1. Trang chủ
  2. » Khoa Học Tự Nhiên

Introduction to Modern Liquid Chromatography, Third Edition part 10 ppt

10 378 0
Tài liệu đã được kiểm tra trùng lặp

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Định dạng
Số trang 10
Dung lượng 313,42 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

Column diameter d c does not directly affect the relationship between N, run time, and P, but smaller column diameters mean smaller volume peaks, which are more affected by extra-column

Trang 1

6

4

2

n (optimum) B/ n

h

An 1/3

Cn

h

(a)

8

6

4

2

F (mL/min)

30 °C 6,000 Da

80 °C

200 Da

h

(b)

n

Figure2.13 Knox equation (a) Plot of reduced plate height h versus reduced velocity ν with

A = 1, B = 2, and C = 0.05 The total plate height h is shown as a solid curve; the three con-tributions of Equation (2.17a) to h are shown as dashed or dotted curves (b) Effect of a change

in solute molecular weight (6000 vs 200 Da) or temperature (80◦C vs 30◦C) on h versus flow rate All plots based on Equation (2.17) with A = 1, B = 2, and C = 0.05.

2.4.1.2 Some Guidelines for Selecting Column Conditions

Column conditions can be defined as column length and diameter, particle size, and

flow rate Together these experimental choices by the user determine the value of

N, run time, and column pressure P In general, larger values of N require longer

run times, but an increase in the available column pressure can be used to improve

the trade-off between N and run time: either larger values of N for the same run time or shorter run times for the same value of N Column diameters can be varied

for different purposes; larger diameters for preparative separations (Chapter 15), or smaller diameters for (1) increased sensitivity when the amount of sample is limited, (2) reduced flow rates for LC-MS (Section 4.14), (3) reduced solvent consumption,

Trang 2

or (4) the use of very high-pressure operation (to minimize problems caused by the

generation of heat within the column; Section 3.5.4.3) Column diameter d c does

not directly affect the relationship between N, run time, and P, but smaller column

diameters mean smaller volume peaks, which are more affected by extra-column peak broadening; small-diameter columns are therefore more demanding in terms

of equipment Similarly pressures higher than about 5000 psi also require special equipment (Section 3.3.5.4) and columns

Apart from the issue of column diameter, the choice of column conditions can

be guided by the theory of Section 2.4.1.1 The following discussion will assume a sample molecular weight<1000 Da, but the separation of larger molecules follows

from the discussion of Figure 2.11b and Section 13.3.1 Consider the following three

possibilities:

A only a single column configuration is available (e.g., a 150× 4.6-mm

column of 5-μm particles)

B a single column configuration is available in different column lengths so

that length can be varied

C there is a wide choice of column lengths and particle size

For case A, only the flow rate can be varied (as in Fig 2.11a), and there is a maximum possible flow rate F that is determined by the pressure limit of the system;

however, it is often desirable to operate at a lower pressure In this situation the

largest value of F should be selected that results in the desired value of N but does

not exceed some maximum pressure For example, for the 5-μm, 150 × 4.6-mm

column of Figure 2.11a, assume that a value of N≥ 10,000 is required, while not exceeding a pressure of 2000 psi It can be seen(o) that a flow rate of≈ 2 mL/min

will yield a value of N = 10,000, at a pressure <2000 psi.

For case B, both the flow rate and column length L can be varied Maximum values of N in minimum time result when the combined choice of values of F and L results in a maximum column pressure From Equation (2.13a) this is equivalent to holding the product FL constant Figure 2.14a illustrates the resulting dependence

of N on run time (values of t0) for three different particle sizes: 2, 5, and 10

μm A conclusion from Figure 2.14a is that the smallest particles (2 micron) are

preferred for the shortest separation times (and correspondingly for smaller values

of the required N) Thus for a value of t0= 1 min, values of N are largest for a

2-μm particle column, while values of N are largest for a 10-μm particle column

when t0= 1000 min Thus, when large values of N are required, larger particles are best; when small values of N are adequate, smaller particles represent a better choice because of shorter run times At intermediate analysis times (e.g., t0> 5 min),

intermediate-size particles (5 micron) can give the required plate count in less time than the smallest particles

The data of Figure 2.14a are recast into the form of a kinetic or Poppe plot [37, 38] in Figure 2.14b, where the time required to generate one theoretical plate (t0/N) is plotted against the value of N This ‘‘time per plate’’ is seen to increase with

the value of N; Figure 2.14b also emphasizes the advantage of small particles when low values of N are required, and vice versa for large values of N The dashed lines

labeled ‘‘10 sec,’’ ‘‘100 sec,’’ and ‘‘1000 sec’’ correspond to conditions of constant

t0; other values of t0 are defined by lines parallel to those shown Poppe plots

Trang 3

10 6

10 5

10 4

103

10 −1

10 −3

10 −2

10 −4

N

t0(min)

d p= 10 mm

(a)

(b)

5 mm

2 mm

5 mm

2 mm

1

t0

N

106

105

104

103

102

d p= 10 mm

P= 2000 psi

t0= 1000 sec

t0= 100 sec

t0= 10 sec

Figure2.14 Variation of N as a function of column length, holding column pressure constant

by varying flow rate (a) N versus t0for different particle sizes; (b) ‘‘Poppe plot’’ for data of (a) Calculated values from Equation (2.17), assuming A = 1, B = 2, and C = 0.05, a viscosity

of 0.6 cP (e.g., 60% acetonitrile-water, 35◦C), D m= 10−5cm2/ sec (corresponds to a sample

molecular weight of 300 Da for the latter conditions), and a column pressure P= 2000 psi

are now widely used to compare column performance, as they take both column efficiency and permeability into account

For case C, we are permitted to vary column length and particle size contin-uously While this will be only approximately true in practice, there is not much penalty for using column lengths and particle sizes that are not too different from the optimum Case C should be considered when the goal is either very fast separation

Trang 4

or very large values of N (extreme separation conditions) The past decade has

witnessed an increasing demand for faster HPLC separation, sometimes involving run times of a minute or less (Section 2.5.3.2) Similarly biochemists are now faced with the need to separate samples that contain thousands of components (Section

13.4.5); in such cases large values of N become more important Whether the goal for difficult separations is minimum run time or maximum N, experimental conditions should be selected that favor maximum N per unit time—as in Figure 2.14b It can be shown that the best choice of column length, particle size, and flow rate will always correspond to a minimum value of h ≈ 2 (or ν ≈ 3) Because

commercial columns are limited to certain lengths and particle sizes, we can only approximate conditions forν ≈ 3 while at the same time achieving a required value

of N in the shortest run time Fortunately, an increase in ν by 2- to 3-fold has only

a small effect on h or N, which gives us considerable flexibility in the selection of a

preferred column length or particle size for a given sample See also [38a]

Figure 2.15, which results from the application of Equation (2.17) withν = 3,

can serve as an example for the selection of conditions that provide maximum N

for a given separation time with some maximum pressure Assume a maximum

pressure P = 2000 psi, and a value of k = 3 for the last peak in the chromatogram (for other values of k, adjust the times in Figure 2.15 in accordance with Eq 2.5).

A value of N= 5000 can be achieved in about 15 sec (marked by• and arrow in

the figure) Conversely, N= 300,000 will require about 17 hr (60,000 sec, marked

by◦and arrow in the figure) In the first case (N= 5000), ≈1-μm particles will be

required In the second case (N= 300,000), 9 μm particles are recommended That

is, fast separations require smaller particles, while large N separations require larger

particles (plus much longer separation times)—as in our preceding discussion of Figure 2.14 An increase in the allowable pressure results in either shorter times for

the same plate number, or larger values of N for the same time—as the reader can verify from Figure 2.15 for P= 20,000 compared with 1000 psi Higher pressures also favor the use of smaller particles, for either faster separation or larger values of

N Very large values of N or very short run times require extreme conditions, as can

be inferred from the plot of Figure 2.15

For various reasons the plots of Figures 2.11 and 2.13 through 2.15 must

be regarded as approximate, as well as varying with other conditions according

to Table 2.4 (e.g., mobile-phase viscosity, sample molecular weight, and tempera-ture) So these figures should be used primarily as qualitative guides for selecting experimental conditions rather than quantitative data for a given separation Flow rates and column lengths corresponding to Figure 2.15 can be determined

as follows: From Equation (2.18) with H = L/N, and h = 2,

L = NH = Nhd p

Here cgs units apply: L and d pin cm From Equation (2.18a) forν ≈ 3,

u= 3D m

Trang 5

N= 10 6

10 5

10 4

10 3

100

10

1

N= 10 5

N= 2 × 10 4

N= 3 × 10 5

N= 5000

N= 1000

Run time

t (sec)

d p= 10 μm

d p= 5

d p= 1

3

2 1.5

Pressure (psi × 0.001)

1 min

1 hr

P= 2000 psi

Figure2.15 Variation of column plate number N with separation time t, column pressure P, and particle diameter d p ( _) lines representing different values of N; ( -) lines

represent-ing different particle sizes; ( .) line representrepresent-ing a column pressure of 2000 psi All values

in figure are for minimum plate height, h = 2 and k = 3 Adapted from [23] for a viscosity of

0.6 cP (e.g., 60% acetonitrile-water, 35◦C) and D m= 10−5cm2/sec (corresponds to a sample molecular weight of 300 for the latter conditions)

with F given by

F=3V mDm

where V m is the column dead-volume (equal to t0F, Eq 2.7a) The units of Equation

(2.21b) are mL/sec (F), mL (V m), cm2/ sec (D m ), cm (L), and cm (d p)

2.4.2 Peak Shape

So far we have assumed symmetrical peaks in the final chromatogram Under ideal conditions a peak will have a Gaussian shape given by

y = (2π) −0.5 e −x2/2 (2.22)

Trang 6

Here x is (t − t R)/σ , where t is time and σ = W/4; y is the peak height Actual

peaks in a chromatogram will usually depart slightly from a symmetrical, Gaussian

shape, typically showing more or less tailing as in Figure 2.16a Peak tailing can

be characterized in either of two ways (Fig 2.16a): by the asymmetry factor A s

(left-hand side of figure) or by the tailing factor TF (right-hand side) Values of A s

and TF are related approximately as

so that values of A s are typically somewhat larger than values of TF (Table 2.5) Figure 2.16b shows how peak shape varies for different values of A s or TF; a value of A s or TF= 1.00 corresponds to a perfectly symmetrical peak The effect

of peak tailing on the separation of two adjacent peaks in the chromatogram is illustrated in Figure 2.16c (where only peak tailing varies, as measured by the value

Peak tailing factor TF

= (A + B)/2A (5% values)

(b) (a)

(c)

10% of peak height

5% of peak height

Peak asymmetry factor A s

= B/A (10% values)

t

Figure2.16 Measurement and effects of peak tailing (a) Definitions of peak asymmetry and tailing factors (A S and TF); (b) peak shape as a function of asymmetry (A S ) and tailing (TF) factors; (c) effect of peak tailing on separation; (d) peak fronting; (e) ‘‘overloaded’’ tailing, in contrast to ‘‘exponential’’ tailing in (a).

Trang 7

Table 2.5

Relation of Peak-Tailing and Peak-Asymmetry Factors

Peak-Asymmetry Factor (at 10%) Peak-Tailing Factor (at 5%)

of TF) The two peaks with TF = A s = 1.0 are well separated from each other As

peak tailing increases, however, their separation progressively worsens Many times

the extent of peak tailing is quite minor (TF < 1.2) and may not be noticed Tailing

of this magnitude (TF ≤ 1.2) will have a negligible effect on separation, except for the case of a large peak followed by a very small peak (compare Fig 2.17d,e) Less often, peak fronting will be observed (TF < 1.0; Fig 2.16d), with a similar,

potentially adverse effect on separation Major peak tailing (TF > 2) can be highly

detrimental to both separation and quantitative analysis; a common requirement for

routine separations is that TF < 2 for all peaks Quantitation based on peak area

(Chapter 11) requires integration of the entire peak; this in turn depends on setting proper integration limits, which can become difficult for the case of severely tailing peaks

If peak tailing with TF ≥ 2 is observed, steps should be taken immediately

to correct the problem—whether for a routine assay procedure or during the development of an HPLC method The remainder of this book will assume symmetric

peaks, unless noted otherwise Several possible causes of peak distortion or tailing exist:

• ‘‘bad’’ column (plugged frit or void; Section 5.8)

• contaminated column (buildup of strongly retained sample contaminants; Section 5.8)

• column overload (too large a sample; Section 2.6.2)

• a sample solvent that is too strong (Section 2.6.1)

• extra-column peak broadening (Sections 2.4.1, 3.9)

• silanol interactions (including contamination of the column packing by trace metals; Sections 5.4.4.1, 7.3.4.2)

• inadequate or inappropriate buffering (Section 7.2.1.1)

• use of a higher column temperature with cold (inadequately thermostatted) mobile phase (Section 2.5.3.1)

See also Section 17.4.5.3 Before attempting to reduce tailing, two kinds of peak

tailing should be distinguished: exponential versus column-overload tailing

Expo-nential tailing is characterized by a gradual return of the peak to baseline, as in

Trang 8

Rs= 1.0 1.5 2.0 1:1

10:1

100:1

1000:1

1000:1

TF = 1.2

1:1000

TF = 1.2

(peaks

reversed)

(a)

(b)

(c)

(d )

(e)

(f )

Figure2.17 Separation as a function of resolution, relative peak size (1:1 10:1, etc.), and (e, f )

peak tailing

Figure 2.16a; column-overload tailing creates a peak with a ‘‘right-triangle’’ appear-ance, as in Figure 2.16e A reduction in sample weight will usually reduce the tailing

of an overloaded peak as in Figure 2.16e, while exponential tailing may not change (or may even increase) when sample size is reduced [39, 40] Usually it is feasible (and desirable) to reduce peak tailing so that TF ≤ 1.5.

Various attempts have been made to describe the shape of tailing peaks in

mathematical terms [41, 42], and to define a plate number N that includes the adverse effect of peak tailing on separation Values of N for tailing peaks that are

calculated by means of Equation (2.9) (and especially Eq 2.9a) tend to overstate column efficiency; actual column performance as measured by the separation of peaks

within the chromatogram usually suggests a lower value of N The Dorsey–Foley

equation is widely used, in an attempt to correct values of N for peak tailing [43]:

N= 41.7(tR/W0.1)2

Here the quantities A and B are defined in Figure 2.16a (at 10% of peak height), and W0.1 is the value of W measured at 10% of peak height.

Trang 9

Still another attempt at a more realistic value of N (taking peak asymmetry

into account) is the so-called 5-σ value:

N= 25



tR

W5σ

2

(2.22c)

Here W5σ is the peak width measured at 4.5% of peak maximum, similar to the

measurement of N from the width at half height (Eq 2.9a) The effect of peak tailing

on N becomes more pronounced for peak widths that are measured closer to the baseline; that is, W1/2(least effect)< W < W5σ(most effect) Equations (2.22b) and (2.22c), as well as other expedients [41], provide at best only approximate measures

of column performance for tailing peaks Peaks that tail cannot be described by

a single generic equation (comparable to Eq 2.22 for symmetrical peaks), as

witnessed by comparing the two peaks of Figures 2.16a and e; consequently no single value of N can accurately describe peak width for peaks with TF > 1.2 For

the characterization of tailing peaks, we recommend a value of N from Equation (2.9a) together with a value of either A s or TF In many cases peak tailing is tracked

as part of system suitability measurements to assess deterioration of the column; for such applications nearly any measure of peak tailing will serve to successfully track changes in peak shape over time

HPLC method development refers to the selection of separation conditions that

provide an acceptable separation of a given sample; for a more detailed (but less up-to-date) account than provided by the present book; see [44] The separation of

two peaks i and j as in Figure 2.10a is usually described in terms of their resolution

Rs= (difference in retention times)/(average peak width):

Rs= 2[t R(j) − t R(i)]

W i + W j

(2.23)

Wi and W j are the baseline widths W for peaks i and j, respectively Better

separation (increased resolution) results from a larger difference in peak retention times and/or narrower peaks Accurate quantitative analysis based on separations

as in Figure 2.17a is favored by baseline resolution, where the valley between the two peaks returns to the baseline For two peaks of similar size (as in Fig 2.17a), baseline resolution corresponds to R s >1.5 For preparative separations (Section

15.3.2), baseline resolution also allows a complete recovery of each peak with a purity of 100%

Examples of resolution are shown in Figure 2.17 for various values of R s(1.0, 1.5, 2.0) and different peak-size ratios ratios (1:1, 10:1, 100:1, 1000:1) As long as

R s > 1.5 and the two peaks are symmetrical (TF = 1.0), it is seen that there is little

overlap of the two peaks—regardless of relative peak size Minor peak tailing as

in Figure 2.17e (TF = 1.2) can lead to a significant loss of resolution when a small peak follows a large peak, and in such cases a value of R s >2 may be necessary.

Trang 10

From a practical standpoint, it is nearly impossible to obtain peaks with TF ≈ 1.0.

Consequently, if a sample component is present in very low concentration, relative

to a closely adjacent major component, it is preferable—if possible—to place the

impurity peak before the main peak, as a means of minimizing the effects of peak tailing on resolution (compare Fig 2.17f,e) Changes in relative peak position, so as

to place a smaller peak in front of a larger peak, can sometimes be achieved during method development by a change in conditions that affect selectivity (Section 2.5.2)

A common goal of HPLC method development is the separation of every peak

of interest from adjacent peaks (with R s≥ 2), normally corresponding to a one-third

safety factor (vs baseline separation with R s ≈ 1.5, for peaks of comparable size) The goal of R s≥ 2 takes into account minor peak tailing, peaks of moderately dissimilar size, and the usual slow deterioration of the column over time—with

an increase in both peak width and tailing Even larger values of R s may be required in some cases The resolution of two peaks can be improved as described

in Sections 2.5.1 through 2.5.3

When more than two peaks are to be separated, the goal is usually R s≥ 2

for the least well-separated peak-pair This peak-pair is referred to as the critical peak-pair, and its resolution is referred to as the critical resolution of the separation;

for example, adjacent peaks 1 and 2 in Figure 2.18a, for which Rs = 0.8, or

in Figure 2.18d, for which R s = 3.9 Method development usually strives for an

acceptable resolution of the critical peak-pair, which then means an adequate

separation (R s ≥ 2) for all other peaks as well In this book, when we refer to the

resolution R s of a chromatogram, we mean the critical resolution (unless stated

otherwise)

For method-development purposes, it is convenient to derive an alternative, approximate expression for resolution from Equations (2.5a) and (2.10a) (assuming equal widths for the two peaks) [45]:

Rs=

 1 4

 

k

(1+ k)



Here resolution is expressed as a function of the retention factor k for the first peak i (term a), the separation factor α (term b), and column efficiency or the plate number

N (term c) The separation factor α (a measure of so-called separation selectivity or

relative retention) is defined as

α = kj

The quantities k i and k j are the values of k for adjacent peaks i and j, as

in Figure 2.10a For this separation k ≡ k1= 1.55, α = 1.12, and N = 11,000; resolution is therefore given by R s = (1/4)(1.55/2.55)(1.12 − 1)(11, 0000.5)= 1.9.

Equation (2.24) states that resolution can be improved in any of three different ways

(by varying k, α, or N), but improving separation selectivity (values of α) is by far the most powerful option.

Equation (2.24) is slightly less accurate than Equation (2.23) but more useful for interpreting chromatograms during method development Numerical values of

Ngày đăng: 04/07/2014, 01:20

TỪ KHÓA LIÊN QUAN