DSpace at VNU: Triple-channel portable capillary electrophoresis instrument with individual background electrolytes for...
Trang 1Triple-channel portable capillary electrophoresis instrument with individual
background electrolytes for the concurrent separations of anionic and cationic species
Thanh Duc Mai, Minh Duc Le, Jorge Sáiz, Hong Anh Duong, Israel Joel Koenka,
Hung Viet Pham, Peter C Hauser
PII: S0003-2670(16)30117-9
DOI: 10.1016/j.aca.2016.01.029
Reference: ACA 234369
To appear in: Analytica Chimica Acta
Received Date: 30 November 2015
Revised Date: 12 January 2016
Accepted Date: 16 January 2016
Please cite this article as: T.D Mai, M.D Le, J Sáiz, H.A Duong, I.J Koenka, H.V Pham, P.C Hauser, Triple-channel portable capillary electrophoresis instrument with individual background electrolytes for
the concurrent separations of anionic and cationic species, Analytica Chimica Acta (2016), doi: 10.1016/
j.aca.2016.01.029.
This is a PDF file of an unedited manuscript that has been accepted for publication As a service to our customers we are providing this early version of the manuscript The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Trang 251 Flushing of the injection
interface 10 s 1 Open 2 Off Off
2 Flushing of the capillary 180 s 1 Open 1 Off Off
3 Flushing of the HV
interface 5 s 2 Closed 2 Off Off
4 Aspiration of sample into
the injection interface 5 s 1 Closed 2 Off On
5 Hydrodynamic injection variable 1 Closed 1 Off Off
6 Flushing of the injection
interface 5s 1 Open 2 Off Off
7 Electrophoretic separation variable 1 Closed 2 On Off
8 Flushing of the injection
interface 10 s 1 Open 2 Off Off
9 Flushing of the capillary 180 s 1 Open 1 Off Off
Trang 26r 2
LOD b (µM)
Precision of Migration Time (% RSD) c (n = 4)
Precision of Peak Area (% RSD) c (n = 4) Channel 1: Inorganic cations
Trang 29Compressed
air
Regulating valve
Injection manifold
Sample GND
W Capillary
HV
Pt
HV interface
Safety cage
C 4 D
W
Vent V1 1 2
BGE
BGE
Pt
Trang 30Pressure-limiting tubing
BGE
Trang 31600 500
400 300
12 3 4 5 6 7 8 9 10
450 400
350 300
250 200
300 250
300 250
Trang 32350 300
250 200
350 300
250 200
K+
Ca2+
Na+Mg2+
450 400
350 300
250 200
300 250
300 250
600 500
400 300
1
2
5 6 7
8 9 3
*
800 700
600 500
400 300
2 3
7 8 94
800 700
600 500
400 300
3 6 9
400 350
300 250
Trang 33300 250
200 150
Ace
*
800 700
600 500
400 300
5 2 1
6
*
800 700
600 500
400 300
4 6
1
Figure 5
450 400
350 300
250 200
300 250
200 150
Cyc Sac
400 350
300 250
200 150
Ace Sac
450 400
350 300
250 200
NH4+K+
Ca2+Na+Mg2+
800 700
600 500
400 300
4 6 8 9
*
450 400
350 300
250 200
Trang 34• Separate background electrolytes allow individual optimization
• The instrument is compact and field portable