:
1 Obtain return merchandise authorization from NeoMedica Customer Service Department Inform NeoMedica of the serial number and mark this serial number on the cartons If the serial number cannot be recognized, the return cannot be accepted Describe briefly reasons for return
2 Freight Charge: customer is responsible for freight charges (including customs) for any returns
Trang 71 Overview
NUA-500 urine analyzer is manufactured under strict supervision This is a kind of precise analyzer, improper operation or any damaged component may affect the results or make damage to the instrument If an operational or instrument problem occurs, an error number may display on the Analyzer screen This section of the manual lists various errors and messages, along with probable causes and recommended remedies to quickly and easily correct the problem If the problem persists after following the recommended remedy, consult Table 2 for common malfunction and remedy
Caution:
Be sure to turn off the power supply before opening the instrument cover or replacing parts Beware of the high voltage inside
2 Common malfunction and remedy
2.1 Malfunctions due to improper installation environment
Consult table 1
Table 1
1 Direct sunshine Strip sensor failure
2 Abnormal temperature Erroneous test result、abnormal
LCD
3 Volatile voltage Device failure
4 Strong electromagnetic interference Device failure
5 Dust, moist and corrosives gas Impaired precise parts
6 Uneven ground of worktable Deviation of strip position
Trang 86
2.2 Common malfunction and remedy
Consult Table 2 for common malfunction and remedy
Table 2 common malfunction and remedy
1 There is no response when
turning on the instrument.
1 There is some problem about the input voltage
2 Power connector is loose or defective
3.1
2
The instrument can not get to
the ready screen when turning
on the power
1 The display connection cable is defective
2 The power circuit is defective
3 The main board is defective
4 Electrical motor system is defective
Push bar doesn’t move right
when a strip is put on the
platform
1 Strip detector is defective
2 The electric motor is defective
3.4
5 Push bar moves right when
there is no strip on the platform
1 Strip sensor is triggered by hand or other foreign object by accident
2 Fixed platform is dirty
3 The strip sensor is defective
3.5
6 Test results cannot be printed
out by the internal printer
1 The printer is OFF
2 The printer is short of paper
3 Printer connection cable is loose or defective
4 Damaged printer
3.6
7 The keys don’t work well 1 The keys are defective 2 The main board is defective
3 Connection cable is loose or defective
3.7
8 Transfer electrical motor doesn’t
work well
1 The socket of motor is loose
2 The electrical motor is defective
3 The main board is defective
3.8
9 Detecting electrical motor
doesn’t work well
1 The socket of motor is loose
1 The electrical motor is defective
2 The main board is defective
3.9
10 Strip transfer mechanism works
abnormal 1 The socket of motor is loose 2 The electrical motor is defective 3.10
Trang 97
3 The main board is defective
11 LED lights don’t work when
detecting 1 The connection cable is loose 2 The LED lights are defective 3.11
12 Serial port output/input doesn’t
work 1 The setting about serial port is wrong 2 The connection cable is loose 3.12
13 Parallel interface output/input
14 USB output/input doesn’t work 1 The setting about serial port is wrong 2 The connection cable is loose 3.14
15 Bar code input doesn’t work
1 The setting about serial port is wrong
2 The connection cable is loose
3.15
16 “T-1”is displayed on the screen
when turning on the power
Self-checking error of the detecting
17 “T-2”is displayed on the screen
when turning on the power
The fixed platform doesn’t be put in position or the limit circuit of fixed platform is damaged
20 “T-4”is displayed on the screen
when turning on the power Light value is abnormal 3.20
21 “T-5”is displayed on the screen
when turning on the power
Initialize error of inner communication
or bad contact of 10-core cable for inner communication
3.21
Note:
Make sure the version of new board is consistent with the replaced board when replacing a circuit board
Trang 108
3 Malfunction and Remedy
3.1 There is no response when turning on the power
Check if the power cord is well
connection
Re-plug the power cord
No Yes
Check if the supply voltage is normal No Make sure the voltage and
socket is normal
Yes Check if the fuses are normal No Replace the fuses
Yes Check if power input socket is normal No Replace the power input
socket
Yes Check if power switch is connection No Replace the power switch
Yes Check if the fuse socket is normal No Replace the fuse socket
Yes Check if the output voltage of switch
power is normal
No Replace the switch power
Check if +5V pin voltage of J15 in main
board is normal See table 3
Re-plug or replace the cable in the socket
Yes
No Replace U5 (LM2576) on main
board
Yes Check if +12V pin voltage of J15 in
main board is normal See table 3
No
Yes Check if the voltage of TP1 on main
board is +3.3V
No Replace U2 (AMS1117) on main
board
Trang 119
Table 3 Test voltage of J15 pins (Reference value)
Voltage value +5V +5V G G +12
Range of voltage value: X(1±4%)V
3.2 The instrument can not get to the ready screen when turning on the power (Self-checking not pass)
Yes
No
The instrument can not get to the ready screen when turning on the power
(Self-checking not pass)
Turn on the power again to check if the instrument can get to the ready screen
Check if all three electrical
motors don’t move at all
Check if supply voltage is normal No
See 3.1
No
Check if “T-1” is displayed Yes See 3.16 Repair welding or replace
U9 in the main board (U9 needs to be programmed)
No Check if “T-2” is displayed Yes See 3.17
No Check if “T-5” is displayed Yes See 3.21
No Repair welding or replace U9 in
the main board (U9 needs to be
Trang 1210
3.3 There is no display on the screen
When turning on the power, electrical motors work
and buzzing, but there is no display
Check if the screen is dark or blue Dark
1 Remove the display part form the main board, test the voltage of J4 (14, 26, 27 pin) on main board
2 If the voltage is not +5V, see 3.1
3 If the voltage is +5V, replace the display part
Blue
Check if the connection between J4
on main board and 30-core cable
on display interface board is good
No
Reconnect the cable
Yes
Adjust the potentiometer RV1 to
check if there is any change in the
display
Yes Readjust the value of potentiometer
RV1
No
Adjust the potentiometer RV1 to
see if the voltage could change
from -12V~-22V
No Replace the potentiometer RV1 or
U10(MC79L12)
Check if the supply power and
output power voltage of U7, U8 is
normal
Yes
No Repair welding or replace U7, U8, U9
(U9 needs to be programmed) on main board
Yes Replace the LCD display
Trang 1311
Figure 1 Main Board
Trang 14put a strip close to the detector
The position of strip sensor is too high, please low the strip sensor
or replace the LED
Yes
No Check if the scan LED light is normal
Check if the connection between strip
sense board to main board (J1) is good
No
No
Reconnection
Yes Check if the LED has any light No Replace the LED
Yes
Initialize error of strip sensor Turn off
the power and wait 30s, and then turn
on the power again
Problem Persists
1 Replace U1 on the strip sensor or
replace the sensor (U1 needs to be
programmed)
2 Replace the strip sensor board
Yes
Trang 1513
3.5 Push bar moves when there is no strip on the platform
Check if the push bar is trigger by
hand or other object accidentally
Yes Do not put a hand or any other
foreign object on the platform
No The fixed platform is too dirty which
results in error of initialization
Yes
Turn off the power and then clean the fixed platform
No Turn off the power, wait 30s, and
then turn on the power again
Problem Persists The strip sense board is defective
Replace or repair the strip sense
board
Figure 2 Strip sensor board
Trang 1614
3.6 The results could not be printed by inner printer
The results could be displayed on the screen but
could not be printed by inner printer
Check if the printer setting is set to be “OFF”
Yes
Set the printer to be “OPEN”
No Check if the printer is short of paper
Yes
Install a new roll of paper
No Check if the connection (J4) between printer and
print board is loose
Yes
Re-plug the cable
No Check if the connection (J5) between main board
and print board is loose
Yes
Re-plug the cable
No Check if the print board
is defective
Yes
Check if paper feeds when turning on the power
No
1 Repair welding or replace U3 (LB1838)
2 Repair welding or replace
Yes
Test if the voltage of VP1 is 8V when the printer is printing and feeding paper No Repair welding or replace
U5(IRF7404) on the print board
1 Repair welding or replace U2(STM8S105K4T6C) (U2 needs to be programmed.)
2 Replace the print board
Yes
No
Repair welding or replace J5,
U9 on main board (U9 needs
to be programmed.)
Trang 1715
3.7 The keys don’t work well
LCD could display but the keys are not sensitive
Open the display cover to check if J3 from
keys to display interface board is well
Problem Persists The keys are defective Replace the keys
Yes
Figure 3 Printer Board
No
Reconnect J4
Trang 1816
Figure 4 Display part
Figure 5 Interface board
Trang 1917
3.8 Transfer electrical motor doesn’t work well
Transfer electrical motor doesn’t work well
Check if the transfer
electrical motor only
works half a circle when
turning on the power
Yes Check if the connection J1 between photo-electricity limiter and strip sensor board is good
No Reconnect J1
No
Check if the transfer
electrical motor works all
along when turning on the
power
Yes
Check if the photo-electricity limiter is defective
Yes
Yes Replace the photo-electricity limiter
Check if the transfer
electrical motor doesn’t
work at all when turning on
the power
No
Yes
Check if the connection J2 between
the motor lead and strip sensor board
is good
No
Reconnect J2
Yes Check if the motor is defective Yes Replace the motor
No
1 Repair welding or replace U2 and U1 on strip sensor board (U1 needs to be
programmed)
2 Repair welding or replace U9 (U9 needs to be programmed)
Check if the connector J1 on main
board is well connection
Yes
No Reconnect J1 on main board
Trang 20Detecting electrical motor doesn’t work well
Check if the motor
only moves towards
outside when turning
on the power
Check if the connection J10 between photo-electricity limiter and main board is good
Reconnect J10
No Check if the motor
only moves towards
inside when turning on
the power
Yes
Check if the photo-electricity limiter is defective
No
1 Repair welding or replace U19 and U9 on the main board (U9 needs to be programmed)
2 Replace the main board
No
Check if the motor doesn’t
work when turning on the
power
Yes Replace the photo-electricity limiter
No
Yes Check if the connection J11 between the
motor lead and the main board is good
No
Reconnect J11
Yes Check if the detecting electrical motor is
defective
Yes
Replace the motor
No
1 Repair welding or replace U9, U19 (U9 needs to be programmed)
2 Replace the main board
Trang 2119
3.10 Strip transfer mechanism works abnormal
Strip transfer mechanism works abnormal
Check if the floating panel shakes when
works two circles when
turning on the power Yes
Check if the connection J13 from photo-electricity limiter to the main board is good
No Reconnect J13
No Check if the electrical
motor works all along
when turning on the
power
Yes Check if the photo- electricity limiter is defective
No
Check if the electrical
motor doesn’t work at all
when turning on the
power
No
Yes Check if the connection J14 between the motor lead
and the main board is good
No
Reconnect J14 Yes
Check if the detecting electrical motor is defective Yes
Replace the motor
No
1 Repair welding or replace U9, U21 on the main board (U9 needs to be programmed)
2 Replace the main board
Trang 2220
3.11 LED lights don’t work when detecting (The light value is weak)
LED lights don’t work when detecting
Get to the adjust screen to open the LED light
Check if all the LED
lights don’t work Yes
Check if J1 on amplified board is well connection
Figure 6 Amplify Board
Check if the 20-core cable from J2 on interface board
to J17 on main board is well connection
Trang 2321
NOTE:
The light value needs to be readjusted after replacing the LED
The way to get to the adjust screen to open the LED: Press F2 for 3s in the menu screen, when “di…” sound is heard, press F1, F2, MENU key in turn, the “Adjust” option will list on the menu Select “Adjust” option to get to the adjust screen, then select LED control to open or close the LED
3.12 Serial port output/input doesn’t work
3.13 Parallel interface output doesn’t work
1 Check if the connection is good
2 Check if the setting is all right
No 1 Replace the connection cable
2 Set the setting to be right according operating manual
Yes Check if the 10-core cable from J9 on
main board to J4 on interface board
is good
No
Reconnect or replace the cable
Yes Check if the interface board is
defective
No Repair welding or replace U9, J9 (U9
needs to be programmed.) Yes
Repair welding or replace D1, U4 on interface board
Check if the 10-core cable from J9 on
main board to J4 on interface board
is good
No Reconnect or replace the cable
Yes Check if the interface board is
defective
No Repair welding or replace U9, J9 (U9
needs to be programmed.) Yes
Repair welding or replace D2, U2, Y1 on interface board (U2 needs to be programmed.)
Trang 2422
3.14 USB output/input doesn’t work
3.15 Bar code input doesn’t work
USB output/input doesn’t work
Check if the 10-core cable from J9
on main board to J4 on interface
board is well connection
on interface board
Repair welding or replace U9, J9
on main board (U9 needs to be programmed.)
No
Check if the bar code is set to be
open
No
Set the bar code to be open
Check if the 10-core cable from J9
on main board to J4 on interface
board is well connection
Yes
No Reconnection or replace the
cable
Yes Check if the interface board is
defective
No Repair welding or replace U9, J9
on main board (U9 needs to be programmed.)
Yes Repair welding replace D2, U2, Y1
on interface board (U2 needs to be
programmed)
Bar code input doesn’t work
Trang 25Still TROUBEL-1
Check if the fixed platform is
installed in the right position
No Install the fixed platform in the right
off the power and wait 30s and then
turn on the power
Still TROUBEL-2
Replace or repair welding U9 on the
main board
Check if the motor could work normally when
turning on the power
No See 3.9 for details
Yes Initialize error of detector Turn off the power
and wait 30s and then turn on the power
Repair welding or replace U9 on main board
Check if the floating panel could be
lifted normally
No
Install the floating panel correctly
Yes
Trang 26No Put the strip in the right position as
the direction in operating manual
Retest the sample
Check if the strip type is right
Yes
Yes
No
Put the right strip type
Check if the light spot is in the
middle of strip when detecting
No Correct the light spot by adjust the
screws in the guide rail
Yes Get to adjust screen to check if the
light values of 550, 620, 720 are
between 2800~3400
Still TROUBLE-3
1 Repair welding or replace U1 on amplified board
2 Repair welding U4, U15, U20, U25 and U9 on main board
Yes Repair welding or replace U25, U15,
U9 on main board or replace the main board (U9 needs to be programmed)
on the white block of fixed
platform
No Get to motor screen to adjust the
motor steps to make the light spot shine on the white block
Yes
Check if the maximum and
minimum light value could be
adjust between 2800 and 3400 in
the adjust screen
No Readjust the maximum and
minimum light value
Trang 27No There is some problem about strip
transmission See 3.4 and 3.10 for details
Yes Check if the brightness of LED is
normal
No See 3.11 for details
Yes Check if the light spot could shine on
the white block of fixed platform No Get to motor screen to adjust the
motor steps to make the light spot shine on the white block
Yes See 3.18 for details
Check if the detecting motor works
normally when turning on the power
No See 3.9 for details
Check if all LED could light normally
when turning on the power
Yes
No See 3.11 for details
Yes Check if the light spot could shine on
the white block of fixed platform
No Get to motor screen to adjust the
motor steps to make the light spot shine on the white block
Yes See 3.18 for details
Trang 2826
3.21“T-5”is displayed on the screen when turning on the power (Initialize error of inner communication or bad contact of 6-core cable for inner communication
Check if the connection between the
main board to strip sensor board is
good
Reconnection or replace the cable
Check if the main board is defective
or the strip sensor board is
defective
No
Yes
Strip sensor board
Repair welding or replace U1 on strip sensor board (U1 needs to be programmed)
Main board Check if the patch diodes D3, D4 on
main board are defective
Yes
Replace D3, D4
No Repair welding or replace U9 (U9
needs to be programmed)
Trang 29MÁY XÉT NGHIỆM NƯỚC TIỂU TỰ ĐỘNG HOÀN TOÀN MODEL: NUA-1500 (NEOMEDICA-NƯỚC SERBIA)
1 Giới thiệu chung
Máy xét nghệm nước tiểu Model: NUA-1500 là máy hoàn toàn tự động và dễ vận hành Tất cả các việc người vận hành phải làm là nạp que đo, đặt bệnh phẩm và nhấn START, tất cả việc còn lại
sẽ do NUA-1500 thực hiện Máy NUA-1500 có thể thực hiện liên tiếp đến 110 bệnh phẩm Bệnh phẩm cấp cứu có thể được đưa vào bất cứ lúc nào như là chạy mẫu bằng tay Trình tự thực hiện của máy là chuyển bệnh phẩm, hút bệnh phẩm, nhả bệnh phẩm trên que đo, rửa kim hút, lấy que đo, nhận diện màu của bệnh phẩm… Hệ thống sử dụng 2 loại que đo là NEOMEDICA 11F và 12F có thể thực hiện đo được đến 14 thông số Tất cả kết quả được in ra bởi máy in gắn sẵn hoặc máy in ngoài
1.1 Đặc tính
1) Có 5 mode chạy test bao gồm: bằng tay, STAT, QC, Calibration, và Online
2) Với các que đo NEOMEDICA 12F và 11F bạn hoàn toàn có thể kết hợp để đạt được toàn bộ
14 thông số
3) Vận hành tự động hoàn toàn: chuyển mẫu, hút mẫu, nhả mẫu, vệ sinh, lấy que đo bơm nước thải, nước cấp…
4) Bộ phận chuyển mẫu: dễ dàng tháo lắp để vệ sinh, bảo dưỡng
5) Cách nạp que đo: Cho toàn bộ hộp vào
6) Cách lấy que: Xoay tròn để nạp que
7) Chuyển que: các que đo được vận chuyển để đo liên tục
8) Nhận diện que đo: kiểu que đo được phận biệt bằng cách nhận diện tại điểm cuối cùng của que đo
9) Nhận diện bệnh phẩm: có lắp đầu đọc mã vạch và có thể nhận diện ID của bệnh phẩm
13) Chức năng khuấy: Chức năng khuấy mẫu
14) Thả mẫu: Thả dạng giọt chính xác tránh tình trạng nhiễm chéo giữa các test với nhau
15) Rửa kim hút: Kim hút được rửa cả 2 mặt bên trong và bên ngoài
16) Lưu dữ liệu: Dung lựng bộ nhớ đến 4GB (khoảng 300.000 kết quả) và có thể lớn hơn theo yêu cầu đặt hàng
17) Xem lại kết quả: Bạn có thể xem lại lịch sử test và các kết quả QC
18) Màn hình LCD cung cấp cho người dùng các thông báo (trạng thái, kết quả) và kết quả có thể được in bằng máy in gắn sẵn hoạc máy in ngoài, Kết quả có chứa cả thông số màu sắc của bệnh phẩm
19) Que thử đã sử dụng: Được chuyển đến hộp đựng que thải một cách tự động
Trang 3022) Không thao tác: hệ thống có thể tự động làm việc liên tục 110 bệnh phẩm mà không cần thêm một thao tác nào khác
23) Chức năng báo lỗi: Khi hệ thống có lỗi, âm thanh báo lỗi được phát ra
1.2 Đặc tính kỹ thuật
1) Nguyên lý đo: Phản xả quang học
2) Các bước sóng: 720, 620, 570, 550, 470 nm
3) Thể tích mẫu cần thiết: nhỏ nhất là 2ml
4) Công suất: 300 test/giờ
5) Số rack tối đa: 11 rack
6) Số ống tối đa trên rack: 10 ống
7) Số que tối đa có thể nạp: 200 que
8) Hiển thị: Màn hình LCD màu cảm ứng
9) Interface: RS232, LPT, LAN, USB, đầu đọc barcode
10) Máy in: Máy in gắn sẵn hoặc máy in ngoài qua cổng USB và LPT
11) Thúng rác: chứa được 500 que đã sử dụng
12) Phương pháp rửa: Bơm – hút rửa bằng áp suất
13) Công suất tiêu thụ điện: 250VA
14) Kích thước: 630 x 626 x 570 (dài x rộng x cao)
Trang 31Mặt trước máy
1 Máy in và nắp : in kết quả
2 Màn hình: Hiển thị giao diện và các thông tin test
3 STAT rack: Cho các mẫu chạy STAT
4 Hệ thống chuyển mẫu: đặt rack và ống mẫu
5 Nắp bên phải: Để nạp các que đo, mở ra để nạp
6 Phím tắt mở nguồn cho máy
Phía sau máy