Part 1 book A practical approach to clinical echocardiography presentation of content: Echocardiography basic principles, technique, Dis; hemodynamic evaluation by EchoDoppler techniques; mitral stenosis, mitral regurgitation, aortic valve stenosis, tricuspid valve, pulmonary valve, evaluation of prosthetic heart valves,... And other content
Trang 3Jagdish C Mohan MD DM FASE
Director of Cardiac Sciences
Fortis Hospital Shalimar Bagh, New Delhi, India
Foreword
Bijoy K Khandheria
JAYPEE BROTHERS MEDICAL PUBLISHERS (P) LTD
New Delhi • London • Philadelphia • Panama
®
Trang 4Jaypee Brothers Medical Publishers (P) Ltd
Headquarters
Jaypee Brothers Medical Publishers (P) Ltd
4838/24, Ansari Road, Daryaganj
New Delhi 110 002, India
Phone: +91-11-43574357
Fax: +91-11-43574314
Email: jaypee@jaypeebrothers.com
Overseas Offices
J.P Medical Ltd Jaypee-Highlights Jaypee Medical Inc.
83, Victoria Street, London Medical Publishers Inc The Bourse
SW1H 0HW (UK) City of Knowledge, Bld 237 111 South Independence Mall East Phone: +44-2031708910 Clayton, Panama City, Panama Suite 835, Philadelphia, PA 19106, USA Fax: +02-03-0086180 Phone: +1 507-301-0496 Phone: +1 267-519-9789
Email: cservice@jphmedical.com
Jaypee Brothers Jaypee Brothers
Medical Publishers (P) Ltd Medical Publishers (P) Ltd
17/1-B Babar Road, Block-B Shorakhute, Kathmandu
Shaymali, Mohammadpur Nepal
Dhaka-1207, Bangladesh Phone: +00977-9841528578
Mobile: +08801912003485 Email: jaypee.nepal@gmail.com
Email: jaypeedhaka@gmail.com
Website: www.jaypeebrothers.com
Website: www.jaypeedigital.com
© 2014, Jaypee Brothers Medical Publishers
The views and opinions expressed in this book are solely those of the original contributor(s)/author(s) and do not necessarily represent those of editor(s) of the book.
All rights reserved No part of this publication may be reproduced, stored or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise, without the prior permission in writing of the publishers.
All brand names and product names used in this book are trade names, service marks, trademarks or registered trademarks of their respective owners The publisher is not associated with any product or vendor mentioned in this book
Medical knowledge and practice change constantly This book is designed to provide accurate, authoritative information about the subject matter
in question However, readers are advised to check the most current information available on procedures included and check information from the manufacturer of each product to be administered, to verify the recommended dose, formula, method and duration of administration, adverse effects and contraindications It is the responsibility of the practitioner to take all appropriate safety precautions Neither the publisher nor the author(s)/ editor(s) assume any liability for any injury and/or damage to persons or property arising from or related to use of material in this book.
This book is sold on the understanding that the publisher is not engaged in providing professional medical services If such advice or services are required, the services of a competent medical professional should be sought.
Every effort has been made where necessary to contact holders of copyright to obtain permission to reproduce copyright material If any have been inadvertently overlooked, the publisher will be pleased to make the necessary arrangements at the first opportunity.
Inquiries for bulk sales may be solicited at: jaypee@jaypeebrothers.com
A Practical Approach to Clinical Echocardiography
First Edition: 2014
ISBN 978-93-5152-140-2
Printed at:
Trang 7tips Although there are no separate chapters on transesophageal echocardiography and real-time 3D imaging, these have been well covered in individual sections wherever appropriate The same principle has been applied to contrast echocardiography.
There is little doubt that cardiovascular ultrasound will continue its revolution, and play an important role in the practice of cardiology as well as medicine This book compiled by Prof Jagdish C Mohan is a welcome addition to the various sources of education in this field It is a must-read book for those wishing to practice the art and science of cardiovascular ultrasound
Bijoy K Khandheria MD FACP FAHA FACC FESC FASE
Adjunct Clinical Professor of MedicineUniversity of Wisconsin, School of Medicine and Public Health
Director, Echocardiography Services Aurora St Luke’s Medical Center, Milwaukee, Wisconsin, USADirector, Echocardiography Center for Research and Innovation
Aurora Research Institute, Milwaukee, Wisconsin, USAPast President, American Society of Echocardiography
Former Chair of Cardiovascular Disease, Mayo Clinic, Arizona
Former Professor of Medicine, Mayo Medical School
Trang 9used this technique year after year for assessment of left ventricular function only Echocardiography represents a strange mixture of several features which have proven prime-time use and an equally numerous techniques which have yet to show clinical utility despite extensive research work spanning decades There is an urgent need to promote that part of science which has robust validation What clinicians require needs to be emphasized with clarity Novices often get unnerved by the available applications on the systems which are more to stand up to the competition rather than
of tangible clinical value However, extensive research has shown applicability of a lot of echocardiographic knowledge and information in day-to-day evidence-based medicine Quantitative and semi-quantitative protocols are gaining ground because these make serial follow-up easily This book is an attempt to make echocardiography simple, practical and easily usable with reproducible data Unnecessary and impractical details have been omitted I would welcome suggestions to make it more readable and meaningful
Jagdish C Mohan
Trang 13• Hemodynamic Evaluation 24
• Doppler Principle for Estimating Velocity 25
• Ohm’s Law of Fluids (The Hagen–Poiseuille Equation) 27
• The Bernoulli’s Equation 28
• Estimation of Pulmonary Artery Pressures and Pulmonary Vascular Resistance 39
SECtIon 2: Valvular Heart Disease
• Anatomy of the Mitral Valve 45
• Functional Morphology of MR 64
• Anatomy of Trileaflet Aortic Valve 83
• Unicuspid Aortic Valve 85
• Bicuspid Aortic Valve 85
• Quadricuspid Aortic Valve 86
• Anatomy of Aortic Stenosis 86
Trang 14• Planimetry of the Aortic Valve Area on Aortic Stenosis 87
• Hemodynamics of Aortic Valve Stenosis 88
• Technical Tips 90
• Low-gradient, Low-flow Aortic Valve Stenosis 93
• Paradoxical Low-flow as with Preserved Ejection Fraction 95
• Subaortic Stenosis 95
• Hemodynamics of Aortic Regurgitation 98
• Functional Anatomy of Aortic Regurgitation 100
• Detection of Aortic Regurgitation by Echo-Doppler Techniques 102
• Volumetric Severity of Aortic Regurgitation 104
• Color Flow Doppler Evaluation of Aortic Regurgitation 104
• Aortic Regurgitation Assessment by Vena Contracta 105
• Proximal Isovelocity Surface Area Flow Convergence Method 105
• Diastolic Flow Reversal in the Descending Aorta and Severity of Aortic Regurgitation 106
• Pressure Half-time of Aortic Regurgitation Signal 107
• M-mode Color Flow Propagation Velocity 108
• Summary of Aortic Regurgitation Assessment by Echo-Doppler Methods 108
• Functional Morphology of the Tricuspid Valve 110
• Conditions Affecting Tricuspid Valve 113
• Remodelling in Tricuspid Regurgitation 114
• Functional TR 114
• Organic Tricuspid Valve Disorders 115
• Tricuspid Valve Stenosis 120
• Assessment of Severity of TR 121
• Vena Contracta 121
• Proximal Isovelocity Surface Area Method 122
• Anterograde Velocity of Tricuspid Inflow 122
• Hepatic Vein Flow in Assessment of TR 123
• Anatomy of the Pulmonary Valve 126
• Pulmonary Valve Disorders 127
• Transesophageal Echocardiography in Pulmonary Stenosis 133
• Consequences and Associations of Pulmonary Stenosis 134
• Assessment of Pulmonary Regurgitation 134
• Consequences of Pulmonary Regurgitation 136
• Ball-in-cage Valve 138
• Tilting or Mono-disk Valve 139
Trang 15SECtIon 3: Systolic and Diastolic Function
• Morphology of the Left Ventricle 165
• Limitations 186
11 Echocardiographic Assessment of Right Ventricular Function:
• Peculiarities of the Right Ventricle 189
• Function Parameters 190
• Right Ventricle Diastolic Function and Pressures 193
• Tricuspid Annular Peak Systolic Velocity 198
• Right Ventricle Strain and Strain Rate Analysis 200
• Physiology of Diastole 204
• Factors Contributing to Diastole 205
• Significance of Diastolic Function 205
• Isovolumic Relaxation Time 206
• Rapid Filling Phase 207
• Deceleration Time of Early Filling Wave (Mitral E- and Pulmonary D-Waves) 207
• Diastasis 208
• Atrial Kick or Contribution 209
• Tissue Motion and Diastolic Function 209
• Left Atrial Volume and Diastolic Function/Dysfunction 210
• How to Perform a Study Focusing on Diastolic Function 211
• Mitral Inflow Velocities 212
• Mitral Annular Velocities 214
• How to Obtain Annular Tissue Velocities 216
• Pulmonary Vein Flow and Diastolic Function 218
• Mitral Flow Propagation by Color M-Mode 221
• Longitudinal Strain, Rotation and Untwisting Rate by Acoustic Speckle Tracking 223
• Diastolic Stress Test 224
Trang 16SECtIon 4: Muscle Mechanics
• Concept of TDI 229
• Labeling of Tissue Velocity Waveforms 229
• Terms Used in Tissue Doppler Imaging 231
• Technical Details of TDI 231
• Myocardial Velocities in Short Axis 234
• Internal Dependency of Velocities 235
• Fundamental Basis of TDI 236
• Tissue Doppler Data Processing for Deformation Imaging 238
• Clinical Utility of TDI 239
• Prognostic Value of TDI in Diverse Cardiac Disorders 250
• Limitations of TDI 251
• Myocardial Deformation 257
• Descriptive Terms 261
• Acoustic Speckle Tracking 262
• Velocity Vector Imaging 264
• How to Perform Two-Dimensional Strain Imaging? 266
• Effects of Ischemia on Regional Deformation Metrics 268
• Early Systolic Longitudinal Lengthening (Paradoxical Longitudinal Systolic Strain) 269
• Post-systolic Longitudinal Shortening 270
• Paradoxical Strain Patterns 272
• Right Ventricular Deformation 273
• Left Atrial Function by Deformation Imaging 273
• Mechanical Properties of the Aorta 274
• Three-dimensional/Four-dimensional Deformation Imaging 274
• Clinical Applications of Strain and Strain Rate Imaging 276
• Limitations 277
• Fundamentals of Torsion 280
• Echocardiographic Methods of Studying Twist 284
• Clinical Applications of Torsion 285
• Aging and Torsion 286
• Aortic Stenosis, Hypertension and Hypertrophic Cardiomyopathy 286
• Heart Failure 287
• Constrictive Pericarditis 287
• Torsion and Cardiac Resynchronization Therapy 288
• Acute and Chronic Ischemia 289
• Grades of Diastolic Dysfunction and Twist 290
Trang 17• Carpentier Classification of Ebstein’s Anomaly 306
• Double-Chambered RV 307
• Sinus of Valsalva Aneurysms 308
• Tetralogy of Fallot 311
• Morphological Variants of TOF 312
• Truncus Arteriosus or Common Arterial Trunk 313
• Subaortic Membranous Stenosis 314
• Physiology and Pathology 314
• Transposition of Great Vessels 315
• Structure of Aorta 318
• Parts of Aorta 318
• Functions of Aorta 320
• Normal Aortic Measurements and Imaging Views 321
• Etiopathogenesis of Aortic Disorders 322
• Congenital Disorders of Aorta 322
• Aortic Coarctation 324
• Aneurysm of Sinus of Valsalva 325
• Aortopulmonary Window 329
• Truncus Arteriosus 329
• Supravalvular Aortic Stenosis 330
• Patent Ductus Arteriosus 330
Trang 18SECtIon 6: Structural Heart Disease
• Myocardial Segment Nomenclature and Coronary Vascular Territory 363
• Right Ventricular Segmentation 363
• Ischemia and Echocardiographic Imaging 365
• Echocardiography in IHD 369
• Evaluation of Myocardial Infarction 369
• Mechanical Complications of Myocardial Infarction 370
• Stress Echocardiography 377
• Cardiac Manifestations of Masses 383
• Cardiac Thrombi and Spontaneous Echo Contrast 384
• European Classification of Cardiomyopathies 397
• Summary of EMF Echocardiographic Features 406
Index 415
Trang 19Fundamentals of Echocardiography
Chapters
Ö Echocardiography: Basic Principles, Technique,
Trang 21Echocardiography is a technique of generating images of
the heart with the help of ultrasound (like skiagraphy is
performed with X-rays) Echocardiography has become
an integral part of clinical examination with its bedside
mobility and utility Practice of echocardiography
mandates the following steps:
• Understanding of ultrasound physics
• Knowledge of the instrumentation
• Acquisition skills
• Interpretative skills
• Sound knowledge of cardiovascular anatomy and
physiologies
• Adequate knowledge of common cardiac pathology
• Reporting, storage and retrieval skills
Ultrasound behaves very differently when it passes
through any tissue Many of the objects and artifacts seen
in ultrasound images are due to the physical properties
of ultrasonic beams, such as reflection, refraction,
diffraction and attenuation Indeed, physical artifacts
are an important element in clinical diagnosis (Fig 1.1)
Appreciating the phenomena created by ultrasound may
greatly benefit the patient in terms of increased accuracy
of interpretation and diagnosis The ultrasound wave
created for sending through the tissues is called incident
wave This is the wave whose behavior is changed by the
tissue
uLtrASound PHYSIcS
Sound
Sound is a form of mechanical energy that consists of waves
of compression and decompression of the transmitting medium, travelling at a fixed velocity (Figs 1.2 and 1.3).Sound is an example of longitudinal waves oscillating back and forth in the direction the sound travels, thus consisting of successive zones of compression
Fig 1.1: Behavior of the sound waves Sound wave after striking a surface gets reflected as well as transmitted into the other medium in a different direction (refracted) Total internal reflection without transmis- sion occurs if it strikes the interface at 90°.
Trang 22(high pressure) and rarefaction (low pressure) The
medium particles, however, show both longitudinal and
transverse oscillations
Longitudinal oscillations: The oscillating particles
of the medium are displaced parallel to the direction of
motion (direction of energy transfer)
Transverse oscillations: The oscillating particles of the
medium are displaced in a direction perpendicular to the
motion of the wave
Ultrasound: Sound waves with a frequency of 20000
c/s (20 KHz) are labeled ultrasound Ultrasound used in
medical diagnosis has a frequency of 1–10 MHz.1–3 These
are beyond the capacity of human ear to perceive
Why use ultrasound in Medicine?
With higher frequencies (shorter wavelengths), the sound tends to move more in straight lines like electromagnetic beams and is reflected like light beams It is reflected by much smaller objects (because of shorter wavelengths) and hence gives good spatial resolution If the wavelength
of the sound is smaller than the object, no noticeable diffraction occurs
wavelength = speed/frequency
Frequency: The frequency of an ultrasound wave
consists of the number of cycles or pressure changes that occur in one second (Fig 1.2) The units are cycles per second or hertz (Hz) Frequency is determined by the sound source only and not by the medium in which the sound is travelling
Propagation speed: Propagation speed is the rate
at which sound can travel through a medium and is typically considered 1,540 m/s for soft tissue The speed
is determined solely by the medium characteristics like density and stiffness Speed is inversely proportional to density and incompressibility
Pulsed ultrasound: Pulsed ultrasound describes a
means of emitting ultrasound waves from a source To achieve the depth of resolution required for clinical uses, pulsed beams are used Typically, the pulses are a millisecond or so long and several thousands are emitted per second (Fig 1 4)
Ultrasound interaction with tissue: As a beam of
ultrasound travels through a material, various things happen to it A reflection of the beam is called an echo,
Fig 1.2: Alternating compression and rarefaction of a medium as
sound passes through it. Fig 1.3:Compression is accompanied by high pressure and rarefaction by low Sound as an oscillating wave of mechanical energy
pressure.
Fig 1.4: The concept of pulsed ultrasound If an interface is closure
to the source of ultrasound, more pulses will be received per second
resulting in better resolution.
Trang 23a critical concept in all diagnostic imaging (Fig 1.5) The
production and detection of echoes form the basis of the
technique that is used in all diagnostic instruments A
reflection occurs at the boundary between two materials
provided that acoustic impedance of the materials is
different This acoustic impedance is a product of the
density and propagation speed
If two materials have the same acoustic impedance,
their boundary will not produce an echo If the difference in
acoustic impedance is small, a weak echo will be produced
and most of the ultrasound will carry on through the
second medium If the difference in acoustic impedance
is large, however, a strong echo will be produced If the
difference in acoustic impedance is very large, all the ultrasound will be totally reflected Typically in soft tissues, the amplitude of an echo produced at a boundary is only a small percentage of the incident amplitudes (Fig 1.6)
Strong reflections or echoes show on the ultrasound image as white and weaker reflections as gray (Figs 1.7A and B)
A sound wave will undergo certain behaviors when it encounters a tissue interface Possible behaviors include:
• Reflection off the tissue interface
• Diffraction around the interface
• Transmission (accompanied by refraction) into the interface or new medium (Fig 1.8)
Fig 1.5: The concept of an echo. Fig 1.6: Soft tissue acoustic interface producing some reflection
and more transmission If a medium has high acoustic impedance, stronger reflection will occur.
Figs 1.7A and B: (A) Tissue interfaces with variable acoustic impedance Higher impedance produces brighter echo due to more reflection
of ultrasound; (B) Refraction: Bending of the sound wave as it enters another medium Refraction is associated with decreasing speed and wavelength.
Trang 24Reflection of sound waves off of surfaces can lead to
one of two phenomena—an echo or a reverberation.
The reception of multiple reflections off of the interface causes reverberations—the prolonging of a sound (Figs 1.9 and 1.10)
Angle of incidence: If a beam of ultrasound strikes the
boundary at right angle, it will be reflected parallel to the transmitting beam and shall produce stronger echo (Fig 1.11) If it strikes a boundary obliquely, the interactions are more complex than for normal incidence (Fig 1.12) The echo will return from the boundary at an angle equal to the angle of incidence The transmitted beam will be deviated from a straight line by an amount that depends on the difference in the velocity of ultrasound
at either side of the boundary This process is known as refraction (Figs 1.1 and 1.7)
Fig 1.8: Decreasing wavelength after refraction in the second medium. Fig 1.9: Multiple linear shadows (arrows) in front of mitral prosthesis in
transesophageal echocardiographic (TEE) view due to rever berations.
Fig 1.10: Reverberations shown by the curved arrow resulting in a
mirror-image artifact.
Fig 1.11: Reflection from a tissue interface at right angle There is stronger reflection resulting in brighter echo Typically, from posterior pericardium.
Fig 1.12: Reflection at an angle equal to the angle of incidence
producing less bright echo.
Trang 25Reflection is of two types (Fig 1.13):
1 Specular reflection: is from a large tissue interface
(smooth boundary between media) and produces
bright echoes These signals are intense and angle
dependent:
2 Acoustic scattering: occurs from smaller objects
Scattering is responsible for tissue texture The signals
are less intense and less angle dependent These
provide tissue signature to the image
Attenuation
As sound waves travel through a medium (e.g tissue or
blood), the intensity weakens or attenuates The degree
of attenuation is expressed in decibels (dB) Absorption
represents a conversion of sound energy to another
form of energy and is the major reason for attenuation
Attenuation is greater for high-frequency sounds, which
result in higher absorption and more scatter Attenuation
coefficient is smallest for the fat and maximum for lungs
Bones also have very high attenuation coefficient
Penetration
Depth to which an ultrasound beam travels into a tissue
is called penetration Besides the tissue characteristic,
penetration is largely dependent upon the frequency of
the ultrasound being applied:
uLtrASound BEAM
The sound beam is the confined, directional beam of ultrasound travelling as a longitudinal wave from the transducer face into the propagation medium Ultrasound beams are made of scan lines These have length (azimuth) along their long axis and width (elevation) along their short axis Beam width should be as narrow as possible to prevent beam width artifacts and which can be achieved
by using lens Ultrasound beams are either steered mechanically or electrically Both rapidly sweep sound waves through tissues (Fig 1.14)
Imaging using ultrasound
Image formation requires an ultrasound machine with appropriate transducers and display, using a cathode-ray tube or a flat panel
Fig 1.13: Reflection from a large interface is called specular
reflection and appears bright while smaller objects (smaller than the
wavelength) produce acoustic scattering.
Fig 1.14: Schematic diagram of an ultrasound beam.
Trang 26A basic echocardiography machine has the following
parts (Fig 1.15):
• Transducer probe: Probe that sends and receives the
sound waves
• Central processing unit (CPU): Computer that does all
of the calculations and contains the electrical power
supplies for itself and the transducer probe
• Transducer pulse controls: Changes the amplitude,
frequency and duration of the pulses emitted from the
transducer probe
• Display: Displays the image from the ultrasound data
processed by the CPU
• Keyboard/cursor: Inputs data and takes measure ments
from the display
• Disk storage device (hard, floppy, CD): Stores the
• Ultrasound transducers, or probes, can be categorized based on their frequency range, low frequency versus high frequency; and the shape of the probe, curved versus linear or sector Linear probes are mostly used for vascular examination while sector probes are used for cardiac examination (Fig 1.17)
• Linear array probes are high-frequency probes frequency probes have less tissue penetration but good near-field image resolution
High-• Curved array probes are low-frequency probes and used mainly for abdominal examination
• Low-frequency probes have greater tissue penetration; however, resolution is compromised (Fig 1.18)
• Ultrasound transducers have a housing that contains inside piezoelectric element
• Main component of a transducer is a piezoelectric element that generates ultrasound waves as it gets deformed by the applied electrical energy As returning ultrasound beam deforms it again, it generates electri-city, which is used to create an image (Fig 1.19)
• Most transducers in use nowadays have synthetic piezoelectrical ceramics, polymers, composites, crystals and so forth, with defined electromechanical coupling and acoustic impedance
Fig 1.15: A schema of an echocardiograph. Fig 1.16: Transducer and its function.
Fig 1.17: Shapes of the transducers.
Trang 27• The new generations of piezoelectric elements have
broad bandwidth for optimal resolution and
penet-ration (Fig 1.20)
• Nominal frequency of a transducer depends upon its
resonance frequency
• Damping and electrical energy modulation generate a
wide range of frequencies (broad bandwidth) around
the nominal frequency
• Frequency of a transducer depends upon electrical
pulse frequency
• The piezoelectric element has a very high acoustic
impedance and works like a reflective surface but
material surrounding it works as a dump
• The multielement transducer generates a main ultrasound beam that produces diagnostic images and accessory beams (side lobes; Fig 1.21)
Side-lobe beams can produce side-lobe artifacts, which need to be recognized (Fig 1.22) Side-lobes may
be minimized by driving the elements at variable voltages
in a process called apodization
• Most transducers have a focal point Up to the focal point, there is near field and beyond that there is far field Far field has increased distance between scan lines and hence lower resolution
• Currently, most transducers are fully sampled matrix array transducers with thousands of elements
Fig 1.18: Relationship between resolution and penetration of various
transducer probes. Fig 1.19:function as a transducer. Piezoelectric element encased in a housing assembly to
Fig 1.20: Broad bandwidth transducer with a wide range of resonant
frequency.
Fig 1.21: Main ultrasound beam flaked by side lobe beams.
Trang 28[compared to 64–256 elements in two-dimensional
(2D) probes] that provide a thick ultrasound beam
(Fig 1.23)
• The probe also has a sound absorbing substance, to
eliminate back reflections from the probe itself; and an
acoustic lens, to help focus the emitted sound waves
Despite these, there can be back reflects from the
transducer producing transducer artifacts (Fig 1.24)
• The shape of the probe determines its field of view, and
the frequency of emitted sound waves determines how
deep the sound waves penetrate and the resolution of
the image
• Transducers of high frequency have thin piezoelectric elements that generate pulses of short wavelength
• Beam-forming (link) is a general processing technique
used to control the directionality of the reception or transmission of a signal on a transducer array
Developments in transducer technology have resulted
in a reduced transducer footprint, improved side-lobe suppression, increased sensitivity and penetration, and the implementation of harmonic capabilities that can be used for both gray-scale and contrast imaging
A conventional 2D phased array transducer is posed of multiple piezoelectric elements that are elect-rically isolated from each other and arranged in a single row Individual ultrasound wave fronts are generated by firing individual elements in a specific sequence with a delay in phase with respect to the transmit initiation time Each element adds and subtracts pulses to generate a single ultrasound wave with a specific direction that constitutes
com-a rcom-adicom-ally propcom-agcom-ating sccom-an line The linecom-ar com-arrcom-ay ccom-an be steered in two dimensions [vertical (axial) and lateral (azimuthal)], while resolution in the Z-axis (elevation) is fixed by the thickness of the image slice, which, in turn, is related to the vertical dimension of piezoelectric elements
Three-dimensional echocardiography (3DE) array transducers are composed of 3,000–9,000 indep-
matrix-endent piezoelectric elements with operating frequencies ranging from 2 to 4 MHz and 5 to 7 MHz for transthoracic echocardiography (TTE) and transesophageal echocar-diographic (TEE), respectively These piezoelectric elements are arranged in a matrix configuration within
Fig 1.22: M-mode of aortic root and the left atrium behind it There
is an image of the posterior LV wall seen within the left atrium due to
side lobe error Note that this image is less echogenic than the original.
Fig 1.23: A fully sampled matrix transducer with pyramidal field of view.
Fig 1.24: Transducer artifact near the apex due to back reflects from
the probe.
Trang 29the transducer to steer the ultrasound beam electronically
The electronically controlled phasic firing of the elements
in that matrix generates a scan line that propagates
radially and can be steered both laterally (azimuth) and
in the elevation to acquire a volumetric pyramid of data
(Fig 1.25)
Multiple-time reception and Parallel
Processing of reflected Beam
Basic imaging by ultrasound uses the amplitude
information in the reflected signal One pulse is emitted,
the reflected signal, however, is sampled more or less
continuously (99% time is for receiving) As the velocity
of sound in tissue is fairly constant, the time between
the emission of a pulse and the reception of a reflected signal is dependent on the distance; that is, the depth
of the reflecting structure The reflected pulses are thus sampled at multiple time intervals (multiple range gating), corresponding to multiple depths, and displayed
in the image as depth Parallel processing for multiple simultaneous reception of reflected waves improves the temporal resolution (Fig 1.26) This parallel processing could be in ratio of 1:16 or more
Spatial resolution
It is the parameter of an ultrasound imaging system that characterizes its ability to detect closely spaced interfaces and displays the echoes from those interfaces as distinct and separate objects If resolution is better, clarity of the image is improved Resolution is of two types:
Axial resolution: Is the minimum required reflector
separation along the direction of propagation required
to produce separate reflections Good axial resolution is achieved with short spatial pulse lengths Short spatial
pulse lengths are a result of higher frequency and higher
damped transducers Therefore, the higher the frequency,
the better is the resolution (Fig 1.27) Axial resolution is also called longitudinal or azimuthal resolution
Lateral resolution: Is the minimum reflector separation
perpendicular to the direction of propagation required
to produce separate reflections Good lateral resolution
is achieved with narrow acoustic beams Wider beams typically diverge further in the far field and any ultrasound beam diverges at greater depth, decreasing lateral
Fig 1.25: The technique of obtaining pyramidal data set with use of a
matrix array transducer. Fig 1.26:channels for a single transmitted beam. Parallel processing of multiple reflected beams in several
Fig 1.27: Axial versus lateral resolution.
Trang 30resolution Therefore, lateral resolution is best at shallow
depths and worse with deeper imaging
Temporal resolution: is the ability to detect that an
object has moved over time For the purposes of medical
ultrasound, temporal resolution is synonymous with
frame rate Typical frame rates in echo imaging systems
are 30–100 Hz The temporal resolution or frame rate = 1/
(time to scan 1 frame) The time to scan one frame is equal
to the pulse repetition period number of scan lines per
frame
Common means of improving frame rate include:
• Narrowing the imaging sector, which decreases the
time it takes to scan one frame
• Decreasing the depth, which decreases the pulse
repetitive frequency
• Decreasing the line density, which requires fewer lines
to scan one frame
• Turning off multifocus, which decreases the number of
pulses needed per line
Focusing
Within transducers, there is a FOCUS that concentrates
the sound beam into a smaller beam area than would
exist otherwise This area of focus is where one obtains the
best images The focus is on the monitor, on the vertical
millimeter scale So when positioning anatomy, make
sure it is in the region of the focus, so best images may be
obtained
It is defined as the peak rarefactional pressure (negative pressure) divided by the square root of the ultrasound frequency The used range varies from 0.05 to 1.9
technique of Image Acquisition
• At the start of an echocardiography examination, the appropriate transducer is selected according to the type of examination and patient’s body habitus
• A higher frequency transducer provides better resolution, but it has a shallower depth of penetration For the pediatric population, the transducer frequency
is usually 5–7.5 MHz, but for adults the transducer frequency at the start of an examination is usually 2–2.5 MHz
• In fundamental imaging, echocardiographic images are created when the transducer receives reflected beams of the same frequency as the transmitted beam, but the interface between tissue and blood can
be delineated better with the reception of harmonic frequencies
• When the only reflected frequency received to create the ultrasound image is equal to a multiple of the transmitted frequency, the technique is called harmonic imaging Myocardial tissues are able to generate harmonic frequencies, and harmonic imaging improves the delineation of the endocardial border (tissue harmonic imaging) As a result, harmonic imaging is usually the imaging modality of choice (Fig 1.28)
• A limitation of harmonic imaging in routine 2D echocardiography is the increased sparkling quality to the ultrasound image and the increased thickness of the endocardial border (Fig 1.29)
• The following should be shown on the screen: The
patient’s identification, blood pressure at the time of the examination, and an electrocardiographic tracing
• The examination of an adult patient usually begins with a depth of 20–25 cm and a wide sector (90°) This also gives an idea about any unusual extracardiac structures After the initial view, adjust the field depth
to use the entire screen to demonstrate the intended cardiovascular images
• A zoom or regional expansion selection (RES) function should be used frequently to visualize a region of interest The zoomed image is also better for
Fig 1.28: Improved signal-to-noise ratio with harmonic imaging
compared to fundamental imaging.
Trang 31making measurements, with less intraobserver and
interobserver variability (Fig 1.30)
• When quantitative measurements are made, review
the acquired image in a cine loop format to identify a
frame at a specific timing of a cardiac cycle Examples
are a midsystolic frame to measure the diameter of the
left ventricular outflow tract, an end-systolic frame to
measure the size of the left atrium and an end-diastolic
frame to measure the wall thickness of the left ventricle
Improving temporal resolution
Specific areas need to be imaged and it may be necessary
to decrease the sector size, which will improve temporal
resolution by increasing the frame rate The gain of the image is controlled by overall gain and regional gain [by time gain compensation (TGC)]
tissue texture
Hyperechoic areas have a great amount of energy from
returning echoes and are seen as white
Hypoechoic areas have less energy from returning
echoes and are seen as gray
Anechoic areas without returning echoes are seen as
black
time Gain compensation
The received ultrasound signal can be amplified by increasing the gain Decreased gain yields a black image and details are masked, while increased gain yields a whiter image
TGC will change the gain factor so that equally reflective structures will be displayed with the same brightness regardless of their depth (Fig 1.31)
TGC allows amplification of ultrasound beams from deeper depths because different amplitudes of ultrasound signals are produced when received from different depths More TGC is required for higher frequency transducers, which create more attenuation
Returning ultrasound waves are referred to as signal, while background artifact is referred to as noise Increasing the gain increases the signal-to-noise ratio
Fig 1.29: Note thickness of the chordae tendineae in harmonic
imag-ing at 1.3 MHz compared to fundamental imagimag-ing at 1.9 MHz All other
parameters are unchanged.
Fig 1.30: Zoom view of the mitral valve to measure the size of annular caseoma.
Fig 1.31: Time gain compensation (TGC) is mounted on the knob
panel of the echocardiograph.
Trang 32Reduces the differences between the smallest and largest
amplitudes of ultrasound images by reducing the total
range without altering the signal ratio
Side Lobe Artifacts
The probe cannot produce a pulse that travels purely in
one direction Pulses also travel off at specific angles These
side lobes are relatively weak and so normally do little
to degrade the image Their effect is only normally seen
faintly superimposed in fluid filled areas that are anechoic
and so do not obscure the weak side lobe reflections The
exception is when a side lobe strikes a particularly strong
reflector at 90° In this case, the reflector can appear within
the image
Partial volume: The slice that makes up the ultrasound
image is 3 dimensional This typically means that fluid
filled areas, where they are very small or adjacent to
soft tissue will not appear anechoic (black) as would be
expected, but often contain low level echoes which can be
mistaken for debris or even soft tissue
Acoustic Shadowing
Tissues with high attenuation coefficients like bone or
prosthesis do not allow passage of ultrasound waves
Therefore any structure lying behind tissue with a high
attenuation coefficient cannot be imaged and will be seen
as an anechoic region (Fig 1.32)
Echocardiographic display
Reflected ultrasound waves (echoes) are detected, filtered and amplified electronically by the system and displayed
on the cathode ray tube in the following manner.4–6
A-mode and M-mode image: Sending and receiving an
ultrasound wave along a single line generates an A-mode (amplitude-mode) or an M-mode image In amplitude mode (A-mode), reflected sound signal is displayed as a vertical spike (voltage amplitude)
In brightness mode (B mode), the amplitude spike is replaced by a dot The dot has a location, certain brightness and also motion Dots of a structure or interface have similar brightness and motion, which can be displayed in two axes: vertical and horizontal If there are dots along a single scan line, it is called M-mode echocardiography
If dots are along multiple scan lines that are steered in a predetermined fashion, the morphology of the structure can be identified when dots are arranged in a plane and this is called cross-sectional or 2D echocardiography “Motion imaging mode (M-mode) is a gray-scale display of amplitude from each depth along a single scan line over time with a high temporal resolution” (Figs 1.33 and 1.34)
2D image is a gray-scale display of amplitude (dots)
from each depth along several scan lines created by steering the ultrasound beam in an imaging plane This provides a real-time tomographic view in the form of a thin slice with spatial resolution at the cost of temporal resolution (Figs 1.35 and 1.36)
Fig 1.32: Acoustic shadow (arrow) due to metallic aortic prosthesis. Fig 1.33: M-mode echocardiogram through the basal part of the LV
Structures are identified by motion patterns.
Trang 33These slices are of less than 1 mm each and can be sagittal, coronal, transverse or oblique.7
3D image is a colorized-display of amplitude over
depth from several imaging planes in a pyramidal volume
as if tomographic slices in various X, Y and Z planes have been stitched together (Figs 1.37 and 1.38)
Real-time 3D images are obtained by matrix array transducers (Fig 1.39)
Fig 1.34: M-mode echocardiogram at the level of aortic valve. Fig 1.35: Two-dimensional echocardiographic long axis view.
Fig 1.36: Two-dimensional echocardiographic view perpendicular to
the long axis as shown in Figure 1.35. Fig 1.37:of length, width and depth. Three-dimensional echocardiographic view with perception
Fig 1.38: Longitudinally reconstructed image from a 3D pyramidal
volume Part of the pyramidal volume is appreciated.
Trang 34• Parasternal (anterior chest wall on either side of
sternum between second and 4th intercostal spaces)
• Apical (over palpable apex)
• Subcostal (below xiphisternum)
• Suprasternal (above manubrium sterni in suprasternal
notch)
• Transesophageal
• Epicardial (during open heart surgery)
• Intracardiac
Axis: is the plane in which the ultrasound beam travels
through the heart:
Long axis: When the ultrasound beam travels
longitudinally through the heart (Fig 1.40)
Short axis: Short axis image is obtained when the
ultrasound beam travels perpendicular to the long axis of
the heart (Fig 1.41)
Views: are window-specific images like apical
four-chamber view or parasternal long axis (PLAX) view or suprasternal view (Fig 1.42)
Sequence of an Echocardiographic Examination
Sequence of an echocardiographic examination is different in adults and children with suspected congenital heart disease In infants and children, the primary window
is subcostal followed by suprasternal with a little use of parasternal views In adults, the following sequence of windows for examination, is usually followed (Fig 1.43):
• Left parasternal window
• Apical window
• Subcostal window
Fig 1.39: Arrangement of piezoelectric elements in a matrix
tansducer shown in a cross-section view.
Fig 1.40: Long axis view.
Fig 1.41: Short axis image (left lower inset) Right panel shows
position transducer. Fig 1.42:sternal notch and oriented anteriorly to obtain long axis view of aorta. Suprasternal view with transducer placed in the
Trang 35supra-• Suprasternal window
• Right parasternal window
To obtain an image, it is common to rotate, tilt or
move the transducer in different directions to get the best
possible views
Parasternal Long Axis View
In this view, ultrasound beam goes through the long axis
of the heart The transducer is placed in the left second to
fourth intercostal space close to the sternum The view and
the structures that it shows are depicted in Figure 1.44
Fig 1.43: Echocardiographic windows and the orientation of the transducer.
Fig 1.44: Parasternal long axis view showing left ventricle (LV); right
ventricle (RV); aortic root (AO); left atrium (LA); thoracic aorta (TA);
mitral valve (MV); aortic valve (AV).
Fig 1.45: Long axis view of the right ventricular outflow tract (RVOT) and the pulmonary artery (PA).
In the left parasternal view, tilting the transducer superiorly and laterally will show the right ventricular outflow tract and the pulmonary artery in the long axis view (Fig 1.45)
Parasternal Short Axis View
From the PLAX view, the transducer is rotated counterclockwise to right angle and tilted up and down
to obtain short axis views at various levels (Figs 1.46
to 1.48)
Trang 36Apical Window
The transducer is placed at the palpable apex or just below
and 4 views can be obtained by rotating and tilting it
1 Apical four-chamber view
2 Apical two-chamber view
3 Apical long axis view
4 Apical five-chamber view
Apical four-chamber view is obtained when the
transducer is placed at the apex and the reference point
toward the right shoulder (Fig 1.49)
When the transducer from the apical four-chamber
position is rotated anticlockwise, apical two-chamber
view can be obtained (Fig 1.50)
Fig 1.46: Parasternal short axis view at mid-LV level. Fig 1.47: Parasternal short axis view at the level of aortic valve.
Fig 1.48: Parasternal short axis view at the level of mitral valve (MV). Fig 1.49: Apical four-chamber view and the structures shown.
Rotating the transducer clockwise from the apical four-chamber position shows apical long axis view (Fig 1.51)
With inferior tilt of the transducer from the apical chamber view, an apical five-chamber view can be shown (Fig 1.52)
Subcostal Window
By placing the transducer in subxiphoid position, subcostal views can be obtained both like that from parasternal and apical windows The transducer needs
to be rotated or tilted Most standard views like the chamber view (Fig 1.53) or the short axis views can also be
Trang 37four-Main structures examined by subcostal views besides the cardiac chambers are:
• Inferior vena cava
• Hepatic veins
• Azygous and anomalous pulmonary veins
• Abdominal aorta and its branchesThe four cardiac chambers, the right ventricular outflow tract, the aorta, and the vena cava can be visualized in the subcostal view Sometimes, it is also possible to visualize
a portion of the abdominal aorta Pointing the transducer toward the right side of the patient would result in a good view of showing the liver and suprahepatic veins as well
as a transverse cross-section of the inferior vena cava (Fig 1.55)
Fig 1.50: Apical two-chamber view with structures shown. Fig 1.51: Apical long axis view.
Fig 1.52: Apical five-chamber view. Fig 1.53: Subcostal four-chamber view (left upper corner) and the
schema (lower left corner).
obtained from a subcostal transducer position The
sub-costal view can be very helpful in patients with bad image
quality or in patients with suspected pericardial effusion
To obtain the subcostal four-chamber view, place the
transducer over the center of the epigastrium and tilt it
downward from the suprasternal notch to the left shoulder
of the patient The image produced will be similar to the
apical four-chamber view The short-axis subcostal view,
however, is similar to the parasternal view and is ideal for
studying the right side of the heart
Subcostal four-chamber view is obtained by pointing the
reference point of the transducer toward the left shoulder
with some inferior tilt to the base of probe By rotating the
transducer, a series of short axis can be obtained (Fig 1.54)
Trang 38Suprasternal View
Suprasternal views (Fig 1.56) are routinely used in
children and adult patients to study:
• Ascending aorta; arch and its branches; descending
thoracic aorta
• Coarctation of aorta
• Patent ducts arteriosus
• Right and left superior vena cavae
• Azygous veins
• Supracardiac variety of anomalous pulmonary venous drainage
• Pulmonary artery bifurcation
transesophageal Echocardiographic Views
The technique of TEE permits visualization of heart without the acoustic interference imposed by the chest wall, ribs and the lungs This is achieved by introduction of
Fig 1.54: Subcostal short axis view showing the entire right heart
structures.
Fig 1.55: Longitudinal view of the inferior vena cava (IVC) from the subcostal view.
Fig 1.56: Suprasternal long axis view showing aorta and bifurcation of
the pulmonary artery Segmental branches of the left pulmonary artery
(LPA) are also seen Right pulmonary artery (RPA).
Fig 1.57: The procedure of TEE in a schematic diagram The probe
is flexible and can be rotated and flexed or retroverted with knobs at the handle.
Trang 39Fig 1.58: Transverse midesophageal four-chamber view (4CV). Fig 1.59: Vertical or Longitudinal midesophageal two-chamber
High esophageal views are for visualizing the great vessels (Fig 1.65) while the transgastric views are used to study the detailed anatomy of the mitral and aortic valves including continuous wave (CW) interrogation of the left ventricular outflow tract
Details of TEE examinations are shown in respective chapters
interference
Its drawbacks are:
• Semi-invasive and hence needs intensive monitoring
• Novel imaging planes need greater technical skills to
interpret
• Not possible in uncooperative patients or those with
esophageal stricture, cervical arthritis, gastric ulcer
and so forth
tEE Probe Manipulation
Probe movements (entire probe moves):
• Advance or withdraw
• Turn right or left
Trang 40Fig 1.60: Midesophageal long axis view showing aorta, left atrium, left
ventricle and the mitral valve. Fig 1.61:aortic valve, atria and the RVOT. Transverse midesophageal view to show interatrial septum,
Fig 1.62: TEE RV inflow view showing tricuspid valve. Fig 1.63: Midesophageal bicaval view showing inferior and superior
vena cavae and the interatrial septum.
Fig 1.64: Midesophageal transverse short axis view showing left atrial
appendage (LAA). Fig 1.65:beyond the sinotubular junction. High esophageal view showing ascending aortic dissection