Trang 1Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 1 of 39
Agilent InfinityLab LC Series
Preventive Maintenance Checklist
Agilent Preventive Maintenance (PM) provides factory recommended service for your systems to assure reliable operation and the accuracy of your results Delivered by highly-trained and certified service engineers using genuine Agilent parts and supplies, Agilent Preventive
Maintenance provides everything you need to reduce unplanned downtime and keep your systems operating at their peak
For more information about Agilent Technologies services please visit our web site using the following URL http://www.agilent.com/home
Customer Information
• Customers should provide all necessary operating supplies upon request of the engineer
• A customer representative should be available to the engineer while performing the
preventive maintenance procedures
• Any parts, not included in the Parts Lists section of this document, are not part of the
recommended Preventive Maintenance service, nor are they included in the price of this service
• If a system requires the use of additional or special procedures and/or parts for the
instrument service, then these must be ordered separately and charged as a repair, which may incur additional costs
Service Engineers Responsibilities
Only complete/printout pages that relate to the system or module being serviced
Complete empty fields with the relevant information
Complete the relevant checkboxes in the checklist using a “X” or tick mark “ ” in the checkbox.Complete Not Applicable check boxes to indicate services that are not delivered as needed.Complete the PM service in the order of the task listed
Complete the Service Review section together with the customer In case of diagnostic tests, record test results in section service review (test results)
Additional Instruction Notes
• Please exercise care when using solvents close to the instrument It could damage the vapor sensor present in the ELSD
Trang 2Record the list of system component product numbers
Trang 3G1364F 1260 Infinity II Analytical Fraction Collector
G5628A Bio-inert Manual Injector
G5654A Bio-inert Quaternary Pump
G5664B 1260 Infinity II Bio-inert Fraction Collector
G5668A Bio-inert Multisampler
G7102A Evaporative Light Scattering Detector
G7104A/C Flexible Pump
G7111A/B Quaternary Pump VL / Quaternary Pump
G7114A/B Variable Wavelength Detector
G7115A Diode Array Detector WR
G7116A/B Multicolumn Thermostat
G7117A/B/C Diode Array Detector FS / Diode Array Detector / Diode Array Detector HS
G7121A/B Fluorescence Detector / Fluorescence Detector Spectra
G7129A/B/C Vialsampler VL/Vialsampler
G7162A/B Refractive Index Detector
G7165A Multiple Wavelength Detector
Trang 4Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 4 of 39
G7104A/C Procedure for 1290 Infinity II Easy Maintenance and LongLife Pump Heads
This procedure requires a torque wrench.
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the pump head assembly
Follow the steps provided in the 1290 Infinity II Easy Maintenance Pump Head TechNote and the 1290 Infinity II Long Life Pump Head TechNote.
Replace the pump seals
Replace the seal wash seals
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Connect the solvent lines
Purge the pump with isopropanol
Replace the high pressure filter frit
Replace the inline filter frit, if installed
Replace the seal wash pump cartridge
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
G7104A Pump Head Exchange Procedure
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the classic pump head assembly
Install the Infinity II Easy Maintenance pump head assembly
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Replace the high pressure filter frit
Replace the inline filter frit, if installed
Replace the seal wash pump cartridge
Connect the solvent lines
Purge the pump with isopropanol
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
Trang 5Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 5 of 39
G7104A Alternative Procedure (Classic Pump Heads)
This procedure requires the 1290 Infinity Pump head service kit.
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the pump head assembly
Follow the steps provided in the Agilent 1290 Infinity Pump Head Maintenance Technote
Replace the pump seals
Replace the seal wash seals and gasket
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Connect the solvent lines
Purge the pump with isopropanol
Replace the high pressure filter frit
Replace the inline filter frit, if installed
Replace the seal wash pump cartridge
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
Trang 6G7110B, G7111A/B, G5654A
Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 6 of 39
Section NOT Applicable
Power off the pump
Remove pump head
Disassemble the pump head
Remove and clean pistons
Clean support rings
Replace piston seals
Replace wash seals and gaskets, if the seal wash option is installed
Reassemble the pump head and install it
Reconnect the AIV to the connector (if installed)
Replace the seal wash peristaltic pump if active seal wash is installed
Power on the module
For standard (PTFE) or PE seals perform the seal wear-in procedure
While running, begin maintenance on the sampler and column compartment if applicable
Purge valve: replace the PTFE frit, replace the seal cap assembly
If seal wash option is installed: Review the correct installation of seal wash function in non-recycle mode Manually activate the seal wash and check if there is a flow
Perform the System Pressure Test (Agilent Lab Advisor software)
Perform the Leak Rate Test (Agilent Lab Advisor software)
Save or print-out results
Reset EMF counters
Trang 7Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 7 of 39
Section NOT Applicable
Power off the pump
Remove pump heads
Disassemble the pump head
Remove and clean pistons
Clean support rings
Replace piston seals
Replace wash seals and gaskets, if the seal wash option is installed
Reassemble the pump heads and install them Reconnect the Active Inlet Valves to the connectors
Replace the seal wash peristaltic pump if active seal wash is installed
Power on the module
For standard (PTFE) or PE seals perform the seal wear-in procedure
While running, begin maintenance on the sampler and column compartment if applicable
Purge valve: replace the PTFE frit, replace the seal cap assembly
If seal wash option is installed: Review the correct installation of seal wash function in non-recycle mode Manually activate the seal wash and check if there is a flow
Perform the system pressure test (Agilent Lab Advisor Software)
Perform a Pump Leak Rate Test (Agilent Lab Advisor software)
Perform a pump elasticity calibration (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
Trang 8Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 8 of 39
G7120A Procedure for 1290 Infinity II Easy Maintenance and LongLife Pump Heads
This procedure requires a torque wrench.
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the pump head assembly
Follow the steps provided in the 1290 Infinity II Easy Maintenance Pump Head TechNote and the 1290 Infinity II Long Life Pump Head TechNote.
Replace the pump seals
Replace the seal wash seals
Replace the filter frit (PTFE frits (pack of 5) (01018-22707)) and seal cap (Seal cap
(5067-4728)) of the high pressure filter assembly
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Connect the solvent lines
Purge the pump with isopropanol
Replace the seal wash pump cartridge
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform the system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
G7120A Pump Head Exchange Procedure
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the classic pump head assembly
Install the Infinity II Easy Maintenance Pump Head assembly
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Connect the solvent lines
Purge the pump with isopropanol
Replace the seal wash pump cartridge
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform the system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
Trang 9Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 9 of 39
G7120A Alternative Procedure (Classic Pump Heads)
This procedure requires the 1290 Infinity Pump head service kit
Section NOT Applicable
Start "Remove/Install Pump Head" procedure (Agilent Lab Advisor Software)
Remove or lift up solvent inlets in order to avoid solvent spilling
Remove the pump head assembly
Follow the steps provided in the Agilent 1290 Infinity Pump Head Maintenance Technote
Replace the pump seals
Replace the seal wash seals and gasket
Replace the filter frit (PTFE frits (pack of 5) (01018-22707)) and seal cap (Seal cap
(5067-4728)) of the high pressure filter assembly
Finish "Remove/Install Pump Head" procedure (Agilent Lab Advisor software)
Connect the solvent lines
Purge the pump with isopropanol
Replace the seal wash pump cartridge
Perform pump leak rate test (Agilent Lab Advisor Software)
Perform the system pressure test (Agilent Lab Advisor Software)
Save or print-out results
Reset EMF counters
Trang 10Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 10 of 39
Section NOT Applicable
Verify the proper function of the drawers
Replace the rotor seal
Replace the needle assembly and the high pressure needle seat assembly
Replace the peristaltic pump cartridge
Check the finger of the vial gripper
Verify the proper installation of the sample loop Especially check if sample loop is moving smoothly with the safety cover installed
Check the proper function of the needle wash Verify correct positioning of the drain tube
Clean the instrument, especially the ventilation
Clean the ICC heater (if applicable)
Clean the Sample Cooler, especially the ventilation and verify that the drain tube is positioned correctly (if applicable)
Clean the Sample Thermostat, especially the ventilation and verify that the drain tube is positioned correctly (if applicable)
After maintaining the multicolumn thermostat or the ICC heater, purge both modules with isopropanol for 5 min
Cap the outlet of the vialsampler with a blank nut ( Blank Nut SL (5067-6127))
Perform a System Pressure Test
Perform a System Pressure Test with ICC heater (if applicable)
Perform a Sample Cooler Test (if applicable)
Perform a ICC Heater Test (if applicable)
Trang 11Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 11 of 39
Section NOT Applicable
Replace rotor seal
Replace the stator face assy (G1328C: valve 5067-4191, G5628A: valve 5067-4158)
Purge injector and column compartment with isopropanol for 5 min
Cap the valve outlet with a blank nut
Perform a (System) Pressure Test Take care for the maximum pressure of the valve and installed loop (PEEK loop max pressure is 200 bar)
Trang 12G7167A/B, G5668A
Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 12 of 39
Section NOT Applicable
Verify the proper function of the drawer(s)
Check the needle position of the Multisampler (see LadAdvisor: Instrument control > Special Commands > Check Needle position) If the needle tip position is off the center perform an auto referencing under operating conditions
Replace the rotor seal(s)
Replace the needle assembly(ies) and the high pressure needle seat assembly(ies)
Verify the proper installation of the sample loops especially check if sample loops are kinked
or bended
Replace the peristaltic pump cartridge (if applicable)
Check the proper function of the needle wash and seat backflushing (if applicable) Verify correct positioning of the drain tube
Clean the instrument, especially the ventilation
Clean the Sample Cooler, especially the ventilation and verify that the drain tube is positioned correctly (if applicable)
Clean the Sample Thermostat, especially the ventilation and verify that the drain tube is positioned correctly (if applicable)
After maintaining the multicolumn thermostat, purge both modules with isopropanol for
Trang 13Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 13 of 39
Section NOT Applicable
Before disconnecting the columns, please record the columns and their position as a reference for re-installation
If a valve is installed, replace the rotor seal (see parts section for details)
Purge LC system with HPLC grade water for 10 min
Perform a System Pressure Test Use a ZDV to install a blank nut onto the heat exchanger's outlet capillary If a valve is installed use the blanking nut to block one of the outlet ports In case of a column selector valve use a ZDV-union to mount the blanking nut on the outlet port capillary with the valve switched to the bypass channel
Reset EMF counter (Valve switches)
Trang 14Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 14 of 39
The maintenance procedure requires NO consumables.
Section NOT Applicable
Inspect flow cell for leaks
Perform holmium oxide test
Perform Intensity Test
Perform Wavelength Verification Test
Perform Dark Current Test
Perform Filter Test
Trang 15G7115A, G7165A
Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 15 of 39
The maintenance procedure requires NO consumables.
Section NOT Applicable
Inspect flow cell for leaks
Perform Filter Test and attach results to checklist
Perform dark current test
Perform holmium oxide test
Perform Intensity Test
Perform Slit Test
Perform Wavelength Verification Test
Trang 16Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 16 of 39
The maintenance procedure requires NO consumables.
Section NOT Applicable
Inspect the tubing from cell outlet to waste (must be a PTFE tubing without restriction)
Inspect flow cell for leaks
Perform the lamp Intensity Test and compare profile (no given limits)
Perform the wavelength verification test using the raman band
Trang 17Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 17 of 39
Section NOT Applicable
Purge the reference and sample cells with water for 5 min
Check diode balance with pure water in both, the reference and the sample cell, instrument stable as specified Adjust optical balance if value >±0.2
Trang 18 Section NOT Applicable
Inspect flow cell for leaks
Check the condition of the positioner cell drawer assembly
Perform Wavelength Verification Test
Perform Intensity Test
Perform Dark Current Test
Perform Slit Test (not available in G7117A/C)
Trang 19Issued: 10-Aug-2018 Revision: 3.4 Copyright: 2017, 2018 Agilent Technologies
Page 19 of 39
Section NOT Applicable
Check communication of Control Software with the instrument
Mass Flow Controller - Check control of gas flow
Lamp Assembly - Check change in baseline from standby to run
Photo Detector Module - Check change in baseline from standby to run
Check autozero function
Evaporator Heater Assembly - Check evaporator temperature increases/decreases
Nebuliser Heater Assembly - Check nebuliser temperature increases/decreases
Peltier Assembly (on G7102A cooled only) - Check for cooling to sub-ambient temperatures
Clean optics block assembly if baseline noise greater than 3 LSD units
If cleaning the Optics did not achieve the results then proceed with the cleaning of the prism assembly
Check for dust around cooling fan, and clean if necessary
Replace the following parts:
Evaporator Cartridge
Gas Inlet Plug Assembly
Clean the Nebuliser Assembly if cleaning prism and optics assembly, and replacing above parts does not resolve baseline noise issues
Return the instrument to initial conditions