May 2011 Mai 2011 IEC 61010 1 (3rd edition – 2010) Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1 General requirements CEI 61010 1 (3ème édition – 2[.]
Trang 1May 2011 Mai 2011
IEC 61010-1 (3rd edition – 2010) Safety requirements for electrical equipment for
measurement, control, and laboratory use –
Part 1: General requirements
CEI 61010-1 (3ème édition – 2010) Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire – Partie 1: Exigences générales
C O R R I G E N D U M 1
French corrections follow English text
Table 6 – CLEARANCES and test voltages for secondary circuits derived from MAINS
CIRCUITS of OVERVOLTAGE CATEGORY II up to 300 V
Replace Table 6 with the following table:
Trang 2May 2011 Mai 2011
Table 6 – CLEARANCES and test voltages for secondary circuits derived from MAINS
CIRCUITS of OVERVOLTAGE CATEGORY II up to 300 V
Secondary
W ORKING VOLTAGE
MAINS voltage, line-to-neutral, OVERVOLTAGE CATEGORY II
≤ 150 V a.c r.m.s
> 150 V ≤ 300 V a.c r.m.s
a.c r.m.s
V
d.c or a.c peak
V
C LEARANCE
mm
Test voltage
V a.c r.m.s
C LEARANCE
mm
Test voltage
V a.c r.m.s
16
33
50
100
150
300
600
1 000
1 250
1 600
2 000
2 500
3 200
4 000
5 000
6 300
8 000
10 000
12 500
16 000
20 000
25 000
32 000
40 000
50 000
63 000
22,6 46,7
70
140
210
420
840
1 400
1 750
2 240
2 800
3 500
4 480
5 600
7 000
8 820
11 200
14 000
17 500
22 400
28 000
35 000
44 800
56 000
70 000
88 200
0,10 0,11 0,12 0,13 0,16 0,39 1,01 1,92 2,50 3,39 4,49 6,02 8,37 10,9 14,0 18,2 23,9 30,7 39,6 52,5 67,9 87,9
117
151
196
258
500
510
520
540
580
770
1 070
1 630
1 960
2 390
2 890
3 520
4 390
5 320
6 590
8 270
10 400
12 900
16 100
20 400
25 300
31 600
40 400
50 300
62 800
79 400
0,50 0,52 0,53 0,61 0,69 0,94 1,61 2,52 3,16 4,11 5,30 6,91 9,16 11,6 14,9 19,1 24,7 31,6 40,5 53,5 68,9 89,0
118
153
198
260
840
850
860
900
940
1 040
1 450
1 970
2 280
2 730
3 230
3 850
4 660
5 610
6 960
8 620
10 700
13 300
16 400
20 700
25 600
32 000
40 700
50 800
63 400
80 000 Linear interpolation is allowed
Trang 3May 2011 Mai 2011
Table K.9 – Minimum values for distance or thickness of solid insulation
Replace Table K.9 with the following table:
Table K.9 – Minimum values for distance or thickness of solid insulation
Line-to-neutral voltage Minimum thickness
a
Minimum distance L (see Figure K.2)
a, b
a These values are independent of the OVERVOLTAGE CATEGORY
b These values apply for BASIC INSULATION, SUPPLEMENTARY INSULATION and REINFORCED INSULATION
Trang 4May 2011 Mai 2011
French text :
Tableau 6 – DISTANCES D’ISOLEMENT et tensions d’essai des circuits secondaires dérivés
des CIRCUITS RÉSEAU en CATÉGORIE DE SURTENSION II jusqu’à 300 V
Remplacer le Tableau 6 existant par le suivant:
Trang 5May 2011 Mai 2011
des CIRCUITS RÉSEAU en CATÉGORIE DE SURTENSION II jusqu’à 300 V
T ENSION DE SERVICE
au secondaire
Tension RÉSEAU , phase-neutre, CATÉGORIE DE SURTENSION II
≤ 150 V alternatif efficace alternatif efficace > 150 ≤ 300 V Alternatif
efficace alternatif crête Continu ou D ’ D ISOLEMENT ISTANCE Tension d’essai D ’ D ISOLEMENT ISTANCE Tension d’essai
16
33
50
100
150
300
600
1 000
1 250
1 600
2 000
2 500
3 200
4 000
5 000
6 300
8 000
10 000
12 500
16 000
20 000
25 000
32 000
40 000
50 000
63 000
22,6 46,7
70
140
210
420
840
1 400
1 750
2 240
2 800
3 500
4 480
5 600
7 000
8 820
11 200
14 000
17 500
22 400
28 000
35 000
44 800
56 000
70 000
88 200
0,10 0,11 0,12 0,13 0,16 0,39 1,01 1,92 2,50 3,39 4,49 6,02 8,37 10,9 14,0 18,2 23,9 30,7 39,6 52,5 67,9 87,9
117
151
196
258
500
510
520
540
580
770
1 070
1 630
1 960
2 390
2 890
3 520
4 390
5 320
6 590
8 270
10 400
12 900
16 100
20 400
25 300
31 600
40 400
50 300
62 800
79 400
0,50 0,52 0,53 0,61 0,69 0,94 1,61 2,52 3,16 4,11 5,30 6,91 9,16 11,6 14,9 19,1 24,7 31,6 40,5 53,5 68,9 89,0
118
153
198
260
840
850
860
900
940
1 040
1 450
1 970
2 280
2 730
3 230
3 850
4 660
5 610
6 960
8 620
10 700
13 300
16 400
20 700
25 600
32 000
40 700
50 800
63 400
80 000 L'interpolation linéaire est permise
Trang 6May 2011 Mai 2011
Tableau K.9 – Valeurs minimum de la distance ou de l’épaisseur de l’isolation solide
Remplacer le Tableau K.9 existant par le suivant:
Tableau K.9 – Valeurs minimum de la distance ou de l’épaisseur de l’isolation solide
Tension phase-neutre alternative efficace
ou continue
Epaisseur minimale
a
Distance minimale L (voir la Figure K.2)
a, b
a Ces valeurs ne dépendent pas de la CATÉGORIE DE SURTENSION
b Ces valeurs s’appliquent pour l’ISOLATION PRINCIPALE, l’ISOLATION SUPPLÉMENTAIRE