 According to the guide line of items, clean the items with adequate detergent and remove blood, species and stains.. Only dried items can be loaded in the sterilizer..  Insuffic

Trang 1

(HOW TO OPERATE HMTS-80)

Quick Operation Guide

HUMANMEDITEK

Customer Service Team

Trang 2

HMTS-80 구구

Chambe

r

ON,

OFF,

CANCEL

Therm

al

Printer

HPS Unit

SMC Card

Drain Unit

Trang 3

According to the guide line of items,

clean the items with adequate detergent

and remove blood, species and stains.

Dry all of medical instruments Only dried items

can be loaded in the sterilizer.

it can be needed to disassemble it Follow the guide line

of the items.

Insufficiently dried items can make some errors

Examine repeatedly sterilized items if they are defected

by chemicals.

the items.

PREPERATION OF ITEMS

Trang 4

Absorbable Material

- Wooden articles, Linen, Paper, Sponge

Disposable suture material

Wet items

Oil

Powder

Instruments with lumen

- Hard Lumens

: An inside diameter of 1 mm or shorter and a length of 500 mm or larger

- Flexible Lumens

: An inside diameter of 1 mm or shorter and a length of 10000 mm or larger

CAN’T BE STERILEZED

Trang 5

LOADING METHOD

Running Instrument

Tray Setting

DOOR

The position of

SCBI Ampoule

Cleaning/ Rinse/ Drying

CI Strip

Loading

in the Chamber

Sealing (temp 110~130℃)

Trang 6

OPERATION

1 Vacuum Ⅰ

37 1:24 8000 38

mTorr

Trang 7

HOW TO CHANGE THE STERILANT

2.Open the side door and remove the old bottle

4 Insert a new bottle matching with the red point

5 Select “SETUP”

button 6 Put PASSWORD(1212) in SET-CLR-PASSWORD-ENT 7 Touch” Bar code” button 8 The lift bottle -pusher automatically comes down

3 Remove the black lid

1 Check the message

Trang 8

HOW TO DRAIN (MANUAL)

1 Turn “ON”

5 Press “Drain”

4 Put in PASSWORD

SET-CLR-PASSWORD-ENT

6 Press “YES”

2 Select “MODE” 3 Select “SETUP”

Trang 9

HANDLING OF THE DRAIN BOTTLE

1 Open the drain door Whenever you change new steriliant for a checking

3 Prepare a bottle with 100ml

of water to the arrow mark

5 Close the door

2 If the bottle is full (to the red line ), Empty the bottle

4 The both arrow marks on the bottle and the unit should

be matched

Trang 10

An empty hydrogen peroxide bottle should be handled

by regulations.

CAUTIONS

Hydrogen peroxide is an acidic chemical It can be caused of

chemical burn on your skin or irritation in your eyes.

If you were touched by hydrogen peroxide on your skin or eyes please flush with large amount of water immediately (at least 30 minutes for eyes, 15minutes for skin) Get medical aid.

A red button on the sterilant bottle should be matched with the red button on the HPS unit.

Before refill and withdrawal of the hydrogen peroxide bottle,

please put personal protective equipment

Trang 11

1 Open the cover 2 Remove the old roll

3 Put the new roll in,

then pull out the paper

through the space

4 Close the cover REPLACEMENT OF PRINTER PAPER

Right direction

Trang 12

PRINT PAPER COUNT SETTING

1 Turn “ON” 2 Select “MODE” 3. Select “SETUP”

6 Press “NEXT” 7 Set count

5 Put in PASSWORD(1212)

SET-CLR-PASSWORD-ENT

Trang 13

CANCEL

Press the Cancel

button

Touch MODE Select MONITOR

After Plasma II & Vent / Dry Stage is completed, the door will

be opened It takes about 8 minutes

After finishing the cycle,

the items must be re-packed and re-sterilized.

Green color 1) Plasma II 2) Vent / Dry

6 1

40

Trang 14

WHEN AN ERROR WAS OCCURED

After a printer paper comes out, the door will be opened.

Wait until the process is finished

When an error is occurred ,

HMTS-80 goes to Plasma Phase II and vent & dry phase automatically (Except Error 1, 10)

Please refer to the Trouble - shooting.

Trang 15

STORAGE OF ITEMS

Staffs only enter to the storage room.

Store sterilized items at least 20cm from the bottom, at

least 50cm from

the ceiling, at least 5cm from the wall.

Place sterilized items on the shelves to identify expiration date and

chemical indicators(CI).

Place sterilized items with short expiration date on the front

Consider that items are contaminated if they are torn or

wet.

Consider that items are contaminated if there is no CI or it is opened.

Do not place sterilized items with non-sterilized one

together.

Trang 16

THE FACTORS OF EXPIRATION DATE OF

STERILIZED ITEMS

The material and type of packages.

The number of handling the sterilized item before its usage.

The number of staff dealing with the sterilized item.

The shape of cabinet(open or close).

The environment of storage.

(eg The status of Cleaning, temperature, humid)

Use of a dust protective cover.

Trang 17

Customer Service Center

+ 82 2 6292 2682/2683

www.hmmdt.com

cs @ hmmdt.com

Trang 18