BEAM PROPERTIES The ultrasound beam propagates as a longitudinal wave from the transducer surface into the propagation medium, and exhibits two distinct beam patterns: • a slightly conv
Trang 1Transducers
Trang 2 Ultrasound is produced and detected with a transducer, composed of one or more ceramic elements with
electromechanical (piezoelectric)
properties
• The ceramic element converts electrical
energy into mechanical energy to produce ultrasound and mechanical energy into
electrical energy for ultrasound detection
Trang 3 Over the past several decades, the transducer assembly has evolved considerably in design, function, and capability, from a single-element resonance crystal to a broadband transducer array of hundreds of individual elements
transducer has major components including the
Trang 5Piezoelectric Materials
A piezoelectric material (often a crystal
or ceramic) is the functional component
of the transducer
• It converts electrical energy into mechanical (sound) energy by physical deformation of the crystal structure
Trang 6 ConverseIy, mechanical pressure
applied to its surface creates electrical energy
• Piezoelectric materials are characterized by a well-defined molecular arrangement of
electrical dipoles (Fig 16-9)
Trang 7 An electrical dipole is a molecular entity containing positive and negative electric charges that has no net charge
• When mechanically compressed by an
externally applied pressure, the alignment of the dipoles is disturbed from the equilibrium position to cause an imbalance of the charge distribution
Trang 8 A potential difference (voltage) is created across the element with one surface
maintaining a net positive charge and
one surface a net negative charge
• Surface electrodes measure the voltage,
which is proportional to the incident
mechanical pressure amplitude
Trang 9 Conversely, application of an external
voltage through conductors attached to the surface electrodes induces the
mechanical expansion and contraction of the transducer element
Trang 10 There are natural and synthetic
piezoelectric materials
• An example of a natural piezoelectric material
is quartz crystal, commonly used in watches and other time pieces to provide a mechanical vibration source at 32.768 kHz for interval
timing
quartz, determined by the crystal cut and machining properties.
Trang 11 Ultrasound transducers for medical imaging applications employ a synthetic piezoelectric ceramic, most often lead-zirconate-titanate
Trang 12 For PZT in its natural state, no piezoelectric
properties are exhibited; however, heating the material past its “Curie temperature” (i.e., 3280
C to 3650 C) and applying an external voltage
causes the dipoles to align in the ceramic
has cooled to below its Curie temperature
• Once the material has cooled, the dipoles retain their alignment
Trang 13 At equilibrium, there is no net charge on ceramic surfaces
• When compressed, an imbalance of charge produces a voltage between the surfaces
electrodes attached to both surfaces, mechanical deformation occurs.
Trang 14 The piezoelectric element is composed
of aligned molecular dipoles
Trang 15 Under the influence of mechanical
pressure from an adjacent medium
(e.g., an ultrasound echo), the element thickness
• Contracts (at the peak pressure amplitude),
• Achieves equilibrium (with no pressure) or
• Expands (at the peak rarefactional pressure),
produce positive and negative surface charge
Trang 17 Surface electrodes (not shown)
measure the voltage as a function of time
Trang 18 An external voltage source applied to the element surfaces causes compression or expansion from equilibrium by
realignment of the dipoles in response to the electrical attraction or repulsion
force
Trang 20Resonance Transducers
Resonance transducers for pulse echo
ultrasound imaging are manufactured to
operate in a “resonance” mode, whereby a
voItage (commonly 150 V) of very short
duration (a voltage spike of ≈ 1 µ sec) is
applied, causing the piezoelectric material to initially contract, and subsequently vibrate at a natural resonance frequency
to the preferential emission of ultrasound waves
whose wavelength is twice the thickness of the
Trang 21 The operating frequency is determined from the speed of sound in, and the
thickness of, the piezoelectric material
• For example, a 5-MHz transducer will have a wavelength in PZT (speed of sound in PZT is
≈ 4,000 m/sec) of
mm meters
m f
c
80 0 10
8 sec
/ 10 5
sec /
Trang 22by the thickness of the transducer equal
to 1/A
Trang 23 To achieve the 5-MHz resonance
frequency, a transducer element
thickness of ½ X 0.8 mm = 0.4 mm is
required
• Higher frequencies are achieved with thinner elements, and lower frequencies with thicker elements
preferentially at a single “center frequency.”
Trang 24Damping Block
The damping block, layered on the back of the piezoelectric element, absorbs the backward directed ultrasound energy and attenuates
stray ultrasound signals from the housing
in create an ultrasound pulse width a short spatial
pulse length, which is necessary to preserve detail
along he beam axis (axial resolution)
Trang 28 Dampening of the vibration (also known
as “ring-down”) lessens the purity of the resonance frequency and introduces a broadband frequency spectrum
• With ring-down, an increase in he bandwidth (range of frequencies) of he ultrasound pulse occurs by introducing higher and lower
frequencies above and below the center
(resonance) frequency
Trang 29 The “Q factor” describes the bandwidth of the sound emanating from a transducer as
where f o is the center frequency and the
bandwidth is the width of the frequency
distribution.
Bandwidth
f
Trang 30 A “high Q” transducer has a narrow
bandwidth (i.e., very little damping) and
a corresponding long spatial pulse
length
• A “low Q” transducer has a wide bandwidth and short spatial pulse length
Trang 31 Imaging applications require a broad
bandwidth transducer in order to achieve high spatial resolution along the direction
of beam travel
• Blood velocity measurements by Doppler
instrumentation require a relatively
narrow-band transducer response in order to
preserve velocity information encoded by
changes in the echo frequency relative to the incident frequency.
Trang 32 Continuous-wave ultrasound transducers have a very high Q characteristic
• While the Q factor is derived from the term
quality factor, a transducer with a low Q does
not imply poor quality in the signal.
Trang 33Matching Layer
The matching layer provides the interface
between the transducer element and the tissue and minimizes the acoustic impedance
differences between the transducer and the
patient
impedances that are intermediate to those of soft
tissue and the transducer material
• The thickness of each layer is equal to one-fourth the wavelength, determined from the center operating frequency of the transducer and speed of sound in the matching layer
Trang 34 For example, the wavelength of sound in
a matching layer with a speed of sound
of 2,000 m/sec for a 5-MHz ultrasound beam is 0.4 mm
• The optimal matching layer thickness is equal
to ¼ λ = ¼ x 0.4 mm = 0 1 mm
coupling gel (with acoustic impedance similar to soft tissue) is used between the transducer and the skin of the patient to eliminate air pockets that
could attenuate and reflect the ultrasound beam.
Trang 35piezoelectric element is intricately machined into a
large number of small “rods,” and then filled with an epoxy resin to create a smooth surface
Trang 38 The acoustic properties are closer to
issue than a pure PZT material, and thus
provide a greater transmission efficiency
of the ultrasound beam without resorting
to multiple matching layers
• Multifrequency transducers have bandwidths that exceed 80% of the center frequency.
Trang 39 Excitation of the multifrequency
transducer is accomplished with a short square wave burst of 150 V with one to three cycles, unlike the voltage spike
used for resonance transducers
• This allows the center frequency to be
selected within the limits of the transducer bandwidth
Trang 40 Likewise, the broad bandwidth response permits the reception of echoes within a wide range of frequencies
• For instance, ultrasound pulses can be
produced at a low frequency, and the echoes received at higher frequency
Trang 41 “Harmonic imaging” is a recently
introduced technique that uses this
ability;
• lower frequency ultrasound is transmitted into the patient, and the higher frequency
harmonics (e.g., two times the transmitted
center frequency) created from the interaction with contrast agents and tissues, are received
as echoes
Trang 42 Native tissue harmonic imaging has certain advantages including greater depth of penetration, noise and clutter removal, and improved lateral spatial resolution
Trang 43Transducer Arrays
The majority of ultrasound systems
employ transducers with many individual rectangular piezoelectric elements
arranged in linear or curvilinear arrays
• Typically, 128 to 512 individual rectangular
elements compose the transducer assembly
the wavelength and a length of several millimeters
Trang 45Linear Arrays
Linear array transducers typically contain
256 to 512 elements; physically these
are the largest transducer assemblies
Trang 46 In operation, the simultaneous firing of’ a small group of ≈ 20 adjacent elements
produces the ultrasound beam
• The simultaneous activation produces a
synthetic aperture (effetive transducer width) defined by the number of active elements
Trang 47 Echoes are detected in the receive mode
by acquiring signals from most of the
transducer elements
• Subsequent “A-line” acquisition occurs by
firing another group of transducer elements
displaced by one or two elements
Trang 48 A rectangular field of view is produced with this transducer arrangement
• For a curvilinear array, a trapezoidal field of
view is produced.
Trang 50 By using time delays in the electrical activarion
of the discrete elements across the face of the transducer, the ultrasound beam can be
steered and focused electronically without
moving the transducer
transducer elements detect the returning echoes from the beam path, and sophisticated algorithms
synthesize the image from the detected data.
Trang 51BEAM PROPERTIES
The ultrasound beam propagates as a longitudinal wave from the transducer surface into the propagation medium, and exhibits two distinct beam patterns:
• a slightly converging beam out to a distance specified by the geometry and frequency of the transducer (the near field), and
• a diverging beam beyond that point (the far field)
Trang 52 For an unfocused, single-element
transducer, the
length of the near field is determined
by the transducer diameter and the frequency of the transmitted sound
Trang 53 For multiple transducer element arrays,
an “effective” transducer diameter is
determined by the excitation of a group of’ transducer elements
• Because of the interactions of each of the
individual beams and the ability to focus
and steer the overall beam, the formulas for a single-element, unfocused transducer are not directly applicable.
Trang 54The Near Field
The near field, also known as the
Fresnel zone, is adjacent to the
transducer face and has a converging beam profile
• Beam convergence in the near field occurs because of multiple constructive and
destructive interference patterns of the
ultrasound waves from the transducer
surface
Trang 55 Huygen’s principle describes a large
transducer surface as an infinite
number of point sources of sound
energy where each point is
characterized as a radial emitter
• By analogy, a pebble dropped in a quiet pond creates a radial wave pattern
Trang 56 As individual wave
patterns interact, the
peaks and troughs from adjacent sources
constructively and
destructively interfere, causing the beam profile
to be tightly collimated in the near field
Trang 57 The ultrasound beam path is thus largely confined to the dimensions of the active portion of the transducer surface, with
the beam diameter converging to
approximately half the transducer
diameter at the end of the near field
Trang 58 The near field length is dependent on the transducer frequency and diameter:
• where d is the transducer diameter, r is the
transducer radius, and λ is the wavelength of ultrasound in the propagation medium
λ λ
2 2
4
r
d length
field
Trang 59 In soft tissue, λ = 1.54mm/f(MHz), and
the near field length can be expressed
as a function of frequency:
( ) ( )
( )mm
MHz mm
d length
field Near
2 2
54 1
4 ×
=
Trang 60 A higher transducer frequency (shorter wavelength) will
result in a longer
near field, as will a larger diameter
element
Trang 61 For a 10-mm-diameter transducer, the
near field extends 5.7 cm at 3.5 MHz and 16.2 cm at 10 MHz in soft tissue
• For a 15-mm-diameter transducer, the
corresponding near field lengths are 12.8 and 36.4 cm, respectively
Trang 62 Lateral resolution (the ability of the
system to resolve objects in a direction perpendicular to the beam direction) is dependent on the beam diameter and is best at the end of the near field for a
single-element transducer
• Lateral resolution is worst in areas close to and far from the transducer surface.
Trang 63 Pressure amplitude characteristics in the near field are very complex, caused by the constructive and destructive
interference wave patterns of the
ultrasound beam
• Peak ultrasound pressure occurs at the end of the near field, corresponding to the minimum beam diameter for a single-element
transducer
Trang 64 Pressures vary rapidly from peak
compression to peak rarefaction several times during transit through the near
field
• Only when the far field is reached do the
ultrasound pressure variations decrease
continuously.
Trang 65 The far field is also known as the
Fraunhofer zone, and is where the beam diverges
• For a large-area single-element transducer, the angle of ultrasound beam divergence, 0, for the far field is given by
d
λ
θ 1 22 sin =
Trang 66 Less beam divergence occurs with frequency, large-diameter transducers
high-• Unlike the near field, where beam intensity varies from maximum to minimum to
maximum in a converging beam, ultrasound intensity in the far field decreases
monotonically with distance.
Trang 67Focused Transducers
Single-element transducers are focused
by using a curved piezoelectric element
or a curved acoustic lens to reduce the beam profile
• The focal distance, the length from the
transducer to the narrowest beam width, is
shorter than the focal length of a non-focused transducer and is fixed
Trang 68 The focal zone is defined as the region over which the width of the beam is less than two times the width at the focal
distance;
• Thus, the transducer frequency and
dimensions should be chosen to match the depth requirements of the clinical situation.
Trang 69Transducer Array Beam
Formation and Focusing
In a transducer array, the narrow
piezoelectric element width (typically less than one wavelength) produces a
diverging beam at a distance very close
to the transducer face
• Formation and convergence of the ultrasound beam occurs with the operation of several or all of the transducer elements at the same
time
Trang 70 Transducer elements in a linear array that are fired simultaneously produce an effective
transducer width equal to the sum of the
widths of the individual elements
destructive interference to produce a collimated beam that has properties similar to the properties
of a single transducer of the same size