1. Trang chủ
  2. » Ngoại Ngữ

lecture 6 gc

5 424 1
Tài liệu đã được kiểm tra trùng lặp

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Định dạng
Số trang 5
Dung lượng 231,71 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

GAS CHROMATOGRAPHYPHAM VAN HUNG, PhD Function „Separation of volatile organic compounds „Volatile –when heated, VOCsundergo a phase transition into intact gas-phase species „Separation o

Trang 1

GAS CHROMATOGRAPHY

PHAM VAN HUNG, PhD

Function

„Separation of volatile organic compounds

„Volatile –when heated, VOCsundergo a phase transition into intact gas-phase species

„Separation occurs as a result of unique equilibria established between the solutes and the stationary phase (the GC column)

„An inert carrier gas carries the solutes through the column

Application

„ Uses

„ Separation and analysis of organic compounds

„ Testing purity of compounds

„ Determine relative amounts of components in mixture

„ Compound identification

„ Isolation of pure compounds (microscale work)

„ Similar to column chromatography, but differs in 3 ways:

„ Partitioning process carried out between Moving Gas Phase

and Stationary Liquid Phase

„ Temperature of gas can be controlled

„ Concentration of compound in gas phase is a function of

the vapor pressure only.

„ GC also known as Vapor-Phase Chromatography (VPC) and

Gas-Liquid Partition Chromatography (GLPC).

Filters/Traps

Ai H

ier Column

Components

„ gas system

„ inlet

„ column

„ detector

„ data system

Data system

Syringe/Sampler Inlets

Detectors Regulators

H

RESET

Schematic Diagram of Gas Chromatography Carrier Gas-Supply

„Carrier gases, which must be chemically inert, include helium, nitrogen, and hydrogen Associated with the gas supply are pressure regulators, gauges, and flow meters In addition, the carrier gas system often contains a molecular sieve to remove water or other impurities

„Must be at a constant flow rate so that retention times & retention volumes may be equated

Trang 2

„A GC syringe penetrates a septum to inject

sample into the vaporization camber

„Instant vaporization of the sample, 280 °C

„Carrier gas transports the sample into the head

of the column

„Purge valve controls the fraction of sample that

enters the column

Split or splitless

„Usually operated in split mode unless sample

limited

„Chromatographic resolution depends upon the

width of the sample plug

„In splitlessmode the purge valve is close for

30-60 s, which means the sample plug is 30-30-60

seconds

„As we will see, refocusing to a more narrow

sample plug is possible with temperature

programming

Column Configurations

„Two general types of columns are encountered

in gas chromatography, packedand open

„Chromatographic columns vary in length from less than 2 m to 50 m or more They are constructed of stainless steel, glass, fused silica,

or Teflon In order to fit into an oven for thermostating, they are usually formed as coils having diameters of 10 to 30 cm

Columns

• Packed

•Capillary

0.32 mm ID

Liquid Stationary phase

Mobile phase (Helium) flowing at 1 mL/min

Open Tubular Capillary Column

15-60 m in length

0.1-5 μm

Trang 3

Polar vs nonpolar

„Separation is based on the vapor pressure and

polarity of the components

„Within a homologous series (alkanes, alcohol,

olefins, fatty acids) retention time increases with

chain length (or molecular weight)

„Polar columns retain polar compounds to a

greater extent than non-polar

„ C18 saturated vs C18 saturated methyl ester

C16:0

C18:0 C18:1

C16:1

C16:0

C18:0 C18:1 C18:2

C16:1

RT (min)

RT (min) Polar column

Non-polar column

Oven

„Programmable

„Isothermal-run at one constant temperature

„Temperature programming -Start at low

temperature and gradually ramp to higher

temperature

„ More constant peak width

„ Better sensitivity for components that are retained

longer

„ Much better chromatographic resolution

„ Peak refocusing at head of column

Typical Temperature Program

Time (min)

50°C

220°C

160°C

Detection Systems

ideal detector for gas chromatography has the

following characteristics:

1 Adequate sensitivity

2 Good stability and reproducibility

3 A linear response to solutes that extends over

several orders of magnitude

4 A temperature range from room temperature

to at least 400oC

Characteristics of the Ideal Detector

5 A short response time that is independent of flow rate

6 High reliability and ease of use

7 Similarity in response toward all solutes or a highly selective response toward one or more classes of solutes

8 Nondestructive of sample

Trang 4

„Flame Ionization Detectors (FID)

„Electron Capture Detectors (ECD)

„Electron impact/chemical ionization (EI/CI)

Mass spectrometry

FIDs

„Effluent exits column and enters an air/hydrogen flame

„The gas-phase solute is pyrolized to form electrons and ions

„All carbon species are reduced to CH2+ions

„These ions collected at an electrode held above the flame

„The current reaching the electrode is amplified

to give the signal

FID

„A general detector for organic compounds

„Very sensitive (10-13g/s)

„Linear response (107)

„Rugged

„Disadvantage: specificity

ECD

„Ultra-sensitive detection of halogen-containing species

„Pesticide analysis

„Other detectors besides MS

„ IR

„ AE

SEMI- QUANTITATIVE ANALYSIS OF FATTY ACIDS

C

C

C Detector Response

Retention Time

14

16

18 Peak Area (cm )

Sample Concentration (mg/ml) 2

4 6 8 10

0.5 1.0 1.5 2.0 2.5 3.0

2

T h e c o n te n t % o f C f a tt y a c id s = C

C + C + C 1 4 ∗ 10 0

1 8

1 6

1 4

1 4

TENTATIVE IDENTIFICATION OF UNKNOWN COMPOUNDS

Response

GC Retention Time on Carbowax-20 (min)

Mixture of known compounds

Hexane Octane Decane 1.6 min = RT

Response

Unknown compound may be Hexane

1.6 min = RT

Trang 5

GC Retention Time on SE-30

Unknown compound RT= 4 min on SE-30

Response

GC Retention Time on SE-30

Hexane RT= 4.0 min on SE-30

1 Very good separation

2 Time (analysis is short)

3 Small sample is needed - μl

4 Good detection system

5 Quantitatively analyzed

DISADVANTAGES OF GAS CHROMATOGRAPHY

Material has to be volatilized at 250C without decomposition

R C OH CH 3 OH H 2 SO 4

O

R C O CH 3

O

CH 2 O C R

CH O C R

CH 2 O C R

O

O

O

CH 3 OH

O

R C O CH 3

CH 3 ONa

Fatty Acids Methylester

Reflux

Volatile in Gas Chromatography

Volatile in Gas Chromatography

Gas Chromatogram of Methyl Esters of Fatty Acids

Ngày đăng: 08/03/2014, 15:35

Xem thêm

TỪ KHÓA LIÊN QUAN

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN