Trang 1Intravascular ultrasound (IVUS) is widely used as a major diagnostic and assessment technique that provides detailed cross-sectional imaging of blood vessels in the cardiac catheterization laboratory The first ultrasound imaging catheter system was developed by Bom and his colleagues in Rotterdam, the Netherland, in 1971.1 By the late 1980s, the first images of human vessels were recorded by Yock and his colleagues.2 Since then, IVUS has become a pivotal catheter-based imaging technology that can provide scientific insights into vascular biology and practical guidance for percutaneous coronary interventions (PCIs) in clinical settings In this chapter, IVUS and the other catheter-based imaging devices—optical coherence tomography (OCT), angioscopy and spectro scopy—are described These newly developed imaging technologies provide supplemental and unique insights into vascular biology as well
INTRAVASCULAR ULTRASOUND
Basics of IVUS and Procedures
The IVUS imaging systems use reflected sound waves
to visualize the vessel wall in a two-dimensional format analogous to a histologic cross-section In general, higher frequencies of ultrasound limit the scanning depth but improve the axial resolution, and current IVUS catheters used in the coronary arteries have center frequencies ranging 20–45 MHz
– Detection of Vulnerable Plaque
– Safety and Limitations
– Future Directions
• Angioscopy – Imaging Systems and Procedures
– Image Interpretation
– Clinical Experience
– Detection of Vulnerable Plaque
– Safety and Limitations
– Future Directions
• Spectroscopy
– Imaging Systems and Procedures
– Experimental Data
– Clinical Experience
– Safety and Limitations
Trang 2There are two different types of IVUS transducer systems: (i) the solid-state dynamic aperture system (the electronically switched multi-element array system) and (ii) the mechanically
rotating single-transducer system (Table 1 and Figs 1A and B)
Several types of artifacts can be observed common or unique to
each system (Figs 2A to D) With both systems, still frames and
video images can be digitally archived on local storage memory
or a remote server using digital imaging and communications in medicine (DICOM) Standard 3.0 Regardless of IVUS system used in the patient, both require preprocedural administration
of intravenous heparin (5,000–10,000 U), or equivalent anticoagulation along with intracoronary nitroglycerin (100–300 µg), to reduce the potential for coronary spasm
Normal Vessel Morphology
The interpretation of IVUS images is possible as the layers of
a diseased arterial wall can be identified separately Particularly
in muscular arteries, such as the coronary tree, the media of the vessel is characte rized by a dark band compared with the
intima and adventitia (Figs 3A and B) Differentiation of the
layers of elastic arteries, such as the aorta and carotid, can be problematic because media are less distinctly seen by IVUS However, most of the vessels currently treated by catheter techniques are muscular or transitional arteries These include the coronary, iliofemoral, renal and popliteal systems Therefore,
it is usually easy to identify the medial layer
FIGURES 1A AND B: Diagrams of two basic imaging catheter designs: (A) solid state and (B) mecha nical (A: bottom) an image obtained using
a solid-state catheter imaging system (B: bottom) an image obtained using a mechanical catheter imaging system
Trang 3TABLE 1
Comparison of two IVUS designs
Solid-state dynamic
aperture system Mechanically rotating single-transducer system
Basics An electronic solid state
catheter system with
multiple imaging elements
at its distal tip, providing
Products One system is
Features The imaging catheter has
64 transducer elements
arranged around the
catheter tip and uses a
Fr (compatible with 6 Fr guide catheters)
Image quality This imaging catheter has
better scanning depth but
poorer axial resolution
compared with the
mechanical systems
Higher frequencies improve the axial resolution Therefore, mechanical transducers have traditionally offered advantages in image quality compared with the solid-state systems Artifacts The guidewire runs inside
the IVUS catheter thereby
preventing guidewire
artifact
This system does not
require flushing with
saline
The guidewire runs outside the IVUS catheter, parallel to the imaging segment, resulting in guidewire artifact This system requires flushing with saline before insertion to eliminate any air in the path of the beam Incomplete flushing artifact may result in poor image quality
Contd
Trang 4Solid-state dynamic aperture system Mechanically rotating single-transducer system
This system eliminates nonuniform rotational distortion (NURD)
Since the solid-state transducer has a zone
to appear in one or more segments of the image The imaging catheters have excellent near-field resolution and do not require the subtraction of
a mask
Others Short
transducer-to-tip distance (10.5 mm) facilitates visualization of distal coronary anatomy
The pullback trajectory is stabilized and it reduces the risk of a nonuniform speed in a continuous pullback
Contd
The relative echolucency of media compared with intima and adventitia gives rise to a three-layered appearance (bright-dark-bright), first described in vitro by Meyer and his colleagues.3 Due to the lack of collagen and elastin compared
to neighboring layers, the media displays lower ultrasound reflection “Blooming”, a spillover effect, is seen in the IVUS image because the intimal layer reflects ultrasound more strongly than the media This results in a slight overestimation of the thickness of the intima and a corresponding under estimation of the medial thickness On the other hand, the media/adventitia border is accurately rendered, because a step-up in echo reflectivity occurs at this boundary and no blooming appears The adventitial and periadventitial tissues are similar enough
in echoreflectivity that a clear outer adventitial border cannot
in angiographically normal segments At the other extreme, patients with a significant plaque burden have thinning of the media underlying the plaque As a result, the media is often indistinct or undetectable in at least some part of the IVUS
Trang 5cross-section This problem is exacerbated by the blooming phenomenon Even in these cases, however, the inner adventitial boundary (at the level of the external elastic lamina) is always clearly defined For this reason, most IVUS studies measure and report the plaque-plus-media area as a surrogate measure for plaque area alone The addition of the media represents only a tiny percentage increase in the total area of the plaque The determination of the position of the imaging plane within the artery is one important aspect of image interpretation For example, an IVUS beam penetrates beyond the coronary artery, providing images of peri vascular structures, including
the cardiac veins, myocardium and pericardium (Figs 4A to
C) These structures provide useful landmarks regarding the
position of the imaging plane because they have a characteristic appearance when viewed from various positions within the arterial tree The branching patterns of the arteries are also
FIGURES 2A TO D: Common IVUS image artifacts: (A) A “halo” or
a series of bright rings immediately around the mechanical IVUS catheter is usually caused by air bubbles that need to be flushed out (B) Radiofrequency noise appears as alternating radial spokes or random white dots in the far-field The interference is usually caused
by other electrical equipment in the cardiac catheterization laboratory (C) Nonuniform rotational distortion (NURD) results in a wedge-shaped, smeared appearance in one or more segments of the image (between
12 O’clock and 3 O’clock in this example) This may be corrected by straightening the catheter and motor drive assembly, lessening tension
on the guide catheter, or loosening the hemostatic valve of the Y-adapter (D) Circumferential calcification causes reverberation artifact between
10 O’clock and 1 O’clock
Trang 6FIGURES 3A AND B: Cross-sectional format of a representative IVUS image The bright-dark-bright, three-layered appearance is seen in the image with corresponding anatomy as defined The “IVUS” represents the imaging catheter in the vessel lumen Histologic correlation with intima, media and adventitia are shown The media has lower ultrasound reflectance owing to less collagen and elastin compared with neighboring layers Since the intimal layer reflects ultrasound more strongly than the media, there is a spillover in the image, resulting in slight overestimation
of the thickness of the intima and a corresponding underestimation of the medial thickness
distinctive and help to identify the position of the transducer
In the left anterior descending (LAD) coronary artery system, for example, the septal perfora tors usually branch at a wider angle than the diagonals On the IVUS scan, the septals appear
to bud away from the LAD much more abruptly than the
diagonals (Figs 5A to D) The branching pattern and perivascular
landmarks, once understood, can provide a reference to the actual orientation of the image in space
IVUS Measurements
The IVUS images have an intrinsic distance calibration, which
is usually displayed as a grid in the image Elec tro nic caliper (diameter) and tracing (area) measure ments can be performed
at the tightest cross section, as well as at reference segments located proximal and distal to the lesion
In everyday clinical practice, where accurate sizing of devices is needed, vessel and lumen diameter measurements are important The maximum and minimum diameters (i.e the major and minor axes of an elliptical cross-section) are the most widely used dimensions The ratio of maximum to minimum diameter defines a measure of symmetry Area measure ments are performed with computer planimetry; lumen area is determined
Trang 8FIGURES 5A TO D: Pullback imaging sequence from mid to proximal portion of the left anterior descending (LAD) artery: (A) The mid and distal portions of the LAD often lie deeper in the sulcus than the proximal LAD and myocardium may be observed The pericardium is seen at the opposite site of myocardium (B and C) The septal branches emerge opposite to the pericardium, but the diagonal branches take off more superiorly The angle between the septal and the diagonal branches usually increases to as much as 180 degrees (D) The left circumflex artery emerges on the same side as the emergence of the diagonal branches
by tracing the leading edge of the blood/intima border, whereas vessel or external elastic membrane (EEM) area is defined as the area enclosed by the outermost interface between media and adventitia Plaque area or plaque-plus-media area is calculated
as the difference between the vessel and lumen areas; the ratio
of plaque to vessel area is termed percent plaque area, plaque
Trang 9burden or percent cross-sectional narrowing Area measurements can be added to calculate volumes using Simpson’s rule with the use of motorized pullback In general, the investigator selects the most normal-looking cross-section (i.e largest lumen with smallest plaque burden) occurring within 10 mm of the lesion with no intervening major side branches as the reference segment.4
Tissue Characterization
The IVUS can provide detailed information about plaque composition Regions of calcification are very brightly echo- reflective and create a dense shadow more peripherally from the catheter, a phenomenon known as “acoustic shadowing”
(Figs 6A to C) Shadowing prevents determination of the true
thickness of a calcific deposit and precludes visualization of structures in the tissue beyond the calcium Reverberation is another charac teristic finding with calcification It causes the appearance of multiple ghost images of the leading calcium
interface, spaced at regular intervals radially (Fig 2D) Like
calcium, densely fibrotic tissue appears bright on the ultrasound scan Fatty plaque is less echogenic than fibrous plaque The brightness of the adventitia can be used as a gauge to discriminate between predominantly fatty from fibrous plaque Therefore, an area of plaque that appears darker than the adventitia is fatty In an image of extremely good quality, the presence of a lipid pool can be inferred from the appearance of
a dark region within the plaque (Figs 7A and B) Furthermore,
the “hot” lesions like ruptured plaques responsible for unstable angina or acute coronary syndromes can be observed by IVUS
(Figs 8A and B)
Recently, the clinical impact of attenuated plaques zed as hypoechoic plaque with ultrasound attenuation despite
characteri-little evidence of calcium has been reported (Figs 9A to C)
These specific plaques are more often seen in patients with acute coronary syndromes than in those with stable angina and are characterized by positive remodeling and nearby calcifica-tion.5 Clinical studies have indicated that attenuated plaques are associated with no reflow and creatine kinase-MB elevation after PCI because of distal embolization.6,7 This novel defined plaque may contain microcalcification, thrombus or cholesterol crystals.8
Visual interpretation of conventional grayscale IVUS images is limited in the detection and quantification of specific plaque components Therefore, computer-assisted analysis of raw radiofrequency (RF) signals in the reflected ultrasound
beam has recently been developed (Figs 10A to C) Virtual
Histology™ (VH) IVUS (Volcano Corporation, Rancho
Trang 11Cordova, California,) is recog nized as the first commercialized
RF analysis technology A classification algorithm developed from ex vivo human coronary data sets can generate color-mapped images of the vessel wall with a distinct color for each category: fibrous, necrotic, calcific and fibro-fatty.9 Another mathematical technique used in RF ultrasound backscatter analysis is Integrated Backscatter (IB) IVUS (YD Corporation, Nara, Japan) This method utilizes IB values, calculated as the average power of the backs cattered ultrasound signal from a sample tissue volume The IB-IVUS system constructs color-coded tissue maps, providing a quantitative visual readout as
FIGURES 7A AND B: Atherosclerotic plaque with lipid pool Lipid pool is defined as an echolucent area within the plaque and observed
at 8-2 O’clock in this IVUS image
FIGURES 8A AND B: Example of plaque rupture On the cross-sectional IVUS images (A), a cavity in contact with the vessel lumen is observed The longitudinal IVUS image (B) shows a spatial representation of the plaque rupture The rupture occurs in an eccentric plaque and has a residual thin flap that probably corresponds to a thin fibrous cap
Trang 13four types of plaque composition: calcification, fibrosis, dense fibrosis and lipid pool.10 Similar to these RF-based tissue characteriza tion techniques, iMap™ (Boston Scientific Inc, Natick, Massachusetts) has recently been introduced as an up-to-date tissue characterization method that is compatible with the latest 40-MHz mechanical IVUS imaging system (as opposed
FIGURES 10A TO C: Color-mapped images of the coronary plaque Conventional grayscale IVUS images (left) (A) Virtual Histology ™ shows
a distinct color for each of the fibrous, necro tic, calcific and fibro-fatty (B) Integrated Backscatter-IVUS can provide a quantitative visual readout
as four types of plaque composition: calcification, fibrous, dense fibrosis and lipid pool (C) iMap™ allows identification of four different types of plaque components (fibrotic, necrotic, lipidic and calcified tissue) with
a confidence level assessment of each plaque component (Source:
Figure A Dr Kenji Sakata)
Trang 14to VH-IVUS with 20-MHz solid-state IVUS system) The iMap allows identification and quanti fication of four different types of atherosclerotic plaque components: fibrotic, necrotic, lipidic and calcified tissues with accuracies at the high level
of confidence (95%, 97%, 98% and 98% for fibrotic, necrotic, lipidic and calcified tissues, respectively).11 Recently, multiple investigators have been trying to elucidate the clinical utility
of RF analysis technology, particularly for prediction of future adverse coronary events Providing regional observations to study predictors of events in the coronary tree (PROSPECT) trial is one of the largest natural history trials to employ three-vessel imaging with VH-IVUS in 700 acute coronary syndrome patients Multivariate analysis identified VH-IVUS determined thin-cap fibroatheroma (TCFA) (common type of vulnerable plaque defined as the presence of a confluent, necrotic core greater than 10% of plaque in contact with lumen at more than
30 degrees) at baseline as one of the independent predictors of future cardiac events (cardiac death, cardiac arrest, myocardial infarction, unstable angina or increasing angina) (HR = 3.35,
P <0.001).12
Insights into Plaque Formation and Distribution
Some of the classic pathologic findings in arterial disease have
been “rediscovered” in vivo by IVUS In a vessel that appears to
have a discrete stenosis by angiography, IVUS almost invariably shows considerable plaque burden throughout the entire length
of the vessel (Figs 11A and B) In fact, IVUS studies have
shown that the reference segment for an intervention which by defi nition is normal or nearly normal angiographically has, on average, 35–51% of its cross-sectional area occupied by plaque The phenomenon of remodeling, first described in human coronary specimens by the pathologist Glagov, is well illustrated
in vivo by IVUS (Figs 12A and B).13,14 The IVUS studies have also added to the original descrip tions in the pathology literature by demonstrating that the remodeling response is in fact bidirectional, with some segments showing the positive remodeling of the typical Glagov paradigm and others showing negative remodeling, or constriction, in the area of lumen
stenosis (Figs 12A and B).14 One important issue in evaluating this heterogeneous process by IVUS is the methodology used
to quantify and categorize arterial remodeling Although remodeling was originally conceptualized as a change in vessel size in response to plaque accumu lation over time, most histomorphometric or IVUS studies have relied on measure-ments of reference sites as a surrogate for the size of the vessel before it became diseased Therefore, results can vary distinctly accord ing to the choice of reference site as well as
Trang 16the manner of addressing vessel tapering.15 Theoretically, the use of the proximal reference, rather than the distal reference or the average of proximal and distal references, should preclude the potential influence of distal flow and pressure disturbance due to the presence of the IVUS catheter in the stenotic site
A remodeling index (the ration of vessel area at the lesion site
to that at the reference site) as a continuous variable may also
be preferable to categorical classifications, because arterial remodeling is considered to be a continuous biologic process
In fact, this remodeling index has been shown to conform to the normal frequency distribution in patients with chronic stable angina.16 The assessment of remodeling is clinically important, not only for optimal therapeutic device sizing but also for risk stratification regarding plaque rupture or evaluating procedural and long-term outcomes of intervention The vulnerable lesions responsible for acute coronary syndromes have usually undergone extensive positive remodeling A histopathologic study by Pasterkamp and his colleagues supported these clinical IVUS observa tions by demonstrating that positive remodeling
FIGURES 12A AND B: IVUS images showing remodeling: (A) Positive remodeling with localized expansion of the vessel in the area of plaque accumulation (B) Negative remodeling or shrinkage where the lesion has
a smaller media-to-media diameter than the adjacent less diseased sites
Trang 17is frequently associated with large, soft, lipid-rich plaques with increased inflammatory cell infiltrate.17 One IVUS study reported an association between preinterventional positive remodeling and creatine kinase elevation after intervention—a marker of distal embolization and future adverse cardiac events.18 Furthermore, other investigators directly showed that preinterventional positive remodeling assessed by IVUS predicts target lesion revascularization after coronary interventions.19Although the predictive values of these parameters in the context
of stenting have not been established with certainty, ventional IVUS may identify lesions with significant positive remodeling, providing triage information for increased risk of unfavorable outcomes and possible need of adjunctive biologic modalities for antirestenosis therapy in specific patients
preinter-Interventional Applications
According to the 2005 American College of Cardiology/American Heart Association/Society for Cardiovascular Angiography and Interventions (ACC/AHA/SCAI) 2005 Guideline Update for PCI, it is reasonable to use IVUS: (a) to assess the adequacy of coronary stent deploy ment, including the extent of apposition and minimum luminal diameter within the stent; (b) to determine the cause of stent restenosis and guide selection of appropriate therapy; (c) to evaluate coronary obstruction in a patient with a suspected flow-limiting stenosis when angiography is difficult because of location and (d) to assess a suboptimal angiographic result after PCI.20 In addition, not only after PCI but also before PCI, IVUS is a useful application to assess lesion characteristics
Preinterventional Imaging
Preinterventional IVUS has been used to clarify situa tions
in which angiography is equivocal or difficult to interpret (especially in ostial lesions or tortuous segments in which the angiogram may not lay out the vessel well for interpretation)
In addition, intermediate coronary lesions identified by angiography (40–70% angiographic stenosis) represent a challenge for revascularization decision-making Although anatomic evaluation does not provide direct estimation of hemodynamic significance of a given coronary lesion, minimum lumen area (MLA) measured by IVUS demonstrated good correlation with results from physiologic assessment The ischemic MLA threshold is 3.0–4.0 mm2 for major epicardial coronary arteries,21,22 and 5.5–6.0 mm2 for the left main coronary artery,23 based on physiologic assessment with coronary flow reserve, fractional flow reserve or stress scintigraphy
Trang 18Validated fractional flow reserve data have shown that deferring interventions in lesions with intermediate severity that are not considered hemodynamically significant (> 0.8 mm2) have a favorable clinical prognosis.24 Similarly, patients with intermediate coronary lesions in whom intervention was deferred based on IVUS findings (MLA >4.0 mm2) showed that the rate
of the composite endpoint was only 4.4% and target lesion revascularization 2.8%.25 As a result, IVUS imaging appears
to be an acceptable alternative to physiological assessment in patients presenting with intermediate coronary lesions Preinterventional IVUS imaging is also useful in determining the appropriate catheter-based intervention strategy With current IVUS catheters, most of the significant stenoses can be safely imaged before intervention providing detailed information about the circumferential and longitudinal extent of plaque as well as the character of the tissue involved This can lead to a change
in interventional strategy in 20–40% of cases.26,27 In particular, the presence, location and extent of calcium can significantly affect the results of balloon angioplasty, atherectomy and stent deployment The amount and distribution of plaque can be accurately determined and may favor atheroablative procedures
as primary or adjunctive treatment Precise measure ments of lesion length and vessel size can guide the optimal sizing of devices to be employed Detailed assessment of target lesion anatomy in the coronary tree is also useful to prevent major side branch encroach ment by intervention
Balloon Angioplasty
The IVUS imaging of percutaneous transluminal coronary angioplasty (PTCA) sites demonstrates plaque disruption or dissection more often than angiography does (40–70% of cases versus 20–40% by angio graphy).28,29 The IVUS is able
to characterize the depth and extent of dissections created by balloon inflation with relatively high accuracy Although the extent of dissections is relatively unpredictable, it is frequently possible to predict where tears will occur, based on certain morphologic features shown by IVUS If a plaque deposit is eccentric, tears usually occur at the junction between the plaque
and the normal wall (Figs 13A and B) This is presumably
because the non-diseased wall is more elastic than the plaque, and, with balloon inflation, it stretches away from the plaque, creating a cleavage plane running either within the media
or within the plaque substance, close to the media Another important factor in determining the location of tears is the presence of localized calcium deposits During balloon inflation, shear forces are highest at the junction between the calcium and the softer, surrounding plaque This creates an “epicenter” for the start of a tear, which then extends out to the lumen In
Trang 19lesions with localized calcification, cutting balloon angioplasty may be preferable, owing to its controlled tearing, to avoid the risk of unfavorable large dissections Creating dissections
in a controlled manner may also be beneficial to lessen acute elastic recoil after balloon angioplasty Data from phase I of the GUIDE trial showed that lesions with tears had less recoil than lesions that had not torn, suggesting that plaque tearing may effectively act to release the diseased segment from the mechanical constriction process caused by the plaque.28
Guidance of Procedures
A direct approach to balloon sizing, based on IVUS images, was pursued by the Clinical Outcomes with Ultrasound Trial (CLOUT) investigators, who reasoned that more aggressive balloon sizing might be more safely accomplished using the
“true” vessel size and plaque characteristics as determined by IVUS.30 In this prospective, nonrandomized study, balloon sizes were chosen to equal the average of the reference lumen and media-to-media diameters for cases in which the plaques were not extensively calcified This led to an average 0.5 mm
“oversizing” of the balloon compared with sizing based on standard angiographic criteria, and resulted in a significant decrease in post-procedure residual stenosis (from 28% to 18%) Importantly, there was no increase in clinically significant complications from this aggressive balloon sizing approach One-year follow-up of this trial showed a late adverse event rate (death, myocardial infarction or target lesion revas culari-zation) of 22%.31 This IVUS-guided aggressive PTCA strategy was expanded and confirmed by two single-center studies of provisional stenting, wherein balloon sizing was performed based on IVUS measure ments of media-to-media diameter at the lesion site.32,33 Angiographic or clinical follow-up of these
FIGURES 13A AND B: Examples of dissections: (A) A superficial dissection starts at 8 O’clock and extends counterclockwise (B) Eccentric plaque with deeper dissection is seen between 4 O’clock and 9 O’clock
A guidewire is seen inside the cavity of dissection
Trang 20studies also showed long-term outcomes equivalent to those
of elective stenting
Bare Metal Stent Implantation
The IVUS clearly visualizes stent struts as bright, distinct echoes Stents essentially provide a rigid scaffold against the force of vessel recoil During stent implantation, axial extrusion
of noncalcified plaque into the adjacent reference zones can occur.34 However, commensurate with the ability of the stent
to enlarge and hold open the treated segment, the extrusion effect in stenting may be more prominent than for balloon angioplasty Extrusion of plaque may also contribute to the step-up/step-down appearance on angiography, as well as some
of the side branch encroachment seen after stent deployment
Guidance of Procedures
The IVUS has identified several stent deployment issues, including incomplete expansion and incomplete apposition
(Figs 14A to C) Incomplete expansion occurs when a portion
of the stent is inadequately expanded com pared with the distal and proximal reference dimen sions, as may occur where dense fibrocalcific plaque is present Incomplete apposition (seen in 3–15% of stent cases) occurs when part of the stent structure is not fully in contact with the vessel wall, possible increasing local flow disturbances and the potential risk for subacute thrombosis
in certain clinical settings Tobis and Colombo developed the current high-pressure stent deployment technique after their
FIGURES 14A TO C: The IVUS-detected problems with stent deployment: (A) Incomplete stent apposition with a gap between a portion
of the stent and the vessel wall between 6 O’clock and 10 O’clock (B) Incomplete stent expansion relative to the ends of the stent and the reference segments (C) An edge tear or “pocket flap” with plaque disruption at the stent margin
Trang 21collaboration in the early 1990s revealed an unexpectedly high percentage of these stent deployment issues.35,36
After stent implantation, tears at the edge of the stent (marginal tears or pocket flaps) occur in 10–15% of cases
(Figs 13A and B).37 These tears have been attributed to the shear forces created at the junction between the metal edge
of the stent and the adjacent, more compliant tissue or to the effect of balloon expansion beyond the edge of the stent (the
“dog-bone” phenomenon) Although minor nonflow-limiting edge dissections may not be associated with late angiographic in-stent restenosis, significant residual dissections can lead to
an increased risk of early major adverse cardiac events.38 The current practice in our laboratory is to determine from the IVUS image whether the tear appears to be flow-limiting (i.e whether there is an extensive tissue arm projecting into the lumen), and,
if so, an additional stent is placed to cover this region
Over the past decade, a number of studies have shown that IVUS-guided stent placement improves the clinical outcome
of bare metal stents.39–44 In the landmark trial, Multicenter Ultrasound-guided Stent Implantation in Coronaries (MUSIC) trial, three main IVUS variables were considered for assessing optimal stent deployment: (1) complete stent apposition over the entire stent length; (2) in-stent minimum stent area (MSA) greater than or equal to 90% of the average of the reference areas or 100% of the smallest reference area and (3) symmetric stent expansion with the minimum/maximum lumen diameter ratio greater than or equal to 0.7.45 This study highlights that appropriate evaluation of stent deployment by IVUS impacts restenosis rate
A subacute thrombosis rate of less than 2% was believed
to represent a reduction compared with non guided deployment, although, with current antiplatelet regimens, similar results can usually be achieved by high-pressure postdilation without IVUS confirmation Nevertheless, a number of studies have suggested a link between suboptimal stent implantation and stent thrombosis, including the predictors and outcomes of stent thrombosis (POST) registry, which demonstrated that 90% of thrombosis patients had suboptimal IVUS results (incomplete apposition, 47%; incomplete expansion, 52% and evidence of thrombus, 24%), even though only 25% of patients had abnormalities
on angiography.46 In a more recent study by Cheneau and his colleagues, these observations were replicated suggesting that mechanical factors continue to contri bute to stent thrombosis, even in this modern stent era, with optimized antiplatelet regimens.47 Although the use of IVUS in all patients for the sole purpose of reducing thrombosis is clearly not warranted given the costs, IVUS imaging should be considered in patients
Trang 22who are at particularly high risk for thrombosis (e.g slow flow)
or in whom the consequences of thrombosis would be severe (e.g left main coronary artery or equivalent)
The MSA, as measured by IVUS, is one of the strongest predictors for both angiographic and clinical restenosis after bare metal stenting.48–50 Kasaoka and his colleagues indicated that the predicted risk of restenosis decreases 19% for every
1 mm2 increase in MSA and suggested that stents with MSA greater than 9 mm2 have a greatly reduced risk of restenosis.49
In the can routine ultrasound improve stent expansion (CRUISE) trial, IVUS guidance by operator preferences increased MSA from 6.25 mm2 to 7.14 mm2, leading to a 44% relative reduction in target vessel revasculari zation at 9 months, compared with angiographic guidance alone.42 In the angiography versus IVUS-directed stent placement (AVID) trial, IVUS-guided stent implantation resulted in larger acute dimensions (7.54 mm2) than angiography (6.94 mm2), without
an increase in complications, and lower 12-month target lesion revascularization rates for vessels with angio graphic reference diameter less than 3.25 mm, severe stenosis at preintervention (> 70% angiographic diameter stenosis), and vein grafts.51However, some IVUS-guided stent trials produced controversial results,52,53 presumably due to differing procedural end points for IVUS-guided stenting, and the various adjunctive treatment strategies that were used in these trials in response to suboptimal results Overall, a meta-analysis of nine clinical studies (2,972 patients) demonstrated that IVUS-guided stenting significantly lowers 6-month angiographic restenosis [odds ratio = 0.75, 95% confidence interval (CI), 0.60–0.94; P = 0.01] and target vessel revasculari zation (OR = 0.62; 95% CI, 0.49–0.78; P = 0.00003), with a neutral effect on death and nonfatal myocardial infarction, compared to an angiographic optimization.54
Insights into Long-term Outcomes
Intimal proliferation rather than chronic stent recoil primarily causes in-stent restenosis Growth of neoin tima is usually greatest in areas with the largest plaque burden,55 and the intimal growth process seems to be more aggressive in diabetic patients.56 The IVUS can be helpful to differentiate pure intimal ingrowth from poor stent expansion in the treatment of in-stent
restenosis (Figs 15A and B) Using serial IVUS immediately
before and after balloon angioplasty for in-stent restenosis, Castagna and his colleagues57 demonstrated in 1,090 consecu tive in-stent restenosis lesions that 38% of lesions had an MSA of less than 6.0 mm2 Even with minimal neointimal hyperplasia, stent underexpansion can result in clinically significant lumen compromise For this type of in-stent restenosis, mechanical optimization is appropriate in most cases
Trang 23The IVUS can also track the response to treatment, with evidence that angioplasty of in-stent restenosis is followed by early lumen loss due to decompression and/or reintrusion of tissue immediately after inter vention This phenomenon was more prominent in longer lesions and in those with greater in-stent tissue burden, perhaps accounting for the worse long-term outcomes in diffuse versus focal in-stent restenosis Direct tissue removal, rather than tissue compression/extrusion through the stent struts, may help minimize early lumen loss due to this phenomenon Several investigators have reported
a considerable reduction in angiographic and/or clinical recurrence of in-stent restenosis in patients with diffuse in-stent restenosis treated with ablative therapies (directional coronary atherectomy, rotational atherectomy or laser angio-plasty) compared with PTCA alone.58–60
Drug-eluting Stent Implantation
In current clinical experience, IVUS observations of antiproliferative drug-eluting stents (DES) have shown a
striking inhibition of in-stent neointimal hyperplasia (Fig 15)
Thus, it comes as no surprise that since the introduction of DES, both the rate of restenosis and need for repeat revascularization
FIGURES 15A AND B: The IVUS images 8 months after stent ment: (A) A conventional bare metal stent shows a considerable amount
deploy-of neointima inside the stent (B) In contrast, significant suppres sion deploy-of instent neointimal proliferation is observed when a drug-eluting stent was used
Trang 24have been dramatically reduced Moreover, both statistical and geographic distri butions of neointimal hyperplasia can be signi ficantly different between biologic (DES) and mechanical (bare metal) stents, despite mechanical performances of DES being similar to those of conventional bare metal stents.61 In general, neointimal volume (as a percentage of stent volume) within bare metal stents follows a near-Gaussian or normal frequency distribution, with a mean value of 30–35% The standard deviation of this statistical distribution represents biologic variability in vascular response to acute and/or chronic vessel injury as a result of inter ventions In contrast, biologic modifica tions through DES often result in a non-Gaussian frequency distribution, with variable degrees of the tail ends Since restenosis corresponds to the right tail at the end of the distribution curve, a discrepancy between mean neointimal volume and binary or clinical restenosis can occur in DES trials Similarly, bare metal stents show a wide individual variation in geographic distribution of neointima along the stented segment, whereas some types of DES demonstrate predilection of in-stent neointimal hyperplasia for specific locations (e.g proximal stent edge) In serial IVUS studies with multi ple long-term follow-ups, neointima within nonresteno tic bare metal stents showed mild regression after 6 months.62 In contrast, both sirolimus- and paclitaxel-eluting stents showed
a slight but continuous increase in neointimal hyperplasia for
up to 4 years.63–65
Guidance of Procedures
The value of MSA remains as a powerful predictor for in-stent restenosis in the DES era.66,67 A recent IVUS work by Sonoda and his colleagues demonstrated that sirolimus-eluting stents showed a stronger positive relation, with a greater correlation coefficient between baseline MSA and 8-month MLA, compared
to control bare metal stents (0.8 vs 0.65 and 0.92 vs 0.59, respectively).66 The utility of IVUS to ensure adequate stent expansion cannot be overemphasized, particularly if there are clinical risk factors for DES failure (e.g diabetes, renal failure)
In this context, preinterventional IVUS can provide useful information about plaque composition In particular, calcified plaque is important to identify, because the presence, degree and location of calcium within the target vessel can substantially affect the delivery and subsequent deployment of coronary stents
(Fig 6) One important advantage of online IVUS guidance is
the ability to assess the extent and distance from the lumen of calcium deposits within a plaque For example, lesions with extensive superficial calcium may require rotational atherectomy before stenting Conversely, apparently significant calcification
Trang 25on fluoroscopy may subsequently be found by IVUS to be distributed in a deep portion of the vessel wall or to have a lower degree of calcification (calcium arc < 180 degrees) In these cases, stand-alone stenting is usually adequate to achieve
a lumen expansion large enough for DES deployment
The stent deployment techniques on clinical outcomes of patients treated with the cypher stent (STLLR) trial demons-trated that geographic miss (defined as the length of injured or stenotic segment not fully covered by DES) had a significant negative impact on both clinical efficacy (target vessel and lesion revascularization) and safety (myocardial infarction) at 1 year after sirolimus-eluting stent implantation.68 These findings suggest that less aggressive stent dilation and complete coverage
of reference disease may be beneficial, as long as significant underexpansion and incomplete strut apposition are avoided Another single center study showed optimal stent longitudinal positioning of sirolimus-eluting stents using unique stepwise IVUS criteria (mainly targeting the sites with plaque burden < 50%) In this study, plaque burden in the reference lesion was the strongest predictor of stent margin restenosis.69 Online IVUS guidance can facilitate both the determination of appropriate stent size and length and the achievement of optimal procedural end points, with the goal being to cover significant pathology with reasonable stent expansion while anchoring the stent ends
in relatively plaque-free vessel segments The efficacy of DES
is related not only to the pharmacological (drug and polymer) kinetics but also to how well the stent is deployed within the coronary artery
Insights into Long-term Outcomes
Several large studies have assessed the impact of IVUS guidance during DES implantation on long-term clinical outcomes In a single-center study of IVUS-guided DES implantation versus propensity score matched control population with angiographic guidance alone, a higher rate of definite stent thrombosis was seen in the angiography-guided group at both 30 days (0.5% vs 1.4%, P = 0.046) and 12 months (0.7% vs 2.0%, P = 0.014).70 In addition, a trend was seen in favor of IVUS guidance in 12-month target lesion revascularization (5.1% vs 7.2%, P = 0.07) In addition, recent results of the revascularization for unprotec-ted left main coronary artery stenosis: comparison of percuta-neous coronary angioplasty versus surgical revascularization (MAIN-COMPARE) registry showed significantly lower 3-year mortality in the IVUS-guidance group as compared with the conventional angiography-guidance group (4.7% vs 16.0%, log-rank P = 0.048) in patients treated with DES.71 Despite the growing evidence of the benefits of IVUS-guided DES
Trang 26implantation, few multicenter studies have been conducted to prove this hypothesis in a randomi zed controlled fashion The Angiographic versus IVUS Optimization (AVIO) study was the first randomized trial designed to establish modern, universal criteria for IVUS optimization of DES implantation in complex coronary lesions.72,73 This study proposed unique optimization criteria in which the target stent area was determined according
to the size of a post-dilation, noncompliant balloon chosen
on the basis of IVUS-measured media-to-media diameters at multiple different sites within the stented segment Post-proce-dure minimum lumen diameter, as the primary endpoint of this study, was significantly larger in the IVUS-guided group, particularly when optimal IVUS criteria were met, with no increased complication as compared to the angiography-guided group (target IVUS criteria met: 2.86 mm, target IVUS criteria not met: 2.6 mm, angiography alone: 2.51 mm) Further studies with a larger population are required to determine whether this acute benefit in complex lesions can translate into improved long-term clinical outcomes
Due to the low incidence of DES failure, clarification of its exact mechanisms awaits the cumulative analysis of large clinical studies Nevertheless, suboptimal deployment or mechanical problems appear to contri bute to the development of both restenosis and throm bosis Particularly, the most common mechanism is stent underexpansion, the incidence of which has been reported as 60–80% in DES failures In a study of 670 native coronary lesions treated with sirolimus-eluting stents, the only independent predictors of angiographic restenosis were postprocedural final MSA and IVUS-measured stent length (OR
sirolimus-of nonrecurrent lesions These observations emphasize the importance of procedural optimization at DES implan tation
for both de novo and in-stent restenosis lesions.
Although published data on DES thrombosis are further limited, one single-center IVUS study reported stent underexpansion (P = 0.03) and a significant residual reference segment stenosis (P = 0.02) as indepen dent multivariate predictors of sirolimus-eluting stent thrombosis (median time,
14 days after implantation).75 The IVUS features of stent thrombosis from another single-center IVUS study appear analogous to the previous observations.76
Trang 27For very late DES thrombosis (> 12 months), another investigator group has suggested incomplete stent apposition as a possible risk factor.77 Late-acquired incomplete stent apposition with DES has been reported in both experimental (paclitaxel)78
and clinical (sirolimus and paclitaxel) studies (Fig 16).79–82Several IVUS studies have indicated that the main mechanism
is focal, positive vessel remodeling (Figs 17A and B).79,81 In addition, there is strong suggestion that incompletely apposed struts are seen primarily in eccentric plaques, and that gaps develop mainly on the disease-free side of the vessel wall Thus, the combination of mechanical vessel injury during stent implantation and biologic vessel injury with pharmacologic agents or polymer in the setting of little underlying plaque may predispose the vessel wall to chronic, pathologic dilation
(Figs 18A and B) Despite a recent meta-analysis suggesting
an increased risk of late/very late stent thrombosis in patients with late-acquired incomplete stent apposition,83 it remains controversial whether this morphologic abnormality indepen-dently contributes to the occurrence of stent thrombosis.84–86 Other IVUS-detected conditions that may be of importance
in DES include nonuniform stent strut distribution and stent
fractures after implantation (Figs 19A and B) Theoretically,
FIGURE 16: Classification of incomplete stent apposition (ISA) The ISA observed at follow-up is either persistent from baseline or late acquired Late-acquired ISA without vessel remodeling is typically seen
in thrombus-containing lesions, whereas late-acquired ISA with focal, positive vessel remodeling is more characteristic to drug-eluting stents
Trang 31both abnormalities can reduce the local drug dose delivered to the arterial wall, as well as affecting the mechanical scaffolding
of the treated lesion segment By IVUS, strut fracture is defined
as longitu dinal strut discontinuity and can be categorized based upon its morphological characteristics: (1) strut separa tion;
(2) strut subluxation or (3) strut intussusceptions (Fig 20).87
A recent IVUS study of 24 sirolimus-eluting stent restenosis cases identified the number of visualized struts (normalized for the number of stent cells) and the maximum interstrut angle as independent multi variate IVUS predictors of both neointimal hyperplasia and MLA.88 In addition, angiographic or IVUS studies have reported the incidence of DES fracture as 0.8–7.7%, wherein in-stent restenosis or stent thrombosis occurred at 22–88%.89 The exact incidence and clinical implica tions of strut fractures remain to be investigated in large clinical studies
Safety
As with other interventional procedures, the possibility of spasm, dissection and thrombosis exists when intra vascular imaging catheters are used In a retrospective study of 2,207 patients, Hausmann and his colleagues identified spasm in 2.9% of patients, and other complications, including dissection, thrombosis and abrupt closure with “certain relation” to IVUS,
in 0.4%.90 Another multicenter European registry revealed 1.1% complications were reported (spasm, vessel dissection
or guidewire entrapment) in a total of 718 examina tions.91These studies were performed with first-generation catheters
in the 1990s, and it is likely (although not documented) that the incidence of spasm and other complications is substantially lower with the current generation of catheters
Future Directions
An interesting approach would be to combine IVUS with
a therapeutic device, such as balloon catheter In 2010, one angioplasty balloon catheter to integrate IVUS imaging
FIGURE 20: Classification of stent strut fracture By IVUS, strut fracture
is defined as longitudinal strut discontinuity and can be categorized based on morphological characteristics
Trang 32(Vibe™ RX, Volcano Corporation, Rancho Cordova, California) gained CE-mark clearance in Europe This new device can provide precise IVUS-guided balloon dilatation with immediate confirmation of interventional results without additional catheters or catheter exchanges.
Another interesting device iteration is “forward-looking” IVUS which can visualize the vessel wall in front of the imaging catheter thereby having the potential to visualize the true and false lumens in chronic total occlusion (CTO) lesions This enhanced visualization could be used to improve CTO crossing
by continually maintaining and directing the catheter or wire toward the true lumen.92
Currently, commercially available IVUS catheters used in the coronary arteries have center frequencies ranging from 20 MHz to 45 MHz with the highest frequency IVUS being the
45 MHz Revolution™ IVUS catheter (Volcano Corporation, Inc., Rancho Cordova, California) In general, higher frequen-cies of ultrasound improve the axial resolution On the other hand, higher frequency IVUS may result in stronger scattering echoes from the blood, hampering visualization of the vessel lumen and thereby limiting the scanning depth Therefore, IVUS frequencies higher than the current 45 MHz range IVUS may have inherent limitations By overcoming these limitations, the next generation higher frequency IVUS catheter will enable
better axial resolution (Figs 21A to C) Theoretically, an increase
of center frequency from 40 MHz to 50 MHz corresponds to a 25% improvement in axial resolution if the design is similar
OPTICAL COHERENCE TOMOGRAPHY
The principal technology was developed and first described
by researchers at the Massachusetts Institute of Technology
in 1991.93 Since then, optical coherence tomography (OCT) has been applied clinically in ophthalmology, dermatology, gastroenterology and urology Currently, intracoronary OCT has
emerged as an in vivo optical microscopic imaging technology,
as it generates real-time tomographic images from back scattered reflections of infrared light Thus, the use of optical echoes by OCT can be regarded as an optical analog of IVUS, with its greatest advantage being its significantly higher resolution (10 times or greater) compared to conventional pulse-echo and other ultrasound-based approaches
Imaging Systems and Procedures
The imaging catheter includes a fiberoptic core with a microlens and prism at the distal tip to generate a focused scanning beam perpendicular to the catheter axis, thereby providing circum-ferential imaging of the arterial wall In standard OCT systems,
Trang 33FIGURES 21A TO C: Comparing variable frequency IVUS images Higher frequency IVUS can produce improved image quality due to
higher resolution (60 MHz IVUS image) (Source: Silicon Valley Medical
Instrument, Inc., CA)
the optical engine includes a superluminescent diode as a source
of low coherent, infrared light, typically with a wave length around 1300 nm The first commercialized intravascular OCT device (St Jude Medical, Inc St Paul, Minnesota) consisted
of a guidewire-based imaging catheter with a profile of 0.014 inches, a proximal low-pressure occlusion balloon catheter, and a system console containing the optical imaging engine and computer for signal acquisition, analysis and image reconstruction The imaging procedure of intravascular OCT
is similar to that of IVUS except that blood must be displaced
by saline or contrast medium while imaging Technically, this
is because the dominant mode of signal attenuation in OCT is multiple scattering, so that additional scattering by red blood
cells results in very large signal loss (Fig 22B) During OCT
image acquisition, blood flow is interrupted by inflating the balloon with a modest amount of liquid flush from the distal flush exit ports of the occlusion catheter The balloon inflation
is performed at a low pressure to avoid unnecessary vessel stretching Although this first generation intravascular OCT
Trang 34system was not approved by the United States Food and Drug Administration, the Fourier-domain OCT system (the so-called second generation OCT system) (St Jude Medical, Inc.) was approved in 2010 Other companies have been develop ing similar rapid-scan OCT systems, referred to as optical frequency-domain imaging (OFDI) This technique measures optical echo time delay using a light source whose light output can be rapidly swept over a range of wavelengths (e.g 1,260–1,360 nm) Fourier transform techniques enable conversion of the frequency-domain (or wavelength dependent) data to be converted to a time-domain representation While first generation OCT (time-domain OCT) systems have a frame rate of 4–20 frames/sec, the Fourier/frequency-domain OCT achieves 80–110 frames/sec acquisition, allowing comprehen-sive scanning of long arterial segments during one bolus flush through the guide catheter without the need for occlusion Since the OCT catheter has a short guidewire lumen at the distal portion of the catheter tip, the guidewire can be seen as
a point artifact with shadowing (Figs 22A to D).
Image Interpretation
The higher resolution of OCT can often provide superior delineation of each structure compared with IVUS The OCT can reliably visualize the microstructure (i.e 10–50 µm, vs 150–200 µm for IVUS) of normal and pathologic arteries Typically, the media of the vessel appears as a lower signal intensity band than the intima and adventitia, providing a
three-layered appearance similar to that seen by IVUS (Figs
23A and B) Atheromatous lesions and fibrous plaques exhibit
homo geneous, signal-rich (highly backscattering) regions; lipid-rich plaques exhibit signal-poor regions (lipid pools) with poorly defined borders and overlying signal-rich bands (corresponding to fibrous caps); and calcified plaques exhibit
signal-poor regions with sharply delineated borders (Figs 24A
to C) The OCT has the advantage of being able to image
through calcium without shadowing, as would be seen with IVUS On the other hand, signal penetra tion through the diseased arterial wall is generally more limited (no more than 2 mm with current OCT devices), making it difficult
to investigate deeper portions of the artery or to track the entire circumference of the media-adventitia interface Plaque
characteristics of OCT versus IVUS are listed in Table 2.
The diagnostic accuracy of OCT for the above plaque characterization criteria was confirmed by an ex vivo study of
307 human atherosclerotic specimens including aorta, carotid, and coronary arteries.94 Independent evaluations by two OCT analysts demonstrated a sensiti vity and specificity of 71–79%
Trang 36and 97–98% for fibrous plaques; 90–94% and 90–92% for
lipid-rich plaques and 95–96% and 97% for fibrocalcific plaques,
respectively (overall agreement vs histopathology, κ values of
0.88–0.84) The interobserver and intra observer reproducibility
of OCT assessment was also high (κ values of 0.88 and 0.91,
respectively)
Clinical Experience
In the first coronary OCT study in humans reported by Jang
and his colleagues, 17 coronary segments in 10 patients were
imaged with 3.2F OCT catheters (modified IVUS catheters)
during intermittent saline flushes through the guide catheter.95
The maximum penetration depth of OCT imaging measured
1.25 mm versus 5 mm for IVUS In vivo axial resolutions,
TABLE 2
Plaque characteristics of OCT versus IVUS
Fibrous Homogeneous Homogeneous
Signal-rich (highly High echogenicity backscattering)
Calcium Sharply delineated borders Very high
Lipid Poorly defined borders Low echogenicity
Signal-poor
FIGURES 23A AND B: Example of cross-sectional image format of
OCT The bright-dark-bright, three-layered appearance is seen in the
image with corresponding anatomy as defined Histologic correlation
with intima, media and adventitia are shown The OCT shows the three
layer appearance of normal vessel wall, with the muscular media being
revealed as a low signal layer comprised between intima and adventitia
Trang 38determined by measuring the full-width half-maximum of the first derivative of a single axial reflectance scan at the surface of the tissue, were 13 ± 3 µm with OCT versus 98 ± 19 µm with IVUS All fibrous plaques, macrocalcifi cations and echolucent regions identified by IVUS were visuali zed in corresponding OCT images Intimal hyperplasia and echolucent regions, which may corres pond to lipid pools, were identified more frequently
by OCT than by IVUS
In addition, recent clinical reports by the same investi gator group showed that intravascular OCT detec ted lipid-rich plaques and thrombus more fre quently in acute myocardial infarction
or unstable angina than in stable angina lesions.96 In a recent study using the first commercialized OCT system, Kubo and his colleagues reported that the plaque rupture and thrombus in patients with acute myocardial infarction were identified more frequently by OCT than by IVUS (73% vs 40%, P = 0.009,
and 100% vs 33%, P < 0.001) (Figs 25A and B).97
Encouraging preliminary results have been reported in the assessment of coronary interventions as well Bouma and his colleagues successfully imaged 42 coronary lesions before and immediately after stenting.98 In this series, OCT detected dissections, instent tissue prolapse and incomplete
stent apposition more often than IVUS (Figs 26A to C) With
a dedicated OCT catheter, Grube and his colleagues reported a follow-up OCT examination 6 months after drug-eluting stent implan tation for the treatment of instent restenosis.99 The high resolution of OCT allowed clear visualization of the overlapped stents (stent-in-stent), distinctly identifying each stent strut as well as a very thin neointimal layer covering the drug-eluting
stent struts (Figs 27A and B) More recently, some investigator
groups have reported that OCT images can visualize the thin neointima on each stent strut and quantify its thickness after
FIGURES 25A AND B: Culprit lesion in the left anterior descending (LAD) artery in a patient with unstable angina: (A) Coronary angiogram shows significant lumen narrowing at the proximal portion of LAD (B) OCT clearly visualizes the protruded thrombus (asterisk) attached to the plaque rupture site (arrow)
Trang 39FIGURES 26A TO C: Stent deployment problems detected by OCT: (A) Incomplete stent apposition, in which there is a gap between a portion
of the stent and the vessel wall between 6 O’clock and 10 O’clock (B) Tissue prolapse between the stent struts at 6 to 7 O’clock (C) An edge tear or “pocket flap” with a disruption of plaque at the stent margin
FIGURES 27A AND B: OCT images 8 months after stent deployment: (A) OCT visualizes the stent struts covered by a thick neointima that appeared as a bright luminal layer surrounding the stent struts after bare metal stent implantation (B) The OCT shows stent struts and thin, bright reflective tissue coverage after drug-eluting stent implantation
drug-eluting stent implantation.100,101 The OCT revealed the majority of stent struts were covered by a thin neointima layer less than 100 µm thick, which is beyond IVUS resolution cap-abilities, 6 months after sirolimus-eluting stent implantation.100
In addition, Otake and his colleagues reported that subclinical thrombus after sirolimus-eluting stenting was significantly associated with longer stents, a larger number of uncovered struts, and greater average of neointimal unevenness score (maximum neointimal thickness in the cross section/average neointimal thickness of the same cross section).102 Although the exact clinical impact of intravascular OCT findings requires
Trang 40systematic evaluation, these preliminary reports have confirmed that this new imaging technology has the potential to provide a new level of anatomic detail, not only as a research technique
but also as a clinical tool (Figs 28A to C) In fact, OCT has
been used as a tool for evaluating neointimal proliferation after commercially available stents or newly developed stents in some multicenter trials.103–106 OCT for DES SAfety (ODESSA) reported the frequency of uncovered stent struts at 6 months
in overlapped segments (sirolimus-eluting stent, 8.7 ± 13.3%; paclitaxel-eluting stent, 8.3 ± 20.9%; zotarolimus-eluting stent, 0.05 ± 0.19%; bare metal stents, 1.8 ± 4.0%) and in non overlapped segments (sirolimus-eluting stent, 7.9 ± 11.3%; paclitaxel-eluting stent, 2.3 ± 4.1%; zotarolimus-eluting stent, 0.01 ± 0.05%; bare metal stents, 0.5 ± 2.2%).103 In the OCT in acute myocardial infarction (OCTAMI) study, uncovered stent struts were reported in 0.00% of zotarolimus-eluting stent and
in 1.98% of bare metal stents (P = 0.13) 6 months after stent implantation in patients with ST-segment elevation myocardial infarction.106
Detection of Vulnerable Plaque
One of the most valuable challenges for OCT is its role in the detection of vulnerable plaque OCT is often able to identify a thin fibrous cap of vulnerable plaque, the thickness of which (<65 µm) is technically below the image resolution of IVUS (~ 150 µm) The TCFA, that is the primary plaque type at the site of plaque rupture, exists in nonculprit lesions and is distributed in all three coronary arteries Fujii and his colleagues performed three-vessel OCT examination in patients with
FIGURES 28A TO C: Strut assessment by OCT in relation to the vessel wall Due to a blooming effect of metal struts, the highest intensity point within the strut image should be used for the measurement Strut apposition to the vessel wall is determined by measuring the distance from the stent strut surface to the vessel wall as compared to the nominal strut thickness (Nominal strut thickness; Cypher ® , Cordis, Johnson and Johnson Co, Miami, FL; 154 µm, TAXUS Liberté ® , Boston Scientific Inc, Natick, MA; 115 µm, Endeavor ® , Medtronic, Santa Rosa, CA; 96 µm, Xience™V, Abbott Vascular, Santa Clara, CA; 89 µm)