Nội dung

Boonprasit Kritpracha, MD Department of Surgery, Prince of Songkla University Aortic Aneurysm Morphology Imaging: A Key to Success Planning for EVAR... Vascular Surgery Exposure befor

Trang 1

Boonprasit Kritpracha, MD

Department of Surgery, Prince of Songkla University

Aortic Aneurysm Morphology Imaging:

A Key to Success

Planning for EVAR

Trang 3

Vascular Surgery

Exposure before 1990

Vascular Surgery Exposure after 1990

A Fundamental Shift

Trang 5

Neck Proximal anchoring zone

Iliac arteries Distal “seal” zone

2 important zones

Trang 6

Imaging is Critical AAA Endograft Planning

Patient selection

Device selection – type, ? Extensions

Additional procedures – IIA coil embolization

Bell-bottom procedure

Surveillance

Endoleak,   aneurysm size

Morphological changes – aorta, device

Device failure – migration, wire fracture, etc

Trang 7

Preoperative imaging studies

Trang 8

AAA Endograft Planning

Trang 9

Axial CT Scan Problems

• Slice Thickness

• Axial slice artifact

• Length error

• Measurement process

Trang 10

Slice Thickness 10 mm

Slice Thickness 2 mm

Trang 11

Tortuosity makes neck appear elliptical

Axial Slice Artifact

Trang 12

Ouriel K, Green RM, et al, JVS 1992

Assuming the aortic neck is round

can lead to “axial slice” error

Trang 13

6.7% had not round neck difference > 4 mm

34 of 120 cases aortic neck not round

Trang 14

Axial 28 mm

33 mm

27 mm

Trang 16

32 mm

20 mm

Diameters of Iliac Arteries

Also need reformatted CT slices!!

Trang 17

Axial CT Scan Problems

• Axial slice artifact

• Measurement process

Trang 19

Axial CT Scan Problems

• Axial slice artifact

• Length error

Trang 20

Diameter Measurements

Measuring on small Axial Slices

contributes to exaggerated errors

Clinical Evidence

% Error: 10%+

Error: >+/- 2.5mm

Trang 22

Parallax & Magnification

Important during procedure - up to 3 cm

Can cause endograft failure

Need consistent system of control

Trang 23

AAA diameter is 61 mm

Thrombus effect

Trang 24

Thrombus effect

Trang 25

78 o

Perception of angle depends on projection

Trang 26

Arteriography length measurement artifact

Trang 27

AAA Endograft Planning

CT scan + 3-D reconstruction

Trang 28

2 Cases