1. Trang chủ
  2. » Khoa Học Tự Nhiên

Introduction to Modern Liquid Chromatography, Third Edition part 93 pps

10 125 0
Tài liệu đã được kiểm tra trùng lặp

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Định dạng
Số trang 10
Dung lượng 143,15 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

17.28b Gradient mixing volume too large relative to gradient volume Use shallower gradient; reduce mixing volume Gradient step-test failure Steps in gradient step-test are of uneven heig

Trang 1

876 TROUBLESHOOTING

Table 17.11

(Continued)

Source

Solution

Gradient distortion

(Fig 17.28b)

Gradient mixing volume too large relative to gradient volume

Use shallower gradient; reduce mixing volume

Gradient step-test

failure

Steps in gradient step-test are of uneven height and/or are distorted

Bubble, check-valve failure, or leak;

Degas mobile phase, clean or replace check valve, fix leak; blocked reservoir frit; bad proportioning valve Gradient

proportioning

valve (GPV) test

failure

GPV test steps>5% Blocked reservoir

frit; restricted solvent supply tubing; bad proportioning valve

Replace frit; clear or replace tubing; replace

proportioning manifold

Dwell-volume

differences

between systems

Changes in retention and/or selectivity between systems

Normal for differences in dwell-volume

Adjust method to compensate for dwell-volume differences Flow-rate check

failure

Flow rate> ±2%

from set value

Bubble; bad check valve or pump seal; leak; wrong compressibility set

Degas mobile phase; clean or replace check valve, replace pump seal; adjust compressibility (or ignore)

Pressure

bleed-down

failure

>15% pressure loss in

10 min

Check valve or pump-seal failure; leak

Clean or replace check valve, replace pump seal; fix leak Retention time

reproducibility

failure

> ±0.05 min for

standard test; more than normal method performance

Bubbles, leaks, check-valve or pump-seal failure; pump or mixing failure

Degas solvents, fix leaks, clean or replace check valve, replace pump seal; run gradient performance test to isolate further Peak area

reproducibility

failure

>1% imprecision in

standard test; more than normal method performance

Autosampler problem

Clean or replace needle; replace seals; see autosampler manual

REFERENCES

1 J W Dolan and L R Snyder, Troubleshooting LC Systems, Humana Press/Springer,

Clifton, NJ, 1989

2 V R Meyer, Pitfalls and Errors of HPLC in Pictures, Wiley-VCH, Weinheim, 2006.

3 J W Dolan, ‘‘LC Troubleshooting,’’ in LCGC, a monthly column (1983-present).

4 http://www.chromforum.org

Trang 2

REFERENCES 877

5 http://www.lcresources.com/resources/TSWiz/

6 S J Williams, J Chromatogr A, 1052 (2004) 1.

7 M D Nelson and J W Dolan, LCGC, 16 (1998) 992.

8 C K Cheung and R Swaminathan, Clin Chem., 33 (1987) 202.

9 C T Mant and R S Hodges, eds., High-Performance Liquid Chromatography of

Peptides and Proteins: Separation, Analysis, and Conformation, CRC Press, Boca Raton,

1991

10 J W Dolan, J R Kern, and T Culley, LCGC, 14 (1996) 202.

11 D W Bristol, J Chromatogr., 188 (1980) 193.

12 P.-L Zhu, L R Snyder, and J W Dolan, J Chromatogr A, 718 (1995) 429.

13 J W Dolan, LCGC, 11 (1993) 640.

14 J W Dolan, LCGC, 9 (1991) 22.

15 M L Ledtje and D Long, Jr., US Patent 5,002,662 (March 26, 1991)

16 J W Dolan, LCGC, 26 (2008) 532.

17 J W Dolan, personal communication

18 L R Snyder and J W Dolan, High-Performance Gradient Elution, Wiley, Hoboken,

NJ, 2007

19 SPD-10Avp UV-Vis Detector Instruction Manual, Shimadzu, Kyoto, 1997.

20 N S Wilson, R Morrison, and J W Dolan, LCGC, 19 (2001) 590.

21 J W Dolan, LCGC, 23 (2005) 370.

22 Reviewer Guidance: Validation of Chromatographic Methods, USFDA-CDER, Nov.

1994, http://www.fda.gov/cder/guidance/index.htm

23 D V McCalley, Anal Chem., 78 (2006) 2532.

24 M T W Hearn, ed., Ion-pair Chromatography Theory and Biological and

Pharma-ceutical Applications, Dekker, New York, 1985.

25 E Rajakyl ¨a, J Chromatogr., 218 (1981) 695.

26 R D Morrison and J W Dolan, LCGC, 23 (2005) 566.

27 J W Dolan, LCGC, 25 (2008) 610.

28 R G Wolcott, J W Dolan, L R Snyder, S R Bakalyar, M A Arnold, and

J A Nichols, J Chromatogr A, 869 (2000) 211.

29 T.-L Ng and S Ng, J Chromatogr., 389 (1985) 13.

30 W R Melander, H.-J Lin, and C Horv ¨ath, J Phys Chem., 88 (1984) 4527.

31 M R Euerby, C M Johnson, I D Rushin, and D A S Sakunthala Tennekoon,

J Chromatogr A, 705 (1995) 229.

32 M R Euerby, C M Johnson, I D Rushin, and D A S Sakunthala Tennekoon,

J Chromatogr A, 705 (1995) 219.

33 J J Gilroy and J W Dolan, LCGC, 22 (2004) 982.

34 T Culley and J W Dolan, LCGC, 13 (1995) 940.

35 J W Dolan, LCGC, 13 (1995) 940.

36 U D Neue, personal communication (1995)

37 T Eidenberger, personal communication (1995)

38 J W Dolan and L R Snyder, Troubleshooting LC Systems, Humana Press, Totowa,

NJ, 1989

39 G Hendriks, J P Franke, and D R A Uges, J Chromatogr A, 1089 (2005) 193.

Trang 4

APPENDIX I

PROPERTIES OF HPLC

SOLVENTS

Solvents are used in HPLC for formulating mobile phases, for dissolving the sample, and for carrying out sample preparation Mobile-phase solvents are of primary concern, because their properties must often fall within narrow limits for acceptable performance However, these same properties also influence the choice

of the sample-injection solvent and solvents used for sample preparation Table I.1 lists several solvent properties that can be important when selecting solvents for an HPLC application Some of these properties have been discussed previously in one

or more sections of this book (second column of Table I.1) The present appendix contains several tables that list values of one or more solvent properties (third column of Table I.1) A brief comment on each solvent property is given in the last column of Table I.1; this serves as an introduction to following sections that deal with individual solvent properties

I.1 SOLVENT-DETECTOR COMPATIBILITY

I.1.1 UV Detection

The mobile phase will preferably have an absorbance A < 0.2 AU at the wavelength

used for detection of the sample; a lower absorbance may mean improved assay precision and better results with gradient elution, but higher absorbances may be acceptable for some isocratic separations Table 1.2 summarizes values of solvent absorbance at different wavelengths (200–260 nm) for solvents that are used for RPC (exclusive of NARP) Very rarely, there may be a reason to use UV detection at

a wavelength<200 nm, for the detection of solutes with low absorptivity at higher

wavelengths

Because water does not absorb at 200 nm or above, the absorbance of aqueous mobile phases that contain these solvents will equal the pure-solvent absorbance times the volume-fraction φ of the B-solvent in the mobile phase For example,

a mobile phase of 25% B would have the following absorbance values A for

different B-solvents at 215 nm: ACN, 0.00 AU; MeOH, 0.09 AU; degassed MeOH, 0.05 AU; THF, 0.22 AU; IPA, 0.07 AU Note that degassing methanol lowers the

Introduction to Modern Liquid Chromatography, Third Edition, by Lloyd R Snyder,

Joseph J Kirkland, and John W Dolan

Copyright © 2010 John Wiley & Sons, Inc.

879

Trang 5

880 PROPERTIES OF HPLC SOLVENTS

Table I.1

Table I 1 Solvent Properties of Interest in HPLC

solvents depend on wavelength required for sample detection Refractive index 4.11 1.3 For RI detection; low values

generally preferred Polarity 2.3.2.1, 6.2.1, 8.2.1 I.4 Determines solvent strength

for 1≤ k ≤ 10

Selectivity 6.3, 8.3.2 I.4 Determines differences in

solvent-type selectivity

injection of large samples

in prep-LC or trace analysis

drop; low values of viscosity desirable

and safety; higher boiling solvents preferred

organic solvent and sample solvent

accurate) formulation of mobile phases by weight

usually critical

concentration of oxygen in the pure solvent, with a resulting decrease in solvent absorbance by about 1/3 over the range 200 to 240 nm When the mobile phase

is degassed, as by helium sparging, the absorbance of other B-solvents will also be lowered, in proportion to the amount of oxygen that is normally present in the solvent Oxygen is more soluble in less polar solvents such as THF and IPA; the absorbance values of Table 1.2 for these solvents may therefore be higher than found

in practice

While water should not absorb light at wavelengths≥200 nm, this assumes that the water has been properly purified Specifications for HPLC-grade water are described in ASTM D1193, but water for use with low-wavelength detection may require total organic carbon (TOC) levels below 50 ppb, as well as high resistivity values Table 1.2 also lists UV absorbance data for some commonly used buffers, and Table 1.3 provides UV cutoff wavelengths for several additional solvents

Trang 6

I.1 SOLVENT-DETECTOR COMPATIBILITY 881

Table I.2

UV Absorbance as a Function of Wavelength of Various Solvents and Buffers Used for RPC

Absorbance at Indicated Wavelength (nm)

Solvents

Acetonitrile 0.06 0.02 0.02 0.01 0.00 0.00 0.00 0.00 0.00 Methanol 1.0+ 1.0 0.53 0.35 0.23 0.10 0.04 0.02 0.01 Methanol (degassed) 1.0+ 0.76 0.35 0.21 0.15 0.06 0.02 0.00 0.00 Tetrahydrofuran 1.0+ 1.0+ 1.0+ 0.85 0.70 0.49 0.30 0.17 0.09 Isopropanol 1.0+ 0.98 0.46 0.29 0.21 0.11 0.05 0.03 0.02

Buffers

Acetate

Acetic acid, 1% 1.0+ 1.0+ 1.0+ 1.0+ 1.0+ 0.87 0.14 0.01 0.00 Ammonium salt 10 nM 1.0+ 0.94 0.53 0.29 0.15 0.02 0.00 0.00 0.00

Carbonate

(NH 4 )HCO 3 , 10 mM 0.41 0.10 0.01 0.00 0.00 0.00 0.00 0.00 0.00

Formate

Sodium salt, 10 mM 1.00 0.73 0.53 0.33 0.20 0.03 0.01 0.01 0.01

Phosphate

H 3 PO 4 , 1% 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00

KH 2 PO 4 , 10 mM 0.03 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00

K 2 HPO 4 , 10 mM 0.53 0.16 0.05 0.01 0.00 0.00 0.00 0.00 0.00 (NH 4 ) 2 HPO 4 , 10 mM 0.37 0.13 0.03 0.00 0.00 0.00 0.00 0.00 0.00 sodium salt, pH-6.8, 10 mM 0.20 0.08 0.02 0.01 0.00 0.00 0.00 0.00 0.00

Trifluoroacetic acid

0.1% in water 1.0+ 0.78 0.54 0.34 0.20 0.06 0.02 0.00 0.00 0.1% in ACN 0.29 0.33 0.37 0.38 0.37 0.25 0.12 0.04 0.01

Source: Data of [1, 2].

It is preferable that the detector response of the mobile phase remains constant during gradient elution This requires that the A-solvent (water) and the B-solvent each respond similarly When UV detection is used at low wavelengths, this may not be the case, especially for B-solvents other than acetonitrile A related problem

is a variation in the absorbance of the buffer or other mobile-phase additives as %B changes Each of these effects is discussed in Section 17.4.5.1

I.1.2 RI Detection

For isocratic separation, the choice of mobile phase is usually not limited for RI detection Detection sensitivity can be increased by selecting a mobile phase whose RI-value is more different than that of sample components (Table 1.3) Detectors based on differential measurement (e.g., RI) cannot be used for gradient elution because of the usual large difference in response for the A- and B-solvents

Trang 7

882 PROPERTIES OF HPLC SOLVENTS

Table I.3

Miscellaneous Solvent Properties

aWavelength at which solvent absorbs 1.0 AU in a 10-mm cell.

bRefractive index.

cBoiling point.

dSolvent strength parameter [4].

I.1.3 MS Detection

The MS interface evaporates the mobile phase, so mobile phases comprising water, organic solvent, and volatile additives are used (i.e., no nonvolatile buffers or salts) Because the mobile phase is removed, UV absorbance is of no concern for LC-MS, but the solvents must be free of particulates

I.2 SOLVENT POLARITY AND SELECTIVITY

Table 1.4 lists solvent properties that affect solvent strength, selectivity, and solu-bility These ‘‘normalized selectivity’’ properties recognize three contributions of the solvent to solute–solvent interaction: solvent hydrogen-bond (H-B) acidity α H

2

H-B basicity β2/Σ, and dipolarity π/Σ The latter parameters are the basis of

Trang 8

I.2SOLVENT POLARITY AND SELECTIVITY 883

Table I.4

Solvent Selectivity Characteristics

H-B Acidity H-B Basicity Dipolarity

α H

aValues from [4].

bPolarity index; values from [5].

cDielectric constant; values from [3].

the solvent-selectivity triangle (Fig 2.9) Solvent polarity P is a measure of overall

solvent polarity Sample solubility tends to correlate with values of P—‘‘like

dissolves like,’’ so samples tend to be more soluble in solvents of similar P Less polar compounds, such as hydrocarbons, will be preferentially dissolved by solvents

with low values of P, and the reverse will be true for solvents with high values of

P The dielectric constant ε similarly correlates with the ability of the solvent to

Trang 9

884 PROPERTIES OF HPLC SOLVENTS

Table I.5

Viscosity of RPC Mobile Phases as a Function of Composition (%B) and Temperature (T ) a

15 1.10 1.43 1.72 1.92 2.00 2.02 1.91 1.69 1.40 1.05 0.63 1.10 1.18 1.23 1.30 1.09 0.98 0.89 0.81 0.70 0.54 0.40

20 1.00 1.32 1.57 1.75 1.83 1.83 1.72 1.52 1.25 0.93 0.60 1.00 1.14 1.10 1.13 0.99 0.90 0.81 0.69 0.56 0.50 0.37

25 0.89 1.18 1.40 1.56 1.62 1.62 1.54 1.36 1.12 0.84 0.56 0.89 1.01 0.98 0.98 0.89 0.82 0.72 0.59 0.52 0.46 0.35

30 0.79 1.04 1.23 1.36 1.43 1.43 1.36 1.21 1.01 0.76 0.51 0.79 0.90 0.87 0.86 0.80 0.74 0.65 0.52 0.45 0.43 0.32

35 0.70 0.92 1.07 1.19 1.24 1.26 1.21 1.09 0.91 0.69 0.46 0.70 0.73 0.78 0.76 0.72 0.68 0.59 0.47 0.43 0.39 0.30

40 0.64 0.82 0.96 1.05 1.11 1.12 1.08 0.98 0.83 0.64 0.42 0.64 0.72 0.70 0.68 0.65 0.62 0.54 0.44 0.41 0.36 0.27

45 0.58 0.75 0.87 0.96 1.00 1.02 0.98 0.89 0.76 0.58 0.39 0.58 0.61 0.64 0.61 0.59 0.58 0.50 0.43 0.38 0.33 0.25

50 0.54 0.71 0.82 0.89 0.93 0.94 0.90 0.82 0.70 0.54 0.37 0.54 0.60 0.60 0.57 0.55 0.53 0.46 0.41 0.36 0.31 0.24

55 0.51 0.67 0.77 0.84 0.88 0.88 0.84 0.76 0.65 0.50 0.36 0.51 0.53 0.56 0.53 0.51 0.49 0.43 0.38 0.34 0.29 0.23

60 0.47 0.61 0.70 0.77 0.81 0.81 0.79 0.72 0.61 0.47 0.33 0.47 0.52 0.53 0.50 0.49 0.46 0.41 0.35 0.37 0.27 0.22

Source: Data from [6, 7].

aThe composition is given as %B (v/v), where B is either methanol (top) or acetonitrile (bottom); for example, the viscosity of 30% methanol water at 30◦C is 1.36 See [8, 9] for viscosity compared to com-position, temperature, pressure, as well as compressibility data.

Table I.6

Density of Solvents for RPC Mobile Phases [3]

Note: Error of 1◦C

Trang 10

I.3 SOLVENT SAFETY 885

dissolve ionized solutes or buffers; high values of ε favor increased solubility for

ionized compounds Table I.5 provides viscosity values for mixtures of MeOH/water and ACN/water as a function of temperature These data are useful in estimating column pressure drop (Eq 2.13) Table I.6 provides densities for some common solvents, to facilitate the more accurate formulation of reversed-phase mobile phases

by weighing each solvent in the mixture

I.3 SOLVENT SAFETY

The solvents commonly used for HPLC are often flammable and moderately toxic Consequently most of these solvents should be stored in a secure, metal cabinet Solvent flammability can be roughly assessed by the flash point, values of which are listed in Table 1.7 Common experience suggests that methanol is only moderately flammable, so that solvents with flash points above 12◦C should not normally present a problem in terms of fire safety However, solvents with lower flash points present a greater danger and should be treated accordingly

Many factors can contribute to solvent toxicity, and solvents other than water should be manipulated in a hood A very rough measure of immediate toxicity is the solvent LD50value (Table 1.7), the administered amount in mg/kg body weight that causes mortality in 50% of the population However, solvents in the laboratory

Table I.7

Flammability and Toxicity Data for Various Solvents

aData from [2].

bMaximum allowable concentration of solvent vapor in the work place air as established by governmental regulation [2].

Ngày đăng: 04/07/2014, 01:20

TỪ KHÓA LIÊN QUAN