• Flows through the tip electrode cathode • Stimulates the heart • Returns through body fluid and tissue to the IPG anode A Unipolar Pacing System Contains a Lead with Only One Electrode
Trang 1Pacing Therapy for Bradyarrhythmias
Trang 2those of past years.
• Compared to the very first
pacemakers paced only in the lower chamber, we’ve had pacemakers that function in both the upper and lower chambers of the heart (DDD/R
pacemaker).
• One of the most notable development was a type of pacemaker that can change its pacing rate because of sensor(s) inside the pacemaker The DDDR pacemaker.
Trang 3• A pacemaker is really a system that has two
parts: a small metal
titanium can containing
the electronic circuitry
and a long-lasting battery, called a pulse generator and an insulated wire,
called a lead
What is a pacemaker?
Trang 4• Contains a battery that
provides the energy for
sending electrical
impulses to the heart
• Houses the circuitry that
Trang 6• Pulse generator: power
Trang 7• Endocardial or transvenous leads
• Myocardial/Epicardial leads
Types of Leads
Trang 8Transvenous Leads Have Different “Fixation” Mechanisms
Trang 9Transvenous Leads
• Active Fixation
– The helix (or screw)
extends into the
endocardial tissue
– Allows for lead positioning
anywhere in the heart’s chamber
Trang 10Myocardial and Epicardial Leads
• Leads applied directly to the
Trang 13Conduction Pathways
• Body tissues and fluids
are part of the conduction
pathway between the
anode and cathode
Tissue
Cathode Anode
Trang 14• Begins in the pulse
generator
• Flows through the lead and
the cathode (–)
• Stimulates the heart
• Returns to the anode (+)
During Pacing, the Impulse:
Impulse onset
*
Trang 15• Flows through the tip
electrode (cathode)
• Stimulates the heart
• Returns through body
fluid and tissue to the IPG (anode)
A Unipolar Pacing System Contains a Lead with Only One Electrode Within the Heart; In This System, the Impulse:
Cathode
Anode
+
Trang 16-Unipolar leads
• Unipolar leads may have
a smaller diameter lead body than bipolar leads
• Unipolar leads usually
exhibit larger pacing
artifacts on the surface ECG
Trang 18• Flows through the tip
electrode located at the end of the lead wire
• Stimulates the heart
• Returns to the ring
electrode above the lead tip
A Bipolar Pacing System Contains a Lead with Two Electrodes Within the Heart In This System, the Impulse:
Cathode
Trang 19Today Pacemakers
Trang 20• One lead implanted in
Trang 21Benefits of Dual Chamber Pacing
• Provide AV Synchrony
• Provide ventricular backup if A-to-V conduction is lost where a single chamber pacing system
cannot.
Trang 22Proven Benefits of Atrial Based Pacing
Study Results
Higano et al 1990
Gallik et al 1994 Santini et al 1991 Rosenqvist et al 1991
Sulke et al 1992
Improved cardiac index during low level exercise (where most patient activity occurs) Increase in LV filling
30% increase in resting cardiac output Decrease in pulmonary wedge pressure Increase in resting cardiac output
Increase in resting cardiac output, especially
in patients with poor LV function Decreased incidence of mitral and tricuspid valve regurgitation
Trang 23Proven Benefits of Atrial Based Pacing
Suppression of atrial dysrhythmias Improved morbidity (less AF, CHF, embolic events) after
3 plus uears, compared to VVI
Trang 24• Stimulate cardiac depolarization
• Sense intrinsic cardiac function
• Respond to increased metabolic demand by providing rate responsive pacing
• Provide diagnostic information stored by the pacemaker
Most Pacemakers Perform Four Functions:
Trang 25Rate Responsive Pacing
Trang 26Rate Response
• Rate responsive (also called rate modulated)
pacemakers provide patients with the ability to vary heart rate when the sinus node cannot provide the appropriate rate
• Rate responsive pacing is indicated for:
– Patients who are chronotropically incompetent (heart rate
cannot reach appropriate levels during exercise or to
meet other metabolic demands)
– Patients in chronic atrial fibrillation with slow ventricular
response
Trang 27Rate Responsive Pacing
• Cardiac output (CO) is determined by the combination
of stroke volume (SV) and heart rate (HR)
• SV X HR = CO
• Changes in cardiac output depend on the ability of the
HR and SV to respond to metabolic requirements
Trang 28Rate Responsive Pacing
• SV reserves can account for increases in cardiac output of up to 50%
• HR reserves can nearly triple total cardiac output in response to metabolic demands
Trang 29Rate Responsive Pacing
• When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate Rate Responsive Pacing Fixed-Rate Pacing
Daily Activities
Trang 30A Variety of Rate Response Sensors Exist
• Those most accepted in the market place are:
– Activity sensors that detect physical movement and increase
the rate according to the level of activity
– Minute ventilation sensors that measure the change in
respiration rate and tidal volume via transthoracic impedance readings
Trang 31Rate Responsive Pacing
• Activity sensors employ a
piezoelectric crystal that
detects mechanical
signals produced by
movement
• The crystal translates the
mechanical signals into
electrical signals that in
turn increase the rate of
the pacemaker
Piezoelectric crystal
Trang 32Rate Responsive Pacing
• Minute Ventilation (MV) is the volume of air introduced into the lungs per unit of time
• MV has two components:
– Tidal volume–the volume of air introduced into the lungs in a
single respiration cycle
– Respiration rate–the number of respiration cycles per minute
Trang 33Rate Responsive Pacing
• Minute ventilation can be measured by measuring the changes in electrical impedance across the chest cavity
to calculate changes in lung volume over time