Designation C1402 − 17 Standard Guide for High Resolution Gamma Ray Spectrometry of Soil Samples1 This standard is issued under the fixed designation C1402; the number immediately following the design[.]
Trang 1Designation: C1402−17
Standard Guide for
This standard is issued under the fixed designation C1402; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1 Scope
1.1 This guide covers the identification and quantitative
determination of gamma-ray emitting radionuclides in soil
samples by means of gamma-ray spectrometry It is applicable
to nuclides emitting gamma rays with an approximate energy
range of 20 to 2000 keV For typical gamma-ray spectrometry
systems and sample types, activity levels of about 5 Bq (135
pCi) are measured easily for most nuclides, and activity levels
as low as 0.1 Bq (2.7 pCi) can be measured for many nuclides
It is not applicable to radionuclides that emit no gamma rays
such as the pure beta-emitting radionuclides hydrogen-3,
carbon-14, strontium-90, and becquerel quantities of most
transuranics This guide does not address the in situ
measure-ment techniques, where soil is analyzed in place without
sampling Guidance for in situ techniques can be found in Ref
( 1 ) and ( 2 ).2 This guide also does not discuss methods for
determining lower limits of detection Such discussions can be
found in Refs (3 ), ( 4 ), ( 5 ), and ( 6 ).
1.2 This guide can be used for either quantitative or relative
determinations For quantitative assay, the results are expressed
in terms of absolute activities or activity concentrations of the
radionuclides found to be present This guide may also be used
for qualitative identification of the gamma-ray emitting
radio-nuclides in soil without attempting to quantify their activities
It can also be used to only determine their level of activities
relative to each other but not in an absolute sense General
information on radioactivity and its measurement may be
found in Refs (7 ), ( 8 ), ( 9 ), ( 10 ), and ( 11 ) and Standard Test
MethodsE181 Information on specific applications of
gamma-ray spectrometry is also available in Refs (12 ) or ( 13 ) Practice
D3649may be a valuable source of information
1.3 The values stated in SI units are to be regarded as
standard No other units of measurement are included in this
standard
1.4 This standard may involve hazardous material,
operations, and equipment This standard does not purport to
address all of the safety concerns, if any, associated with its use It is the responsibility of the user of this standard to establish appropriate safety and health practices and deter-mine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in
accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for the Development of International Standards, Guides and Recom-mendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2 Referenced Documents
2.1 ASTM Standards:3
C859Terminology Relating to Nuclear Materials
C998Practice for Sampling Surface Soil for Radionuclides
C999Practice for Soil Sample Preparation for the Determi-nation of Radionuclides
C1009Guide for Establishing and Maintaining a Quality Assurance Program for Analytical Laboratories Within the Nuclear Industry
D3649Practice for High-Resolution Gamma-Ray Spectrom-etry of Water
D7282Practice for Set-up, Calibration, and Quality Control
of Instruments Used for Radioactivity Measurements
E181Test Methods for Detector Calibration and Analysis of Radionuclides
IEEE/ASTM-SI-10Standard for Use of the International System of Units (SI) the Modern Metric System
2.2 ANSI Standards:4
N13.30Performance Criteria for Radiobioassay
N42.14Calibration and Use of Germanium Spectrometers for the Measurement of Gamma-Ray Emission Rates of Radionuclides
N42.23American National Standard Measurement and As-sociated Instrumentation
IEEE-325Standard Test Procedures for Germanium Gamma-Ray Detectors
1 This guide is under the jurisdiction of ASTM Committee C26 on Nuclear Fuel
Cycle and is the direct responsibility of Subcommittee C26.05 on Methods of Test.
Current edition approved June 1, 2017 Published July 2017 Originally approved
in 1998 Last previous edition approved in 2009 as C1402 – 04 (2009) DOI:
10.1520/C1402-17.
2 The boldface numbers in parentheses refer to the list of references at the end of
this standard.
3 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
4 Available from American National Standards Institute (ANSI), 25 W 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959 United States
Trang 23 Terminology
3.1 Except as otherwise defined herein, definitions of terms
are as given in Terminology C859
4 Summary of Guide
4.1 High-resolution germanium detectors and multichannel
analyzers are used to ensure the identification of the
gamma-ray emitting radionuclides that are present and to provide the
best possible accuracy for quantitative activity determinations
4.2 For qualitative radionuclide identifications, the system
must be energy calibrated For quantitative determinations, the
system must also be shape and efficiency calibrated The
standard sample/detector geometries must be established as
part of the efficiency calibration procedure
4.3 The soil samples typically need to be pretreated (for
example, dried), weighed, and placed in a standard container
For quantitative measurements, the dimensions of the container
holding the sample and its placement in front of the detector
must match one of the efficiency-calibrated geometries If
multiple geometries can be selected, the geometry chosen
should reflect the detection limit and count rate limitations of
the system Qualitative measurements may be performed in
non-calibrated geometries
4.4 The identification of the radionuclides present is based
on matching the energies of the observed gamma rays in the
spectrum to computer-based libraries of literature references
[see Refs (14 ), ( 15 ), ( 16 ), ( 17 ), or ( 18 )] The quantitative
determinations are based on comparisons of observed count
rates to previously obtained counting efficiency versus energy
calibration data, and published branching ratios for the
radio-nuclides identified
5 Significance and Use
5.1 Gamma-ray spectrometry of soil samples is used to
identify and quantify certain gamma-ray emitting
radionu-clides Use of a germanium semiconductor detector is
neces-sary for high-resolution gamma-ray measurements
5.2 Much of the data acquisition and analysis can be
automated with the use of commercially available systems that
include both hardware and software For a general description
of the typical hardware in more detail than discussed in Section
7, see Ref (19 ) For best practices on set-up, calibration, and
quality control of utilized spectrometry systems, see Practice
D7282
5.3 Both qualitative and quantitative analyses may be
per-formed using the same measurement data
5.4 The procedures described in this guide may be used for
a wide variety of activity levels, from natural background
levels and fallout-type problems, to determining the
effective-ness of cleanup efforts after a spill or an industrial accident, to
tracing contamination at older production sites, where wastes
were purposely disposed of in soil In some cases, the
combination of radionuclide identities and concentration ratios
can be used to determine the source of the radioactive
materials
5.5 Collecting samples and bringing them to a data acqui-sition system for analysis may be used as the primary method
to detect deposition of radionuclides in soil For obtaining a representative set of samples that cover a particular area, see Practice C998 Soil can also be measured by taking the data acquisition system to the field and measuring the soil in place (in situ) In situ measurement techniques are not discussed in this guide
6 Interferences
6.1 In complex mixtures of gamma-ray emitters, the degree
of interference of one nuclide in the determination of another
is governed by several factors Interference will occur when the photopeaks from two separate nuclides overlap within the resolution of the gamma-ray spectrometer Most modern analy-sis software can deconvolute multiplets where the separation of
any two adjacent peaks is more than 0.5 FWHM (see Refs (20 ) and (21 )) For peak separations that are smaller than 0.5
FWHM, most interference situations can be resolved with the
use of automatic interference correction algorithms (22 ).
6.2 If the nuclides are present in the mixture in very unequal radioactive portions and if nuclides of higher gamma-ray energy are predominant, the interpretation of minor, less energetic gamma-ray photopeaks becomes difficult due to the high Compton continuum and backscatter
6.3 True coincidence summing (also called cascade sum-ming) occurs regardless of the overall count rate for any radionuclide that emits two or more gamma rays in coinci-dence Cobalt-60 is an example where both a 1173-keV and a 1332-keV gamma ray are emitted from a single decay If the sample is placed close to the detector, there is a finite probability that both gamma rays from each decay interact within the resolving time of the detector resulting in a loss of counts from both full energy peaks Coincidence summing and the resulting losses to the photopeak areas can be considerable (>10 %) before a sum peak at an energy equal to the sum of the coincident gamma-ray energies becomes visible Coincidence summing and the resulting losses to the two individual photo-peak areas can be reduced to the point of being negligible by increasing the source to detector distance or by using a small detector Coincidence summing can be a severe problem if a well-type detector is used See Test MethodsE181and (7 ) for
more information
6.4 Random summing is a function of count rate (not dead time) and occurs in all measurements The random summing rate is proportional to the total count squared and to the resolving time of the detector and electronics For most systems, uncorrected random summing losses can be held to less than 1 % by limiting the total counting rate to less than
1000 counts/s However, high-precision analyses can be per-formed at high count rates by the use of pileup rejection circuitry and dead-time correction techniques Refer to Test Methods E181for more information
7 Apparatus
7.1 Germanium Detector Assembly—The detector should
have an active volume of greater than 50 cm3, with a full width
at one half the peak maximum (FWHM) less than 2.0 keV for
Trang 3the cobalt-60 gamma ray at 1332 keV, certified by the
manufacturer A charge-sensitive preamplifier should be an
integral part of the detector assembly
7.2 Sample Holder Assembly—As reproducibility of results
depends directly on reproducibility of geometry, the system
should be equipped with a sample holder that will permit using
reproducible sample/detector geometries for all sample
con-tainer types that are expected to be used at several different
sample-to-detector distances
7.3 Shield—The detector assembly should be surrounded by
a radiation shield made of material of high atomic number
providing the equivalent attenuation of 100 mm (or more in the
case of high background radiation) of low-activity lead It is
desirable that the inner walls of the shield be at least 125 mm
distant from the detector surfaces to reduce backscatter and
annihilation radiation If the shield is made of lead or has a lead
liner, the shield should have a graded inner shield of
appropri-ate mappropri-aterials, for example, 1.6 mm of cadmium or tin-lined
with 0.4 mm of copper, to attenuate the induced 88-keV lead
fluorescent X-rays The shield should have a door or port for
inserting and removing samples The materials used to
con-struct the shield should be prescreened to ensure that they are
not contaminated with unacceptable levels of natural or
man-made radionuclides The lower the desired detection capability,
the more important it is to reduce the background For very low
activity samples, the detector assembly itself, including the
preamplifer, should be made of carefully selected low
back-ground materials
7.4 High-Voltage Power/Bias Supply—The bias supply
re-quired for germanium detectors usually provides a voltage up
to 65000 V and 1 to 100 µA The power supply should be
regulated to 0.1 % with a ripple of not more than 0.01 % Noise
caused by other equipment should be removed with r-f filters
and power line regulators
7.5 Amplifier—A spectroscopy amplifier which is
compat-ible with the preamplifier If used at high count rates, a model
with pile-up rejection should be used The amplifier should be
pole-zeroed properly prior to use
7.6 Data Acquisition Equipment—A multichannel
pulse-height analyzer (MCA) with a built-in or stand-alone
analog-to-digital converter (ADC) compatible with the amplifier
output and pileup rejection scheme The MCA (hardwired or a
computer-software-based) collects the data, provides a visual
display, and stores and processes the gamma-ray spectral data
The four major components of an MCA are: ADC, memory,
control, and input/output The ADC digitizes the analog pulses
from the amplifier The height of these pulses represents energy
deposited in the detector The digital result is used by the MCA
to select a memory location (channel number) which is used to
store the number of events which have occurred at the energy
The MCA must also be able to extend the data collection time
for the amount of time that the system is dead while processing
pulses (live time correction)
7.7 Count Rate Meter—It is useful but not mandatory to
have a means to measure the total count rate for pulses above
the amplifier noise during the measurement If not provided by
the MCA, a separate count rate meter may be used for this purpose In the absence of a rate meter, count rates that are too high to provide reliable results may also be detected by monitoring the system dead time or peak resolution, or both
7.8 Pulser—Required only if random summing effects are
corrected with the use of a stable pulser (23 ) and ( 24 ).
7.9 Computer—Most modern gamma-ray spectrometers are
equipped with a computer for control of the data acquisition as well as automated analysis of the resulting spectra Such computer-based systems are readily available from several commercial vendors Their analysis philosophies and capabili-ties do differ from each other somewhat See ANSI N42.14 for
a series of tests on how to tell if a particular gamma-spectrometry software package has adequate analysis capabili-ties In addition to the analysis capabilities, it is important to consider the overall user interface and architecture of the software For small-scale operations, a few samples per week,
a user interface that requires a lot of user intervention is sufficient For larger-scale operations, with hundreds of samples per week on multiple detectors, a software package that permits some kind of batch processing and automated operation is recommended
8 Container for a Test Sample
8.1 Sample holders and containers must have a reproducible geometry Considerations include commercial availability, ease
of use and disposal, and the containment of radioactivity for protection of the working environment, personnel, and the gamma-ray spectrometer from contamination For small soil samples (up to a few grams), plastic bottles are convenient containers, while large samples (up to several kilograms), which require greater sensitivity, are frequently packaged in Marinelli beakers For analyzing low-energy gamma rays at close geometries, the consistency of the wall thickness of the sample container facing the detector becomes an important factor in the variability of the analysis results
8.2 Measurements may require precautions to prevent the loss of volatile radionuclides For example, the direct determi-nation of radium-226 in soil by the measurement of the 609-keV gamma ray of bismuth-214 assumes secular equilib-rium between radium-226 and its bismuth-214 progency and that the radon-222 daughter was not lost from the sample 8.3 A beta absorber consisting of about 6 mm of aluminum, beryllium, or plastic should be placed between the detector and sample for samples that have significant quantities of high-energy beta emitters
9 Calibration and Standardization
9.1 Overview:
9.1.1 Commission and operate the instrumentation and de-tector in accordance with the manufacturer’s instructions and best practices such as may be contained in Practice D7282 Initial set-up includes all electronic adjustments to provide constant operating conditions consistent with the application and life expectancy of the calibrations The analog-to-digital converter gain and range, amplifier gain, and zero-level must
be adjusted to yield an optimum energy calibration Both the
Trang 4energy and efficiency calibration must be accomplished with
radioactive sources covering the entire energy range of interest
(6, 7 and Test MethodsE181) Subsequent efficiency
calibra-tions and source analyses are performed with the same gain
settings and the same high-voltage setting Prepare efficiency
calibration standards by weighing an appropriate amount of a
radionuclide standard solutions containing 100 to 10 000 Bq
each onto a soil matrix in an appropriate container, drying it,
and mixing thoroughly Standardized dried soil and bottom
sediment are also available from the U.S National Institute of
Standards and Technology (NIST) or other appropriate sources
which can be used directly or diluted with ambient soil to a
measured weight or volume Prepare blank sources containing
the same quantity of unspiked soil to account for any naturally
occurring radionuclides that may be present Commercially
available epoxy soil-equivalent standards with an appropriate
mixture of radionuclides can also be used It should be noted
that soils that contain high atomic number materials will
significantly alter the expected self-attenuation
9.1.2 Follow the manufacturer’s instructions, limitations,
and cautions for the setup and the preliminary testing for all of
the spectrometry equipment to be used in the analysis This
equipment would include, as applicable, detector, power
supplies, preamplifiers, amplifiers, multichannel analyzers, and
computing systems For example, ensure that the detector has
had ample time (typically 6–8 h) to cool down after the first
filling with liquid nitrogen before turning on the high voltage
Also, ensure that the high-voltage bias supply is set for the
recommended operating voltage and the correct polarity
9.1.3 Place an appropriate weight of standardized dried soil
in an appropriate soil matrix in a sealed container and place the
container at a desirable and reproducible source-to-detector
distance The standard (traceable to a designated standards
organization) should provide enough counts in each calibration
peak (typically 20 000 or more, see Test Methods E181 or
ANSI N42.14) in a reasonable amount of time (4–12 h) In all
radionuclide measurements, the volumes, shape, and physical
and chemical characteristics of all the samples and standards
and their containers must be as identical as practical for the
most accurate results For situations where it is not possible or
practical to produce standards that are identical to the samples,
standard matrices that are different from the sample matrices
have been found to provide acceptable results when coupled
with attenuation correction methods
9.2 Energy and Shape Calibration:
9.2.1 The energy and shape calibration (the peak
gamma-ray energy versus channel number of the multichannel analyzer
and peak shape versus the peak gamma-ray energy) of the
detector system is determined at a specific gain setting
(typi-cally 0.5 keV/channel) using standards containing known
radionuclides The peak shape calibration may involve only
calculating the peak resolution (full-width-at-half-maximum,
or FWHM), or include other, nonsymmetrical components as
well The standards should be in sealed containers and should
emit at least eight different gamma-ray energies covering the
range of interest, usually from 20 to 2000 keV, in order to test
for system linearity If the calibration is performed with only
the radionuclides of interest, fewer gamma-ray energies can
also be used Energy and shape calibration can be performed without NIST traceable sources
9.2.2 Verify the radionuclide purity of the standard periodi-cally to ensure against accidental contamination or the pres-ence of long-lived impurities by comparing the observed gamma rays with the data published in the literature Careful adherence to precautions and certificate calibration instructions are necessary when using the calibration standards
9.2.3 Calibrate a multichannel analyzer for energy, shape, and efficiency to cover the energy range or interest If the range
of interest is from 20 to 2000 keV, adjust the gain of the system until the centroid of the cesium-137 photopeak, 661.6 keV, is about one-third full-scale Leaving the gain constant, locate at least three other photopeaks of different energies within the energy range of interest Determine and record the peak centroid for each of the four gamma energies A linear relationship between the gamma-ray energies and their channel numbers should be observed if the equipment is operating properly Calculate the slope and intercept of the line using a least-squares calculation If the spectrometry system is computerized, follow the appropriate manufacturer input in-structions for the determination of the slope and intercept 9.2.4 If the system is being calibrated with the radionuclides
of interest, fewer lines may be used for calibration and the linearity of the MCA is not an issue as long as the peaks of interest are identified and quantified consistently
9.3 Effıciency Calibration:
9.3.1 Efficiency calibration must be performed with sources that are traceable to a national standards laboratory, such as NIST A mixed gamma-ray standard for both energy and efficiency calibration containing Am-241, Cd-109, Co-57, Ce-139, Hg-203, Sn-113, Sr-85, Cs-137, Y-88, and Co-60 is available from many commercial source manufacturers who provide NIST traceable sources The gamma-ray energies of this mixed standard as well as some other commercially available NIST traceable radionuclides that are suitable for efficiency calibration (and energy and shape calibration) are shown in Table 1 As another example, an antimony-125/ europium-154,155 mixture from NIST (SRM 4275B or its replacement) has 19 major photopeaks between 100 and 1600 keV
TABLE 1 Radionuclides Useful for Energy Calibration
Trang 59.3.2 For environmental or low-activity samples (0.01 to 1
Bq/g), typically, 300 to 500 g of prepared soil are used If a
fixed volume is used, the mass will vary according to the
density High-density samples may cause significant
self-absorption of low-energy gamma rays and degrade the
higher-energy gammas Therefore, it is important to calibrate the
detector with standards of the same geometry, composition,
and density, or use appropriate attenuation and geometry
correction algorithms
9.3.3 Accumulate a gamma-ray spectrum using sealed,
cali-brated radioactivity standards until there are approximately
20 000 net counts (see Test MethodsE181or ANSI N42.14) in
each full-energy calibration peak provided that this does not
require an excessive amount of time To achieve reasonable
count times, smaller peak sizes may also be used Compare the
length of the count to the half-life of the radionuclide of
interest If the duration of the count is a significant portion of
the half-life (>10 %), a correction factor must be applied for
decay during the count Refer to Standard Test MethodsE181
or Ref (6 ) for additional information To correct the results to
the start of the measurement due to decay during spectrum
acquisition, use the following equation:
where:
K = multiplicative corrective for peak area or intensity,
λ = nuclide decay constant (in the same time units as T),
and
T = data collection elapsed clock time (that is, real time, not
live time)
9.3.4 The equation shown in 9.3.3 is not sufficient if the
count rate changes significantly during data acquisition, as
might happen if a short-lived nuclide is the main source of the
activity Other methods, such as the Virtual Pulser (25 ) and add
“N” Method (26 ) may be used for a varying count rate
situation
9.3.5 Correct the radioactivity standard source gamma-ray
emission rate for the decay from the time of standardization to
the start of data acquisition Many commercial and
noncom-mercial data analysis software packages will do this
automati-cally as part of the efficiency calibration
9.3.6 Calculate the full-energy peak efficiency, E f, as
fol-lows:
E f 5N p
where:
E f = full-energy peak efficiency (counts recorded per
gamma ray emitted),
N p = net gamma-ray count rate in the full-energy peak of
interest (counts per second), and
N g = gamma-ray emission rate (gamma rays per second)
9.3.7 If the standard source is calibrated for activity rather
than emission rate, the gamma-ray emission rate is given as
follows:
where:
A = number of nuclear decays per second, and
P g = emission probability for the gamma ray per nuclear decay
9.3.8 Many modern spectrometry systems are computerized, and the determination of the gamma-ray effi-ciencies is accomplished automatically at the end of an appropriate counting interval Refer to the manufacturer’s instructions for specific input requirements It is necessary for the user to determine the basis of the system analysis and its limitations
9.3.9 Plot the values for the full-energy peak efficiency (as determined in 9.3.6) versus gamma-ray energy The plot will allow the determination of efficiencies at energies for which standards are not available Many computerized systems pro-vide such plotting capabilities as part of the overall function-ality of the system Computerized systems also typically provide a variety of different calculational models to automati-cally calculate the efficiencies at any energy The calibration curve, regardless of whether it is calculated by hand or by a computerized system, generally should not be used for peak energies beyond the first and last calibration points If it is necessary to use the calibration curve in such a manner, one must pay particular attention to establishing appropriate uncer-tainties for any peak energies outside the calibration range Alternatively, calibrate using standards of the radionuclides of interest to obtain direct calibration factors for them without establishing an efficiency curve
10 Measurement Control
10.1 A properly run laboratory must have a measurement control program to verify that the detection system is in calibration See GuideC1009or Ref (27 ) for further guidance
on laboratory measurement control programs
10.2 As a minimum, the following periodic checks should
be made
10.2.1 Each day or prior to each measurement, energy, resolution, and efficiency response should be checked using at least two different gamma-ray energies If the energy calibra-tion slope and intercept are essentially unchanged, the energy calibration data are assumed to remain valid If an appreciable change in the slope or intercept is evident, the reason should be determined and corrected
10.2.2 Once the efficiencies for the various sample sizes, matrices, and source-to-detector distances have been established, it is not necessary to repeat the process unless there is a change in resolution or system configuration, or a new sample size, matrix, or geometry is added, or the detector has failed and has been returned to service after it has been repaired However, a complete check of the efficiency and energy calibrations should be done periodically (typically annually) Similar tests should be used to determine loss of
resolution or efficiency (28 ).
10.2.3 Ideally, a measurement of the room background should be made before and after any series of determinations 10.2.4 A measurement of a standard or sample with a known concentration to provide a measurement bias check Radioac-tive decay of the standard must be taken into account
Trang 610.2.5 Periodic replicate measurements of a standard or
sample to determine that the precision reported by the analysis
method, such as a computer program, is appropriate
10.3 It is recommended that control charts and other
peri-odic statistical analysis of the precision and bias data be used
11 Sample Measurements
11.1 After the spectrometer system has been set up and the
energy and efficiency calibrations performed, unknown
samples can be measured
11.2 Soil samples are collected by methods in accordance
with Practice C998 and prepared for analysis in accordance
with PracticeC999 An appropriate aliquot of soil is transferred
to the sample container (8.1) and positioned in the same
manner as was done during system calibration (Section9)
11.3 Measure the activity of the sample for a period of time
long enough to acquire a gamma-ray spectrum which will meet
the minimum acceptable counting uncertainty for the
radionu-clides of interest
12 Calculation
12.1 Ambient background peak areas must be subtracted if
their contributions are not negligible In many experiments, the
background may not affect the results but is still monitored to
ensure the integrity of the system The method presented here
is not the only acceptable one but is compatible with available
computational hardware and should be used to verify the
validity of commercial software
12.2 The underlying aim of this method is to subtract the
continuum or baseline from the spectral data where it underlies
a photopeak of interest For operator-directed calculations, the
choice of the baseline level may be straightforward The
simplest way, using a plot of the spectral data, is to draw a
straight line, using judgement and experience, that best
de-scribes the baseline The baseline data can be read directly
from the plot and subtracted A variety of computer programs
have accomplished this but details are not included in this
guide
12.3 Photopeaks lying on a sloping baseline, or one with
curvature, will be analyzed, regardless of method, with
in-creased uncertainty Use of data from these peaks should be
limited to those cases where there is no other alternative
Photopeaks that overlap with each other will also increase the
uncertainty of the final result In the case where use of
overlapping peaks cannot be avoided and software programs
are not available, the experimenter may estimate the areas by
assuming that the ratio of the peak areas is equal to the ratio of
the peak heights, but this may introduce a sizable error
Computer programs separating overlapping peaks with varying
degrees of success may be found in Refs (29 ), ( 30 ) The current
quality of peak reduction and deconvolution software available
from the major counting system manufacturers is adequate for
most situations
12.4 The radioactive decay process is governed by Poisson
statistics In Poisson statistics, the variance in N accumulated
events acquired with a detector is simply N The standard
deviation is the square root of the variance
12.5 The areas of well-resolved spectral peaks can be determined by summing the data above the underlying con-tinuum For narrow peaks, the continuum can be well-approximated by interpolation of the background line The peak area can then be calculated from the following equation:
C 5 P 2F N p
2 S B1
N B11
B2
where:
C = net peak area (counts),
P = total counts in the peak including the
continuum,
N p = number of channels summed for P,
B1 and B2 = number of counts in continuum regions on
either side of peak, and
N B1 and N B2 = number of channels summed to determine B1
and B2, respectively.
12.6 The variance in quantity f that is a function of n independent variables x is approximated by the following
equation:
σ 2~f!5(i51
n
S αf
αx iD2
12.7 Hence, the propagated variance of the area C is given
by the following expression (remembering that the variance of
P = P, variance of B1 = B1, and variance of B2 = B2):
σ 2~C!5 P1F S N p
2·N B1D2
·B1G1F S N p
2·N B2D2
·B2G (6)
where:
σ(C) = standard deviation of Peak Area C.
12.8 In order to determine radionuclide activity concentrations, the photopeak areas, corrected for background and interferences, are divided by the count time and efficiency for the energy of the gamma ray being calculated to yield gammas per second for the peak of interest If, as is the case for some radionuclides, the gamma-ray abundance is 100 %, division by the detector efficiency converts counts per second for the photopeak of interest to decays per second (bequerels) for the nuclide If not, the gammas per second are converted to disintegrations per second by dividing the gammas per second
by the gammas per disintegration, for the nuclide and photo-peak of interest The results are then corrected for sampling or decay, or both, as demanded by the application The activity of
a particular radionuclide may be calculated using the following equation:
where:
A = activity concentration, Bq/g,
ε = efficiency of the spectrometer for the gamma ray of
interest,
K µ = correction factor to accommodate for the attenuation
in the sample matrix compared to the attenuation in the matrix of the standard source,
B = number of gamma rays emitted per decay,
e –λTw = decay correction,
Trang 7W = weight of sample, g,
T = count time, s, (live time), and
T w = decay time
The energies, half-lives, and gammas per disintegration for
typical radionuclides that might be present in soil samples are
available in Refs (14-18 ).
12.8.1 The activity of the ith radionuclide can also be
calculated as follows:
A i 5 k i C
where:
k i = calibration factor for the ith radionuclide.
13 Precision and Bias
13.1 Precision:
13.1.1 Precision of the method is influenced by random
counting uncertainties, and interferences in the spectra of the
individual components with each other as well as the sampling
uncertainty The more complex the spectrum, the greater are
the errors, and in general, major components can be determined
more precisely than minor ones Precision may be improved
with increased counting time and by taking as large a sample of
the soil volume being analyzed as possible Variations in
sample vial geometry and positioning will affect the precision
of the measurement as well
13.1.2 To illustrate typical precision of a gamma-ray
spec-trometry system, a check source was counted ten consecutive
times without removing the source from the detector system
The ten single operator counts of ten minutes each provide a
measure of system repeatability The results are listed inTable
2
13.1.3 The same source was used and another set of 10
measurements of 10 min each were made on successive days to
capture more sources of variation This process involved the
normal day-to-day system operational checks and should
provide a measure of the variability of operating procedures
The results are listed inTable 3
13.2 Bias:
13.2.1 The calibration of standard sources, including errors
introduced in using a standard radioactive solution or aliquot
thereof, to prepare a working standard for bias correction may
result in a bias The full-energy peak efficiency at a given energy determined from the calibration function may introduce
a bias
13.2.2 The single-operator bias of a gamma-ray spectrom-etry system was estimated by measuring the NIST Rocky Flats Soil Number 1 (SRM 4353) and the NIST River Sediment (SRM 4350 B) six different times The results are shown in
Table 4
13.3 Sources of Error:
13.3.1 Variation of the physical geometry of the sample and its relationship with the detector will produce both qualitative and quantitative variations in the gamma-ray spectrum To adequately account for these geometry effects, efficiency cali-brations (and occasionally also energy calicali-brations) should be designed to duplicate all conditions including source-to-detector distance, sample shape and size, and sample matrix When it is not possible to have a calibration source that duplicates the sample matrix, the difference between the calibration matrix and the sample matrix as well as its height can be corrected for by a transmission measurement and a calculation and use of a geometry dependent attenuation factor
(see Ref (31 )).
13.3.2 Since some spectrometry systems are calibrated for various size sources at many different source-to-detector distances, a wide range of activity levels can be measured by the same detector For high-activity samples (for example,
>106Bq), which may have resulted from a spill or accident, a large source-to-detector distance (for example >1 m) may be used The large source-to-detector distance for high-activity samples will reduce the overall rate and thus minimize the random summing problems in the spectrum A larger source-to-detector distance will also remove coincidence summing, if present, regardless of count rate
13.3.3 Electronic problems, such as loss of resolution and random summing, may be minimized by keeping the overall count rate below 2000 counts/s For most soil samples, a high count rate is not a problem Some care may be needed in preparing or purchasing calibration standards so that their count rate in the measurement geometries stays below the desired limit Total counting time is governed by the radioac-tivity of the sample, the detector-to-source distance, and the acceptable Poisson counting distribution uncertainty
TABLE 2 Precision of Repeated Gamma-Ray Measurements in Counts per Minute (CPM) Without Removing the Sample Between
Measurements
Measurement
Trang 813.3.4 The density of the sample is another factor that can
affect quantitative results Errors from this source can be
avoided by preparing the standards for calibration in matrices
with a composition and density comparable to the sample being
analyzed The important factor is the linear attenuation which
is the product of the density and the mass attenuation
Attenuation correction methods to correct for differences in the
densities of the calibration source and actual samples, such as
the use of a transmission source, are acceptable to use if they
can be demonstrated to give good results
14 Keywords
14.1 coincidence summing; gamma-ray; high-purity
germa-nium; HPGe; photopeak; Poisson; radionuclides; shield; soil
TABLE 3 Precision of Reproducibility of Gamma-Ray Measurements in Counts per Minute (CPM) While Removing the Sample from the
Measurement Position Between Measurements
Measurement
Trang 9(1) Beck, H L., DeCampo, J., and Gogolak, C V., “In-Situ Ge(Li) and
NaI(Tl) Gamma-Ray Spectrometry,” USDOE Report HASL-258,
1972.
(2) Miller, K M., Shebell, P., and Klemic, G A., “In-Situ Gamma-Ray
Spectrometry for the Measurement of Uranium in Surface Soils,”
Health Physics, Vol 67, 1994, p 140.
(3) Currie, L A., “Limits for Qualitative Detection and Quantitative
Determination,” Analytical Chemistry, Vol 40, 1968, p 586.
(4) Currie, L A., “Lower Limit of Detection: Definition and Elaboration
of a Proposed Position for Radiological Effluent and Environmental
Measurements,” NUREG/CR-4007, 1984.
(5) ANSI N13.30-1986, Performance Criteria for Radiobioassay.
(6) Brodsky, A., “Accuracy and Detection Limits for Bioassay
Measure-ments in Radiation Protection - Statistical Considerations,”
NUREG-1156, 1986.
(7) Debertin, K., and Helmer, R G., Gamma- and X-Ray Spectrometry
with Semiconductor Detectors, North-Holland Publishing Co., 1988.
(8) Parker, J L., “General Topics in Passive Gamma-ray Assay,” Passive
Nondestructive Assay of Nuclear Materials, (D Reilly, N Ensslin, H.
Smith Jr., and S Kreiner, eds) NUREG/CR-5550, 1991, p 95.
(9) Knoll, G F., Radiation Detection and Measurement, 2nd ed., John
Wiley & Sons, New York, 1989.
(10) Handbook of Radioactive Measurement Procedures, NCRP Report
No 58, 1985, 2nd ed., available from the National Council on
Radiation Protection and Measurements, 7910 Woodmont.,
Washington, DC 20014.
(11) ANSI N42.14-1991, Calibration and Use of Germanium
Spectrom-eters for the Measurement of Gamma-Ray Emission Rates of
Radionuclides.
(12) Proceedings of Computers in Activation Analysis and Gamma-Ray
Spectroscopy, held in Mayaguez, Puerto Rico, April 30-May 4, 1978.
Available from the U S National Technical Information Center as
CONF-2780421, 1979.
(13) Helmer, R G., Putnam, M H., and McCullagh, C M., “Nuclide
Activities Determined from Gamma-Ray Spectra from Ge Detectors:
A review with GAUSS VII as the Example,” Nucleare Instruments &
Methods A242, 1986, p 427.
(14) Kocher, D C., “ Radioactive Decay Data Tables,” DOE/TIC-11026,
1981, available from U S National Technical Information Service.
(15) Browne, E., and Firestone, R B., (V S., Shirley, ed.), Table of
Radioactive Isotopes, John Wiley & Sons, New York, 1986.
(16) Erdtmann, G., and Soyka, W., The Gamma Rays of the
Radionuclides, Verlag Chemie, Weinhem, New York, 1989.
(17) Heath, R L., “ Gamma-Ray Spectrum Catalogue, Ge(Li) and Si(Li)
Spectrometry,” ANCR-1000-2, Vol 2, 1974.
(18) Firestone, R B., (V S Shirley, ed.), Table of Isotopes, 8th ed, John
Wiley & Sons, New York, 1996.
(19) Parker, J L., “Instrumentation for Gamma-Ray Spectroscopy,”
Passive Nondestructive Assay of Nuclear Materials, (D Reilly, N.
Ensslin, H Smith Jr., and S Kreiner, eds.) NUREG/CR-5550, 1991,
p 65.
(20) Koskelo, M J., and Mercier, M T., “Verification of Gamma Spectroscopy Programs: A Standardized Approach,” Nuclear Instru-ments & Methods Vol A299, 1990, p 318.
(21) Mercier, M T., and Koskelo, M J., “Verification of Gamma
Spectroscopy Programs: Accuracy and Detectability,” Journal of
Radioanalysis & Nuclear Chemistry, Vol 160, 1992, p 233.
(22) Koskelo, M J., and Mercier, M T., “Verification of Gamma Spectroscopy Programs: Multiplet Area Problems and Solutions,”
Journal of Radioanalysis and Nuclear Chemistry, Vol 193, 1995, p.
211.
(23) Wiernik, M., “Normal and Random Pulse Generators for the
Cor-rection of Dead-Time Losses in Nuclear Spectrometry,” Nuclear
Instruments & Methods, Vol 96, 1971, p 325.
(24) Debertin, K., and Schotzig, U., “Limitations of the Pulser Method for
Pile-Up Corrections in Ge(Li)-Spectrometry,” Nuclear Instruments
& Methods, Vol 140, 1977, p 337.
(25) Harms, J., “Automatic Dead-time Correction for Multichannel
Pulse-Height Analyzers at Variable Counting Rates,” Nuclear
Instru-ments & Methods, Vol 53, 1967, p 192.
(26) Westphal, G P., “Real-Time Corrections of Counting Losses in
Nuclear Pulse Spectroscopy,” Journal of Radioanalysis Chemistry,
Vol 70, 1982, p 387.
(27) ANSI N42.23, Measurement Quality Assurance for Radioassay Laboratories.
(28) Gehrke, R J., Cline, J E., and Heath, R L., “Determination of Relative Photopeak Efficiency and System Linearity for Ge(Li)
Gamma-Ray Spectrometers,” Nuclear Instruments & Methods, Vol
91, 1970, p 349.
(29) Routti, J T., and Prussin, S G., “Photopeak Method for the Computer Analysis of Gamma-Ray Spectra from Semiconductor
Detectors,” Nuclear Instruments & Methods, Vol 72, 1969, p 125.
TABLE 4 Results of NIST Rocky Flats Soil Number 1 (SRM 4353) and NIST River Sediment (SRM 4350 B) (Each measured value is
an average of 6 measurements of 1000 min each The estimated measured uncertainty includes only the counting statistics The
NIST uncertainty is the total uncertainty.)
SRM 4353 (Bq/ g) Radionuclide Measured
Value
NIST Value
Ratio Measured/NIST
Cs-137 0.0190 ± 0.0010 0.0176 ± 0.0008 1.08 Ra-226 (Bi-214) 0.0430 ± 0.0030 0.0430 ± 0.0028 1.00 Th-228 (Tl-208) 0.083 ± 0.002 0.0708 ± 0.0036 1.17 Th-232 (Ac-228) 0.0690 ± 0.005 0.0693 ± 0.0035 1.00
SRM 4350 B (Bq/g) Co-60 0.00464 ± 0.0024 0.00464 ± 0.00023 1.00
Cs -137 0.031 ± 0.002 0.0290 ± 0.0018 1.07 Eu-152 0.037 ± 0.003 0.0305 ± 0.0012 1.21 Eu-154 0.0035 ± 0.0022 0.00378 ± 0.00057 0.93 Ra-226 (Bi-214) 0.034 ± 0.002 0.0358 ± 0.0036 0.95
Trang 10(30) Gunnink, R., and Niday, J B., “Computerized Quantitative Analysis
by Gamma-Ray Spectrometry,” Lawerence Livermore Laboratory,
Report UCRL-51061, 1972.
(31) Parker, J L., “Attenuation Correction Procedures,” Passive
Nonde-structive Assay of Nuclear Materials, (D Reilly, N Ensslin, H.
Smith, Jr., and S Kreiner, eds.) NUREG/CR-5550, 1991.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org) Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/