Microscopy and Surface Analysis– Optical microscopy – Confocal microscopy – Electron microscopy SEM and TEM, related methods – Scanning probe microscopy STM and AFM, related methods – X-
Trang 1Lecture Date: March 11 th , 2008
Microscopy and Surface Analysis 1
Reading Assignments for Microscopy and
Surface Analysis
Characterization by Spectroscopy and Microscopy”
Microscopies in Chemistry,” J Phys Chem., 1996, 100,
13103-13120
Trang 2Microscopy and Surface Analysis
– Optical microscopy
– Confocal microscopy
– Electron microscopy (SEM and TEM, related methods)
– Scanning probe microscopy (STM and AFM, related methods)
– X-ray fluorescence (from electron microscopy)
– Auger electron spectrometry
– X-ray photoelectron spectrometry (XPS/UPS/ESCA)
– Other techniques:
Secondary-ion mass spectrometry (SIMS)
Ion-scattering spectrometry (ISS)
IR/Raman methods
Why Study Surfaces?
system, and the techniques available are extremely
powerful
reactions
Trang 3X-ray wavelengths/energies, have been combined with
microscopy to create some of the most powerful analytical
tools available…
Imaging Resolution and Magnification
(x)
-Optical Microscopy
Electron Microscopy
Probe Microscopy
Trang 4Optical Microscopy - History
thousands of years Chinese tapestries dating from 1000
B.C depict eyeglasses
the first theoretical study of the lens.
telescope.
possible John and Zaccharias Jannsen (Holland) made
the first commercial and first compound microscopes.
many great discoveries….
Modern Optical Microscopy in Chemistry
developed, the compound
microscope was applied to
the study of chemical
crystals.
(1880): can see boundaries
between materials with
different refractive indices,
while also detecting
isotropic and anisotropic
materials
http://www.microscopyu.com/articles/polarized/polarizedintro.html
Trang 5Optical Microscope Design
characterized NA
(numerical aperature)
– The numerical aperture of a
microscope objective is a
measure of its ability to
gather light and resolve fine
specimen detail at a fixed
object distance
– Large NA = finer detail =
better light gathering http://www.microscopyu.com/articles/polarized/polarizedintro.html
Diagram from Wikipedia (public domain)
not changed much in
300 years
– But the lenses are more
perfect – free of
abberations
The Diffraction Limit
small point of light is not a
point – it is an “Airy” disk
with concentric bright/dark
conventional optical microscope is limited by Fraunhofer diffraction
at the entrance aperture of the objective lens
Airy disk
Resolved Not resolved
Trang 6The Diffraction Limit
microscopy Its lateral resolution is limited to ~200 nm.
– The need for the light-gathering objective lens and its aperture in
a conventional microscope leads to a diffraction limit
overcome the diffraction limit A fiber tip with an aperture <100
nm is scanned over a sample (forming the basis of techniques
like NSOM/SNOM – the scanning near-field optical
microscope).
– NSOM is now being using in conjunction with AFM – to study
“nano-phototonics”
– Resolution now down to ~30 nm
Confocal Scanning Microscopy
CSM) was first proposed by Marvin Minsky in 1957.
point is illuminated and focused at a time The light
reflected (or produced e.g by fluorescence) is detected
for just that point Light from out-of-focus areas is
suppressed A complete image is formed by scanning.
– Greater depth of field from images
– Images are free from out-of-focus blur
– Greater signal-to-noise ratio (for a spot – but images take time!)
– Better effective resolution (diffraction limit)
M Minsky, “Memoir on inventing the confocal scanning microscope” Scanning, 1988, 10, 128-138.
Trang 7Confocal Scanning Microscopy: Imaging Types
scanning:
– disks with staggered holes – block all but a certain lateral portion of
the sample beam
Laser Confocal Scanning Microscopy
fluorescent dye to stain
biological samples
Diagram from http://www.cs.ubc.ca/spider/ladic/images/system.gif
Trang 8Laser Confocal Scanning Microscopy
Diagram from http://www.cs.ubc.ca/spider/ladic/images/system.gif
Laser Confocal Scanning Microscopy
Diagram from http://www.cs.ubc.ca/spider/ladic/images/system.gif
Trang 9IR Microscopy and Spectroscopy
– However, this is at the expense of resolution (limited to ca 10 um),
in contrast with scanning techniques Resolution in FTIR imaging is
of course limited by the diffration limit, which is even worse for IR
wavelengths.
Figure from J L Koenig, S Q Wang, and R Bhargava, Anal Chem., 73, 361A-369A (2001)
IR Microscopy: Image Analysis
color-coding peaks based on their IR frequency (a)
chosen via a scatter plot (c) of
every point in the image vs two
(or more) frequencies, followed
by location of the center-of-gravity
and possible statistical analysis
constructed (b)
Figure from J L Koenig, S Q Wang, and R Bhargava, Anal Chem., 73, 361A-369A (2001)
Trang 10IR Microscopy: Polymer Chemistry Applications
material science, chemical and biochemical applications
dissolution of a polymer by a solvent
Figure from J L Koenig, S Q Wang, and R Bhargava, Anal Chem., 73, 361A-369A (2001)
IR Microscopy: Polymer Chemistry Applications
molecular motion process is observed for polymers (e.g
polymethylstyrene) above their entanglement mwt:
Figure from J L Koenig, S Q Wang, and R Bhargava, Anal Chem., 73, 361A-369A (2001)
Trang 11Raman Microscopy
inherent resolution than IR
(uses lasers at shorter
optical wavelengths)
(must still scan the sample)
– this does have its
advantages though
systems for combined 3D
visualization and
spectroscopy
Raman Microscopy: Forensic Applications
one that is not so obvious is for forensic analysis of
colored fibers
fingerprint, and the complex spectra obtained from dye
mixtures can be used to determine if two fibers are
from the same origin
– The individual dyes used in fabics are varied, and their ratios
are especially varied (even from batch to batch!)
LC or ESI MS on dye-containing extracts from fibers
For more about forensic Raman microscopy, see: T A Brettell, N Rudin and R Saferstein, Anal Chem., 75, 2877-2890 (2003).
Trang 12Electron Microscopy (EM)
microscopy (SEM) – an
electron beam is scanned
in a raster pattern and
“reflected” effects are
different manner in EM
Bottom photo - http://www.mos.org/sln/sem/velcro.html Top photo - http://emu.arsusda.gov/snowsite/default.html
Velcro (x35) Ice crystals
optical SEM
Electron Microscopy: Basic Design
Electrongun (1-30 keV)
Magneticlenses andscanning coils
Sample
Detectors
Detectors
electronsphotons
electrons
Computer
Trang 13Electron Microscopy: Resolution
are impossible to discern with optical microscopy?
accelerated by a 10 kV potential:
nm 0.0123 m
10 23 1
) V C)(10 10
60 1 )(
kg 10 2(9.11
s J 10 63 6
2 2
2
11
4 19 - 31
-34
2 2 1
m m h
m
eV v
eV mv
EM can see >10000x more detail than visible light!
Note: Resolution
is limited by lens aberations!
Electron Microscopy: Resolution
EM can in some cases be moving pretty close to the
speed of light.
nm 10 7 3
) 1
)(
V C)(10 10
60 1 )(
kg 10 2(9.11
s J 10 63 6
) 1 ( 2 2
3
) 10 3 ( kg) 10 9.11 ( 2
) V C)(10 10 60 1 ( 4
19 - 31
-34 2
2 8 -31 4 19 - 2
mc eV
meV
h eV
m m h
At high potentials, EM can see atomic dimensions
Using the
relativistically
corrected form of the
previous equation:
Trang 14Electron Microscopy: Sample-Beam Interactions
control how both SEM and
TEM (i.e STEM) operate:
– Formation of images
– Spectroscopic/diffractometric
analysis
eight) types of sample-beam
interactions (which can be
confusing and hard to
remember!)
sample-beam interactions:
– bulk specimen interactions (bounce off sample – “reflected”)
– thin specimen interactions (travel through sample- “transmitted”)
SEM: Sample-Beam Interactions
Backscattered Electrons (~30 keV)
Caused by an incident electron colliding with an
atom in the specimen which is almost normal to
the incident electron’s path The electron is then
scattered "backward" 180 degrees
Backscattered electron intensity varies directly
with the specimen's atomic number This
differing production rates causes higher atomic
number elements to appear “brighter” than lower
atomic number elements This creates contrast
in the image of the specimen based on different
average atomic numbers
Backscattered electrons can come from a wide
area around the beam impact point (see pg 552
of Skoog) – this also limits the resolution of a
SEM (along with abberations in the EM lenses)
Trang 15SEM: Sample-Beam Interactions
Secondary Electrons (~5 eV)
Caused by an incident electron passing "near"
an atom in the specimen, close enough to
impart some of its energy to a lower energy
electron (usually in the K-shell) This causes a
slight energy loss, a change in the path of the
incident electron and ionization of the electron in
the specimen atom The ionized electron then
leaves the atom with a very small kinetic energy
(~5 eV) One incident electron can produce
several secondary electrons
Production of secondary electrons is closely
linked to sample topography Their low energy
(~5 eV) means that only electrons very near to
the surface (<10 nm) are detected They also
don’t suffer from the backscattered electron
lateral resolution problem depicted in Fig 21-16
of Skoog Changes in topography in the sample
that are larger than this sampling depth can
change the yield of secondary electrons via
indirect effects (called collection efficiencies)
Electron Microscope: Image Formation
From JEOL Application Note, V E Robertson, ca 1985.
Trang 16SEM: Sample-Beam Interactions
Auger Electrons (10 eV – 2 keV)
Caused by relaxation of an ionized atom after a
secondary electron is produced The lower
(usually K-shell) electron that was emitted from
the atom during the secondary electron process
has left a vacancy A higher energy electron from
the same atom can drop to a lower energy, filling
the vacancy This leaves extra energy in the atom
which can be corrected by emitting a
weakly-bound outer electron; an Auger electron
Auger electrons have a characteristic energy,
which is unique and depends on the emitting
element Auger electrons have relatively low
energy and are only emitted from the bulk
specimen from a depth of several angstroms
SEM: Sample-Beam Interactions
X-ray Emission
Caused by relaxation of an ionized atom after a
secondary electron is produced Since a lower
(usually K-shell) electron was emitted from the
atom during the secondary electron process an
inner (lower energy) shell now has a vacancy A
higher energy electron can "fall" into the lower
energy shell, filling the vacancy As the electron
"falls" it emits energy in the form of X-rays to
balance the total energy of the atom
X-rays emitted from the atom will have a
characteristic energy which is unique to the
element from which it originated
X-ray (elemental) mapping of sample surfaces is
a common applications and a very powerful
analytical approach
Trang 17SEM: Sample-Beam Interactions – X-rays
SEM: Sample-Beam Interactions
Cathodoluminescence (CL)
Caused by electron hole pairs, which are created
by the electron beam in certain kinds of materials
When the pairs recombine, cathodoluminescence
(CL) can result CL is the emission of
UV-Visible-IR light by the recombination effect CL is usually
very weak and covers a wide range of
wavelengths, and requires high beam currents,
lowering resolution and challenging detector
systems!
CL signals typically result from small impurities in
an otherwise homogeneous material, or lattice
defects in a crystal
CL can be used effectively for some analytical
problems Some “random” examples:
– Differentiation of anatase and rutile
– Studying ferroelectric domains in sodium niobate
– Location of subsurface crazing in ceramics
– Forensic analysis of glasses
Trang 18TEM: Sample-Beam Interactions (Thin Sample)
Unscattered Electrons
Incident electrons which are transmitted
through the thin specimen without any
interaction occurring inside the specimen
Used to image - the transmission of
unscattered electrons is inversely proportional
to the specimen thickness Areas of the
specimen that are thicker will have fewer
transmitted unscattered electrons and so will
appear darker, conversely the thinner areas
will have more transmitted and thus will
appear lighter
TEM: Sample-Beam Interactions (Thin Sample)
Elastically-Scattered Electrons
Incident electrons that are scattered (deflected
from their original path) by atoms in the
specimen in an elastic fashion (without loss of
energy) These scattered electrons are then
transmitted through the remaining portions of
the specimen
Electrons follow Bragg's Law and are
diffracted All incident electrons have the
same energy (and wavelength) and enter the
specimen normal to its surface So all incident
electrons that are scattered by the same
atomic spacing will be scattered by the same
angle These "similar angle" scattered
electrons can be collated using magnetic
lenses to form a pattern of spots; each spot
corresponding to a specific atomic spacing,
This pattern can then yield information about
the orientation, atomic arrangements and
phases present in the area being examined
Trang 19TEM: Sample-Beam Interactions (Thin Sample)
Inelastically-Scattered Electrons
Incident electrons that interact with sample atoms
inelastically (losing energy during the interaction)
These scattered electrons are then transmitted
through the rest of the sample
Inelastically scattered electrons have two uses:
1 Electron Energy Loss Spectroscopy (EELS): The
amount of inelastic loss of energy by the incident
electrons can be used to study the sample
These energy losses are unique to the bonding
state of each element and can be used to extract
both compositional and bonding (i.e oxidation
state) information on the sample region being
examined
2 Kakuchi bands: Bands of alternating light and dark
lines caused by inelastic scattering, which are
related to interatomic spacing in the sample
These bands can be either measured (their width
is inversely proportional to atomic spacing) or
used to help study the elasticity-scattered electron
pattern
Electron Optics: Electron Source (Gun)
Positive electrical potential applied to the anode
The filament (cathode) is heated until a stream of
electrons is produced
The electrons are then accelerated by the positive
potential down the column (can be up to 30 kV)
A negative electrical potential (~500 V) is applied
to the Wehnelt cap
Electrons are forced toward the column axis by
the Wehnelt cap
Electrons collect in the space between the
filament tip and Wehnelt cap (a space charge or
“pool”)
Those electrons at the bottom of the space
charge (nearest to the anode) can exit the gun
area through the small (<1 mm) hole in the
Wehnelt cap
These electrons then move down the column
towards the EM lens and scanning systems
Figure from http://www.unl.edu/CMRAcfem/interact.htm
Thermionic electron gun – how it works:
The results:
- Electrons are emitted from
a nearly perfect point source (the space charge)
- The electrons all have similar energies (monchromatic)
- The electrons will travel parallel to the column axis