1. Trang chủ
  2. » Kỹ Thuật - Công Nghệ

Bsi bs en 14154 3 2005 + a2 2011

90 1 0

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Water Meters — Part 3: Test Methods And Equipment
Trường học British Standards Institution
Chuyên ngành Standards
Thể loại Standard
Năm xuất bản 2005
Thành phố Brussels
Định dạng
Số trang 90
Dung lượng 1,4 MB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Cấu trúc

  • 5.1 General requirements for the test installation (10)
    • 5.1.1 Location (10)
    • 5.1.2 Test water (11)
    • 5.1.3 Calibrated reference device (11)
    • 5.1.4 Freedom from spurious influences (11)
    • 5.1.5 Group testing of meters (12)
  • 5.2 Static pressure test (12)
    • 5.2.1 Object of test (12)
    • 5.2.2 Preparation (12)
    • 5.2.3 Test procedure (12)
    • 5.2.4 Acceptance criteria (13)
  • 5.3 Determination of intrinsic errors (of indication) (13)
    • 5.3.1 Object of test (13)
    • 5.3.2 Preparation (13)
    • 5.3.3 Test procedure (14)
    • 5.3.4 Acceptance criteria (14)
  • 5.4 Error (of indication) tests (15)
    • 5.4.1 Object of test (15)
    • 5.4.2 Preparation (15)
    • 5.4.3 Test Procedure (15)
    • 5.4.4 Acceptance criteria (15)
  • 5.5 Absence of flow test (16)
    • 5.5.1 Object of test (16)
    • 5.5.2 Preparation (16)
    • 5.5.3 Test procedure (16)
    • 5.5.4 Acceptance criteria (16)
  • 5.6 Water temperature test (within ROC) (16)
    • 5.6.1 Object of test (16)
    • 5.6.2 Preparation (16)
    • 5.6.3 Test procedure (16)
    • 5.6.4 Acceptance criteria (17)
  • 5.7 Overload water temperature test (17)
    • 5.7.1 Object of test (17)
    • 5.7.2 Preparation (17)
    • 5.7.3 Test procedure (17)
    • 5.7.4 Acceptance criteria (17)
  • 5.8 Water pressure test (17)
    • 5.8.1 Object of test (17)
    • 5.8.2 Preparation (18)
    • 5.8.3 Test procedure (18)
    • 5.8.4 Acceptance criteria (18)
  • 5.9 Verification of flow profile sensitivity classes (18)
    • 5.9.1 Object of test (18)
    • 5.9.3 Test procedure (18)
    • 5.9.4 Acceptance criteria (19)
  • 5.10 Tests on ancillary devices of a water meter (21)
    • 5.10.1 Object of test (21)
    • 5.10.2 Preparation (21)
    • 5.10.3 Procedure (21)
    • 5.10.4 Acceptance criteria (21)
  • 5.11 Pressure loss test (21)
    • 5.11.1 Object of test (21)
    • 5.11.2 Preparation (21)
    • 5.11.3 Test procedure (22)
    • 5.11.4 Acceptance criteria (22)
  • 5.12 Reverse flow test (22)
    • 5.12.1 Meters designed to measure reverse flows (22)
    • 5.12.2 Meters not designed to measure reverse flows (22)
  • 5.13 Endurance tests (23)
    • 5.13.1 Continuous flow test (23)
    • 5.13.2 Discontinuous flow test (26)
  • 6.1 General requirements (29)
    • 6.1.1 Environmental classification (29)
    • 6.1.2 Electromagnetic environments (29)
    • 6.1.3 Reference conditions (29)
    • 6.1.4 Test volumes for measuring error (of indication) of a water meter (30)
    • 6.1.5 Influence of the water temperature (30)
    • 6.1.6 Requirements for environmental tests (30)
    • 6.1.7 Equipment under test (EUT) (30)
  • 6.2 Climatic and mechanical environment (32)
    • 6.2.1 Dry heat (non-condensing) (32)
    • 6.2.2 Cold (33)
    • 6.2.3 Damp heat, cyclic (condensing) (34)
    • 6.2.4 Vibration (random) (36)
    • 6.2.5 Mechanical shock (37)
  • 6.3 Electromagnetic environment (38)
    • 6.3.1 Electrostatic discharge (38)
    • 6.3.2 Radiated radio frequency/Electromagnetic fields (39)
  • 6.4 Power supply (41)
    • 6.4.1 A.C. power voltage variation (41)
    • 6.4.2 A.C. voltage dips and short interruptions (42)
    • 6.4.3 Surge immunity (44)
    • 6.4.4 Electrical fast transients/Burst (45)
    • 6.4.5 D.C. power voltage variation (46)
  • 6.5 Static magnetic field (48)
    • 6.5.1 Object of test (48)
    • 6.5.2 Preparation (48)
    • 6.5.3 Test procedure (48)
    • 6.5.4 Acceptance criteria (48)
  • 6.6 Interruption in battery supply (49)
    • 6.6.1 Object of test (49)
    • 6.6.2 Test procedure (49)
    • 6.6.3 Acceptance criteria (49)
  • A.1 Irregularity in water velocity field (50)
  • A.2 Flow disturbers (50)
  • C.1 Principle (65)
  • C.2 Description of the test rig (65)
  • C.3 Pipe-work (65)
    • C.3.1 Description (65)
    • C.3.2 Test section (66)
    • C.3.3 Precautions to be taken during tests (66)
    • C.3.4 Special arrangements for the installation of meters (66)
    • C.3.5 Cyclic distortion of the meter (68)
    • C.3.6 Major factors affecting the measurement of errors of indication (69)
  • D.1 Principle (70)
  • D.2 Pressure loss test equipment (70)
    • D.2.1 General (70)
    • D.2.2 Measuring section (70)
  • D.3 Test procedure (75)
    • D.3.1 Determination of pressure loss attributable to pipe lengths for water meters (measurement 1)73 (75)
    • D.3.2 Measurement and calculation of the actual ∆∆∆ P of a water meter (measurement 2) (76)
  • E.1 General (77)
  • E.2 Types of reference devices (77)
    • E.2.1 Calibration using volumetric vessels (77)
    • E.2.2 Calibration using reference meters (master meters) (78)
    • E.2.3 Calibration using calibrated tubes (79)
    • E.2.4 Calibration by weighing method (80)
    • E.2.5 Other calibration methods (80)

Nội dung

7 EN 61000-4-11:1994, Electromagnetic compatibility EMC — Part 4-11: Testing and measurement techniques — Voltage dips, short interruptions and voltage variations immunity tests IEC 60

Trang 1

Water meters —

Part 3 : Test methods and equipment

ICS 91.140.60

12&23<,1*:,7+287%6,3(50,66,21(;&(37$63(50,77('%<&23<5,*+7/$:

Trang 2

National foreword

This British Standard is the official English language version of

EN 14154-3:2005+A2:2011 It supersedes BS EN 14154-3:2005 which is withdrawn

The start and finish of text introduced or altered by amendment is indicated in the text by tags Tags indicating changes to CEN text carry the number of the CEN amendment For example, text altered by CEN amendment A1 is

indicated by !"

The UK participation in its preparation was entrusted by Technical Committee CPI/30, Measurement of fluid flow in closed conduits, to Subcommittee

CPI/30/7/1, Water meters

A list of organizations represented on this subcommittee can be obtained on request to its secretary

This publication does not purport to include all the necessary provisions of a contract Users are responsible for its correct application

Compliance with a British Standard cannot confer immunity from legal obligations.

Amendments/corrigenda issued since publication

31 July 2011 Implementation of CEN amendment A2:2011

This British Standard was

published under the authority

of the Standards Policy and

Trang 3

NORME EUROPÉENNE

ICS 91.140.60

English Version

Water meters - Part 3: Test methods and equipment

Compteurs d'eau - Partie 3: Méthodes et équipement

d'essai

Wasserzähler - Teil 3: Prüfverfahren und -einrichtungen

This European Standard was approved by CEN on 26 July 2004 and includes Amendment 1 approved by CEN on 6 March 2007 and Amendment 2 approved by CEN on 22 February 2011

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member

This European Standard exists in three official versions (English, French, German) A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same status as the official versions

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom

EUROPEAN COMMITTEE FOR STANDARDIZATION

C O M I T É E U R O P É E N D E N O R M A L I S A T I O N

E U R O P Ä I S C H E S K O M I T E E FÜ R N O R M U N G

Management Centre: Avenue Marnix 17, B-1000 Brussels

© 2011 CEN All rights of exploitation in any form and by any means reserved

worldwide for CEN national Members

Ref No EN 14154-3:2005+A2:2011: E

Trang 4

Contents

Page

Foreword 5



1



Scope 6



2



Normative references 6



3



Reference conditions 7



4



Administrative and external examination for approval programme 7



5



Performance tests for approval programme 8



5.1



General requirements for the test installation 8



5.1.1



Location 8



5.1.2



Test water 9



5.1.3



Calibrated reference device 9



5.1.4



Freedom from spurious influences .9



5.1.5



Group testing of meters 10



5.2



Static pressure test 10



5.2.1



Object of test 10



5.2.2



Preparation 10



5.2.3



Test procedure 10



5.2.4



Acceptance criteria 11



5.3



Determination of intrinsic errors (of indication) 11



5.3.1



Object of test 11



5.3.2



Preparation 11



5.3.3



Test procedure 12



5.3.4



Acceptance criteria 12



5.4



Error (of indication) tests 13



5.4.1



Object of test 13



5.4.2



Preparation 13



5.4.3



Test Procedure 13



5.4.4



Acceptance criteria 13



5.4.5



#Interchange test on all types of cartridge and meters with exchangeable metrological units $ 13



5.5



Absence of flow test 14



5.5.1



Object of test 14



5.5.2



Preparation 14



5.5.3



Test procedure 14



5.5.4



Acceptance criteria 14



5.6



Water temperature test (within ROC) 14



5.6.1



Object of test 14



5.6.2



Preparation 14



5.6.3



Test procedure 14



5.6.4



Acceptance criteria 15



5.7



Overload water temperature test 15



5.7.1



Object of test 15



5.7.2



Preparation 15



5.7.3



Test procedure 15



5.7.4



Acceptance criteria 15



5.8



Water pressure test 15



5.8.1



Object of test 15



5.8.2



Preparation 16



5.8.3



Test procedure 16



5.8.4



Acceptance criteria 16



5.9



Verification of flow profile sensitivity classes 16



5.9.1



Object of test 16



5.9.2



Preparation 16



Trang 5

3

5.9.3



Test procedure 16



5.9.4



Acceptance criteria 17



5.10



Tests on ancillary devices of a water meter 19



5.10.1



Object of test 19



5.10.2



Preparation 19



5.10.3



Procedure 19



5.10.4



Acceptance criteria 19



5.11



Pressure loss test 19



5.11.1



Object of test 19



5.11.2



Preparation 19



5.11.3



Test procedure 20



5.11.4



Acceptance criteria 20



5.12



Reverse flow test 20



5.12.1



Meters designed to measure reverse flows 20



5.12.2



Meters not designed to measure reverse flows 20



5.13



Endurance tests 21



5.13.1



Continuous flow test 21



5.13.2



Discontinuous flow test 24



6



Performance tests related to influence quantities 27



6.1



General requirements 27



6.1.1



Environmental classification 27



6.1.2



Electromagnetic environments 27



6.1.3



Reference conditions 27



6.1.4



Test volumes for measuring error (of indication) of a water meter 28



6.1.5



Influence of the water temperature 28



6.1.6



Requirements for environmental tests 28



6.1.7



Equipment under test (EUT) 28



6.2



Climatic and mechanical environment 30



6.2.1



Dry heat (non-condensing) 30



6.2.2



Cold 31



6.2.3



Damp heat, cyclic (condensing) 32



6.2.4



Vibration (random) 34



6.2.5



Mechanical shock 35



6.3



Electromagnetic environment 36



6.3.1



Electrostatic discharge 36



6.3.2



Radiated radio frequency/Electromagnetic fields 37



6.4



Power supply 39



6.4.1



A.C power voltage variation 39



6.4.2



A.C voltage dips and short interruptions 40



6.4.3



Surge immunity 42



6.4.4



Electrical fast transients/Burst 43



6.4.5



D.C power voltage variation 44



6.5



Static magnetic field 46



6.5.1



Object of test 46



6.5.2



Preparation 46



6.5.3



Test procedure 46



6.5.4



Acceptance criteria 46



6.6



Interruption in battery supply 47



6.6.1



Object of test 47



6.6.2



Test procedure 47



6.6.3



Acceptance criteria 47



Annex A (normative) Irregularity in water velocity field 48



A.1



Irregularity in water velocity field 48



A.2



Flow disturbers 48



Annex B (informative) Examples of methods and components used for testing Concentric Water Meters 59



Annex C (normative) Equipment and method used to determine meter error 63



C.1



Principle 63



C.2



Description of the test rig 63



Trang 6

C.3



Pipe-work 63



C.3.1



Description 63



C.3.2



Test section 64



C.3.3



Precautions to be taken during tests 64



C.3.4



Special arrangements for the installation of meters 64



C.3.5



Cyclic distortion of the meter 66



C.3.6



Major factors affecting the measurement of errors of indication 67



Annex D (informative) Equipment and methods for pressure loss test 68



D.1



Principle 68



D.2



Pressure loss test equipment 68



D.2.1



General 68



D.2.2



Measuring section 68



D.3



Test procedure 73



D.3.1



Determination of pressure loss attributable to pipe lengths for water meters (measurement 1)73



D.3.2



Measurement and calculation of the actual ∆∆∆P of a water meter (measurement 2) 74



Annex E (informative) Characteristics of reference devices 75



E.1



General 75



E.2



Types of reference devices 75



E.2.1



Calibration using volumetric vessels 75



E.2.2



Calibration using reference meters (master meters) 76



E.2.3



Calibration using calibrated tubes 77



E.2.4



Calibration by weighing method 78



E.2.5



Other calibration methods 78



Annex ZA (informative) !Relationship between this European Standard and the Essential Requirements of EU Directive 22/2004/EC on Measuring Instruments" 79



Bibliography 86



Trang 7

Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights

This document includes Amendment 1, approved by CEN on 2007-03-06 and Amendment 2, approved by CEN on 2011-02-22

This document supersedes #EN 14154-3:2005+A1:2007$

The start and finish of text introduced or altered by amendment is indicated in the text by tags ! " and # $ This European Standard has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s)

For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this document

The standard consists of 3 parts The other parts are:

 Part 1: General Requirements

 Part 2: Installation and conditions of use

In developing a new Standard, CEN/TC 92 aimed to harmonise it with existing standards and recommendations for water meters, to accommodate new technologies and anticipating the requirements of the Directive 22/2004 EC on Measuring Instruments

According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom

Trang 8

1 Scope

This document applies to water meters intended for residential, commercial, light industrial and industrial use, and specifies the test parameters and the test methods for water meters, irrespective of the design technologies, as specified in #EN 14154-1:2005+A2$, used to meter the actual volume of clean cold potable water or heated water, flowing through a fully charged, closed conduit These water meters shall incorporate devices, which indicate the integrated volume

In the case where water meters having a value of Q 3 > 160 m3/h, the test schedule may make provisions for modification of the Reference Conditions, to meet individual test laboratory limitations, when testing specifically for endurance or for performance under Influence Quantities

Meters thus tested shall be marked so as to unambiguously indicate part compliance with this document To augment this marking the meter manufacturer shall, in addition, be obliged to fully disclose the specific non compliance(s) due to the test laboratory limitations

2 Normative references

The following referenced documents are indispensable for the application of this document For dated references, only the edition cited applies For undated references, the latest edition of the referenced document (including any amendments) applies

#EN 14154-1:2005+A2:2011$, Water meters — Part 1: General requirements

#EN 14154-2:2005+A2:2011$, Water meters — Part 2: Installation and conditions of use

EN 60068-1:1994, Environmental testing — Part 1: General and guidance

EN 60068-2-1:1993, Environmental testing — Part 2: Tests, Test A: Cold

EN 60068-2-2:1993, Environmental testing — Part 2: Tests, Test B: Dry heat

EN 60068-2-30:1999, Environmental testing — Part 2: Tests — Test Db and guidance: Damp heat, cyclic

((12 + 12) h cycle)

EN 60068-2-31:1993, Environmental testing — Part 2: Tests — Test Ec: Drop and topple, primarily for

equipment-type specimens

EN 60068-2-47:1999 Environmental testing — Part 2-47: Test methods — Mounting of components, equipment

and other articles for vibration, impact and similar dynamic tests

EN 60068-2-64:1994, Environmental testing — Part 2: Test methods — Test Fh: Vibration, broad-band random

(digital control) and guidance

EN 60068-3-1:1999, Environmental testing — Part 3: Background information, Section 1: Cold and dry heat tests

EN 61000-4-2:1995, Electromagnetic compatibility (EMC) — Part 4-2: Testing and measurement techniques —

Electrostatic discharge immunity test

EN 61000-4-3:1996, Electromagnetic compatibility (EMC) — Part 4-3: Testing and measurement techniques —

Radiated, radio-frequency, electromagnetic field immunity test

EN 61000-4-4:1995, Electromagnetic compatibility (EMC) — Part 4-4: Testing and measurement techniques —

Electrical fast transient/burst immunity test

EN 61000-4-5:1995, Electromagnetic compatibility (EMC) — Part 4-5: Testing and measurement techniques —

Surge immunity test

Trang 9

7

EN 61000-4-11:1994, Electromagnetic compatibility (EMC) — Part 4-11: Testing and measurement techniques —

Voltage dips, short interruptions and voltage variations immunity tests

IEC 60068-3-4:2001, Environmental testing — Part 3-4: Supporting documentation and guidance — Damp heat

tests

ISO 4185, Measurement of fluid flow in closed conduits — Weighing method

OIML D4, Installation and Storage Conditions for Cold Water Meters

OIML D7, The Evaluation of Flow Standards and Facilities Used for Testing Water Meters

OIML P7, Planning of Metrology and Testing Laboratories

Ambient temperature range: 15 °C to 25 °C*

Ambient relative humidity range: 45 % to 75 %*

Ambient atmospheric pressure range: 86 kPa to 106 kPa (0,86 bar to 1,06 bar)

Power supply voltage (mains a.c.): Nominal voltage (U nom) ± 5 %

Power supply frequency: Nominal frequency (f nom) ± 2 %

Power supply voltage (battery): A voltage U in the range; Umin ≤ U ≤ Umax

During each test, the temperature and relative humidity shall not vary by more than 5 °C or 10 % respectively within the reference range

* Here the ambient temperature and/or ambient relative humidity exceed the above-mentioned ranges, the effect

on the error (of indication) shall be taken in account

4 Administrative and external examination for approval programme

The meter and its technical documentation shall be examined before entering the rest of the test program The administrative examination shall demonstrate that the water meter fulfils the respective requirements of Clause 4 of

#EN 14154-1:2005+A2:2011$, which can be verified by administrative work and shall show that the meter is manufactured in conformity with it

Trang 10

The following documents shall be examined, as far as is relevant, for the respective water meter:

 general description of the water meter and its working principle;

 conceptual design, general assembly drawings;

 subassemblies drawings, circuits, etc.;

 metrological characteristics claimed;

 indicated device and dial drawings;

 security sealing plan drawings;

 list of mechanical components and materials;

 list of electronic components and specifications;

 description of the software;

 descriptions and explanations, necessary to understanding the above, including the operation of the instrument;

 results of design calculations examinations etc.;

 test reports

The following aspects of the meter design shall be examined on at least one meter sample:

 meter designation;

 meter size and overall dimensions;

 meter end connections;

 indicating device;

 supplementary devices;

 protection devices: verify that the water meter includes protection devices, which can be sealed to prevent, both before and after correct installation, dismantling or modification of the meter or its adjustment device, without damaging these devices

5 Performance tests for approval programme

5.1 General requirements for the test installation

5.1.1 Location

The environment chosen for the meter tests shall be in accordance with the principles elaborated in OIML P7 document, “Planning of Metrology and Testing Laboratories”, and shall be free from unintended disturbing influences, for example ambient temperature variation and vibration

Annexes C and E give detailed requirements for the test installation and equipment

Trang 11

The water pressure upstream of the meter shall not vary, during the text, by more than 10 %

Pressure at the entrance to the meter shall not exceed the MAP for the meter

The maximum uncertainty in the measurement of pressure (or pressure loss) shall be ± 5 % of the measured value

5.1.2.3 Flow rate

The relative variation in the flowrate during each test (not including starting and stopping) shall not exceed:

 ± 2,5 % from Q1 to Q2 (not inclusive);

 ± 5,0 % from Q2 (inclusive) to Q4

The average flowrate value is the actual volume passed during the test divided by the time

5.1.2.4 Temperature

The temperature of the water during the test shall not change by more than 5 °C

The maximum uncertainty in the measurement of temperature shall not exceed ± 2 °C

5.1.3 Calibrated reference device

5.1.3.1 Overall uncertainty of the actual volume

When a test is conducted, the expanded uncertainty of the actual volume shall not exceed 1/5 of the applicable maximum permissible error for pattern approval, initial verification and subsequent verification

The uncertainty shall be estimated according to the “Guide to the expression of uncertainty in measurement” with a

coverage factor of k = 2

5.1.3.2 Minimum volume of the calibrated reference device

The minimum volume permitted depends on requirements determined by the test starting and stopping effects (timing error), and the type and design of the indicating device (value of the verification scale interval)

5.1.4 Freedom from spurious influences

Test rigs shall be so designed, constructed and used, that the performance of the rig itself shall not contribute significantly to the test error To this end, high standards of rig maintenance and adequate supports and fittings are necessary to prevent vibration of the meter, the test rig and its accessories

Trang 12

The test rig environment shall be such that the reference conditions of the test are met

As part of the validation process, periodic intercomparisons between test rigs shall be carried out in accordance with OIML International Document D 7

5.1.5 Group testing of meters

Meters are tested either individually or in groups In the latter case the individual characteristics of the meters shall

be precisely determined Interaction between meters, and test rigs, shall be eliminated

When meters are tested in series, the pressure at the exit of each meter shall be sufficient to prevent cavitation

5.2 Static pressure test

5.2.1 Object of test

To verify that the water meter can withstand the specified hydraulic test pressure, without leakage or damage, according to its Maximum Admissible Pressure (MAP) class (see 5.1.2 of #EN 14154-1:2005+A2:2011$)

5.2.2 Preparation

a) Install the meters in the test rig either singularly or in batches;

b) bleed the test rig pipe-work and the water meter of air;

c) ensure that the test rig is free from leaks;

d) ensure that the supply pressure is free from pressure pulsations

5.2.3 Test procedure

5.2.3.1 In-line meters

a) Increase the hydraulic pressure to 1,6 × MAP of the meter and hold it for 15 min;

b) examine the meters for physical damage, for external leaks and for leaks into the indicating device;

c) increase the hydraulic pressure to 2 × MAP and hold it for 1 min The flowrate shall be zero during the test; d) examine the meters for physical damage, for external leaks and for leaks into the indicating device

Additional requirements:

e) increase and decrease the pressure gradually without pressure surges;

f) apply only the reference temperature for this test

5.2.3.2 Concentric meters – seal integrity test

The test procedure in 5.2.3 also applies to pressure testing of concentric water meters; however the seals located at the concentric meter/manifold interface, shall also be tested to ensure that undisclosed internal leaks between the inlet and outlet passages of the meter do not occur

When the pressure test is carried out the meter and manifold shall be tested together

A pressure of 2 × ∆P is applied to the meter inlet side of the seal

Trang 13

To verify that the meter complies with the requirements in 7.6 of #EN 14154-1:2005+A2:2011$ and the effects

of the meter orientation on the error (of indication)

5.3.2 Preparation

The method described here for determining the meter errors (of indication) is the so-called “collection” method, in which the quantity of water passed through the water meter is collected in one or more collecting vessels and the quantity determined volumetrically or by weighing, as described in Annex C Other methods may be used, provided the requirements of uncertainty are met

The checking of the errors (of indication) consists in comparing the volume indications given by the meter under reference conditions against a calibrated reference device

For the purpose of these tests, the meter should be tested without its temporary supplementary devices attached (if any) Tests to determine the influence of attaching temporary supplementary devices to the meter are described

in 5.10

5.3.2.1 Orientation of water meter(s)

The position of the meters (spatial orientation) shall be as indicated by the manufacturer and they shall be mounted

in the test rig as appropriate:

a) if the meters are marked ‘H’ mount the connecting pipework with the flow axis in the horizontal plane during the test (indicating device positioned on top);

b) if the meters are marked ‘V’ mount the connecting pipework with the flow axis in the vertical plane during the test;

c) if the meters are not marked either ‘H’ or ‘V’:

1) at least one meter from the sample shall be mounted with the flow axis vertical, with flow direction from bottom to top;

2) at least one meter from the sample shall be mounted with the flow axis vertical and flow direction from top

to bottom;

3) at least one meter from the sample shall be mounted with the flow axis at an intermediate angle to the vertical and horizontal (chosen at the discretion of the approving authority);

4) the remaining meters from the sample shall be mounted with the flow axis horizontal;

5) where the meters have an indicating device, which is integral with the body of the meter, at least one of the horizontally mounted meters, shall be oriented with the indicating device positioned at the side and the remaining meters shall be oriented with the indicating device positioned at the top;

d) the tolerance on the position of the flow axis for all meters, whether horizontal, vertical or at an intermediate angle, shall be ± 5°

Trang 14

NOTE In the case of the meters, where the number of meters presented to test is less than four, supplementary needed meters will be taken from the basis population or the same meter will be submitted to different positions test

a) Determine the intrinsic errors (of indication) of the water meter (in the measurement of the actual volume), shall

be determined for at least the following seven flowrates, the error (of indication) at each flowrate being determined twice:

b) Test the water meter without its supplementary devices (if any) attached;

c) during a test hold all other influence factors at reference conditions;

d) measure the errors (of indication) at other flowrates if required, depending on the shape of the error curve; e) calculate the relative error (of indication) for each flowrate

5.3.4 Acceptance criteria

a) The errors observed for each of the seven flowrates shall not exceed the maximum permissible errors If the error observed on one or more meters is greater than the maximum permissible error at one flowrate only, the test at that flowrate shall be repeated The test shall be declared satisfactory if two out of the three results lie within the maximum permissible error and the arithmetic mean of the results for the three tests at that flowrate

is less than or equal to the maximum permissible error;

b) if all the errors (of indication) of the water meter have the same sign, at least one of these errors shall not exceed one half of the maximum permissible error In all cases this requirement shall be applied equitably with respect to the water supplier and the consumer

Trang 15

a) Take the intrinsic error of the water meter for the defined flowrate(s);

b) apply the influence quantity to be tested;

c) the error (of indication) of the meter at the defined flowrate(s) shall be taken as the arithmetic mean of the “n” error measurements during or after exposure to influence quantities, where “n” is at least one but may be

increased at the request of the manufacturer, prior to the test;

d) during a test hold all other influence quantities at reference conditions Reference pressure may be less than 0,2 MPa (2,0 bar) provided that this pressure is kept constant during all tests

EN 14154-2:2005+A2:2010, 5.1.3.1 – are not permitted

5.4.5.3 Test procedure

2 cartridge meters or exchangeable metrological units to be tested in 5 connection interfaces of every compatible interface type, resulting in 10 accuracy curves for each type of compatible interface Test flow rates according to the definition in 5.3.3 of this European Standard

5.4.5.4 Acceptance criteria

All accuracy curves shall be positioned within MPE at all times

Trang 16

Error variation within the five tests ± 0,5 MPE maximum if standard connection interfaces are being used on cartridge meters On meters with exchangeable metrological units standard interfaces are being used exclusively.$

5.5 Absence of flow test

a) Fill the meter with water, purging out all air;

b) ensure there is no flow through the measurement transducer;

c) observe the meter index for 15 min;

d) fully discharge the water from the meter;

e) observe the meter index for 15 min;

f) during the test, the reference conditions for all influence quantities other than flowrate shall be maintained

Apply the installation and operational requirements described in Annex C

For temperatures greater than 90 °C, where it is not possible to work in open atmosphere, it is recommended to use, as reference device, a reference meter or equivalent system working under pressure, calibrated at, or corrected to, the test temperature and pressure

Trang 17

15

c) the inlet water temperature related to the relevant temperature classes are specified in Table 1;

d) the water temperature shall be maintained within a tolerance of ± 5 °C, except for the test at the meter’s MAT which shall have a tolerance of + 0 / − 5 °C;

e) during the test, the reference conditions for all other influence quantities shall be maintained

Table 1 — Test water temperatures

The relative error (of indication) of the meter shall not exceed the applicable MPE

5.7 Overload water temperature test

This test only applies to meters with a MAT ≥ 50 °C

5.7.1 Object of test

To verify that the meter performance is not affected after exposure to an elevated, overload, water temperature, as required in #8.5$ of #EN 14154-1:2005+A2:2011$

5.7.2 Preparation

Apply the installation and operational requirements described in Annex C

The test is to be performed on one meter only

5.7.3 Test procedure

a) Expose the meter to a flow of water at a temperature of MAT+10 °C for a period of 1 h after the meter has reached temperature stability;

b) after recovery, determine the error (of indication) of the meter at the reference flowrate;

c) during the test, the reference conditions for all other influence quantities shall be maintained

5.7.4 Acceptance criteria

a) The meter functionality with regard to volume totalisation shall remain unaffected;

b) additional functionality, as indicated by the manufacturer, shall remain unaffected;

c) the error (of indication) of the meter shall not exceed the applicable MPE

5.8 Water pressure test

5.8.1 Object of test

To measure the effects of internal water pressure on the errors (of indication) of the meter The meter shall comply with the requirements in #5.1.2$ of #EN 14154-1:2005+A2:2011$

Trang 18

5.8.2 Preparation

Apply the installation and operational requirements described in Annex C

The water pressure shall be measured upstream of the meter inlet for MAP evaluation and downstream of the meter outlet for mAP evaluation

5.8.3 Test procedure

a) Determine the error (of indication) of at least one meter at a flow rate of Q2 with the inlet water pressure held firstly at the mAP (0,3 bar ± 5 %) and then at the MAP (+ 0 / − 10 %);

b) during the test, hold all other influence factors at reference conditions;

c) calculate the relative error (of indication) for each inlet water pressure

5.8.4 Acceptance criteria

The error (of indication) of the meter shall not exceed the applicable MPE

5.9 Verification of flow profile sensitivity classes

Some types of water meter, e.g volumetric water meters (that is, involving measuring chambers with mobile walls), such as oscillating piston or nutating disc meters have been shown to be insensitive to upstream installation conditions, hence this test is not applicable

The manufacturer shall specify the flow disturbance class for the meter, being no longer than 15 × DN upstream and 5 × DN downstream

5.9.1 Object of test

To verify that the meter complies with the requirements in 5.3 of #EN 14154-1:2005+A2:2011$

flow upstream and downstream of the meter are measured

NOTE 2 Types 1 and 2 disturbance devices are used in the tests to create left-handed (sinistrorsum) and right-handed (dextrorsum), rotational velocity fields (swirl), respectively The flow disturbance is of a type usually found downstream of two 90° bends directly connected at right angles A type 3 disturbance device creates an asymmetric velocity profile usually found downstream of a protruding pipe joint or a gate valve not fully opened

5.9.2 Preparation

Apply the installation and operational requirements described in Annex C In addition, the conditions described in 5.9.3 shall be applied

5.9.3 Test procedure

a) Using the types 1, 2 and 3 flow disturbers specified in Annex A, determine the error (of indication) of the meter

at a flowrate between 0,9 Q3 and Q3, for each one of the six installation conditions specified in Figure 1; b) during each test, all other influence factors shall be held at the reference conditions

Additional requirements:

c) For meters where the manufacturer has specified installation lengths of straight pipe of at least 15 × DN upstream and 5 × DN downstream of the meter, no external flow straighteners are allowed;

Trang 19

f) devices within the water meter having flow straightening functions shall not be considered to be a

“straightener” in the context of these tests;

g) some types of water meter which have been proven to be unaffected by flow disturbances upstream and downstream of the meter may be exempted from this test by the approving authority

5.9.4 Acceptance criteria

The relative error (of indication) of the meter shall not exceed the applicable MPE for any of the velocity field tests

Trang 20

Key

Type 1 disturber: − Swirl generator sinistrorsum

Type 2 disturber: − Swirl generator dextrorsum

Type 3 disturber: − Velocity profile flow disturber

Trang 21

19

5.10 Tests on ancillary devices of a water meter

5.10.1 Object of test

To verify that the meter complies with the requirements in #4.4.5$ of #EN 14154-1:2005+A2:2011$

Two types of test are required:

a) where supplementary devices can be attached temporarily to the meter, e.g for testing or data transmission purposes, the error of indication of the meter shall be measured with the supplementary device(s) fitted to ensure that the errors of indication do not exceed the MPE’s;

b) for permanently fitted and temporarily fitted supplementary devices, the indications of volume from the supplementary device/s shall be checked to ensure that the readings do not differ from those of the primary indication

5.10.2 Preparation

a) The installation and operational requirements described in Annex C shall apply and the meter orientation shall

be in accordance the markings on the meter;

b) temporary supplementary devices shall be fitted either by the manufacturer or in accordance with the manufacturer’s instructions;

c) where the output from a supplementary device is an electric signal consisting of a pulse stream in which a single pulse corresponds to a finite volume, the pulses may be totalled on an electronic totalizer which, when connected, has no significant influence on the electrical signal

5.11 Pressure loss test

Trang 22

5.11.3 Test procedure

Where it has been established that the pressure loss of the meter will follow the square law, the pressure loss shall

be tested at Q3 only

When it is suspected that a pressure loss peak occurs below Q3, the pressure loss shall be determined between

Q1 - Q3, starting at Q1 and increasing the flowrate by max 0,1 × Q3 After Q3 is reached, the flow rate shall be decreased by max 0,1 × Q3

Concentric meters, of any type and measuring principle, shall be tested together with its respective manifold

5.11.4 Acceptance criteria

The pressure loss measurement at any flow rate within the ROC shall not exceed the maximum value of the pressure loss corresponding to the class declared by the manufacturer

5.12 Reverse flow test

5.12.1 Meters designed to measure reverse flows

 error of indication (according to provisions in 5.3);

 flow disturbances (according to provisions in 5.9);

 pressure loss (according to provisions in 5.11);

 endurance (according to provisions in 5.13)

5.12.1.4 Acceptance criteria

The test is satisfactory if at any test the respective acceptance criteria of 5.3, 5.9 and 5.11 are fulfilled

5.12.2 Meters not designed to measure reverse flows

Trang 23

Test flow rate

Test water temperature Type of test

Number of interruptions

Duration

of pauses

Period of operation

at test flowrate

Duration of start-up and rundown

a [Q3] is the number equal to the value of Q3 expressed in m 3 /h

5.13.1 Continuous flow test

5.13.1.1 Object of test

To verify that the water meter is durable when subjected to continuous, permanent and overload flow conditions The meter shall complies with the requirements in #8.9$ of #EN 14154-1:2005+A2:2011$

Trang 24

The test consists of subjecting the meter to a constant flowrate of even Q3 or Q4 according to the prescriptions layed down in #5.13$, Table 2

5.13.1.2 Preparation

5.13.1.2.1 Description of the installation

The installation consists of:

a) a water supply (non-pressurised, pressurised tank; pump; etc.);

b) one or more isolating valves;

c) a device for measuring the water temperature at the meter inlet;

d) means for checking the flowrate and duration of the test;

e) devices for measuring pressure at the inlet and outlet

The different devices shall not cause cavitation phenomena

c) carry out the following tests:

1) For meters with Q3 ≤ 16 m3/h, run the meter at a flowrate of Q4for a period of 100 h;

2) For meters with Q3 > 16 m3/h, run the meter at a flowrate of Q4 for a period of 200 h then at Q3for a period

Trang 25

23

g) for each flowrate, subtract the error (of indication) obtained before the test (step a)) from the error (of indication) obtained after the test (step f))

5.13.1.3.1 Tolerance on flowrate

The flowrate shall be kept constant throughout the test at a predetermined level

The relative variation of the flowrate values during each test shall not exceed ± 10 % (except when starting and stopping)

5.13.1.3.2 Tolerance on test timing

The specified duration of the test is a minimum value

5.13.1.3.3 Tolerance on discharged volume

The volume indicated at the end of the test shall not be less than that determined from the product of the specified nominal flowrate of the test and the specified nominal duration of the test

To satisfy this condition, sufficiently frequent corrections to the flowrate shall be made The water meters on test may be used to check the flowrate

5.13.1.3.4 Test readings

During the test the following readings from the test rig shall be recorded at least once every 24 h period, or once for every shorter period if the test is so divided:

a) water pressure upstream of the meter(s);

b) water pressure downstream of the meter(s);

c) water temperature upstream of the meter(s);

d) flowrate;

e) readings of the test meter/s;

f) volume passed by the meter(s)

5.13.1.4 Acceptance criteria

After the continuous endurance test:

a) The variation in the error curve shall not exceed:

 3 % for flowrates in the lower zone (Q1 ≤ Q < Q2); and

 1,5 % for flowrates in the upper zone (Q2 ≤ Q ≤ Q4)

For the purpose of determining these requirements the mean values of the errors (of indication) at each flowrate, shall apply;

b) the error curves shall not exceed a maximum error limit of:

 ± 6 % for flowrates in the lower zone (Q1 ≤ Q < Q2); and

 ± 2,5 % for flowrates in the upper zone (Q2 ≤ Q ≤ Q4) for meters intended to meter water with a temperature between 0,1 °C and 30 °C; or

Trang 26

 ± 3,5 % for flowrates in the upper zone (Q2 ≤ Q ≤ Q4) for meters intended to meter water with a temperature greater than 30 °C

5.13.2 Discontinuous flow test

This test is applied only to meters with Q3 ≤ 16 m3/h and combination meters

5.13.2.1 Object of test

To verify that the water meter is durable when subjected to cyclic flow conditions

The test consists of subjecting the meter to the specified number of starting and stopping flowrate cycles of short

duration, the constant test flowrate phase of each cycle being kept at the specified flowrate (Q3) throughout the duration of the test

5.13.2.2 Preparation

5.13.2.2.1 Description of the installation

The installation consists of:

a) a water supply (non-pressurised, pressurised tank; pump; etc.);

b) pipe-work

5.13.2.2.2 Pipe-work

The meters can be arranged in series or in parallel, or the two systems can be combined

In addition to the meter(s), the piping system consists of:

a) one flow-regulating device (per line of meters in series, if necessary);

b) one or more isolating valves;

c) a device for measuring the temperature of the water upstream of the meters;

d) means for checking: the flowrate, the duration of cycles and the number of cycles;

e) one flow-interrupting device for each line of meters in series;

f) devices for measuring pressure at the inlet and outlet

The various devices shall not cause cavitation phenomena or other types of parasitic wear of the meter(s)

5.13.2.2.3 Precautions to be taken

The meter(s) and connecting pipes shall be suitably bled of air

The flow variation during the repeated opening and closing operations shall be progressive, so as to prevent water hammer

5.13.2.2.4 Flowrate cycles

A complete cycle comprises the following four phases:

a) period from zero to test flowrate Q3;

Trang 27

25

b) period at constant test flowrate Q3;

c) period from the test flowrate Q3 to zero;

d) period at zero flowrate

The test programme shall specify the number of flowrate cycles, the duration of the four phases of a cycle, and the total volume to be discharged

5.13.2.3 Test procedure

5.13.2.3.1 For all types of meters

a) Before commencing the continuous endurance test, measure the errors (of indication) of the meters as described in 5.3 and at the same flowrates;

b) mount the meters either singly or in batches in the test rig in the same orientations as those used in the determination of the intrinsic error (of indication) tests (see 5.3.2.1);

c) during the tests, hold the meters within their rated operating conditions and with the pressure downstream of the meters high enough to prevent cavitation in the meters;

d) adjust the flowrate to within the specified tolerances;

e) run the meters at the conditions shown in Table 2;

f) following the discontinuous endurance test, measure the final errors (of indication) of the meters as described

in 5.3 and at the same flowrates;

g) calculate the relative errors (of indication) for each flowrate;

h) for each flowrate, subtract the value of intrinsic error (of indication) obtained before the test (step a)) from the error (of indication) obtained after the test (step g))

5.13.2.3.2 Specific test for combination meters

After the test described in 5.13.2.3.1, the combination meter shall undergo an endurance test simulating service conditions, under the following conditions:

a) test flowrate: at least twice the change-over flowrate (Q x), determined using increasing flowrates;

b) type of test: discontinuous;

c) number of interruptions: 50 000;

d) duration of stop: 15 s;

e) duration of running at the test flowrate: 15 s;

f) duration of acceleration and deceleration: minimum 3 s, maximum 6 s

5.13.2.3.3 Tolerance on flowrate

The relative variation of the flow values shall not exceed ± 10 % outside the opening, closing and stoppage periods The meters on test may be used to check the flowrate

Trang 28

5.13.2.3.4 Tolerance on test timing

The tolerance on the specified duration of each phase of the flow cycle shall not exceed ± 10 %

The tolerance on the total test duration shall not exceed ± 5 %

5.13.2.3.5 Tolerance on the number of cycles

The number of cycles shall not be less than that stipulated, but shall not exceed this number by more than 1 %

5.13.2.3.6 Tolerance on discharged volume

The volume discharged throughout the test shall be equal to half the product of the specified nominal test flow times the total theoretical duration of the test (operating periods plus transient and stoppage periods with a tolerance of ± 5 %)

This precision can be obtained by sufficiently frequent corrections of the instantaneous flows and operating periods

5.13.2.3.7 Test readings

During the test the following readings from the test rig shall be recorded at least once every 24 h period, or once for every shorter period if the test is so divided:

a) water pressure upstream of the meter(s);

b) water pressure downstream of the meter(s);

c) water temperature upstream of the meter(s);

d) flowrate;

e) duration of the four phases of the cycle of the discontinuous flow test;

f) number of cycles;

g) readings of the test meter/s;

h) volume passed by the meters

5.13.2.4 Acceptance criteria

After the cyclic endurance test:

a) the variation in the error curve shall not exceed:

 3 % for flowrates in the lower zone (Q1 ≤ Q < Q2); and

 1,5 % for flowrates in the upper zone (Q2 ≤ Q ≤ Q4)

For the purpose of determining these requirements the mean values of the errors (of indication) at each flowrate, shall apply:

b) the error curves shall not exceed a maximum error limit of:

 ± 6 % for flowrates in the lower zone (Q1 ≤ Q < Q2); and

 ± 2,5 % for flowrates in the upper zone (Q2 ≤ Q ≤ Q4) for meters intended to meter water with a temperature between 0,1 °C and 30 °C; or

Trang 29

These performance tests are additional to the tests described in Clause 5 and apply to complete meters, to separable parts of a water meter, and, if required, to ancillary devices

When the effect of one influence quantity is being evaluated, all other influence quantities are to be held at the reference conditions (see 3)

The pattern approval tests specified in this clause may be carried out in parallel with the tests specified in Clause 5, using an example of the same model of the water meter, or its separable parts

 class B for fixed meters installed in a building;

 class C for fixed meters installed outdoors;

 class I for mobile meters

The applicant for pattern approval may also indicate specific environmental conditions in the documentation supplied to the metrology service, based on the intended use of the meter In this case, the metrology service will carry out performance tests at severity levels corresponding to these environmental conditions These severity levels shall not be less than class B

In all cases the metrology service shall verify that the conditions of use are met

6.1.2 Electromagnetic environments

Water meters are divided into two electromagnetic environment classes:

 class E1 residential, commercial and light industrial;

 class E2 industrial

6.1.3 Reference conditions

Reference conditions are listed in Clause 3

Trang 30

6.1.4 Test volumes for measuring error (of indication) of a water meter

Some influence quantities should have a constant effect on the error of indication of a water meter and not a proportional effect related to the measured volume

In other tests the effect of the influence quantity applied to a water meter is related to the measured volume In these cases, in order to be able to compare results obtained in different laboratories, the test volume for measuring

the error of indication of the meter shall correspond to that delivered in 1 min at the over-load flowrate Q4

However, some tests may require more than one minute, in which case they shall be carried out in the shortest possible time, taking into consideration the measurement uncertainty

6.1.5 Influence of the water temperature

Dry heat, cold and damp heat tests are concerned with measuring the effects of ambient air temperature on the performance of the meter However, the presence of the measurement transducer, filled with water, may also influence heat dissipation in electronic components

If the meter has a value Q3 ≤ 16 m3/h, the meter should have water passing through it at the reference flowrate and the error (of indication) of the meter shall be measured with the electronic parts and the measurement transducer subjected to the reference conditions

Optionally, a simulation of the measurement transducer may be used for testing all electronic components Where simulated tests are used, they shall replicate the effects caused by the presence of water for those electronic devices which are normally attached to the flow sensor, and the reference conditions shall be applied during the tests

6.1.6 Requirements for environmental tests

The following requirements are associated with the environmental tests and the relevant IEC/CENELEC standards are listed in the appropriate sections of this document:

a) preconditioning of the Equipment Under Test (EUT);

b) any deviations in the procedure from the relevant IEC/CENELEC standard;

c) initial measurements;

d) state of the EUT during conditioning;

e) severity levels, values of the influence factor and duration of exposure;

f) measurements required and/or the loading during conditioning;

g) recovery of the EUT;

h) final measurements;

i) the acceptance criteria for the EUT passing a test

Where no IEC/CENELEC standard exists for a specific test, the essential requirements for the test are given in this document

6.1.7 Equipment under test (EUT)

For the purpose of testing, the EUT shall be categorised as one of the cases, A to E, according to the technology described in subclauses 6.1.7.1 to 6.1.7.4, and the following requirements shall apply:

 case A: no performance test (as mentioned in this section) is required;

Trang 31

29

 case B: the EUT is the complete meter, the test shall be carried out with water in the volume or flow sensor;

 case C: the EUT is the measurement transducer, the test shall be carried out with water in the volume or flow sensor;

 case D: the EUT is the electronic calculator including the indicating device or the ancillary device The test shall be carried out with water in the volume or flow sensor;

 case E: the EUT is the electronic calculator including the indicating device or the ancillary device, the test may

be carried out with simulated measurement signals without water in the volume or flow sensor

6.1.7.1 Positive displacement meters and turbine water meters

a) The meter is not fitted with electronic devices: Case Ab) measurement transducer and the electronic calculator including the indicating device are in

c) measurement transducer is separate from the electronic calculator, but not fitted with

electronic devices:

Case A

d) measurement transducer is separate from the electronic calculator and fitted with electronic

e) electronic calculator including the indicating device is separate from the measurement

transducer and simulation of the measurement signals is not possible: Case D

f) electronic calculator including the indicating device is separate from the measurement

transducer and simulation of the measurement signals is possible: Case E

6.1.7.2 Electromagnetic water meters

a) The measurement transducer and the electronic calculator including the indicating device are

b) flow sensor, consisting only of the pipe, the coil and the two meter electrodes, is without any

c) measurement transducer including the flow sensor is separate from the electronic calculator

d) electronic calculator including the indicating device is separate from the measurement

transducer and simulation of the measurement signals is not possible:

Case D

6.1.7.3 Ultrasonic water meters, Coriolis water meters, Fluidic water meters, etc

a) The measurement transducer and the electronic calculator including the indicating device are

b) measurement transducer is separate from the electronic calculator and fitted with electronic

c) electronic calculator including the indicating device is separate from the measurement

transducer and simulation of the measurement signals is not possible: Case D

Trang 32

6.1.7.4 Ancillary devices

a) The ancillary device is a part of the meter, a part of the measurement transducer or a part of

b) ancillary device is separate from the meter, but not fitted with electronic devices: Case A

c) ancillary device is separate from the meter, a simulation of the input signals is not possible: Case Dd) ancillary device is separate from the meter, a simulation of the input signals is possible: Case E

6.2 Climatic and mechanical environment

6.2.1 Dry heat (non-condensing)

Table 3 — Influence factor: Dry heat (non-condensing)

Influence factor: Dry heat (non-condensing)

The testing arrangements are those described in the following publications:

EN 60068-2-2:1993, +A1:1993, +A2:1994, Basic environmental testing procedures — Part 2: Tests,

Test B: Dry heat

Guidance on testing arrangements is given in:

EN 60068-3-1:1999, Environmental testing — Part 3: Background information, Section 1: Cold and dry heat

tests

EN 60068-1:1994, Environmental testing — Part 1: General and guidance

6.2.1.3 Test procedure (in brief)

a) No pre-conditioning is required;

b) measure the error of indication of the EUT at the reference flowrate and at the following test conditions:

1) at the reference air temperature of (20 ± 5) °C, before conditioning the EUT;

2) at an air temperature of (55 ± 2) °C, after the EUT has been stabilized at this temperature for a period of

2 h;

3) at the reference air temperature of (20 ± 5) °C, after recovery of the EUT;

Trang 33

31

c) calculate the relative error (of indication) at each test condition;

d) check that the EUT is functioning correctly

During the application of the influence factor:

a) all the functions of the EUT shall operate as designed; and

b) the relative error of indication of the EUT, at the test conditions, shall not exceed the MPE of the “upper zone“

6.2.2 Cold

Table 4 — Influence factor: Cold

Influence factor: Cold

The testing arrangements are those described in the following publications:

EN 60068-2-1:1993, +A1:1993, +A2:1994, Environmental testing — Part 2: Tests, Test A: Cold.

Guidance on testing arrangements is given in:

EN 60068-3-1:1999, Environmental testing — Part 3: Background information, Section 1: Cold and dry heat

tests

EN 60068-1:1994, Environmental testing — Part 1: General and guidance

Trang 34

6.2.2.3 Test procedure (in brief)

a) Do not pre-condition the EUT;

b) measure the error (of indication) of the EUT at the reference flowrate (actual or simulated) and at the reference air temperature;

c) stabilize the air temperature at either - 25 °C (severity level 3) or + 5 °C (severity level 1) for a period of 2 h; d) measure the error (of indication) of the EUT at the reference flowrate (actual or simulated) at an air temperature of either - 25 °C (severity level 3) or + 5 °C (severity level 1);

e) measure the error of indication of the EUT at the reference flowrate (actual or simulated), and at the reference air temperature, after recovery of the EUT;

f) calculate the relative error (of indication) at each test condition;

g) check that the EUT is functioning correctly

During the application of the influence factor:

a) all the functions of the EUT shall operate as designed; and

b) the relative error (of indication) of the EUT, at the test conditions, shall not exceed the MPE of the “upper zone“

6.2.3 Damp heat, cyclic (condensing)

Table 5 — Influence factor: Damp heat, cyclic (condensing)

Influence factor: Damp heat, cyclic (condensing)

Trang 35

The testing arrangements are those described in the following publications:

EN 60068-2-30:1999, Environmental testing — Part 2: Tests — Test Db and guidance: Damp heat, cyclic

(12 h + 12 h cycle)

IEC 60068-3-4:2001, Environmental testing — Part 3-4: Supporting documentation and guidance — Damp

heat tests

6.2.3.3 Test procedure (in brief)

The requirements for, the performance of the test equipment, conditioning and recovery of the EUT, and exposure

of the EUT to cyclic temperature changes under damp heat conditions, are described in EN 60068-2-30

The test programme consists of steps a) to f) below

a) Pre-condition the EUT;

b) expose the EUT to cyclic temperature variations between the lower temperature of 25 °C and the upper temperature 55 °C (environmental classes C and I) or 40 °C (environmental class B) Maintaining the relative humidity above 95 % during the temperature changes and during the phases at low temperature, and at 93 %

at the upper temperature phases Condensation should occur on the EUT during the temperature rise;

c) allow the EUT to recover;

d) after recovery, measure the error (of indication) of the EUT at the reference flowrate;

e) calculate the relative error (of indication);

f) check that the EUT is functioning correctly

Additional requirements:

g) the power supply to the EUT is switched off during steps a), b) and c);

h) when measuring the errors (of indication), the installation and operational conditions described in 5.3.2 shall be followed and the reference conditions shall be applied unless otherwise specified Meters not marked either ‘H’ or

‘V’, only will be tested with the flow axis in horizontal orientation Meters with two reference temperatures, will be only tested at lowest reference temperature

6.2.3.4 Acceptance criteria

After the application of the influence factor and recovery:

a) all the functions of the EUT shall operate as designed;

b) the relative error of indication of the EUT, at the test conditions, shall not exceed the MPE of the “upper zone“

Trang 36

6.2.4 Vibration (random)

Table 6 — Disturbance: Vibration (random)

Disturbance: Vibration (random)

6.2.4.1 Object of test

To verify that the meter complies with the requirements in #8.6$ of #EN 14154-1:2005+A2:2011$, after the application of random vibrations

6.2.4.2 Preparation

The testing arrangements are those described in the following publications:

EN 60068-2-64:1994, Environmental testing — Part 2: Test methods — Test Fh: Vibration, broad-band

random (digital control) and guidance

EN 60068-2-47:1999 +Err:2000, Environmental testing — Part 2-47: Test methods — Mounting of components,

equipment and other articles for vibration, impact and similar dynamic tests

6.2.4.3 Test procedure (in brief)

a) Mount the EUT on a rigid fixture by its normal mounting means, such that the gravitational force acts in the same direction as it would in normal use However, if the gravitational effect is insignificant, and the meter is not marked H or V, the EUT may be mounted in any position;

b) apply random vibrations, over the frequency range 10 Hz to 150 Hz, to the EUT, in three mutually perpendicular axes in turn, for a period of at least 2 min per axis;

c) allow the EUT a period for recovery;

d) examine the EUT for correct functioning;

e) measure the error (of indication) of the EUT at the reference flowrate;

f) calculate the relative error (of indication)

Additional requirements:

g) where the flow sensor is included in the EUT, it shall not be filled with water during the application of the disturbance;

h) the power supply to the EUT shall be switched off during steps a), b) and c);

i) during the application of the vibrations the following conditions shall be met:

Trang 37

35

 Total RMS level: 7 m.s-2

 ASD level 10 to 20 Hz: 1 m2.s-3

 ASD level 20 to 150 Hz: -3 dB/octave

j) when measuring the errors (of indication), the installation and operational conditions described in 5.3.2 shall be followed and the reference conditions shall be applied unless otherwise specified Meters not marked either ‘H’ or

‘V’, only will be tested with the flow axis in horizontal orientation Meters with two reference temperatures, will be only tested at lowest reference temperature

6.2.4.4 Acceptance criteria

After the application of the disturbance and recovery:

a) all the functions of the EUT shall operate as designed;

b) the error (of indication) of the EUT, at the test conditions, shall not exceed the MPE of the “upper zone“

6.2.5 Mechanical shock

Table 7 — Disturbance: Mechanical shock

Disturbance: Mechanical shock

The testing arrangements are those described in the following publication:

EN 60068-2-31:1993, Basic environmental testing procedures — Part 2: Tests — Test Ec: Drop and topple,

primarily for equipment-type specimens

EN 60068-2-47:1999 +Err:2000, Environmental testing — Part 2-47: Test methods — Mounting of components,

equipment and other articles for vibration, impact and similar dynamic tests

6.2.5.3 Test procedure (in brief)

a) The EUT shall be placed on a rigid level surface in its normal position of use and tilted towards one bottom edge until the opposite edge of the EUT is 50 mm above the rigid surface However, the angle made by the bottom of the EUT and the test surface shall not exceed 30°;

b) allow the EUT to fall freely onto the test surface;

c) repeat steps a) and b) for each bottom edge;

d) allow the EUT a period for recovery;

e) examine the EUT for correct functioning;

Trang 38

f) measure the error (of indication) of the EUT at the reference flowrate;

g) calculate the relative error (of indication)

Additional requirements:

h) where the flow sensor is part of the EUT, it shall not be filled with water during the application of the disturbance;

i) the power supply to the EUT shall be switched off during steps a), b) and c);

j) when measuring the errors (of indication), the installation and operational conditions described in 5.3.2 shall be followed and the reference conditions shall be applied unless otherwise specified Meters not marked either ‘H’ or

‘V’, only will be tested with the flow axis in horizontal orientation Meters with two reference temperatures, will be only tested at lowest reference temperature

6.2.5.4 Acceptance criteria

After the application of the disturbance and recovery:

a) all the functions of the EUT shall operate as designed; and

b) the relative error of indication of the EUT, at the test conditions, shall not exceed the MPE of the “upper zone“

6.3 Electromagnetic environment

6.3.1 Electrostatic discharge

Table 8 — Disturbance: Electrostatic discharges

Disturbance: Electrostatic discharges

Number of test cycles:

At each test point, at least ten direct discharges shall be applied at intervals of at least 10 s between discharges, during the same measurement or simulated measurement

For indirect discharges, a total of ten discharges shall be applied on the horizontal coupling plane, and a total of ten discharges for each of the various positions of the vertical coupling plane

6.3.1.1 Object of test

To verify that the meter complies with the requirements in #8.7$ of #EN 14154-1:2005+A2:2011$, during the application of direct and indirect electrostatic discharges

6.3.1.2 Preparation

The testing arrangements are those described in the following publication:

EN 61000-4-2:1995, +A1:1998, +A2:2001, Electromagnetic compatibility (EMC) — Part 4-2: Testing and

measurement techniques — Electrostatic discharge immunity test

Trang 39

37

6.3.1.3 Test procedure (in brief)

a) Measure the error (of indication) of the EUT before applying the electrostatic discharges;

b) charge a capacitor of 150 pF capacitance by means of a suitable DC voltage source, then discharge the capacitor through the EUT by connecting one terminal of the supporting chassis to ground and the other via

a 330 Ω resistor, to surfaces of the EUT which are normally accessible to the operator The test includes the paint penetration method, if appropriate;

c) measure the error (of indication) of the EUT during the application of the electrostatic discharges;

d) calculate the error (of indication) of the EUT for each test condition;

e) calculate the significant fault by subtracting the relative error (of indication) of the meter measured before applying the electrostatic discharges from that measured after the applying the electrostatic discharges

Additional requirements:

f) during the measurement of the error of indication the EUT shall be subjected to the reference flowrate;

g) when measuring the error (of indication), the installation and operational conditions described in 5.3.2 shall be followed and the reference conditions shall be applied unless otherwise specified Meters not marked either ‘H’ or

‘V’, only will be tested with the flow axis in horizontal orientation Meters with two reference temperatures, will be only tested at lowest reference temperature

6.3.1.4 Acceptance criteria

a) After the application of the disturbance, all the functions of the EUT shall operate as designed;

b) the difference between the relative error of indication, obtained during the application of the electrostatic discharges and that obtained before the test, under reference conditions, shall not exceed one half of the maximum permissible error of the “upper zone"

6.3.2 Radiated radio frequency/Electromagnetic fields

Table 9 — Disturbance: Electromagnetic radiation

Disturbance: Electromagnetic radiation

The testing arrangements are those described in the following publication:

EN 61000-4-3:1996, +A1:1998, +A2:2001, Electromagnetic compatibility (EMC) — Part 4-3: Testing and

measurement techniques — Radiated, radio-frequency, electromagnetic field immunity test

Trang 40

6.3.2.3 Test procedure (in brief)

a) Measure the intrinsic error (of indication) of the EUT at reference conditions before applying the electromagnetic field;

b) apply the electromagnetic field in accordance with the additional requirements of a) to e) given below;

c) start a new measurement of the error (of indication) for the EUT;

d) step the carrier frequency until the next carrier frequency in #Table 10$, is reached, in accordance with additional requirements of e) given below;

e) stop the measurement of the error (of indication) for the EUT;

f) calculate the relative error (of indication) of the EUT during the application of the electromagnetic field;

g) calculate the significant fault as the difference between the intrinsic error (of indication) from step a) and the error (of indication) from step f);

h) change the polarisation of the antenna;

l) the preferred transmitting antennae are, a bi-conical antenna for the frequency range 26 MHz to 200 MHz and

a log-periodic antenna for the frequency range 200 MHz to #2 000 MHz$;

m) the test is performed as 20 partial scans with vertical antenna and 20 partial scans with horizontal antenna The start and stop frequencies for each scan are listed in #Table 10$;

#Table 10 — Start and stop carrier frequencies

Ngày đăng: 14/04/2023, 08:04

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN