Trang 1How to Set Up your
GX Series Microlog
Trang 2Module Objective(s)
To instruct the new GX Series Microlog User on how to set the Microlog’s “system” settings and “global
measurement” parameters.
Trang 3Two General “Setup” Screens
Config uration Screen for Global Measurement Settings
Setup Mode for System Settings
Trang 4Setup Mode - “System” Settings
Trang 5Setup Screen Fields
• Right arrow to open field.
• Up/Down arrow to select
new setting.
• Left arrow, or Enter
button to accept change
and close field.
apply new settings
Trang 6“Setup” Fields
Route Memory – Internal or Card
Timeout - 5 min , 10 min , 15 min , 20 min , or Off
IrDA – Disabled or Enabled (for use with ActiveSync operations) Baud Rate – for serial communications / must match MA
Flow Control – Off or Hardware
Default Units – English or Metric (e.g., in/sec or mm/sec)
Date Format – DD/MM/YYYY ; MM/DD/YYYY ; YYYY/MM/DD
Time Zone – for use with ActiveSync operations
Date / Time - set current date and time
Trang 7“Setup” Fields (continued)
Memory – Delete All Routes or Initialize
• Delete All Routes - preserves your system preferences, but deletes ROUTE and non-ROUTE measurements and data
• Initialize - erases all stored data, but preserves the system settings and parameters and also refreshes the active memory if the Microlog is experiencing memory
related problems
Purge Cache – On or Off
• On - quickest resetting of ROUTE data, but requires more Microlog memory
• Off – select Off if memory issues exist
Route Font – Big or Small (Small setting displays hierarchy level
icons)
Language - select language for GX Series Microlog displays
Trang 8Configuration “Global Measurement” Settings
both ROUTE and Non ROUTE
provide access to Config settings
Trang 9“Config” Fields
Freq Units – Hz , CPM , or Orders
Fast Averaging – On or Off
• On – display is updated as often as possible
• Off – only final averaged spectrum displays
Auto Advance – On or Off (display spectrum during ROUTE
collection?)
• On – quickly advance through ROUTE data collection without displaying spectral
results
• Off – display the resulting spectrum for each spectral measurement
Aquisition – Single Shot or Continuous (stop measurements after averaging process, or repeat measurements until Enter is pressed)
Trang 10“Config” Fields (continued)
Input Range – Auto or Manual (set amplitude range scaling method)
• Auto –automatically scales optimal amplitude range for the input signal
• Manual – applies the measurement’s Full Scale setting as in MA
Autoranging – Conservative , Aggressive , or Fixed
• Conservative – slower data collection, with higher quality data
• Aggressive – provides faster data collection, with greater potential for settling error
• Fixed – no autoranging is performed Input range fixed according to the transducer type and sensitivity
Machine Speed – Normal or Slow
• Normal – select for measurements on shafts rotating at or above 600 RPM.
• Slow – select if ski slope or settling problems occur on shafts rotating below 600
Trang 11“Config” Fields (continued again)
Settling Delay – if ski slope problems occur, a settling delay of 2 or 3 seconds is recommended
High Pass Filter – Fixed or Variable
• Fixed – the High-pass Filter setting is always 1.1 Hz.
• Variable – the Microlog uses the Low Frequency Cut-off setting from
Machine Analyst software.
Route ICP – Never or Always (specified for your sensor, disable or enable sensor ICP)
Trang 12“Config” Fields (continued again still)
Memory – Purge , Delete ROUTE , Delete all ROUTEs , or Initialize
measurements.
system settings.
NonROUTE POINTs stored within the Microlog Preserves system settings, and refreshes memory if the Microlog is experiencing memory-like problems.
Fill Spectrum – On or Off
Trigger Level – Automatic or Fixed
setting in the NonROUTE mode’s User setup screen applies.
Trang 13Quick Review