Apart from the commercially used Thermalox, BeOR is a new type of Beryllium Oxide fabricated by a different company in Turkey. As it is shown by other groups, this dosimeter yields two overlapping main dosimetric TL peaks in the temperature region of 100–300 ◦C and one peak at 350 ◦C.
Trang 1Available online 1 February 2022
1350-4487/© 2022 Elsevier Ltd All rights reserved
Recombination pathways in a BeO yielding two main dosimetric TL peaks
aAristotle University of Thessaloniki, Physics Department, Nuclear Physics and Elementary Particles Physics Section, GR, 54124, Thessaloniki, Greece
bInstitute of Nanoscience and Nanotechnology, NCSR Demokritos, GR, 15310, Ag Paraskevi, Athens, Greece
cCondensed Matter Physics Laboratory, Physics Department, University of Thessaly, GR, 35100, Lamia, Greece
A R T I C L E I N F O
Keywords:
BeO R
Activation energy
de-localized and localized pathways
Isothermal decay
Peak shape methods
Fractional glow technique
A B S T R A C T Apart from the commercially used Thermalox, BeOR is a new type of Beryllium Oxide fabricated by a different company in Turkey As it is shown by other groups, this dosimeter yields two overlapping main dosimetric TL peaks in the temperature region of 100–300 ◦C and one peak at 350 ◦C The aim of the present study is the calculation of the activation energies of these two peaks with the use of specific protocols, including the Frac-tional Glow Technique (FGT), an Isothermal Decay signal (PID) and the Peak Shape Methods (PSM) Also, these experimental procedures were conducted in two different readers, namely the Harshaw and the Risø reader, for better repeatability and in order to minimize the error of the calculations Moreover, another goal is to correlate the activation energy values with the recombination pathways that are present in the BeOR and its corresponding peaks Specifically, both peaks have activation energies around 1–1.3 eV in most cases, although there are some kinds of indications for a difference in the pathways, as the high-temperature region TL main peak seems to follow the localized recombination via tunneling, in contrast with the low-temperature region TL main peak which sticks to the delocalized model with the passing via the conduction band
1 Introduction
Beryllium oxide (BeO) is a well-established dosimeter and is highly
used as medical, environmental, and personal dosimeter, due to its
unique characteristics (Nieto, 2016; Ogorodnikov et al., 2016; Tochilin
et al., 1969) First of all, the tissue equivalence (Zeff = 7.14, almost
identical to that of Zeff = 7.42 of the biological tissues) of BeO is
considered ideal for an excessive number of devices in a variety of
medical fields, which use radiation for diagnostics or therapy (Watanabe
et al., 2010) Even if it is described as an insulator, it possess a thermal
conductivity which even exceeds that of some metals This fact ensures a
swift and unfluctuating transfer of heat throughout the stage of the
readout of the TL signal, and a TL sensitivity close to that of the
commonly used TLD-100 (Becker et al., 1970) Additionally, its
annealing is easier and simpler as it does not have peaks at the low
temperature region This tough ceramic material is insoluble in water,
insensitive to most chemical agents, has a high resistance to the
influ-ence of mechanical forces, and can be formed at high temperature
en-vironments into rods or discs (Kiiko et al., 1999), having ready
availability and low cost While being solid, free of dust, BeO is nearly
innocuous and the possible health risks accrued from it being machined,
squashed, or converted to its reactive dispersed form are minimal (Stockinger, 1966)
For the thermoluminescence and dosimetry uses, the Ceramic BeO (product name; Thermalox@ 995, fabricated by Brush Beryllium Co.) is the most common of the alternative commercially used BeO Ceramics, as
it offers quality assurance, an almost stable response regarding the long- term uses and is mostly free of noise signals (Crase and Gammage,
1975) Gammage and Garrison (1974) pointed out Al2+substitutional for Be as the TL trapping entity in Thermalox@ 995, while its elemental composition (ppm) as it is given by the manufacturing company is mainly Si, Mg and Fe, with Ca, Al, Ti, and Cu at lower concentrations Its commercial name is Thermalox 995 BeO (BeOT), and yields one main Thermoluminescence (TL) dosimetric peak at the dosimetric region between the temperatures of 100–250 ◦C as well as one peak at 350 ◦C This type of BeO was investigated, for its OSL (Bulur and Goksu, 1998), LM-OSL (Bulur and Yeltik, 2010) and TR-OSL (Bulur and Saraç, 2013; Bulur, 2014) properties, as well as for the phototransfer and the prop-erties of the TL signals (Bulur, 2007) On the same note, Yukihara (2011, 2019a, 2019b) investigated the TR-OSL and TT-OSL signals of BeOT and proposed a new protocol for the usage of BeO chips with the help of automated OSL readers, such as Risø TL/OSL (Yukihara et al., 2016)
* Corresponding author
E-mail address: pavkonst@physics.auth.gr (P.G Konstantinidis)
Contents lists available at ScienceDirect Radiation Measurements
journal homepage: www.elsevier.com/locate/radmeas
https://doi.org/10.1016/j.radmeas.2022.106716
Received 19 November 2021; Received in revised form 23 January 2022; Accepted 29 January 2022
Trang 2Recently, Altunal et al (2018), synthesized four different
nano-powders of BeO with the help of the sol-gel method and studied
their OSL properties, including the thermal stability, the minimum dose
that can be detected and the fading of the signals The same team found
out that the BeO pellet sintered at 1100 ◦C has a TL peak around 200 ◦C
In another study of Altunal et al (2019), some BeO pellets were
syn-thesized using the precipitation method at three different sintering
temperatures It was found out that the higher the sintering temperature,
the higher the OSL signal was Finally, Altunal et al (2020), who
investigated the effects of the different impurities in Beryllium oxide
ceramics, found that a BeO dosimeter with Dysprosium or Erbium
im-purities in its Bravais lattice form, yields a double peak at the dosimetric
region
However, a new type of BeO ceramic, originating from a Turkish
private company (Radkor), with laboratory code name BeOR yields two separate TL peaks at the dosimetric temperature region of 100–300 ◦C and one peak at 350 ◦C (Fig 1a) BeOR was studied extensively by As¸lar
et al (2021) with a plethora of techniques, which besides TL and opti-cally stimulated luminescence (OSL) include electron spin resonance (ESR), scanning electron microscopy (SEM) coupled with energy dispersive x-ray spectroscopy (EDX) and x-ray diffraction analysis (XRD) Moreover, the same team conducted some experiments on a variety of the luminescence dosimetric properties of BeOR, such as dose-response, minimum detectable dose threshold, thermal stability and quenching, reproducibility, and bleaching These authors concluded that the difference in the TL glow curve between the two types of do-simeters could arise from different sintering and/or calcination appli-cations during the manufacturing process or from the increased content
of Cr, Mg and Sn
On the other hand, Polymeris et al (2020) concentrated on the recombination pathways during the readout stage regarding four different known dosimeters, including BeOT It is known that these pathways can be either delocalized, via the conduction band, or local-ized, involving tunneling through an excited state of the trapped elec-trons These authors have proposed a series of experiments in order to compare the value of the activation energy of each peak and to distin-guish its recombination pathway, by including the fractional glow technique (FGT), an isothermal decay signal (PID) and the peak shape methods (PSM)
The main aim of the present study is to use the methodology that is proposed by Polymeris et al (2020) in order to generate experimental data for the case of both main dosimetric peaks of BeOR Specifically, it would be beneficial to exploit this methodology to the case of one single material that yields two different and overlapping TL glow peaks with different recombination pathways By conducting the three series of experiments, it is possible to calculate the activation energy of both main dosimetric peaks at different scenarios and thus being able to compare them The second aim of this study, as the protocol suggests, is
to distinguish the recombination pathways that are available in the BeOR dosimeter, after discriminating them
2 Apparatus and experimental procedure
The inorganic compounds of BeOR are colorless under normal con-ditions and are created to have dimensions of 4 × 4 × 1mm The do-simeters that are used in the experiments are either un-preheated or preheated at 160 ◦C
All of the experiments were conducted in two different readers, namely a Harshaw-3500 and a Risø TLD-reader The Harshaw-3500 TLD-reader includes only neutral density filters, taking advantage of a nitrogen atmosphere and a low constant heating rate of 2 ◦C/s in order
to minimize the thermal lag that may occur On the other hand, the Risø TLD-reader (Model DA-20) is equipped with an EMI 9235QB PM Tube, and the filter used in front of the PMT is a 7.5 mm Hoya U-340 filter (270–380 nm, FHWM 80 nm) The irradiations were accomplished via
a90Sr/90Y beta source with a dose rate of 0.53 Gy/min One of the dif-ferences of the two readers lies in the thermal contact, of which the Harshaw reader is superior This may lead to a temperature lag in the case of the Risø measurements resulting in a shift of the TL glow curves
Fig 1a shows that the BeO fabricated by Radkor indeed has a double
TL main dosimetric peak The activation energies of each one between these two overlapping TL traps were calculated using Fractional Glow and Initial Rise Techniques, Peak Shape Methods, and analysis of Isothermal TL
The prompt isothermal decay (PID) experiments were conducted in both a Harshaw-3500 (Tmax =400 ◦C) and a Risø (Tmax =500 ◦C) TLD- Reader, according to the following protocol:
1 Test dose and TL up to Tmax
Fig 1 a) TL signals of BeOR measured using the (1) Risø-reader and (2)
Harshaw-Reader BeOR has two different overlapping peaks at the main
dosi-metric region At the same figure, glow curve (3) represents BeO Thermalox (b)
Examples of Isothermal and Residual TL signals of BeOR according to steps 3
and 4 of the corresponding protocol In all curves, the arithmetic corresponds to
the temperature at which Isothermal TL was measured, while the index
rep-resents the corresponding step of the protocol For the case of the Isothermal
signals the x axis corresponds to the temperature up to the maximum intensity
(170 ◦C and 215 ◦C), while the decaying part is plotted versus the stimulation
time As for the Residual TL curves, the x axis corresponds only to the
temperature
Trang 32 Dose of 0.275 Gy for Harhsaw and 0.5 Gy for Risø The experiment
was performed two times in different chips, namely with and without
a preheat with a constant heating rate up to 160 ◦C in order to
remove the unstable low temperature TL peaks and to ensure that the
only remaining peak from the two overlapping ones in the dosimetric region will be the higher temperature one
Fig 2 Residual TL glow curves according to step 4 corresponding to (a) Risø equipment for un preheated sample, (b) Risø equipment for preheated sample, (c)
Harshaw reader for un preheated sample, (d) Harshaw reader for preheated sample In all cases the residual curves show a slight shift towards higher temperatures with increasing stimulation temperature (illustrated as arrows)
Fig 3 Normalized Isothermal decay signals for the
case of (a) Harshaw reader for un preheated sample, (b) Harshaw reader for preheated sample, (c) Risø equipment for un preheated sample, (d) Risø equip-ment for preheated sample Isothermal decay curves fall clearly within two distinguished groups The first one includes curves with different shape according to the Isothermal temperature (illustrated as arrows); this group is prominent in figures a (where the cor-responding decay curves are highlighted within ellipsoidal for temperature region 110–165 ◦C) and c The second group consists of decay curves that, despite their different stimulation temperature, indi-cate shape that coincides This phenomenon is prominent in all four plots and specifically in plot a is highlighted as a rectangle within the temperature region of 165–220 ◦C
Trang 43 Temperature increase up to Ti with a constant heating rate, in which
the isothermal decay will take place for 110 s, in order to record the
PID signal
4 TL readout up to Tmax (400 ◦C or 500 ◦C, depending on the reader) to
require the remnant Residual TL signal (RTL) from step 3
5 Repeat steps 2–4 for Ti: 110–240 ◦C with a step of 5 ◦C for the
Harshaw reader and 10 ◦C for the Risø one
One example of the measurement sequence of the aforementioned
protocol is Fig 1b, in which the first part of the signal, up to the
maximum intensity, is used for the Initial Rise method, while the second
decaying part can be used for the Isothermal Decay Method Step 4 is
very important, as the Residual Curves can be used for the Initial Rise
method as well as for the Peak Shape Methods (PSM) for the preheated
versions
The Fractional Glow Technique (FGT) protocol is conducted in Risø
equipment according to the following:
1 Test Dose (TD) of 0.38 Gy and record TL glow curve up to Ti =40 C
2 Record TL glow curve up to Ti C, varying from 50 ◦C to 500 ◦C in
steps of 10 ◦C
3 Method of analysis
As it was mentioned, the activation energies of the two TL glow peaks
of the main dosimetric region of BeOR are calculated with different
methods, which are proposed by Polymeris et al (2020), in two different
readers in order to accumulate more data and thus better results
3.1 Initial Rise method (IR)
The Initial Rise method (IR) was firstly given by Garlick and Gibson
(1948), simulated by Kitis et al (2017) and is the one of the most
important techniques for the calculation of the activation energy of a
trap It was generalized, by recording the luminescence during both the
heating and cooling stages, and accepted as the Fractional Glow
tech-nique (FGT) by Gobrecht and Hofmann (1966) and Tale (1981) In the
present study, a consecutive Initial Rise method in order to increase the
temperature step wise (or the heating stage of the FGT) is applied,
ac-cording to which the TL intensity in the region of negligible disturbance
of trap population is given by the following equation:
I(T)∝n0se
(
−E/ kT
)
(1) Where s(s− 1) is the frequency factor, E(eV) is the activation energy and T
(K) is the temperature According to equation (1), the plot of ln(T)
versus 1/kT represents a straight line, the slope of which is the activation
energy -E In the present study the Initial rise method was applied to the
first region of Fig 1b up to the maximum intensity and to the RTL data of
Fig 2
3.2 Isothermal decay method (PID)
The isothermal decay method can be described by three identical models as (1) delocalized recombination model (Kitis and Vlachos,
2013), (2) localized recombination model (Kitis and Pagonis, 2018) and (3) localized tunneling recombination model (Kitis and Pagonis, 2013)
In all aforementioned models the decay constant λ can be given by the following equation (Chithambo and Niyonzima, 2014):
Fig 4 Example of deconvolution of an Isothermal decay at 150◦C for the
Harshaw un preheated case Two components are used for a better fit
Fig 5 lnτ for the dominant component of the Isothermal Decay signal for a) the entire temperature region, in which the activation energies are calculated with the help of the corresponding slopes and b) a close up at higher temperatures
Trang 5Once again the s(s− 1) is the frequency factor, E(eV) is the activation
energy, and the T i is the temperature that is used in the step 3 of the
aforementioned PID protocol The deconvolution of the PID signals can
be accomplished with the general order formula, introduced by Kirsh
and Chen (1991):
I(t) = I0
[
1 + (b − 1) t
τ
]−b b− 1
(3)
In this equation b∕=1 is the kinetic order, I(t) is the luminescence
intensity and τ =1/λ is the mean lifetime Thus, the decay constant λ can
be calculated by the fitting of Fig 4, with the help of the spreadsheets of
Afouxenidis et al (2011) By replacing λ with τ in equation (2) and with
the use of the natural algorithm, the plot of ln( τ ) versus 1/kT is a straight
line with a slope that represents the activation energy E of the trap
However, in order to ensure a good fit for the signals, it is essential that
the Figure Of Merit (FOM %) -introduced by Balian and Eddy (1977)-
should be as low as possible These authors concluded that a FOM of 3%
or lower is highly desirable, while a FOM of 10% and above is highly
unrecommended as an acceptable value For values between 3% and
10% the whole fit should be revised with some alterations in the initial
parameters Nevertheless, the FOM value is given by the following equation:
i
[⃒⃒Y exp− Y fit
⃒
A
]
(4)
In this case Yexp regards the experimental data, Y fit is the theoretical
data resulting from the whole procedure and A is the area of the fitted
curve of the signal that is analyzed
3.3 Peak shape methods (PSM)
The peak shape methods are used in order to calculate the activation energy of a TL peak with the condition of it being isolated Thus, these methods will be applied only in the cases of the preheated samples, in order to analyze one TL glow peak at a time The peak shape methods take advantage of the geometrical characteristics of the TL glow curve,
mainly the maximum peak temperature T m and the high and low
tem-peratures (T 1 and T 2 accordingly) sides of the glow curve in the half maximum intensity Chen (1969) and Kitis and Pagonis (2007) have independently created peak shaped equations that give similar but no identical values for the activation energies based on the general order kinetics model Thus, the presenting calculations in this work are based
on both groups of PSM
4 Results and discussion
Fig 2 shows all the residuals according to the Isothermal decay protocol and can be used for two experimental analyses All of the peaks
in the case of Risø have a higher Tm temperature (a deviation of 40–50 ◦C) and are clearly wider than those of the Harshaw cases, even though the experiments were conducted in both readers with a heating
Fig 6 a) Results of the Fractional glow technique; three different plateaus
become prominent for the entire glow curve b) Activation energies calculated
via all different experiments of Initial Rise for all samples, based on the first half
of the PID signal (IR) and the Residual TL curves (RTL)
Table 1
Activation energies of the two main dosimetric traps, resulting from four different analyses for both readers N corresponds to the total number of the curves that were used in the whole analysis For each equipment the first arithmetic indicates the total number of un preheated curves, while the second the total number of preheated curves
Activation
un preheated 1.26 ±
0.02 1.02 ±0.04 1.27 ±0.03 1.08 ±
0.01 preaheated 1.27 ±
0.01 1.07 ±0.01 1.26 ±0.03 1.12 ±
0.03
Chen 1.00 ±
0.19 1.19 ±0.19 1.06 ±0.20 1.05 ±0.12 1.14 ±
0.15
1.11 ± 0.12 Kitis 1.03 ±
0.2 1.03 ±0.3 1.03 ±0.4 1.18 ±0.08 1.18 ±
0.09
1.18 ± 0.10
165 C1 165- 215 140- C2
210
C1 140-
165
C1 165-
215
C2 140-
210
un preheated 0.96 0.05 0.56 0.96 0.06 0.68 preheated 0.96 0.14 0.59 – 0.1 0.44
(eV)
130–280 1.25 ±
0.04 320–390 1.36 ±
0.04 440–490 1.90 ±
0.09
Trang 6rate of 2 ◦C/s The aforementioned phenomenon can be ascribed to the
thermal gradient between the sample and the heater plates (
Kon-stantinidis et al., 2020; Bilski et al., 2014) When a Harshaw reader is
being used, there is a direct heating of the sample, as it is placed on the
corresponding planchette (Bilski et al., 2014) On the other hand, in the
case of a Risø reader, the sample is firstly placed inside the stainless-steel
cup, which is then placed on the heating planchette of the Risø reader
This geometry leads to an indirect heating of the used dosimeter or
sample Having in mind the previous analysis, it can be assumed that
there is smaller thermal gradient in the Harshaw readers and thus, the
values of Tm and FWHM acquired by this reader are closer to the actual
form of the TL glow curve/peaks As for the comparison of the signals
between the two readers, it is worth mentioning that the TL curves of the
experiments of the Risø system are more intense than those of the
Harshaw 3500 reader, despite the fact that the Risø system has a Hoya
U-340 filter and the Harshaw has none, as there is a difference within the
same applied dose For the Initial Rise method all four figures can be
utilized, while only the preheated versions are used in order to acquire
the necessary data for the Peak Shape Methods Also, the clearly visible
temperature shift of the second glow peak can be a first indicator of a
possible localized recombination pathway without the passing via the
conduction band
Fig 3 shows the four different versions of the isothermal decay
sig-nals, which follow the same trend The difference in the starting
tem-perature between the two readers (10–20 ◦C) is due to the thermal lag, as
mentioned and discussed before Also, the preheated samples are used
mainly for the analysis of the second peak of the main dosimetric region
Moreover, these figures are the second indicator of the different
path-ways that are used in the two overlapping peaks of the BeOR dosimeters
Specifically, the de-localized model pathway can be found up to 165 ◦C,
while the localized one is at temperatures higher than 165 ◦C, as can be
seen from the example of Fig 3a Fig 4 presents an example of the
deconvolution analysis of these PID signals, with the use of the
spreadsheets of Afouxenidis et al (2011) and equations (2) and (3) As
BeOR exhibits two overlapping peaks, the use of two different
compo-nents, namely C1 and C2, is essential for the experimental and the
theoretical signals to be in agreement with a FOM value lower than 3%
It must be noted that the study focuses on the dominant C1, as C2
rep-resents a slower component to the deconvolution process
From the Isothermal Decay signal (PID) study, the activation
en-ergies of the first component in Fig 5a in the region of 130–165 ◦C seems
to be near 0.96 eV, regarding the first peak of the BeOR in the dosimetric
region From the data of Fig 5b, it is concluded that at temperatures
beyond 165 ◦C the analysis does not give a satisfactory result, indicating
once again the localized pathway of the second peak Moreover, Fig 6a
shows the results of the Fractional Glow Technique through the
tem-perature range of 50–500 ◦C There are 3 different plateaus, namely at
130–280 ◦C, 320–390 ◦C and 440–490 ◦C The main dosimetric peaks
are located within the first plateau, both having activation energies of
1.25 eV From Fig 6b one can see the activation energies that accrues
from all the experimental procedures of the Initial Rise that were
con-ducted in the present study, for the un preheated and the preheated
samples The same results can be viewed in Table 1 and are in agreement
with the findings of As¸lar et al (2021) From the Initial Rise method of
the residuals both versions seem to have an activation energy of 1.26eV,
while the IR method from the isothermal signals are set to be at 1.08eV
Lastly, the peak shape methods of the residuals of the preheated samples
have visible differences between the two readers For the case of Risø,
the activation energy is at 1.03eV while for the Harshaw it is at 1.18eV
5 Conclusions
In the present study, the methodology proposed by Polymeris et al
(2020) was used in order to generate specific experimental data for the
case of a newly developed BeO dosimeter with two main dosimetric TL
peaks In the framework of this methodology, the activation energy of
each one between these two peaks was calculated using Isothermal Decay, Fractional Glow Technique (or Initial Rise) and Peak Shape Methods According to the analysis of the present study, BeOR yields two main dosimetric peaks with two different recombination pathways For the case of the main TL trap corresponding to the low-temperature re-gion, all techniques yield similar values of activation energy, namely around 1–1.30 eV, providing strong experimental indications towards a recombination pathway via the conduction band On the contrary, regarding the main TL trap corresponding to the high-temperature re-gion, Isothermal TL provides values of activation energies below 0.2 eV, while the other two techniques around 1.10–1.30 eV In a nutshell, the effective use of the aforementioned methodology was successfully exploited even for the case of one single material that yields two different and overlapping TL glow peaks with different recombination pathways Discrimination could be achieved using solely the Isothermal
TL and neither Initial Rise technique nor the Peak Shape Methods could identify localized recombination pathways
Declaration of competing interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper
References
Afouxenidis, D., Polymeris, G.S., Tsirliganis, N.C., Kitis, G., 2011 Computerized curve deconvolution of TL/OSL curves using a popular spreadsheet program Radiat Protect Dosim 149, 363–370
Altunal, V., Yegingil, Z., Tuken, T., Depci, T., Ozdemir, A., Guckan, V., Nur, N., Kurt, K., Bulur, E., 2018 Optically stimulated luminescence characteristics of BeO nanoparticles synthesized by sol-gel method Radiat Meas 118, 54–66 Altunal, V., Guckan, V., Ozdemir, A., Sotelo, A., Yegingil, Z., 2019 Effect of sintering temperature on dosimetric properties of BeO ceramic pellets synthesized using precipitation method Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 441, 46–55
Altunal, V., Guckan, V., Yu, Y., Dicker, A., Yegingil, Z., 2020 A newly developed OSL dosimeter based on beryllium oxide: BeO:Na,Dy,Er J Lumin 222, 117140 As¸lar, E., S¸ahiner, E., Polymeris, G.S., Meriç, N., 2021 Thermally and optically stimulated luminescence properties of BeO dosimeter with double TL peak in the main dosimetric region Appl Radiat Isot 170, 109635
Balian, H.G., Eddy, N.W., 1977 Figure-of-merit (FOM), an improved criterion over the normalized chi-squared test for assessing goodness-of-fit of gamma-ray spectral peaks Nucl Instrum Methods 145 (2), 389–395
Becker, K., Cheka, J.S., Oberhofer, M., 1970 Thermally stimulated exoelectron emission,
TL and impurities in LiF and BeO Health Phys 19, 391–403 Bilski, P., Gieszczyk, W., Obryk, B., Hodyr, K., 2014 Comparison of commercial thermoluminescent readers regarding high-dose high-temperature measurements Radiat Meas 65, 8–13
Bulur, E., 2007 Photo-transferred luminescence from BeO ceramics Radiat Meas 42, 334–340
Bulur, E., 2014 More on the TR-OSL signal from BeO ceramics Radiat Meas 66, 12–20 Bulur, E., Goksu, H.Y., 1998 OSL from BeO ceramics: new observations from an old material Radiat Meas 29, 639–650
Bulur, E., Saraç, B.E., 2013 Time-resolved OSL studies on BeO ceramics Radiat Meas
59 Bulur, E., Yeltik, A., 2010 Optically stimulated luminescence from BeO ceramics: an LM- OSL study Radiat Meas 45, 29–34
Chen, R., 1969 On the calculation of activation energies and frequency factors from glow curves J Appl Phys 40, 570–585
Chithambo, M.L., Niyonzima, P., 2014 On isothermal heating as a method of separating closely collocated thermoluminescence peaks for kinetic analysis J Lumin 155, 70–78
Crase, K.W., Gammage, R.B., 1975 Improvements in the use of ceramic BeO for TLD Health Phys 29, 739–746
Gammage, R.B., Garrison, n A.K., 1974 Investigation of EPR, TLD and TSEE of BeO ceramic In: Proc 4 th Int Conf Luminescence Dosimetry Institute of Nuclear Physics, Krakow, p 263
Garlick, G.F.J., Gibson, A.F., 1948 The electron trap mechanism of luminescence in sulfide and silicate phosphors Proc Phys Soc Lond 60, 574–590
Gobrecht, H., Hofmann, D., 1966 Spectroscopy of traps by fractional glow technique
J Phys Chem Solid 27, 509–522 Kiiko, V.S., Zolotukhina, L.V., Zabolotskaya, E.V., Dmitriev, I.A., Makurin, Yu N., 1999 Defects in beryllium ceramics Glass Ceram 56 (11), 346–349
Kirsh, Y., Chen, R., 1991 Analysis of the blue phosphorescence of X-irradiated albite using a TL-like presentation Nucl Tracks Radiat Meas 18, 37–40
Kitis, G., Pagonis, V., 2007 Peak shape methods for general order thermoluminescence glow-peaks: a reappraisal Nucl Instrum Methods Phys Res B 262, 313–323
Trang 7Kitis, G., Pagonis, V., 2013 Analytical solution for stimulated luminescence emission
from tunneling recombination in randomly distributions of defects J Lumin 137,
109–115
Kitis, G., Pagonis, V., 2018 Localized transition models: a reappraisal Nucl Instrum
Methods Phys Res B 432, 13–19
Kitis, G., Vlachos, N., 2013 General semi-analytical expressions for TL, OSL and other
luminescence stimulation modes derived from OTOR model using the Lambert W
function Radiat Meas 48, 47–54
Kitis, G., Pagonis, V., Tzamarias, E.E., 2017 The influence of competition effects on the
initial rise method during thermal stimulation of luminescence: a simulation study
Radiat Meas 100, 27–36
Konstantinidis, P., Tsoutsoumanos, E., Polymeris, G.S., Kitis, G., 2020
Thermoluminescence response of various dosimeters as a function of irradiation
temperature Radiat Phys Chem 177, 109156
Nieto, Azorin J., 2016 Present status and future trends in the development of
thermoluminescent materials Appl Radiat Isot 117, 135–142
Ogorodnikov, I.N., Petrenko, M.D., Ivanov, V Yu, 2016 Low-temperature luminescence
and thermoluminescence from BeO: Zn single crystals Opt Mater 62, 219–226
Polymeris, G.S., Pagonis, V., Kitis, G., 2020 Investigation of thermoluminescence processes during linear and isothermal heating of dosimetric materials J Lumin
222, 117142 Stockinger, H.E (Ed.), 1966 Beryllium: its Industrial Hygiene Aspects Academic Press, New York
Tale, I.A., 1981 Trap spectroscopy by the fractional glow technique Phys Stat sol 66, 65–75
Tochilin, E., Goldstein, N., Miller, W.G., 1969 BeO as a thermoluminescent dosimeter Health Phys 16 (issue 1), 1–7
Watanabe, S., Gundu Rao, T.K., Page, P.S., Bhatt, B.C., 2010 TL, OSL and ESR studies on beryllium oxide J Lumin 130, 2146–2152
Yukihara, E.G., 2011 Luminescence properties of BeO optically stimulated luminescence (OSL) detectors Radiat Meas 46, 580–587
Yukihara, E.G., 2019a Characterization of the thermally transferred optically stimulated luminescence (TT-OSL) of BeO Radiat Meas 126, 106132
Yukihara, E.G., 2019b Observation of strong thermally transferred optically stimulated luminescence (TT-OSL) in BeO Radiat Meas 121, 103–108
Yukihara, E.G., Andrade, A.B., Eller, S., 2016 BeO optically stimulated luminescence dosimetry using automated research readers Radiat Meas 94, 27–34