If the anatomic areas are overexposed, the optical densities fall within the Use of high tube voltage results in a results in a decrease in the visibility of detail in a much lower radia
Trang 1Clinical Applications of Basic X-ray Physics Principles1
This artide meets the
criteria for 1.0 credIt
the AMA Physician’s
Recognition Award.
To obtain credit, see
the questionnaire on
pp 725-730.
LEARNING
OBJECTIVES
After reading tbis article
and taking the test, the
reader will:
selection of x-ray exposure
factors influences image
density and contrast.
. Be able to identify
causes of unsharpness in a
radiograph and learn how
the visibility of detail is
affected by technique and
equipment factors.
. Be able to identify the
four basic types of x-ray
generators and to select
the appropriate type for a
particular radiographic
application.
various scatter reduction
mechanisms with respect
to image quality and
patient dose
current, and exposure time-determine the basic characteristics of radiation
I #{149} INTRODUCTION
An understanding of basic x-ray physics includes knowledge of the principles of x-ray
principles to produce an image that allows the radiologist to visualize the internal
tech-nique factors and equipment design for a particular clinical examination
gener-ally has a detrimental effect on the other factors Image evaluation must also include
Index terms: Physics ‘ Radiography
RadloGraphics 1998; 18:731-744
‘From the Department of Diagnostic Radiology, Mayo Clinic, 200 First St, SW, Rochester, MN 55905 From the AAPM/ RSNA Physics Tutorial at the 1996 RSNA scientific assembly Received March 1 3, 1998; revision requested May 27 and re-ceived November 24; accepted November 28 Address reprint requests to the author.
©RSNA, 1998
Trang 2732 #{149}Imaging & Therapeutic Technology Volume 18 Number 3
demon-strate how radiographic density varies with different milliampere-second values The normal radiograph (b)
was acquired at 70 kVp and 16 mAs With the kilovoltage unchanged, the underexposed radiograph (a) was
ra-diographic density varies with different kilovoltage values The normal radiograph (e) was acquired at 70 kVp
and 16 mAs With the miffiampere-seconds unchanged, the underexposed radiograph (d) was acquired with a
kVp The 10-kVp rule is demonstrated by observing that the densities of the underexposed radiographs (a, d) are similar and the densities of the overexposed radiographs (c, f) are similar
expo-sure Therefore, it is important to consider
patient exposure as tow as possible
size, x-ray generator type, and scatter rejection
addi-tional information on principles of basic x-ray
physics and image quality (1-7)
FACTORS
den-sity and contrast of the radiograph and the
pa-tient exposure
current and exposure time, expressed as
also affect density, but in this case increasing
rays through the patient As a result, small
mdli-ampere-seconds and tube voltage that results
known as the 10-kVp rule: An increase of 10 kVp is equivalent to doubling the milliampere-seconds
density similar to that achieved by reducing the
milliampere-seconds from 16 mAs to 8 mAs,
32 mAs It should be noted that the 10-kVp rule does not apply for radiographs acquired at <60
the extremities
contrast is defmed as the difference in
relative radiation intensities of the x-ray beam
exiting the patient The subject contrast is
adjacent background tissue The penetrability,
ef-fective energy of the x-ray beam: Higher-energy
tow-energy beams do Because x-ray beam energy is
directly affected by changing the tube voltage, the latter is a major factor in determining radio-graphic contrast
Trang 3a d.
b.
f.
-c.
Trang 4-4
Shoulder
C,)
0
C)
C.
0
log Relative Exposure
Figure 2 Effect of tube voltage on contrast and dose (a) High-contrast radiograph of a skull
phan-torn was acquired at 70 kVp and 60 mAs (b) Low-contrast radiograph of a skull phantom was
ac-quired at 100 kVp and 9 mAs In addition to a reduction in contrast, the increase in kilovoltage results
in a large reduction in patient exposure The entrance skin exposure for the low-contrast radiograph
is 1 1 5 mR (0.297 x 10’ C/kg), whereas the skin exposure produced in the high-contrast radiograph is
Figure 3 Characteristic curve for a screen-film
represented by the solid and dashed lines A larger
dii-ference in optical densities between the two areas
indi-cates higher contrast is present in the image When the
anatomic areas are properly exposed, the optical
densi-ties fall within the linear portion of the characteristic
curve and the contrast is greatest If the anatomic areas
are overexposed, the optical densities fall within the
Use of high tube voltage results in a
results in a decrease in the visibility of detail in
a much lower radiation exposure to the patient
compared with that needed in the 70-kVp
beam is absorbed by the patient
Milliampere-Seconds Selection
Trang 5.
May-June 1998 Schueler #{149}RadioGrapbics U 735
C.
film radiography, both underexposure
characteristic curve (or Hurter and Driffield
[H&D] curve) describes the relationship
has three regions that correspond to different
exposure levels For low- and high-exposure
levels, the slope of the curve is relatively small
steep slope Within the straight-line portion,
)‘
Figure 4. Loss of contrast due to improper exposure Underexposed radiograph of a skull phantom acquired at 70 kVp and 30 mAs (a) and the overexposed radiograph acquired at 70
kVp and 1 20 mAs (c) are lower in contrast corn-pared with the normal exposure acquired at 70
kVp and 60 mAs (t)
anatomic areas wilt be the largest for the
ex-posure results in densities that lie in the toe or
Un-der- and overexposure on contrast in a clinical
The choice of focal spot size primarily
also influences the amount of motion blur in an
tube current and tube voltage settings, thereby
affecting the exposure time In addition, the
may influence the focal spot size, radiation
pro-vided by the x-ray tube
rays Instead, it is a rectangular region of finite
size This causes a point in an object to appear
Trang 6SOD
SID
OlD
1
6a.
Object Plane
Bf
blurred on the image The amount of blur in
Bf = F x (OlD/SOD),
where F = focal spot size, OlD = object-image
compare the focal spot blur to the size of the
object itself, we calculate the blur in the plane
of the object (Bf) Bf0 is determined by
The effect of focal spot size and magnifica-tion on blur in a clinical image is demonstrated
fo-cal spot blur wifi increase as the focal spot size
mar-gins are indistinct and some fme structures
Focal Spot
5.
Figures 5, 6 (5) Focal spot blur Diagram illustrates how the focal spot blur in the image _
closer to the focal spot OlD = object-image distance, SID = source-image distance, SOD = source-object distance
(6) Effect of focal spot size and magnification on blur (a, b) Radio-graph of the sella turcica, obtained with a small focal spot of nominal size 0.3 mm (measured size, 0.5 mm) (a), exhibits greater detail than the radiograph obtained with a large focal spot of nominal size 1.0
mm (measured size, 1 8 mm) (b) Both a and b have the same
nominal size 1 0 mm (measured size, 1 8 mm) but with the object in
corn-pared with b, even though a large focal spot was used
ing Bf by the magnification M (M = SID/SOD,
Bf0 = Bf/M
Trang 71 1.2 1.4 1.6 1.8 2 2.2 2.4
Magnification
0.1
0
a.
Figure 7 Blur in the object plane as a function of magnification (a) A radiographic system with a 1 0-mm
fo-cal spot and a high speed screen has a minimum composite blur at a magnification of 1.5 (b) A system with a
1 0-mm focal spot and a detail screen has a minimum composite blur when no magnification is used
b.
Magnification
magnification (Fig 6c). When there is no
magni-fication (M = 1), the focal spot blur is zero if
focal spot closer to the object, the focal spot
blur will increase
contribute to the total image blur in a
in-teracting with the intensifying screen Because
thickness of the screen phosphor layer A
(Br) of approximately 0.7 mm, whereas the
blur from a thin, detail screen is 0.2-0.3 mm
As with focal spot blur, it is more clinically
size of the object itself The receptor blur in
magnification:
Br = Br/M.
mag-nified
recep-tor blur It is calculated as the square root of
re-lationship between total image blur and
mag-nification is used, receptor blur dominates
the total image blur For this system, a magnifi-cation of 1.5 will produce the sharpest radio-graph Figure 7b demonstrates the
1.0-mm focal spot are used The detail screen
this system, the sharpest radiograph wifi be
ills-tance
Trang 8Anode
Focal Spot Track
Anode
Angle
I
Cathode
I I I
I
I Effective Focal
I
to-tal image blur is patient motion Motion blur is
possible However, the selection of technical
may result in an increase in focal spot blur We
the anode and results in a tower heat capacity
be-tween geometric unsharpness and motion blur
small focal spot should be used to reduce focal
large focal spot with higher tube current and
with respect to the central axis of the x-ray
angles that range from 7#{176}to 20#{176}.This
minimiz-ing the effective focal spot size, which is the
the filament as the anode rotates As a result,
rays
For a given effective focal spot size, the
provide the highest heat capacity, but radiation
Figure 8. Diagram depicts the side view of a rotat-ing anode assembly of an x-ray tube The angle
be-tween the anode surface and the central axis is
de-fined as the anode angle The effective focal spot size is the length and width of the x-ray beam pro-jected down the central axis
a larger area, but the rate of heat dissipation is
spot track In clinical practice, the choice of
radio-graphic examination For applications requiring
high heat capacity and small field coverage,
small anode angles are used General radiogra-phy requires large field coverage, so large
an-ode angles are needed
U GENERATOR SELECTION
qual-ity and patient exposure Just as with x-ray
com-promise of various factors depending on the
basic types of generators are single phase,
Trang 9po-Track Width Track Width
/
I I I
/
/ / / I
Field Coverage Field Coverage
-0
Figure 9. Comparison of small and large anode angles For a given effective focal spot
size, the choice of the anode angulation is a trade-off between heat capacity and field
cover-age (a) Diagram depicts the side view of an anode with a small angle The small angle
pro-vides a large focal spot track width for a high heat capacity, but the resultant radiation field
coverage is limited (b) Diagram depicts the side view of an anode with a large angle The
large angle provides larger radiation field coverage, but the rate of heat dissipation is low
because of the small width of the focal spot track
I fsJ#{149}SstJ#{149}SsJ#{149}CsJ\ Single phase
(two pulse)
(twelve pulse)
High frequency
Constant potential
Figure 10 Diagram illustrates representative
volt-age waveforms for single-phase, three-phase
potential x-ray generators
tential Selection criteria include minimization
low unit cost
Time
percentage difference between the maximum
single-phase generator exhibits a 100% voltage
genera-tor has a lower voltage ripple of 1 3%-25%.
Three-phase, 12-pulse generators have a ripple
of 3%- 10%, which is similar to the voltage
radiograph at a certain kilovoltage selection
low-energy x rays that do not contribute to the
Trang 10Primary X-rays
Image Receptor
con-stant potential generator provides the most
substantial reduction in patient dose
Generator types with large voltage ripple
also require longer exposure times, which
re-sults in greater motion blur This is because the
con-stant potential generator is capable of the
shortest exposure pulses of approximately 0.5
msec
An additional generator property affecting
times can be shorter for a desired
milliampere-seconds
reduce the number of retakes Reproducibility
is also critical for digital subtraction
angiogra-phy because differences in tube voltage
in-complete subtraction
High-frequency and constant potential
superior to that available with single- or
three-phase power output depends directly on
limits the ability to compensate for sudden line
correct for variations from the desired settings
grid placed between the patient and image receptor
helps reduce the amount of scattered radiation that reaches the receptor
con-stant potential generators provide the lowest
size and high cost of the system
size and relatively less expensive In addition,
high-frequency generators can be designed to run from either single- or three-phase line
for mobile radiographic units
A large fraction of the x rays entering a patient
undergo Compton interactions, which produce
emit-ted in all directions, but they tend to be
of the primary beam is increased When the
tissue, the radiographic density change between
the soft tissue and bone should be very large
scat-tered x rays, which strike the image receptor
Trang 11May-June 1998 Schueler U RadioGraphics #{149}741
rays that reach the receptor Two of these
meth-oils are the use of grids or an air gap
Collima-tion to limit the volume of irradiated tissue also
Grids
level of scattered radiation reaching the image
receptor is use of grids (9) A grid is constructed
of alternating strips oftead and nonabsorbing
ar-ranged to transmit only those x rays directed in
a line from the x-ray source (Fig 1 1) X rays
di-rected at an angle are preferentially absorbed by
large fraction of the scattered radiation is
entrance skin exposure for the radiograph
sub-stantially tower at 33 mR (0.085 x 10 C/kg)
ra-tio grids and higher energy exposures
scat-tered radiation that reaches the image receptor
is to place a gap between the patient and the
Trang 12Primary X-rays
Patient
primary beam, a large fraction will not strike
the receptor if it is separated from the patient
by a sufficient distance (Fig 13) However,
pri-mary x rays directed in a line from the x-ray
distance is 15-45 cm, which wifi also
Figure 14
Both grid and air gap techniques are effective
results in less tube loading In addition, the
the source-to-patient distance used
air gap technique is used in cerebral
angiogra-phy, the geometric magnification is generally
adjusted to 1 5- 1 8 Some radiologists prefer
posi-tioning the patient closer to the x-ray source is
this case, because a large source-image
and the resulting magnification is slight
Air
Gap
$
Figure 13 Cross-sectioned diagram shows how
an air gap placed between the patient and image
re-ceptor helps reduce the amount of scattered radia-tion that reaches the receptor
x-ray beam Object characteristics include thick-ness, physical density, and effective atomic number (Z). Two tissue areas that have either
different thicknesses, densities, or Z will
characteristics include tube voltage and the
gen-erator, plus filtration (ie, added and inherent
attenuating material in the path of the x-ray
kilovolt-age has a substantial effect on image contrast
In addition, factors such as filtration and volt-age waveform shape also alter the distribution
of x-ray energies in the beam
to-gether to produce subject contrast is
Trang 13Substance
Effective Atomic Number
Physical Density (g/cm3)
tions of x rays in tissue The probability of
pri-manly on tissue density, with very little
occurrence of photoelectric interactions has a
strong dependence on atomic number and
x-ray energy The probability of photoelectric
in-teractions increases as Z increases and
de-creases as energy increases
Imaging of soft tissue requires special attention
to radiographic technique Muscle, tissue
flu-ids, and fat have relatively low effective atomic
ton interactions are more likely to occur when
a patient is imaged with a high-energy x-ray beam (>40 keY) However, Compton interac-tions are not as effective as photoelectric inter-actions for distinguishing tissue types To
x-ray beam must be used An example of soft-tissue radiography is mammography in which technique factors of 25-30 kVp are used with a special x-ray tube anode and filter to produce
low-energy x rays
Contrast agents can be introduced into the body to change the attenuation of the object being imaged so that the subject contrast is
materials contain iodine and barium
den-sity Because the size of the structures normally
usually small, we need to maximize the differ-ential absorption Use of 60-75 kVp is prefer-able because it produces many x rays just