They also allow for simultaneous acquisition and display of multiple orthogonal 2DE images that can be electronically steered in the lateral plane or the plane of elevation.. This mode c
Trang 1There are several modes of 3DE acquisition; the appropriate mode must be decided based on the clinical question While the nomenclature of different modes is specific to each vendor, there are many shared features
Enhanced Two-Dimensional Echocardiography Capabilities
Matrix transducers enable rotation of the 2DE plane incrementally by 360 degrees They also allow for simultaneous acquisition and display of multiple orthogonal 2DE images that can be electronically steered in the lateral plane or the plane of elevation This mode can be used to assess the location, size and shape of atrial and ventricular septal defects, the morphology of valves and outflow tracts, and the nature of regurgitation (Fig 20.3, Video 20.4)
FIG 20.3 Simultaneous two-dimensional echocardiography (2DE) color
Trang 2ventricular septal defect in the parasternal long axis in the left panel, which
is the reference image The operator can touch a single button to activate
biplane (X-plane) mode, which defaults to an (adjustable) view that is rotated 90 degrees clockwise from the reference image In this case, the
result is a simultaneous 2DE color flow demonstration of the ventricular
septal defect in both a parasternal long-axis view and parasternal short-axis view, with a frame rate of 24 frames per second LA, Left atrium; MV,
mitral valve; RA, right atrium; TV, tricuspid valve.
Real-Time or Live Three-Dimensional
Echocardiography
Live 3DE provides a narrow, adjustable pyramidal volume of data It minimizes the potential for motion and artefacts that are seen with other modes of 3DE, as described below Acceptable temporal resolution with live 3DE comes at the expense of field of view, which can be a disadvantage in complex defects where multiple spatial relationships need to be determined Cardiac structures that are relatively static, such as atrial or ventricular septal defects, lend themselves well
to live 3DE and are particularly useful when guiding interventions, for example Some vendors allow adjustments to prioritize temporal resolution at the expense
of line density
Electrocardiographically Gated Acquisitions
Over Multiple Beats (Full Volumes)
Full volume 3DE provides a wide pyramidal 3DE volume that is acquired over multiple consecutive cardiac cycles Full volumes are frequently used in
pediatric imaging because they allow the acquisition of a sufficiently large field
of view with good temporal resolution In young children, these datasets are prone to artefacts related to breathing and other movement This is not an issue
in children who can hold their breath or are being ventilated under general
anesthesia because ventilation can be briefly suspended While the capability to acquire a full volume of data over a single cardiac cycle is now available, the current temporal resolution of this latter mode is not acceptable in children
Three-Dimensional Color Flow
Trang 3of grayscale data To ensure adequate temporal resolution, these acquisitions have to be performed over several cardiac cycles