measurements the solution for the design for n=4 angle blocks is as follows:The solution for the reference standard is shown under the first column of thesolution matrix; for the check s
Trang 1Block sizes Angle blocks normally come in sets of
1, 3, 5, 20, and 30 seconds
1, 3, 5, 20, 30 minutes
1, 3, 5, 15, 30, 45 degreesand blocks of the same nominal size from 4, 5 or 6 different sets can becalibrated simultaneously using one of the designs shown in this catalog
Design for 4 angle blocks
Restraint The solution to the calibration design depends on the known value of a
reference block, which is compared with the test blocks The reference block
is designated as block 1 for the purpose of this discussion
Check
standard
It is suggested that block 2 be reserved for a check standard that is maintained
in the laboratory for quality control purposes
Trang 2Series 1: 2-3-2-1-2-4-2Series 2: 5-2-5-1-5-3-5Series 3: 4-5-4-1-4-2-4Series 4: 3-4-3-1-3-5-3
Measurements
for 6 angle
blocks
Series 1: 2-3-2-1-2-4-2Series 2: 6-2-6-1-6-3-6Series 3: 5-6-5-1-5-2-5Series 4: 4-5-4-1-4-6-4Series 5: 3-4-3-1-3-5-3
The equations explaining the seven measurements for the first series in terms
of the errors in the measurement system are:
with B a bias associated with the instrument, d is a linear drift factor, X is the
value of the angle block to be determined; and the error terms relate torandom errors of measurement
Trang 3The check block, C, is measured before and after each test block, and the
difference measurements (which are not the same as the differencemeasurements for calibrations of mass weights, gage blocks, etc.) areconstructed to take advantage of this situation Thus, the 7 readings arereduced to 3 difference measurements for the first series as follows:
For all series, there are 3(n - 1) difference measurements, with the first
subscript in the equations above referring to the series number The differencemeasurements are free of drift and instrument bias
Design matrix As an example, the design matrix for n = 4 angle blocks is shown below.
1 1 1 1
0 1 -1 0 -1 1 0 0
0 1 0 -1
0 -1 0 1 -1 0 0 1
0 0 -1 1
0 0 1 -1 -1 0 1 0
0 -1 1 0 The design matrix is shown with the solution matrix for identificationpurposes only because the least-squares solution is weighted (Reeve) toaccount for the fact that test blocks are measured twice as many times as thereference block The weight matrix is not shown
Trang 4measurements the solution for the design for n=4 angle blocks is as follows:
The solution for the reference standard is shown under the first column of thesolution matrix; for the check standard under the second column; for the firsttest block under the third column; and for the second test block under thefourth column Notice that the estimate for the reference block is guaranteed
to be R*, regardless of the measurement results, because of the restraint that
is imposed on the design Specifically,
Solutions are correct only for the restraint as shown
Trang 5For n blocks, the differences between the values for the blocks measured in
the top ( denoted by "t") and bottom (denoted by "b") positions are denotedby:
The standard deviation of the average (for each block) is calculated fromthese differences to be:
s test = K1s1where K 1 is shown under "Factors for computing repeatability standarddeviations" for each design and is the repeatability standard deviation asestimated from the design Because this standard deviation may seriouslyunderestimate the uncertainty, a better approach is to estimate the standarddeviation from the data on the check standard over time An expandeduncertainty is computed according to the ISO guidelines
Trang 62 Measurement Process Characterization
2.3 Calibration
2.3.4 Catalog of calibration designs
2.3.4.5 Designs for angle blocks
2.3.4.5.1 Design for 4 angle blocks
Trang 7Y(21) 0 -5.0516438 -1.2206578 2.2723000
Y(22) 0 7.3239479 7.3239479 9.3521166
Y(23) 0 -1.2206578 -5.0516438 2.2723000
Y(31) 0 -1.2206578 2.2723000 -5.0516438
Y(32) 0 7.3239479 9.3521166 7.3239479
Y(33) 0 -5.0516438 2.2723000 -1.2206578
R* 1 1 1 1 R* = VALUE OF REFERENCE ANGLE BLOCK
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
Trang 82 Measurement Process Characterization
2.3 Calibration
2.3.4 Catalog of calibration designs
2.3.4.5 Designs for angle blocks
2.3.4.5.2 Design for 5 angle blocks
Trang 9Y(13) 0.00000 2.48697 -0.89818 -4.80276 -0.78603 Y(21) 0.00000 -5.48893 -0.21200 -1.56370 3.26463 Y(22) 0.00000 5.38908 5.93802 4.71618 7.95672 Y(23) 0.00000 -0.89818 -4.80276 -0.78603 2.48697 Y(31) 0.00000 -0.21200 -1.56370 3.26463 -5.48893 Y(32) 0.00000 5.93802 4.71618 7.95672 5.38908 Y(33) 0.00000 -4.80276 -0.78603 2.48697 -0.89818 Y(41) 0.00000 -1.56370 3.26463 -5.48893 -0.21200 Y(42) 0.00000 4.71618 7.95672 5.38908 5.93802 Y(43) 0.00000 -0.78603 2.48697 -0.89818 -4.80276 R* 1 1 1 1 1.
R* = VALUE OF REFERENCE ANGLE BLOCK
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
Trang 102 Measurement Process Characterization
2.3 Calibration
2.3.4 Catalog of calibration designs
2.3.4.5 Designs for angle blocks
2.3.4.5.3 Design for 6 angle blocks
Trang 11Y(11) 0.0000 3.2929 -5.2312 -0.7507 -0.6445 -0.6666
Y(12) 0.0000 6.9974 4.6324 4.6495 3.8668 3.8540
Y(13) 0.0000 3.2687 -0.7721 -5.2098 -0.6202 -0.6666
Y(21) 0.0000 -5.2312 -0.7507 -0.6445 -0.6666 3.2929
Y(22) 0.0000 4.6324 4.6495 3.8668 3.8540 6.9974
Y(23) 0.0000 -0.7721 -5.2098 -0.6202 -0.6666 3.2687
Y(31) 0.0000 -0.7507 -0.6445 -0.6666 3.2929 -5.2312
Y(32) 0.0000 4.6495 3.8668 3.8540 6.9974 4.6324
Y(33) 0.0000 -5.2098 -0.6202 -0.6666 3.2687 -0.7721
Y(41) 0.0000 -0.6445 -0.6666 3.2929 -5.2312 -0.7507
Y(42) 0.0000 3.8668 3.8540 6.9974 4.6324 4.6495
Y(43) 0.0000 -0.6202 -0.6666 3.2687 -0.7721 -5.2098
Y(51) 0.0000 -0.6666 3.2929 -5.2312 -0.7507 -0.6445
Y(52) 0.0000 3.8540 6.9974 4.6324 4.6495 3.8668
Y(53) 0.0000 -0.6666 3.2687 -0.7721 -5.2098 -0.6202
R* 1 1 1 1 1 1
R* = VALUE OF REFERENCE ANGLE BLOCK
FACTORS FOR REPEATABILITY STANDARD DEVIATIONS
SIZE K1
Trang 132 Measurement Process Characterization
temperature of the bath with a standard resistance thermometer at the
beginning, middle and end of each run of K test thermometers The test
thermometers themselves are measured twice during the run in thefollowing time sequence:
where R1, R2, R3 represent the measurements on the standard resistance
thermometer and T1, T2, , T K and T'1, T'2, , T' K represent the pair
of measurements on the K test thermometers.
Assumptions
regarding
temperature
The assumptions for the analysis are that:
Equal time intervals are maintained between measurements onthe test items
Trang 14deviations
The residual variance is given by
.The standard deviation of the indication assigned to the ith testthermometer is
and the standard deviation for the estimates of shift and drift are
respectively
Trang 152 Measurement Process Characterization
2.3 Calibration
2.3.4 Catalog of calibration designs
2.3.4.7 Humidity standards
Humidity standards The calibration of humidity standards
usually involves the comparison ofreference weights with cylinderscontaining moisture The designs shown
in this catalog are drift-eliminating andmay be suitable for artifacts other thanhumidity cylinders
List of designs
2 reference weights and 3 cylinders
●
Trang 162 Measurement Process Characterization
2.3 Calibration
2.3.4 Catalog of calibration designs
2.3.4.7 Humidity standards
2.3.4.7.1 Drift-elimination design for 2
reference weights and 3 cylinders
OBSERVATIONS 1 1 1 1 1
Y(1) +
Y(2) +
Y(3) +
Y(4) +
Y(5) - +
Y(6) - +
Y(7) +
Y(8) +
Y(9) - +
Y(10) +
RESTRAINT + +
CHECK STANDARD + DEGREES OF FREEDOM = 6 SOLUTION MATRIX DIVISOR = 10 OBSERVATIONS 1 1 1 1 1
Trang 17Y(1) 2 -2 0 0 0
Y(2) 0 0 0 2 -2
Y(3) 0 0 2 -2 0
Y(4) -1 1 -3 -1 -1
Y(5) -1 1 1 1 3
Y(6) -1 1 1 3 1
Y(7) 0 0 2 0 -2
Y(8) -1 1 -1 -3 -1
Y(9) 1 -1 1 1 3
Y(10) 1 -1 -3 -1 -1
R* 5 5 5 5 5
R* = average value of the two reference weights FACTORS FOR REPEATABILITY STANDARD DEVIATIONS WT K1 1 1 1 1 1 1 0.5477 +
1 0.5477 +
1 0.5477 +
2 0.8944 + +
3 1.2247 + + +
0 0.6325 + -
Explanation of notation and interpretation of tables
Trang 182 Measurement Process Characterization
2.3 Calibration
2.3.5 Control of artifact calibration
Purpose The purpose of statistical control in the calibration process is to
guarantee the 'goodness' of calibration results within predictable limitsand to validate the statement of uncertainty of the result Two types ofcontrol can be imposed on a calibration process that makes use ofstatistical designs:
Control of instrument precision or short-term variability
1
Control of bias and long-term variability
Example of a Shewhart control chart
estimates the basic precision of the instrument Designs should bechosen to have enough measurements so that the standard deviationfrom the design has at least 3 degrees of freedom where the degrees of
freedom are (n - m + 1) with
n = number of difference measurements
The check standard should be of the same type and geometry as itemsthat are measured in the designs These artifacts must be stable andavailable to the calibration process on a continuing basis There should
be a check standard at each critical level of measurement For example,for mass calibrations there should be check standards at the 1 kg; 100 g,
10 g, 1 g, 0.1 g levels, etc For gage blocks, there should be check
Trang 19standards at all nominal lengths.
A check standard can also be a mathematical construction, such as thecomputed difference between the calibrated values of two referencestandards in a design
Database of
check
standard
values
The creation and maintenance of the database of check standard values
is an important aspect of the control process The results from eachcalibration run are recorded in the database The best way to record thisinformation is in one file with one line (row in a spreadsheet) of
information in fixed fields for each calibration run A list of typicalentries follows:
Trang 202 Measurement Process Characterization
separate control chart is required for each instrument except whereinstruments are of the same type with the same basic precision, in whichcase they can be treated as one
The baseline is the process standard deviation that is pooled from k = 1, , K individual repeatability standard deviations, , in the database,each having degrees of freedom The pooled repeatability standarddeviation is
with degrees of freedom
Trang 21
UCL Usually, only the upper control limit is of interest because we are
primarily interested in detecting degradation in the instrument'sprecision A possible complication is that the control limit is dependent
on the degrees of freedom in the new standard deviation and iscomputed as follows:
The quantity under the radical is the upper percentage point from the
F table where is chosen small to be, say, 05 The other two termsrefer to the degrees of freedom in the new standard deviation and thedegrees of freedom in the process standard deviation
Limitation
of graphical
method
The graphical method of plotting every new estimate of repeatability on
a control chart does not work well when the UCL can change with each
calibration design, depending on the degrees of freedom The algebraicequivalent is to test if the new standard deviation exceeds its controllimit, in which case the short-term precision is judged to be out ofcontrol and the current calibration run is rejected For more guidance,see Remedies and strategies for dealing with out-of-control signals
As long as the repeatability standard deviations are in control, there isreason for confidence that the precision of the instrument has notdegraded
Trang 222 Measurement Process Characterization
Need for
monitoring
precision
The precision of the balance is monitored to check for:
Slow degradation in the balance
The standard deviations over time and many calibrations are tracked and monitored using a
control chart for standard deviations The database and control limits are updated on a yearly or bi-yearly basis and standard deviations for each calibration run in the next cycle are compared with the control limits In this case, the standard deviations from 117
calibrations between 1975 and 1985 were pooled to obtain a repeatability standard
deviation with v = 3*117 = 351 degrees of freedom, and the control limits were computed
at the 1% significance level.
read mass.dat t id y bal s ds let n = size s
y1label MICROGRAMS x1label TIME IN YEARS xlimits 75 90
x2label STANDARD DEVIATIONS ON BALANCE 12 characters * blank blank blank
lines blank solid dotted dotted let ss=s*s
let sp=mean ss let sp=sqrt(sp) let scc=sp for i = 1 1 n let f = fppf(.99,3,351)
Trang 23let f=sqrt(f) let sul=f*scc plot s scc sul vs t
Further
interpretation
of the control
chart
However, it is also clear from the pattern of standard deviations over time that the precision
of the balance was gradually degrading and more and more points were approaching the control limits This finding led to a decision to replace this balance for high accuracy calibrations.
Trang 242 Measurement Process Characterization
2.3 Calibration
2.3.5 Control of artifact calibration
2.3.5.2 Control of bias and long-term
The check standard values are denoted by
The baseline is the process average which is computed from the checkstandard values as
The process standard deviation is
with (K - 1) degrees of freedom.
Trang 25with denoting the upper critical value from the
t-table with v = (K - 1) degrees of freedom.
let v1 = k-1let t = tppf(alphau, v1)return the following value:
THE COMPUTED VALUE OF THE CONSTANT T =0.2570583E+01
C > UCL
Trang 26to see if the failure was a chance occurrence Check standard valuesthat remain in control, especially over a period of time, provideconfidence that no new biases have been introduced into themeasurement process and that the long-term variability of the processhas not changed.
Out-of-control
signals that
recur require
investigation
Out-of-control signals, particularly if they recur, can be symptomatic
of one of the following conditions:
Change or damage to the reference standard(s)