7 Information to be included in the relevant product standard The following information, which is required to conduct the environmental tests, shall be included in the relevant product
Terms and definitions
3.1.1 intruder alarm system alarm system that detects and indicates the presence, entry or attempted entry of an intruder into supervised premises
3.1.2 hold-up alarm system alarm system designed to permit the deliberate creation of an alarm condition in the case of a hold-up
3.1.3 social alarm system alarm system providing facilities to summon assistance for use by persons who can be considered to be living at risk
Fixed equipment refers to devices that are securely attached to a support or anchored in a designated location, making them immobile This includes items that lack carrying handles and possess a weight that prevents easy relocation, such as a control panel for an intruder alarm system that is mounted on a wall.
Movable equipment refers to non-fixed devices that are typically not in use when their location is altered An example of this is a local unit or controller for a social alarm system, which can be placed on a tabletop.
3.1.6 portable equipment equipment designed to be in operation while being carried (e.g access control "Smart card" badge, electronic key, social alarm trigger device carried by the user)
3.1.7 preconditioning treatment of a specimen before conditioning with the object of removing or partly counteracting the effects of its previous history
3.1.8 conditioning exposure of a specimen to environmental conditions in order to determine the effect of such conditions on the specimen
3.1.9 recovery treatment of a specimen after conditioning in order that the properties of the specimen may be stabilized before measurement.
Abbreviations
EMC: electromagnetic compatibility ppm: parts per million
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
This part of IEC 62599 specifies the tests and severities to be used for each of the following environmental classes:
I Indoor but restricted to residential/office environment (e.g living rooms and offices)
(e.g sales floors, shops, restaurants, stairways, manufacturing and assembly areas, entrances and storage rooms)
III Outdoor but sheltered from direct rain and sunshine, or indoor with extreme environmental conditions (e.g garages, lofts, barns and loading bays)
Classes I, II, III and IV are progressively more severe, and therefore class IV equipment may be used in class III applications, etc
Classes III and IV can be modified with the suffix "A" to create environmental classes IIIA and IVA, specifically designed for the extremely cold conditions in northern Europe These classes maintain the same specifications as their counterparts, with the primary distinction being the conditioning temperature for cold environments.
Testing for Classes IIIA and IVA will follow the same procedures as Classes III and IV, with the exception of the operational and temperature change tests These specific tests will utilize a lower conditioning temperature as specified in the relevant tables (refer to section 10.3.4).
Unless otherwise specified, the atmospheric conditions in the laboratory shall be the standard atmospheric conditions for measurements and tests, specified in 5.3.1 of IEC 60068-1:1988, as follows:
– air pressure: 86 kPa to 106 kPa
To ensure accurate measurements, it is crucial to minimize variations in parameters that significantly impact the results during a series of tests conducted on a single specimen.
Unless otherwise stated, the tolerances for the environmental test parameters shall be as given in the basic reference standards for the test (e.g the relevant part of IEC 60068-2)
7 Information to be included in the relevant product standard
To conduct environmental tests, the relevant product standard must include essential information such as the equipment class (fixed, movable, or portable), the mounting arrangements for the specimen, any deviations from the specified test procedures or severities, initial measurements or inspections required before conditioning (like a functional test), and the necessary state of the specimen during conditioning, including its configuration and operating conditions.
This document is licensed to Mecon Limited for internal use at the specified locations in Ranchi and Bangalore, and is supplied by the Book Supply Bureau It outlines the necessary monitoring, measurements, and inspections required during the conditioning process, including functional tests when applicable Additionally, it details the final measurements and inspections to be conducted post-conditioning, such as functional tests and visual inspections, along with any special recovery conditions needed prior to these assessments The document also specifies the pass/fail criteria and provides a test schedule that allocates specimens to each test.
The following points should be taken into account during the drafting of the product standard making reference to this standard:
– The information, a) to h) above, may differ from test to test or between types of tests (e.g between operational and endurance tests)
Certain equipment may face limitations that prevent the standard functional tests from being conducted during the conditioning phase of specific tests.
In certain scenarios, such as when placed in an environmental chamber, it may be essential to perform a reduced functional test or forgo the functional test entirely during conditioning Additionally, some tests do not allow for a functional test during conditioning due to the transient or evolving nature of the conditioning process.
– The product standards should indicate whether any memory back up batteries should remain connected during endurance tests, and if so whether the memory contents should be retained
Object of the test
To demonstrate the ability of the equipment to function correctly at high ambient temperatures, which may occur for short periods in the anticipated service environment.
Principle
The test involves subjecting the specimen to high temperatures for a duration that ensures temperature stability, allowing for the execution of functional tests and monitoring.
'Free air' conditions are simulated for heat dissipating specimens to allow for self-heating effects.
Test procedure
General
The test apparatus and procedure shall generally be as described in IEC 60068-2-2:2007
Gradual temperature change tests will be conducted, utilizing Test Bd for heat dissipating specimens as defined in IEC 60068-2-2, while Test Bb will be applied to non-heat dissipating specimens.
The dry heat operational test may be combined with the dry heat endurance test by omitting the recovery and the functional test in between.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 1:
Equipment class Fixed, movable and portable
Environmental class I II and III IV
The test conducted at 70 °C evaluates the impact of solar heat radiation using straightforward methods If these methods are deemed inadequate, an alternative is to utilize the simulated solar radiation test, which assesses temperature rise during operation as outlined in Clause 24.
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any measurements mandated by the product standard must be conducted within the final 30 minutes of this period.
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate the ability of the equipment to withstand long-term ageing effects.
Principle
The test consists of exposing the specimen to high temperatures for a long period to accelerate ageing effects.
Test procedure
General
The test apparatus and procedure shall generally be as described in IEC 60068-2-2:2007
Use test Bb for non heat issipating specimens
The dry heat endurance test may be combined with the dry heat operational test by omitting the recovery and the functional test in between.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Mount the specimen as specified in the product standard The specimen shall not be supplied with power during conditioning.
Conditioning
Apply the appropriate severity of conditioning shown in Table 2:
Equipment class Fixed, movable and portable
Environmental class I, II and III IV
The test conducted at 70 °C effectively accounts for the impact of solar heat radiation If this method is deemed inadequate, an alternative is to utilize the simulated solar radiation test, which measures temperature rise during operation as outlined in Clause 24.
Measurements during conditioning
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate the ability of the equipment to function correctly at low ambient temperatures appropriate to the anticipated service environment.
Principle
The test involves subjecting the specimen to low temperatures for an adequate duration to achieve temperature stability, enabling the performance of functional tests and monitoring.
'Free air' conditions are simulated for heat dissipating specimens to allow for self-heating effects.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-1:2007
Gradual temperature change tests will be conducted, utilizing Test Ad for heat dissipating specimens as defined in IEC 60068-2-1, and Test Ab for non-heat dissipating specimens.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 3:
Equipment class: Fixed, movable and portable
Environmental class I II III and IV
Duration (h) 16 16 16 a This temperature is –40 °C for classes IIIA and IVA (see Clause 4).
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any measurements mandated by the product standard must be conducted within the final 30 minutes of this period.
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate the ability of portable equipment to function correctly when exposed to temperature shocks when carried back and forth between normal and cold ambient temperature.
Principle
The test consists of exposing the specimen to a succession of changes of temperature The specimen is moved from one test chamber to another.
Test procedure
General
The test apparatus and procedure shall generally be as described in IEC 60068-2-14:2009
Test Na with rapid change of temperature, with prescribed time of transition, shall be used.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Conditioning
Apply the appropriate severity of conditioning shown in Table 4:
Environmental class I II III and IV
Change-over time t 2 (min) 2 to 3 2 to 3 2 to 3
Number of cycles 4 4 4 a For classes IIIA and IVA (see Clause 4), make this temperature –40 °C.
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any measurements mandated by the product standard must be conducted within the first 10 minutes of both the high and low temperature conditioning phases of the final cycle.
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
12 Damp heat, steady state (operational)
Object of the test
To demonstrate the ability of the equipment to function correctly at high relative humidity
(without condensation) which may occur for short periods in the service environment.
Principle
The test consists of exposing the specimen to a constant temperature and high relative humidity in such a manner that condensation does not occur on the specimen
The period of exposure is chosen to allow surface effects due to adsorption to be identified.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-78:2001
The damp heat, steady-state operational test may be combined with the damp heat, steady- state endurance test by omitting the recovery and the functional test in between.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 5:
Equipment class Fixed, movable and portable
Environmental class I II, III and IV
The damp heat cyclic operational test is essential for this condition If the damp heat cyclic test is not performed, the environmental class I test must be conducted for classes II, III, and IV.
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any required measurements specified by the product standard should be conducted in the final 30 minutes of this period.
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
13 Damp heat, steady state (endurance)
Object of the test
The equipment's resilience to long-term humidity effects in its operational environment is crucial, as it can lead to alterations in electrical properties due to moisture absorption, chemical reactions, and galvanic corrosion.
Principle
The test consists of exposing the specimen to a constant temperature and high relative humidity in such a manner that condensation does not occur on the specimen
A long period of exposure is chosen to allow effects due to absorption and chemical changes to take place.
Test procedure
General
The testing apparatus and methodology must adhere to the guidelines outlined in IEC 60068-2-78:2001 It is permissible to merge the damp heat, steady-state endurance test with the damp heat, steady-state operational test by eliminating the recovery phase and the functional test in between.
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen as specified in the product standard The specimen shall not be supplied with power during conditioning.
Conditioning
Apply the appropriate severity of conditioning shown in Table 6:
Equipment class Fixed, movable and portable
Environmental class I, II, III and IV
Measurements during conditioning
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate the immunity of the equipment to an environment with high relative humidity, where condensation occurs on the equipment.
Principle
The test involves subjecting the specimen to cyclic temperature fluctuations between 25 °C and a designated upper limit of either 40 °C or 55 °C During the high-temperature phase, relative humidity is kept at (93 ± 3)%, while it remains above 80% during the low-temperature and transitional phases The temperature increase rates are designed to ensure condensation forms on the specimen's surface.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-30:2005, using the variant No 2 test cycle and controlled recovery conditions
The damp heat cyclic operational test may be combined with the damp heat cyclic endurance test by omitting the recovery and the functional test in between
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 7:
Equipment class Fixed, movable and portable
Environmental class I II III and IV
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any measurements mandated by the product standard must be conducted within the final 30 minutes of the high-temperature phase of the last cycle.
Final measurements
After the recovery period, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate the ability of the equipment to withstand the long-term effects of high humidity and condensation.
Principle
The test consists of exposing the specimen to cyclic temperature variations between 25 °C and
The temperature is set at 55 °C, with relative humidity levels maintained at (93 ± 3) % during the high temperature phase and above 80 % during the low temperature and transitional phases The temperature increase rates are designed to ensure that condensation forms on the surface of the specimen.
Test procedures
General
The test apparatus and procedure shall be as described in IEC 60068-2-30:2005, using the variant no 2 test cycle and controlled recovery conditions
The damp heat cyclic endurance test may be combined with the damp heat cyclic operational test by omitting the recovery and the functional test in between
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen as specified in the product standard The specimen shall not be supplied with power during the conditioning.
Conditioning
Apply the appropriate severity of conditioning shown in Table 8:
Equipment class Fixed, movable and portable
Environmental class I and II III and IV
Measurements during conditioning
Final measurements
After the recovery period, subject the specimen to the final measurements required by the product standard
Object of the test
To demonstrate that the equipment is appropriately protected against the ingress of water.
Principle
The test evaluates the specimen's resistance to water through various application methods, tailored to the specific protection level needed This includes exposure to falling drops at angles up to 15° from vertical, water sprayed from all directions, and total immersion for portable devices as specified by the manufacturer.
Test procedure
General
The testing apparatus and methodology must adhere to the guidelines outlined in IEC 60068-2-18:2000 For dripping water tests, the Ra2 procedure should be implemented, while the Rb2.1 or Rb2.2 procedures are designated for spraying water tests Total immersion tests should follow the Rc1 procedure.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Ensure the specimen is mounted in its intended orientation and operating condition as per the product standard, following the manufacturer's installation instructions Utilize any provided weather protection accessories along with the appropriate cables and cable glands.
Conditioning
Apply the appropriate severity of conditioning shown in Table 9 or 10:
Equipment class Fixed and movable
Environmental class I and II III IV
Test procedure Ra2 Rb1.1 or Rb1.2
Water flow/nozzle (dm 3 min –1 ) 0,07 % ± 5 %
Water flow rate (dm 3 min –1 ) 10 % ± 5 %
IPX2 IPX4 a From all directions with the shield removed b 3 minutes per m 2 surface area with a minimum of 15 min
Environmental class I and II III and IV Optional a
Test procedure Ra2 Rb1.1 or Rb1.2 Rc1
Water flow/nozzle (dm 3 min –1 ) 0,07 % ± 5 %
Water flow rate (dm 3 min –1 ) 10 % ± 5 %
The IEC 60529 classification includes ratings such as IPX2, IPX4, and IPX7, which indicate the device's resistance to water immersion This classification applies when the manufacturer specifies that the device can withstand immersion in water Testing is conducted from all directions with the shield removed, requiring a duration of 3 minutes per square meter of surface area, with a minimum exposure time of 15 minutes.
Measurements during conditioning
Monitor the specimen during the conditioning period to detect any change in status
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Final measurements
After the conditioning, subject the specimen to the final measurements required by the product standard and inspect it for any damage or water ingress
Before final measurements, any specific recovery conditions, such as specimen drying, must be outlined in the product standard Additionally, the standard should clarify whether water entry into the specimen enclosure is permissible If this is not specified, the manufacturer should indicate if water penetration is acceptable.
Object of the test
To demonstrate the ability of the equipment to withstand the corrosive effects of sulphur dioxide as an atmospheric pollutant.
Principle
The test involves subjecting the specimen to an atmosphere rich in sulphur dioxide while maintaining a constant temperature and high relative humidity It is crucial to keep the surface temperature of the specimen above the dew point; however, the presence of hygroscopic materials, either on the specimen or as byproducts of corrosion, may result in condensation.
Test procedure
General
The test apparatus and procedure shall be generally as described in IEC 60068-2-42:2003, except for the relative humidity of the test atmosphere, which shall be maintained at (93 ± 3) % instead of (75 ± 5) %.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen according to the product standard, ensuring it is not powered during conditioning Connect untinned copper wires of the appropriate diameter to sufficient terminals for conducting the functional test.
After the conditioning, without making further connections to the specimen.
Conditioning
Apply the appropriate severity of conditioning shown in Table 11:
Equipment class Fixed, movable and portable
Environmental class I II III IV
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Measurements during conditioning
Final measurements
Immediately After the conditioning, subject the specimen to a drying period of 16 h at 40 °C,
≤ 50 % RH, followed by a recovery period of at least 1 h at standard laboratory conditions
Then subject the specimen to final measurements as required by the product standard Inspect the specimen visually for mechanical damage both externally and internally
Object of the test
To demonstrate an adequate level of protection against corrosion for equipment exposed to the elements.
Principle
The test involves placing the specimen in its standard position and subjecting it to a designated number of salt mist spray cycles, each followed by a storage period in humid conditions.
Test procedure
General
The test apparatus and procedure shall comply with the requirements of IEC 60068-2-52:1996.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
The specimen must be installed in its designated orientation following the manufacturer's guidelines, utilizing any provided weather protection accessories along with the suitable cables and cable glands.
The specimen must remain unpowered during conditioning, but it should be equipped with untinned copper wires of the correct diameter connected to adequate terminals This setup allows for functional testing to be conducted after conditioning without the need for additional connections to the specimen.
Conditioning
Apply the appropriate severity of conditioning shown in Table 12:
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Equipment class Fixed, movable and portable
Environmental class I, II and III IV
Salt mist exposure: Salt (NaCl) concentration (%) a pH of salt solution
Measurements during conditioning
Final measurements
After conditioning, let the specimen cool for 1 to 2 hours under standard laboratory conditions before conducting the final measurements as per the product standard Additionally, check the specimen for any mechanical damage, both externally and internally.
NOTE Any special cleaning procedure(s) required before the measurements (e.g drying of the specimen) should be specified in the product standard
Object of the test
To demonstrate the immunity of the equipment to mechanical shocks, which are likely to occur, in the service environment.
Principle
The test involves applying a series of shock pulses to the specimen through standard mounting points Each shock pulse is characterized by its maximum acceleration amplitude, duration, and the specific shape of the acceleration/time relationship Although the selected pulse shape, such as a half sine wave, may not occur in a pure form in real-world scenarios, it offers a consistent method for simulating the effects of more realistic shock events.
The shock amplitude (peak acceleration) is related to the mass of the specimen, as shown in
Figure 1, in order to limit the energy imparted to heavier specimens.
Test procedure
General
The test apparatus and procedure shall generally be as described in IEC 60068-2-27:2008, for a half sine wave pulse, but with the peak acceleration related to specimen mass as indicated below
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 13:
Equipment class Fixed, movable a and portable a
Environmental class I, II, III and IV
Peak acceleration  (m s –2 ), related to the specimen mass M (kg) b : M < 4,75
For movable and portable devices, the shock test can be omitted if false alarms are not tolerated during the free fall test The number of pulses per direction is three, and both directions (+ and -) must be tested across three mutually perpendicular axes, as illustrated in Figure 1.
Pe ak a cce le ra tio n  ( m s –2 )
Figure 1 – Graph showing peak acceleration versus specimen mass
Measurements during conditioning
Monitor the specimen during the conditioning period to detect any change in status
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Final measurements
After the conditioning, subject the specimen to the final measurements required by the product standard and inspect it visually for mechanical damage both externally and internally
Object of the test
The article highlights the importance of assessing the immunity of both fixed and movable equipment against mechanical impacts that may occur on their surfaces during normal service conditions It emphasizes the need for these equipment types to be capable of withstanding such impacts effectively.
Principle
The test consists of subjecting the specimens to impacts from a small hemispherical hammer head on any exposed surface of the specimen.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-75:1997 for test Ehb
Impacts shall be applied to all accessible surfaces of the specimen, unless otherwise specified in the product standard
NOTE For certain devices, it may be necessary for the product standard to restrict the surfaces to be impacted
For all such surfaces, three blows shall be applied to any point(s) considered likely to cause damage to or impair the operation of the specimen
To avoid any influence from previous series of three blows, it is essential to ensure that each series is independent If there is uncertainty about the impact of earlier blows, the defect should be ignored, and an additional series of three blows should be conducted on a new specimen at the same location.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 14:
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Equipment class Fixed and movable
Environmental class I, II and III IV
Number of impacts per point 3 3
Similar to IEC 62262 classification IK04 IK06
Measurements during conditioning
Monitor the specimen during the conditioning period to detect any change in status.
Final measurements
After the conditioning, subject the specimen to the final measurements required by the product standard and inspect it visually for mechanical damage both externally and internally
Object of the test
This article aims to showcase the resilience of portable equipment against mechanical impacts that may occur on its surface during typical service conditions, highlighting its ability to endure such challenges effectively.
Principle
The test consist of subjecting the specimens to falls from a specified height onto a surface of concrete or steel.
Test procedure
General
The test apparatus and procedure shall be as described in Procedure 1 of IEC 60068-2-
The specimen in its operating condition for use shall be allowed to fall freely.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 15:
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Environmental class I, II, III and IV I, II, III and IV
The product committee must evaluate the severity of the drop test based on the application and the probability of equipment being dropped For instance, a drop height of 0.5 meters is appropriate for a local transmission unit used in a social alarm system, typically positioned on a tabletop in a residential setting In contrast, a lower severity test or the absence of testing may be acceptable in more controlled environments.
Measurements during conditioning
Monitor the specimen during the conditioning period to detect any change in status.
Final measurements
After the conditioning, subject the specimen to the final measurements required by the product standard and inspect it visually for mechanical damage both externally and internally
Object of the test
To demonstrate the immunity of the equipment to vibration at levels appropriate to the service environment.
Principle
The test involves exposing the specimen to sinusoidal vibrations at levels and frequency ranges that reflect its service environment This includes performing a sweep cycle across the frequency range in both directions—minimum to maximum and back to minimum—while assessing each of the specimen's main functional modes, such as quiescent, alarm, and fault warning conditions, across three mutually perpendicular axes.
Test procedure
General
The test apparatus and procedure shall be generally as described in IEC 60068-2-6:2007
Vibration testing will be conducted sequentially along three mutually perpendicular axes, with one axis positioned perpendicular to the equipment's normal mounting plane.
The vibration operational test can be integrated with the vibration endurance test, allowing the specimen to undergo operational test conditioning followed by endurance test conditioning across each axis.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition(s), as specified in the product standard.
Conditioning
Apply the appropriate severity of conditioning shown in Table 16:
Equipment class Fixed, movable and portable
Environmental class I II, III and IV
Frequency range (Hz) 10 to 150 10 to 150
Number of sweep cycles / axis / functional mode 1 1
Measurements during conditioning
Monitor the specimen during the conditioning period to detect any change in status.
Final measurements
After conditioning the specimen in all three axes, conduct the final measurements as specified by the product standard and perform a visual inspection for any mechanical damage, both externally and internally.
Object of the test
To demonstrate the ability of the equipment to withstand the long-term effects of vibration at levels appropriate to the environment.
Principle
The test involves exposing the specimen to sinusoidal vibrations across a frequency range that simulates the service environment, but at an elevated level to expedite the effects of these vibrations.
Test procedure
General
The test apparatus and procedure shall be generally as described in IEC 60068-2-6:2007, for vibration endurance by sweeping
The vibration shall be applied in each of three mutually perpendicular axes in turn One of these axes shall be perpendicular to the normal mounting plane of the equipment
The vibration endurance test can be integrated with the vibration operational test, allowing the specimen to undergo operational test conditioning followed by endurance test conditioning sequentially along each axis.
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen as specified in the product standard The specimen shall not be supplied with power during the conditioning.
Conditioning
Apply the appropriate severity of conditioning shown in Table 17:
Equipment class Fixed, movable and portable
Environmental class I II, III and IV
Frequency range (Hz) 10 to 150 10 to 150
Number of sweep cycles/axis 20 20
Measurements during conditioning
Final measurements
After conditioning the specimen in all three axes, conduct the final measurements as specified by the product standard and perform a visual inspection for any mechanical damage, both externally and internally.
24 Simulated solar radiation, temperature rise (operational)
This test serves as an alternative to the dry heat (operational) test for group IV While the dry heat test is typically sufficient, it may not accurately simulate the effects of solar heat radiation, particularly in cases where the test specimen is equipped with a solar shield In such instances, the simulated solar radiation test can be utilized.
Object of the test
To demonstrate the ability of the equipment to function correctly when exposed to the thermal effects of solar radiation under the conditions experienced at the surface of the earth.
Principle
The test involves subjecting the specimen to an irradiance of 1,120 W m⁻² while gradually varying the ambient temperature in a diurnal cycle Since the focus is solely on the thermal effects of solar radiation, any spectral distribution of the radiation source may be utilized, provided that adjustments are made for the absorptance factor of the specimen.
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-5:1975, for procedure
A and where only the thermal effects of solar radiation are of interest.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen and place it in the configuration and operating condition, as specified in the product standard.
Conditioning
The temperature-radiation-time relationship during conditioning shall be in accordance with
Figure 1 of IEC 60068-2-5:1975 Two 24 h cycles shall be performed, with an upper temperature of 40 °C at free field conditions
Apply the appropriate severity of conditioning shown in Table 18:
Equipment class Fixed, movable and portable
Environmental class I, II and III IV
Measurements during conditioning
During the conditioning period, it is essential to monitor the specimen for any changes in status Additionally, any further measurements mandated by the product standard should be conducted in the final 30 minutes of the last irradiation period.
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
25 Simulated solar radiation, surface degradation (endurance)
NOTE This test should only be selected where it is necessary to assess specific materials or components for which degradation due to solar radiation is considered to be critical.
Object of the test
To demonstrate the ability of the equipment surface to withstand the surface degradation effects of solar radiation under the conditions experienced at the surface of the earth.
Principle
The test involves subjecting the specimen to an irradiance of 1120 W m\(^{-2}\) using a Xenon radiation source Either the entire test specimen or samples from its surface can be utilized for the testing process.
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Test procedure
General
The test apparatus and procedure shall be as described in IEC 60068-2-5:1975, for procedure C.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen as specified in the product standard The specimen shall not be supplied with power during the conditioning.
Conditioning
The radiation source is aligned to produce a direction of incident radiation of 90°, and adjusted to produce an irradiation of 1 120 W m –2 perpendicular to the test specimen
The temperature-radiation-time relationship during conditioning shall be in accordance with
Apply the appropriate severity of conditioning shown in Table 19:
Equipment class Fixed, movable and portable
Environmental class I, II and III IV
Measurements during conditioning
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
NOTE This test should only be selected where it is necessary to assess specific enclosures for which the ingress of dust is considered to be critical.
Object of the test
To demonstrate that specific enclosures are adequately protected against the ingress of fine dust The test is not suitable for simulation of natural or induced environments.
Principle
The test involves subjecting a specimen to an air flow rich in non-abrasive powder of a defined particle size Specimens from a specific category are evaluated under internal air pressure that is lower than the surrounding atmospheric pressure to facilitate the testing process.
This document is licensed to Mecon Limited for internal use in Ranchi and Bangalore, supplied by Book Supply Bureau The specified quantity of talcum powder, with a maximum grain size of 75 μm, ensures a high and uniform dust density while preventing the ingress of powder.
Test procedure
General
The test apparatus and procedure shall be as described in Table II of IEC 60529:1989 (and
Amendment 1:1999), for the dust test for the first characteristic numerals 5 and 6.
Initial measurements
Before conditioning, subject the specimen to the initial measurements required by the product standard.
State of specimen during conditioning
Mount the specimen according to the product standard, ensuring it is typically in a non-operating condition However, the relevant specification may require the specimen to be powered on and/or operated during testing.
Make sure that seals and other means for dust protection are situated as specified by the manufacturer It shall be determined whether the specimen belongs to category 1 or category 2
Category 1: Air pressure drop in the specimen below the surrounding air caused by intermittent operation or change of temperature of the surrounding air will occur
Category 2: Air pressure drop in the specimen below the surrounding air will not occur.
Conditioning
Apply the appropriate severity of conditioning shown in Table 20 or 21:
Category 1: The pressure inside the specimen is maintained below the surrounding air by a vacuum pump
Equipment class Fixed, movable and portable
Environmental class (I, II, III and IV) a
The IEC 60529 classification, specifically IP5X or IP6X, is applicable to various environmental classes, but it is primarily utilized to evaluate enclosures where dust ingress is deemed critical The acceptance criteria for this assessment are determined by the product committee.
Category 2: The test specimen is not connected to a vacuum pump
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Equipment class Fixed, movable and portable
Environmental class (I, II, III and IV) a
The IEC 60529 classification, specifically IP5X or IP6X, is applicable to various environmental classes, but it is primarily utilized to evaluate enclosures where dust ingress is deemed critical The selection of this test is contingent upon the acceptance criteria established by the product committee.
Measurements during conditioning
Final measurements
After a recovery period of at least 1 h at standard laboratory conditions, subject the specimen to the final measurements required by the product standard
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
IEC 60068-3 (all parts), Environmental testing – Part 3: Background information
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
7 Renseignements à inclure dans la norme de produit appropriée 49
8.3.3 Etat de l'échantillon pendant l'épreuve 51
9.3.3 Etat de l'échantillon pendant l'épreuve 52
10.3.3 Etat de l'échantillon pendant l'épreuve 53
11 Variations de température (essai fonctionnel) 53
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
11.3.3 Etat de l'échantillon pendant l'épreuve 54
12 Essai continu en chaleur humide (essai fonctionnel) 54
12.3.3 Etat de l'échantillon pendant l'épreuve 55
13 Essai continu de chaleur humide (essai d'endurance) 55
13.3.3 Etat de l'échantillon pendant l'épreuve 56
14 Essai cyclique de chaleur humide (essai fonctionnel) 56
14.3.3 Etat de l'échantillon pendant l'épreuve 57
15 Essai cyclique de chaleur humide (essai d'endurance) 57
15.3.3 Etat de l'échantillon pendant l'épreuve 58
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
16.3.3 Etat de l'échantillon pendant l'épreuve 59
17 Anhydride sulfureux (SO2) (essai d'endurance) 60
17.3.3 Etat de l'échantillon pendant l'épreuve 61
18 Brouillard salin, essai cyclique (essai d'endurance) 61
18.3.3 Etat de l'échantillon pendant l'épreuve 62
19.3.3 Etat de l'échantillon pendant l'épreuve 63
20.3.3 Etat de l'échantillon pendant l'épreuve 65
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
21.3.3 Etat de l'échantillon pendant l'épreuve 66
22.3.3 Etat de l'échantillon pendant l'épreuve 67
23.3.3 Etat de l'échantillon pendant l'épreuve 69
24 Rayonnement solaire simulé, élévation de température (essai fonctionnel) 69
24.3.3 Etat de l'échantillon pendant l'épreuve 70
25 Rayonnement solaire simulé, dégradation de surface (essai d'endurance) 70
25.3.3 Etat de l'échantillon pendant l'épreuve 71
26 Résistance à la poussière (essai d'endurance) 71
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
26.3.3 Etat de l'échantillon pendant l'épreuve 72
Figure 1 – Graphique illustrant l'accélération crête en fonction de la masse de l'échantillon 64
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
SYSTÈMES D'ALARME – Partie 1: Méthodes d’essais d'environnement
1) La Commission Electrotechnique Internationale (CEI) est une organisation mondiale de normalisation composée de l'ensemble des comités électrotechniques nationaux (Comités nationaux de la CEI) La CEI a pour objet de favoriser la coopération internationale pour toutes les questions de normalisation dans les domaines de l'électricité et de l'électronique A cet effet, la CEI – entre autres activités – publie des Normes internationales, des Spécifications techniques, des Rapports techniques, des Spécifications accessibles au public (PAS) et des
The IEC Publications are developed by study committees, which allow participation from any national committee interested in the subject matter International, governmental, and non-governmental organizations also collaborate with the IEC on these projects Additionally, the IEC works closely with the International Organization for Standardization (ISO) under terms established by an agreement between the two organizations.
2) Les décisions ou accords officiels de la CEI concernant les questions techniques représentent, dans la mesure du possible, un accord international sur les sujets étudiés, étant donné que les Comités nationaux de la CEI intéressés sont représentés dans chaque comité d’études
3) Les Publications de la CEI se présentent sous la forme de recommandations internationales et sont agréées comme telles par les Comités nationaux de la CEI Tous les efforts raisonnables sont entrepris afin que la CEI s'assure de l'exactitude du contenu technique de ses publications; la CEI ne peut pas être tenue responsable de l'éventuelle mauvaise utilisation ou interprétation qui en est faite par un quelconque utilisateur final
4) Dans le but d'encourager l'uniformité internationale, les Comités nationaux de la CEI s'engagent, dans toute la mesure possible, à appliquer de faỗon transparente les Publications de la CEI dans leurs publications nationales et régionales Toutes divergences entre toutes Publications de la CEI et toutes publications nationales ou régionales correspondantes doivent être indiquées en termes clairs dans ces dernières
5) La CEI elle-même ne fournit aucune attestation de conformité Des organismes de certification indépendants fournissent des services d'évaluation de conformité et, dans certains secteurs, accèdent aux marques de conformité de la CEI La CEI n'est responsable d'aucun des services effectués par les organismes de certification indépendants
6) Tous les utilisateurs doivent s'assurer qu'ils sont en possession de la dernière édition de cette publication
7) Aucune responsabilité ne doit être imputée à la CEI, à ses administrateurs, employés, auxiliaires ou mandataires, y compris ses experts particuliers et les membres de ses comités d'études et des Comités nationaux de la CEI, pour tout préjudice causé en cas de dommages corporels et matériels, ou de tout autre dommage de quelque nature que ce soit, directe ou indirecte, ou pour supporter les cỏts (y compris les frais de justice) et les dépenses découlant de la publication ou de l'utilisation de cette Publication de la CEI ou de toute autre Publication de la CEI, ou au crédit qui lui est accordé
8) L'attention est attirée sur les références normatives citées dans cette publication L'utilisation de publications référencées est obligatoire pour une application correcte de la présente publication
9) L’attention est attirée sur le fait que certains des éléments de la présente Publication de la CEI peuvent faire l’objet de droits de propriété intellectuelle ou de droits analogues La CEI ne saurait être tenue pour responsable de ne pas avoir identifié de tels droits de propriété et de ne pas avoir signalé leur existence
La Norme internationale CEI 62599-1 a été établie par le comité d'études 79 de la CEI:
Systèmes d’alarme et de sécurité électronique
La présente norme est basée sur l’EN 50130-5 (1995) et ses amendements 1 (1998) et 2
Le texte de cette norme est issu des documents suivants:
FDIS Rapport de vote 79/276/FDIS 79/292/RVD
Le rapport de vote indiqué dans le tableau ci-dessus donne toute information sur le vote ayant abouti à l'approbation de cette norme
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
Cette publication a été rédigée selon les Directives ISO/CEI, Partie 2
Une liste de toutes les parties de la série CEI 62599, présentées sous le titre général
Systèmes d’alarme, peut être consultée sur le site web de la CEI
La série CEI 62599 comprend actuellement la présente Partie 1 qui concerne les méthodes d’essai d’environnement, et la Partie 2, qui couvre les exigences pour l’immunité CEM
The committee has determined that the content of this publication will remain unchanged until the stability date specified on the IEC website at "http://webstore.iec.ch" regarding the relevant publication data On that date, the publication will be updated.
• remplacée par une édition révisée, ou
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
The purpose of environmental testing is to ensure that equipment operates correctly in its intended environment and continues to do so over a reasonable period However, alarm system hardware is installed in a wide range of environments, making it nearly impossible to test every aspect related to the most extreme environmental conditions.
The tests and severity levels outlined in this section of IEC 62599 are designed to provide a series of feasible assessments to determine the equipment's ability to withstand the most likely mechanical failures in its intended environment, typically found in industrial and commercial settings It is important to note that additional precautions may be necessary for specific installations where certain environmental factors may be identified as unusually severe A particular additional severity level has been incorporated into the cold test to address the unique cold conditions encountered in the extreme northern regions of Europe.
These tests are designed to demonstrate defects arising from realistic usage environments However, significant failures often result from gradual changes under actual operating conditions To conduct these tests within reasonable timeframes and at lower costs, it may be necessary to accelerate these changes by intensifying conditions, such as increasing the level of an environmental agent or extending its duration or frequency of application.
Les essais contenus dans la présente norme sont par conséquent divisés en deux types d’essai:
In these tests, the sample is subjected to conditions that mimic its intended operating environment The purpose of these tests is to demonstrate the equipment's ability to withstand and function correctly under normal usage conditions, as well as to show its immunity to specific aspects of that environment Consequently, the sample is operational, its conditions are monitored, and it can be functionally tested during the evaluation process related to these tests.
In these tests, the sample is subjected to more severe conditions than those encountered in normal usage to accelerate the effects of the typical operating environment The purpose of these tests is to demonstrate the materials' ability to withstand the long-term effects of the operating environment Since the test aims to study residual effects rather than immediate effects, the sample is typically not powered or controlled during the testing period.
This standard serves as a foundational document for environmental testing, which can be referenced in specific product standards for alarm system components within this application area To ensure consistency among these standards, working groups developing specific product standards should select the recommended tests and severity levels for the appropriate classes of equipment and environment, unless there are valid technical reasons to deviate.
LICENSED TO MECON LIMITED - RANCHI/BANGALORE, FOR INTERNAL USE AT THIS LOCATION ONLY, SUPPLIED BY BOOK SUPPLY BUREAU.
SYSTÈMES D'ALARME – Partie 1: Méthodes d’essais d'environnement