The “feedback loop” in the analytical approach is maintained by a quality assurance program Figure 15.1, whose objective is to control systematic and random sources of error.1–5The under
Trang 1705
Quality Assurance
I n Chapter 1 we noted that each field of chemistry brings a unique
perspective to the broader discipline of chemistry For analytical
chemistry this perspective was identified as an approach to solving
problems, which was presented as a five-step process: (1) Identify and
define the problem; (2) Design the experimental procedure; (3) Conduct
an experiment and gather data; (4) Analyze the experimental data; and
(5) Propose a solution to the problem The analytical approach, as
presented thus far, appears to be a straightforward process of moving
from problem-to-solution Unfortunately (or perhaps fortunately for
those who consider themselves to be analytical chemists!), an analysis is
seldom routine Even a well-established procedure, carefully followed,
can yield poor data of little use.
An important feature of the analytical approach, which we have
neglected thus far, is the presence of a “feedback loop” involving steps
2, 3, and 4 As a result, the outcome of one step may lead to a
reevaluation of the other two steps For example, after standardizing a
spectrophotometric method for the analysis of iron we may find that its
sensitivity does not meet the original design criteria Considering this
information we might choose to select a different method, to change
the original design criteria, or to improve the sensitivity.
The “feedback loop” in the analytical approach is maintained by a
quality assurance program (Figure 15.1), whose objective is to control
systematic and random sources of error.1–5The underlying assumption
of a quality assurance program is that results obtained when an
analytical system is in statistical control are free of bias and are
characterized by well-defined confidence intervals When used
properly, a quality assurance program identifies the practices necessary
to bring a system into statistical control, allows us to determine if the
system remains in statistical control, and suggests a course of corrective
action when the system has fallen out of statistical control.
The focus of this chapter is on the two principal components of a
quality assurance program: quality control and quality assessment In
addition, considerable attention is given to the use of control charts for
routinely monitoring the quality of analytical data.
Trang 2706 Modern Analytical Chemistry
Quality control encompasses all activities used to bring a system into statistical
control The most important facet of quality control is a set of written directives scribing all relevant laboratory-specific, technique-specific, sample-specific, method-specific, and protocol-specific operations.1,3,6 Good laboratory practices
de-(GLPs) describe the general laboratory operations that need to be followed in any analysis These practices include properly recording data and maintaining records, using chain-of-custody forms for samples that are submitted for analysis, specifying and purifying chemical reagents, preparing commonly used reagents, cleaning and calibrating glassware, training laboratory personnel, and maintaining the laboratory facilities and general laboratory equipment.
Good measurement practices (GMPs) describe operations specific to a
tech-nique In general, GMPs provide instructions for maintaining, calibrating, and using the equipment and instrumentation that form the basis for a specific tech- nique For example, a GMP for a titration describes how to calibrate a buret (if nec-
1 Identify the problem
Determine type of information needed
(qualitative, quantitative,characterization, or fundamental)Establish context of the problem
2 Design the experimental procedure
Establish design criteria (accuracy, precision,
scale of operation, sensitivity, selectivity,
cost, speed)
Identify interferents
Select method
Establish validation criteria
Establish sampling strategy
Qualityassuranceprogram
3 Conduct an experimentCalibrate instruments and equipmentStandardize reagents
Gather data
4 Analyze the experimental dataReduce or transform dataAnalyze statisticsVerify resultsInterpret results
5 Propose a solutionConduct external evaluation
Q
ua
l
it
i
ty
Figure 15.1
Schematic diagram of the analytical approach to problem solving, showing the role of the quality assurance program
quality control
Those steps taken to ensure that an
analysis is under statistical control
good laboratory practices
Those general laboratory procedures
that, when followed, help ensure the
quality of analytical work
good measurement practices
Those instructions outlining how to
properly use equipment and
instrumentation to ensure the quality of
measurements
quality assurance
The steps taken during an analysis to
ensure that the analysis is under control
and that it is properly monitored
Trang 3Chapter 15 Quality Assurance 707
standard operations procedure
The procedure followed in collecting andanalyzing samples and in interpreting theresults of an analysis
essary), how to fill a buret with the titrant, the correct way to read the volume of
titrant in the buret, and the correct way to dispense the titrant.
The operations that need to be performed when analyzing a specific analyte in a
specific matrix are defined by a standard operations procedure (SOP) The SOP
describes all steps taken during the analysis, including: how the sample is processed
in the laboratory, the analyte’s separation from potential interferents, how the
method is standardized, how the analytical signal is measured, how the data are
transformed into the desired result, and the quality assessment tools that will be
used to maintain quality control If the laboratory is responsible for sampling, then
the SOP will also state how the sample is to be collected and preserved and the
na-ture of any prelaboratory processing A SOP may be developed and used by a single
laboratory, or it may be a standard procedure approved by an organization such as
the American Society for Testing and Materials or the Federal Food and Drug
Ad-ministration A typical SOP is provided in the following example.
EXAMPLE 15.1
Provide an SOP for the determination of cadmium in lake sediments by atomic
absorption spectrophotometry using a normal calibration curve.
SOLUTION
Sediment samples should be collected using a bottom grab sampler and stored
at 4 °C in acid-washed polyethylene bottles during transportation to the
laboratory Samples should be dried to constant weight at 105 °C and ground
to a uniform particle size The cadmium in a 1-g sample of the sediment is
extracted by adding the sediment and 25 mL of 0.5 M HCl to an acid-washed
100-mL polyethylene bottle and shaking for 24 h After filtering, the sample is
analyzed by atomic absorption spectrophotometry using an air–acetylene
flame, a wavelength of 228.8 nm, and a slit width of 0.5 nm A normal
calibration curve is prepared using five standards with nominal concentrations
of 0.20, 0.50, 1.00, 2.00, and 3.00 ppm The accuracy of the calibration curve is
checked periodically by analyzing the 1.00-ppm standard An accuracy of ± 10%
is considered acceptable.
Although an SOP provides a written procedure, it is not necessary to follow the
procedure exactly as long as any modifications are identified On the other hand, a
protocol for a specific purpose (PSP), which is the most detailed of the written
quality control directives, must be followed exactly if the results of the analysis are
to be accepted In many cases the required elements of a PSP are established by the
agency sponsoring the analysis For example, labs working under contract with the
Environmental Protection Agency must develop a PSP that addresses such items as
sampling and sample custody, frequency of calibration, schedules for the preventive
maintenance of equipment and instrumentation, and management of the quality
assurance program.
Two additional aspects of a quality control program deserve mention The first
is the physical inspection of samples, measurements and results by the individuals
responsible for collecting and analyzing the samples.1For example, sediment
sam-ples might be screened during collection, and samsam-ples containing “foreign objects,”
such as pieces of metal, be discarded without being analyzed Samples that are
dis-carded can then be replaced with additional samples When a sudden change in the
protocol for a specific purpose
A precisely written protocol for ananalysis that must be followed exactly
Trang 4708 Modern Analytical Chemistry
performance of an instrument is observed, the analyst may choose to repeat those measurements that might be adversely influenced The analyst may also decide to reject a result and reanalyze the sample when the result is clearly unreasonable By identifying samples, measurements, and results that may be subject to gross errors, inspection helps control the quality of an analysis.
A final component of a quality control program is the certification of an lyst’s competence to perform the analysis for which he or she is responsible.7Before
ana-an ana-analyst is allowed to perform a new ana-analytical method, he or she may be required
to successfully analyze an independent check sample with acceptable accuracy and precision The check sample should be similar in composition to samples that the analyst will routinely encounter, with a concentration that is 5 to 50 times that of the method’s detection limit.
The written directives of a quality control program are a necessary, but not a cient, condition for obtaining and maintaining an analysis in a state of statistical control Although quality control directives explain how an analysis should be properly conducted, they do not indicate whether the system is under statistical
suffi-control This is the role of quality assessment, which is the second component of a
quality assurance program.
The goals of quality assessment are to determine when a system has reached a state
of statistical control; to detect when the system has moved out of statistical control; and,
if possible, to suggest why a loss of statistical control has occurred so that corrective tions can be taken For convenience, the methods of quality assessment are divided into two categories: internal methods that are coordinated within the laboratory and exter- nal methods for which an outside agency or individual is responsible The incorpora- tion of these methods into a quality assurance program is covered in Section 15C.
The most useful methods for quality assessment are those that are coordinated by the laboratory and that provide the analyst with immediate feedback about the sys- tem’s state of statistical control Internal methods of quality assessment included in this section are the analysis of duplicate samples, the analysis of blanks, the analysis
of standard samples, and spike recoveries.
Analysis of Duplicate Samples An effective method for determining the precision
of an analysis is to analyze duplicate samples In most cases the duplicate samples
are taken from a single gross sample (also called a split sample), although in some cases the duplicates must be independently collected gross samples The results
from the duplicate samples, X1 and X2, are evaluated by determining the difference,
The steps taken to evaluate whether an
analysis is under statistical control
duplicate samples
Two samples taken from a single gross
sample and used to evaluate an analytical
method’s precision
Trang 5Chapter 15 Quality Assurance 709
Table 15.1 Selected Quality Assessment Limits for the Analysis of Waters and Wastewaters
Limits for Spike Recovery (d )r When [Analyte] < 20×MDL (d )r When [Analyte] > 20×MDL
Abbreviation: MDL = method’s detection limit.
where diis the difference between the ith pair of duplicates The degrees of freedom
for the standard deviation is the same as the number of duplicate samples If
dupli-cate samples from several sources are combined, then the precision of the
measure-ment process must be approximately the same for each The precision obtained is
then compared with the precision needed to accept the results of the analysis.
EXAMPLE 15.2
To evaluate the precision for the determination of potassium in blood serum,
duplicate analyses were performed on six samples, yielding the following
Trang 6The Analysis of Blanks The use of a blank was introduced in Chapter 3 as a means of correcting the measured signal for contributions from sources other than the analyte The most common blank is a method, or reagent blank, in which an analyte-free sample, usually distilled water, is carried through the analysis using the same reagents, glassware, and instrumentation Method blanks are used to identify and correct systematic errors due to impurities in the reagents and con- tamination in the glassware and instrumentation At a minimum, method blanks should be analyzed whenever new reagents are used, although a more frequent analysis provides an ongoing monitoring of the purity of the reagents A new method blank should also be run whenever a sample with a high concentration of the analyte is analyzed, because any residual carryover of the analyte may contami- nate the glassware or instrumentation.
When samples are collected in the field, the method blank may be augmented with field and trip blanks.8A field blank is an analyte-free sample carried from the
laboratory to the sampling site At the sampling site the blank is transferred to a clean sample container, exposing it to the local environment, preserved, and trans- ported back to the laboratory for analysis Field blanks are used to identify and
correct systematic errors due to sampling, transport, and analysis Trip blanks are
analyte-free samples carried from the laboratory to the sampling site and returned
to the laboratory without being opened A trip blank is used to identify and correct systematic errors due to cross-contamination of volatile organic compounds during transport, handling, storage, and analysis.
Analysis of Standards The analysis of a standard containing a known tion of analyte also can be used to monitor a system’s state of statistical control Ide- ally, a standard reference material (SRM) should be used, provided that the matrix
concentra-of the SRM is similar to that concentra-of the samples being analyzed A variety concentra-of appropriate SRMs are available from the National Institute of Standards and Technology (NIST) If a suitable SRM is not available, then an independently prepared synthetic sample can be used if it is prepared from reagents of known purity At a minimum,
a standardization of the method is verified by periodically analyzing one of the bration standards In all cases, the analyte’s experimentally determined concentra- tion in the standard must fall within predetermined limits if the system is to be con- sidered under statistical control.
cali-Spike Recoveries One of the most important quality assessment tools is the ery of a known addition, or spike, of analyte to a method blank, field blank, or sam-
recov-ple To determine a spike recovery, the blank or sample is split into two portions,
and a known amount of a standard solution of the analyte is added to one portion.
The concentration of the analyte is determined for both the spiked, F, and unspiked portions, I, and the percent recovery, %R, is calculated as
where A is the concentration of the analyte added to the spiked portion.
A blank prepared in the laboratory that
accompanies a set of sample containers
in the field and laboratory
spike recovery
An analysis of a sample after spiking with
a known amount of analyte
Trang 7Chapter 15 Quality Assurance 711
EXAMPLE 15.3
A spike recovery for the analysis of chloride in well water was performed by
adding 5.00 mL of a 25,000-ppm solution of Cl–to a 500-mL volumetric flask
and diluting to volume with the sample Analysis of the sample and the spiked
sample resulted in chloride concentrations of 183 ppm and 409 ppm,
respectively Determine the percent recovery of the spike.
SOLUTION
The concentration of the added spike is calculated by taking into account the
effect of dilution.
Thus, the spike recovery is
Spike recoveries on method blanks and field blanks are used to evaluate the
general performance of an analytical procedure The concentration of analyte added
to the blank should be between 5 and 50 times the method’s detection limit
Sys-tematic errors occurring during sampling and transport will result in an
unaccept-able recovery for the field blank, but not for the method blank Systematic errors
occurring in the laboratory, however, will affect the recoveries for both the field and
method blanks.
Spike recoveries for samples are used to detect systematic errors due to the
sample matrix or the stability of the sample after its collection Ideally, samples
should be spiked in the field at a concentration between 1 and 10 times the expected
concentration of the analyte or 5 to 50 times the method’s detection limit,
whichever is larger If the recovery for a field spike is unacceptable, then a sample is
spiked in the laboratory and analyzed immediately If the recovery for the
labora-tory spike is acceptable, then the poor recovery for the field spike may be due to the
sample’s deterioration during storage When the recovery for the laboratory spike
also is unacceptable, the most probable cause is a matrix-dependent relationship
be-tween the analytical signal and the concentration of the analyte In this case the
samples should be analyzed by the method of standard additions Typical limits for
acceptable spike recoveries for the analysis of waters and wastewaters are shown in
Table 15.1.7
Internal methods of quality assessment should always be viewed with some level of
skepticism because of the potential for bias in their execution and interpretation.
For this reason, external methods of quality assessment also play an important role
in quality assurance programs One external method of quality assessment is the
certification of a laboratory by a sponsoring agency Certification is based on the
successful analysis of a set of proficiency standards prepared by the sponsoring
agency For example, laboratories involved in environmental analyses may be
re-quired to analyze standard samples prepared by the Environmental Protection
Trang 8712 Modern Analytical Chemistry
Agency A second example of an external method of quality assessment is the tary participation of the laboratory in a collaborative test (Chapter 14) sponsored by
volun-a professionvolun-al orgvolun-anizvolun-ation such volun-as the Associvolun-ation of Officivolun-al Anvolun-alyticvolun-al Chemists Finally, individuals contracting with a laboratory can perform their own external quality assessment by submitting blind duplicate samples and blind standard sam- ples to the laboratory for analysis If the results for the quality assessment samples are unacceptable, then there is good reason to consider the results suspect for other samples provided by the laboratory.
In the previous section we described several internal methods of quality assessment that provide quantitative estimates of the systematic and random errors present in an analytical system Now we turn our attention to how this numerical information is incorporated into the written directives of a complete quality assurance program Two approaches to developing quality assurance programs have been described9: a prescriptive approach, in which an exact method of quality assessment is prescribed; and a performance-based approach, in which any form of quality assessment is ac- ceptable, provided that an acceptable level of statistical control can be demonstrated.
With a prescriptive approach to quality assessment, duplicate samples, blanks, dards, and spike recoveries are measured following a specific protocol The result for each analysis is then compared with a single predetermined limit If this limit is exceeded, an appropriate corrective action is taken Prescriptive approaches to qual- ity assurance are common for programs and laboratories subject to federal regula- tion For example, the Food and Drug Administration (FDA) specifies quality as- surance practices that must be followed by laboratories analyzing products regulated by the FDA.
stan-A good example of a prescriptive approach to quality assessment is the protocol outlined in Figure 15.2, published by the Environmental Protection Agency (EPA) for laboratories involved in monitoring studies of water and wastewater.10Indepen- dent samples A and B are collected simultaneously at the sample site Sample A is split into two equal-volume samples, and labeled A1 and A2 Sample B is also split into two equal-volume samples, one of which, BSF, is spiked with a known amount
of analyte A field blank, DF, also is spiked with the same amount of analyte All five samples (A1, A2, B, BSF, and DF) are preserved if necessary and transported to the laboratory for analysis.
The first sample to be analyzed is the field blank If its spike recovery is ceptable, indicating that a systematic error is present, then a laboratory method blank, DL, is prepared and analyzed If the spike recovery for the method blank is also unsatisfactory, then the systematic error originated in the laboratory An ac- ceptable spike recovery for the method blank, however, indicates that the systematic error occurred in the field or during transport to the laboratory Systematic errors
unac-in the laboratory can be corrected, and the analysis contunac-inued Any systematic rors occurring in the field, however, cast uncertainty on the quality of the samples, making it necessary to collect new samples.
er-If the field blank is satisfactory, then sample B is analyzed er-If the result for B is above the method’s detection limit, or if it is within the range of 0.1 to 10 times the amount of analyte spiked into BSF, then a spike recovery for BSF is determined An
Trang 9unacceptable spike recovery for BSFindicates the presence of a systematic error
in-volving the sample To determine the source of the systematic error, a laboratory
spike, BSL, is prepared using sample B and analyzed If the spike recovery for BSLis
acceptable, then the systematic error requires a long time to have a noticeable effect
on the spike recovery One possible explanation is that the analyte has not been
properly preserved or has been held beyond the acceptable holding time An
unac-ceptable spike recovery for BSLsuggests an immediate systematic error, such as that
due to the influence of the sample’s matrix In either case, the systematic errors are
fatal and must be corrected before the sample is reanalyzed.
If the spike recovery for BSFis acceptable, or if the result for sample B is below
the method’s detection limit or outside the range of 0.1 to 10 times the amount of
analyte spiked in BSF, then the duplicate samples A1and A2are analyzed The results
for A1and A2are discarded if the difference between their values is excessive If the
difference between the results for A1and A2is within the accepted limits, then the
results for samples A1and B are compared Since samples collected from the same
sampling site at the same time should be identical in composition, the results are
discarded if the difference between their values is unsatisfactory, and accepted if the
difference is satisfactory.
DF recoverywithin limits
DL recoverywithin limits
B > MDL, or
B > 0.1 × [spike], and
B < 10 × [spike]
Systematicerror inlaboratory
Systematicerror infield
YesNo
YesNo
Immediatesystematicerrors
dependentsystematicerrors
Time-YesNo
No
A1 – B
within limits
PoorReplication
NoYes
A1 – A2
within limits
BSF recoverywithin limits
BSL recoverywithin limits
Poorduplicatesamples
Figure 15.2
Example of a prescriptive approach to quality assurance Adapted from Environmental Monitoring and SupportLaboratory, U.S Environmental Protection Agency, “Handbook for Analytical Quality Control in Water andWastewater Laboratories,” March 1979
Trang 10This protocol requires four to five evaluations of quality assessment data before the result for a single sample can be accepted; a process that must be repeated for each analyte and for each sample Other prescriptive protocols are equally demand- ing For example, Figure 3.7 in Chapter 3 shows a portion of the quality assurance protocol used for the graphite furnace atomic absorption analysis of trace metals in aqueous solutions This protocol involves the analysis of an initial calibration verifi- cation standard and an initial calibration blank, followed by the analysis of samples
in groups of ten Each group of samples is preceded and followed by continuing ibration verification (CCV) and continuing calibration blank (CCB) quality assess- ment samples Results for each group of ten samples can be accepted only if both sets of CCV and CCB quality assessment samples are acceptable.
cal-The advantage to a prescriptive approach to quality assurance is that a single sistent set of guidelines is used by all laboratories to control the quality of analytical results A significant disadvantage, however, is that the ability of a laboratory to pro- duce quality results is not taken into account when determining the frequency of col- lecting and analyzing quality assessment data Laboratories with a record of producing high-quality results are forced to spend more time and money on quality assessment than is perhaps necessary At the same time, the frequency of quality assessment may
con-be insufficient for laboratories with a history of producing results of poor quality.
In a performance-based approach to quality assurance, a laboratory is free to use its experience to determine the best way to gather and monitor quality assessment data The quality assessment methods remain the same (duplicate samples, blanks, standards, and spike recoveries) since they provide the necessary information about precision and bias What the laboratory can control, however, is the fre- quency with which quality assessment samples are analyzed, and the conditions in- dicating when an analytical system is no longer in a state of statistical control Fur- thermore, a performance-based approach to quality assessment allows a laboratory
to determine if an analytical system is in danger of drifting out of statistical trol Corrective measures are then taken before further problems develop.
con-The principal tool for performance-based quality assessment is the control chart In a control chart the results from the analysis of quality assessment samples
are plotted in the order in which they are collected, providing a continuous record
of the statistical state of the analytical system Quality assessment data collected over time can be summarized by a mean value and a standard deviation The fundamen- tal assumption behind the use of a control chart is that quality assessment data will show only random variations around the mean value when the analytical system is
in statistical control When an analytical system moves out of statistical control, the quality assessment data is influenced by additional sources of error, increasing the standard deviation or changing the mean value.
Control charts were originally developed in the 1920s as a quality assurance tool for the control of manufactured products.11Two types of control charts are commonly used in quality assurance: a property control chart in which results for single measurements, or the means for several replicate measurements, are plotted sequentially; and a precision control chart in which ranges or standard deviations are plotted sequentially In either case, the control chart consists of a line represent- ing the mean value for the measured property or the precision, and two or more boundary lines whose positions are determined by the precision of the measure- ment process The position of the data points about the boundary lines determines whether the system is in statistical control.
714 Modern Analytical Chemistry
control chart
A graph showing the time-dependent
change in the results of an analysis that is
used to monitor whether an analysis is in
a state of statistical control
Trang 11Chapter 15 Quality Assurance 715
Construction of Property Control Charts The simplest form for a property control
chart is a sequence of points, each of which represents a single determination of the
property being monitored To construct the control chart, it is first necessary to
de-termine the mean value of the property and the standard deviation for its
measure-ment These statistical values are determined using a minimum of 7 to 15 samples
(although 30 or more samples are desirable), obtained while the system is known to
be under statistical control The center line (CL) of the control chart is determined
by the average of these n points
The positions of the boundary lines are determined by the standard deviation, S, of
the points used to determine the central line
with the upper and lower warning limits (UWL and LWL), and the upper and lower
control limits (UCL and LCL) given by
UWL = CL + 2S LWL = CL – 2S UCL = CL + 3S LCL = CL – 3S
EXAMPLE 15.4
Construct a property control chart for the following spike recovery data (all
values are for percentage of spike recovered).
The mean and the standard deviation for the 20 data points are 99.4 and
1.6, respectively, giving the UCL as 104.2, the UWL as 102.6, the LWL as
96.2 and the LCL as 94.6 The resulting property control chart is shown in
Trang 12Property control charts can also be constructed using points that are the mean value, X –i, for a set of r replicate determinations on a single sample The mean for the ith sample is given by
where Xij is the jth replicate The center line for the control chart, therefore, is
To determine the standard deviation for the warning and control limits, it is
neces-sary to calculate the variance for each sample, si2.
The overall standard deviation, S, is the square root of the average variance for the
samples used to establish the control plot.
Finally, the resulting warning and control limits are
X j
i
ij j r
Sequence
20LCLLWL
CL = X
UWLUCL
1041031021011009998979695
Figure 15.3
Property control chart for Example 15.4
Trang 13Chapter 15 Quality Assurance 717
Table 15.2 Statistical Factors for the Upper
Warning Limit and Upper Control Limit
Constructing a Precision Control Chart The most common measure of precision
used in constructing a precision control chart is the range, R, between the largest
and smallest results for a set of j replicate analyses on a sample.
To construct the control chart, ranges for a minimum of 15–20 samples
(prefer-ably 30 or more samples) are obtained while the system is known to be in statistical
control The line for the average range, R – , is determined by the mean of these
n samples
The upper control line and the upper warning line are given by
where fUCLand fUWL(Table 15.2) are statistical factors determined by the number
of replicates used to determine the range Because the range always is greater than
or equal to zero, there is no lower control limit or lower warning limit.
EXAMPLE 15.5
Construct a precision control chart using the following 20 ranges, each
determined from a duplicate analysis of a 10-ppm calibration standard
Trang 14The average range for the 20 duplicate samples is 0.177 Because two replicates
were used for each point, the UWL and UCL are
UWL = (2.512)(0.177) = 0.44 UCL = (3.267)(0.177) = 0.58
The complete control chart is shown in Figure 15.4.
The precision control chart is strictly valid only for the replicate analysis of identical samples, such as a calibration standard or a standard reference material Its use for the analysis of nonidentical samples, such as a series of clinical or environ- mental samples, is complicated by the fact that the range usually is not independent
of the magnitude of Xlargeand Xsmall For example, Table 15.3 shows the relationship between R and the concentration of chromium in water. – 10Clearly the significant difference in the average range for these concentrations of Cr makes a single preci- sion control chart impossible One solution to this problem is to prepare separate precision control charts, each of which covers a range of concentrations for which R –
is approximately constant (Figure 15.5).
Interpreting Control Charts The purpose of a control chart is to determine if a tem is in statistical control This determination is made by examining the location
sys-of individual points in relation to the warning limits and the control limits, and the distribution of the points around the central line If we assume that the data are normally distributed, then the probability of finding a point at any distance from the mean value can be determined from the normal distribution curve The upper and lower control limits for a property control chart, for example, are set to ± 3S, which, if S is a good approximation for σ , includes 99.74% of the data The proba-
bility that a point will fall outside the UCL or LCL, therefore, is only 0.26% The
718 Modern Analytical Chemistry
10.00
0.60
R
Sequence
20R
UWLUCL
Trang 15Figure 15.5
Example of the use of subrange precisioncontrol charts for samples that span arange of analyte concentrations Theprecision control charts are used for (a) low concentrations of analyte;
(b) intermediate concentrations of analyte;and (c) high concentrations of analyte
most likely explanation when a point exceeds a control limit is that a systematic
error has occurred or that the precision of the measurement process has
deterio-rated In either case the system is assumed to be out of statistical control.
Rule 1 A system is considered to be out of statistical control if any single point
exceeds either the UCL or the LCL.
The upper and lower warning limits, which are located at ± 2S, should only be
ex-ceeded by 5% of the data; thus
Rule 2 A system is considered to be out of statistical control if two out of three
consecutive points are between the UWL and UCL or between the LWL and
LCL.
When a system is in statistical control, the data points should be randomly
dis-tributed about the center line The presence of an unlikely pattern in the data is
an-other indication that a system is no longer in statistical control.4,12Thus,
Rule 3 A system is considered to be out of statistical control if a run of seven
consecutive points is completely above or completely below the center line
(Figure 15.6a).
Table 15.3 Average Range for Duplicate Samples for Different
Concentrations of Chromium in Water
Trang 16Figure 15.6
Examples of property control charts that
show a run of data (highlighted in box)
indicating that the system is out of statistical
CL
LWLLCL
CL
LWLLCL
(a)
(b)
(c)
Trang 17Rule 4 A system is considered to be out of statistical control if six consecutive
points are all increasing in value or all decreasing in value (Figure 15.6b) The
points may be on either side of the center line.
Rule 5 A system is considered to be out of statistical control if 14 consecutive
points alternate up and down in value (Figure 15.6c) The points may be on
either side of the center line.
Rule 6 A system is considered to be out of statistical control if any obvious
“nonrandom” pattern is observed.
The same rules apply to precision control charts with the exception that there are
no lower warning and lower control limits.
Using Control Charts for Quality Assurance Control charts play an important role
in a performance-based program of quality assurance because they provide an easily
interpreted picture of the statistical state of an analytical system Quality assessment
samples such as blanks, standards, and spike recoveries can be monitored with
prop-erty control charts A precision control chart can be used to monitor duplicate samples.
The first step in using a control chart for quality assurance is to determine the
mean value and the standard deviation (except when using the range) for the quality
assessment data while the system is under statistical control These values must be
es-tablished under the same conditions that will be present during the normal use of the
control chart Thus, preliminary data should be randomly collected throughout the
day, as well as over several days, to account for short-term and long-term variability.
The preliminary data are used to construct an initial control chart, and discrepant
points are determined using the rules discussed in the previous section Questionable
points are dropped, and the control chart is replotted As the control chart is used, it
may become apparent that the original limits need adjusting Control limits can be
re-calculated if the number of new data points is at least equivalent to the amount of data
used to construct the original control chart For example, if 15 points were initially
used, the limits can be reevaluated after 15 additional points are collected The 30
points are pooled together to calculate the new limits A second modification can be
made after a further 30 points have been collected Another indication that a control
chart needs to be modified is when points rarely exceed the warning limits In this case
the new limits can be recalculated using the last 20 points.
Once a control chart is in use, new quality assessment data should be added
at a rate sufficient to ensure that the system remains in statistical control As
with prescriptive approaches to quality assurance, when a quality assessment
sample is found to be out of statistical control, all samples analyzed since the last
successful verification of statistical control must be reanalyzed The advantage of
a performance-based approach to quality assurance is that a laboratory may use
its experience, guided by control charts, to determine the frequency for
collect-ing quality assessment samples When the system is stable, quality assessment
samples can be acquired less frequently.
15D KEY TERMS
control chart (p 714)
duplicate samples (p 708)
field blank (p 710)
good laboratory practices (p 706)
good measurement practices (p 706)
Trang 18722 Modern Analytical Chemistry
Few analyses are so straightforward that high-quality results are
easily obtained Good analytical work requires careful planning
and an attention to detail Creating and maintaining a quality
as-surance program is one way to help ensure the quality of analytical
results Quality assurance programs usually include elements of
quality control and quality assessment
Quality control encompasses all activities used to bring a
sys-tem into statistical control The most important facet of quality
control is written documentation, including statements of good
laboratory practices, good measurement practices, standard
oper-ating procedures, and protocols for a specific purpose
Quality assessment includes the statistical tools used to mine whether an analysis is in a state of statistical control and, ifpossible, to suggest why an analysis has drifted out of statisticalcontrol Among the tools included in quality assessment are theanalysis of duplicate samples, the analysis of blanks, the analysis ofstandards, and the analysis of spike recoveries
deter-Another important quality assessment tool, which provides anongoing evaluation of an analysis, is a control chart A control chartplots a property, such as a spike recovery, as a function of time Re-sults exceeding warning and control limits, or unusual patterns ofdata indicate that an analysis is no longer under statistical control
15E SUMMARY
Bell, S C.; Moore, J “Integration of Quality Assurance/
Quality Control into Quantitative Analysis,” J Chem Educ.
1998, 75, 874–877.
The use of several QA/QC methods is described in this
article, including control charts for monitoring the
concentration of solutions of thiosulfate that have been
prepared and stored with and without proper
preservation; the use of method blanks and standard
samples to determine the presence of determinate error
and to establish single-operator characteristics; and the use
of spiked samples and recoveries to identify the presence
of determinate errors associated with collecting and
Marcos, J.; Ríos, A.; Valcárcel, M “Practicing Quality Control
in a Bioanalytical Experiment,” J Chem Educ 1995, 72,
947–949
This experiment demonstrates how control charts and ananalysis of variance can be used to evaluate the quality ofresults in a quantitative analysis for chlorophyll a and b inplant material
The following three experiments introduce aspects of quality assurance and quality control.
1 Make a list of good laboratory practices for the lab
accompanying this course (or another lab if this course does
not have an associated laboratory) Explain the rationale for
each item on your list
2 Write directives outlining good measurement practices for
(a) a buret, (b) a pH meter, and (c) a spectrophotometer
3 A method for the analysis of lead in industrial wastewater has
a method detection limit of 10 ppb The relationship between
the analytical signal and the concentration of lead, asdetermined from a calibration curve is
Smeas= 0.349×(ppm Pb)
Analysis of a sample in duplicate gives Smeasas 0.554 and0.516 Is the precision between these two duplicatesacceptable based on the limits shown in Table 15.1?
4 The following data were obtained for the duplicate analysis of
a 5.00-ppm NO3 standard
15G PROBLEMS
Trang 19Calculate the standard deviation for the analysis of these
duplicate samples If the maximum limit for the relative
standard deviation is 1.5%, are these results acceptable?
5 Gonzalez and colleagues developed a voltammetric procedure
for the determination of tert-butylhydroxyanisole (BHA) in
chewing gum.13Analysis of a commercial chewing gum gave
results of 0.20 mg/g To evaluate the accuracy of their results,
they performed five spike recoveries, adding an amount of
BHA equivalent to 0.135 mg/g to each sample The
experimentally determined concentrations of BHA in these
samples were reported as 0.342, 0.340, 0.340, 0.324, and 0.322
mg/g Determine the % recovery for each sample and the
average % recovery
6 A sample is to be analyzed following the protocol shown in
Figure 15.2, using a method with a detection limit of 0.05
ppm The relationship between the analytical signal and the
concentration of the analyte, as determined from a calibration
curve is
Smeas= 0.273×(ppm analyte)Answer the following questions if the limits for a successful
spike recovery are ±10%
(a) A field blank is spiked with the analyte to a concentration
of 2.00 ppm and returned to the lab Analysis of the spikedfield blank gives a signal of 0.573 Is the spike recovery forthe field blank acceptable?
(b) The analysis of a spiked field blank is found to be
unacceptable To determine the source of the problem,
a spiked method blank is prepared by spiking distilledwater with the analyte to a concentration of 2.00 ppm
Analysis of the spiked method blank gives a signal of0.464 Is the source of the problem in the laboratory
or in the field?
(c) The analysis for a spiked field sample, BSF, was found to be
unacceptable To determine the source of the problem, thesample was spiked in the laboratory by adding sufficientanalyte to increase the concentration by 2.00 ppm
Analysis of the sample before and after the spike gave
signals of 0.456 for B and 1.03 for sample BSL Consideringthese data, what is the most likely source of the systematicerror?
7 The following data were obtained for the repetitive analysis of
8 The following data were obtained for the repetitive spike
recoveries of field samples.15
Sample % Recovery Sample % Recovery Sample % Recovery
Trang 20724 Modern Analytical Chemistry
The following texts and articles may be consulted for an additional
discussion of the various aspects of quality assurance and quality
control
Amore, F “Good Analytical Practices,” Anal Chem 1979, 51,
1105A–1110A
Barnard, Jr A J.; Mitchell, R M.; Wolf, G E “Good Analytical
Practices in Quality Control,” Anal Chem 1978, 50,
1079A–1086A
Cairns, T.; Rogers, W M “Acceptable Analytical Data for Trace
Analysis,” Anal Chem 1993, 55, 54A–57A.
Taylor, J K Quality Assurance of Chemical Measurements Lewis
Publishers: Chelsea, MI, 1987
Additional information about the construction and use of controlcharts may be found in the following sources
Miller, J C.; Miller, J N Statistics for Analytical Chemistry, 3rd ed.
Ellis Horwood Limited: Chichester, England, 1993
Ouchi, G I “Creating Control Charts with a Spreadsheet
1 Taylor, J K Anal Chem 1981, 53, 1588A–1596A
2 Taylor, J K Anal Chem 1983, 55, 600A–608A.
3 Taylor, J K Am Lab October 1985, 67–75.
4 Nadkarni, R A Anal Chem 1991, 63, 675A–682A.
5 Valcárcel, M.; Ríos, A Trends Anal Chem 1994, 13, 17–23.
6 ACS Committee for Environmental Improvement, “Principles of
Environmental Analysis,” Anal Chem 1983, 55, 2210–2218.
7 American Public Health Association, Standard Methods for
the Analysis of Water and Wastewater, 18th ed Washington,
D.C., 1992
8 Keith, L H Environmental Sampling and Analysis: A Practical Guide.
Lewis Publishers, Chelsea, MI, 1991
9 Poppiti, J Environ Sci Technol 1994, 28, 151A–152A.
10 Environmental Monitoring and Support Laboratory, U.S
Environmental Protection Agency, “Handbook for Analytical QualityControl in Water and Wastewater Laboratories,” March 1979
11 Shewhart, W A Economic Control of the Quality of Manufactured
Products Macmillan: London, 1931.
12 Mullins, E Analyst 1994, 119, 369–375.
13 Gonzalez, A.; Ruiz, M A.; Yanez-Sedeno, P.; et al Anal Chim Acta
1994, 285, 63–71.
14 American Public Health Association, Standard Methods for the
Analysis of Water and Wastewater, 18th ed Washington, D.C., 1992.
Data from Table 1030:I on page 1–10
15 American Public Health Association, Standard Methods for the
Analysis of Water and Wastewater, 18th ed Washington, D.C., 1992.
Data adapted from Table 1030:II on page 1–10
15I REFERENCES
Trang 22726 Modern Analytical Chemistry
For example, the proportion of the area under a normal distribution curve that lies to the right of a deviation of 0.04 is 0.4840, or 48.40% The area to
the left of the deviation is given as 1 – P Thus, 51.60% of the area under the normal distribution curve lies to the left of a deviation of 0.04 When the deviation is negative, the values in the table give the proportion of the area under the normal distribution curve that lies to the left of z; therefore,
48.40% of the area lies to the left, and 51.60% of the area lies to the right of a deviation of –0.04.
Appendix 1B
t-Tablea
Value of t for confidence interval of: 90% 95% 98% 99%
Critical value of ötö for αvalues of: 0.10 0.05 0.02 0.01
Trang 24728 Modern Analytical Chemistry
F-Table for Two-Tailed Test at α = 0.05
Trang 25KSbOC4H4O6 324.92 compound prepared by drying KSbOC4H4O6·1/2H2O
at 110 °C and storing in desiccator
Bi2O3 465.96 not considered a primary standard
continued
constant weight at 110 °C Most compounds can be dissolved in dilute acid (1:1 HCl or 1:1 HNO3), with heating if necessary; some of the compounds are water- soluble.
A ll compounds should be of the highest available
purity Metals should be cleaned with dilute acid to
remove surface impurities and rinsed with distilled water.
Unless otherwise indicated, compounds should be dried to
Trang 26730 Modern Analytical Chemistry
MnSO4• H2O 169.01 not considered a primary standard; may be dried at
110 °C without loss of hydrated water
Source: Information compiled from Moody, J R.; Greenberg, R R.; Pratt, K W.; et al Anal Chem 1988, 60, 1203A–1218A; and Smith, B W.; Parsons,
M L J Chem Educ 1973, 50, 679–681.
Trang 28732 Modern Analytical Chemistry
Source: All values are from Martell, A E.; Smith, R M Critical Stability Constants, Vol 4 Plenum Press: New York, 1976 Unless otherwise stated, values
are for 25 °C and zero ionic strength.
Appendix 3B
Acid Dissociation Constants
Trang 29Compound Conjugate Acid pKa Ka
OH
NH3
NH3 OCHCH2CNH2COOH
NH3CHCH2COOHCOOH
COOH
CH2NH3
OH
OH
Trang 30Acid Dissociation Constants—continued
OH
NH3CHCH2SHCOOH
NOHHON
CH3
H3C
CH2COOHHOOCH2C
Trang 31Compound Conjugate Acid pKa Ka
CHCH2COOH
NH3
HN
NH3CHCH2CH(CH3)2COOH
NH3CHCH2CH2CH2CH2NH3COOH
Trang 32Acid Dissociation Constants—continued
OH
NH3CHCH2CH2SCH3COOH
NO2
Trang 33Compound Conjugate Acid pKa Ka
NH3CHCH2C6H5COOH
COOH
COOH
NH2
COOHN
H2
Trang 34Acid Dissociation Constants—continued
COOH
OH
NH3CHCH2OHCOOH
Trang 35Compound Conjugate Acid pKa Ka
Source: All values are from Martell, A E.; Smith, R M Critical Stability Constants, Vol 1–4 Plenum Press: New York, 1976 Unless otherwise stated, values
are for 25 °C and zero ionic strength Values in parentheses are considered less reliable.
NH3
HN
OHCHCH2
COOH
NH3
NH3CHCH(CH3)2COOH
Trang 36Metal–Ligand Formation Constants—continued
Trang 38Metal–Ligand Formation Constants—continued
Source: All values are from Martell, A E.; Smith, R M Critical Stability Constants, Vol 1–4, Plenum Press: New York, 1976 Unless otherwise stated, values
are for 25 °C and zero ionic strength Values in parentheses are considered less reliable.
742 Modern Analytical Chemistry
N
NH+
Trang 39BrO3–+ 6H++ 6e– t Br–+ 3H2O 1.478
Cd2++ 2e– t Cd(s) –0.4030Cd(CN)42–+ 2e– t Cd(s) + 4CN– –0.943Cd(NH3)42++ 2e– t Cd(s) + 4NH3 –0.622
Cl2(g) + 2e– t 2Cl– 1.396ClO–+ H2O + e– t 1⁄2Cl2(g) + 2OH– 0.421 1 M NaOHClO–+ H2O + 2e– t Cl–+ 2OH– 0.890 1 M NaOHHClO2+ 2H++ 2e– t HOCl + H2O 1.64
ClO3–+ 2H++ e– t ClO2(g) + H2O 1.175ClO3–+ 3H++ 2e– t HClO2+ H2O 1.181ClO4–+ 2H++ 2e– t ClO3–+ H2O 1.201
continued
Trang 40MnO4–+ 8H++ 5e– t Mn2++ 4H2O 1.51MnO4–+ 2H2O + 3e– t MnO2(s) + 4OH– 0.60
Hg2++ 2e– t Hg(l) 0.85352Hg2++ 2e– t Hg22+ 0.911
744 Modern Analytical Chemistry