The air sample enters through the air inlet, is drawn into the sensor by way of the narrow annulus where it comes into intimate contact with the solution contained on the electrode suppo
Trang 1Full Terms & Conditions of access and use can be found at http://www.tandfonline.com/action/journalInformation?journalCode=uawm16
Journal of the Air Pollution Control Association
ISSN: 0002-2470 (Print) (Online) Journal homepage: http://www.tandfonline.com/loi/uawm16
Nitrogen Dioxide Detection Using a Coulometric Method
Royal E Rostenbach & Robert G Kling
To cite this article: Royal E Rostenbach & Robert G Kling (1962) Nitrogen Dioxide Detection Using a Coulometric Method, Journal of the Air Pollution Control Association, 12:10, 459-463, DOI:
10.1080/00022470.1962.10468113
To link to this article: http://dx.doi.org/10.1080/00022470.1962.10468113
Published online: 19 Mar 2012.
Submit your article to this journal
Article views: 389
View related articles
Trang 2NITROGEN DIOXIDE DETECTION Using a COULOMETRIC METHOD
ROYAL E ROSTENBACH, Former Research Director, and ROBERT G KLING,
Experimental Chemist, Mast Development Company, 2212 East 12th Street, Davenport, Iowa
Instrumentation for the
quan-titative measurement of nitrogen
di-oxide is used today in areas of smog
control and industrial hygiene It is
well known that nitrogen oxides are
emitted from motor vehicles' exhaust,
chimneys, stacks, and other combustion
processes By photochemical and
ther-mal reactions nitrogen dioxide combines
with ozone and organic chemicals
resulting in unpleasant atmospheric
smog mixtures In the field of
propel-lants nitrogen dioxide is a fuel oxidizer
As a low cost selective oxidizer for
industry, nitrogen dioxide (or nitrogen
tetroxide) is currently gaining
recogni-tion with a resultant increase of
produc-tion in tank car quantities The
toxic level presence of this gas in any
working area is an industrial hygiene
problem, and its unwanted presence in
any area may present an air-pollution
problem
The adaptability of the Mast Model
724 Series Meter for the monitoring of
nitrogen dioxide has been
demon-strated Wartburg, Brewer, and
Lodge1' 2 have shown that the Meter
operates extremely stable in the range
of 0 to 3 ppm of nitrogen dioxide
when other oxidizing agents are
ab-sent Van Nattan, Drake, et al.,3
have demonstrated the adaptability of
the'Meter as a monitor up to 25 ppm of
nitrogen dioxide Within the past
year the authors have demonstrated
that the Meter will sense accurately
0-5000 ppm of nitrogen dioxide
The Mast Nitrogen Dioxide Meter
is based upon a coulometric system
This is the same one which has been so
successful in the sensing of ozone
Only a brief description of the sensing
system is included here as it has been
reported upon heretofore.4 Figure 1
shows a typical construction of a
micro-coulomb sensor including the basic
components essential.to operation A
chemical solution containing the proper
amounts of reagents is pumped into the
sensor The solution flows in a thin
film down the electrode support, upon which are wound many turns of a fine platinum wire cathode and a single turn of a platinum wire anode, and is deposited in the waste reservoir The air sample enters through the air inlet, is drawn into the sensor by way
of the narrow annulus where it comes into intimate contact with the solution contained on the electrode support, and exits by means of air pump "P."
A small potential is supplied across the cathode and anode by a battery
"B." The current flow is measured by
the microammeter, or a strip chart recorder is used to measure the voltage drop across a range resistor placed
in the circuit
In this coulometric system the current
is measured and this related to the concentration of nitrogen dioxide
When a sensing solution containing potassium iodide is used, iodine is released in solution by the chemical reaction of the nitrogen dioxide Sim-ilarly when a bromide solution (with
no iodide present) is used, bromine is released The current flows, and is measured, only when released iodine or bromine is present in the sensing solution
Some of the typical reactions taking place in the bromide sensing solution are:
NO 2
2NoO4
NO + H2O + Br2 (1)
2 HNO + 2 NO + Br2 (2)
* Presented at the 55ttt Annual Meeting
of APCA, Sheraton-Chicago Hotel, May
20-24, 1962, Chicae-o, Illinois
At normal atmospheric conditions of temperature and pressure the gas present is primarily nitrogen dioxide
Hence, the first reaction is more typical
of what is taking place Nitrogen dioxide and nitrogen tetroxide are usually considered together There is a rapid disassociation and association which may be written as a reversible reaction:
N2O4 ?± 2 NO2 The reaction is dependent upon tem-perature and pressure Nitrogen
tetrox-ide is a colorless gas while nitrogen dioxide is a brown gas
The sensing solution system feeds the solution to the sensor at the rate of approximately 2.5 ml per hour This
is done by a micro-feed pump The pump is driven by a synchronous motor operating a cam which rotates one revolution every hour The bellows
of the pump is actuated by the cam movement An elastic diaphragm sep-arates the liquid in the bellows from the solution being pumped The solution moves through the pump chamber and
is controlled by elastic check valves One complete pumping cycle takes one hour which is governed by the cam and motor noted above Reservoirs, 125
ml capacity, are provided for the new and used sensing solution because the solution is used only once
In the pumping cycle, fluid in the bel-lows is displaced, thus filling and ex-panding a flexible diaphragm The expansion of the diaphragm in turn displaces solution in the solution cham-ber, which flows through the outlet check valve During the refill portion
of the cycle the spring loaded bellows draws liquid from the diaphragm, causing it to collapse This in turn draws solution from the solution reser-voir through the inlet check valve
Fig 1 Microcoulomb sensor.
Trang 3LLJ
LU
a
o
u
a:
U
o
CO
Z
LU
0 1000 2000 3000 4000 5000
NITROGEN DIOXIDE CONC ( P P M / VOL.)
Fig 2 High concentration nitrogen dioxide measurement with a 1 0 % potassium bromide solution.
The air sample is drawn continuously
through the microcoulomb sensor by a
small precision constant-volume piston
pump The pump is driven by a 300
rpm synchronous electric motor The
pump is adjusted to operate at 140 ml
per min capacity
The instrument requires only
115-volt, 60-cycle power supply and a small
1.34-volt mercury battery The ac
cur-rent is used by the motor for the
solu-tion pump and air pump The battery
supplies a dc potential to the sensor
The microammeter is in series with
the sensor and indicates the current
flowing through the sensor When the
microammeter is by-passed, the elec-trical signal of the sensor can be con-ducted through a phone jack inserted into the socket of a remote recording meter In this way a continuous record
of the sensor's operation may be made
A battery-operated portable unit was developed for nitrogen dioxide measurements at locations where an external power supply is not available
This unit was designed to operate from two rechargeable nickel-cadmium bat-teries or from a 115-volt, 60-cycle power supply The batteries used were 12 volts each (Burgess CD29) and were connected in parallel
.3 4 5 APPLIED SENSOR VOLTAGE
Fig 3 Effect of sensor voltage on sensor current.
.6
A 12 VDC Veo rpm Haydon motor was used to drive the solution pump while drawing an average current of
10 ma A 12 VDC Amglo reed-con-trolled 300 rpm motor was used to drive the air pump while drawing an average current of 50 ma With a total drain of 60 ma, the two batteries, when fully charged, were capable of powering the unit continuously for up
to 16 hours An eight-hour minimum battery-powered period between charge periods was desired This was ob-tained with the two batteries More cells in parallel would have increased this period but also would have in-creased instrument size and weight somewhat
A simple charging circuit was con-structed with a small transformer and
a 500-ma silicon rectifier plus resistors and filter capacitors This charging unit delivered about 95 ma to the two batteries The unit could be operated while batteries were charging In such use, the two motors drew 60 ma current leaving 35 ma available for recharging the two Burgess CD-29 cells at the same time
The sensing cell, solution pump, and air pump used on the battery-powered unit are the same as those used on the Model 724 series ac-powered units and were described previously
Wartburg, et al.,1 and Lodge2 have reported upon the use of the Mast Model 724-1 Ozone Meter as an in-strument for the quantitative detec-tion of nitrogen dioxide For example, the contents of a photolysis chamber made up of hydrocarbons, aldehydes, ozone, and nitrogen oxides were passed through a coarse glass frit heated to about 200 °C Under these conditions the instrument measured nitrogen di-oxide in the range of 0 to 3 ppm in direct relation to the meter reading These quantities of nitrogen dioxide were confirmed by chemical laboratory determinations The response to nitro-gen dioxide on a molar volume basis was 11.6% of that to ozone Van Nattan, Drake, et al.,3 demon-strated the adaptability of the same model meter as a monitor up to 25 ppm of nitrogen dioxide when ozone was practically absent The foregoing investigations used a sensing reagent buffered to pH 7 with phosphates and containing two percent potassium iodide, five percent potassium bromide, and three percent ethylene glycol
Effenberger5 has shown that nitrogen dioxide has greater oxidizing efficiency
at lower sensing solution pH Gluckauf,
et al.,6 have reported work along similar lines on the reaction:
NO2 + 2H+ + 2 I - - *
H2O + NO + 12
In addition, an iodide solution is more readily oxidized than is a bromide
Trang 4solution in a coulometric system.
In other words, a given nitrogen dioxide
concentration will result in a
con-siderably smaller current when a
bro-mide solution is used in place of an iodide
solution With either solution, the
sensor output is proportional to the
nitrogen dioxide passing through the
sensing cell Therefore, in the effort
to sense and monitor nitrogen dioxide
in the range of 0-1000 and 0-2000
ppm, sensing solutions containing
bro-mide ions at a lower pH of 5.9 were
investigated These solutions were 20
and 10% aqueous potassium bromide
solutions
Operating characteristics of the
micro-coulomb sensor as used in the Mast
Meters are that:
(1) A fixed dc voltage is applied
across the sensor cathode and
anode;
(2) The chemical solution flows over
the electrodes at a fixed flow rate
(1.25 ml/hr typical for ozone
meters and 2.5 ml/hr typical for
nitrogen dioxide meters) ;
(3) The gas sample containing an
oxidant to be measured flows
through the sensing cell at a fixed
flow rate (140 ml/min typical for
ground level applications)
The above characteristics when
con-trolled at fixed conditions enable the use
of the Mast microcoulomb sensing cell
in accurate oxidant measurements
This is true when the
oxidation-reduc-tion reacoxidation-reduc-tion of the total oxidant
con-tent in the gas sample stream is 100%
complete during rapid passage of the
gas sample stream through the sensing
cell or when the oxidation-reduction
reaction is only partially complete,
even less than 1% complete, during
passage through the sensing cell For
example, the oxidation of an iodide
solution by ozone in the gas sample is
approximately 100% complete as the
reactions occur at a very rapid rate
Thus, with conditions of 0.2 VDC
sensor voltage, 1.25 ml/hr KI solution
flow rate, and 140 ml/min gas sample
flow rate, a concentration of five pphm
by volume of ozone results in a sensor
current of one ua Using the same KI
solution, flow rate, and sensor voltage at
the same gas sample flow rate, a
con-centration of about one ppm by volume
of nitrogen dioxide results in a one ua
sensor current The slower reaction
rate of NO2 required a concentration of
20 times that of ozone to obtain the
same one ua sensor current This
suggests that the NO2-KI reaction in
the sensing cell proceeded to only
five percent completion during the
passage of gas sample through the cell
Also, the oxidation of a bromide
solution by ozone in the gas sample is
only partially completed under similar
conditions With sensor voltage of
DC
o
to
Z
uu to
0 100 2 0 0 300 4 0 0 5 0 0
N I T R O G E N D I O X I D E C O N C ( P P M / V O L
Fig 4 Medium concentration nitrogen dioxide measurement with an acidified bromide solution.
0.56 VDC, a solution flow rate of 2.5 ml per hour of 10% KBr solution, and a gas sample flow rate of 140 ml/min, a concentration of about 125 pphm or 1.25 ppm by volume of ozone will result
in a sensor current of one ua With the same conditions of 0.56 VDC sensor voltage, 2.5 ml/hr flow rates of 10%
KBr solution, and 140 ml/min gas sample flow rate, a concentration of about 20 ppm by volume of nitrogen
dioxide in the gas sample will result in
a current of one ua Thus with a bromide solution as used in the above conditions the 03-KBr reaction ap-parently proceeds to about four percent completion and the NO2-KBr reaction apparently proceeds to about one quarter of one percent completion
To restate, with certain sensing cell factors controlled, the sensing cell output current is representative of oxidant
Fig.
0 2 4 6 8 10 12 NITROGEN DIOXIDE CONC ( P P M / V O L )
5 Low concentration nitrogen dioxide measurement with an acidified bromide solution.
461
Trang 5* 2 " '
n
Fig 6 Nitrogen dioxide sample delivery
apparatus.
concentration in the gas sample whether
or not the oxidant reacts to completion
with the sensor solution By choosing
proper conditions of sensor voltage,
solution composition, solution flow rate,
and gas sample flow rate, the Mast
microcoulomb sensor can be used in the
construction of instruments to
continu-ously measure various concentrations of
ozone or nitrogen dioxide
The original goal of this project was
to develop an instrument to
con-tinuously monitor NO2 over a range of 0
to 250 ppm by volume for unattended
periods of 30 days minimum The KI
solutions were not considered suitable
for measuring as great a concentration
as 250 ppm, although it was realized
that use of a very low (approximately 50
ohms) range resistor with a suitable dc
amplifier circuit would allow for
measure-ments as high as 200 ppm of NO2
Early success was obtained in
measur-ing higher NO2 concentrations with a
solution of 20% potassium bromide and
6% ethylene glycol in water This
solution was used for several months
while efforts were directed to securing
test data that could lead to a dependable
system capable of reproducible NO2
measurements
The slower reaction rate of the
bro-mide solution allows measurement of
high concentrations of oxidant; for
example, 5000 ppm/vol or more of
NO2 Figure 2 is a graph of nitrogen dioxide measurements in the 0-5000 ppm range obtained with a 10% potas-sium bromide solution
When gas measurements cause the magnitude of sensor current to vary, there is a possibility of large sensor currents causing a drop in sensor voltage of 10% or more It was desired
to employ a sensor voltage that could
be varied considerably without causing
a variation in magnitude of the sensor output signal for a given gas sample composition Figure 3 is a graph of data obtained by measuring the sensor output signal while sampling a stable gas sample from a large polyethylene bag and varying the sensor voltage
As shown by the plateau on this graph, varying the sensor voltage from 0.43 to 0.65 volt resulted in the same sensor output signal On this basis, a sensor voltage of 0.55 to 0.57 volt was chosen for applications of the microcoulomb sensor requiring a bromide solution containing no iodide ions All sub-sequent NO2 measurement investiga-tions were made with an applied sensor voltage of 0.55 to 0.57 volt This voltage was obtained usually from a small 1.34 volt mercury cell A zener diode-controlled dc voltage from an
ac source was also proved satisfactory
in one system tested
Some salting out was observed in the 20% KBr solutions A change of sensor solution to 10% KBr in water with no ethylene glycol added was made and there was no apparent change in performance Other solution composi-tions could be used in the microcoulomb sensor for nitrogen dioxide
A potassium bromide solution con-taining citric acid was found to have a sensitivity between that of the iodide and the bromide solutions Other solutions could probably have been investigated with probability of suc-cess The use of mercury salts, how-ever, is not recommended in this type
of sensor solution
Figure 4 is a graph of tests in the 0 to
500 ppm range using a modified acidic bromide solution of pH 2 This solu-tion consisted of 10% potassium bro-mide, 1% citric acid, and 3% ethylene
glycol Figure 5 is a graph of tests in the 0 to 10 ppm range using the same modified bromide solution It is be-lieved that improved low concentra-tion measurements in ranges of 0 to
1 ppm, and 0 to 10 ppm of nitrogen dioxide, could be obtained with a modified iodide solution
The matter of nitrogen dioxide sampling for calibration is a unique one The method was used originally by Wright Air Development Center and
is shown in Fig 6 A 100-ml glass syringe is held in an upright position with clamps A small rotor is mounted
on the plunger The rotor and plunger are turned by means of a stream of compressed air or gas When an appropriate volume of nitrogen dioxide gas is injected into the syringe holding some air, and when more air is drawn into an exact volume, a capillary tube is attached by means of an adapter to the normal opening of the syringe When the rotor and plunger are turned, a small constant volume per unit of time
of nitrogen dioxide is released to the Nitrogen Dioxide Meter
The Nitrogen Dioxide Meter was calibrated using the above described apparatus Specific amounts of nitro-gen dioxide used over definite time peri-ods led to the formation of the calibra-tion and other curves shown herein Solution temperature is recognized as
a factor The results reported here were conducted within the range of 70°
to 95 °F No correction was made within this temperature range Be-cause the sensing solution is a salt solution it is not planned to operate below 35 °F without means of heating the instrument
Since nitrogen dioxide is a liquid below 21.3 °C and strongly attracted
to moisture, sample preparation for meter calibration has been found difficult at low ambient temperatures and at very high relative humidities Since many of the investigations were made at or near 100% relative humidity, care had to be exercised in maintaining the sample transfer syringes free of moisture droplets The presence of a small droplet of water in a syringe was found capable of altering the NO2
Fig 7 Mast Model 724-11 nitrogen dioxide
meter with a recorder, front view.
462
Fig 8 Mast Model 724-11 nitrogen dioxide meter, front view, cover open. Fig 9 Mast Model 724-11 nitrogen dioxidemeter, rear view, cover open.
Trang 6concentration of a sample in either
direction The moisture could absorb
a considerable portion of an NO2 sample
and then later release a considerable
quantity of NO2 to a more dilute gas
sample
This test program led to the
develop-ment of the Mast Model 724-11
Nitro-gen Dioxide Meter This instrument
is similar in size and construction to
the Mast 724-2 Ozone Meter Figure 7
is a front view of the Meter and a strip
chart recorder, and shows the
micro-ammeter on the front cover Figure 8
is a front view of the Nitrogen Dioxide
Meter with front cover open and shows
the microcoulomb sensor, control knob,
and waste solution reservoir Figure 9
is a rear view photo with the rear cover
removed and shows the solution pump,
air pump, and fresh solution reservoir
The interrelationship of ozone and
nitrogen dioxide as detected by the
Mast Model 724 Series Meters has been
covered Some observations have been
made of the interference of sulfur dioxide
Using the modified bromide solution
containing citric acid, the Nitrogen
Dioxide Meter was found to give very
low positive readings to quantities of
sulfur dioxide in an air sample A
sample of 390 ppm gave a reading of
only 0.28 microampere Successive air
dilutions indicated that even this low
output signal was representative of the
sulfur dioxide concentration, as 195
ppm gave a 0.18 ua reading, 98 ppm gave
a 0.11 ua reading, and 49 ppm gave a
0.06 ua reading The same solution
would have required about 2 ppm of
nitrogen dioxide to give a 0.28 ua
reading, as obtained from 390 ppm
of sulfur dioxide Preliminary tests
were made in measuring nitrogen
dioxide and sulfur dioxide mixtures
When the ratio of sulfur dioxide to
nitrogen dioxide was one part to eight
parts by volume, the interference was
not significant and the nitrogen dioxide
reading appeared normal As the
pro-portion of sulfur dioxide was increased,
a negative interference was observed
resulting in lowered nitrogen dioxide
readings This interference appeared
greater at higher gas concentrations
For example, at a one to two ratio the
output signal for 330 ppm was reduced
by about one third but appeared norma1
for 40 ppm of nitrogen dioxide with the
same ratio or 20 ppm of sulfur dioxide
present A one to one ratio of sulfur
dioxide reduced a 330 ppm nitrogen
dioxide reading by 60%, and a three to
one ratio of sulfur dioxide reduced a
165 ppm nitrogen dioxide reading by
about 63% Normal readings were
again obtained as soon as the sulfur
dioxide content was reduced or omitted
from the gas sample
The sensitivity to sulfur dioxide was
very low with the solution composition
and other sensor characteristics chosen
for nitrogen dioxide measurement in these tests It appears feasible that these factors could be varied to increase sensitivity to sulfur dioxide if desired, and a similar instrument be developed for the continuous measurement of sulfur dioxide This approach is sched-uled for preliminary investigation in the near future
On the matter of performance, the response time of a sustained level of
2000 ppm of nitrogen dioxide is 50%
of full reading in 0.2 minutes, 87% in one minute, and full reading response occurs in less than five minutes The detector recovers in less than 12 seconds after removal of nitrogen dioxide
Recovery down to 20% of an actual concentration reading occurs in less than
60 seconds Full recovery occurs in less than five minutes
The above response time values are considered typical for a straight bro-mide solution Faster and slower values have been observed Usually 85% of full reading is obtained in less than one-half minute with an acidified bro-mide solution Various factors can affect this response time Conditions
or characteristics of various sensors and sensor gas sample inlet tubes are among these factors Clean, dry Teflon sur-faces would be considered an optimum carrier for NO2 sampling
REFERENCES
1 A F Wartburg, A W Brewer, and James P Lodge, Jr., "Evaluation of a Coulometric Ozone Sensor." Presented
at 138th Meeting, American Chemical Society (September 1960).
2 James P Lodge, Jr Private communi-cations (May 16, 1960, and April 24, 1961).
3 W R Van Nattan, Tamas A Drake,
et al., "NOs Calibration for Mast De-velopment Company's Ozone Meter MDC Model 724-1," a Martin Com-pany, Denver, Colorado, communica-tion (November 1960).
4 Gifford M Mast, "Research and De-velopment on the Instrumentation of Ozone Sensing," Instrument Society of America, 18-SF60 (May 1960).
5 E Effenberger, L Anal Chemic, 134:
106-109 (1951).
6 E Gluckauf, H G Neal, G R Marten,
and F A Paneth, J of Chemical
Society, 48: 2045(1946).
CLASSIFIED
SENIOR RESEARCH SCIENTIST (AIR POLLUTION) Conducts
origi-nal research in air pollution field Re-quire college graduation and three years of professional research experience
in the field Experience requirement varies for holders of master and doctoral degrees in field Salary range $10,520
to $12,575 All Civil Service benefits
Write: Mr Richard H Mattox, Direc-tor of Personnel, New York State Department of Health, 84 Holland Avenue, Albany, New York for applica-tion procedure
PHS TO SPONSOR TRAINING COURSES
The Public Health Service, through its Division of Air Pollution, will conduct the one-week training course,
"Measurement of Airborne Radio-activity," December 10 to 14, 1962, at Cincinnati, Ohio The course is de-signed for engineers It focuses atten-tion on sampling and analysis of radio-activity in air, with time devoted to interpretation of the resulting measure-ments, in terms of health significance Completion of the course, "Basic Radiological Health" (given in Cin-cinnati, October 22 to November 2 and
at Rockville, Maryland, November
26 to December 7) is recommended prior training
The Public Health Service, through its Division of Occupational Health, will conduct a series* of related training courses November 26 through December
21, 1962, at the Occupational Health Research and Training Facility, Cin-cinnati, Ohio "Industrial Hygiene Engineering," designed for industrial hygienists and engineers in the field of occupational health, and "Industrial Hygiene Chemistry," for chemists and chemical engineers in this field, are given concurrently, November 26 to December 7 Trainees in both courses meet together for the first week for instruction in industrial hygiene and medicine, toxicology, and principles pertaining to the evaluation of the environment Meeting separately the second week, work for the hygienists covers temperature and humidity meas-urements, illumination, noise measure-ment and control, and industrial ventila-tion; for the chemists, laboratory analyses for lead, free silica, and solvents; and spectroscopy, polarog-raphy, X-ray diffraction, electron microscopy, and gas chromatography Related courses, "Industrial Exhaust Ventilation" and "Emission Spectro-graphic Techniques" follow on December 10 to 21 and 10 to 20 re-spectively The first provides valuable training for design engineers and hygiene engineers charged with evaluation of the industrial environment The second acquaints the trainee with the principles and applications of emission spectro-graphic methods as applied to analysis
of environmental and biological samples Descriptions of these courses are
given in the Training Program Bulletin.
Applications or requests for information should be addressed to the Chief, Training Program, Robert A Taft Sanitary Engineering Center, 4676 Columbia Parkway, Cincinnati 26 Ohio, or to a PHS Regional Office