Bsi bs en 01159 3 2003

30 2 0

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Advanced Technical Ceramics — Ceramic Composites, Thermophysical Properties — Part 3: Determination Of Specific Heat Capacity
Trường học British Standards Institution
Chuyên ngành Advanced Technical Ceramics
Thể loại British Standard
Năm xuất bản 2003
Thành phố London
Định dạng
Số trang 30
Dung lượng 321,4 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Cấu trúc

  • 4.1 Principle (7)
  • 4.2 Apparatus (7)
  • 4.3 Standard reference materials (7)
  • 4.4 Test specimens (7)
  • 4.5 Calibration of calorimeter (7)
    • 4.5.1 General (7)
    • 4.5.2 Electrical Calibration (8)
    • 4.5.3 Calibration using standard reference material (8)
  • 4.6 Test procedures (8)
    • 4.6.1 Test without a crucible (8)
    • 4.6.2 Test with a crucible (8)
    • 4.6.3 Description of test (9)
  • 4.7 Calculations (9)
    • 4.7.1 General (9)
    • 4.7.2 Determination of the calorimetric calibration factor (10)
    • 4.7.3 Determination of mean specific heat capacity C p (10)
  • 5.1 Principle (11)
    • 5.1.1 General (11)
    • 5.1.2 Stepwise heating method (11)
    • 5.1.3 Continuous heating method (11)
  • 5.2 Apparatus (12)
    • 5.2.1 Differential scanning calorimeter (12)
  • 5.3 Standard reference materials, SRM (12)
  • 5.4 Test specimens (12)
  • 5.5 Temperature calibration (12)
  • 5.6 Test procedure for the determination of C p (13)
    • 5.6.1 General (13)
    • 5.6.2 Method 1: Measurements requiring the knowledge of the K factor (13)
    • 5.6.3 Method 2: Measurements requiring the use of a reference standard material (SRM) (14)
  • 5.7 Calculation of results (15)
    • 5.7.1 Method requiring the knowledge of the K factor (15)
    • 5.7.2 Method using an SRM (16)

Nội dung

www bzfxw com BRITISH STANDARD BS EN 1159 3 2003 Advanced technical ceramics — Ceramic composites, thermophysical properties — Part 3 Determination of specific heat capacity The European Standard EN 1[.]

Trang 1

Advanced technical ceramics — Ceramic composites,

thermophysical properties —

Part 3: Determination of specific heat capacity

The European Standard EN 1159-3:2003 has the status of a British Standard

ICS 81.060.20

12&23<,1*:,7+287%6,3(50,66,21(;&(37$63(50,77('%<&23<5,*+7/$:

Trang 2

`,,,,`,,`,,,,,,,``,`,,,``,``-`-`,,`,,`,`,,` -This British Standard was

published under the authority

of the Standards Policy and

This British Standard is the official English language version of

EN 1159-3:2003 It supersedes DD ENV 1159-3:1995 which is withdrawn.The UK participation in its preparation was entrusted to Technical Committee RPI/13, Advanced technical ceramics, which has the responsibility to:

A list of organizations represented on this committee can be obtained on request to its secretary

Cross-references

The British Standards which implement international or European

publications referred to in this document may be found in the BSI Catalogue

under the section entitled “International Standards Correspondence Index”, or

by using the “Search” facility of the BSI Electronic Catalogue or of British

— aid enquirers to understand the text;

— present to the responsible international/European committee any enquiries on the interpretation, or proposals for change, and keep the

Amendments issued since publication

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 3

`,,,,`,,`,,,,,,,``,`,,,``,``-`-`,,`,,`,`,,` -EUROPÄISCHE NORM

April 2003

Détermination de la capacité thermique spécifique

Hochleistungskeramik - Keramische Verbundwerkstoffe, thermophisikalische Eigenschaften - Teil 3: Bestimmung

der spezifischen Wärmekapazität

This European Standard was approved by CEN on 2 January 2003.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the Management Centre or to any CEN member.

This European Standard exists in three official versions (English, French, German) A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal, Slovakia, Spain, Sweden, Switzerland and United Kingdom.

EUROPEAN COMMITTEE FOR STANDARDIZATION

C O M I T É E U R O P É E N D E N O R M A L I S A T I O N

E U R O P Ä I S C H E S K O M I T E E F Ü R N O R M U N G

Management Centre: rue de Stassart, 36 B-1050 Brussels

© 2003 CEN All rights of exploitation in any form and by any means reserved

worldwide for CEN national Members.

Ref No EN 1159-3:2003 E

Trang 4

Contents

page

Foreword 3

1 Scope 4

2 Normative references 4

3 Terms and definitions 4

4 Method A - Drop calorimetry 5

4.1 Principle 5

4.2 Apparatus 5

4.3 Standard reference materials 5

4.4 Test specimens 5

4.5 Calibration of calorimeter 5

4.5.1 General 5

4.5.2 Electrical Calibration 6

4.5.3 Calibration using standard reference material 6

4.6 Test procedures 6

4.6.1 Test without a crucible 6

4.6.2 Test with a crucible 6

4.6.3 Description of test 7

4.7 Calculations 7

4.7.1 General 7

4.7.2 Determination of the calorimetric calibration factor 8

4.7.3 Determination of mean specific heat capacity Cp 8

5 Method B - Differential scanning calorimetry 9

5.1 Principle 9

5.1.1 General 9

5.1.2 Stepwise heating method 9

5.1.3 Continuous heating method 9

5.2 Apparatus 10

5.2.1 Differential scanning calorimeter 10

5.3 Standard reference materials, SRM 10

5.4 Test specimens 10

5.5 Temperature calibration 10

5.6 Test procedure for the determination of Cp 11

5.6.1 General 11

5.6.2 Method 1: Measurements requiring the knowledge of the K factor 11

5.6.3 Method 2: Measurements requiring the use of a reference standard material (SRM) 12

5.7 Calculation of results 13

5.7.1 Method requiring the knowledge of the K factor 13

5.7.2 Method using an SRM 14

6 Test report 16

Annex A (normative) Drop calorimetry - Determination of the calibration factor using standard reference material 19

Annex B (informative) Standard reference material 20

Annex C (informative) Materials for calorimeter calibrations 25

Bibliography 26

Copyright British Standards Institution Reproduced by IHS under license with BSI - Uncontrolled Copy Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 5

Foreword

This document (EN 1159-3:2003) has been prepared by Technical Committee CEN/TC 184 "Advanced technical

ceramics", the secretariat of which is held by BSI

This European Standard shall be given the status of a national standard, either by publication of an identical text or

by endorsement, at the latest by October 2003, and conflicting national standards shall be withdrawn at the latest

by October 2003

This document supersedes ENV 1159-3:1995

EN 1159 Advanced technical ceramics – Ceramic composites, thermophysical properties consists of three parts:

 Part 2: Determination of thermal diffusivity

 Part 3: Determination of specific heat capacity

Annex A is normative Annexes B and C are informative

According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following

countries are bound to implement this European Standard: Austria, Belgium, Czech Republic, Denmark, Finland,

France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal,

Slovakia, Spain, Sweden, Switzerland and the United Kingdom

Trang 6

4

1 Scope

This part of EN 1159 describes two methods for the determination of the specific heat capacity of ceramic matrix

composites with continuous reinforcements (1D, 2D, 3D)

Unidirectional (1D), bi-directional (2D) and tridirectional (XD, with 2 < x ≤ 3)

The two methods are:

 method A: drop calorimetry;

 method B: differential scanning calorimetry

They are applicable from ambient temperature up to a maximum temperature depending on the method: method A

may be used up to 2 250 K, while method B is limited to 1 900 K

NOTE Method A is limited to the determination of an average value of the specific heat capacity over a given temperature

range and can give a larger spread of results

2 Normative references

This European Standard incorporates by dated or undated reference, provisions from other publications These

normative references are cited at the appropriate places in the text, and the publications are listed hereafter For

dated references, subsequent amendments to or revisions of any of these publications apply to this European

Standard only when incorporated in it by amendment or revision For undated references the latest edition of the

publication referred to applies (including amendments)

EN 60584-1, Thermocouples - Part 1: Reference tables (IEC 60584-1:1995)

ENV 13233:1998, Advanced technical ceramics – Ceramic composites – Notations and symbols

3 Terms and definitions

For the purposes of this European Standard, the following definitions and those given in ENV 13233:1998 apply

3.1

specific heat capacity, Cp

amount of heat required to raise the temperature of a mass unit of material by 1 K at constant temperature and

pressure

dT

dQ m

mean specific heat capacity, Cp

amount of heat required to raise the temperature of a mass unit of a material from temperature T1 to temperature

T2 at a constant pressure, divided by the temperature range (T2 – T1) expressed in K

3.3

representative volume element (R.V.E.)

the minimum volume which is representative of the material considered

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 7

The mean specific heat capacity is determined from the measured amount of heat required to maintain the

temperature constant in the second chamber Transfer of the test piece shall be done under conditions as close as

possible to adiabatic conditions

4.2 Apparatus

4.2.1 Drop calorimeter, there are several types of drop calorimeters They include one (or more) conditioning

chambers and measuring chambers which can be operated under controlled atmosphere and which are all

equipped with a temperature control system which allows a temperature stability of less than 1 K

The conditioning chamber shall have a homogeneous temperature zone size greater than the test specimen size

The measuring chamber shall have a homogeneous temperature zone of a sufficient length to accept several

specimens and a sufficient thermal inertia to limit the temperature disturbance, due to the drop

Heat transfer by radiation during the drop shall be avoided as far as possible

4.2.2 Balance, with an accuracy of 0,1 mg for test pieces over 10 mg and an accuracy of 0,01 mg for test pieces

below 10 mg

4.2.3 Temperature detectors, thermocouples in accordance to EN 60584-1 shall be used for the measurement

of temperature up to 1 920 K

For higher temperature, infrared detectors or any other suitable device may be used

4.2.4 Data acquisition system, the sampling period during the test shall be less than 0,5 s.

4.3 Standard reference materials

Standard reference materials which can be used for calibration purposes are listed in annex B

4.4 Test specimens

The test specimens shall be representative of the material

NOTE This criterion is generally met by test specimens containing the maximum number of representative volume

elements, compatible with the volume of the crucible, if this number is less than five, several solutions are possible:

a) the test specimens should have an exact number of representative volume elements;

b) the material should be ground to powder and a specimen taken from this powder However this solution will

lead to results which may differ from results obtained on solid test pieces and should be used only if no other

solution is possible;

c) the material should be cut into specimens and a number of similar test specimens should be tested and an

average value determined

4.5 Calibration of calorimeter

4.5.1 General

Calibration of calorimeters, may be done according to two different methods The first consists in dissipating a

known amount of thermal power using a calibrated resistor introduced in the second chamber of the calorimeter In

Trang 8

The calibration factor is the ratio of a known amount of thermal power dissipated in the resistor to the steady state

calorimetric output signal and is measured at temperature T2

NOTE 1 The method using power dissipation in a resistor is limited to1 350 K

NOTE 2 This method can only be used if the sensitivity of the calorimeter is not affected by the filling of the measuring

chamber

4.5.3 Calibration using standard reference material

This calibration is called “drop calibration” A specimen made from a standard reference material with a known

specific heat capacity is dropped according to the test procedures described in section 4.6 (See annex B for

standard reference material) This allows determination of the calibration factor (see annex A)

4.6 Test procedures

NOTE The avoidance of interaction between the test specimen and the calorimetric conditioning and measuring chambers

can require the use of a sealed crucible

4.6.1 Test without a crucible

4.6.1.1 Test with drop calibration

The test without a crucible and with drop calibration is done in the following order:

R, T, R, T, R, T, R

with

R = test of standard reference material, and;

T = test of test specimen

Carry out each test as described in 4.6.3

4.6.1.2 Test with electrical calibration

The test without a crucible and with calibration using power dissipation in a resistor is done in the following order:

 calibration of calorimeter;

 test on three test specimens

Carry out each test as described in 4.6.3

4.6.2 Test with a crucible

4.6.2.1 General

The mass of all empty crucibles used for the test shall not differ by more than 5 %

4.6.2.2 Test with drop calibration

The test with a crucible and with drop calibration is carried out in the following order:

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 9

C, C + R, C + T, C, C + R, C + T, C, C + R, C + T, C

with

C is the test with the empty crucible;

C + R = test of crucible plus standard reference material;

C + T = test of crucible plus test specimen

Carry out each test as described in 4.6.3

4.6.2.3 Test with electrical calibration

The test with a crucible and with calibration using power dissipation in a resistor is done in the following order:

 calibration of calorimeter;

 carry out the following sequence:

C, C + T, C, C + T, C, C + T, C

with

C is the test with the empty crucible;

C + T = testwith crucible plus test specimen

Carry out each test as described in 4.6.3

4.6.3 Description of test

The test piece (test specimen, standard material or empty crucible) and reference material shall be dried at

(110 ± 5) °C until the difference in weight of two successive weighings is lower than 0,2 mg:

 measure the mass when a crucible is not used with an accuracy of ± 0,1 mg or ± 0,1 % whichever is the

smaller;

 when a crucible is used, measure the mass of each assembly dropped, (empty crucible, crucible and standard

reference material, crucible and test specimen);

 place the test piece (test specimen, standard material or empty crucible) in the conditioning chamber at

temperature T1 and wait for a sufficient period (in the order of 15 min), to reach thermal equilibrium of the test

piece with its environment Measure T1 and T2 start recording the calorimetric signal before the test piece is

dropped Drop the test piece Stop the record when the steady state output signal is reached

4.7 Calculations

4.7.1 General

The change in heat Q corresponding to the drop of the test piece is related to the area A under the calorimetric

output signal by the following equation

A K

Q = ⋅

where

K is the calorimeter calibration factor

Trang 10

8

4.7.2 Determination of the calorimetric calibration factor

4.7.2.1 Electrical calibration (see annex A)

A

H

signaloutput

ic calorimetrthe

underarea

dissipatedheat

4.7.2.2 With standard reference material

See annex B

4.7.3 Determination of mean specific heat capacity Cp

The mean specific heat capacity is determined using the following formula:

)

,,

1 2

2 1 i i 2

1 p

(

1

T - T

T T Q m

= T T

C

where

T1 is the initial temperature at which test pieces, are conditioned;

T2 is the calorimeter temperature;

Qi (T1,T2) is the heat variation between T1 and T2;

mi is the mass of the test piece, determined by weighing;

(

1 2

)

p T , T

C mean specific heat capacity between T1 and T2

The subscript i has a different meaning depending on the type of test piece:

 i = c for an empty crucible;

 i = t for a test piece;

 i = t + c for a test piece and crucible

A K C

t

t pt

with crucible

)

)

1 2 t

c t +

c pt

T - (T m

A - (A K

=

C

with

At is the value of integration of calorimetric output signal of test specimen;

Ac is the value of integration of calorimetric output signal of crucible;

Ac+t is the value of integration of calorimetric output signal of test specimen plus crucible

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 11

`,,,,`,,`,,,,,,,``,`,,,``,``-`-`,,`,,`,`,,` -5 Method B - Differential scanning calorimetry

5.1.2 Stepwise heating method

The mean specific heat capacity Cp

(

T1,T2

)

is measured in a temperature interval defined by two isothermal levels

T1 and T2 The heat QE which is necessary to change the temperature from T1 to T2 is determined by integratingthe thermal power PE with respect to time The corresponding heat QE is:

(

1 2

)

c o) 2 1)

p t E

C is the mean specific heat capacity of the test specimen;

Co is the heat capacity of the calorimeter;

Cc is the heat capacity of the crucible

Another experiment for the determination of the base line is performed using an identical imposed heatingsequence with the empty crucible The corresponding heat QB is given by:

B E 2

1 p

T - (T m

Q - Q

= T - T C

5.1.3 Continuous heating method

Temperature is increased linearly versus time at a constant heatingrate ß Using the same notation as in 5.1.2 thethermal power PE supplied at every moment to the system is:

(

t p c o

)

β

t

c m C C C S

= S

K⋅ c c o)

Trang 12

C

t

c t c p

5.2.1 Differential scanning calorimeter

5.2.1.1 There are two types of differential scanning calorimeters operating on power compensation and heatflux principles, both designed to operate under adiabatic conditions

Both comprise two measuring cells housed in a furnace which provides overall system heating One cell containsthe test specimen and its crucible, the other contains an empty crucible only

5.2.1.2 Power compensation type: each cell has an additional heater to compensate for the temperaturevariations from the overall heating programme The power which is supplied to either cell heater to maintain equaltemperatures during heating is measured

5.2.1.3 Heat flux type: power is exchanged between each cell and its respective surrounding, during theheating programme The difference in power exchange between the two cells is measured

5.2.2 Balance, with an accuracy better than 0,1 mg.

5.2.3 Temperature detectors, thermocouples in accordance with EN 60584-1 shall be used for themeasurement of temperature

5.2.4 Data acquisition system, the time duration between two successive measurements shall be less

than 0,5 s

5.3 Standard reference materials, SRM

Standard reference materials shall be used for calibration An example is given in annex B

5.4 Test specimens

The test specimens shall be representative of the material

NOTE This criterion is generally met by test specimens containing the maximum number of representative volumeelements, compatible with the volume of the crucible, if this number is less that 5, several solutions are possible:

a) the test specimens should have an exact number of representative volume elements;

b) the material should be ground to powder and a specimen taken from this powder However this solution willlead to results which may differ from results obtained on solid test specimens and should only be used if noother solution is possible;

c) the material should be cut into pieces and a number of similar test pieces should be tested and an averagevalue determined

5.5 Temperature calibration

A temperature calibration curve for the furnace using the same heating rate as for the determination of the specificheat capacity is established by using the melting points of standard reference materials (see for example annex C)

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 13

`,,,,`,,`,,,,,,,``,`,,,``,``-`-`,,`,,`,`,,` -Thermocouples shall be calibrated in accordance with EN 60584-1.

5.6 Test procedure for the determination of C

p

5.6.1 General

Depending on the necessity to use or not a calibration factor K for the calorimeter, two methods can be used:Method 1: measurements requiring the knowledge of the K factor; in that case, care shall be taken in order toensure that the calibration is valid for all the measurements to be done

NOTE Generally, this can be done by running a test using a test specimen with well-known properties

Method 2: measurements requiring the use of a reference standard material during a series of tests

5.6.2 Method 1: Measurements requiring the knowledge of the K factor

5.6.2.1 Determination of the K factor

The calibration factor K is obtained by electrical calibration It is determined from the ratio of a known amount ofpower dissipated in a resistor to the steady state calorimetric output signal

5.6.2.2 Measurements with the specimen for the determination of the Cp

5.6.2.2.1 General

A series of measurements shall always be referenced to a base line measurement performed under identicalexperimental conditions as the other measurements in the series The type of crucible used depends on the type ofthe test specimen and on the temperature range and shall be the same for the series of measurements The mass

of all empty crucibles used in the series shall not differ by more than 5 %

5.6.2.2.2 Test sequence for the stepwise heating method (see Figure 1)

Generation of the base line:

1) weigh the two empty crucibles to the nearest 0,1 mg;

2) place the two crucibles in the calorimeter;

3) set the calorimeter heating rate, initial and final temperature, and cooling rate;

NOTE Generally the heating rate is in the range 1 K/min to 20 K/min

4) heat to an initial temperature, and wait for the temperature to be stabilised at the initial temperature;5) heat at a constant rate to final temperature of the first step while recording the calorimeter output signaluntil the final temperature is reached and stabilised in order to obtain a base line;

6) repeat 3 to 5 for the number of steps required;

7) cool down to initial temperature;

8) remove the crucibles from the measurement cell

Trang 14

Measurements using a test specimen

Weigh the test specimen and place it in the crucible to be located in the measurement cell Repeat operations 2

to 8 of the above paragraph on generation of the base line Repeat this procedure for a minimum of three testspecimens

5.6.2.2.3 Test sequence for the continuous heating method (see Figure 2)

Generation of the base line:

1) weigh the two empty crucibles to the nearest 0,1 mg;

2) place the two crucibles in the calorimeter;

3) set the calorimeter heating rate, initial and final temperature, and cooling rate;

NOTE Generally the heating rate is in the range 1 K/min to 20 K/min

4) heat to an initial temperature, and wait for the temperature to be stabilised at the initial temperature;5) heat at a constant rate to final temperature of the first step while recording the calorimeter output signaluntil the final temperature is reached and stabilised in order to obtain a base line;

6) cool down to initial temperature;

7) remove the crucibles from the measurement cell

Measurements using a test specimen

Weigh the test specimen and place it in the crucible to be located in the measurement cell Repeat operations 2

to 7 of the above paragraph on generation of the base line Repeat this procedure for a minimum of three testspecimens

5.6.3 Method 2: Measurements requiring the use of a reference standard material (SRM)

5.6.3.1 General

The two methods described in the following paragraphs 5.6.3.2 and 5.6.3.3 require each:

 measurements with two empty crucibles for the generation of the baseline;

 measurements with one empty crucible and one crucible with the SRM;

 measurements with one empty crucible and one crucible with the test specimen

5.6.3.2 Test sequence for the stepwise heating method (see Figure 3)

Generation of the base line

1) weigh the two empty crucibles to the nearest 0,1 mg;

2) place the two crucibles in the calorimeter;

3) set the calorimeter heating rate, initial and final temperature, and cooling rate;

NOTE Generally the heating rate is in the range 1 K/min to 20 K/min

4) heat to an initial temperature, and wait for the temperature to be stabilised at the initial temperature;

Copyright British Standards Institution

Reproduced by IHS under license with BSI - Uncontrolled Copy

Document provided by IHS Licensee=Bureau Veritas/5959906001, 11/08/2004 00:17:52 MST Questions or comments about this message: please call the Document

Trang 15

`,,,,`,,`,,,,,,,``,`,,,``,``-`-`,,`,,`,`,,` -5) heat at a constant rate to final temperature of the first step while recording the calorimeter output signaluntil the final temperature is reached and stabilised in order to obtain a base line;

6) repeat 3 to 5 for the number of steps required;

7) cool down to initial temperature;

8) remove the crucibles from the measurement cell

Measurement with a test specimen or with a SRM

Weigh the test specimen or the SRM and place it in the crucible to be located in the measurement cell Repeatoperations 2 to 8 of the above paragraph on generation of the base line Repeat this procedure for a minimum ofthree test specimens

5.6.3.3 Test sequence for continuous heating method (see Figure 4)

Generation of the base line

1) weigh the two empty crucibles to the nearest 0,1 mg;

2) place the two crucibles in the calorimeter;

3) set the calorimeter heating rate, initial and final temperature, and cooling rate;

NOTE Generally the heating rate is in the range 1 K/min to 20 K/min

4) heat to an initial temperature, and wait for the temperature to be stabilised at the initial temperature;

5) heat at a constant rate to final temperature of the first step while recording the calorimeter output signaluntil the final temperature is reached and stabilised in order to obtain a base line;

6) cool down to initial temperature;

7) remove the crucibles from the measurement cell

Measurements using a test specimen or an SRM

Weigh the test specimen or the SRM and place it in the crucible to be located in the measurement cell Repeatoperations 2 to 6 of the above paragraph on generation of the base line Repeat this procedure for a minimum ofthree test specimens

5.7 Calculation of results

5.7.1 Method requiring the knowledge of the K factor

5.7.1.1 Stepwise heating method (see Figure 1)

NOTE The use of a computer with an adapted software greatly simplifies data acquisition and treatment, and its use isrecommended

For the considered temperature interval the shaded areas, A, are the integrals of the output signal, s, with respect

c = (C + C (T - T A

K

Ngày đăng: 13/04/2023, 21:52