01285776 PDF BRITISH STANDARD BS EN 60118 2 1996 BS 6083 2 1996 Implementing Amendment No 1, not published separately IEC 118 2 1983 (including Amendments 1 1993 and 2 1997 to IEC 118 2 1983) Hearing[.]
Trang 1Amendment No 1, not published separately
IEC 118-2: 1983
(including Amendments 1:1993 and 2:1997 to IEC 118-2:1983)
The European Standard EN 60118-2:1995 with the inclusion of
amendment A2:1997 has the status of a British Standard
ICS 17.140.50
Trang 2BS EN 60118-2:1996
This British Standard, having
been prepared under the
direction of the Electrotechnical
Sector Board, was published
under the authority of the
Standards Board and
comes into effect on
15 September 1996
© BSI 06-1999
The following BSI references
relate to the work on this
standard:
Committee reference EPL/29
Draft for comment 84/20636 DC
British Association of OtolaryngologistsBritish Hearing Aid Industry AssociationBritish Medical Association
British Society of AudiologyBritish Telecommunications plcConfederation of British IndustryDepartment of Health
Department of Trade and Industry (National Physical Laboratory)Health and Safety Executive
Institute of AcousticsInstitute of Sound and Vibration ResearchInstitution of Electrical Engineers
Medical Research CouncilMinistry of DefenceRoyal Aeronautical SocietyRoyal National Institute for Deaf peopleSociety of Environmental EngineersUniversity of Exeter
Amendments issued since publication
Amd No Date Comments
9753 January
1998 Indicated by a sideline in the margin
Trang 3PageCommittees responsible Inside front cover
Trang 4BS EN 60118-2:1996
National foreword
This British Standard has been prepared by Technical Committee EPL/29 and is
the English language version of EN 60118-2:1996 Hearing aids Part 2: Hearing
aids with automatic gain control circuits, including amendment A2:1997
published by the European Committee for Electrotechnical Standardization (CENELEC) It is identical with IEC 118-2, second edition 1983, together with its Amendments 1:1993 and 2:1997, and supersedes BS 6083-2:1984, which is withdrawn
From 1 January 1997, all IEC publications have the number 60000 added to the old number For instance, IEC 27-1 has been renumbered as IEC 60027-1 For a period of time during the change over from one numbering system to the other, publications may contain identifiers from both systems
A British Standard does not purport to include all the necessary provisions of a contract Users of British Standards are responsible for their correct application
Compliance with a British Standard does not of itself confer immunity from legal obligations.
Cross-references
Publication referred to Corresponding British Standard
EN 60118-0:1993(IEC 118-0:1983) BS EN 60118 Hearing aids BS EN 60118-0:1993 Measurement of electroacoustical
IEC 263:1982 BS 6397:1983 Specification for scales and sizes for
plotting frequency characteristics and polar diagrams
IEC 268-8:1973 BS 6840 Sound system equipment
Part 8:1988 Methods for specifying and measuring the
characteristics of automatic gain control devices
a Superseded by EN 60118-1:1995 (IEC 118-1:1995), for which the corresponding British Standard
Trang 5UDC 534.773.2:621.395.92:621.395.665:620.1:621.317.6 Supersedes HD 450.2 S1:1984 Descriptors: Electromedical device, hearing aid, automatic gain control, definitions, measurement procedures
English version Hearing aids Part 2: Hearing aids with automatic gain control circuits
(includes amendment A2:1997) (IEC 118-2:1983 + A1:1993 + A2:1997)
Appareils de correction auditive
Partie 2: Appareils de correction auditive
comportant des commandes automatiques de
gain (inclut l’amendement A2:1997)
(CEI 118-2:1983 + A1:1993 + A2:1997)
Hörgeräte Teil 2: Hörgeräte mit automatischer Verstärkungsregelung
(Enthält Änderung A2:1997) (IEC 118-2:1983 + A1:1993 + A2:1997)
This European Standard was approved by CENELEC on 1994-03-08;
amendment A2 was approved by CENELEC on 1997-07-01 CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of
a national standard without any alteration
Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Central Secretariat or to any
CENELEC member
This European Standard exists in three official versions (English, French,
German) A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the
Central Secretariat has the same status as the official versions
CENELEC members are the national electrotechnical committees of Austria,
Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy,
Luxembourg, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and
United Kingdom
CENELEC
European Committee for Electrotechnical StandardizationComité Européen de Normalisation ElectrotechniqueEuropäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B-1050 Brussels
Trang 6EN 60118-2:1995
© BSI 06-1999
2
Foreword
The text of the International Standard
IEC 118-2:1983, prepared by IEC TC 29,
Electroacoustics, was approved by CENELEC
as HD 450.2 S1 on 1984-09-11
This Harmonization Document was submitted to
the formal vote for conversion into a European
Standard and was approved by CENELEC
as EN 60118-2 on 1994-03-08
The text of amendment 1:1993 to the
International Standard IEC 118-2:1983 was
submitted to the formal vote and was approved
by CENELEC as amendment A1 to EN 60118-2
on 1994-03-08 without any modification
Having first withheld the publication
of EN 60118-2 and its A1, the Technical Board
of CENELEC has allowed on 1995-09-20 the
circulation of the definitive version
of EN 60118-2 with incorporation of its
amendment A1
The following dates were fixed:
Annexes designated “normative” are part of the
body of the standard In this standard, Annex ZA is
normative Annex ZA has been added by
The following dates were fixed:
Annexes designated “normative” are part of the body of the standard Annexes designated
“informative” are given for information only In this standard, Annex ZA is normative and
Annex B and Annex C are informative Annex ZA has been added by CENELEC
— latest date by which the
have to be withdrawn (dow) 1996-07-01
— latest date by which the amendment has to be implemented at national level by publication of an identical national
standard or by endorsement (dop) 1998-04-01
— latest date by which the national standards conflicting with the amendment have to be withdrawn (dow) 1998-04-01
Trang 74.1 Automatic gain control (AGC) 5
4.2 Steady-state input/output graph 5
4.3 Lower AGC limit or AGC threshold 5
4.4 Compression ratio (between specified
input sound pressure level values) 5
4.5 Dynamic output characteristics 5
4.8 AGC activated frequency response 6
4.9 Operating frequency range of the AGC 6
4.10 Overall root-mean-square sound
pressure level (overall r.m.s SPL) 6
4.11 One-third-octave band level 6
4.12 Auto-spectrum (power spectrum) 6
4.13 Cross-spectrum (GAB) 6
4.15 Synchronous analysis 6
5 Steady-state input/output graph 6
5.1 Graph showing the relation between
input sound pressure level and output
sound pressure level 6
8 Effect on steady-state and dynamic
performance with respect to variation
in battery or supply voltage 8
9 AGC activated frequency response of
hearing aids with a single channel
AGC circuit in operation using pure
10 Frequency response of hearing aids
with AGC circuits in operation using
steady-state broad-band input signals 9
PageAnnex A (informative) Examples of
employing digitally controlled input SPL and a band rejection filter 15Figure 5 — Nominal values with upper
and lower limits (± 2 dB) for one-third-octave band levels of the noise input signal at the
Table 1 — Nominal values for the one-third-octave band levels of the noiseinput signal at the test point 11
Trang 84 blank
Trang 91 Scope
1.1 This standard applies to the hearing aids of any
type with automatic gain control (AGC) circuits
This standard gives uniform methods for specifying
dynamic and static performance characteristics of
hearing aids with AGC circuits together with the
relevant methods of measurement for these
characteristics
This standard is confined to a description of the
different characteristics and the relevant methods
of measurement It does not attempt to specify
performance requirements
1.2 This standard includes devices which have
compression and/or limiting properties with respect
to the envelope of the input signal Devices which
control the long-term average output level are also
included
a) AGC is employed to obtain compression, or the
reduction of the dynamic range of the sound at
the output, with the object of preserving the
integrity of the input waveform
b) AGC circuits instead of clipping devices are
often used for limiting purposes
A limiting effect occurs when the input/output
characteristic flattens out at higher input levels
Limiting action is mainly used as a means of
preventing excessive output sound from the hearing
aid from reaching the listener’s ear
1.3 This standard does not include:
a) Expanders
b) Clipping devices, which cut off the signal peaks
above a certain level; such devices differ basically
from AGC circuits, which, in a steady state, tend
to preserve the waveform of the input signal
NOTE An AGC circuit with very short recovery time may
cause considerable distortion, especially in the low-frequency
range This should be given special attention.
2 Object
2.1 The purpose of this standard is to facilitate
measurements of certain characteristics of hearing
aids with AGC circuits that are not described
elsewhere in IEC Publication 118-0: Hearing Aids,
Part 0: Measurement of Electroacoustical
Characteristics, and which are considered necessary
for a physical description of the function of the
automatic gain control
2.2 In general, the methods of measurement
recommended are those which are considered to be
the most directly related to the characteristics This
does not exclude the use of other stated methods
which will give equivalent results
3 Conditions
3.1 General conditions
Reference is made to IEC Publication 268-8: Sound System Equipment, Part 8: Automatic Gain Control Devices
Measurements other than those described herein and that are stated in IEC Publication 118-0 can be performed in accordance with that publication, but with AGC operating, provided the operating conditions are stated
3.2 Throughout this standard, all sound pressure levels are referred to 20 4Pa
4 Explanation of terms
4.1 Automatic gain control (AGC)
A means in a hearing aid by which the gain is automatically controlled as a function of the magnitude of the envelope of the input signal or other signal parameter
NOTE Throughout this standard, reference is made to the use
of acoustic inputs However, where appropriate, additional measurements may be made with an electromagnetically induced input.
4.2 Steady-state input/output graph
The graph illustrating the output sound pressure level as a function of the input sound pressure level for a specified frequency, both expressed in decibels
on identical linear scales (Figure 1, page 12)
4.3 Lower AGC limit or AGC threshold
The input sound pressure level which, when applied to the hearing aid, gives a reduction in the gain of 2 ± 0.5 dB with respect to the gain in the linear mode (Figure 1)
4.4 Compression ratio (between specified input sound pressure level values)
Under steady-state conditions, the ratio of an input sound pressure level difference to the corresponding output sound pressure level difference, both expressed in decibels (Figure 1)
4.5 Dynamic output characteristics
The output sound pressure envelope shown as a function of time when an input sound signal of a predetermined frequency and level is modulated by
a square envelope pulse with a predetermined pulse amplitude (Figure 2, page 13)
Trang 10EN 60118-2:1995
4.6 Attack time
The time interval between the moment when the
input signal level is increased abruptly by a stated
number of decibels and the moment when the
output sound pressure level from the hearing aid
with the AGC circuit stabilizes at the elevated
steady-state level within ± 2 dB (Figure 2, page 13)
4.6.1 Attack time for the normal dynamic
range of speech
The attack time, as defined in Sub-clause 4.6,
when the initial input sound pressure level
is 55 dB and the increase in input sound pressure
level is 25 dB
4.6.2 High level attack time
The attack time, as defined in Sub-clause 4.6,
when the initial input sound pressure level
is 60 dB and the increase in input sound pressure
level is 40 dB
4.7 Recovery time
The time interval between the moment when the
stated input signal level is reduced abruptly to a
level a stated number of decibels lower after the
AGC amplifier has reached the steady-state
output under elevated input signal conditions,
and the moment when the output sound pressure
level from the hearing aid stabilizes again at the
lower steady-state level within ± 2 dB (Figure 2)
4.7.1 Recovery time for the normal dynamic
range of speech
The recovery time, as defined in Sub-clause 4.7,
when the initial input sound pressure level
is 80 dB and the decrease in input sound pressure
level is 25 dB
4.7.2 High level recovery time
The recovery time, as defined in Sub-clause 4.7,
when the initial sound pressure level is 100 dB and
the decrease in input sound pressure level is 40 dB
4.8 AGC activated frequency response
The frequency response when the AGC circuit is
activated by a specified AGC activating signal
4.9 Operating frequency range of the AGC
For a specified input sound pressure level, the
frequency range in which the AGC threshold is
reached or exceeded
4.10 Overall root-mean-square sound pressure
level (overall r.m.s SPL)
The root-mean-square sound pressure level with
measurement bandwidth equal to the frequency
range covered by the one-third-octave frequency
bands (see IEC 61260) from 200 Hz to 8 000 Hz
4.11 One-third-octave band level
The level of that part of the signal contained within
a band one-third-octave wide as defined
in IEC 61260
4.12 Auto-spectrum (power spectrum)
The power spectrum of either the input signal (GAA)
to or the output signal (GBB) from a hearing aid in the frequency domain It is computed by multiplying the Fourier transform of the signal by the complex conjugate of the Fourier transform of the same signal
4.13 Cross-spectrum (GAB)The degree to which the same signal frequencies are mutually present in the input and output of a hearing aid It is computed by multiplying the complex conjugate of the Fourier transform of the input signal to the hearing aid by the Fourier transform of the output signal from the hearing aid
4.14 Coherence
A number ranging from 0 to 1 showing to what degree the output from a hearing aid is correlated to the input Coherence for a random noise test signal
is reduced by non-linearity and by system noise The coherence is calculated from the auto- and
cross-spectrum averages as follows:
Coherence =
4.15 Synchronous analysis
Analysis which is synchronized with the period of the input signal, for example with the periodicity of pseudo-random noise
5 Steady-state input/output graph
5.1 Graph showing the relation between input sound pressure level and output sound pressure level
The graph shall have the input sound pressure level
as abscissa and the output sound pressure level as ordinate, both expressed in decibels on linear scales having divisions of identical size
NOTE In the input/output graph of an AGC device, different portions may be distinguished:
— Below the lower AGC limit the slope is essentially 45° (linear amplifier mode).
— Above this limit, the graph curves over in a portion having
a decreasing slope, often followed by another portion having a nearly flat slope (AGC mode).
— At very high input levels, a flat or sloping portion may be followed by a portion with a steeper slope, generally due to saturation of the AGC circuit.
Trang 115.2 Methods of measurement
The gain control is adjusted to its maximum setting
Any adjustable gain control after the AGC-loop
shall be adjusted in such a manner that overload of
the hearing aid is avoided
An input sound signal of frequency 1 600 Hz
or 2 500 Hz when appropriate, is applied at the
lowest possible level consistent with an adequate
signal-to-noise ratio of preferably more than 10 dB
The input sound pressure level is increased up
to 100 dB in sufficiently small steps, and the
corresponding output sound pressure level is
measured after steady-state conditions have been
reached The graph is plotted with the input sound
pressure level as abscissa and the output level as
ordinate, as described in Sub-clause 5.1.
Where separate adjustable controls exist, such
as AGC, gain or output controls, which will
influence the shape and other characteristics of the
steady-state input/output graph, it is recommended
that input/output graphs be plotted, when useful,
for various additional stated setting of such
controls
6 Dynamic output characteristics
6.1 Characteristics to be measured
The purpose of this test is to determine the dynamic
characteristics of the AGC circuit, particularly
attack and recovery times It should be emphasized
that all these characteristics will depend on test
frequency as well as on such factors as signal level,
control settings and battery voltage
6.2 Methods of measurement
6.2.1 Dynamic output characteristics for
speech levels
At the maximum setting of the gain control an input
signal of 1 600 Hz or 2 500 Hz when appropriate
with a sound pressure level of 55 dB is applied Any
adjustable gain control after the AGC loop shall be
adjusted in such a manner that overload of the
hearing aid is avoided
This signal is modulated by a square envelope pulse
raising the input level by 25 dB The pulse length
shall be at least five times longer than the attack
time being measured If more than a single pulse is
applied, the interval between two pulses shall be at
least five times the longest recovery time being
measured
NOTE 1 This test may be carried out at various control settings
as stated in Sub-clause 5.2.
NOTE 2 If lower gain control settings are applied, the method
of obtaining these settings shall be clearly specified.
NOTE 3 The loudspeaker employed for the measurement of
dynamic output characteristics as in Clause 6 must be
NOTE 4 The output signal should be monitored on a device such as an oscilloscope, the time constants of which are considerably shorter than those being measured.
NOTE 5 When very short response times are to be measured, the response time of the source shall be reported.
NOTE 6 For half-wave rectifying AGC circuits, the attack time
is dependent upon the polarity of the first half wave of the test signal after the onset of the modulating square wave envelope Depending upon the polarity, a shorter or longer attack time will occur This is best demonstrated in the case of an instantaneous rise in the amplitude occurring at a zero crossing of the test signal.
6.2.2 Dynamic output characteristics for high level input
At the maximum setting of the gain control an input signal of 1 600 Hz or 2 500 Hz when appropriate with a sound pressure level of 60 dB is applied Any adjustable gain control after the AGC loop shall be adjusted in such a manner that overload of the hearing aid is avoided This signal is modulated by
a square envelope pulse raising the input level
by 40 dB The pulse length shall be at least five times the attack time observed
If more than a single pulse is applied, the interval between two pulses should be at least five times the longest recovery time being measured
NOTE See Notes 1 to 6 of Sub-clause 6.2.1.
7 Non-linear distortion
7.1 Transients
The signal may be distorted during the attack time and recovery time by transients as well as by unwanted low-frequency modulation caused by instabilities The effect of these phenomena on the listener is not sufficiently understood to allow a recommendation for measuring transient distortions to be made
7.2 Harmonic distortion 7.2.1 Characteristics to be specified
The purpose of this test is to determine the harmonic distortion as a function of the input sound pressure level after steady-state conditions have been reached
7.2.2 Methods of measurement
Harmonic distorsion is measured in accordance with the test procedure described in
IEC Publication 118-0
NOTE This test may be carried out at various control settings
as mentioned under Sub-clause 5.2.
7.3 Intermodulation distortion
Intermodulation distortion is measured in accordance with the test procedure, described
in IEC Publication 118-0
Trang 12EN 60118-2:1995
8 Effect on steady-state and dynamic
performance with respect to variation
in battery or supply voltage
In accordance with IEC Publication 118-0,
Sub-clause 7.8, it is recommended that the change
in the following performance parameters be tested
with respect to variation in battery or supply
voltage: steady-state input/output graphs as
mentioned in Clause 5, dynamic-output
characteristics, attack and recovery times as
mentioned under Clause 6, and non-linear
distortion as mentioned in Clause 7.
9 AGC activated frequency response
of hearing aids with a single
channel AGC circuit in operation
using pure tone signals
9.1 Introduction
9.1.1 Measurement below the AGC threshold
It is important to know the frequency response of
hearing aids when the automatic gain control is
active Measurements carried out in accordance
with 7.4 of IEC 118-0 can be employed, provided
that the input signal levels at all measuring
frequencies are below the AGC thresholds
9.1.2 Measurement above the AGC threshold
When measurements are taken above the AGC
threshold with moderately slow scanning speeds
and the attack and decay times in current use, they
can be considered at any frequency as steady-state
measurements
From the set of comprehensive frequency response
curves obtained using this method of measurement,
a steady-state input/output graph can be
constructed for any frequency
9.2 General conditions
Throughout this standard all sound pressure levels
specified refer to 20 4Pa When appropriate, sound
pressure level will be abbreviated SPL
NOTE Throughout the standard, reference is made to the use of
acoustic input However, when appropriate, additional
measurements may be made with electromagnetic or electric
inputs in accordance with IEC 118-1 and IEC 118-6.
Test results obtained by the substitution method
(see 4.2 of IEC 118-0) shall be considered basic.
9.3 Test equipment
The test equipment shall comply with clause 5
of IEC 118-0 Figure 3 and Figure 4 give schematic
illustrations of the measuring equipment
A tracking bandpass filter, centred at a centre
frequency fc, shall be inserted in the measuring system The 3 dB bandwidth shall not be greater than 10 % of the centre frequency, and the 20 dB bandwidth shall not be greater than 20 % of the
centre frequency For frequencies higher that 4 fc or
less than fc/4,fc being the centre frequency, the attenuation shall be greater than 40 dB For
frequencies higher than 8 fc or lower than fc/8, the attenuation shall be equal to or greater than 60 dB
NOTE In certain cases it may be necessary to use a filter with
an attenuation higher than 40 dB or 60 dB, respectively, in order not to increase the frequency range relative to the activating frequency where no reliable results can be obtained.
When automatic swept frequency recording using a compressor system to maintain a constant input sound pressure level is employed, a second tracking bandpass filter complying with the specifications given above shall be inserted in the feed-back loop
NOTE When synchronizing the tracking bandpass filter to the swept frequency, consideration should be given to the influence of the transmission time from the sound source to test point, the bandwidth of the filter and the scanning speed.
Alternatively, a band rejecting filter tuned to the AGC activating signal frequency and having an attenuation equal to or greater than 50 dB at the
centre frequency (fc) may be employed For
frequencies greater than 1,05 fc or less than fc/1,05, the attenuation shall be less than 3 dB
9.4 Test conditions
The AGC control should be set for maximum AGC effect (i.e lowest AGC threshold) Otherwise the test conditions shall comply with the specifications
stated in clause 6 of IEC 118-0, as applicable, The
settings shall be stated
9.5 Measurement
Data should be quoted for that part of the frequency range between 200 Hz and 8 000 Hz over which the output of the hearing aid falls by at least 10 dB when the signal sources are switched off
9.5.1 Determination of the operating range of the AGC
Test procedure
a) Adjust the gain control to full on and set other
controls in accordance with 9.4 of this standard.
b) Apply a sinusoidal input signal and vary its frequency, keeping the input SPL constant
at 50 dB, 60 dB, 70 dB, 80 dB and 90 dB or until the AGC threshold has been exceeded
c) Plot the output SPL versus frequency at a constant input SPL for each of the input sound pressure levels