Optimization of a flow injection hydride generation atomic absorption spectrometric method for the determination of arsenic, antimony and selenium in iron chloride/sulfate-based water tr
Trang 1Optimization of a flow injection hydride generation atomic absorption spectrometric method for the determination of arsenic, antimony and selenium in iron chloride/sulfate-based water
treatment chemical Teemu Näykkia,∗, Paavo Perämäkia, Jyrki Kujalab, Anna Mikkonenb
aDepartment of Chemistry, University of Oulu, P.O Box 3000, FIN-90014 Oulu, Finland
bKemira Chemicals Oy, Oulu Research Centre, P.O Box 171, FIN-90101 Oulu, Finland
Received 26 September 2000; received in revised form 3 March 2001; accepted 19 March 2001
Abstract
The flow injection hydride generation technique together with atomic absorption spectrometry was used for the determi-nation of arsenic, antimony and selenium in the iron-based water treatment chemical FeClSO4 Thiourea, l-cysteine and potassium iodide–ascorbic acid were used as masking agents to diminish the interference caused by the very high iron con-centrations in the samples These reagents act also as prereductants for As(V) and Sb(V) Thiourea andl-cysteine did not prevent the signal depression caused by such high iron content, but potassium iodide–ascorbic acid eliminated iron interference well even up to 2500 mg Fe l−1 The limits of detection (LODS) in aqueous solutions containing no iron were 0.037g l−1, 0.121 g l−1and 0.131g l−1for As, Sb and Se, respectively The linear dynamic range was 0–10g l−1for As and 0–30g l−1 for Sb and Se The precision relative standard deviation was expressed as 2.6% for As, 4.4% for Sb and 2.9% for Se The precision determinations were done on the FeClSO4matrix at the level 0.5–0.8g l−1for the elements to be analyzed The accuracies of the methods were tested by using two standard reference materials (SRM 361, LA Steel and SRM 2074, river sediment) The concentrations obtained for As, Sb and Se were very close to the certified values © 2001 Elsevier Science B.V All rights reserved
Keywords: Atomic absorption spectrometry; Hydride generation; Flow injection; Arsenic; Antimony; Selenium; Iron interference; Water
treatment chemical
1 Introduction
It is commonly known that the transition metals
in-terfere in the determination of hydride-forming
ele-ments when the hydride generation (HG) technique is
used [1,2] Different mechanisms have been suggested
∗Corresponding author Present address: Finnish Environment
Institute (FEI), Research Laboratory, Hakuninmaantie 4-6,
FIN-00430 Helsinki, Finland Fax: +358-9-4030-0890.
E-mail address: teemu.naykki@vyh.fi (T Näykki).
for the interference effects observed [3,4] The pre-dominant mechanism is probably due to the reaction
of the interfering transition metal ions with the NaBH4
reductant, and the precipitate which is formed is able
to capture and catalytically decompose the evolved hydrides [3] For instance, Bax et al [4] and Bye [5] have stated that the precipitates are probably not el-emental metals, but rather metal borides Lugowska and Brindle [6] investigated the redox processes oc-curring in transition metal solutions during reduction
by NaBH4 Also they found, boron(III) as boride-like
0003-2670/01/$ – see front matter © 2001 Elsevier Science B.V All rights reserved.
PII: S 0 0 0 3 - 2 6 7 0 ( 0 1 ) 0 1 0 0 1 - 7
Trang 2species causing the suppression of the signal in
hy-dride generation [6]
Iron is a common transition metal that is present
at high concentrations in many type of samples
However, only a few studies have been published
where the iron concentrations of the samples are as
high as in this work Narsito et al [7] found that
the interference caused by 200 mg l−1 iron was
re-moved when 1% thiourea was used as a prereductant
Wickström et al [8] studied hydride generation and
the complexation reactions in an alkaline sample
solution The hydrides formed were evolved by
sub-sequent acidification of the sample solution They
found that 0.3 mol l−1 tartrate completely removed
iron interference at the 500 mg Fe l−1level and when
the iron concentration was 5000 mg l−1, the signal
was depressed by only 15% However, dissociation of
the metal complex can occur during the acidification
of the sample Thus, the formation of free metal ions
and precipitation is possible Boampong et al [9]
used 3% l-cystine in 5 mol l−1 hydrochloric acid as
a masking agent to prevent the interferences of very
high concentrations of iron Welz and Sucmanova [10]
proposed the use of l-cysteine instead of l-cystine
as a masking agent Earlier Chen et al [11] had
ob-served thatl-cysteine removed iron interference up to
1000 mg l−1 even though the solution became turbid
due to the formation ofl-cystine when the iron(III)
concentration was greater than 200 mg l−1
Gener-allyl-cysteine has been proved to be very useful for
preventing iron interferences, but nearly all
previ-ous publications have focused on iron concentrations
below 1000 mg l−1 Finally, it is worth noting that
iron(III) has also been used for minimizing
interfer-ence effects of the other transition metals in the HG
technique The high positive potential of the reduction
of iron(III) to iron(II) suggests a preferential
reduc-tion of this species More seriously interfering metal
ions, such nickel(II) and copper(II), will be reduced
to the metals and precipitated only after all the iron
has been reduced [12,13]
The extent of the transition metal interference
de-pends very much on the HG system used Severe
in-terference was observed when we earlier used the
batch type hydride generation system described by
Siemer and Hagemann [14] The flow injection (FI)
technique is less prone to transition metal
interfer-ence [10,15] There are at least two main reasons for
this observation When using the FI system instead of the batch system, the concentration of the reductant is usually lower and formation of the interfering precip-itates, e.g borides, is decreased Another reason can
be called kinetic discrimination The reduction of the hydride-forming elements is fast and the reaction is completed before the reduction of the transition metal ion to the interfering species Also the separation of the hydrides from the sample matrix is very fast in a gas–liquid separator
In this study, the parameters for the determina-tion of As, Sb and Se from the PIX-110TMchemical (FeClSO4) using flow-injection hydride generation (FI-HG) were optimized The PIX-110TMwater coag-ulant is a FeClSO4solution manufactured by Kemira Chemicals Oy, containing approximately 12.0 wt.% (180 g l−1) of Fe3 +, 22.3% of SO
4 − and 7.5% of
Cl− Since As, Sb and Se are toxic, the European
Committee for Standardization (CEN) has laid down the maximum limits for these elements and therefore
it is essential to monitor their low amounts in water treatment chemicals [16]
2 Experimental
2.1 Instrumentation
A Perkin-Elmer model 5100 Zeeman atomic ab-sorption spectrometer equipped with a Perkin-Elmer model FIAS-400 flow injection system and an AS90 autosampler, controlled by Perkin-Elmer AA WinLab version 2.61 software, was used for the measurements Silicone pump tubes (NaBH4: 1.14 mm i.d.; carrier HCl: 1.52 mm i.d.; sample loading and waste: 3.17 mm i.d.) were used throughout this study, and all other tubing was 1 mm i.d PTFE An electrically heated quartz tube was used as an atomizer Electrodeless dis-charge lamps, operated from an external power supply (Perkin-Elmer EDL system 2) were used for the mea-surements The instrumental parameters are shown in Tables 1 and 2; they were mainly chosen according to the manufacturer’s recommendations
2.2 Reagents and standard solutions
All reagents were of analytical-reagent grade un-less otherwise stated Ultrapure water (18 M cm−1),
Trang 3Table 1
AA spectrometer settings used for the hydride
generation-AAS-measurements
Sb: 217.6 Se: 196.0 Lamp current (mA) As and Sb: 400
Se: 280 Measured signal Integrated absorbance
Sb and Se: 18
Quartz cell temperature ( ◦C) 900
Quartz cell length/i.d (mm/mm) 180/7.5
Reaction coil length (mm) 300
Background correction No
prepared with a USF Elga Maxima purification
sys-tem, was used throughout All glassware and plastic
containers were soaked in (1+1) nitric acid overnight
and rinsed five times with de-ionized water and five
times with ultrapure water prior to use
The working standard solutions were prepared
daily by diluting the 1000 mg l−1 standard stock
so-lution of As(V) (Merck) and the 4000 mg l−1Sb(III)
solution (Merck) To prepare the selenium standard
stock solution (1000 mg l−1) appropriate amounts
of selenium salts NaHSeO3 (Merck, for Se(IV)) or
Na2SeO4·10H2O (Merck, for Se(VI)) were dissolved
in ultrapure water A few drops of concentrated
hy-drochloric acid were added before dilution
The iron stock solution was prepared from
PIX-110TM, because it was observed to contain less
arsenic than the commercial iron(III) chloride
stan-dard The 20,000 mg l−1stock solution of Fe(III) was
prepared by diluting 16,5289 g (1000 g ≈ 0.667 ml)
of PIX-110TMsolution (purity not certified) to 100 ml
with 4.7 mol l−1HCl.
Table 2
FIAS-400 flow injection program used for the hydride
generation-AAS-measurements
Flow injection program
Step Time (s) Speed of pump
1 (rpm)
Speed of pump
2 (rpm)
Sodium tetrahydroborate solution (0.1–0.3% (w/v)) was prepared daily by dissolving the appropriate amount of NaBH4powder (Fluka) in ultrapure water The solution was stabilized with NaOH (Merck) Hydrochloric acid (0.03–4.7 mol l−1, Merck) was
used in samples and as a carrier solution Different concentrations of HCl were used in the optimization
of the operating conditions
The stock solution of potassium iodide was pre-pared by dissolving 5 g of potassium iodide (Merck) in
100 ml of ultrapure water The ascorbic acid stock so-lution was prepared by dissolving 5 g ofl(+)-ascorbic acid (Merck) in 100 ml of ultrapure water
Thiourea stock solution (2.0 mol l−1) was prepared
by dissolving 38 g of thiourea, (J.T Baker, purity >
99%) in 250 ml of 1.5–4.7 mol l−1HCl.
l-Cysteine stock solution was prepared by dis-solving 5 g of l-cysteine (Merck, for biochemistry, purity > 99%) in 100 ml of 0.03–0.09 mol l−1 HCl.
Different concentrations of l-cysteine and/or HCl were used to optimize the operating conditions The standard reference materials (SRMs) 361 Low-alloy Steel and 2704 Buffalo River Sediment were obtained from the National Institute of Stan-dards and Technology (NIST, Gaithersburg, MD) The samples were prepared by dissolving 1 g of the SRMS in 30 ml of aqua regia (HCl:HNO3, 3:1 (v/v))
2.3 Procedure
The FI system used consisted of two peristaltic pumps: one was used for pumping the carrier and re-ductant solutions (HCl and NaBH4, respectively) and the other for pumping the sample solution (Fig 1) The carrier flow rate was 9 ml−1min and the
reduc-tant flow rate was 5 ml−1min The 500l sample loop was used in all measurements
The optimal instrumental parameters for As, Sb and
Se are given in Tables 3 and 4 Potassium iodide (0.5% (w/v)) and ascorbic acid (1.0% (w/v)) were added as prereductants for arsenic and antimony prior to anal-ysis The prereduction time used was 15 min Sele-nium(VI) was reduced to selenium(IV) by heating the sample in a beaker with HCl (4.7 mol l−1) at 90◦C for
20 min
The standard addition method was used when PIX-110TMwas analyzed Dilutions of 1–225 ml (As)
or 1–72 ml (Sb and Se) were made before analysis
Trang 4Fig 1 Flow schematic of FIAS-400.
Iron was added to calibration standards to match
the iron content in the certified reference materials
when the accuracy of the method was tested A more
detailed description of the method optimization and
validation is given in the next section
Table 3
Optimal experimental conditions for the measurement of As and
Sba
values
HCl (carrier and samples) (mol l −1) 4.7 1.5
Carrier gas (Ar) flow rate (ml min −1) ∼65 ∼100
a 0.5% (w/v) KI, 1.0% (w/v) ascorbic acid, 0.3% (w/v) NaBH 4
in 0.05% (w/v) NaOH.
Table 4
Optimal experimental condition for the measurement of Se
values HCl (carrier and samples) (mol l −1) 4.7
Carrier gas (Ar) velocity (ml min −1) ∼65
Prereduction temperature ( ◦C) 90
Prereduction time (min) 20
(w/v) NaOH
3 Results and discussion
3.1 Selection of the prereduction agent 3.1.1 Arsenic and antimony
At the beginning of the study, the instrumental pa-rameters were optimized using only aqueous As stan-dards The same parameters were used for Sb because
of the similar chemical nature of As and Sb Opti-mum levels for carrier gas flow, reductant concentra-tion, carrier-HCl concentration and the concentrations
of the prereductants were investigated using two-level factorial designs Measurements were performed with-out an interfering iron matrix The results were pro-cessed with Modde for Windows version 3.0 software (Umetri AB) [17] Later, iron was added to the stan-dard solutions and the optimization was repeated using the parameters obtained from the previous stage The optimal concentrations of the various prereductants were studied using iron contents up to 5000 mg l−1.
When optimizing the prereduction conditions, the sample matrix contained 1500 mg l−1 of iron The
optimal concentration of the ascorbic acid was in-vestigated first and then kept at a constant level (0.8% (w/v)) when optimizing the KI concentra-tion (Fig 2A) When the sample matrix contained
5000 mg l−1 iron 1.0% (w/v) ascorbic acid gave the
Trang 5Fig 2 The effect of KI and ascorbic acid concentrations in the determination of As, when the iron concentration of the samples was: (A) 1500 mg l −1 and (B) 5000 mg l−1 Arsenic concentration in the samples was 1g l −1 HCl concentration of samples and carrier was
4.7 mol l −1.
maximum sensitivity The concentration of ascorbic
acid was kept at a constant level and the optimal KI
concentration was found to be 0.8% (w/v) (Fig 2B)
The sensitivity obtained with thiourea andl-cysteine
was worse than that with KI Using an iron
concen-tration, <200 mg l−1 and l-cysteine as the masking
agent, the recovery of the signals of arsenic and
anti-mony were excellent, as, for example, Welz [10] and
Chen [11] have reported The use of concentrated iron
solutions caused a severe signal depression despite
the presence ofl-cysteine; using a 1500 mg l−1 iron
Fig 3 Variation of peak area for 1 g l −1As as a function of prereduction time.
matrix, the depression of the As signal was ony 50%
of the signal measured without iron
The prereduction times needed with the different prereductants were investigated by measuring the re-sponse of 1g As(V) l−1solutions as a function of the
prereduction time (Fig 3) For each measurement the amount of As in the matrix/prereduction agents was subtracted from the results When thiourea was used, the prereduction of As(V) occurred immediately With l-cysteine and KI–ascorbic acid the maximum sensi-tivity was achieved within 15 min Fig 3 shows also
Trang 6the influence of the prereductants on the sensitivity.
The highest sensitivity was obtained with KI–ascorbic
acid When KI–ascorbic acid was used a higher
sen-sitivity (7%) was obtained in the presence of an iron
matrix The reasons for this effect were not studied
further
Memory effects were investigated by performing
about 10 successive atomizations from the samples
containing 1g As l−1 and 5000 mg Fe l−1
Immedi-ately after, 10 successive atomizations were performed
without the iron matrix For each measurement, the
amount of As in the matrix/prereduction agents was
subtracted from the results Memory effects were not
observed in the presence of any prereduction agent
Once again a dramatic decrease in the sensitivity
was observed when using prereductants other than
KI–ascorbic acid When KI–ascorbic acid was used
in the presence of 5000 mg Fe l−1, the sensitivity was
only 14% lower than that for the aqueous As standard
Using thiourea or l-cysteine to prevent 5000 mg l−1
iron interference, depression of As-signals were over
75% compared to the signal measured without iron
The existence of the matrix effect was studied by
observing the slopes of the calibration graphs obtained
with and without the iron matrix The matrix effect
was observed when using thiourea and l-cysteine in
the presence of very high iron concentrations The
de-crease in the calibration slope caused by the iron
ma-trix was from 0.32 to 0.17 when thiourea was used
as prereductant A similar effect was observed
us-ing l-cysteine, but with KI–ascorbic acid the
mea-surements were practically free of the matrix effect
(change of slope 0.31–0.30)
3.1.2 Selenium
At first the optimal HCl concentration,
prereduc-tion time and prereducprereduc-tion temperature without an
interfering iron matrix were studied using two-level
factorial designs Selenium(VI) was used in these
experiments Prereductants, e.g thiourea, can reduce
Se(VI) to the elemental state and the formation of
hy-dride becomes impossible [18] Therefore, different
prereduction agents for selenium were not examined
The heating of the sample solution together with
hydrochloric acid has been observed to be the most
practical way to reduce Se(VI) to Se(IV) Using the
optimal parameters obtained from the previous stage,
the optimization was repeated in the presence of iron
Table 5 The factors and their levels used for the optimization of parameters for As and Sb by CCI design a
Factor Low level ( −) High level ( +)
Argon (ml min −1) 50 (75) 100 (125)
KI (% (w/v)) 0.1 (0.5) 1.0 (1.5) Ascorbic acid (% (w/v)) 0.1 (0.5) 1.0 (1.5)
a The star distance used was 0.35; values in brackets are for Sb The axial points of the design can be calculated using the equation: (mean level) ± [(high level) − (mean level)] × star distance.
The most suitable prereduction agent, KI–ascorbic acid, was used in the final optimization of the oper-ating conditions for As and Sb by central composite inscribed (CCI) design In this design, each factor had five-levels The low and high levels of the design are given in Table 5 The factor levels used for the opti-mization of Se parameters by CCI design are given in Table 6
3.2 Optimal parameters for As, Sb and Se
The final optimal parameters for As, Sb and Se are shown in Tables 3 and 4 The memory effects and the matrix effects for selenium were investigated us-ing the optimal parameters as previously described for
As and Sb As in the determinations of As and Sb,
a memory effect was not observed in the presence of
5000 mg Fe l−1 However, the matrix effect was
ob-served and, according to the results, the iron matrix increased the sensitivity of the selenium determina-tion by about 45% when the iron concentradetermina-tion of the samples was 2500 mg l−1 Using higher iron
concen-trations, the signals of the selenium determinations
Table 6 The factors and their levels used for the optimization of parameters for Se by CCI-design a
Factor Low level ( −) High level ( +)
a The star distance used was 0.50; the axial points of the design can be calculated using the equation: (mean level) ± [(high level)
− (mean level)] × star distance.
Trang 7were depressed and the sensitivity was lower The
in-creased sensitivity was probably due to an oxidizing
and complexing effect of Fe(III) It is known that
el-emental chlorine can sometimes be produced in the
sample solutions According to the investigations
car-ried out by Krivan et al Se(IV) was slowly converted
to non-reactive Se(VI) due to back-oxidation by
ele-mental chlorine [19,20] If some iron is present in the
sample solution, it will form complexes with chloride,
thus preventing the formation of elemental chlorine
This may partly explain the better sensitivity obtained
for selenium in the samples-containing iron
The results also showed that when a lower
prereduc-tion temperature (50◦C) was used, the selenium signal
from the aqueous standard solution was greater to that
from the solution-containing iron When the
tempera-ture was raised to 70–90◦C, the selenium signal from
the iron solution was larger Thus, there is some
evi-dence that selenium may be lost more easily from an
aqueous solution (as a volatile chloride or hydride, for
example), than from the solution having an oxidizing
iron(III) Anyway, prereduction of selenium with HCl
is a common practice in HG [18]
To minimize the matrix effects, the determination of
selenium in PIX-110TMwas carried out using the
stan-dard addition method by adding 0.5–2.0g l−1Se(VI)
prior to prereduction
3.3 Validation of the methods
Detection limits (blank+3σ) and quantitation limits
(blank+ 10σ ) for As, Sb and Se were determined
in aqueous solutions that contained no iron matrix
The detection and quantitation limits are presented in
Table 7 The linear dynamic ranges for As, Sb and
Se (Table 8) were monitored in aqueous solutions
Precision and reproducibility was determined using
the standard addition method
Table 7
Detection limits (blank+3σ) and quantitation limits (blank+10σ )
in aqueous solutions for As, Sb and Se
Element Limit of detection
( g l −1) Limit of quantitation(g l −1)
Table 8 Linear dynamic ranges for As, Sb and Se in aqueous solutionsa Element Linear range
( g l −1) Calibration equation R
2 (n= 6) Arsenic 0–10 y = 0.3516x + 0.0557 0.9981
Antimony 0–30 y = 0.1173x − 0.0008 0.9997
Selenium 0–30 y = 0.1055x − 0.0307 0.9964
ay: integrated absorbance, x: concentration ing l −1.
3.4 Precision
The precision of the methods was tested using one-way variance analysis The determinations were performed according to the arrangement in Fig 4 The concentrations of As, Sb and Se in the samples were between 0.5 and 0.8g l−1 In the determination
of As the samples prepared contained 800 mg Fe l−1.
In the determination of Sb and Se the iron content of the samples was 2500 mg l−1.
The total variance of the results was composed of the residual variance and the influence of the day on which the measurement was made (s2
tot= s2 res+ s2 day)
A one-way F-test was used to test if the two variances
differed significantly from each other [21] According
to the F-tests, the variance arising from the
measure-ment day (s2
day) appeared to be significantly greater than the variance of the residual error for arsenic and selenium (Table 9) Considering the low concentra-tions of the element determined it was concluded that the precisions obtained (2–4% relative standard devi-ation, R.S.D.) were excellent
3.5 Reproducibility
The reproducibilities of the methods when car-ried out by more than one analyst were evaluated
Fig 4 Experimental arrangement used in the determination method precision.
Trang 8Table 9
F-test values, the variance components and the precisions for As, Sb and Se using the experimental arrangement in Fig 4
Element Mean (n= 25)
( g l −1) Variance of residualerrors (s2
res )
Variance of days (s2 day )
Precision F-test (MSdays /Ms res )
g l −1 R.S.D (%) Fcalc Ftheor(4, 20)a
a Degrees of freedom (numerator, denominator).
Fig 5 Experimental arrangement used in determination of
repro-ducibility.
according to Fig 5 The total variance of the results
is composed of the residual variance, the influence
of the day of measurement and the influence of the
analyst (s2
tot= s2
res+ s2
day+ s2 ana) According to F-tests
(Table 10), the variance arising from the
measure-ment day was significantly greater than the variance
of the residual error only for arsenic The effect of the
analyst was insignificant For antimony and selenium,
the only significant source of deviation was found to
be the residual errors Thus, the total variance of the
determinations equaled that of the residual error The
Table 11
The variance components and the overall reproducibilities for As, Sb and Se using the experimental arrangement presented in Fig 5
Element Mean value(n= 20) Variance of residual errors Variance of days Variance of analysts Reproducibility
(s2 res ) (s2
day ) (s2
ana ) g l −1 R.S.D (%)
a ins.: insignificant.
Table 10
F-test values for As, Sb and Se using the experimental arrangement
in Fig 5
Element F-test (MSdays /MS res ) F-test (MSanalyst /MS days )
Fcalc Ftheor (2, 16)a Fcalc Ftheor (1, 2)a
As 29.48 3.63 5.3 × 10 −4 18.51
a Degrees of freedom (numerator, denominator).
variance components and the reproducibilities for As,
Sb and Se are presented in Table 11 The discrepancies observed in the results were obviously due to the small number of measurement days and analysts (two)
3.6 Accuracy
The accuracies of the methods were tested using the standard reference materials SRM 361 Low-alloy Steel and Buffalo Rivel Sediment SRM 2704 Three replicates were prepared for all determinations and 5–8 successive atomizations were performed from each sample The certified and measured concentrations are shown in Table 12 The results obtained are in good agreement with the certified values
Trang 9Table 12
Analytical results of SRMs (mg kg −1) Mean± 95% confidence limits
Certified value Found Certified value Found Certified value Found
2704 Sediment 23.4 ± 0.8 22.8 ± 0.7 3.79 ± 0.15 3.77 ± 0.32 1.12 ± 0.05 1.09 ± 0.06
a Value uncertified.
Table 13
Limits for As, Sb and Se laid down by CEN for type 1 (highest
purity) iron chloride sulfate solutions
Element Limits for FeClSO 4 mg kg −1Fe(III)
3.7 Contents of antimony, arsenic and selenium in
PIX-110TM
The methods developed were used to measure the
As, Sb and Se concentrations in the PIX-110TM
The concentration of arsenic in the diluted sample
(1 ml/225 ml) was higher than the limit of
quantita-tion, and the arsenic content of 52g l−1was obtained
(0.29 mg kg−1Fe(III); calculated using a PIX-110TM
density of 1.5 g ml−1and Fe(III) content of 12 wt.%).
The measured concentrations for antimony and
sele-nium in diluted samples (1 ml/72 ml) were lower than
the limits of quantitation, so it was concluded that
results for selenium were<0.12 mg kg−1Fe(III) and
for antimony<0.10 mg kg−1Fe(III).
The European Committee for Standardization
(CEN) has laid down certain limits for toxic
sub-stances in different types of iron chloride sulfate
solutions used for the treatment of water intended for
human consumption The measured arsenic, antimony
and selenium concentrations in PIX-110TM were
lower than the limits shown in Table 13 and therefore
it fulfills the requirements for a type 1 (chemical of
highest purity) iron chloride sulfate solution [22]
4 Conclusions
KI–ascorbic acid was found to be the most
ef-fective prereductant/masking agent for As and Sb
interference when the iron content was very high The use of thiourea or l-cysteine did not prevent the signal depression caused by high iron content KI–ascorbic acid eliminated iron interference even
up to 2500 mg Fe l−1 The FI-HG system used was
found to be very sensitive and the samples could be diluted more than with the batch type hydride gener-ation system used earlier The absence of the memory effects also showed the excellent performance of the FI-system
The methods developed for As, Sb and Se are rapid, sensitive and can be used in routine analysis of sam-ples containing high iron concentrations The results obtained for the standard reference materials showed that the methods are accurate when very low analyte levels are measured in high iron matrix concentrations
Acknowledgements
The corresponding author is grateful to Kemira Chemicals Oy for financial support for this work
References
[1] G.F Kirkbright, M Taddia, Anal Chim Acta 100 (1978) 145.
[2] B Welz, M Melcher, Analyst 109 (1984) 569.
[3] B Welz, M Schubert-Jacobs, J Anal At Spectrom 1 (1986) 23.
[4] D Bax, J Agterdenbos, E Worrell, J.B Kolmer, Spectrochim Acta 43B (1988) 1349.
[5] R Bye, Talanta 33 (1986) 705.
[6] E Lugowska, I.D Brindle, Analyst 122 (1997) 1559 [7] Narsito, J Agterdenbos, D Bax, Anal Chim Acta 244 (1991) 129.
[8] T Wickström, W Lund, R Bye, J Anal At Spectrom 10 (1995) 803.
[9] C Boampong, I.D Brindle, X Le, L Pidwerbesky, C.M.C Ponzoni, Anal Chem 60 (1988) 1185.
Trang 10[10] B Welz, M Sucmanova, Analyst 118 (1993) 1425.
[11] H Chen, I.D Brindle, S Zheng, Analyst 117 (1992) 1603.
[12] B Welz, M Melcher, Analyst 109 (1984) 577.
[13] R Bye, Anal Chim Acta 192 (1987) 115.
[14] D.D Siemer, L Hagemann, Anal Lett 8 (1975) 323.
[15] M Yamamoto, M Yasuda, Y Yamamoto, Anal Chem 57
(1985) 1382.
[16] European Committee for Standardization, Chemicals Used
for Treatment of Water Intended for Human Consumption —
Iron(III) Chloride Sulfate, prEN 891, 1988, p 8.
[17] A.B Umetri, P.O Box 7960, S-90719 Umeå, Sweden [18] H Uggerud, W Lund, J Anal At Spectrom 10 (1995) 405 [19] V Krivan, J Anal At Spectrom 7 (1992) 155.
[20] V Krivan, K Patrick, B Welz, M Melcher, Anal Chem 57 (1985) 1703.
[21] D.L Massart, B.G.M Vandeginste, S.N Deming, Y Michotte,
L Kaufman, Chemometrics: A Textbook, 1st Edition, Elsevier, Amsterdam, 1988, pp 59–74.
[22] R Nilsson, Vesitippa, Publication of Kemira Chemicals Oy Kemwater 12 (2) (1996).