The differential refractometer, or refractive index RI detector, responds to a difference in the refractive index of the column effluent as it passes through the detector flow cell.. The R
Trang 11 2 3 4 5 6
5
7
1
4
6
(a)
(b)
Figure4.19 Response of chemiluminescent nitrogen detector (CLND) for different
com-pound types (a) Response of 50 ng nitrogen-equivalent of 1, N,N-dimethyl aniline; 2, nitrobenzene; 3, miconazole nitrate; 4, nicotinamide; 5, 4-acetamidophenol; 6, glycine; 7, caffeine (b) Response of 1, 1 ng nitrogen-equivalent of standard to 1 mg injected solvents:
2, acetone; 3, ethyl acetate; 4, hexane; 5, isopropanol; 6, methanol; 7, water Adapted from
[29]
both the concentration of the chiral compound and its molecular structure Optical
rotary dispersion (ORD) detectors operate on a similar principle to polarimeters,
but use lower wavelengths (e.g., the 365-nm mercury emission line), which in
theory should give stronger signals Circular dichroism (CD) detectors are based
on measuring the difference in absorption of right and left circularly polarized light when an analyte passes through the flow cell For strong CD signals, it is desired that the analyte have a chromophore with absorption in the 200 to 420 nm range
In the example of Figure 4.21 [31], the response of CD, ORD, and UV detection
is compared for the chiral chromatographic separation of ibuprofen enantiomers
The CD detector generates peaks with signal to noise (S /N) about 5-fold larger than
the ORD detector, but only half that of the UV detector Note that the two chiral detectors produce both negative and positive peaks Another study [32] compared the response of PL, ORD, and CD detection for 6 pharmaceutical compounds For naproxen, CD was about 6-fold more sensitive than PL, and 24-fold more than
Trang 2Asp
0 50 100 150 200 250
0 200 400 600 800 1000
Time (min)
Pro Pro
Ala Ala
Thr
Thr Glu
Glu
Cys Gln
Gly
Gly Ser
Ser Asn
Asn
Hyp
Asp
(b)
(a)
Figure4.20 Response of chemiluminescent nitrogen detector (CLND) for amino acids
(a) 10 μL injection of a 0.1 mM standard solution of 13 underivatized amino acids; (b) 10 μL
injection of wine filtered through a 1000 Da filter Adapted from data of [30]
ORD The relative response for the various test compounds varied, but CD was superior in all cases
The differential refractometer, or refractive index (RI) detector, responds to a difference in the refractive index of the column effluent as it passes through the detector flow cell The RI detector is a bulk-property detector that responds to all solutes, if the refractive index of the solute is sufficiently different from that of the mobile phase The most popular RI detector design is the deflection refractometer
illustrated in Figure 4.22a Light from the source lamp (typically tungsten) is
directed through a pair of wedge-shaped flow cells One cell is the reference cell, typically containing a trapped (static) sample of mobile phase; column effluent is directed through the sample cell As the light passes through the detector cells,
Trang 30 5
Time (min)
2000
–2000
0
–500
–1500
–1000
0 400
800
1200
CD (230 nm)
ORD
UV (265 nm)
0
(a)
(b)
(c)
Figure4.21 Comparison of response of circular dichroism (CD), optical rotation (ORD), and
UV detectors for a 10μg injection of ibuprofen (a) CD at 230 nm, S/N = 49.6; (b) ORD,
S /N = 10.9; (c) UV at 265 nm, S/N = 113.4 Adapted from [31].
lamp
slit reference cell
photodiode
detector output
(a)
photodiode array
(b)
Figure4.22 Schematic of a deflection refractive index (RI) detector (a) Dual-photodiode detector; (b) photodiode array detector (lamp and flow cell not shown) Dashed lines show
optical path
it is refracted differently, depending on the instantaneous conditions in the cell
A pair of photodiodes measures the change in refraction (position of the beam)
of the light passing through the flow cell and converts this to an output voltage
The conventional RI detector uses two photodiodes, as shown in Figure 4.22a As
the refractive index of the sample solution changes, the light is deflected so that the amount of light reaching each photodiode changes More recent application
Trang 4Table 4.7
Characteristics of Refractive Index Detectors
Excellent versatility; all solutes can be detected
Moderate sensitivity
Generally not useful for trace analyses
Not useful for gradient elution
Efficient heat-exchanger required
Sensitive to temperature changes
Reliable, fairly easy to operate
Nondestructive
of photodiode-array technology to the RI detector allows multiple photodiodes to
be used for detection, as shown in Figure 4.22b This configuration is claimed to
improve the dynamic range of the RI detector and increase detector sensitivity Table 4.7 summarizes the characteristics of RI detectors Because they respond
to all solutes, these devices have excellent versatility if the mobile phase is properly selected For maximum RI detector sensitivity, the mobile phase should have a refractive index as different from the solute as possible (Table I.3 of Appendix I) However, even under optimum conditions RI detectors possess only modest sensi-tivity Although this detector generally is not useful for trace analysis, it is possible under optimum conditions to quantify peaks at the 0.1% concentration level A severe limitation of RI detectors is that they are unsuitable for use with gradient elution, since it would be exceedingly difficult to match the refractive indexes of the
reference and sample streams (see exception in the discussion of Fig 4.25a, Section
4.12.1) Even isocratic mobile-phase composition changes that are insignificant with
UV detection can show up as baseline noise or ripple For best results hand-mixed mobile phases will give quieter baselines than those prepared by on-line mixing Despite the sensitivity limitation and impracticality in gradient elution, the differ-ential refractometer is widely used, particularly in size-exclusion chromatography, where sensitivity is not as critical
The sensitivity of RI detectors to temperature change also represents a severe limitation Current models of RI detectors have been carefully designed to minimize temperature fluctuations through the use of constant-temperature detection envi-ronments and efficient heat exchangers to thermally equilibrate the mobile phase stream with the detector For best performance, the RI detector should be turned
on at all times, or allowed to warm up for at least two hours prior to use Another good tip is to insulate the tubing connecting the column to the detector so as
to minimize temperature fluctuations Refractometers are convenient and reliable, although generally not as trouble free and easy to operate as UV detectors
RI baseline drift can result when changing from one bottle of ‘‘pure’’ solvent
or ‘‘identical’’ hand-mixed mobile phase to another Baseline drift can be severe when different solvents are involved, until the first solvent is completely flushed from the HPLC equipment and column To maintain a homogeneous composition
of the mobile phase during a series of runs, a sufficiently large volume of mobile
Trang 55 10 Time (min) 0
0 10 20 30 40
treosulfan
barbital (I.S.)
Figure4.23 Refractive index detector response for 560μg/mL treosulfan 1 hr after onset of intravenous infusion; barbital is used as an internal standard Adapted from data of [33]
phase should be formulated, with continuous stirring within the reservoir For acceptable baseline stability, any change in the solvent composition (due to degassing, evaporation, water vapor pickup, etc.) normally should be avoided
In the past RI detectors based on a Fresnel design or interferometric detection were available, but the deflection refractometer is most popular today For many years, the RI detector was the only option for ‘‘universal’’ detection with HPLC Today, light-scattering detectors (Section 4.12) are replacing RI detectors for many applications Low-wavelength UV detection (<210 nm) also provides better
sensi-tivity than RI for many compounds that have very weak UV absorbance at higher wavelengths (see Fig 4.25 for some comparisons of UV, RI, and ELSD responses) The sensitivity of RI detection usually precludes its use in routine drug monitoring, but in some cases it has proved useful for the determination of drug concentrations in biological samples, for example, when high drug concentrations are present, and other detection techniques have failed In the example of Figure 4.23 the RI detector is used to measure treosulfan (L-threitol-1,4-methanesulfonate) levels
in pediatric plasma [33] Treosulfan is an antitumor drug that is toxic to stem cells, and is administered intravenously prior to a stem cell transplant to kill all the native stem cells Figure 4.23 shows a chromatogram for 560-μg/mL treosulfan in pediatric plasma following infusion of the drug; adjusting for sample preparation, this is equivalent to an injection of 83μL of plasma
In recent years improvements in light-scattering detectors have led to their replacing the refractive index (RI) detectors for many applications One reason for this trans-formation (which has also boosted the practicality of mass spectrometric detectors)
Trang 6Table 4.8
Comparison of Refractive Index and Light-Scattering Detectors
Note: +, good; 0, intermediate; −, very poor; na, does not apply (detector designed for qualitative infor-mation, not sensitivity).
aRI, refractive index; ELSD, evaporative light-scattering detector; CNLSD, condensation nucleation light-scattering detector; LLSD, laser light-scattering detector.
is the ability of the light-scattering detector to efficiently nebulize the column effluent
and evaporate the mobile phase The most popular is the evaporative light-scattering
detector (ELSD) The condensation nucleation light-scattering detector (CNLSD) is
a modification of the ELSD that can provide increased performance On the high-end
of the price range are laser light-scattering detectors (LLSD), which occupy more of
a specialty application niche than the ELSD and CNLSD A comparison of some of the properties of the refractive index and light scattering detectors is presented in Table 4.8
4.12.1 Evaporative Light-Scattering Detector (ELSD)
Evaporative light-scattering detectors (ELSD) are based on evaporation of the mobile phase, followed by measurement of light scattered by particles of nonvolatile analyte The ELSD principle is illustrated in Figure 4.24 Column effluent is nebulized in a stream of nitrogen or air and evaporated in a heated drift tube, leaving nonvolatile particles suspended in the carrier gas stream Light scattered by the particles is detected by a photodetector mounted at a fixed angle from the incident beam The ELSD should respond to most compounds that are analyzed by HPLC, but sensitivity may decrease for more volatile analytes Detector response is related to the absolute quantity of analyte present, not its spectral properties
The ELSD, like the refractive index (RI) detector, is considered universal, so it has potential to be used for ‘‘any’’ sample ELSD has the advantage over RI of having
a response independent of the solvent, so it can be used with gradient elution and
is insensitive to temperature or flow-rate fluctuations The selection of the mobile phase for ELSD has similar restrictions as mass spectral detectors (Section 4.14)
in that the mobile phase must be volatile and free of nonvolatile additives (e.g., phosphate buffer) Once the ELSD is adjusted (e.g., carrier flow rate, drift-tube temperature) for the mobile-phase conditions, it should provide acceptably stable operation Linearity is somewhat limited (10- to 100-fold), but with the selection of appropriate calibration standard concentrations, ELSD can be useful for quantitative work over a wider range in analyte concentration
Trang 7nebulizer gas
light-scattering cell
photocell
analyte
nebulizer
from column
heated drift tube
Figure4.24 Schematic of an evaporative light scattering detector (ELSD)
In general, ELSD provides a 10- to 100-fold improvement in sensitivity over the RI detector, with detection limits of 1- to 100-ng on-column For some samples
the sensitivity gain can be much greater, as is seen in Figure 4.25a for the separation
of a triglyceride sample with detection by ELSD, whereas the UV detector at 205 nm and the RI detector do not respond to the triglycerides Note that this separation is via gradient elution in the nonaqueous reversed-phase (NARP) mode (Section 6.5) Whereas water/organic gradients are not suitable for RI detection, acetonitrile and dichloromethane are sufficiently similar in refractive index that a changing mixture
can be tolerated by the RI detector The chromatograms of Figure 4.25b illustrate
the superiority of the ELSD over the RI detector for a polyethylene sample analyzed
by high-temperature (160◦C) GPC
4.12.2 Condensation Nucleation Light-Scattering Detector (CNLSD)
The condensation nucleation light-scattering detector (CNLSD) is an enhancement of the standard ELSD for improved sensitivity and linear range Following evaporation
of the mobile phase, a saturated stream of solvent is added to the particles in the carrier gas The particles act as condensation nuclei and the solvent condenses onto the particles, causing them to grow to a size where they are more easily detected
by light-scattering detection [34] Early work in this field [34] used butanol vapor, but current instrumentation uses water as the condensing solvent The applications
of the CNLSD are the same as those for the ELSD In general, the CNLSD gives 10- to 100-fold improvement in sensitivity over the classic ELSD configuration Manufacturer’s applications literature [35] shows detection of inorganic ions (Li+,
Na+, K+) at 0.5-ng on-column, linearity for sucrose of three orders of magnitude, and five orders of magnitude of dynamic range
Trang 85
Figure4.25 Comparison of ELSD detector response (a) ELSD versus refractive index
(RI) and UV at 205 nm for triglyceride sample Shimadzu Premier C18 column; acetoni-trile/dichloromethane gradient; 1 mL/min; 30◦C (b) ELSD versus RI for the analysis of
polystyrene standards by high-temperature (160◦C) GPC; 200μg sample on PL-Gel Mixed
B column Sample molecular weights: 1, 2,560,000 Da; 2, 320,000 Da; 3, 59,500 Da; 4, 10,850 Da; 5, 580 Da (a) Courtesy of Shimadzu Corporation; (b) courtesy of Varian Polymer
Laboratories
4.12.3 Laser Light-Scattering Detectors (LLSD)
Laser light-scattering detectors (LLSD; also called multi-angle light-scattering,
MALS) generally refer to HPLC detectors that make light-scattering measure-ments in solution, as opposed to the ELSD or CNLSD systems that measure light scattered by particles suspended in a gas LLSD use a laser light source directed
on the flow cell as the sample passes through in the mobile phase Scattered light
is measured at multiple angles (e.g., 3–18 different angles) and can be used, with the proper mathematical transformations, to determine the mass of the analyte in the absence of reference standards These detectors are useful in conjunction with size-exclusion chromatography (see Chapter 13) for the determination of molec-ular weights of synthetic polymers and biological molecules in the range of 103
to 106 Da Figure 4.26 shows superimposed UV chromatograms (280 nm) for a protein kinase fragment and three protein standards (ADH trimer, BSA and ADH
monomer) Also shown are the LLSD-determined molecular weights (y-axis; 3
sep-arate runs) The kinase has a theoretical mass of 53.5 kDa, whereas the molecular weight of the kinase peak by LLSD is about 108,000, indicating that this is a dimer peak The expected molecular weights of the standards are 141,000 (ADH), 67,000
Trang 9LLSD molecular wt (
Elution volume (mL)
LLSD UV Kinase
250
200
150
100
50
ADH
ADH sub-unit BSA
dimer
BSA
Figure4.26 Size-exclusion separation of several proteins, with detection by laser
(multi-angle) light-scattering detector (LLSD) and UV at 280 nm Molecular weights by LLSD
are plotted on the y-axis Kinase, BSA, and ADH each run separately Adapted from Wyatt
Technology Corporation
(BSA), and 35,000 Da (ADH sub-unit), which closely match values by LLSD in Figure 4.26 The BSA dimer (135,000 Da) is observed to elute earlier (23.3 mL) than ADH (24.7 mL) despite its lower molecular weight This demonstrates the greater accuracy of LLSD for molecular-weight determinations, compared to values from size-exclusion measurements (Sections 13.8, 13.10.3.1)
The corona-discharge detector, also called the charged-aerosol detector (CAD)
is classified as a universal HPLC detector because it responds to most analytes The function of the CAD is illustrated in the schematic diagram of Figure 4.27 Column effluent is nebulized and the mobile phase is evaporated, the same as by the evaporative light-scattering detector (Section 4.12.1) or the mass spectrometer (Section 4.14) Analytes in the gas phase are then mixed with a stream of nitrogen gas that has been positively charged by a corona-discharge device The charge is transferred to the analyte particles, and high-mobility charged species are removed
in an ion trap to improve signal quality The remaining charged analyte ions generate
a signal that is read by an electrometer
The CAD is sensitive to nearly any compound that is sufficiently less volatile than the mobile phase so that remains in the gas phase after the mobile phase is evaporated As with other evaporative detectors, the mobile phase is restricted to volatile components (e.g., no phosphate buffer); it also requires particles that can be charged in the detector CAD has been applied to sugars and other carbohydrates
as an alternative detector to RI or ELSD, with detection limits (S /N = 3) for
oligosaccharides of 5-ng on-column and a dynamic range of >104 [36] The example of Figure 4.28 shows that the CAD can be applied to impurities analysis at
Trang 10gas in
nebulizer
analyte vaporizes
charge transfer
corona needle
from column
ion trap
electrometer
Figure4.27 Schematic of the corona discharge detector
Figure4.28 Response of corona-discharge detector to 10μg on-column of sulfadimethoxine
(6) plus 5 ng on-column each of related substances: 1, sulfaguanidine; 2, sulfamerazine; 3, sul-famethazine; 4, sulfamethizole; 5, sulfamethoxazole; and 6, sulfadimethoxin Adapted from
data of [39]
the 0.05% level relative to the active pharmaceutical ingredient (API) [37] In this case 5-ng on-column of 5 related sulfonamide drugs (Fig 4.28, peaks 1–5) are easily detected in the presence of 10-μg on-column of sulfadimethoxine (6)
Hyphenated HPLC detectors refer to the coupling of an independent analytical
instrument (e.g., MS, NMR, FTIR) to the HPLC system to provide detection The