Static artefacs are generated by " chemical shift" differences see chemical shift and, or different susceptibilities.. Chemical shift artefactChemical shift artefact: In spine, abdomen a
Trang 1Artefacts in MRI
From http://members.lycos.nl/mri/Nieuw/
artefacteneng.htm
Artefacts in NMR are useless pixels in a picture wich do not contribute
to the picture itself It is caused by the patient or external factors
Patient-related artefacts are static (constant during measurement) or dynamic Static artefacs are generated by " chemical shift" differences (see chemical shift) and, or different susceptibilities Dynamic artefacts are caused by breathing, heartbeat, bloodflow or any other movement made by the patient External artefacts are hardware related or caused
by external factors
And more from http://www.hull.ac.uk/mri/lectures/Gpl%20web%20page/ gpl_page#RF
Trang 2Chemical shift artefact
Chemical shift artefact: In spine, abdomen and eyes were tissue is
surrounded with fat and after Gadolineum (because fat precesses almost at the same frequency as pathology tissue) In the frequentie-encoding direction the spatial resolution of the signal is determined by the frequency of this
signal In the same organ water is slighty different in frequency from fat (3.5 ppm) An MR-scanner will mismap this frequency difference Fat is being
shifted in the frequency encoding direction In other words: Pixels are placed
in the wrong voxel) It is mainly at high field strenghts chemical shift plays an important role Fatsuppression (STIR/prepuls) will eliminate the artefact
Shimming the magnetic field is essential
Chemical shift contours is seen in Gradient echo sequence only At 1.0 T fat and water spins are in and out of phase every 3,4 ms after the alpha pulse
So TE is 3,6 the sequence is IN phase, TE is 7,2, the sequence is OUT of phase.
Trang 3Susceptibility artefact:
Susceptibility artefact: Caused
by microscopic differences in
susceptibility for the magnetic
field Bone/air or metal/tissue
combinations give this artefact
Defasing of spins and frequency
shift of surrounding tissue is the
result You can recognize the
artefact by the dark and white
disturbances of the tissue The
more defasing (long TE) the more
susceptibility Gradient echo´s do
not use a 180° refasing pulse, so
there´s a lot of defasing there
Here´s a gradient echo picture of
the knee with a screw in the tibia
Trang 4Aliasing/Wrap around
Aliasing/Wrap around: If the FOV is smaller than the object The
anatomy outside the FOV will be misplaced inside the picture
because it gives a signal in the receiver coil These data will be
encoded but will be"undersampled" and misplace IN the FOV
"Oversampling" is this problems solution In the frequency encoding direction the signal is sampled twice as fast. In the fase encoding direction more fase encodings will be generated (scantime is rising) Changing the frequency and fase encoding direction is sometimes a solution too
Trang 5Flow artefacs
Flow artefacs: Caused by blood/liqor Important are: Flow direction, flowspeed and the sequence used One can eliminate these artefacs
by using prepulses to saturate these moving protons