RESULTS AND DISCUSSION Determination of Zn2+ and Cu2+ in rain water and leachate samples collected from LSSC Determination of Cu2+ in CRM samples of waste and ground water... EUH-3 Envir[r]

Trang 1

Presented by Thuan Hoa Nguyen

Department of Chemistry, Faculty of Science and Technology, Thammasat University

Sequential injection system with in-line solid phase extraction and soil minicolumn for determination of zinc and copper

in soil leachates Justyna Paluch, Raquel B.R Mesquita, Víctor Cerdà, Joanna Kozaka, Marcin Wieczorek, Antóni O.S.S Rangel

Trang 2

I ntroduction

Trang 3

CD solution Healthy body

The liver

The kidneys

The nervous system

 toxic to the cytologic and physiologic

industrial, urban, agricultural,

gardening, forestry

zinc copper Levels of can be tolerated without risk to health Copper and Zine up to 1.5 and 5 mg/liter

The effective monitoring of waters affected by soil leachates

Trang 4

3 Mass spectrometry (ICP-MS)

4 Neutron activation analysis (NAA)

4

Trang 6

at 482 nm

Trang 7

Experiment

Trang 8

CD solution

SIA - SPE method for copper and zinc determination

a) Off-line laboratory scale soil core (LSSC)

b) In-line soil mini-column (ILSM)

8

Trang 9

Two types of soil were used:

 Forest soil in Northwest Portugal

 Agricultural soil in Northwest Portugal

acidified to pH ≈ 2 before determination

laboratory scale soil core

Trang 10

CD solution

The In-Line Soil Mini-columns (ILSM)

Two types of soil were used:

 Forest soil in Northwest Portugal

 Agricultural soil in Northwest Portugal

(h = 3 cm, d = 2 cm) with surface soil with about 3 g

acidified to pH ≈ 2 before determination

10

Trang 11

CD solution

SIA - SPE method for copper and zinc determination

a) Off-line laboratory scale soil core (LSSC)

b) In-line soil mini-column (ILSM)

Trang 13

Buffer

Off-line laboratory scale soil core (LSSC)

The Chelex 100 chelating resin

( Zn2+)

Trang 15

RESULTS and

DISCUSSION

Trang 16

RESULTS AND DISCUSSION

CD solution

Bi parametric determination of zinc and copper

Calibration curves for Zn 2+

for the sum of Zn 2+ + Cu 2+

Calibration curves for Cu 2+

Concentration values for Zn 2+

for Cu 2+

16

Trang 17

RESULTS AND DISCUSSION

Trang 18

RESULTS AND DISCUSSION

Trang 19

RESULTS AND DISCUSSION

CD solution

Interference assessment

% interference = absorbance for analyte (interfering ion) – absorbance for analyte

absorbance for analyte x 100

The developed SI-SPE method can be used for determination

of zinc and copper in rain water and soil leachate samples with

No significant interferences.

Trang 20

RESULTS AND DISCUSSION

CD solution

Accuracy assessment

 Both analytes can be determined with satisfactory accuracy

20

Trang 21

RESULTS AND DISCUSSION

CD solution

Determination of Zn2+ and Cu 2+ in rain water and leachate samples collected from

LSSC Determination of Cu2+ in CRM samples of waste and ground water

Trang 22

RESULTS AND DISCUSSION

Trang 23

RESULTS AND DISCUSSION

CD solution

 Analytical Application of the As-Obtained CDs

Trang 24

CD solution

CONCLUSIONS

• Small amount of sample and reagent

• Low detection limits

Trang 25