Tài liệu hướng dẫn vận hành access 2 This document is not intended to replace the information in your instruments Instructions for Use, Quick Reference Guide or Help System Documentation. Information in any of the above documentation supersedes information in any other manual.
Trang 1Access 2 Immunoassay System
In-Lab Training Manual
_
Trang 3Access 2 Immunoassay System
WARNINGS AND PRECAUTIONS
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate the instrument
Beckman Coulter, Inc urges its customers and employees to comply with all national health and safety standards such as the use of barrier protection This may include, but is not limited to, protective eyewear, gloves, suitable laboratory attire when
operating or maintaining this or any other automated laboratory equipment
INTENTION FOR USE
This document is not intended to replace the information in your instruments Instructions for Use, Quick Reference Guide or Help System Documentation Information in any of the above documentation supersedes information in any other manual
REVISION STATUS
Version 1.0 (October 2017)
Access 2 Immunoassay System Software version 3.5.0
Access 2 Immunoassay System Instructions for Use C02062AA
Access 2 Immunoassay System Operators Guide C02060AB
Access 2 Immunoassay System Reference Manual C02061AA
TRADEMARKS
Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc in the United States and other countries
All other trademarks are the property of their respective owners
Trang 5Access 2 Immunoassay System In-Lab Training Manual
Table of Contents
Chapter 1 Analyzer Overview 9
System Hardware 10
Access 2 Systems with Enhanced Features 12
Communicating with the Access 2 13
Sample Processing Sequence 14
Assay Technology 15
Chapter 2 Software Overview 19
Common Screen Elements 20
Main Menu Function Buttons 24
Software Navigation 25
Keyboard Function Keys 26
Filters 27
Printing 27
Using the Help System 27
Chapter 3 Daily Start Up 29
Check System Status 30
Check Zone Temperatures 30
Check System Supplies: Wash Buffer 30
Check System Supplies: Empty the Liquid Waste 31
Substrate Handling Conditions 33
Check System Supplies: Prime Substrate 33
Check System Supplies: Load Reaction Vessels 35
Check System Supplies: Change the RV Waste Bag 37
Reagent Pack Handling Conditions 38
Reagent Inventory 38
Check System Supplies: Load Reagent Packs 39
Check System Supplies: Unload Reagent Packs 42
Confirm System Backup 42
Setup an Automatic Backup 43
Inspect the Fluidics Module 44
Trang 6Clean the Wash Carousel Probe Exteriors 45
Daily Clean System Routine 46
Special Clean Routine 47
Calibration Test Requests 48
Quality Control Test Requests 49
Chapter 4 Sample Management 51
Sample Racks and Containers 52
Calculating Sufficient Sample Volume 52
Processing LIS Test Requests 55
Processing Manual Test Requests 56
Additional Sample Programming Features 57
Rerunning a Test 58
Manually Add on Test(s) 58
Chapter 5 Data Management 59
Calibrator Setup 60
Reviewing Calibration Data 61
Setting up Quality Controls 63
Reviewing Quality Control Data 65
Reviewing Patient Data 67
Test Result Reports 69
Applying a Test Results Filter 71
Chapter 6 Weekly Maintenance 73
Clean the Instrument Exterior 74
Inspect the Liquid Waste Bottle for Wear 74
Check for Fluid in the Waste Filter Bottle 75
Insect and Clean the Primary Probe 76
Replace Aspirate Probes 77
Clean the Aspirate Probes 79
System Verification following Weekly Maintenance 82
Run the System Check Routine 83
Reviewing and Printing System Check Data 84
Chapter 8 Troubleshooting 85
Enabling the Utility Assay 86
Run the Utility Assay 86
Resolving Sample Errors 87
Trang 7Troubleshooting Events 88
Troubleshooting System Check Failures 89
Troubleshooting Calibration Failures 91
Troubleshooting Quality Control Failures 92
Test Result Flags 93
Chapter 9 Resources 95
Beckman Coulter Support Personnel 95
References 96
Chapter 10 In-Lab Training Competency Exercise 99
In-Lab Training Competency Exercise Reference Key 111
Trang 9Chapter 1 Analyzer Overview
Introduction
The Access 2 is an automated immunoassay system that measures analytes in
samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories The system is for in vitro diagnostic use only The Access 2 performs automated analysis of serum, plasma, urine, and amniotic fluid This chapter will cover the following:
System hardware components
Sample processing overview
Assay technology and process overview
Understanding the function of hardware components in the processing of samples on the system can assist in efficient operation, as well as with performing maintenance and troubleshooting
Practice
Use the information presented in the following pages to identify the hardware
components on the system
Review the sample processing sequence overview
Identify the assay technology and the processing steps
NOTE: Additional learning materials can be found on the Beckman Coulter website www.beckmancoulter.com
Trang 10System Hardware
Carousel Module Main Pipettor Analytical Module Fluidics Module
Trang 11System Hardware (continued)
Internal Bar Code Reader
Cup / Tube Detector
Incubator Track and Belt
Wash / Read Carousel
Wash Pump and Valve
Probe Wash Tower
Substrate Pump, Valve and Heater
Vacuum Pump and Reservoir
Peristaltic Waste Pump
Fluids Tray (Wash Buffer Reservoir, Waste Filter Bottle, Liquid Waste Bottle and Substrate Bottles)
Trang 12Access 2 Systems with Enhanced Features
If your Access 2 system has a serial number ≥ 57000, some of the external components are configured differently and a status light bar has been added All of the system processes are the same and diagrams are included in this manual to illustrate the
differences in the configurations An internal reagent bar code reader has been added
to facilitate loading of multiple reagent packs The software on these systems reflects
this capability The system is also PROService ready, please contact Technical
Support if you are interested in utilizing this feature
System Modes and their associated Status Light Bar colors
System Mode Status
Trang 13Communicating with the Access 2
Following the communication sequence in the table below will insure that the instrument will update supply levels, track expirations and process samples accurately
Tell = Function button selection by operator
Wait = Instrument may be busy performing previously requested action System readiness will be indicated when a pop-up box appears with
instructions
Do = Operator performs action as directed by pop-up box instruction
Done = Operator selects Done (Yes) or Cancel (No)
*In many instances Run will need to be selected to initiate system action
Trang 14Sample Processing Sequence
NOTE: The order of sample and reagent addition is assay specific
1
The operator places samples into an appropriate rack Using the Access 2 software
(TELL…WAIT…DO…DONE), the operator loads the sample rack onto the Sample
3 The Cup/Tube Detector determines if a sample is present The Internal Bar Code
Reader scans the sample IDs Test orders are matched with each rack position
4
The sample to be tested (in the Sample Carousel) and the appropriate reagent pack (in the Reagent Carousel) are moved in line with the Main Pipettor Module and the row of reaction vessels (RVs) in the RV Shuttle
5 The Primary Probe is cleaned using the Ultrasonic Transducer and flushed with wash buffer in the Probe Wash Tower
6 The Primary Probe moves along the Pipettor Gantry towards the sample and the
Ultrasonic Transducer is used for level sensing
7
The Primary Probe aspirates the assay-specific sample volume then dispenses the sample aliquot into the first RV in the RV Shuttle The Primary Probe is cleaned at the Probe Wash Tower
8
The Primary Probe moves to the Reagent Carousel and the Ultrasonic
Transducer is used to re-suspend the paramagnetic particles in the reagent pack
The Primary Probe aspirates reagent and dispenses the reagent into the RV
containing the sample aliquot The Primary Probe is cleaned at the Probe Wash
Tower
9
The Primary Probe moves to the Reagent Carousel to aspirate the second reagent
(conjugate) and then dispenses the conjugate into the RV containing the sample and
paramagnetic particles The RV contents are mixed by the Ultrasonic Transducer
10 The Primary Probe is cleaned at the Probe Wash Tower
11 The RV Shuttle transfers the RV into the Incubator Belt
12 The RV remains in the Incubator Belt (while the belt moves back and forth) for an
assay-specific incubation time during which the antibody-antigen reactions occur
13 After the incubation time is complete, the RV is transferred to the Wash/Read
Carousel
14
Three washing steps (Wash Arm moves up and down as wash buffer dispense is
followed by aspiration in a magnetic field) are performed These washes remove the unbound material from the RV while retaining material that is bound to the
paramagnetic particles After the last wash, Substrate is added
15
The contents of the RV are incubated while the RV is moved toward the
Luminometer During this incubation, substrate reacts with alkaline phosphatase
bound to the paramagnetic particles to produce light The Luminometer measures
the light in Relative Light Units (RLUs)
16 The RV is transferred from the Wash/Read Carousel back into the Incubator Belt
17 When the RV aligns with the RV Shuttle, the RV is transferred by the RV Shuttle from the Incubator Belt into the RV Waste Bag (by way of the waste chute)
Trang 15Assay Technology
Immunoassay is an analytical method that uses antigen-antibody reactions to detect or measure a specific analyte in a sample of body fluid
The Access 2 uses enzyme-mediated chemiluminescence
Chemiluminescence is a chemical reaction in which one of the end products is the emission of light
The light emitted is measured in Relative Light Units (RLU)
pipettor into a reaction vessel (RV) The main pipettor mixes the contents in the
RV
2 The RV is incubated for the required time for the assay
3 Following incubation, the RV is transferred to the wash/read carousel where it is washed three times in a magnetic field to remove unbound material
4 Substrate is dispensed into the RV
5 The RV is incubated to allow the bound analyte complexes to react with the substrate
6 The light generated by the reaction is read by the Luminometer The RLU
reading is converted to analyte concentration using the calibration curve
Access 2 Enzyme-Mediated Chemiluminescence Process
Trang 16Assay Technology (continued)
Sandwich Assay
Coated paramagnetic particles bind to an analyte to be measured in a sample
A second reagent (called conjugate) also binds to the analyte
The analyte is bound between the particle on one side and the conjugate on the other (like a sandwich)
The conjugate has alkaline phosphatase bound to it This enzymatically triggers the substrate to produce a glowing light
The Relative Light Units (RLUs) produced are directly proportional to the
amount of analyte in the sample (the more RLUs, the higher the analyte
concentration)
The system uses a calibration curve to convert the measured light emission to
an analyte concentration
Trang 17Assay Technology (continued)
Competitive Binding Assays
Coated paramagnetic particles bind to an analyte to be measured in a sample
A second reagent (called conjugate) also binds to the coated paramagnetic particles (the enzyme labeled conjugate is the same analyte as the analyte being measured)
The analyte from the sample and the conjugate compete for binding sites on the coated paramagnetic particles
The conjugate has alkaline phosphatase bound to it, this enzymatically triggers the substrate to produce a glowing light
The Relative Light Units (RLUs) produced are inversely proportional to the
amount of analyte in the sample (the more RLUs, the lower the analyte
concentration)
The system uses a calibration curve to convert the measured light emission to
an analyte concentration
Trang 19Chapter 2 Software Overview
Introduction
The software is used by the operator to interact with the analyzer and perform system operations Being familiar with the software can assist in understanding and efficiently operating the system
This chapter will cover the following:
Common Screen Elements
Main Menu Function Buttons
Select Main Menu Function Buttons to navigate and review the screen specific
information displayed there
Access the Help button to identify the information available on the system
Select a Show Me Help Link Button to display an illustration contained within a help procedure
Use a filter button to increase or reduce the amount of information displayed in the Test Results screen
Return to the Main Menu from any other screen by selecting F9 on the keyboard
Trang 20Common Screen Elements
The Access 2 displays certain elements on every screen so you can continuously
monitor the system’s status
Trang 21Common Screen Elements (continued)
No new tests are scheduled, but currently scheduled tests continue processing
The system is not ready to process samples and will require initialization
NOTE: On Access 2 systems equipped with a status light bar, the light bar color will change to match the system operating mode background color
System Command Buttons
System command buttons are used to run, pause and stop the instrument
Select to process samples or Run a maintenance routine
Select to Pause the instrument The instrument stops pipetting after it
finishes pipetting the current sample and processing continues on samples already in progress
Select to Stop the instrument If a popup is confirmed, the system
immediately stops processing and cancels any tests in progress
Trang 22Common Screen Elements (continued)
System Status Button Background Colors
Most of the system status buttons are color coded to alert the operator to a system condition that requires attention
NEUTRAL Normal operational conditions No Action is required
YELLOW CAUTION condition or event has occurred that requires investigation
RED WARNING condition or event has occurred that requires immediate
attention
Status Button examples:
Neutral Substrate status button No action is required, there is sufficient volume and time to expiration for the on board bottle of substrate reagent
Caution RV status button This condition change indicates a low number of RVs remaining on board the system RVs should be loaded soon to accommodate test requests and prevent the status from progressing to a Red Warning condition
Warning Event Log status button The Event Log status button should
be selected to determine if an event has occurred that requires the operator’s attention before continuing with operations
Status Line
This area displays the instrument serial number, software version and the current date and time It may also display other information such as what activity the system is currently performing or a message prompt for the operator
Trang 23Common Screen Elements (continued)
System Status Buttons
The ten system status buttons can be viewed at the top of every screen The buttons change color to inform the operator of a Caution (Yellow)or Warning (Red)condition
or event Select each button to view its related supply status
Liquid Waste
Changes color (Red) to indicate a critical high level
Event Log
Changes color (Yellow or Red)
to indicate that a caution or warning event has occurred Select to view informational, caution and warning events generated by the system
Substrate
Changes color (Yellow or Red) to indicate a low volume of tests remaining Select to view the Supplies screen.
Work Pending
Changes color (Yellow) to indicate a sample is required Select to view details
Reaction Vessels (RVs)
Changes color (Yellow or Red) to indicate a low number of remaining RVs Select to view the Supplies screen.
Supplies Required
Changes color (Yellow) to indicate that supplies and/or calibrations are needed to process the requested tests Select to view details.
RV Waste Bag
Changes color (Yellow or Red) to indicate a low number of available space in the RV waste bag Select
to view the Supplies screen.
Help
Select to display specific topics or to access the entire Help system Button remains Neutral (does not change color).
Trang 24screen-Main Menu Function Buttons
The Access 2 system displays a row of eight function buttons across the bottom
View or print test results
View and manage system supplies, waste containers and reagent packs
Configure quality controls and review quality control results
Add/delete calibrator lots and view/print calibration reports
Review system temperatures, enable/disable Utility Assay, review / print results for non-routine maintenance procedures
Access a menu for system initialization, fluidics priming and additional diagnostic procedures
Access a menu for system configuration, assay configuration, and
PC Administration functions (e.g., data backup, system shutdown)
Trang 25Software Navigation
The Access 2 Immunoassay System has three navigation methods that you can use in any combination: touch screen, mouse and keyboard
allow for easy movement between screens
appears on the right side of a screen
Scroll bars appear when there is more information than can be contained in the viewing pane
Trang 26Keyboard Function Keys
The keyboard function keys are located at the top of the keyboard and a template has been applied to simplify identification (P/N 973041)
The following table lists the keyboard equivalents you can use to perform
common actions
Keyboard
F1 – F8 Perform the indicated function or go to the indicated screen
Down
Arrow Display a list in the field
Pause Pause the system
Shift-Pause Stop the system
Screen Print what is currently displayed on the screen
Scroll Lock View the Event Log
Trang 27Filters
allowing the user to expand or reduce the amount of information displayed
Many filters have a sort order feature for selecting the display order of the filtered data
Custom filters can be created to easily find frequently sought information
Printing
Print function buttons are available on many screens; select the
function button on the screen to print a screen specific report
The Print Screen key on the keyboard will print a copy of the screen being
displayed
The Help Window Button Bar is available for printing when in the Help system
Using the Help System
Access the Help System by selecting the Help Button on the screen or by
selecting F12 Help on the keyboard
When you select Help from a screen the information displayed is specific to the screen you are on
Green Text Linkscan be selected to link to the topic named in the text
contents of the Help System
The Help System is intended as a reference only You must return to the User Interface (UI) to perform any function
Exit the Help System by selecting the button in the upper right corner of the help window or selecting any visible portion of the UI beyond the help
window
Trang 28Using the Help System (continued)
The nine Help Link Buttons depicted in the table below indicate links to related
Flowchart
Select to display an overview or procedure flowchart You can find this button at the top of a procedure window
Example
Select to display a relevant example
Select to display another topic containing more information about the selected topic
Forward Select to display a related topic
More Info Select to display another topic containing
more information about the selected topic
Show Me
Select to display an illustration of the procedure or step You can find the button anywhere in the Help system text, where a graphic would be helpful
Navigation
Select to display the associated screen or window navigation topic Navigation topics include directions on how to access the screen or window, lists of required and optional fields, and keyboard equivalents
Screen View
Select to display a graphic of the associated screen or window After the screen or window is displayed, you can move the cursor over the graphic to find out what areas of the graphic are selectable
Glossary Select to display the definition of the
glossary term (used only in the Glossary)
Trang 29Chapter 3 Daily Start Up
Introduction
A daily maintenance should be performed on the Access 2 Immunoassay System after the longest period of inactivity and prior to processing patient samples The following system checks and maintenance tasks are required for daily start up:
Check System Status
Inspect Fluidic Module
Clean the Wash Carousel Probe Exteriors
Run the Daily Clean System or Special Clean Routine
Document task performance on the Maintenance Log
Step/Action Table Information
Step/action tables are provided for each daily start up task in this chapter These
step/action tables are available as a quick reference for the daily start up procedures Review your instruments’ Instructions for Use for any Notes, Cautions, and Warnings as well as step-by-step instructions For complete instructions to the daily start up
procedures, refer to:
From ANY screen:
Based on your laboratory workflow, which shift will be in charge of daily
maintenance?
Trang 30Check System Status
Checking the system status is comprised of three tasks: Check Zone Temperatures, Check System Supplies and System Backup
Check Zone Temperatures
2 Verify that the temperatures in the Zone Temperature fields are within
their expected ranges
NOTES
Out of range temperatures are displayed in red
Contact Technical Support if any of the zone temperatures are out of range
Check System Supplies: Wash Buffer
The wash buffer bottle drains into the wash buffer reservoir The wash buffer bottle needs to be replaced when the bottle is empty You can change the wash buffer at any time Follow the step/action table below to change the wash buffer bottle
1 Mix the contents of the new Wash Buffer Bottle by gently inverting the
sealed bottle three or four times
3 Lift the empty bottle from the fluids tray reservoir
and screw the dispense cap assembly on the new bottle
into the reservoir receptacle
Trang 31Check System Supplies: Empty the Liquid Waste
NOTE: If your system is plumbed to a drain, this maintenance task will not be
necessary
No software interaction is required to empty the liquid waste bottle Depending on the system mode, follow the appropriate step/action table below Your laboratory
procedures will determine how to dispose the contents of the Liquid Waste Bottle
Procedure A – Access 2 Ready OR Not Ready
1 Detach one tubing on the bottle cap by pressing the button on the quick
disconnect (see image on following page) while pulling on the tubing
2 Immediately connect this tubing by snapping it into the quick disconnect
on the cap of a clean liquid waste bottle
4 Make sure that the clean, connected liquid waste bottle is correctly
placed into the fluids tray
5 Dispose of the liquid waste bottle contents Thoroughly rinse the bottle
and cap with tap water and invert to dry
Procedure B – Access 2 Running OR Paused
1 Remove the liquid waste bottle cap with the two tubes still attached to the
cap (see image on following page)
2 Immediately screw this cap onto a clean liquid waste bottle
CAUTION Take care to orient the underside of the cap away from you Liquid waste
is continuously expelled while the system is processing samples
3 Make sure that the clean, connected liquid waste bottle is correctly
placed into the fluids tray
4 Dispose of the liquid waste bottle contents Thoroughly rinse the bottle
with tap water and invert to dry
Trang 32Empty the Liquid Waste (continued)
NOTE: The fluids tray containing the liquid waste bottle has two configurations (A or B)
The connections to the liquid waste bottle are similar
Trang 33Substrate Handling Conditions
Equilibration prior to use
(unopened) (room temperature) 15-30°C Minimum 18 hours Maximum 14 days
In use (opened) External fluids tray substrate position Maximum 14 days
Check System Supplies: Prime Substrate
Substrate should be primed once every 24 hours; by either loading a new bottle or priming the onboard bottle
Substrate priming takes approximately eight minutes During this time the
instrument software cannot be used, however the Help System can be accessed
To load a new bottle of substrate follow the step/action table below
1 Confirm that the system is in the Ready mode
4 WAIT until the screen displays the Change Substrate window
5 Enter the bar code of the new unopened equilibrated substrate bottle by
scanning the bottle with the handheld bar code reader
6 Remove the cap from the new equilibrated substrate bottle
NOTE Avoid substrate contamination by not popping bubbles, contacting bottle neck
area, touching inside the supply cap or wiping the straw
7
Unscrew the substrate supply cap from the bottle in use and immediately screw
it onto the new bottle Place the new bottle into the fluids tray A new bottle of substrate contains 600 tests
8 Select Done F1 Select YES F1 to request priming The system mode screen
will change from Ready to Running while the prime is being performed
Trang 34Prime Substrate (continued)
If sufficient volume is available in the onboard bottle, prime the substrate following the
step/action table below
4 Select the Substrate box Leave the default setting of 4 priming cycles
5 Select Start Priming F2 32 tests are used when priming through the
fluidics
Done F8 to return to the Main Menu
NOTE: There are two configurations (A or B) of the substrate location in the fluids tray
Trang 35Check System Supplies: Load Reaction Vessels
Access 2 RV cartridges contain 98 reaction vessels attached to a support spine
Maximum capacity is 3 cartridges for a total of 294 reaction vessels
4 Confirm that all RVs are firmly seated on the cartridge spine by pressing
down on the spine before you load the cartridge
on the next page)
from the cartridge case Load into the RV Supply Area
7
Release the cartridge spine from the RVs by pressing down firmly on the RV Load Door Open the RV Load Door and remove the empty spine to prevent possible damage to the RV rake
NOTE Only load full cartridges Fill in any missing RVs noted when spine is
removed
Trang 36Load Reaction Vessels (continued)
Trang 37Check System Supplies: Change the RV Waste Bag
Waste RVs are collected on the system in a waste bag Follow the step/action table
below to change the RV waste bag through the software
5 Unfold a new waste bag, grasp its sides, and pull open gently to expand
it
collar until it is free of the waste chute
NOTE RV waste vessels will continue to be ejected (every 36 seconds) when the instrument mode is
Running or Paused
7 Remove the RV in the exit position of the waste chute to ensure that it
does not get pushed back into the system
8
Place the new bag into position by sliding the plastic collar on the bag into the slot on the waste chute Press down firmly to be sure that the collar is completely seated
biohazardous waste bag according to lab policy
count will update to 300 spaces available
NOTE From the Ready mode, the system will not start sample processing
without a waste bag in place
Trang 38Reagent Pack Handling Conditions
The reagent pack
has been previously
Bar Coded: Pack Name, Lot Number, Serial Number and Lot Expiration Date
Non Bar Coded: Number of Tests, Number of Days Open, Pack Expiration, Calibration Expiration, Slot Position and Instrument
Available Actions
From the Supplies screen, select Reagent Inventory F7 to view information
for individual reagent packs
Load Reagent Pack F1 Select to load a reagent pack
Unload Reagent Pack F2 Select to unload a reagent pack
Delete Pack F5 Select to delete a pack from inventory
Print F7 Select to print reagent inventory
Trang 39Filter F8 Select to see All, On-board or Off-board packs
Check System Supplies: Load Reagent Packs
Follow the step/action table below to load reagent packs and manually enter bar code
information
2 Select Load Reagent Pack F1 If the system is running, it will pause to
allow for reagent pack loading
4 Slide the carousel door to the left to open it Open the reagent carousel
door
NOTES
Gently invert unpunctured reagent packs several times prior to loading
Do not invert punctured reagent packs, as contents may leak
NOTE If the bar code will not scan, enter the characters above the pack bar
code into the Bar Code field.
6 Close the reagent carousel door Slide the carousel door to the right to
close it
Trang 40Load Reagent Packs (continued)