Once the catheter is in the distal pulmonary artery, the original floppy, torque-controlled, guide wire that was used to manipulate across the ductus is replaced with the stiffest possibl
Trang 1over the wire If an adequate angiogram cannot be
obtained by either of these techniques, a second, small
ret-rograde catheter should be introduced A freeze-frame
“road map” of this angiogram of the ductus is stored for
reference use
After the angiogram has been obtained, an end-hole
catheter with a smooth tip is advanced over the wire
very gently, through the ductus and into the distal branch
pulmonary artery Once the catheter is in the distal
pulmonary artery, the original floppy, torque-controlled,
guide wire that was used to manipulate across the ductus
is replaced with the stiffest possible exchange length
guide wire that the delivery balloon/stent that will be
used to deliver the stent to the ductus, will accommodate
The exact length of the stent used in the PDA is
deter-mined from the angiogram of each particular ductus
Ideally the stent must cover, and extend slightly past both
ends of the ductus but, at the same time, should not extend
too far into the lumen of either the pulmonary artery or
the aorta Once the delivery wire is secure through the
ductus and well into a distal pulmonary branch, the
sec-ond angiographic catheter is positioned adjacent to the
aortic end of the ductus The correct stent/balloon is
chosen and prepared on the catheterization table and the
prostaglandin infusion is stopped The patient is observed
closely for at least 30 minutes, and assuming no acute or
severe deterioration of the patient during that time, the
aortogram is repeated Usually the ductus constricts and
changes in configuration rapidly once the infusion of
prostaglandin is stopped Occasionally the rebound
con-striction of the ductus is very severe and very rapid which,
in turn, necessitates either the reinstitution of the
pro-staglandin infusion or the rapid introduction of, and the
implant of, the stent The ductus tissues are friable, so
the balloon/stent introduction must be very gentle and
very precise, and the stent must pass easily through the
ductus Because of these factors, some operators prefer to
prepare the balloon stent and position the stent/balloon
across the ductus before stopping the prostaglandin
infu-sion and then wait for the prostaglandin to wear off with
the deflated balloon/stent already in place across the
duc-tus The second catheter adjacent to the aortic end of the
ductus is essential at this stage of the procedure in order to
allow very rapid and precise final positioning of the stent
within the ductus just before deployment
With any deterioration of the patient, or after 30
min-utes, the angiogram is repeated, the stent/balloon
posi-tion adjusted, and the balloon/stent is inflated to its
advertised pressure or until all indentations in the
bal-loon/stent have disappearedawhichever comes first The
balloon is deflated immediately and rapidly With a
pro-perly implanted stent, the patient immediately will
oxy-genate better and stabilize as soon as the balloon is
deflated The balloon is withdrawn carefully out of the
stent/ductus and with the wire still through the stent andductus, the aortogram is repeated to visualize the ad-equacy of the coverage of the stent in the ductus and the
new flow to the pulmonary arteries If there are any areas
of the ductus, particularly at either end of the ductus,which are not covered by the stent, a second (or more)stent(s) is/are implanted during the same catheterization.The additional stent(s) is/are placed overlapping the first
stent and covering all of the ductal tissues entirely Any
“exposed” ductus tissue is notorious for constricting andclosing completely, even with a stent in the remainder ofthe ductus For the implant of an additional stent immedi-ately after the implant of the first stent, all of the cathetersand wires are already in place and minimal furthermanipulation or time in the catheterization laboratory isnecessary for the implant of an additional stent Once theductus is covered adequately and the patient is stable, thedelivery wire is withdrawn from the stent and the catheterwithdrawn The infants are maintained on 21 mg aspirinper day
Stenting of the ductus in patients with pulmonary atresia with intact ventricular septum
In infants with pulmonary atresia and intact ventricularseptum, along with the opening of the pulmonary valve, asystemic to pulmonary “shunt” can be performed in thecatheterization laboratory by implanting a stent in theductus arteriosus The stenting of the ductus is addressedafter the pulmonary valve has been perforated and dilatedsuccessfully This allows the delivery and implant of thestent into the ductus arteriosus through a venous routeand the use of a smaller catheter system in the artery.When a stent is to be implanted in the ductus during thecatheterization, before any intervention on the ductus and
if not administered earlier when the lines were lished, the patient is administered heparin systemically.After the pulmonary valve has been perforated anddilated, the final balloon dilation catheter for the pul-monary valvotomy is removed over the guide wire.Following the perforation and balloon dilation of anatretic pulmonary valve, a long guide wire usually hasalready been advanced through the ductus creating
estab-a “through-estab-and-through” route from estab-a femorestab-al vein,through the right heart and ductus, down the descendingaorta and out through a femoral artery sheath/catheter
If the through-and-through “rail” wire was not lished during the valve perforation/dilation, the “rail” isestablished at this time with an exchange length wirewhich the proposed balloon catheter for stent deliverywill accommodate The balloon dilation catheter used forthe pulmonary valve is replaced with a long 4- or 5-Frenchend-hole catheter, which is advanced through the venoussheath, over the original guide wire and into the pulmonary
Trang 2estab-artery or ductus (wherever the end of the wire is
posi-tioned) If the tip of the prograde venous catheter is in the
pulmonary artery, it is manipulated through the ductus
into the descending aorta and maneuvered through and
out of the femoral artery sheath with the help of a
torque-controlled wire Occasionally the prograde catheter in the
pulmonary artery cannot cross the ductus easily into the
descending aorta In that circumstance a floppy tipped
wire and then a snare catheter is maneuvered from the
aorta, through the ductus and into the pulmonary artery.
A soft tipped wire is advanced through the venous
catheter prograde into the main pulmonary artery and is
snared with a Microvena™ snare (ev3, Plymouth, MN),
which has been passed through the retrograde catheter
into the main pulmonary artery The snared prograde
wire and the prograde venous catheter are withdrawn
through the ductus and out through the femoral arterial
sheath with the retrograde snare Once both ends of the
catheter are available outside of the body, the original
floppy tipped exchange wire is removed and replaced
with a stiff exchange wire of the maximum wire diameter
that the particular coronary stent/balloon catheter or
other pre-mounted stent/balloon that will be used for
implanting the stent in the ductus, requires This wire is
secured outside of the body at both the arterial and
venous ends
An equally effective alternative is to advance the floppy
tipped wire from the retrograde catheter through the
duc-tus and into the main pulmonary artery and to introduce
the snare catheter into the pulmonary artery from the
venous route The retrograde wire then is withdrawn
through the right heart and out through the venous sheath
to complete the through-and-through wire
Even when previously visualized very adequately, a
repeat aortogram with the injection adjacent to the ductus
is performed either through a Tuohy™ adapter attached
to the hub of the catheter over the through-and-through
wire or through a separate prograde or retrograde
catheter positioned in the descending aorta adjacent to the
ductus This angiogram of the ductus is imperative
because of spasm or distortion of the ductus caused by the
various wire/catheter manipulations through it The
diameter and length of the ductus are remeasured very
accurately As with the implant of a stent in any other
duc-tus, the goal is to “line” the entire lumen of the ducduc-tus,
including covering both ends of the ductus with the stent
Similarly to the patients with pure ductus-dependent
pul-monary circulations, a 4 mm diameter stent is used in
these patients Once the through-and-through wire is
established and the appropriate balloon/stent for the
particular ductus is chosen and prepared for delivery, the
prostaglandin infusion is discontinued Assuming no
acute or sudden deterioration in the infant’s saturation/
hemodynamics, the infant is observed for at least
30 minutes while off the prostaglandin infusion and theaortogram repeated Often the combination of the irrita-tion of the through-and-through wire/catheter, the pre-ceding manipulations during the balloon dilations of the valve, and the discontinued prostaglandin infusiondistorts the ductal anatomy significantly The ductus anatomy is re-examined carefully and the measurementsrepeated
If the ductus acutely goes into spasm or the infant riorates significantly when the prostaglandin infusion isstopped, the prostaglandin infusion is restarted and thestent is delivered to the ductus before re-stopping theprostaglandin as described previously
dete-The pre-mounted stent on a 4 mm balloon dilation
catheter is introduced over the venous end of the
through-and-through wire and advanced over the wire, throughthe right ventricle, pulmonary valve and into the ductusarteriosus Although small stents notoriously are poorlyvisible, or even invisible in large adult patients, in smallinfants they are seen clearly in both the collapsed and theexpanded states Once the balloon/stent combination is inposition in the ductus, and before the stent is expanded inthe ductus, the anatomy of the whole ductus and, in par-ticular, the exact location and diameter of the areas of min-imal ductal diameter are re-imaged angiographically Ifthere is not a second catheter in the aorta, the aortogram isaccomplished by advancing a 4- or, preferably, a 5-French,end and side-hole catheter retrograde over the arterial end of the through-and-through wire The catheter isadvanced just to the aortic end of the ductus and a pres-sure, Tuohy™ “Y” adaptor is attached over the wire at theproximal end of the retrograde catheter An angiogram isperformed over the wire through this retrograde catheter,the tip of which should be adjacent to the ductus and thestent/balloon catheter coming from the other end of thewire With the anatomy precisely identified and recorded,the retrograde catheter is withdrawn back into the des-cending aorta as the stent/balloon catheter is positionedexactly in the ductus
The stent should cover both ends of the ductus
com-pletely including, in particular, the area of the narrowestportion of the ductus, which usually is at the pulmonaryend of a long tortuous ductus When satisfied with thestent length and location, the stent is expanded byinflation of the balloon to its full diameter or to the recom-
mended maximum pressure of the balloonawhichever comes firstafollowed by a rapid deflation A repeat
angiogram is recorded with injection through the aortic
catheter before the deflated balloon is withdrawn from the
stent positioned in the ductus When satisfied that thestent is fully inflated and fixed in the ductus, the deflatedballoon is withdrawn cautiously out of the stent, over thewire and out of the body The retrograde catheter is re-advanced to the aortic end of the stent and a final repeat
Trang 3angiocardiogram is recorded through this catheter before
the wire is withdrawn Similarly to the other stents in the
neonatal ductus, if there is an area of the ductus which
is not covered completely, a second, overlapping stent
should be implanted at that time
When satisfied with the position, the adequacy of
expansion of the stent and the “lining” of the entire ductus
by the stent(s), an end-hole catheter is introduced over the
venous end of the wire and advanced into the pulmonary
artery The wire is carefully withdrawn out of the stent
through the venous catheter
The hemodynamics and anatomy are carefully
re-assessed and a decision is made whether an atrial
sept-ostomy is to be performed Hemodynamically, an atrial
communication in the presence of some elevation of the
right atrial pressure enhances left ventricular filling and
systemic output, albeit with some systemic desaturation
and at the expense of some forward flow through the right
ventricle/pulmonary arteries On the other hand, with at
least a moderate sized right ventricle, the right ventricular
pressure near normal, and even in the presence of
signi-ficant tricuspid valve regurgitation, a restrictive atrial
communication and the associated elevated right atrial
pressure should enhance right ventricular filling and, in
turn, encourage forward blood flow through the right
ventricle, pulmonary artery, and lungs Until more
definitive data are available on which ventricles will grow
with adequate flow, and with what type of stimulus, this
remains an on the spot judgment decision, which must be
made in the catheterization laboratory, during each
indi-vidual case
Stenting of the ductus in the hypoplastic left
heart syndrome
The patent ductus is essential for the systemic output in
the infant with severe left heart obstructive lesions and,
particularly, an associated “hypoplastic left heart”
syn-drome In most cases, the open ductus is maintained with
prostaglandins until the infant undergoes the first stage
“Norwood” surgical palliation In that surgery, the entire
ductal tissue is excised purposefully and widely when
the distal pulmonary artery is anastomosed to the aorta
and no catheter intervention on the patent ductus should
be considered when a “Norwood” surgical palliation is
considered
An alternative approach to the “Norwood” and “single
ventricle” approach to the patient with a hypoplastic
left heart is an orthotopic cardiac transplant However,
when the infant is “listed” and awaiting the transplant,
the ductal patentcy must be maintained until the
trans-plant is performedaoften for weeks or months To
main-tain the patency of the ductus medically requires a
continuous, precisely controlled, intravenous (IV) infusion
of prostaglandin The maintenance of the IV and the precise control of the rate of the prostaglandin in aneonate require a 24/7 neonatal intensive care environ-ment Even in this environment the infant is in a very pre-carious situation An alternative technique is to maintainthe ductal patency with an intravascular stent20,21.Once the decision is made to “list” the patient for atransplant, then the implant of a stent in the ductus should
be considered immediately The longer the patient waits,the lower the pulmonary resistance becomes and thesicker the infant becomes The longer the patient remains
on prostaglandin, the greater the likelihood of systemicinfection and the more friable the ductal tissues become.The infant with a “hypoplastic left heart” who is toundergo the implant of a stent in the ductus is taken to thecatheterization laboratory with the ductus patency main-tained with the prostaglandin infusion The infant is intub-ated and ventilated on 17–18% oxygen If the patient doesnot have an indwelling arterial line, a femoral artery iscannulated with a 20-gauge Quick-Cath™, and a multi-purpose, end and side-hole catheter is introduced through
a sheath in the femoral vein This catheter is manipulatedthrough the right ventricle to the pulmonary artery and tothe ductus arteriosus A biplane angiogram is recorded
in the PA and lateral views with the injection directly intothe ductus This angiogram is to visualize the exact dia-meter, length and configuration of the ductus If necessary,the X-ray tubes are re-angled to “cut the ductus on edge”more precisely and the angiogram repeated Precise meas-urements are made of the length and diameter of the duc-tus from the views that elongate the ductus optimally.Freeze-frame images of the desired views are stored for use as “road maps” during the implant of a stent intothe ductus
The end and side-hole catheter is advanced through theductus into the distal descending aorta A 0.018″ or 0.035″exchange length guide wire (depending upon which stentand which balloon catheter is to be used), is advanced farinto the distal aorta, the tip of the wire is fixed in the ilio-femoral artery, and the catheter is removed over the wire
If there appears to be any distortion of the pulmonaryartery–ductus–descending aorta anatomy by the wire, arepeat angiogram is recorded over the wire and throughthe catheter in the ductus before the catheter is removed.This angiogram performed with the catheter over the wire is accomplished by injecting the contrast through
a Tuohy™ high-pressure “Y” adaptor attached to the hub
of the catheter while the catheter is still positioned overthe wire
A stent is chosen that is as large in diameter as the tus/descending aorta can accommodate and long enough
duc-to extend completely through the ductus The expandedstent must extend from well within the pulmonary artery,through the ductus, to well into the descending aorta
Trang 4Unlike the stents in the ductus for pulmonary atresia
patients, where a very controlled flow using a small stent
(3– 4 mm) is desired, in patients with hypoplastic left
heart, the stent must be large enough to accommodate all
of the cardiac output and not create any resistance to this
flow because of restriction from a limited diameter of the
stent Depending upon the patient’s size, this usually
requires an 8–12 mm diameter stent The pre-mounted
standard “Large” Genesis™ stents (Johnson & Johnson–
Cordis Corp., Miami Lakes, FL) are satisfactory for this
use up to 10 mm in diameter The Genesis™ stents are
available pre-mounted in diameters up to 9 mm and in
lengths of 19, 29 and 39 mm These stents can be
intro-duced through a short, 6-French sheath and advanced
over a 0.035″ guide wire without the necessity of
pre-positioning a long sheath across the lesion “Large”
Genesis™ stents are very flexible and conform to the
cur-vature of the ductus None of the coronary artery stents
are suitable for this use because of the maximum diameter
of 4 to 5 mm The Express™ Biliary LD stents (Medi-Tech,
Boston Scientific, Natick, MA) are available pre-mounted
in similar sizes, but have larger side “cells” as a
conse-quence of their open-cell design, which may be a problem
for any use in the ductus because of the propensity for
rapid ingrowth of the ductal tissues through any space
Once the wire is in place and the stent is ready for
deliv-ery, the prostaglandin infusion is stopped The stent is
advanced over the wire and positioned across the ductus
The balloon/stent is maintained in this position while
monitoring the patient’s distal arterial pressure and
sys-temic saturation for 30 minutes or until the patient’s
hemodynamics begin to deteriorate The idea is to allow
the large diameter, often patulous, ductus to constrict
enough after the prostaglandin is stopped to hold the
stent in place The blood pressure and systemic saturation
decrease as the prostaglandin wears off and the ductus
begins to close Unless a second venous catheter has been
introduced into the pulmonary artery or the stent/balloon
was delivered to the ductus through a long sheath, an
angiogram in the pulmonary artery to verify the status of
the ductus and the stent position is not possible at this time
After the 30 minutes or with any deterioration of the
patient, the stent’s position is compared to the
freeze-frame “road map” images of the ductus, and when in the
appropriate position, the balloon is inflated slowly to
deploy the stent precisely in the ductus Once fully
inflated, the balloon is deflated rapidly The large
dia-meter stent should fix very securely in place in the ductus
The inflation/deflation is repeated one, or more, times
and then the balloon is removed from the body over
the wire
An end- and side-hole catheter is advanced over the
wire and into the stent An angiogram is recorded within
the ductus (stent), injecting over the wire with the use of a
Tuohy™ “Y” connector on the catheter If the stent is notfully expanded, if it is at all unstable, or if there are areas ofductal tissue that are “exposed” or “uncovered” by thestent, a further dilation of the stent or the implant of anadditional stent will be necessary to cover the ductal tis-sue completely If a second stent is necessary and the wire still is in place, the additional stent can be implantedduring this same procedure without any significant further or repeated manipulations through the freshlyimplanted stent
Once satisfied with the location and expansion of the stent(s), the wire is removed from the catheter and
the catheter withdrawn The infant is maintained off the
prostaglandin infusion Depending upon the status of thepulmonary vascular resistance, the infant usually can
be discharged and observed as an outpatient until a donor heart becomes available Even with a very satisfac-tory stent in the ductus, the infant still needs very close follow-up As the pulmonary vascular resistance fallssignificantly, the infant’s lungs can become flooded, with
a proportionate decrease in the effective systemic cardiacoutput through the ductus
The presence of the stent in the ductus does not interferewith a cardiac transplant procedure since the stented duc-tus and the adjacent tissues are removed at the time of the
transplant The main disadvantage to stenting the ductus in
an infant with hypoplastic left heart syndrome is when a
donor heart does not become available of for some other
reason, the decision for the course of treatment is reversedand a “Norwood” procedure is required A large stent inthe ductus arteriosus complicates, or even possibly ren-ders impossible, the usual “Norwood” surgery
Complications of arterial stents
Complications of systemic arterial stents include ation or extension of any of the complications of balloondilation of any vessel including, particularly, coarctation
exagger-of the aorta or dilation exagger-of other arterial lesions At the
same time, since the implant of a stent requires no dilation of the artery, complications related to dilation with
over-the balloons are very rare
Iatrogenic obstructions caused by the implant of stentsthat have a maximum diameter that will be too small for
the eventual size of the adult aorta, should be a totally
avoidable “complication” of stents used in the treatment
of coarctation of the aorta and other arteries There noware stents available that can be dilated to 25+ mm, whichmakes these stents large enough in diameter for any adultproximal descending aorta These potentially larger dia-meter stents should be used in any growing patient witheven the potential of having the aorta in the area of coarc-tation grow to larger than 16–18 mm in diameter A stent
Trang 5with a limited diameter, which creates an obstruction
because it is too small for the aorta, creates an obstruction
that is much more difficult to repair than the usual native
or “residual” post-operative obstruction When the patient
is too small to implant a stent that can be dilated to the
anticipated diameter of the adult aorta, it is preferable to
treat the patient with balloon dilation without a stent or
with surgical repair initially A stent can be implanted
later to “finalize” the correction of any residual lesion
fol-lowing prior dilation or surgery
A very serious complication of the catheter treatment of
coarctation of the aorta using balloon dilation with or
without stent implant is injury to the central nervous
sys-tem (CNS) during the catheterization procedure CNS
injury most likely occurs from air and/or clot
emboliza-tion coming from catheters, sheaths or wires that are
pre-sent in the systemic circulation proximal to the carotid
and vertebral vessels Meticulous attention to clearing
fluid lines, catheters and sheaths of any air and/or clot,
keeping guide wires in the circulation “covered” with a
catheter that is maintained on a continual flush for as long
as possible, the continual flushing of all catheters and
sheaths that are positioned in the heart, and the routine
use of systemic heparin, particularly during “left heart”
procedures, should reduce or eliminate CNS problems
Direct injury from the tip of a wire positioned in a cranial
artery has been implicated in central nervous system
injury Never positioning a wire tip in either a carotid or a
vertebral artery eliminates this particular possibility
Probably the most common complication of the use of
stents in the arteries is injury to the local artery at the site
of catheter/stent introduction with subsequent
compro-mise of arterial blood flow in the involved extremity
These injuries occasionally are unavoidable because of the
necessarily very large sized balloon catheters/sheaths
that are used to deliver stents to the aorta Meticulous care
of the arteries, which was discussed earlier in this chapter
and in Chapter 4, is the best prevention and, in turn, best
treatment of this problem Local anesthesia is used
liber-ally and repeatedly around the artery and surrounding
tissues before, during, and at the end of the procedure
Local anesthesia is administered even if the patient is
receiving general anesthesia A precise, single-wall
punc-ture is utilized for the entrance into the artery Although
an indwelling sheath often is as much as 2–3 French sizes
larger than the dilation balloon or balloon catheter that
is used for the stent delivery, an indwelling sheath, which
is relatively fixed in the artery, is always less traumatic to
the artery than a constantly moving catheter or a bare,
rough, “folded” balloon/stent being introduced into
and withdrawn out of the artery When the sheath is
removed from the artery, the puncture site is compressed
manually and personally while continually monitoring
the puncture site for bleeding and, at the same time, the
artery distal to the puncture for a pulse This hemostasiscan take 30, 60, or more minutes, but should be accom-
plished before the patient leaves the observation and care
of the catheterizing physician!
Stent displacement is relatively common during theimplant of stents in coarctation of the aorta There often is
a large discrepancy in diameter between the aorta imal to the coarctation and the aorta distal to the coarcta-tion A stent that is expanded to a diameter that is calculated to fix the stent in the aorta proximal to thecoarctation, can easily become free-floating in the muchlarger distal aorta At the same time, expanding the stent
prox-to a diameter satisfacprox-tory for the diameter of the aorta tal to the coarctation would split the smaller-diameteraorta that is proximal to the obstruction In order to avoiddisplacement, the stent is implanted with the majority ofthe stent positioned in the aorta proximal to the narrow-
dis-ing and no attempt is made at “approximatdis-ing” the distal
end of the stent to the larger distal aorta during the initialimplant procedure If a stent dislodges and moves furtherdistally in the aorta, the stent is purposefully repositioned
in the distal aorta so that it does not compromise vital sidebranches, and then it is expanded and fixed in the moredistal location
Tears in the arterial wall or flaps off the intima/mediaare less common, or at least, less recognized, during stentimplants in systemic arteries, including congenital coarc-tation of the aorta, than with balloon dilation alone ofthese same lesions When the proper sized stents are usedafter the artery is measured properly and very accurately,the adjacent aorta should not be “over-dilated” at all and,
at the same time, any slight disruption of the intima/medial tissue within the area of the stent is compressedback against the wall of the aorta by the stent As theendothelium and neointima develop over and around the stent, the combination of the wall “thickening” by the
“new endothelial tissues”, the scarring as a consequence
of any tears that did occur, and the “metal scaffolding” ofthe stent within the area, all together create a very solidarterial wall The artery in the precise area of the stent isvery non-compliant, but no more so than a surgical scarinvolving the same area!
Aneurysms have occurred acutely during the implant
of stents in coarctations of the aorta These occurred morecommonly with the use of the larger Palmaz™ stents( Johnson & Johnson, Warren, NJ) and occurred predomin-antly (only?) when the stents were dilated acutely to their final large diameters with a single inflation of a large diameter balloon Acute aneurysms are not reportedwith the sequential dilation of stents to their precise, eventual large diameters, or with the use of stents that donot develop sharp tips at their ends as they expand.Aneurysms following the implant of stents in coarctation
of the aorta are still being studied, and should be looked
Trang 6for in every patient who undergoes a stent implant in
the aorta
Tears or ruptures of the aorta occur with dilation of
native coarctation, re-coarctation of the aorta, and middle
aortic syndrome, but should not occur with the
conserva-tive implant of the correct size stent in coarctations.
Meticulous, accurate measurements of the lesion itself
and the adjacent vessels, and avoiding oversized
dila-tions/stents compared to the size of the lesion itself and
the adjacent vessel, presumably should prevent this
com-plication during stent implant In cases of very severe
stenosis of classic coarctation or middle aortic stenosis,
a staged dilation/implant during several sequential
catheterizations is utilized to avoid splitting very
nar-rowed vessels by a single dilation to a very large diameter
A stent that is still narrowed within a vessel can always be
dilated further at a later date Once the artery/aorta is
split or torn, there is little or no “turning back”, although
several operators have reported on the successful use of a
covered stent as an emergency “bail-out” therapy in the
catheterization laboratory22,23
Stents implanted in coarctation of the aorta have been
reported to fracture or kink24 This probably is a result of
the type of stent used in the area There have been no
adverse events from these findings and a recurrent
nar-rowing as a result of a fracture or kink can be treated with
the implant of an additional stent within the original stent
The implant of stents into the patent ductus of newborn
infants has its own specific complications These
compli-cations are in addition to the inherent complicompli-cations of
extensive catheter manipulations in very sick newborn or
small infants As with all complications, the best
treat-ment is prevention by paying meticulous attention to the
details of the procedure and the use of known, established
techniques until newer/better techniques are proven
Irritation and spasm of the ductus is a potential problem
with any manipulation around or through the ductus
This spasm is not always responsive to prostaglandin
infusion or re-infusion Should intractable ductal spasm
occur, having the equipment ready for immediate
deploy-ment of a stent is the best treatdeploy-ment This, however, is not
a guarantee of successful recovery since the “mass” of
even the “tiny” stent/balloon occasionally cannot be
advanced through the ductus once it begins to spasm
Disruption/tears of the ductal tissues is another potential
with stent implant into the patent ductus, particularly
when the stent delivery is rushed The tissue is inherently
very friable and cannot tolerate rough handling
When disruption of the ductus does occur, it usually is
catastrophic
Stent displacement during implant into the “patulous”
ductal tissue in an infant is a real problem Prevention, by
the use of a slow meticulous positioning and by waiting
for the prostaglandin to wear off before deploying the
stent, is the best treatment When a stent displaces fromthe ductus, an attempt is made to capture the stent on aballoon and reposition it back into the ductus A new bal-loon which is at least 1 mm larger than the maximumdiameter of the stent is more effective for “capturing” an
errant coronary stent When a coronary stent becomes placed from the ductus and is positioned or implanted in any other artery (even very peripherally), the coronary
dis-stent, unequivocally, eventually will produce a verysignificant stenosis in that vessel because of its very smallmaximum diameter If a displaced stent cannot be cap-tured and reimplanted successfully in the ductus orremoved from the patient with a catheter, the stent should
be removed surgically from the errant vessel within a fewdays after the attempted implant procedure, unless theerrant vessel is considered “expendable”
The majority of the complications of stent implants inarterial locations are eliminated by the use of extremelyaccurate measurements, a conservative diameter at theinitial implant, and by paying meticulous attention to the
details of every step of the procedure The morbidity and
complications of dilation with intravascular stent implantfor systemic arteries appear to be comparable to or evenless than surgical therapy of these same lesions Dilationwith stent implants in coarctation and other congenitalsystemic arterial lesions still represents a “new” treat-ment, which requires decades of follow-up to determineits real efficacy and safety
References
1 Dotter CT et al Transluminal expandable nitinol coil
stent grafting: preliminary report Radiology 1983; 147(1):
259–260.
2 Palmaz JC et al Atherosclerotic rabbit aortas: expandable
intraluminal grafting Radiology 1986; 160: 723–726.
3 Shaffer KM et al Intravascular stents in congenital heart
dis-ease: short- and long-term results from a large single-center
experience J Am Coll Cardiol 1998; 31(3): 661–667.
4 Morrow WR et al Balloon angioplasty with stent tion in experimental coarctation of the aorta Circulation 1994;
implanta-89(6): 2677–2683.
5 Grifka RG et al Balloon expandable intravascular stents:
aortic implantation and late further dilation in growing
minipigs Am Heart J 1993; 126(4): 979–984.
6 Suarez de Lezo J et al Balloon-expandable stent repair of
severe coarctation of the aorta Am Heart J 1995; 129(5):
1002–1008.
7 Rosenthal E, Qureshi SA, and Tynan M Stent implantation
for aortic recoarctation Am Heart J 1995; 129(6): 1220–1221.
8 Bulbul ZR et al Implantation of balloon-expandable stents
for coarctation of the aorta: implantation data and short-term
results Cathet Cardiovasc Diagn 1996; 39(1): 36–42.
9 Cheatham JP Stenting of coarctation of the aorta Catheter
Cardiovasc Interv 2001; 54(1): 112–125.
Trang 710 Mendelsohn AM et al Stent redilation in canine models of
congenital heart disease: pulmonary artery stenosis and
coarctation of the aorta Cathet Cardiovasc Diagn 1996; 38(4):
430– 440.
11 Cheatham J et al Early experience using endovascular stents
in children with coarctation of the aorta: promising results
but proceed with caution (abstr) Cardiol Young 1998;
9(Suppl 1:11): (abstr).
12 Suarez de Lezo J et al Immediate and follow-up findings
after stent treatment for severe coarctation of the aorta Am J
Cardiol 1999; 83(3): 400–406.
13 Sapin SO, Rosengart RM, and Salem MM Chest pain
dur-ing stentdur-ing of a native aortic coarctation: a case for acute
intercostal muscle ischemia and rhabdomyolysis Catheter
Cardiovasc Interv 2002; 57(2): 217–220.
14 Thanopoulos BV et al Long segment coarctation of the
thor-acic aorta: treatment with multiple balloon-expandable stent
implantation Am Heart J 1997; 133(4): 470–473.
15 Tyagi S et al Percutaneous transluminal angioplasty for
stenosis of the aorta due to aortic arteritis in children Pediatr
Cardiol 1999; 20(6): 404–410.
16 Redington AN and Somerville J Stenting of aortopulmonary
collaterals in complex pulmonary atresia Circulation 1996;
94(10): 2479–2484.
17 El-Said HG et al Stenting of stenosed aortopulmonary
collaterals and shunts for palliation of pulmonary atresia/
ventricular septal defect Catheter Cardiovasc Interv 2000;
49(4): 430–436.
18 Gibbs JL et al Stenting of the arterial duct: a new approach to
palliation for pulmonary atresia Br Heart J 1992; 67(3):
240–245.
19 Michel-Behnke I et al Stent implantation in the ductus
arte-riosus for pulmonary blood supply in congenital heart
dis-ease Catheter Cardiovasc Interv 2004; 61(2): 242–252.
20 Ruiz CE et al Brief report: stenting of the ductus arteriosus as
a bridge to cardiac transplantation in infants with the
hypoplastic left-heart syndrome N Engl J Med 1993; 328(22):
1605–1608.
21 Slack MC et al Stenting of the ductus arteriosus in
hypoplas-tic left heart syndrome as an ambulatory bridge to cardiac
transplantation Am J Cardiol 1994; 74(6): 636–637.
22 Khan MS and Moore JW Treatment of abdominal aortic pseudoaneurysm with covered stents in a pediatric patient.
Catheter Cardiovasc Interv 2000; 50(4): 445–448.
23 Tyagi S, Rangesetty UC, and Kaul UA Endovascular ment of aortic rupture during angioplasty for aortic in-stent
treat-restenosis in aortoarteritis Catheter Cardiovasc Interv 2003;
58(1): 103–106.
24 Ledesma M et al Stent fracture after stent therapy for aortic
coarctation J Invasive Cardiol 2003; 15(12): 719–721.
Trang 8Occlusion of abnormal or persistent arterial or
arterioven-ous structures or vessels feeding vascular leaks or tumors
by catheter embolization techniques has been utilized for
over thirty years1 The embolization techniques were
developed and perfected primarily by the vascular
radi-ologists working in the abdominal viscera, gastrointestinal
areas and the central nervous system, particularly in “end
artery” vessels Many materials and devices, including the
patient’s own clotted blood, Gelfoam™, colloidal plugs,
“glues”, detachable balloons and coil occlusion devices
have been used for these peripheral occlusions1– 6
In the pediatric and congenital heart population there
are numerous abnormal congenital and acquired vascular
communications and intravascular “leaks” which require
or, at least, can be benefited by transcatheter occlusion
The occlusion of these vascular lesions in pediatric and
congenital heart patients has been performed in the
cath-eterization laboratory for over two decades The
abnor-mal flow through these communications usually results
in significant abnormalities of the underlying
hemody-namics and compromises the patient’s symptomatic and
hemodynamic status The abnormal vascular
communica-tions which are encountered in pediatric and congenital
heart patients include traumatic fistulae, systemic to
pul-monary artery collaterals, systemic arteriovenous fistulae,
pulmonary arteriovenous fistulae, coronary arterial-cameral
fistulae, perivalvular leaks and a variety of residual,
surgi-cally created systemic to pulmonary artery
communica-tions including Blalock–Taussig, modified Blalock–Taussig,
Potts, and Waterston/Cooley shunts
There are numerous different catheter-delivered devices
and techniques available for the occlusion of these
abnor-mal vascular structures There is no single device
applic-able for every lesion and multiple devices may be suitapplic-able,
and used, for any one lesion These devices/materials are
used either by themselves or (frequently) in combination
with one or more of the other devices The specific sion device used depends upon the type, size and location
occlu-of the communication/leak as well as the availability occlu-of
a particular device either locally or as approved, in theparticular country Some of these devices are designedspecifically for a particular lesion and are discussed indetail in other chapters in the description of the occlusion
of the specific intracardiac defect These same descriptionsare not repeated in this chapter
Since most of the devices can be utilized for the sion of multiple different structures and many of theabnormal vascular communications can be occluded withseveral different devices, each of these miscellaneous vas-cular lesions and the separate vascular occlusion devicesthat are used for that lesion are included in the discussion
occlu-of the particular lesions in this chapter The multipledevices themselves that are available for these occlusionsare discussed initially in this chapter, before the details oftheir use in the various lesions for which they can be used
Devices/equipment for vascular occlusions
Occlusion coils
Stainless steel occlusion coils are the most widely used
of the catheter-delivered occlusion devices and have hadthe longest continued use in pediatric and congenitalheart lesions They are particularly useful for small or tor-tuous vessels and have gained enormous popularity anduse for the catheter occlusion of the patent ductus arterio-sus (PDA) The specific coils used for PDA occlusion and the modifications of the delivery system/techniquesspecifically for the PDA are discussed separately and indetail in Chapter 27 (“PDA Occlusion”) Many of thesemodifications, which were developed specifically forPDA occlusions, are useful for the occlusion of generalvascular structures
including perivalvular leaks
Trang 9Occlusion coils are available as specific lengths of
vari-ous sizes (gauges) of stainless steel, spring guide wire The
“guide” wires are pre-formed during manufacturing so
that they will coil into a cylindrical tube of a specific
dia-meter of the wire in their “resting” state The number of
loops of coil and the length of the “tube” of coils depends
upon the original length of the straightened wire Most
of the “occlusion coils” have multiple tiny threads or
fila-ments of nylon fabric intertwined within the windings
of the spring wires to promote better thrombosis within
a vessel
Occlusion coils are best suited for the occlusion of long
and/or tortuous vessels with irregular internal diameters,
and especially those which have a significant narrowing
somewhere along the course of the vessel Since a
success-ful occlusion is expected to cut off all blood flow through
the vessel, the vessels being occluded should not be the
sole blood supply to a particular area of tissue unless
necrosis of the tissues that are supplied by the vessel is
desired
The coil occluder with the most extensive use in pediatric
and congenital heart patients is the standard Gianturco™
coil The Gianturco™ coil is a length of special stainless
steel spring guide wire in which the stiffening “core” wire
is pre-formed to curl into a “coil” or “wire cylinder” of a
specific diameter in its “free” or “resting” state These
coils are available in multiple sizes of the spring wire,
multiple diameters of the coil (“cylinder”) and multiple
lengths of the coil wire The Gianturco™ coil has multiple
fine nylon fiber segments embedded within the windings
of the spring wire in order to increase the thrombogenicity
of the implanted coil Gianturco™ coils are available in
spring wire diameters of 0.025″, 0.035″, 0.038″ and now, an
additional coil of 0.052″ wire diameter, in lengths between
1.2 and 15 cm and in coil diameters from as small as 2 mm to
as large as 20 mm in diameter The very smallest diameter,
short coils are available only in the 0.025″ diameter wires
while very large diameter coils are available only in the
more recently available 0.052″ wires The total length of
the straightened segment of coil wire in conjunction with
the particular diameter of the preformed loops of the coils,
determine the number of loops which are formed by any
particular length of coil Each Gianturco™ coil comes
from the manufacturer in a straight metal introducer tube
The internal lumen and the length of the introducer tube
are specific for the diameter of each wire and the
straight-ened length of the wire
The mass of the wire of the coil itself creates a
mechan-ical occlusion and the embedded nylon fibers add to the
thromboses in the area where the coil is deposited and, in
turn, occlude the vessel or communication The coils are
best suited for deposit into tubular vessels that have some length and vessels that have an area of narrowingsomewhere within the channel of the vessel or the abnor-mal communication A stenosis distally in the channel ofthe vessel prevents even coils that are undersized frommigrating out of the target vessel and embolizing to anarea or vital organ distally beyond the vessel
The coils are delivered through polyethylene, end-hole
only, catheters, which have an internal diameter which
is just slightly greater than the diameter of the wire of the spring coil Other end-hole only catheters manufactured
from materials that impart a smooth or slick inner surfaceand that are slightly larger in their internal diameters than the wire of the coil can be used for coil delivery The
catheter for coil delivery must not have side holes Side
holes allow the potentially curved tip of the coil to catch
in, or pass into, a side hole of the catheter as the tip of thecoil crosses the side hole The tip of a coil catching in a sidehole of the catheter will prevent the coil from being deliv-ered through the tip of the catheter Since the standardGianturco™ coils have no attachment to the delivery wire,the coil catching in a side hole also prevents any retrieval
of the coil without totally removing the delivery catheter.Both the material of the catheter and the internal diameter
of the catheter are critically important in order to preventthe coil from “binding” within the lumen of the catheterduring the delivery through the catheter
A catheter that is smaller in internal diameter than thecoil wire obviously does not allow the coil with its imbed-ded fibers to be introduced into, or advanced through, the catheter A catheter with an internal diameter signific-antly larger than the diameter of the wire of the coil allowsthe coil to bend and partially “coil” within the catheter orallows the pusher wire to push past the coil instead ofactually “pushing” the coil through the catheter Eitheroccurrence will cause the coil to bind within the catheter.The delivery catheters are available with many pre-formed tip configurations in order to facilitate entry intospecific areas Straight delivery catheters, which the operator can pre-shape to suit his particular needs, arealso used to deliver coils End-hole, only, floating ballooncatheters can be used to deliver the coils to certain locations or in particular circumstances The inflated balloon helps to fix the tip of the catheter in place and/or
prevents portions of the coil from extending back into
a more proximal main vascular channel The catheterlumen of the floating balloon catheter obviously must be
of a slightly larger internal diameter than the diameter of
the coil wire that is being delivered through the balloon
catheter
The coil is introduced into the proximal hub of thedelivery catheter through the straight metal “loader” as astraight length of wire The straightened coil is pushed out
of the loader, into the delivery catheter and through the
Trang 10delivery catheter with a teflon-coated, spring guide wire
of the same or similar wire size as the coil wire The coil is
delivered by pushing it completely through and out of the
distal end of the delivery catheter As the coil is extruded
out of the delivery catheter, it immediately begins to form
the small loop of its predetermined “coil” diameter as it
opens into its coiled configuration Once the extrusion
from the catheter starts with the standard Gianturco™ or
the 0.052″ coils, there is no way of withdrawing the coil
back into the catheter or stopping or reversing the
deliv-ery Even if the coil is noted to be in an unsatisfactory
posi-tion as it starts to extrude from the catheter, it can only be
extruded completely and then retrieved with a separate
retrieval catheter and system
When choosing the appropriate occlusion coil, the
diameter, the length and the general configuration of
the vessel to be occluded are imaged angiographically
The length and diameters of the vessel are measured
very accurately on the angiograms The Gianturco™ coil
occludes the vessel by the creation of an irregular mass of
the coil wire and the nylon fiber strands that are
incorpor-ated within the wire in which a thrombus forms The coil
used should be 1–2 mm larger in diameter than the vessel
that is to be occluded The slightly larger diameter results
in the coil unraveling in an irregular configuration within
and across the vessel lumen rather than into a neat
“donut” like cylinder or smoothly coiled configuration
If the diameter of the coil is far larger than the diameter of
the vessel, the coil does not “coil” at all, but rather tends to
align straight within the vessel lumen and, in turn, does
not form an effective occlusive mass in the vessel When
the coil is extruded from the catheter, it not only must
have the appropriate diameter and length to fix to the
walls of the vessel, but also must not be excessively long
When the coil is too long, it can extend proximally out of
the target vessel and into the more central feeder vessel,
which potentially can be back into the normal circulation
and interfere with vital structures If, at the other extreme,
the diameter of the formed coil is too small for the vessel,
the coil rolls up into a tight “donut”, does not occlude the
entire vessel, and is likely to tumble distally or even out of
the desired vessel Once one coil is secured within a
ves-sel, additional coils of different sizes and/or diameters
can, and frequently are, intertwined within or deposited
proximal to the original coil to complete the occlusion
Even when used in tandem, but without a distal
narrow-ing or some other type of device for fixation, the standard
Gianturco™ coil generally is only usable in tubular
struc-tures of no more than 7–8 mm in their distended diameter
For larger vessels and vessels without an area of discrete
stenosis, either the standard Gianturco™ coils are used in
conjunction with other intravascular occlusion devices or
the 0.052″ coils are used initially to begin the occlusion
of the vessel
Occlusion coils can be deposited into long vessels thathave a discrete distal narrowing, where the coil thenlodges in place at the narrowing, although it is not “fixed”against the wall In that circumstance, coils can be
“floated” into the vessel, one after another to create a mass
of coils, proximal to the stenosis in the vessel When thistechnique is used, the final coil in the vessel should be of a
slightly larger diameter then the diameter of the vessel in
order to wedge the last coil against the walls of the vessel.The one “fixed” coil assures that the “loose” coils packedwithin the vessel do not “float” back out of the vessel intothe vital circulation
Currently, by far the most common use of theGianturco™ coil in congenital heart lesions is for the clo-sure of the patent ductus arteriosus This is an entirelyseparate subject and is discussed in detail in Chapter 27and is not covered in this chapter at all There are manyabnormal vessels, collaterals and persistent surgically cre-ated systemic to pulmonary artery shunts, which fre-quently are associated with more complex lesions Thesevessels require occlusion when the additional systemicflow competes with normal pulmonary flow, particularlywhen the abnormal communication persists after themajor intracardiac defect has been corrected These communications traditionally required surgical divisionduring the corrective procedure or as a separate, later, surgical procedure When the occlusion of these defects
is performed surgically during the intracardiac repair, itsignificantly prolongs or complicates the surgery Most ofthese abnormal communications now are occluded withGianturco™ coils either before or shortly after the majorsurgery7 With the use of coils, further extensive extrasurgery or repeat surgery is unnecessary for the elimina-tion of persistent systemic to pulmonary artery collaterals
or for any surgically created systemic to pulmonary arteryshunts that are present at the time of, or following, the
“total” correction
Other lesions in which the coils are useful are venous fistulae, including systemic coronary-cameral,peripheral arteriovenous fistulae as well as pulmonaryarteriovenous fistulae These lesions can produce eitherleft to right or right to left shunts In these lesions it is critical to identify a stenotic or “end” vessel into which thedevice can be fixed very securely in order to reduce thedangers of embolization to an essential more distal vessel
arterio-or vital structure in the systemic circulation
The 0.052 inch stainless steel coil
In order to provide a more robust coil, a more occlusivecoil and a coil particularly for use in the patent ductusarteriosus, the 0.052″ stainless steel coil was developed.The 0.052″ coil is a larger, stiffer version of the standardGianturco™ coil with the wire of the coil being the heavier
Trang 11gauge 0.052″ diameter The 0.052″ coil provides a much
sturdier occluding device for the miscellaneous vessels
and is particularly useful for the occlusion of much larger
vessels with higher pressures and higher flows8 The use
of the 0.052″ coils for the occlusion of the patent ductus
arteriosus for which they were developed, is discussed in
detail in Chapter 27
In an end vessel or in a vessel with definite distal
stenosis, the 0.052″ coils are delivered with a “free release”
technique exactly like the standard, smaller sized,
Gianturco™ coils However, when choosing the diameter
of the 0.052″ coil for a particular constrained vessel, the
diameter of the coil used should be no more than 1–1.5 mm
larger than the stretched diameter of the vessel being
occluded The 0.052″ coils are very rigid and have a much
greater tendency to form into a symmetrical “donut”
shape after they are extruded, and they have very little
tendency to form into irregular or elongated shapes In a
vessel that is significantly small in diameter for the coil
and non-elastic, the 0.052″ coil is likely to elongate into
an almost straight wire rather than to bunch up into an
irregular coil
Because of the larger wire size alone, the 0.052″ coil
requires a delivery “catheter” of a larger internal diameter
for implant than the standard Gianturco™ coils The
necessity of using the larger delivery catheters can
pro-hibit the use of the 0.052″ coils in infants and very small
children This becomes an even greater problem when the
0.052″ coil embolizes away from the implant location and
must be retrieved To overcome the necessity of using
the larger French sized delivery catheters, the 0.052″ coil is
usually delivered through a 4- or 5-French, long,
transsep-tal type sheath with a radio-opaque band at the tip rather
than through a separate delivery catheter This allows the
delivery of the larger diameter coil wire without an
over-all increase in outside diameter of the delivery system The
long sheaths have the disadvantages of having less
flex-ibility, less ability to bend at acute angles and, as a
con-sequence, a greater tendency to kink than most “delivery
catheters” This tendency to kink compromises the access
to vessels or lesions arising at acute angles off the major
vessel (aorta) When the long 4- or 5-French sheath is used
to deliver the 0.052″ coil, it is advisable to introduce the
long sheath into the peripheral entry vessel through a
short, 6- or 7-F sheath Then, if the totally extruded coil
needs to be withdrawn, it can be withdrawn into the
larger, short 6- or 7-French sheath, which is already in the
vessel In the larger patient, the 0.052″ coil, of course, can
be delivered through a 6- or 7-French guiding catheter
The “bioptome controlled” delivery technique is
prefer-able to the “free release” technique for the delivery of the
0.052″ coils to all locations, but particularly when there are
any concerns about the proper “seating” or possible distal
embolization of this coil Bioptome-controlled delivery is
described in detail, later in this chapter and in Chapter 27
on “PDA Occlusion”
Delivery techniques for the Gianturco™
“Free-release” technique for coil delivery
An end-hole only delivery catheter is chosen with a shape
of the tip of the catheter that will facilitate entry into thespecific area to be occluded and of an internal diameter to
match the diameter of the coil wire being used The
deliv-ery catheter/sheath is manipulated and advanced as far
as possible into the vessel to be occluded The tip of thedelivery catheter/sheath is fixed securely in the vessel at,
or distal to the site for, implant of the occlusion device.Extra effort should always be made to ensure that thedelivery catheter/sheath is positioned very deep into thevessel that is to be occluded Often during the process ofadvancing the coil through the catheter or extruding thecoil wire out of the tip of the delivery catheter/sheath, the tip of the delivery catheter/sheath can be pushedbackward in the vessel The catheter/sheath must be farenough into the vessel initially to allow for this
Once the catheter is positioned properly, the thin lar end of the coil introducer is introduced into the prox-imal hub of the pre-positioned delivery catheter or sheath.The coil introducer should be introduced through a wireback-bleed/flush device that previously was attached tothe proximal hub of the coil delivery catheter/sheath Theback-bleed/flush device allows a continual flush of thedelivery catheter or sheath during the introduction of the coil and during any exchange of pusher wires, which, inturn, “lubricates” the lumen of the catheter The stiff end
tubu-of a straight teflon-coated spring guide wire (tubu-of the samesize as the internal diameter of the delivery catheter/sheath and the same size or slightly larger [if possible]
than the wire of the coil itself) is introduced into the
prox-imal end of the straight tubular coil introducer As thestraight spring guide wire is introduced into the proximalend of the coil introducer, the coil is pushed through andout of the distal end of the introducer tube and into thedelivery catheter/sheath by the “pusher” spring guide
wire The coil wire has no attachment or connection to the
pusher wire so that once the introduction of the coil into
the catheter has begun, the coil can be advanced only within
the catheter Standard Gianturco™ or 0.052″ coils cannot
be withdrawn at all! Once the coil is completely within the proximal end of the delivery catheter/sheath, the coil introducer is withdrawn from the catheter over the
“pusher” wire and pulled back to the proximal end of, oroff, the pusher wire The teflon-coated spring pusher wire
is reversed and the soft end of the wire is advanced within
Trang 12the catheter, which, in turn, pushes the coil (still as
a straight segment of wire) through the length of the
delivery catheter/sheath Once the wire and coil are
well within the catheter, the catheter, pusher wire and
enclosed coil wire are observed carefully on fluoroscopy
as the coil is advanced within the catheter or long sheath
The coil is distinguishable from the pusher wire by its
slightly “undulating” configuration and the different
dens-ity of the wire of the coil compared to the pusher wire As
a check of the two wires, a small radiolucent “interspace”
can be created between the proximal end of the coil and
the distal end of the pusher wire by withdrawing the
pusher wire several millimeters while observing under
fluoroscopy
Once the distal end of the coil has reached the tip of the
catheter/long sheath, the coil begins to extrude from the
tip of the catheter/sheath by continuing to advance
the “pusher” wire As the coil is extruded from the tip of
the catheter or sheath, the “coil wire” immediately begins
to curl into its pre-formed coiled configuration and/or
push the tip of the delivery catheter/sheath away from
the site Exactly how the coil positions itself within the
vessel depends upon the size relationship of the diameter
of the pre-formed coil and the internal diameter of the
ves-sel at that particular location
It is important to choose a coil diameter of the standard
Gianturco™ coil that is 15–20% larger than the stretched
or expanded internal diameter of the vessel at the area
where the coil is to be implanted If the coil diameter is
smaller than the vessel diameter, the coil wire “coils” into
a tight, smooth circular coil with the appearance of a small
“donut” and is likely to embolize further along in the
ves-sel When it tumbles and if it does become lodged in the
more distal vessel, the lumen within the “donut” of the
coiled wire can line up with the vessel lumen and prevent
effective occlusion On the other hand, when the diameter
of the coil is much larger than the diameter of the vessel
lumen, then the coil cannot “coil” on itself at all and
stretches out longitudinally in the lumen of the vessel and
again, probably will not occlude the vessel When the coil
diameter is so large that the coil cannot “coil” at all, as the
pusher wire and coil wire are advanced through the
catheter or sheath and as the coil wire is pushed out of
the tip of the delivery catheter or sheath, the coil wire
re-mains straight and actually pushes the tip of the delivery
catheter/sheath back in the vessel The catheter or sheath
tip can be pushed completely back out of the vessel into
which the coil is being delivered This leaves some, or all,
of the coil extending proximally out of the target vessel
The ideal coil/vessel size relationship allows the coil to
partially coil on itself yet partially stretch out into a very
irregular shape or mass When the coil is to be delivered
into a very critical and short segment of vessel, the
“dis-tensibility” of the particular vessel can be tested precisely
at that location by a low-pressure inflation of a smallangioplasty balloon that is slightly larger in diameter thanthe vessel at that area
After the successful implant of a coil, a small injection ofcontrast is performed through the delivery catheter orsheath to verify the degree of occlusion Often, it is neces-sary to deposit more than one coil, and often even mul-tiple coils of different sizes, in any single vessel to completethe occlusion As long as there is room in the vessel, addi-tional coils are deposited at the same location, through thesame delivery catheter or sheath, and during the sameprocedure until the vessel is occluded completely
In most cases of the occlusions of discrete vessels, ard coils are extruded into a confined segment of the ves-sel and are released automatically by the “free-release”technique as they exit the delivery catheter The majority
stand-of vessels that are occluded (except the PDA), have somedistal tortuosity or narrowing on which the extruded coilbecomes trapped The very accurate “teetering” across thenarrowest segment of a vessel that is required for coilocclusion of the PDA, seldom is encountered in the major-ity of other vascular/small vessel occlusions With accur-ate information about the size of the vessel to be occludedand when the proper size and type of coil is used for theocclusion of discrete vessels, there is full expectation thatthe coil will “fold up” into a compacted, irregular shape,reorient and move (usually distally) after its release andthen lodge securely in the vessel When there is a highlikelihood of this type of seating in an appropriate vesselthere is little need for the more complicated and expensivedetachable/retractable coils Although unnecessary forthe majority of coil deliveries for vascular occlusions, con-trol of the release and retrievability of the coil, however,does become essential in some circumstances
Special techniques for the delivery or
to improve the safety of their delivery
When the coil must be delivered in a very specific site in
order not to occlude adjacent or branching vessels that arecritical for the supply of essential viable tissues (e.g coron-ary branches adjacent to a coronary-cameral fistula), then a controlled release/retrievable system must beused When the vessel more proximal to the area beingoccluded is significantly wider than the area of the vesselwhere the coils are being implanted or when the coils aredelivered near to the aortic entrance of the vessel and asthe vessel becomes full of coils, then a retractable or con-trolled release coil is desirable or even essential Similarly,
if the vessel has only minimal length or the coil is at alllarge in diameter relative to the vessel diameter, there is
a high probability that the coil will elongate too much ing delivery and will extend back into the central vessel
Trang 13dur-(aorta!) as it is extruded from the delivery catheter, and a
control of the release of the coil is essential Even when
multiple coils are being “wadded” into a long vessel, easy
and immediate retrievability of the last and most
prox-imal coil always is desirable
When the free-release coil is too small in diameter for
the vessel, it “balls-up” into its “donut” configuration,
“bounces around” in the vessel, which is too large, and
easily can embolize backward and out of the vessel into
the central, critical vessel With standard free-release
Gianturco™ coils, there is absolutely no control over this
tumbling, once the extrusion of the coil has started
“Retrieval” of a free-release Gianturco™ coil after the
extrusion of the coil has started, thereafter, represents a
“foreign body” retrieval of a coil which has embolized to a
distal location from the site of implant!
Special techniques to achieve control over
steel coils
In order to overcome the non-retrievability, “all or
noth-ing” delivery of standard Gianturco™ and 0.052″ coils,
multiple modifications of the delivery system/technique
have been developed With a controlled attach/release coil,
the fit and fixation of the completely extruded coil can be
tested once the coil is extruded completely into the vessel
before the coil is released from the attached delivery
wire/cable If the position/fixation is not satisfactory, the
coil can be withdrawn back into the delivery catheter and
repositioned before it is released, or the coil can be
with-drawn totally out of the body and the procedure restarted
with a more appropriately sized coil When there is a
residual leak through the coils and even when the vessel
proximal to the coils is short or otherwise not ideal, by
using a controlled-release coil system, an additional coil
can be extruded safely into earlier coils in the vessel and
tested for both occlusion and fixation before release If
either fixation or occlusion is not satisfactory, the
addi-tional coil can still be withdrawn from the original coil(s),
although care must be taken because each successive coil
tends to become entangled with any previous coil(s)
The various commercially available and the self-made
techniques for controlled release of the coil, all are
effect-ive to some degree at accomplishing retrievability of the
coil during delivery The minute details of all of these
modifications are described in detail in Chapter 27 on
“PDA Occlusion”, where these coil “control” systems
are more of a necessity The specific uses of the various
“controlled-release” systems for general vascular
occlu-sions are discussed here
Because of the problem of a “no-return” delivery once
the extrusion of standard Gianturco™ and 0.052″ coils
is started and, in the long absence of a commercially
available and viable “detachable coil” in the United States,there have been several very innovative techniques devel-oped in order to overcome this shortcoming and make theGianturco™ and 0.052″ coils safer and more reliable
Latson catheter modification for Gianturco™
coil delivery
The Latson™ modification of the delivery catheter provides
some degree of control and retrievability for the standardGianturco™ coil9 The tip of the delivery catheter is heatedand then pulled into a tapered tip over a short, solid,
smooth wire or “mandril” of exactly the diameter of the bare coil spring wire of the particular coil The mandril is
removed and the pulled tip of the catheter is cut off at thenarrowest area of the “pulled taper” on the catheter Thiscreates an opening in the tapered tip of the catheter, which
is tight around the bare spring wire of the coil and very
tight around the coil wire when the nylon fabric is ded in the wire of the coil This modified tip configura-tion is now available commercially as the Latson™ multipurpose catheter: #248498 (Cook Inc., Bloomington,IN) and as the modified Vertebral catheter: #WN27750(Mallinckrodt Inc., St Louis, MO)
embed-As a result of the narrowed orifice of the tip of thecatheter, the coil now is gripped very tightly as it passes
through the narrowed orifice of the tightened, modified
tip of the catheter As a consequence of the tight grip
on the wire, it is necessary to apply considerable force
to the delivery/pusher wire to extrude the standardGianturco™ coil through the tip of the Latson™ catheter.Because of this tight grip on the coil created by themodified tip of the catheter, and unless the coil hasbecome entrapped on something in the vessel (a previouscoil!), the now “dangling” coil, which has been extruded
almost entirely but is now gripped tightly by the tip of the
modified catheter, can be withdrawn from a vessel andout of the body along with the catheter when the deliverycatheter is withdrawn through a peripheral introductorysheath
During the withdrawal of the Latson™ catheter, the coilcannot be withdrawn back into the Latson™ catheter butrather the coil, which is now mostly extruded, will be
“dangling” at the tip of the catheter, exposed in the
circu-lation and not “protected” as it is being withdrawn within
the circulation The coil cannot be repositioned or reusedwith this catheter while it still is within the circulation Tostart over, the coil and delivery catheter are withdrawncompletely out of the body through the peripheral intro-ductory sheath, and a new coil is introduced through thesame delivery catheter
The technique for the delivery of a Gianturco™ coil to avessel using the Latson™ modification is almost identical
to the delivery of the “free-release” coils Usually, but not
Trang 14necessarily, the retrograde approach is used with the
Latson™ catheter Once in position the coil is pushed out
of the catheter with the pusher wire, but now with the
nec-essary considerable additional force on the pusher wire
Because of the force required, more attention must be paid
to the extrusion process in order to extrude the coil in
small, controlled increments in order to deliver the coil to
the precise location The grip on the coil by the tip of
the catheter is dependent completely on the very accurate
“tolerances” of the internal diameter of the tip of the
catheter whether the catheter is built by hand or by
com-mercial manufacturers The Latson™ catheter tip is not as
strong or dependable as any of the other attach/release
mechanisms and does not allow the reuse of the coil
during the same procedure without first withdrawing the
coil out of the body completely Although the Latson™
catheter does provide some extra safety in the delivery of
a standard Gianturco™ coil, the necessary tolerances and
the inability to withdraw the coil into the catheter are
dis-advantages to this catheter modification for coil delivery
Balloon-assisted coil delivery to branch vessels
Floating end-hole balloon catheters occasionally are used
to deliver any 0.035″ or smaller Gianturco™ coils to
abnor-mal branch vessels Balloon catheters are used with
several different techniques for the delivery of the coils
The end-hole, floating balloon catheter may be the only
catheter that can be manipulated into the vessel by the
particular operator In that circumstance, usually, once in
the vessel, the balloon is deflated within the vessel and the
end-hole balloon catheter is used like any other end-hole
catheter for the delivery of the coil Because of the “ragged”
surface of the deflated balloon over the tip of the catheter,
extra attention is necessary during the withdrawal of the
catheter after the coil has been delivered to prevent the
balloon from snagging on the fabric strands that are
dang-ling from the freshly implanted Gianturco™ coil
The end-hole floating balloon catheter is also used with
the balloon inflated securely in the vessel in order to wedge
coils forcibly into distal locations or to keep the coils from
extruding back out of the target vessel For these
pur-poses, the balloon catheter is manipulated into the vessel
to be occluded and then the balloon is inflated tightly
against the walls within the particular vessel With the
balloon catheter fixed tightly within the vessel, the
“free-release” Gianturco™ coil is extruded into the vessel distal
to the balloon When the inflated balloon is fixed tightly
enough in the vessel, the coils can be pushed into the
vessel with some force to “pack” them more tightly in the
vessel This usually represents a delicate balance between
the fixation of the balloon in the vessel and the force used
against the pusher wire If too much force is applied, the
inflated balloon can be pushed out of the vessel along
with the coils that are being packed into the vessel If the entrance of the vessel being occluded is close to theentrance of a vital branch vessel, some of the coils canextend into the adjacent vessel or the entire coil canembolize distally into the more central circulation!
Bioptome-assisted coil delivery
The bioptome-assisted delivery of coils for PDA sion, and in particular the 0.052″ coils, is discussed indetail in Chapter 27 (“PDA Occlusion”)8 The bioptome-assisted technique is effective for the precise control of the delivery of any Gianturco™ coil to locations otherthan the PDA where the exact localization of the coil isextremely critical Bioptome-assisted delivery is equally
occlu-as effective when using the more frequently used 0.038″coils as it is with the 0.052″ coils Bioptome-controlleddelivery is offered an alternative to the commercially
manufactured, detachable coils, which in their more robust form are only available outside of the USA The bioptome
technique allows complete retrievability of the coil at anytime until the purposeful release of the grasp with thebioptome The bioptome technique is used as an adjuncttechnique to the “free-release” technique in the placement
of the coils into very specific locations or the “final” coils
in more critical or precarious locations
The bioptome attachment to the coil requires the lumen
or internal diameter (ID) of the delivery system to be at
least 1.3 mm (4-French ID) A 4-French long sheath with
a back-bleed valve and distal radio-opaque marker is
used as the minimum sized delivery system for
bioptome-controlled delivery However, a 5-French sheath allows
a “more comfortable” delivery and more reliable drawal of the coil/bioptome back into the delivery systemwhen necessary particularly with the 0.052″ coils As aconsequence, a 5-French sheath with distal radio-opaquemarker is usually used with bioptome-controlled coilsexcept in the very smallest infants
with-An end-hole catheter is introduced through a 6- or 7-French short sheath and advanced well into the vessel
to be occluded The end-hole catheter is replaced with
a stiff, exchange length wire The 4- or 5-French longsheath/dilator that is to be used to deliver the coil/biop-tome is introduced over the wire and through the short
introductory sheath and advanced until the tip of the sheath is significantly past the “target” area within the ves-
sel to be occluded Because of the use of a sheath for thedelivery and the stiffness of the bioptome itself, bioptome-controlled delivery does have some limitations as towhich vessels it can be used in Often, if there is significanttortuosity of the vessel proximal to the implant site, thebioptome-controlled technique cannot be used
The bioptome technique requires a special tion of the occlusion coil, which is used in order for the
Trang 15prepara-coil to be grasped with the bioptome All Gianturco™
and 0.052″ coils consist of a very fine stainless steel wire
tightly wound to form the larger “spring guide” type
wire of the advertised coil “wire” diameter One end of
the “spring” wire is sealed and polished with a small
“weld”, while the opposite end is open with a very
tiny hollow tube created by the fine wires The coils,
including the 0.052″ coils, come loaded in a metal “loading
tube” with the welded end of the coil at the distal end of
the loader
The stiff end of a separate 0.038″ spring guide wire
is introduced into the proximal end of the coil holder/
loader and advanced until just 1–2 mm of the distal, closed
(welded) end of the coil is exposed exiting the distal end of
the metal loading tube This exposes the tiny welded tip,
which seals the end of the fine coiled “wire tube” of which
the “spring coil” is made While holding the short more
proximal and exposed portion of the coil tightly between
the fingers, the small welded tip (only) is grasped tightly
with a forceps and pulled 0.25 to 0.5 mm away from the
remainder of the spring coil which is held by the fingers
of the other hand This maneuver separates the small
welded ball approximately 0.5 mm away from the
wind-ings of the coiled portion of the “spring wire”, but still
attached by a single, stretched-out, strand of the fine, very
stiff wire This separation of the welded “ball” allows the
bioptome to grasp the “ball” firmly while at the same time
allowing the bioptome jaws to close almost completely
over the stretched out strand of fine wire This allows the
outside diameter of the combination of the closed,
3-French bioptome jaws, over and holding the welded
“ball” of even the 0.052″ coil, still to be less than 4-French
in outside diameter
The closed bioptome jaws over the tip of the coil,
however, do not fit back within the original metal loader
The combination must be withdrawn into a separate
slightly larger “loader” in order to be front-loaded into the
delivery sheath Before the coil is grasped with the
biop-tome, the bioptome is passed through the separate slightly
larger metal loading tube (which comes with the 0.052″
coils) or through a segment of a 4-French short sheath
which can be used as a loader equally as well as the new,
larger metal loading tube which is provided only with the
0.052″ coils The 4-French sheath must be long enough
to contain the entire straightened length of whichever coil
is being used After the bioptome is passed through the
separate metal loading tube or the 4-French sheath and
during the attachment to the “prepared coil”, the metal
loading tube or length of sheath is withdrawn on the
biop-tome shaft back to its proximal control handle Once the
new “loading tube” is back on the shaft of the bioptome
catheter, the bioptome jaws are opened and then closed
tightly around the “prepared” separated, welded ball at
the end of the coil
With one operator/assistant holding the bioptome jaws closed tightly over the “ball” at the tip of the coil, the “new”, larger diameter “loading tube” through which the bioptome was previously passed, is advancedover the bioptome jaws which now have the tip of the coil grasped within them The entire coil is drawn out
of the original metal loading tube directly into the newlarger “loading tube” or segment of short sheath The coil is withdrawn completely into this new tube until
the now distal (originally proximal) end of the coil is just within the tip of the new loading tube The bioptome-
controlled coil is now ready to be delivered from the newloading tube into the pre-positioned delivery sheath Thedistal end of the loading tube is introduced into the back-bleed valve at the hub of the pre-positioned deliverysheath and the bioptome catheter (with the attached coil)
is advanced into the loader/sheath and the shaft of thedelivery sheath
The bioptome jaws are held closed very tightly whilethe coil is advanced to the end of the delivery sheath Theentire delivery is observed closely on fluoroscopy Thecoil and the bioptome jaws can be visualized very clearly
on fluoroscopy The sheath is withdrawn slightly until thetip of the sheath is exactly in position for implanting the coil The bioptome catheter and coil are advancedtogether, extruding the coil out of the tip of the sheath Thecombination bioptome/coil is advanced together with the sheath held in position until the bioptome jaws with
the grasped coil are just 1–2 mm within the tip of the sheath
At this point the security of the extruded coil is tested byvery slight, gentle, to-and-fro movement of the sheath and bioptome catheter together The degree of occlusioncan be tested by a contrast injection into the involved ves-sel through a second catheter This obviously can only beaccomplished if such a catheter is in place If not satisfiedwith the fixation of the coil in the vessel or the degree ofocclusion, the coil can easily be withdrawn into (and, ifdesired, out of) the sheath, and the procedure restarted in
a new location of the delivery sheath or with a differentcoil When satisfied with the fixation in the vessel and thedegree of occlusion, the bioptome catheter is advanceduntil the jaws have advanced completely out of the tip
of the sheath The bioptome jaws are opened, releasing the coil
The advantages of the bioptome-controlled delivery ofcoils are the obvious complete control over the actualrelease of the coil and the complete retrievability of the
coil until the moment of its purposeful release The major
disadvantages to this technique are the requirement for
a slightly larger delivery system, the lower flexibility ofthe long sheath delivery system for entering more tortu-ous locations or vessels arising at acute angles off themajor feeding vessel (aorta), and the added expense of thebioptome
Trang 16Special commercial modifications of the delivery
system or the coil to control the delivery of the
Gianturco™ type coil
Detachable™ coils: Jackson™ coils, Cook detachable
and Flipper™ coils
Outside the United Statesai.e in countries not under US
FDA jurisdictionathe Jackson™ (or Detachable™) coil is
available commercially from Cook™ Inc., Europe This is
a safe, detachable or controlled-release variation of the
standard Gianturco™ coil The Detachable™ coil is in
widespread, standard, use for PDA occlusion as well as all
other types of vascular occlusions, particularly in Europe10
The European Detachable™ coil is not available in the
United States The Detachable™ coil is slightly less robust
than the standard Gianturco™ coil, however, for the
occlusion of most vascular lesions other than the PDA,
this relative flimsiness is of little consequence The
relat-ively simple attach/release mechanism of the Detachable™
coil gives complete control over the release and
retract-ability of the coil, making vascular occlusions in even the
most precarious vessels a safer and more effective
pro-cedure The Detachable™ coil, with its specific pusher/
delivery wire is, however, slightly more expensive than
the standard “free-release” Gianturco™ coil
The Detachable™ coil essentially is a standard
Gianturco™ coil that has been fitted with a screw
mechan-ism for the purposeful attachment and release from the
delivery or “pusher” wire The European Detachable™
coil comes from the manufacturer as a “set” containing the
coil, a special clear loader, a special delivery/pusher wire
and a fine movable mandril, which passes through the
delivery wire and the coil The clear loader has a slight
funnel at the proximal end where the “female” screw
mechanism in the coil is located The Detachable™ coil
itself outwardly has the same appearance as a standard
Gianturco™ coil, with the distal end of the occlusion coil
sealed with a rounded, smooth and polished “weld”
while the proximal end of the coil is hollow The open,
proximal end of the straightened occluder coil is “squared
off”, hollow and appears slightly irregular on very close
inspection The fine wire windings within the proximal,
hollow end of the coil form a “female screw thread” for
the attachment of the “pusher” wire The special attach/
release delivery or “pusher” wire consists of a spring
guide wire of the same diameter as the coil to be used The
pusher wire has a long, tapered “male” screw mechanism
as an integral part of its distal end and comes with a
removable “torquing” device for purposeful rotation
of the wire/screw tip The coil attaches to the delivery/
pusher wire with this very simple screw mechanism
In addition, the special delivery/pusher wire has no
fixed “core” wire, but is hollow throughout its entire length,
including through the fine screw at the distal end The
small lumen throughout the wire allows a very fine,smooth, steel “mandril” or stiffening wire to pass com-pletely through, out of the end of the pusher wire andthrough the coil The fine, totally removable “mandril”wire, which is approximately 10 cm longer than the com-bined pusher wire and the straight coil wire, comes pack-aged within each pusher wire The mandril passingwithin the delivery/pusher wire and the coil acts as a stiff-ening “core” wire The mandril has a short segment of
“spring” wire attached at its proximal end to serve toidentify the proximal end and as a “hub” for moving andtorquing the mandril Detachable™ coils commerciallycome stretched out as a straight spring wire positioned
within a thin, clear, straight loading tube of approximately
the same ID as the OD of the coil
Once the delivery catheter is positioned properly andsecured in place, the coil is attached to the pusher wire.The clear loader containing the occluder coil is inspectedvery carefully Each end of the occluder coil is positionedslightly more than one centimeter within the ends of theloader The proximal and distal ends of the occludercoil/loader are identified The screw end of the deliv-ery/pusher wire is introduced into the proximal (fun-neled) end of the loader The mandril wire is advancedthrough the delivery/pusher wire and 8–10 mm beyondthe distal tip of the “screw” mechanism of the delivery/pusher Very carefully and without pushing the coil for-ward in the loader, the mandril is introduced into theproximal end of the coil by gentle trial and error probing
and then advanced 8–10 mm into the hollow coil The
delivery/pusher wire is advanced over the mandril untilthe screw at the tip of the delivery/pusher wire enters into the proximal end of the coil When the screw tip hasengaged in the proximal end of the coil, the coil andpusher wire are attached by two and a half clockwiseturns on the pusher wire or on the coil loader when it isgripped facing the pusher wire The screw at the tip of thepusher wire enters and engages with the proximal end
of the coil Once the screw has tightened within the coil,
the delivery/pusher wire is turned counter-clockwise one half turn to slightly loosen the attachment Gentle, and very slight to-and-fro motion of the pusher wire within
the loader is used to test the attachment of the coil to thepusher wire Holding the pusher wire, the core/mandrilwire is advanced the remainder of the way into the coiluntil the mandril reaches the distal, closed tip of thestraightened coil within the loader
The distal end of the clear coil loader is introducedthrough a wire back-bleed/flush device into the hub ofthe previously positioned delivery catheter The deliv-ery/pusher wire with the enclosed mandril is advancedinto the loader, which, in turn, pushes the attached,straightened Detachable™ coil out of the loader and into
Trang 17the delivery catheter The mandril remains in the
deliv-ery/pusher wire as the pusher wire is advanced The
distal end of the catheter is observed continually on
fluoroscopy as the coil advances through the catheter to
the tip of the catheter and as the coil is advanced out of the
tip of the catheter into the desired location for implant by
advancing the delivery/pusher wire further The mandril
is maintained within the delivery wire and coil until the
coil is at least partially out of the delivery catheter and the
tip of the coil is in or slightly distal to the correct position
for implant The mandril still within the coil keeps the coil
straight and allows easy adjustments in the position of the
coil Once the coil is partially out of the catheter and in
proper position for implant, the mandril is withdrawn
slowly, allowing the coil to curl into its “coiled”
configura-tion in the vessel/lesion The pusher catheter and coil are
advanced further while the mandril is withdrawn to just
within the proximal end of the coil and until the coil is totally
extruded from the catheter and formed into its full coil
configuration At this point the coil is tested for its proper
location and its fixation in the vessel by slight to-and-fro
motion of the pusher wire or with an angiogram When
satisfied with the fixation of the coil and the occlusion of
the vessel, the small “vise” is attached to the proximal end
of the delivery/pusher wire and the vise with the pusher
wire/mandril is rotated counter-clockwise until the coil
detaches from the pusher wire The release of the coil as
the coil is being detached is observed very closely under
fluoroscopy to ensure a smooth “unscrewing”, and that
there is no binding or twisting of the still-attached coil
If the Detachable™ coil is not in the exact, desired
posi-tion, at any time before it is purposefully “unscrewed”, it
can easily be withdrawn back into the delivery catheter
As the coil is withdrawn into the catheter, the mandril
remains positioned proximal to and outside of the coil
If the same coil is to be repositioned using the same
catheter, the mandril is re-advanced into the coil (which
now is straightened within the catheter) before the coil
is re-extruded into the vessel This retrievability during
delivery adds total control and a significant degree of
safety to the occlusion of any vascular structure The total
European Detachable™ coil systems are slightly more
complicated and significantly more expensive than the
standard Gianturco™ coils by themselves
Flipper™ coils
The Flipper™ coil (Cook Inc., Bloomington, IN) is
an attempt in the US at a version of the European
Detachable™ coil The earlier US version of a detachable
stainless steel coil of a size considered reasonable to use in
PDA occlusions was a catastrophe The original “larger”
detachable coil which became available in the US was
made from a smaller, less robust spring wire, had far fewer
incorporated “thrombotic” fibers, and had a bizarre,
unnecessarily complicated attach/release mechanism As
a consequence this coil had very little use The Flipper™coil is a revised version of the Cook, Inc., US detachablecoil, which became available in 2001 with an attach/releasemechanism similar to the European Detachable™ coils.The Flipper™ coil wires are 0.035″ wires and still do nothave comparable robustness nor the occlusive capabilities
of the 0.038″ standard Gianturco™ coils Flipper™ coilsare available in 3–12 cm lengths and with coil diametersbetween 3 and 8 mm Current Flipper™ coils have morefiber strands embedded in the coil wire than the original
US version of the detachable coil, and for a 0.035″coil doseem to have comparable occlusive properties to thesmaller 0.035″ standard Gianturco™ coils Flipper™ coilsrequire a delivery catheter with a 0.041″ inner diameter,which generally is a 5-French catheter
The lack of robustness of the 0.035″ coil compared to the0.038″ coil is a limitation of this coil for its use for primarilyclosing a patent ductus arteriosus, but the controllabil-ity and retrievability of the Flipper™ coil make it veryappealing for most other vascular occlusions or for theimplant of additional coils for the “final and total” occlu-sion of a patent ductus which already has a larger, securecoil in place
The packaging, attaching, loading, delivery and release
of the Flipper™ coils are all similar to the EuropeanDetachable™ coils described above
Alternative occlusion coils including “Micro” coils
In addition to the standard and large sized Gianturco™
type coils, there is a large variety of other small coils able for small vessel occlusion The alternative coils are
avail-delivered through significantly smaller and more flexiblecatheters, and as a consequence can be placed in more circuitous and distal locations None of the smaller alter-native coils are as robust as the stainless steel Gianturco™coils
Target™ platinum coils and their delivery technique
Target™ Coils (Target Therapeutics, Fremont, CA) are similar in concept to the Gianturco™ coils They aresegments of very fine spring guidewire-like wires withthrombogenic fibers intertwined in the spaces betweenthe wire coils of the spring wire There, the similarityends Target™ coil wires are made of very small diameter0.014″ and 0.018″ platinum wires In spite of their verysmall diameters, and because of the platinum material,these coils are very radio-opaque and easily visible underfluoroscopy in the catheterization laboratory Because ofthe material and tiny size of these coils, they are moreflexible and pass through tortuous catheters/vessels eas-ier than Gianturco™ coils Target™ coils do not open into
Trang 18a cylindrical “coil” in their relaxed state but rather, into
complex “helix” configurations which vary between 4
and 7 mm in diameter Because of their size and less
“fiber material”, these coils are more compressible in a
particular vessel and, in turn, not as occlusive as the larger
stainless steel Gianturco™ coils
Target™ coils are delivered through specific, very
small (3-French), very flexible Tracker™ delivery catheters
(Target Therapeutics, Fremont, CA) The Tracker™
catheters are small and “trackable” but are
non-radio-opaque throughout their entire length except for a tiny
radio-opaque marker at the very tip of the catheter This
tip marker identifies the most distal location of the tip
but, once the delivery/pusher wire is removed from the
Tracker™ catheter, the course of the catheter proximal to
the tip is invisible! Tracker™ catheters are advanced to the
orifice of the vessel to be occluded through a larger
torque-controlled 5-French, guiding catheter From the guiding
catheter, the Tracker™ catheter is advanced through
the vessel over a special, very fine, Dasher™ guide wire
(Target Therapeutics, Fremont, CA) The Dasher™ guide
wire is pre-positioned in the target vessel Because of its
torque control, its very floppy tip and its very small size,
this wire can be manipulated very readily to very distal
locations and through very tortuous and long vessels
To deliver the Target™ coils, a pre-formed,
torque-controlled, 5-French guiding catheter which accommodates
the Tracker™ catheter is positioned in the orifice of the
vessel to be occluded or, alternatively, into a trunk vessel
off of which the target vessel arises The special Dasher™
tracking wire, which is 0.014″ in diameter and 175 cm long
with a very flexible tip, is passed through the guiding
catheter and into the target vessel The specifically curved
floppy tip of the Dasher™ wires can be manipulated
select-ively through very tortuous and very circuitous, distal
vessels using a “torquer vice” on the stiff shaft of the
wire to help to direct the tip of the wire to the specific
loca-tion An appropriately sized Tracker™ (“Tracker™-18”,
150 cm) catheter (Target Therapeutics, Fremont, CA) is
then advanced over the Dasher™ wire and through the
guiding catheter to the orifice of the target vessel As the
Tracker™ catheter is being advanced over the wire a
continual flush is maintained through the guide catheter
and the Tracker™ catheter through the special double
“Y” adaptors attached to the proximal ends of both the
Tracker™ and the guiding catheters
When the Tracker™ catheter is being advanced over the
wire beyond the tip of the guiding catheter and through
the more tortuous areas of the vessel, the Tracker™
catheter is advanced several centimeters at a time while
alternately withdrawing or tightening the Dasher™
wire very slightly In this way the Tracker™ catheter is
“inched” along the Dasher™ wire without displacing the
wire All of this time, only the tip of the Tracker™ catheter
is visible over the wire Occasionally the Tracker™catheter does not follow over the Dasher™ wire and thewire is pulled out of position In that circumstance, oftenthe very flimsy Tracker™ catheter with just the tip of the exposed, very floppy Dasher™ wire maintained justbeyond the tip of the catheter can be advanced (manipu-lated) together as a unit back to the appropriate location.Once the Tracker™ catheter has reached the desired loca-tion, the Dasher™ wire is withdrawn very carefully andvery slowly from the Tracker™ catheter The Tracker™catheter is maintained on a slow constant flush while theopaque tip of the catheter is observed continuously onfluoroscopy to be sure that the tip is not being displacedduring the removal of the wire Once the wire is removed,
again, only the opaque marker at the tip of the Tracker™
catheter will be visible A small, slow hand injection of
contrast is performed through the Tracker™ catheter andrecorded on biplane angiography or stored fluoroscopy inorder to “road map” the actual “course” of the Tracker™catheter for future reference After the “road map” is
recorded, the Tracker™ catheter is flushed very slowly but
very thoroughly to clear it of all contrast material A rapid
or forceful flush of the Tracker™ catheter can easily “blow”the tip of the Tracker™ catheter back out of the vessel.Each Target™ coil comes straightened in a separate,tiny, metal tubular holder/introducer The introducertube is flushed gently and the end, which does not have ahub, is fitted into the straight segment of the “Y” connec-tor at the proximal end of the Tracker™ catheter The coil
is pushed out of the introducer and into the proximal end
of the Tracker catheter with the short “plunger tool”which comes with the coil Once the plunger tool has beenadvanced to the “hilt”, the coil is completely within the
Tracker™ catheter as a short, straight, free segment of
wire Target™ coils fortunately are very radio-opaque.The plunger tool and the introducer tube are removed
The stiff end of the special “coil pusher wire” is introduced
into the proximal end of the Tracker™ catheter and
advanced approximately 30 cms This “pre-advances” the
coil that distance within the Tracker™ catheter The stiffend of the coil pusher is withdrawn, the pusher wirereversed, and the soft nylon distal end of the pusher
“wire” introduced into the Tracker™ catheter All of this
time the Tracker™ catheter and guiding catheter are tained on a slow continuous flush As the coil pusher (soft
main-nylon end first) is advanced in the Tracker™ catheter, thecoil is advanced proportionately through the catheter.This part of the delivery is observed particularly carefully
on the fluoroscopy Extreme care is taken as the coil isadvanced within the Tracker™ catheter to ensure that the tip of the Tracker™ catheter is not withdrawn evenslightly by inadvertent traction applied to it When theadvancing coil and coil pusher wire begin to enter curvesand bends in the course of the Tracker™ catheter, the
Trang 19advancing coil and coil pusher wire tend to “straighten”
the Tracker™ which, if not compensated for by advancing
the proximal end of the Tracker™ catheter, can pull the tip
of the Tracker™ catheter back out of the desired position
The course of the coil, which now is visible through
the “invisible” catheter, is compared to the previously
obtained angiographic “road map” of the catheter course
This is to help ensure that the coil and pusher are not
dis-placing the “invisible” catheter during the introduction
of the coil
When the coil reaches the tip of the Tracker™ catheter,
by further advancing the coil pusher, the coil is extruded
from the distal end of the catheter and is free in the vessel
to be occluded Like the standard Gianturco™ coils, the
Target™ coils have no attachment to the pusher wire
and once extrusion begins, the coil cannot be withdrawn
Also like the Gianturco™ coils, the way the coils unwind
and coil in the vessel depends upon the relative size of the
coil compared to the vessel size In order to accomplish
complete occlusion of a vessel, usually many coils or
mul-tiple sizes of the coils are used As with other occlusions
and whenever possible, the total occlusion of the vessel
should be accomplished completely before the procedure
is abandoned
Controlled release micro coils
There are two varieties of controlled release systems for
the “micro” coils, which do provide retrievability of the
coils even after they have started to be extruded The two
mechanisms are entirely different, but both are
manu-factured by Target™ Therapeutics (Target Therapeutics,
Fremont, CA) Although the primary use of these coils is
by neuroradiologists, they are equally suitable for small
vessel occlusions in the pediatric/congenital population
The first of these controlled release micro coils is
the Guglielmi™ electrolytically detachable coil (GDC)
(Target Therapeutics, Fremont, CA) These are, as the
name implies, released from the delivery wire by a micro
current delivered through the wire, which, in effect, melts
a connection between the delivery/pusher wire and the
coil The second type of controlled release coil is the
Interlocking Target™ coil (Target Therapeutics, Fremont,
CA) These coils have a unique system of micro machined,
overlapping pins or “couplers” which are compressed
within a tiny delivery catheter until the final millimeter of
the coil is extruded from the tip of the catheter Both of
these controlled release micro coils otherwise are used in
similar circumstances to the other micro coils
Tornado™ coils
The Cook Tornado™ coils (Cook Inc., Bloomington, IN)
functionally are almost a cross between the standard
Gianturco™ coils and the Target™ coils As their nameimplies, these coils are shaped like a tiny tornado “funnel”which along with their size and material gives them morecompressibility and, supposedly, more coil “exposure” tothe lumen of the vessel These coils are constructed of plat-inum wire, which is more radio-opaque than stainlesssteel wire of comparable size The coil wires are available
in both 0.018″ and 0.025″ spring wire sizes, which, in turn,are less robust than the standard Gianturco™ coils Eachcoil has multiple strands of tiny synthetic fibers embed-ded along the coil wire to facilitate thrombosis TheTornado™ coil sizes are labeled corresponding to thelargest and the smallest diameters at the ends of the “fun-nel” For example, a 6/2 coil has a 6 mm diameter largeend, which tapers down to a 2 mm diameter small end.The 0.018″ Tornado™ coils are available in sizes from 3/2
to 10/4 with lengths of the straightened coil varying from
2 to 14 cm, while the 0.025″ Tornado™ coils come in sizesfrom 5/3 to 10/5, with lengths of the straightened coilfrom 4 to 12 cm From the standard packaging of theTornado™ coils, the small end of the “funnel” is deliveredfirst from the “coil” holder and loads first By special orderthe Tornado™ coils can be packaged so the large end isdelivered first Tornado™ coils are delivered throughcatheters with internal diameters of 0.025″ or 0.032″,which allow the use of very small, trackable catheters.The 3-French Slip-Cath™ (Cook Inc., Bloomington, IN)
is ideal for the delivery of these coils The Slip-Cath™ isvery flexible and has a very slick hydrophilic coating,which allows it to track through very circuitous and smallvessels Its small outer diameter allows it to be advanced
to the orifice of the vessel that is to be occluded through atorque-controlled, 5- or 6-French guiding catheter Thenylon 3-French infusion catheter (Cook Inc., Bloomington,IN) with a radio-opaque tip is less flexible than the Slip-Cath™ but also makes a good delivery catheter for delivery
to more proximal areas in small, but less tortuous vessels.The technique for loading and delivering theTornado™ coil is similar to the delivery of a “free-release”Gianturco™ coil There is no attach/release mechanism.The coils are introduced into the pre-positioned deliverycatheter from their loader and advanced through thecatheter with a 0.025″ pusher wire When there is a verytortuous vessel proximal to the site of delivery, a very softtipped pusher wire is used A concomitant slow, continu-ous flush through the delivery catheter assists in advanc-ing the coil through the catheter The smaller Tornado™coils occasionally can be advanced through the deliverycatheter by a strong flush on the catheter alone The strongflush, however, may cause the tip of the catheter to recoilout of position, making the “flush” delivery much lessreliable and less precise As with other micro coil occlu-sions, usually more than one coil essentially is alwaysrequired to complete the occlusion After one or more
Trang 20coils are delivered, a small hand angiogram is performed
through the delivery catheter to check the degree of
occlu-sion If there still is any residual leak, additional coils are
delivered until the vessel is occluded or there is no more
space in the vessel for additional coils
Nester™ coils
The Nester™ coils (Cook Inc., Bloomington, IN) are still
another variety of coil for vascular occlusions These are
0.035″ platinum coils with synthetic fibers entwined in the
coil wire In their relaxed, free shape, the coils have a
cylindrical configuration Nester™ coils are all 14 cm in
length and available in 4–12 mm diameter “cylinders”
These coils are softer than even the comparable diameter
Tornado™ coils The Nester™ coils are designed to
occlude by bunching up and packing into a relatively
large (1–2 cm) mass within the vessel
The Nester™ coils are delivered through a catheter with
a 0.035″ or 0.038″ inner diameter and advanced through
the delivery catheter with a 0.035″ teflon-coated spring
guide “pusher” wire identically to the delivery of the
Gianturco™ free-release coils The standard Nestor™
coils have no attach/release mechanism so, like the
Gianturco™ coils, once extrusion of the coil begins, there
is no turning back Because they are so soft and flimsy,
these coils are used mostly as additional or supplemental
“stuffing” coils rather than primary occlusion devices
They are added as “packing” to complete the occlusion
within a previous, stiffer coil or other device already
placed in a vessel
Particulate materials and non-coil intravascular
devices available for vascular occlusions
There are several materials and intravascular devices in
addition to the intravascular coils that are available and
occasionally used for the occlusion of abnormal vascular
communications
Particulate materials
Vascular radiologists use particulate materials
exten-sively for the acute occlusion of vessels and abnormal
communications These materials usually are used to
control bleeding from a specific vessel and/or to totally
occlude vessels with the intent of necrosing the tissues
“downstream” from the occlusion (tumors, neoplasms)
The particulate materials are useful mostly for very
dif-fuse, but, at the same time, “end artery” lesions The
mater-ials used include autologous blood clots, pieces of
Gelfoam™ and pieces of Polyvinyl Alcohol (Ivalon™)
The autologous blood clots and Gelfoam™ usually only
provide a temporary occlusion and for that reason seldom
are used in congenital vascular lesions where a more manent” occlusion is desired
“per-The autologous clots are just thatasmall solid particles
of the patient’s recently clotted blood, which are injectedfrom a syringe through a catheter that is pre-positioned
in the culprit vessel1 The “clot particles” that are usedshould be slightly larger than the vessel that is to beoccluded The already clotted blood acts as a temporaryocclusive mass to occlude the small, preferably “end” ves-sel The clot usually thrombolyses over a short period oftime, however, it usually remains in place long enough toallow acute bleeding from the vessel to stop and occasion-ally long enough for the vessel to thrombose
Gelfoam™ is an insoluble material made from driedpork skin gelatin and formed into porous foam sheets Thefoam material can absorb many times its own weight
in blood and/or other fluids When left in the tissues
it resorbs into the body within 4–6 weeks Gelfoam™ is
approved to enhance coagulation on the surface of
bleed-ing tissues but is not “approved” for intravascular use Itcomes in sterile sheets, which can be cut to any desiredsize or shape, and the pieces can be compressed into verysmall particles
Very small, cut and compressed pieces of Gelfoam™are soaked in a dilute solution of contrast and then areforced into the end vessel that is to be occluded through acatheter by a strong flush with a syringe2 In the vessel, the
“wad” of Gelfoam™ expands and absorbs blood to create
an occlusive mass Gelfoam™ is useful to “complete” or
“finalize” the occlusion of a vessel/fistula which has been started with another occlusion device Obviously,this soft gelatin like material cannot be used to occludehigh-flow and/or high-pressure vessels, and like autolog-ous clot, the Gelfoam™ itself may not create a permanentocclusion The Gelfoam™ particles themselves are notradio-opaque, so embolization to distal locations is onlyapparent from any contrast solution that might beretained in the particles and/or signs or symptoms thatare produced by the embolized particles
Ivalon™ (Polyvinyl-alcohol or PVA) foam particles are available for a similar use to the autologous clots orGelfoam™ pieces The Ivalon™ does provide a more per-manent occlusion but, as small soft particles, still is usefulonly for end vessel and/or “completion” of other vesselocclusions Ivalon™ particles are available commerciallyfrom Cook Inc (Bloomington, IN) as dried particles inmultiple different particle sizes between the smallest50–100 micron and the largest 2000–2800 micron sizes toaccommodate multiple different vessel sizes The driedIvalon™ particles are mixed with contrast material tosoften them and give them some radio-opacity for injec-tion into the circulation The particles mixed with dilutecontrast are drawn out of their sterile container into
a syringe and injected with the same syringe into the
Trang 21desired site through a pre-positioned catheter Similar to
the autologous clots and Gelfoam™, there is little control
over the location where the particles are delivered and
essentially, no retrievability
“Intravascular” glues
There was a transient interest by several different
investigators for the use of the tissue adhesive isobutyl
2-cyanoacrylate (Bucrylate™) as an occlusive/embolic
material to occlude large vessels/vascular
communica-tions4 The Bucrylate™ polymerizes into a solid mass
instantaneously on contact with blood The area for
deposit of the Bucrylate™ must be “isolated” from the
surrounding circulating blood and/or the Bucrylate™
must be injected very specifically into the area
Bucrylate™ was very difficult to handle and to control
during delivery both during animal testing and during
several clinical uses If injected too fast it embolized
dis-tally and/or backward into proximal branches and, in
doing so, occluded all areas that it entered If injected
too slowly, it occluded the injecting catheter when only
partially extruded The difficulties with delivery of
Bucrylate™ compared to other occlusion materials/
devices, and the unknowns about long-term carcinogenic
effects of Bucrylate™ in humans, led to it being
aban-doned for human trials in the United States
Occlusion devices for large vascular
communications
Gianturco-Grifka Vascular Occlusion Device™
(GGVOD™)
Although it does contain spring guide wire and is
ap-proved for use in the US as a variation of the Gianturco™
coil, the Gianturco-Grifka Vascular Occlusion Device™
(GGVOD™) (Cook Inc., Bloomington, IN) has little
resem-blance in either appearance or use to the Gianturco™
coils The GGVOD™ is a nylon bag or sack of a
predeter-mined fixed diameter into which a specific length of
spring wire is wadded to achieve a tense, fixed diameter,
mass of spring wire11 The wire for the packing of the bag
of the GGVOD™ is a spring guide wire with the stiffening
core and safety wires removed The bag only serves to
contain the mass of wire within the fixed diameter of the
bag The bags are available in diameters of 3, 5, 7 and
9 mm, each coming with a specific length of “packing”
spring guide wire The bags are somewhat flattened and
elongated so that they do not form a circular or spherical
configuration in their length or cross section The stated
diameter is the largest measured cross-section of the
par-ticular bag and not the actual circumferential diameter
of the bag
The GGVOD™ is usable only in tubular vascular tures that are at least 1.5 times longer than their respectivediameters The GGVOD™ is held in the vessel by theradial force created by the wad of spring wire within the bag and exerted against the surrounding vessel wall.The bag diameter should be at least one millimeter largerthan the diameter of the vessel to be occluded If the par-ticular vessel is very distensible, then a larger diameterbag to vessel ratio is used
struc-The technique for delivery of the GGVOD™ to mostabnormal vascular communication is similar to the deliv-ery of the GGVOD™ to the PDA (as described subse-quently in Chapter 27) with one major exception Whendelivering the GGVOD™ to a collateral or branch vesselcompared to the delivery to the PDA, the delivery sheathinitially and, in turn, the bag is delivered exactly to the
implant site and all of the wire extrusion into the bag is
directly into the site where the bag is to be implanted This
is in comparison to the delivery to the usual patent ductus,
where the sheath and the bag are advanced beyond the
implant location and some of the filler wire is extrudedinto the bag in this distal location before the bag/deliverycatheter is withdrawn back into the specific site within theductus for the implant The fixation of the bag and/or theability to push all of the filler wire into the bag depend(s)upon a relatively precise and, at the same time, tight fit ofthe filled bag into the vessel/channel As with the delivery
of coils to critical locations, the vessel/channel can be
“sized” with a small angioplasty balloon that is slightlylarger than the vessel/channel and inflated at a low pres-sure in the vessel/channel
Because of the complexity of the delivery and release
of the GGVOD™ and the critical importance of each vidual step, the details of the GGVOD™ delivery to vas-cular communications other than the PDA are listed here
indi-in a tabular, “cook book” form
Components of the GGVOD™ system
• A long, 8-French, valved outer delivery sheath with adistal marker band
• The nylon “sack” or “bag”: in 3, 5, 7 and 9 mm sizes(diameters)
• An inner, stiff walled, pusher catheter to which the bag
• A second, stiff, middle, “bag release” catheter which ispre-positioned over the pusher catheter
• A complex attach/release system which joins all ofthese components
Trang 22• A short length of 8-French “peal away” sheath which
comes over the distal ends of the two catheters and the
attached “bag”
Six major steps in the implant of a GGVOD to an
abnormal vascular communication (except a PDA)
• The long delivery sheath is advanced into the abnormal
vascular communication until the sheath tip is just within
the area where the “sack” is to be implanted
• The “sack” is introduced into the long sheath through
the “peal away” sheath
• The “sack” is positioned precisely in the area to be
occluded and the sheath is withdrawn off the sack
• The filler wire is pushed into the sack
• The pusher wire is separated from the filler wire
• The filled “sack” is separated from the pusher catheter
Details of the six major steps for the GGVOD
delivery to an abnormal vascular communication
Step 1: The long delivery sheath is positioned in
vascular communication with the tip of the sheath
1–2 cm beyond the area where the sack is to be
implanted
• The abnormal vascular channel is entered with an
end-hole catheter which is advanced as far beyond the area to
be occluded as possible
• A second catheter from a second vascular entrance site
is placed adjacent to the proximal (in the direction of flow)
end of the communication
• The end-hole catheter is replaced with a 0.035″ stiff
exchange wire
• The special 8-French, valved delivery sheath/dilator
set with a marker band at the distal tip of the sheath is
advanced over the wire into the abnormal channel until the
tip of the sheath is just beyond the area of the channel where
the GGVOD™ is to be implanted Unless the delivery
sheath tip can be advanced just distal to the location for
occlusion, the GGVOD™ cannot be used
• The dilator and wire are removed, the system is
cleared of air and clot and then the sheath is flushed
thoroughly
Step 2: The “sack” and delivery system introduction
into the long sheath
• All of the components of the GGVOD system are
inspected carefully
• The “sack”, which is attached to the pusher catheter
and the wire/delivery/release system, is introduced into
the pre-positioned long sheath as a single unit through the
short “peal away” sheath
• The short “peal-away” sheath is removed after the
“sack” is completely within the sheath
• The “sack” is advanced with the attached delivery system
to the tip the sheath (but still within the sheath tip) to theexact location where the GGVOD™ is to be implanted
Step 3: The “sack” is positioned in the vessel/
channel
• With the pusher catheter held in place, the sheath iswithdrawn off the “sack” which now is in the precise position for implant The “sack” itself is invisible, but the attaching band on the proximal neck of the “sack” isvisible and defines the proximal limit of the “sack”
• While observing on fluoroscopy, several loops of the
“filler” wire are advanced very loosely into the “sack” by
advancing the stiff pusher wire 5–10 cm The “sack” is not
packed tightly with the filler wire at this point
• The position of the “sack” is checked with a smallangiogram through the second catheter positioned in oradjacent to the proximal end of the channel
Step 4: The filler wire is pushed into the sack
• With the “sack” in the proper position, the remainder
of the filler wire is fed completely into the “sack” This is
checked on fluoroscopy being sure that the connection
point of the filler wire with the pusher wire is within the sack This is identified by a difference in X-ray densities at
the connection point
• The exact “sack” position in the vessel and the degree ofocclusion are checked by another angiogram through thesecond catheter
Step 5: The pusher wire is separated from the filler wire
• When sure the filler wire is completely within the
“sack” and while observing on fluoroscopy, the wire
release mechanism is activated by pushing forward
on the two side “loops” of the special attach/release handle This detaches the pusher wire from the filler wire
• The pusher wire is withdrawn into the pusher catheter athis releases and separates the pusher wire from the
“sack” This withdrawal of the pusher wire should be
very slowauntil absolutely sure that the pusher wire is
separated from the filler wire
Step 6: The pusher catheter is separated from the sack
• The stiffer, middle, release catheter, which is over the
pusher catheter but within the sheath, is advanced snuglyagainst the neck of the “sack”
Trang 23• With the tip of the release catheter held exactly in place
against the sack, the inner “pusher” catheter is withdrawn
forcibly from within the neck of the “sack” (If the release
catheter moves at all, the position of the “sack” very likely
will move with it!!)
• The forceful withdrawal of the “pusher catheter”
releases the “sack” in the channel entirely free from the
delivery system
The advantages of the GGVOD™ for vascular occlusions
are the same as its advantages for occlusion of the PDA
The GGVOD™ is excellent for the occlusion of relatively
large and/or high flow vascular channels or structures
that have any associated length In this type of lesion, the
GGVOD™ usually produces immediate, complete
occlu-sion Even the largest “sack” is delivered through only an
8-French sheath, which is reasonable in all patients past
infancy Once delivered and before it is released
purpose-fully, if the “filled bag” does not fit in the vessel, is loose
in the vessel or otherwise does not occlude the vessel, the
wire can be withdrawn and the bag repositioned or even
replaced with a bag of a different size The larger diameter
GGVOD™ devices also are applicable to some larger,
high flow vessels and channels where coils alone are not
sufficientafor example in large abnormal venous
chan-nels or large pulmonary arteriovenous malformations
One of the greatest advantages of the GGVOD™ is that
it is approved for this use and is available in the United
States
The GGVOD certainly is not a “universal” device,
only being applicable to tubular vascular lesions that are
somewhat longer than they are in diameter Another
dis-advantage of the GGVOD for vascular occlusions relates
to the size and stiffness of the delivery sheath Since most
small vessels requiring occlusion arise from the aorta and
require arterial access, the 8-French delivery sheath
repres-ents a significant disadvantage for use in infants or small
children In addition to its diameter, the GGVOD™ is
delivered through a long sheath which is not as flexible as
most delivery catheters, so delivery of the device into
acutely angled and/or tortuous vessels is difficult, if not
impossible Replacing the delivery sheath that comes with
the GGVOD™ with a Flexor™ type sheath overcomes
some of this problem A final disadvantage is that the
GGVOD™ is quite complicated to use and when used
infrequently, the delivery/release technique must be
“relearned” with each use
There have been rare complications encountered with
the use of the GGVOD™ even in the occlusion of vascular
structures other than a PDA On several occasions the
filler wire either could not be or was not pushed entirely
into the “sack” and/or was pulled partially out of the
“sack” as the pusher wire was withdrawn (incompletely
released?), resulting in a segment of the filler wire
extend-ing into the lumen of the vascular channel after the release
of the “sack” In most occlusions other than the PDA, thiscreates little or no problem since the goal is to occlude theentire channel If the filler wire extends back into a neces-sary or vital feeding channel and cannot be “wadded”back into the channel that is being occluded, the wire can
be captured with a vascular snare The wire can be bentback and forth repeatedly against the orifice of the targetvessel and, in doing so, broken off from the more distalwire, which is within the target vessel If the wire cannot
be broken, the entire wire can be withdrawn from the sackand out of the body This, unfortunately, leaves the “sack”empty and often destined to embolize to a more distallocation Under these circumstances and before with-drawing the wire completely out of the “sack”, the “sack”should be grasped with a separate retrieval device and/orthe vessel distal to the “sack” selectively occluded with atemporary occlusion balloon unless it is determined thatthe more distal vessel is expendable if permanentlyoccluded with the embolized empty sack
When undersized and/or in a very compliant vessel,the entire, full “sack” can embolize distally from itsdesired site in the abnormal channel Occasionally thismerely occludes the desired channel in a different, but still effective location If, however, the embolized “sack”migrates into the central or other vital areas of the circula-tion, the embolized sack must be removed, which cannot
be accomplished as a full sack First, the full bag is snared
to hold it in position If the neck of the sack can be snared,
it should not be pulled with any force Unusual tension onthe neck can detach the only radio-opaque part of the bagitself!! Once the bag is grasped, a hole is “chewed” in thebag with a bioptome forceps introduced from a separatevessel and through a slightly larger sheath Once a hole
is created in the full “sack”, loops of the packed filler wire usually will extrude immediately through the hole,and/or while “chewing” the hole, a loose portion of thefiller wire becomes exposed and is grabbed The graspedwire then is carefully withdrawn with the forceps whilethe “sack” is held with the snare or other retrieval device.The emptied bag is then withdrawn utilizing the devicethat is holding the sack
In spite of the disadvantages and rare complications,the GGVOD™ has some very specific uses and it should
be available to any laboratory heavily engaged in peutic catheterizations
thera-Spring wire alone
The steel wire of a spring guide wire itself is genic This property of the wire causes unwanted throm-bosis on the wires within the blood stream and withinsheaths and catheters Thrombosis on spring guide wiresrepresents a continual potential for embolic complications.This same property of the wire can be used purposefully
Trang 24thrombo-to promote thrombosis and/or occlusion in very large
abnormal chambers (aneurysms) or channels, either alone
or packed on top of other occlusion devices In order for a
spring guide wire to be used as a stimulant to thrombosis
and vascular occlusion, the wire should be capable of being
wadded into a fairly compact mass Flexibility of the
wire is accomplished by removing the straight, stiffening
safety core wire(s) from inside of a standard stainless steel
spring guide wire Once the safety core is removed from a
spring guide wire, the remaining spring wire becomes
very soft and can be compacted easily into a tight mass
Once a guide wire has the core wire removed, several
meters of 0.025–0.038″ spring wire can be delivered into a
2–3 cm diameter space Once packed within any confined
area that has little or no flow within it, such as a
peduncu-lated aneurysm, thrombosis and obliteration of the area
occur very effectively
Like a “free-release” coil, long spring wires used in this
fashion cannot be retracted back into the delivery catheter
once the extrusion of the wire has started If part of the
wire inadvertently protrudes out of the specific area being
“packed”, usually the loose end of the wire can be
recap-tured with a snare catheter and can be withdrawn
com-pletely if it cannot be positioned ideally
Amplatzer™ Vascular Plug
The Amplatzer™ Vascular Plug (AGA Medical Corp.,
Golden Valley, MN) is the latest addition to the
armamen-tarium for the occlusion of abnormal vascular
commun-ications and is available even in the United States This
unique device is still another modification of the Nitinol™
wire weave or “basket-like” occlusion devices already in
common use for the occlusion of the patent ductus, atrial
septal defects, patent foramen ovale and ventricular
sep-tal defects The Amplatzer™ Vascular Plugs are small,
cylindrical, Nitinol™ wire mesh “plugs”, which are
avail-able in various diameters, but have no flanges or retention
disks at either end of the “plug” The plugs are
manufac-tured of 144 strands of a finer, 0.004″ Nitinol™ wire
Unlike most of the other Amplatzer™ occlusion devices,
the vascular plugs have no polyester disks within them
and, in turn, rely on the fine mesh of the Nitinol™ metal
mesh for occlusion of the vascular structure Similar to
the other Amplatzer™ devices, the plugs do have similar
metal posts or markers at the center of each end of the
plug, with a female micro screw within the post at
the proximal end of the device The female micro screw
allows attachment with an identical screw mechanism to a
standard Amplatzer™ delivery cable The plugs achieve
their fixation in the particular vessel by the tension against
the wall of the vessel by the expansion of the device
against the wall of the vessel/structure similarly to the
other Amplatzer™ devices
The plugs are available in diameters from the smallest
of 4 mm increasing in 2 mm increments up to a maximum
of 16 mm in diameter The 4–10 mm plugs are 7 mm longwhile the 12–16 mm plugs are 8 mm in length The finermetal wires and the lack of polyester disks allow the plugs
to be implanted through smaller delivery systems thanthe comparably sized Amplatzer™ PDA and VSD occlu-sion devices The 4–8 mm diameter devices are deliveredthrough a 5-French, 0.056″ internal diameter sheath, the
10 & 12 mm devices through 6-French, 0.067″ internaldiameter sheaths and the 14 & 16 mm devices are deliv-ered through 8-French, 0.088″ internal diameter sheaths.The Amplatzer™ Vascular Plugs are usable for theocclusion of multiple different venous and arterial struc-tures, all of which, however, must have some “tubular”configuration This includes residual Blalock–Taussigtype shunts, the tubular PDA, systemic to pulmonary collaterals, systemic to pulmonary vein communications,pulmonary arteriovenous fistulae, and even some peri-valvular leaks The diameter of the plug that is usedshould be 2–3 mm larger than the diameter of the lumenthat is to be occluded The Amplatzer™ Plugs, like theother Amplatzer™ devices, have the advantage of beingfully retrievable until they are purposefully released,which makes a test occlusion with the device possiblebefore the operator is committed to releasing the plug
In addition to “the tubular” characteristics of the lesion,the use of the Amplatzer™ Vascular Plug depends uponthe ability to maneuver the delivery catheter into (andpast) the lesion that is to be occluded The current deliv-ery sheath along with the relatively stiff delivery cable precludes the use of the plugs distally in very tortuouslocations, while, at the same time, the retrievability of theplug allows its placement more proximally in tortuousvessels where other non-controlled release devices would
abnor-is dabnor-iscussed in detail for its use for PDA occlusion inChapter 27 Unfortunately, this device is only available
outside of the United States and in a limited clinical trial in
the US Nit-Occlud™ devices are tightly wound coils ofNitinol™ wire, which are pre-shaped to conform to sev-eral different shapes of “typical” conical PDAs Like itspredecessor, the Duct-Occlud™ device, the Nit-Occlud™devices have no fibers intertwined in their coil windings
to help promote thrombosis/occlusion Nit-Occlud™
Trang 25devices depend entirely on the tightness of the coil
con-figurations, the mass of Nitinol™ wire, and the
thrombo-genicity of the metal itself to occlude the vessels The
precise control/release mechanism of the Nit-Occlud™
device makes it useful in situations in vascular channels
where a precise and very controlled positioning is
essen-tial Like the bioptome control for the delivery of standard
stainless steel coils, the attach/release system for the
Nit-Occlud™ makes the delivery system somewhat stiffer
than the usual Gianturco™ type coil delivery catheters
which, in turn, makes these devices less useful for lesions
in more tortuous vessels and/or circuitous locations Where
it is available outside of the US (and hopefully sometime
eventually in the US) the Nit-Occlud™ device should be
considered for the occlusion of larger and unusual and/or
abnormal vessels/vascular communications/leaks where
the placement of the occluder is very precarious
Catheter delivered detachable vascular occlusion
balloons
Several types of detachable occlusion balloon were
avail-able around the world and even in the United States in the
past Although still available in some locations around the
world, none of the detachable occlusion balloons are
available any longer in the US market
B-D Mini-Balloon™
The B-D Mini-Balloon™ occlusion device (Becton Dickson
Co.) probably had the widest use in congenital heart
defects when it was available12 These were very tiny
occlusion devices The balloons were 1 mm in diameter
when deflated and up to 5.3 mm in diameter when
inflated They were delivered “hydraulically” by a unique
(and complex) delivery system The deflated balloon came
attached to the tip of a very soft, flexible, 1 mm diameter
delivery catheter This catheter, with the attached deflated
balloon, was coiled into a spherical “delivery chamber”
These delivery chambers were flattened at two opposite
ends, giving the chamber the appearance of a toy
“rotat-ing top” The delivery chamber had an inlet and outlet
port at the opposite flatter ends of the chamber
A larger, more maneuverable, guiding catheter, which
accommodates the 1 mm delivery catheter, is maneuvered
into the origin of the vessel to be occluded The outlet port
of the delivery chamber which contained the delivery
catheter and attached balloon was attached to the
prox-imal end of the guiding catheter with a Lure-lock
connec-tion By a rapid, forceful, hand flush into an inlet port
in the chamber, the delivery catheter with the deflated
balloon attached at the tip was hydraulically forced into
and through the guiding catheter, and from there into the
vessel The catheter with the deflated balloon actually
floated along with the flow of the fluid and blood in thevessel With this delivery technique, the tiny deflated bal-loon and catheter would traverse almost any bend, curve
or loop throughout the course of the vessel Usually thecatheter with the deflated balloon traveled completelythrough the vessel to, or even past, the end of the vessel.There was, however, no way to direct the course of the bal-loon/catheter purposefully if there was a branch or bifur-cation in the vessel Once delivered, the fine soft, deliverycatheter with the attached deflated balloon was with-drawn slowly until the deflated balloon reached a positionprecisely at the site within the vessel where the occlusionwas to occur
The balloon then was inflated with a predeterminedamount of contrast solution, which was diluted to beexactly isotonic With this inflation and with the balloonstill attached, the degree of occlusion and, of more import-ance, the tightness of the fixation of the balloon within thevessel were tested If the balloon migrated distally and/ordid not fix securely in a suitable position in the vessel, itwas deflated and withdrawn to a different more proximalarea or withdrawn completely If the fixation and occlu-sion seemed satisfactory, then the balloon was furtherinflated with several tenths of a ml more of the dilute contrast, which, in turn, fixed the balloon in place moresecurely and occluded the vessel The delivery catheterwas pulled away forcefully from and out of the “neck” ofthe “fixed” balloon, hopefully leaving the balloon in place.The B-D Mini-Balloon™ had a self-sealing valve in theneck of the balloon, which kept the balloon inflated afterthe delivery catheter was pulled out of the valve The iso-tonic contrast within the balloon would neither leach innor out of the balloon, once the balloon was inflated inposition Occasionally, even with this technique, the tinyballoons would work loose and migrate (embolize) more
distallyausually with no consequences as long as they
remained in the same vessel A 10 mm occlusion balloon
on a 2 mm diameter catheter, which was delivered withthe same system, was developed at about the time theentire system/concept was abandoned
The B-D Mini-Balloon™ had the capability of being able
to be delivered through very tortuous channels to wise inaccessible locations At the same time, the deliverytechnique was quite complex and initially was difficult for many operators to master, so most operators wouldchoose an alternative device (coil) unless the locationmandated the use of the occlusion balloon In addition,because of the balloon material, these occlusion balloonshad a definite and relatively short (one year) shelf lifebefore the materials of the balloon deteriorated andbecame unusable There have been other detachable
other-“mini-balloons” used sporadically and in isolated casesfor the occlusion of congenital heart lesions, but like the B-D Mini-Balloon™ they have been withdrawn from the
Trang 26market, not because of complications, but because of the
very small overall market, the small demand for the more
complicated balloon occlusion techniques, and the
avail-ability of alternative devices which are simpler to use
A second larger Goldvalve™ Detachable Balloon
Embolization Device™ (DB™) was, and may still be,
available in Japan The Goldvalve™ balloons are
manu-factured of latex and are available in multiple sizes
varying from a small 6 mm cylindrical to a large 15 mm
diameter, 30 mm long balloon The Goldvalve™ balloons
attached to a 2–3-French delivery catheter are delivered
through a 9-French catheter The 9-French delivery catheter
is advanced over a long, stiff wire all of the way to the site
for the implant/occlusion and the balloon advanced
through the delivery catheter to the site Once in its proper
position the balloon is filled with a 2-hydroxy-ethyl
methacrylate (HEMA) solidifying solution Two different
liquid components of the HEMA compound are
intro-duced into the balloon, and because of a glucose oxidase
in one of the liquids, the liquids within the balloon
spon-taneously solidified over 40–60 minutes to form a solid
occlusive device13 Once the material solidifies, the
9-French outer catheter is pushed against the inflated
bal-loon, the attached small delivery catheter is pulled against
the 9-French outer catheter to free it from the balloon, and
both catheters are removed
Because of the necessity of delivering the large 9-French
delivery catheter to the lesion, these balloons had no
advantages in either delivery or occlusion capabilities
compared to standard occlusion coils and other occlusion
devices Because of the relative unknowns of the HEMA
materials in the human body over long periods of
time, these DB™ devices probably will not make it to the
United States
At this time there are no balloon occlusion devices
commercially available on the US market for permanent
occlusion of vessels and abnormal vascular channels
Amplatzer™ PDA & VSD devices for the occlusion
of abnormal vascular channels
The Amplatzer™ PDA and VSD occlusion devices
(AGA Corp., Golden Valley, MN) are better suited for the
occlusion of larger vascular channels than many of the
previously discussed devices, and are relatively recent
additions to the devices available for the occlusion of
abnormal and/or unwanted vascular channelsaeven in
the United States The Amplatzer PDA and VSD devices
are described in detail in Chapters 27 and 30, respectively
There also is the Amplatzer™ Plug (AGA Corp., Golden
Valley, MN), which is discussed earlier in this chapter and
is somewhat similar to the VSD and PDA devices
The three devices vary only a little from each other
in their design All three are roughly cylindrical, short,
closed tubes of a fine weave of Nitinol™ memory metal
The VSD devices have small 2 mm lips or rims at each end
of a “cylinder” and the cylinder or hub is not tapered Thecentral “tube” of the PDA device is tapered slightly andhas a rim, but only at the more distal end, which has theslightly larger diameter The “Plug” is a short cylindricaltube with no “rim” at either end and is manufactured of afiner Nitinol™ wire Which one of the devices is prefer-able for any particular vascular channel, depends uponthe particular anatomy of the lesion and the flow withinthe particular channel
The Amplatzer™ PDA and VSD devices as well as the Plug, all occlude by the same mechanism TheAmplatzer™ devices all self expand into their “resting”configurations in a vascular channel when extruded fromtheir delivery sheath When the channel is smaller indiameter than the particular device that is used, the deviceapplies pressure against the vessel walls outward and allaround the lumen to hold the device in place and fill thelumen of the channel Although none of these devices wasdesigned specifically for other types of vascular lesions orchannels, this mechanism of occlusion does make themideal for the occlusion of larger abnormal vascular chan-nels/leaks, which are broad, not necessarily very long,and which have high flow through them
As opposed to the delivery to the PDA or VSD, wherethe Amplatzer™ devices are positioned more or lessstraddling the lesion, their delivery to vascular channelsand/or fistulae usually involves implanting the deviceswithin the body of the channel or vessel or even in a channel which tapers further distally As a consequence,these devices, when expanded, potentially fix even more securely into the walls of the channel rather thanwhen “teetering” on a specific area of narrowing Sincethese devices are designed to fit into a specific sizeddefect, when used in vascular channels they often cannotopen immediately to their designed, fully expanded size and/or shape In those circumstances, they initiallyassume unusual configurations within the vessels, butgenerally occlude the abnormal channel very completely.With the “memory” of the Nitinol™ metal in these devices,the devices eventually will achieve their designed con-figuration even within a constrained vessel!
Both the Amplatzer™ VSD and ASD devices have been used in a variety of abnormal vascular channels inpatients outside of the United States and in a few extenu-ating, compassionate circumstances in the US Now theAmplatzer™ muscular VSD device, the Amplatzer™PDA devices and the Amplatzer™ Plugs are available foruse in the United States This does give physicians aroundthe world including in the US, a better choice of occlusiondevices in order to be able to use the most appropriatedevice for the occlusion of unusual and/or large vascularchannels
Trang 27Atrial septal occluders for miscellaneous vascular
occlusions
Occasionally there are vascular channels that require
occlusion that are very large in diameter and relatively
short All of these lesions are large enough in diameter to
be hemodynamically significant, yet so short in length
that they will not accommodate any of the more common
and accepted vascular occlusion devices without
com-promising the adjacent vital blood channels that are
connected by the abnormal communication These very
large diameter channels include but are not limited to
the afferent vessels of pulmonary arteriovenous fistulae,
anomalous left superior vena cava draining into the left
atrium, and persistent azygos/hemiazygos veins
follow-ing “Glenn” type shunts In these cases, an Amplatzer™
ASD device (AGA Corp., Golden Valley, MN), an
Amplatzer™ PFO device (AGA Corp., Golden Valley,
MN), a Rashkind™ PDA device (USCI, Glens Falls, NY) a
CardioSEAL™ ASD device (NMT Medical Inc., Boston,
MA) and a Sideris Button™ device (Pediatric Cardiology
Custom Medical Devices, Athens, Greece) all can be of an
appropriate size and configuration to fit into, and all of
these devices have been used to occlude, these unusual
communications Some persistent aortic pulmonary
com-munications that were created surgically
(Waterston-Cooley or Potts shunts) are very short, they have high
flow, and they are present between very large,
high-pres-sure and very essential vessels As a consequence, they
require a very low-profile and sturdy device for occlusion
These devices and their delivery to their designed
lesions are described in detail in Chapters 28 (“ASD
Occlusion”) and Chapter 29 (“PFO, Fenestration, and
Baffle Occlusions”) In the large diameter, short and more
or less, “flat” lesions between two large, adjacent “tubes”
(great arteries), the positioning and fixation of the device
in the vessel are critical in order to prevent “movement” of
the device during or after implant The occlusion of the
short aortopulmonary vascular communications often
involves a more circuitous route to the lesion in order to
deliver a device, and the implant procedure usually
requires a very unusual and/or acute angle for the
deliv-ery of the device This makes the delivdeliv-ery to these unusual
locations more difficult and the release/implant of the
device more precarious Once the delivery sheath is in
position through the lesion while at the same time, not
kinking the sheath, the particular device extrusion and
implant is similar to the device delivery into an atrial
sep-tal defect/PFO
Caval filters or Cook “Spider”
Rarely there are very large diameter, abnormal, venous
channels and/or communications with very compliant
walls that require occlusion Because these veins distendvery easily and extensively, none of the previously de-scribed devices can be fixed securely in the vessels In thesevessels, the Spider™ device (Cook Inc., Bloomington, IN)
is used as a “scaffold” or framework for the implant ofother devices, particularly coils The Spider™ “springs”into a broad, spread-out configuration as it is pushed out
of its delivery catheter and attaches to the walls of the vessel by tiny, very sharp hooks in addition to the springtension of the arms of the device against the vessel wall.The Spider™ itself is not occlusive, but other devices such as coils are intertwined in the Spider™ to form anocclusive mass
Stents as a scafold for occlusion devices
Although designed for, and used primarily for, ing the patency of vessels, specially prepared intravascu-lar stents are used to assist in the occlusion of very largediameter channels/vessels14 The stents are used in amanner similar to the previously described Spider™ as a
maintain-“scaffold” for coils or other occlusions devices By fixingthe center of the stent at a restricted diameter by means
of a circumferential suture tied around the stent when
it is mounted on the balloon, a large diameter stent can
be expanded purposefully into an “hour-glass” shape.The suture is “woven” into the spaces between the struts
of the collapsed stent and tied around the stent in a cumference with a small fixed diameter before the stent
cir-is mounted on the delivery balloon The diameter of thesuture around the stent maintains the center of the stentexpanded only to the desired restricted diameter of thesuture When the stent is expanded on a large diameterballoon (15+ mm), the ends of the stent flare widely withthis “narrowing” at the center of the stent creating an
“hour-glass” configuration of the stent With the center
of the stent fixed at the narrow diameter by the suture, thetendency of the ends of the stent to extend and alignalmost perpendicular to the walls of the particular vessel/channel is exaggerated even more than usual The verysharp ends of each strut of particularly the J & J P _ _ 8 or
P _ _ 10 stents create sharp fixation points into the wall ofthe vessel This fixes the stent very securely in the vessel
A stent can be implanted using this technique and using
a P _ _ 10 stent, into vessels/channels up to 25 mm indiameter Once the stent is in place, then coils and/orocclusion devices can be intertwined into or packed intothe “afferent end” of the stent/vessel to achieve the occlu-sion of the vessel
Covered stents
Another mechanism with which intravascular stents areused to “occlude” vessels is by the use of “covered stents”
Trang 28to purposefully occlude side or branching vessels and/or
tears in the wall off a main channel/vessel The covered
stent will cover all areas that are adjacent to the stent
and prevent flow into those vessels and other abnormal
openings/tears This is a very selected use and requires
the availability of a covered stent of the exact diameter of
the torn central or main channel or vessel The covered
stents are available commercially only for adult
peri-pheral vascular use and/or in investigational studies For
use in the isolated congenital patient, commercially
manufactured covered stents have been available only for
compassionate use in very extenuating circumstances
More frequently the stents are modified extensively from
the commercially available covered stents and/or hand
fashioned using ultrathin expanded
polytetrafluoroethy-lene (ePTFE, Gore-Tex™) membranes (W L Gore &
Associates, Flagstaff, AZ) over existing intravascular
stents of appropriate size for the involved vessel15 The
ultrathin ePTFE material has a considerable capability of
stretching acutely so that each diameter of ePTFE tubing
can be used with several sizes of stents However, none of
the means of acquiring and producing covered stents for
congenital heart patients in the United States are rapid
enough or satisfactory for emergency use For such use,
the stent should be available immediately from the
inven-tory of the catheterization laborainven-tory for emergency use
Except in emergencies, the covered stents probably only
should be used in fully-grown patients since there is no
suggestion or information about the ability to re-dilate the
covering materials on the stents even months, much less
years, after implant and/or in the presence of significant
growth of the vessels At the same time, in a life-threatening
emergency, the use of a covered stent that was rapidly
available and of a size to fit even a small vessel would be
preferable to massive blood loss and exsanguination from
an acute vascular tear
Occlusion procedures for specific lesions
Systemic to pulmonary collaterals
The systemic to pulmonary collateral vessels originate
from the aorta or from major branches off the aorta They
usually are relatively long, tortuous, tubular lesions of
varying diameters throughout their course and usually
have one or more discrete narrowing as they course
distally into the pulmonary parenchyma Because of this
anatomy the occluding device usually does not have to
“fix” tightly against the vessel wall and, in fact, almost is
expected to move distally a small distance after release
The anatomy of these lesions makes them the ideal lesions
for occlusion with “free-release” Gianturco™ coils The
most difficult part of the occlusion of these lesions
frequently is the initial access into, and then obtaining a
secure position well within, the vessel The cannulation of
some of the collaterals often requires a great deal of gination as well as a large inventory of specific cathetersand/or deflector wires Very often, a more secure accessinto a collateral off the descending aorta is achieved whenthe catheter is approaching from the ascending aorta tothe descending aorta This approach to the collateral ves-sel can be accomplished through a co-existing ventricularseptal defect, through a transseptal atrial septal puncture
ima-in conjunction with the use of a floatima-ing balloon catheter,
or by introduction of the catheter into a brachial artery.The collateral vessels that come directly off the aortausually are entered directly with the catheter for the deliv-ery of the coil Those vessels arising more remotely off
a brachiocephalic branch of the aorta, those with a verytortuous course, and/or those where the occlusion must
be accomplished in a more circuitous and distal location
in the vessel, often require the use of a separate “guidecatheter” to engage the orifice of the main collateral ves-sel Then, a small, very flexible, or even “floating” smallerdelivery catheter such as the Target Tracker™ catheter(Target Therapeutics, Fremont, CA) or the Cook Slip-Cath™ (Cook Inc., Bloomington, IN) is passed throughthe guide catheter to deliver the smaller coils and maneu-vered to a more secure distal location into the vessel that is
to be occluded
The exact type and size of coil that is used depend upon the access to the vessel as well as the length and thediameter of the vessel The “free-release” Gianturco™ andTornado™ coils usually are ideal for the smaller vesselswith sufficient length to accommodate several “com-pacted” coils placed in series In the same size but shorterlength vessels or in vessels where multiple coils alreadyhave been implanted and are filling the vessel back to
or near the orifice of the entrance from the main vessel,
a controlled release coil or device is desirable Using a controlled release coil during the extrusion of the finalcoil, if the entire coil does not fit completely and securelywithin the vessel to be occluded, the detachable coil can bewithdrawn, repositioned, “packed” into the vessel, oreven completely and simply withdrawn
The 0.052″ coils (Cook Inc., Bloomington, IN), larly with a bioptome-controlled release system, are verygood for the occlusion of larger diameter collaterals
particu-in smaller patients As long as the vessel that is to beoccluded does not arise too acutely off the aorta, thesecoils can be delivered through as small as a 4-Frenchsheath The controlled release allows repositioning, oreven complete withdrawal of the coil if either the posi-tion and/or occlusion are not satisfactory In the larger patient with a large, longer collateral, the Gianturco-Grifka Vascular Occlusion Device™ (GGVOD™) (CookInc., Bloomington, IN) and the Amplatzer™ plug (AGA
Trang 29Medical Corp., Golden Valley, MN) are excellent choices
for these occlusions The use of the GGVOD™ and the
larger diameter plugs is dependent upon being able to use
an 8-French delivery sheath safely through an arterial
approach and to have the stiff sheath make the acute curve
from the aorta into the collateral
Surgical shunts
Occasionally, systemic to pulmonary shunts persist after
“corrective” surgery and require closure following a
pre-vious repair of the underlying congenital lesion Although
the “Blalock” and “modified Blalock” shunts are similar to
the collateral vessels in that they are tubular “vessels”, the
remaining characteristics of the shunts are very different
The shunt, which is created surgically, usually arises at
a very acute angle off the base of a brachiocephalic
vessel, which, in turn, makes access into the shunt from
the arterial approach very difficult The “modern” shunts
usually are made of synthetic materials that are very
non-compliant, they usually are uniform in diameter with no
tortuosity, no narrowing and no distal tapering, which
makes the fixation of an occlusion device within a Blalock
type shunt depend entirely upon the firm pressure of the
occluder against the walls of the shunt
The “free-release” Gianturco™ coil is still the most
commonly used device for transcatheter occlusion of
these lesions The larger, 0.038″ or preferably, 0.052″ coils
are used to ensure firm pressure against the walls of
the shunt Ancillary maneuvers frequently are employed
to assure the fixation of the coils in these shunts
Occasionally, there is a narrowing of the pulmonary artery
at the entrance/anastomosis site of the shunt into the
pul-monary artery In this circumstance and when the patient
is large enough to accommodate it, an intravascular stent
can be implanted in the pulmonary vessel in the area of
the stenosis, which will cross the exit orifice of the shunt.
This effectively “jails” the exit of the shunt and allows
coils to be packed into the shunt against the struts of the
stent, which are crossing the orifice of the shunt into the
pulmonary artery
When there is no stenosis of the shunt and/or no
steno-sis of the pulmonary artery (or the patient is not a
candid-ate for a pulmonary artery stent), an angioplasty type
balloon can be inflated in the pulmonary artery across the
distal end of the shunt for the purpose of temporarily
occluding flow through the shunt while the coils are being
implanted The balloon inflation is maintained as coils or
other devices are placed in the shunt The inflated balloon
occludes the distal end of the shunt and stops blood flow
through the shunt while coils are packed and intertwined
with each other within the shunt Unfortunately, this
tech-nique is not foolproof Unless the coils are packed very
tightly against the walls of the shunt and/or when the
coils are placed in a shunt with a very uniform diameterand/or no distal narrowing, the implanted coils still canembolize into the pulmonary artery when the balloon
is deflated
A sturdy, 0.038″ or even 0.052″ controlled release coil is
the preferred device for the residual surgical shunt whichhas no narrowing within its lumen The controlled releasecoil is implanted in the shunt and, before the coil is
released, tension on the attached delivery wire is relaxed.
In this way the coil is allowed to reposition itself withinthe shunt and usually, if it is not going to hold, the coil willpass through the shunt and to the pulmonary artery whilestill attached to the delivery, attach/release wire Whenthis occurs with the coil still attached, the coil can be with-drawn into the delivery catheter
Another alternative for occluding residual surgicalshunts is to deliver a much sturdier device into the shunt from the venous/pulmonary artery approach
A 0.052″ coil, a Gianturco-Grifka Bag™, or even a smallAmplatzer™ Vascular Plug can be delivered from thevenous approach The venous and pulmonary arteryapproach may have a straighter course to the shunt Thevenous introduction and straighter course allow the use of
a larger and/or stiffer delivery system
An end-hole catheter introduced into the venous system is manipulated into the pulmonary artery wherethe shunt enters, into the shunt from the pulmonary artery end of the shunt, through the shunt and into theaorta The catheter is replaced with an exchange length,long floppy tipped, Super Stiff™ wire The specific deliv-ery catheter/sheath for the device to be used is advancedover the wire through the venous system and into theshunt The dilator and wire are removed, the sheath iscleared of all air and/or clot and placed on a slow flush
Once the delivery catheter/sheath is in place through the shunt, the device is delivered to the site within the shunt
for implant/occlusion With the device maintained in theprecise position, the sheath is withdrawn, which, in turn,deploys the device in the shunt The degree of occlusion isverified with an angiogram, injecting into the arterial end
of the shunt through a separate catheter
Occasionally, entry with any catheter or wire into theshunt from the pulmonary artery end is very difficult
or even impossible In that circumstance, an exchangelength, floppy tipped, 0.035″ wire is advanced through anend-hole catheter, which is passed retrograde from theartery, through the shunt and into the pulmonary artery.The end of the wire is snared in the pulmonary artery with
a snare introduced from a vein and advanced through theright heart to the pulmonary artery Once snared, the wire
is withdrawn through the right ventricle, right atrium andout of the venous system, creating a through-and-throughwire, which passes through the shunt The delivery system for the occlusion device can then be advanced over
Trang 30the wire from the venous approach and into the shunt for
venous delivery of the desired device
Short, large-diameter aorto-pulmonary
communications
Congenital aorto-pulmonary (A-P) windows and
resid-ual, direct aortic to pulmonary artery (Cooley–Waterston
or Potts) shunts can be occluded with devices in the
catheterization laboratory Coils are unsatisfactory for
these lesions because of the very short, broad, “window”
nature of these communications The most satisfactory
device for these “window-like” shunts is one of the
“double umbrellas” The double umbrella (Rashkind™
PDA, Clamshell™ and CardioSEAL™ ASD) devices, the
Sideris double umbrella modification of the Inverted
Button™, and a modified Amplatzer™ ASD device all
have been used to occlude this type of lesion16 –18
A-P window type communications are occluded from
either the arterial or venous approach The choice of
approach depends upon the vascular access for the
particu-lar approach and which approach gives the best “angle”
from the vessel into the lesion for the implant of the device
and, in turn, the best control over the particular delivery
system Because these communications essentially are
part of, and parallel to, the walls of the adjacent vessels
and the vessels have relatively small diameters compared
to a cardiac chamber, there usually is no good angle for the
delivery of any one of these devices! The devices initially
used for these lesions required an 11-French delivery
sheath, however, the Amplatzer™ PFO device potentially
can be delivered through a smaller sheath Because of the
sheath size, the venous approach is preferred when all of
the other factors in the choice in the approach are
other-wise equal
The congenital or residual surgical A-P communication
is identified and measured accurately by echocardiogram
and by selective angiograms with injection immediately
adjacent to the systemic arterial side of the lesion and/or
actually within the communication The X-ray tubes are
angled so that the communication is “cut on edge” at least
in one plane Once identified and measured, the defect is
crossed with an end-hole catheter Preferably, a
through-and-through, artery to vein wire “rail” is created with
an exchange length wire The catheter/wire can be
intro-duced from either the arterial or venous approach and
then captured with a snare catheter, which is introduced
from the opposite vascular access When there is any
question about the size and/or configuration of the defect,
the communication can be sized and “shaped” using a
NuMED™ sizing balloon (NuMED Inc., Hopkinton, NY)
advanced over the wire It is advantageous, if not
essent-ial, to have a second systemic arterial catheter to be used
for “positioning” angiograms during the procedure
For occlusion of the “window type” communicationwith any of these devices, a long delivery sheath first is
positioned through the defect with great care taken not
to kink the sheath because of the unfavorable anglesinvolved The through-and-through wire facilitates thedelivery considerably of the delivery sheath/dilatorthrough the communication Once the delivery sheath isthrough the defect and the wire and dilator are removed,the “umbrella” type devices are delivered exactly as for the occlusion of a PDA with the old Rashkind™ PDA device or the occlusion of a fenestration with aCardioSEAL™ device The distal umbrella/disk isopened on the “distal” side of the defect The sheath, thedelivery system and the device are withdrawn as a unit,pulling the open umbrella/disk tightly against the defect.The device may have to be pulled very firmly against thedefect, in which case, the device becomes markedly dis-torted, very similar to the delivery of an umbrella/diskduring the closure of a fenestration, particularly wherethere is a long communication Usually, the vessel wallssurrounding and adjacent to A-P communications arefairly rigid and the defects are small, so that a marked dis-tortion of the device from the strong “pull” on the deviceagainst the vessel wall is possible during the deliverywithout pulling the device through the defect When thecenter “hinge” or “hub” of the device is within and/orthrough the defect and the proximal legs/disk still arewithin the delivery sheath but completely through thedefect, the sheath alone is withdrawn, opening the proximalumbrella/disk on the more “proximal” side of the defect.After angiographic confirmation with an injection on thesystemic arterial side of the lesion, the device is released.The device often reorients itself markedly after releasefrom the torsion of an acutely angled delivery system
Coronary cameral fistulae
Closure of most coronary-cameral fistulae with a delivered device in the cardiac catheterization laboratorynow is the standard approach for these lesions The fistu-lae arise off either the right, left and/or any branch orcombination of branches off the coronary arterial system.They usually have a fairly long and often tortuous courseand empty into a right-sided cardiac chamber or the pul-monary artery They rarely empty into the left atrium oreven the left ventricle The choice of device and deliverytechnique depends upon the specific anatomy of the fistu-lae, the length of the fistulous tract and the distance awayfrom the true and essential coronary arteries Test occlu-sion of the fistula with a balloon is recommended in allcases to help to define the anatomy and distensibility ofthe fistulous tract, and to determine if there is any myocar-dial tissue dependent on the blood supply off the fistula.This is tested by the inflation of a small Swan™ balloon in
Trang 31catheter-the fistula tract while observing catheter-the electrocardiogram
very closely for any ST-T wave changes
A retrograde, end-hole arterial catheter is introduced
into the “feeding” coronary artery and maneuvered
into the fistula as far as possible using torque-controlled
and/or Glide™ wires as necessary The fistula is defined
angiographically with selective coronary angiograms into
the feeding coronary artery or, preferably, into the orifice
of the fistula itself Multiple views are obtained to define
the course, branchings, narrowing and the site of
empty-ing of the fistula Often there is a narrowempty-ing and/or an
acute tortuosity in the fistula tract, which will help to fix
an occlusion device
An exchange guide wire is passed far into, and if
pos-sible, completely through the fistula The original catheter
is replaced over the wire with a small Swan™ end-hole,
balloon catheter or a very short, small balloon angioplasty
catheter, which is to be used for sizing and test occlusion
Several (multiple) low-pressure test inflations at different
locations along the fistulous tract are performed with the
balloon It is useful to have a second arterial catheter
already in position in the aortic root or even in the
select-ive coronary artery in order to perform small selectselect-ive
angiograms in the aortic root and/or in the coronary
artery proximal to the fistula and the site of test occlusion
In addition to the degree of occlusion of the fistula, this
injection demonstrates additional tracts off the fistula
and any significantly under-perfused areas of the
myo-cardium distal to the balloon occlusion
When the fistula is to be closed from the retrograde
arter-ial approach, the balloon catheter in the fistula is replaced
over the exchange wire with an end-hole only,
torque-controlled catheter The end-hole catheter is passed as far
distally in the fistula as is possible A “free-release” type
coil is satisfactory for coronary-cameral fistulae only when
(1) there is no myocardial tissue dependent on the flow
from the proximal fistula, (2) there is a long tract of the
fistula that is distal to the origin of the fistula off the true
coronary system and (3) the delivery catheter can be
manipulated reasonably easily and far distally into the
fistula With lesions that fulfill these criteria, the coils
usu-ally are delivered from the retrograde approach through a
catheter maneuvered selectively into the fistula from the
feeding coronary artery
When delivered retrograde, the occlusion device is
implanted as far distally in the fistulous tract and as far
away from the orifice of the fistula off the true coronary
artery as possible The distal location keeps the coil away
from the true coronary vessel and allows the deposit of
additional coils more proximally in the fistula Once the
delivery catheter is in position for the coil delivery, a
select-ive angiogram is recorded in the fistula through this
catheter to verify the precise location A coil for occlusion
of the fistula is chosen which is 1–2 mm larger in diameter
than the “balloon sized” diameter of the lumen of thefistula The coil must expand against the walls of thefistula in order to fix the coil in place in a non-stenotic loca-tion In a fistula with a significant narrowing and/or anarea of acute tortuosity more distally in the fistula, a coil is
chosen just larger than the diameter proximal to the
nar-rowing or to straddle a specific narnar-rowing similar to theretrograde implant of a coil into the “narrowing” of thePDA How the coil is delivered into the fistula dependsupon the precise anatomy, and is different in each case
At least 2–3 minutes after the first coil is implanted, thedegree of occlusion is tested with a repeat selective injec-tion into the fistula through the coil delivery catheter.Frequently there are residual leaks through one, or evenseveral, coils in a coronary-cameral fistula Leaving asmall residual leak through the fistulae, particularly withthe coils in place, increases the patient’s likelihood ofdeveloping endarteritis in the future, and when there is a
“jet” of residual leak through the coils, there also is a highchance of creating hemolysis
Several (or multiple) more coils are deposited in thefistula sequentially until the flow through the fistula isstopped The choice of the additional coils depends uponthe residual anatomy and the remaining proximal length
of the fistula Additional coils are intertwined with the initial (earlier) coils to prevent their movement backwardtoward the coronary artery When the most proximal coilswill be close to the entrance of the fistula from the cor-
onary artery, a controlled release coil and technique are used
for the final coil(s)
The prograde venous (retrograde into the fistula)approach for the transcatheter closure of some coronary-cameral fistulae has several distinct advantages over theretrograde arterial approach Fewer catheter manipula-tions are required in and/or through the somewhat pre-carious true coronary arteries and the central arterialsystem in general Larger and stiffer (less bendable) deliv-ery systems can be introduced from the venous approachthan from the retrograde arterial approach This extendsthe options to a larger variety of occlusion devices that can be used to occlude coronary-cameral fistulae Fromthe venous approach, the delivery sheaths for the large 0.052″ coils and Duct-Occlud™ device as well as the largersheaths for the Gianturco-Grifka™ Bag, the Amplatzer™Vascular Plug or the Amplatzer™ muscular VSD occlud-ers can be positioned in the fistula without kinking the sheaths and without compromising the systemic arter-ial entrance site and/or the proximal coronary arteriesthemselves
In order to establish the venous access to the fistula, italmost always is necessary first to create an arterial tovenous, through-and-through wire The catheter/wire isintroduced from the arterial approach, into the coronaryartery, through the fistula and into the venous circulation,
Trang 32where the wire is snared with a catheter introduced from a
vein The introduction site for the venous catheter should
be from the peripheral vein (femoral or jugular) that will
give the best angle of approach into the distal end of the
fistula An end-hole retrograde catheter is introduced into
the feeding coronary artery and from there, manipulated
into the origin of the fistula A torque-controlled, floppy
tipped or curved tip Glide™ wire is maneuvered through
the fistula and into the “exit chamber” of the fistula within
the heart The distal end of the wire is snared at the
“venous” end, where it exits the fistula and is exteriorized
through the venous entry site
The retrograde catheter should remain over the arterial
end of the wire while the delivery catheter or sheath/
dilator is advanced from the venous end of the
through-and-through wire and into the fistula The arterial catheter
over the wire protects the tissues within the coronary
artery from damage by the bare wire and allows a
contin-uous flush around the wire to prevent clotting in the
cor-onary system The tip of the delivery catheter/sheath is
positioned according to both the anatomy of the fistula
and what type of device is being used In general, the tip
of the delivery catheter/sheath is positioned just
prox-imal (in the direction of blood flow in the fistula) to the
location where the device(s) is(are) to be implanted Once
the delivery catheter/sheath is securely in position, the
through-and-through wire is withdrawn through the
arter-ial catheter and the catheter is cleared carefully of any air
and/or clot The retrograde arterial catheter remains
posi-tioned in the coronary artery in the end of the fistula or at
least in the coronary artery just proximal to the fistula The
retrograde catheter is used to perform sequential selective
angiograms during the positioning and delivery of the
device(s) and/or to add additional occlusion devices after
the implant of the original device
An occlusion device/coil is chosen which is appropriate
for the particular anatomy of the coronary fistula When
delivered from the venous approach the choice of type
and size of device is much greater, allowing the delivery
of very retrievable devices such as the Amplatzer™
vas-cular plug and the GGVOD™ devices When delivered
from the venous approach, the device is implanted in the
fistula straddling an area of narrowing or proximal (in the
direction of blood flow) to any areas of narrowing and/or
sharp bends in the fistulous tract, but at the same time
distal to any true coronary artery branches off the fistula
The one disadvantage to the venous approach is that
sub-sequent devices that are delivered through the same
delivery system, must be implanted “downstream” or
dis-tal in the flow from the original device and often into a
larger diameter area of the fistula than the site of the
original device The “downstream” devices, which are
predominately coils, may not fix in the vessel as securely,
and complete occlusion is less likely Any subsequent
device and, in particular, coils delivered in a stream” location must be significantly larger than thediameter of the fistula in that location and must be “inter-twined” and attached to the original device
“down-It always is preferable to implant additional coils/devices proximal to the original device If additional coils
are to be delivered from the venous side and are to be
placed proximal to the first device, a delivery catheter
must be advanced through and/or past the original device and
into the more proximal fistula without dislodging theoriginal device/coil This is accomplished using a TerumoGlide™ (Boston Scientific, Natick, MA) wire and a tiny,end-hole-only catheter similar to crossing a freshlyimplanted coil in a PDA As the coil is being extrudedfrom the catheter, which has passed through/past theoriginal device/coil, the catheter is withdrawn into, oradjacent to, the original device/coil in order to catch orentwine the additional coil(s) on the original device/coil
A preferable technique for the implant of additionalcoils proximal to the original device, which was deliveredfrom the venous approach, and particularly when the firstdevice is implanted distally in the fistula away from theentrance of the fistula off the coronary artery, is to deliver
additional coils “upstream” through the retrograde catheter
that is already positioned in the proximal part of the
fistula Using a controlled-release system additional coils
are “stacked on top of ” the original device to plete the occlusion Often, with a very large coronary-cameral fistula, this combination of venous and arterialapproaches is the planned procedure
com-In certain circumstances during the catheter occlusion
of coronary-cameral fistulae it is imperative to use a controlled-release device for the delivery of the occlusiondevice The presence of a particularly short fistulous tractand particularly one with a wide orifice off the true coronary system necessitates a controlled-release device
A fistula of any length into which multiple coils areimplanted and when the next, and most proximal, coilmust be implanted near the true coronary artery, is theother definite indication for a controlled-release system
In these circumstances, the coil and/or any other type ofdevice must be retrievable immediately in the event thatthe device or part of the device extends back into thelumen of the true coronary artery during the delivery ofthe device The tip of a coil with its attached filamentstrands dangling into the lumen of a true coronary artery
is very likely to shed thrombotic emboli into the high flow
in the coronary artery and can result in infarction ofmyocardium distal to that area
Pulmonary arteriovenous fistulae
Pulmonary arteriovenous (AV) fistulae can be congenital
or “acquired”, they can be isolated or multiple, they can
Trang 33occur in thousands and they can be large or small These
lesions can be occluded with catheter-delivered devices.
How they are occluded and with what particular device
are determined by the exact anatomy of the lesion(s) Each
pulmonary AV “fistula” may have several afferent and
efferent vessels The central communicating vessels usually
are small, tortuous and restrictive, but can be relatively
large and even form an almost straight communication
from the pulmonary artery to the pulmonary vein
The rare, isolated, congenital pulmonary AV fistula
usually is large, with multiple and usually tortuous
affer-ent pulmonary arterial and efferaffer-ent pulmonary venous
channels The size(s) of the major central AV
communica-tion(s) through the fistula is(are) determined by detailed
angiography An occluding device significantly larger
than the major central communicating AV channel must
be used initially to prevent the occlusion device from
passing through the fistula and into the pulmonary
vein/systemic circulation The goal is to occlude the major
afferent channel(s) while, at the same time, occluding no,
or as few as possible, normal pulmonary arterial vessels,
which supply adjacent normal lung tissue In the presence
of multiple channels, test occlusions of the various
affer-ent channels using a balloon angiographic catheter or a
Swan™ balloon catheter is performed to determine the
largest and/or most significant feeding vessel(s) In most
cases, the actual communication between the artery and
vein is relatively small and one or many standard
“free-release” Gianturco™ coils can be used for the occlusion
When the communication is large, the large 0.052″ coils,
the Gianturco-Grifka™ bag, the Amplatzer™ PDA, VSD
devices or vascular Plug or even an “umbrella” occluding
device all can be used to at least begin the occlusion in
these lesions Once a larger device is in place, standard
Gianturco™ coils can be packed proximally to the larger
device in the afferent channel to complete the occlusion of
that particular vessel/segment
Biplane fluoroscopic and angiographic systems are
absolutely essential for the treatment of these lesions The
branching pulmonary arteries are very “three-dimensional”
and arise at multiple angles to, and at different planes
from, each other The branches of the arteries to be
occluded must be cannulated selectively, and then
sub-selectively A single-plane system does not provide the
depth relationships for this precise cannulation Probing
with a single-plane system results in exponential increases
in the radiation use/exposure and of the amount of
contrast used for localizing and cannulating any specific
lesion
It is desirable to have a second venous angiographic
catheter, which is positioned selectively in the pulmonary
artery and immediately adjacent to the delivery catheter
for the occlusion device Once the precise afferent vessel
has been identified, it is cannulated selectively with an
end-hole catheter The catheter is replaced with anexchange length, short tipped, 0.035″ Super Stiff™ wire.The delivery dilator/sheath set with a sheath appropriatefor the device to be used is advanced over the wire into
the major afferent vessel and as close to the fistulous munication as possible Multiple normal pulmonary artery
com-branches, which supply vital normal lung tissue, usuallyarise off the major afferent pulmonary artery proximal tothe fistula The more proximally the occlusion is per-formed in the afferent pulmonary artery that feeds thefistula, the greater the number of vital vessels that supplythe normal adjacent functional lung parenchyma will
be occluded
The occlusion device is introduced into the preciselypositioned sheath and extruded as accurately as is pos-sible into the exact location and released Before release ofthe device, an angiogram is performed through the sec-ond venous catheter to determine the effectiveness of thefirst occlusion device and to determine the location(s) ofthe next most important afferent vessel(s) With residualflow through the fistula from this artery, either a secondlarge device is implanted through the same deliverycatheter or a coil delivery catheter is introduced throughthe delivery sheath and the occlusion is completed withsupplemental coils packed proximal to the original device
in the afferent vessel of the fistula
Once the first feeding artery is occluded, the deliverysheath/catheter is repositioned into the next most import-ant feeding vessel and the procedure repeated Once theinitial, larger vessels are occluded, the remaining vesselsmay be occluded more satisfactorily with a differentand/or smaller device In that case the original deliverysheath/catheter is exchanged for the delivery sheath/catheter for the different device or the new deliverycatheter is advanced to the lesion through the originaldelivery sheath/catheter The procedure is repeated untilthe individual fistula is occluded Each time an occlusiondevice is implanted, it is important to perform the implant
as close to the lesion as possible
In the presence of multiple pulmonary arteriovenousfistulae, patients often are very cyanotic and very symp-tomatic At the same time, there usually are too many sep-arate pulmonary AV fistulae to consider occluding all ofthem In the very symptomatic patient, the largest andmost important AV lesions are identified with detailedangiography and an attempt is made to occlude as many
of the major fistulae as possible without further mising the remaining normal lung tissue This, of course,does not cure the patient, but often, enough of the lesionscan be occluded to provide a significant, albeit temporary,symptomatic palliation for the patient The technique isthe same for multiple fistulae as for an isolated largefistula The exact device and the technique used are deter-mined from the selective angiographic anatomy of the
Trang 34compro-individual lesions With each successful occlusion of one
or more significant lesions, the patient’s systemic arterial
saturation rises slightly If this does not occur after the
occlusion of three or four more significant fistulae, the
patient probably will not receive benefit from further
occlusions
The multiple diffuse (thousands) of tiny pulmonary
arteriovenous fistulae seen frequently after isolated
sup-erior caval to pulmonary artery anastomoses are not
directly amenable to occlusion therapy The exception is
when there are a few separate large fistulae associated
with the other multiple diffuse tiny fistulae These
larger systemic to pulmonary fistulae can, and should, be
occluded Even with the multiple small fistulae, occlusion
of associated larger pulmonary fistulae provides some
symptomatic relief for the patient
The approach to the multiple tiny pulmonary
arterio-venous fistulae is to anticipate their development, and in
turn, prevent their occurrence by proceeding with the
completion of the connection of the inferior vena cava so
that hepatic blood flow reaches both lungs Even after these
multiple tiny pulmonary A-V fistulae have appeared, the
patient’s systemic saturations seem to improve with the
completion of the inferior caval connection to both lungs,
even though the multiple tiny fistulae do not, objectively,
seem to regress
Systemic artery to pulmonary artery and
veno-venous collaterals post “Fontan corrections”
The complete caval-pulmonary repair of single ventricle
lesions has created a whole new spectrum of systemic
artery to pulmonary artery and systemic vein to
pul-monary vein collaterals/fistulae These are usually small
but can be multiple and/or very large, especially some of
the hepatic vein to pulmonary venous fistulae Regardless
of the size of an individual fistula, each fistula provides
“competition” to the flow to the already precarious,
“for-ward” pulmonary blood flow, produces extra volume on
the “single ventricle” and/or creates a “right to left” shunt
with systemic desaturation of the patient These abnormal
vascular communications usually are not obvious from
clinical or noninvasive studies They are detected only
during a cardiac catheterization and then, only when they
are looked for specifically with selective angiography
These abnormal fistulae arise from any systemic artery or
vein within the thorax, neck or even from the abdomen
The arteries communicate with the pulmonary arteries,
the pulmonary veins and/or even the systemic veins The
systemic venous channels communicate with the
pul-monary veins and/or with other systemic veins draining
to an entirely different systemic venous bed The arterial
communications usually are small tortuous channels
but on aortography or systemic venous angiography, the
pulmonary arteries and/or veins light up with contrastlong before the normal “recirculation” from the pul-monary arteries has time to reach the pulmonary veins.When a catheterization is performed for any reason onany of these single ventricle, “Fontan” type patients, theselesions should be looked for specifically while the patient
is in the catheterization laboratory and when found,occluded
The major challenge in occluding these lesions is ing and then cannulating their origins off the systemicarterial or systemic venous systems Once located andcannulated, they usually are occluded quite simply withfree-release coils or other available occlusion devices Thesmaller of these lesions are ideal for the small Tornado™(Cook Inc., Bloomington, IN) or Target™ (Target Thera-peutics, Fremont, CA) coils and delivery systems Thehepatic vein to pulmonary venous fistulae usually aremuch larger and require one or more of the larger occlud-ing systems Fortunately, access to these hepatic to pul-monary venous channels is usually directly from thecentral hepatic veins, allowing introduction of the larger,stiffer delivery systems for the large coils, Gianturco-Grifka™ bags (Cook Inc., Bloomington, IN) or one of theAmplatzer™ occlusion devices (AGA Medical Corp.,Golden Valley, MN)
locat-In 2004, most of the abnormal vascular communicationswere treated in the catheterization laboratory with theavailable devices even in the United States With the addition of the devices that are on the horizon in protocolstudies in the US and/or routinely available in the rest ofthe world, even more of these lesions will be amenable tocatheter therapy The catheter management of all abnor-mal vascular communications will be safer and morewidely distributed with the introduction of a wider choice
of devices that are more appropriate for specific lesions
Occlusion of aortic root/coronary sinus to left ventricular tunnels
Congenital tunnels from the aorta to the left ventricle arerare fistulous tracts, which usually present in infancy Thecombination of the clinical findings of a nondescript systolic murmur, which is followed immediately by a longdiastolic, decrescendo murmur over the base of the heartand in association with a widened pulse pressure areindistinguishable from the findings of aortic valve regur-gitation The diagnosis of an aortic to left ventricular tun-nel is made by echocardiogram and confirmed with anaortic root angiocardiogram A biplane angiogram is nec-essary to determine the precise origin, the course and theexit of the tunnel and to obtain accurate measurements
of the diameter of various areas of the tunnel Generally,these lesions have a fairly straight origin off the aorticroot, are some distance from the aortic valve and coronary
Trang 35arteries and, as a consequence, are very amenable to
occlusion with a catheter-delivered device
Once the tunnel has been identified angiographically,
it is cannulated selectively with a floppy tipped spring
guide wire and then a catheter over the wire Selective
angiograms are recorded with injections directly into the
tunnel in order to define the anatomy of the tunnel more
precisely The X-ray tubes are angled in order to align the
tunnel on edge and in its longest dimension If there is any
question about the anatomy and/or diameters of the
tun-nel, the guide wire is passed through the tunnel into the
left ventricle and a Swan™ floating balloon catheter is
advanced over the wire into the tunnel The balloon is
inflated very gently and at a low pressure within the
tun-nel and, if the tuntun-nel has any significant length, in several
different locations along the course of the tunnel The
bal-loon inflation in the tunnel provides information about
the distensibility of the fistulous tract and how the patient
will tolerate an occlusive device in the tunnel The
com-bined angiographic and “sizing” information determines
the most appropriate device for occlusion of the tunnel A
static sizing balloon would give the same or better
infor-mation, but would require the use of a 9-French sheath in
the artery!
The tunnels usually are large relative to the patient’s
size and all of the tunnels from aorta to left ventricle have
a high-velocity flow through them This requires the use
of a sturdy occlusion device, which will fix securely in
the high-pressure, high-flow tract Because of the usual
small size of the patient, usually either a
controlled-release, large coil and/or a small Amplatzer™ PDA or
Vascular Plug is/are used A coil with a
bioptome-controlled release (in the US) requires a 5-French long
sheath for true controllability and the smallest sizes of
both the Amplatzer™ PDA and Vascular Plug devices
(4 mm) pass through 5-French sheaths The Detachable™
coils, available in the rest of the world, utilize a 4-French
delivery catheter, but these coils often are not robust
enough for these lesions In the rare larger patients, larger
devices including the GGVOD™ can be considered for the
occlusion of these tunnels
A wire is maneuvered from the retrograde catheter,
through the tunnel and into the left ventricle The
retro-grade catheter is replaced over this wire with the delivery
sheath/dilator or catheter that is to be used for the device
chosen for the particular tunnel The delivery sheath is
advanced through the tunnel, to the left ventricle or, at
least, to the distal end of the tunnel Once the delivery
sheath is positioned securely in the tunnel, the wire and
the dilator are removed from the delivery sheath The
sheath is cleared passively and meticulously of all air
and/or clot as with all other delivery systems particularly
in the systemic circulation The appropriate device is
loaded into, and advanced to the end of the sheath within
the tunnel The sheath and occlusion device are drawn together until the device, which is still within thesheath, is positioned in the desired location for implant
with-in the tunnel If the device bewith-ing used is a coil, the coil
is extruded from the sheath into the tunnel until theattach/release mechanism of the coil is just within the tip of the sheath With the other devices, the deliverycatheter/cable and device are fixed in position and thesheath withdrawn off the device until the device is fullydeployed in the tunnel
The fixation of the occlusion device is tested by fairlyvigorous to-and-fro motion of the delivery system, and
a small, hand injected angiogram is performed with aninjection of contrast through the delivery sheath Thedelivery sheath still is over the delivery catheter/cablewith the tip of the sheath just proximal to the devicewithin the tunnel and even a slow, small injection demon-strates the position of the occlusion device and the degree
of occlusion of the tunnelaparticularly if the tunnel is
occluded satisfactorily When satisfied with the fixationand the degree of occlusion, the release mechanism of theparticular device is activated By using a device with acontrolled-release system, if the fixation of the device andocclusion are not satisfactory, the device readily can bewithdrawn into the delivery system and the procedurerestarted with a more appropriate size or type of device
mul-of the “sewing ring” mul-of the prosthetic valve, which usuallyplaces the fistulous communication away from the func-tioning “leaflets” of the prosthetic valve
The aortic prosthetic perivalvular leak cally is similar to a coronary sinus to left ventricular fistula(previously in this chapter), although perivalvular leaksfrequently are multiple and usually occur in significantlylarger and older patients The diagnosis of, and the number of aortic perivalvular leak(s) are suggested bytransthoracic echocardiography and/or from biplane,aortic root aortograms The treatment of perivalvularleaks is monitored in the catheterization laboratory with
hemodynami-both selective biplane angiograms, which are performed
actually within the perivalvular leak, and by esophageal echo (TEE) The aortic perivalvular leak isaccessed with a retrograde catheter from the aorta above the prosthetic valve The orifice of the perivalvu-lar leak is usually located directly off the ascending aorta adjacent to the “ring” of the prosthetic aortic valve
Trang 36trans-The location(s) of the aortic perivalvular leak(s) is(are)
identified with TEE
Once an aortic perivalvular leak is suspected and then
identified, the leak is cannulated selectively with a
retro-grade, end-and-side hole catheter The exact location of the
aortic opening of the perivalvular leak in relation to the
aortic valve prosthesis, its shape, its maximum and minimal
diameters, its length and its exit site into the left ventricle
all are defined very accurately with selective biplane
angiograms with injection directly into the fistulous tract,
with the X-ray tubes angled to elongate the tract
max-imally The perivalvular leaks often are short and have
an elliptical shape adjacent to the rigid, prosthetic valve
“sewing ring” Once the precise anatomy and size are
defined angiographically and correlated with the TEE
findings, an occlusion device is chosen which will occlude
the perivalvular leak, but, at the same time will not extend
out of the orifice of the fistula and interfere with the
func-tion of the adjacent prosthetic valve
An exchange wire is placed through the retrograde
catheter, into and through the tract of the perivalvular
leak and into the left ventricle The original catheter is
replaced over the exchange wire with the specific delivery
system for the device to be used and the delivery system is
advanced deep into the tract of the perivalvular leak The
tract of the perivalvular leak and the delivery system
are interrogated continuously with the TEE It is useful, if
not absolutely necessary, to introduce a second,
retro-grade catheter into the aortic root in order to perform
small, repeated, selective angiograms during the actual
delivery of the device, which is being performed through
the other retrograde catheter/delivery system
When the fistulous tract has any length and has any
constriction within its course, a controlled-release, large
coil often is the ideal device for these lesions, but only
when the coil can be implanted completely within the tract
so that none of the attached coil fibers extend out of the
tract and into the area of the moving leaflets of the
prosthesis
For the short, broader perivalvular leaks, the smaller
diameter Amplatzer™ PDA, the Amplatzer™ muscular
VSD occluders or the newer Amplatzer™ Vascular Plug
are used The choice of which of these devices to use
depends upon the size of the proximal orifice and the
presence of any narrowing along the course of the leak In
spite of the usual elliptical shape of the orifice of these
perivalvular leaks, the Amplatzer™ devices, when
ini-tially implanted, often conform to the shape of the tract
The Amplatzer™ devices are placed either at the aortic
orifice of the leak or within the tract of the leak utilizing
the “ridge” created by the valve ring as the “central
nar-rowing” to straddle with the “waist” of the device The
exact location where the occlusion device is implanted
within a perivalvular leak depends upon the length,
diameter and configuration of the perivalvular leak, theproximity to the prosthetic leaflets and, often, a trial anderror deployment of the device The “lips” or “rims” of theAmplatzer™ PDA and muscular VSD devices usually areshort enough that they do not interfere with the leaflets ofthe prosthesis, even when the “rim” of the device is out-side of the orifice of the fistulous tract
Once a coil or one of the Amplatzer™ occluders hasbeen delivered into the tract, but before its release, arepeat biplane angiogram in, or very close to, the aorticorifice of the leak is performed to visualize the tract, tovisualize the degree of occlusion, and to check the position
of the coil/device in relation to the movement of the thetic valve leaflets The area is interrogated with theTEE/Doppler™ to identify the distance of the particularcoil/device away from the leaflets of the prosthetic valveand to check for residual leak through the tract When theocclusion is satisfactory and there is no interference withthe function of the prosthetic valve as seen on the TEE, thecoil/device is released The angiogram and the TEE arerepeated after the release of the coil/device to confirm the degree of occlusion, changes in the position of thecoil/device, the freedom from interference with the pros-thetic valve apparatus, and/or the presence of any addi-tional perivalvular leaks
pros-Any residual and/or additional perivalvular leak isaddressed during the same catheterization procedure.Residual leaks through devices can result in hemolysisand, if not closed during the initial implant procedure,may have to be closed later as an emergency In addition,when performing additional occlusions during the sameprocedure, all of the anatomy is fresh in the operator’smind and the wires and catheters already are in place forthe implant of an additional occluder, even though it may
be a different type from the original occluder
Small PDA “double umbrella” occlusion devices, which could be implanted completely within the tract ofperivalvular leaks, were used in the past for the occlusion
of perivalvular aortic leaks, however, all of the doubleumbrella devices require a much larger and stiffer deliv-ery system As a consequence, the double umbrelladevices on a rigid frame no longer are used for the occlu-sion of aortic perivalvular leaks
Perivalvular leaks around a prosthetic mitral valve resent significantly greater challenges for transcatheterocclusion Mitral perivalvular leaks frequently are multi-ple and they are more difficult to image precisely by eitherecho or angiography Both prograde access through atransseptal atrial septal puncture to the left atrium andretrograde access from the left ventricle are used for thecatheter occlusions of mitral perivalvular leaks Devices
rep-in perivalvular mitral valve leaks are delivered andimplanted from the left atrial (outlet) end of the leak,which is against the direction of the flow of the blood
Trang 37through the mitral perivalvular leak As with the
occlu-sion of the aortic perivalvular leaks, the delivery and the
implant of occlusion devices for mitral perivalvular leaks
are visualized and controlled by both selective biplane
angiography and TEE/Doppler
Mitral perivalvular leaks initially are identified by
transthoracic echocardiography and biplane, left
ventric-ular angiography Once the perivalvventric-ular mitral leak is
defined angiographically and a decision is made to
perform a transcatheter occlusion, a pre-curved,
end-and-side hole, retrograde catheter is positioned in the left
ventricle, a prograde, long, transseptal sheath is
posi-tioned in the left atrium (usually by transseptal atrial
puncture), and a TEE probe is introduced to visualize the
mitral annulus The specific leak(s) is(are) identified, then
they are defined more definitively by TEE/Doppler™
interrogation and more selective biplane angiograms
with the injections preferably performed directly into the
perivalvular tract from either a prograde or retrograde
approach Selective views of the angiograms in the left
ventricle and/or selective views in the tract of the leak
are displayed as “road maps” of the overall anatomy of
the area
A Judkins™ left coronary catheter or another “J”
shaped catheter is introduced retrograde into the left
vent-ricle The 180° curve on the pre-curved, retrograde, left
ventricular catheter helps to direct the tip of the catheter
toward the mitral annulus or, more specifically, toward
the left ventricular (entrance) end or into the left
ventricu-lar “entrance” of the perivalvuventricu-lar leak(s) in order to obtain
more “selective” injections
Using the angiographic “road maps” and TEE, from the
venous approach, a separate 6- or 7-French, end-and-side
hole torque-controlled catheter with a fixed slight angle
at the tip is maneuvered, usually along with a
torque-controlled guide wire, into the left atrial (exit) end of the
perivalvular leak In order to prevent extensive,
unneces-sary and “blind” maneuvering, biplane fluoroscopy is
absolutely essential for this maneuvering in order to
direct the catheter purposefully both from side to side and
from front to back The catheter is advanced well into, or
through, the fistulous tract The X-ray tubes are angled as
perpendicular to the long axis of the catheter (and tract)
as possible and to place the catheter, which is positioned
in the tract, at one “edge” of the valve ring as seen in at
least one view A high-pressure, Tuohy™ “Y” adaptor is
attached over the wire to the hub of the catheter and a
biplane angiogram recorded while injecting through this
catheter, over the wire and directly into the tract of the
perivalvular leak The biplane angiograms define the
diameter, shape and length of the tract and demonstrate
the distance of the proximal and distal openings of the
tract from the prosthetic valve leaflets The diameters,
lengths and distances are measured very accurately on
both the angiogram and on the TEE images The properdevice to occlude the perivalvular leak is chosen on thebasis of these angiographic measurements and the cor-responding TEE images The original torque wire, whichstill is through the catheter, is replaced with a long, pre-formed, floppy tipped Super Stiff™ (Boston Scientific,Natick, MA) wire, which is looped in the ventricle afterpassing through the perivalvular tract
The choice of the device used to occlude a perivalvularmitral valve leak is similar to the choice of device used
to occlude perivalvular aortic valve leaks The
avail-able devices include large, controlled-release coils, the
Amplatzer™ PDA, muscular VSD and Vascular Plugdevices, and the Gianturco-Grifka Vascular OcclusionDevices (GGVOD™) In addition to these devices, thelarger and stiffer delivery systems that are necessary toaccommodate the “double umbrella” devices, can be tol-erated from the venous/transseptal approach, allowingthe use of virtually any occlusion device that is suited forthe particular anatomy Once the exact device is chosen, theappropriate delivery system/catheter is delivered to theperivalvular leak from the prograde/transseptal/leftatrial approach The specific delivery system/catheter forthe occlusion device can be delivered to the perivalvular
leak over the Super Stiff™ wire through the original
trans-septal sheath, or the new long delivery sheath/dilator isintroduced over the exchange wire after removing theoriginal transseptal sheath If the original transseptalsheath does not accommodate the delivery system/catheter for the device that is chosen, the originaltransseptal sheath is replaced with a larger diameter longtransseptal sheath which is one or two French sizes largerthan the particular delivery system that is to be used forthe device delivery The larger diameter transseptalsheath is pre-curved before it is introduced to conform tothe course from the inferior vena cava, through the rightatrium and atrial septum and to a position just over themitral annulus The large long sheath positioned in the leftatrium provides extra support for the delivery catheter,and when the larger sheath is advanced to the left atrialend of the perivalvular leak over the delivery catheter, the large sheath allows “postioning” angiograms to beobtained just at the distal end of the leak during the deliv-ery of the device without the necessity of another catheter
in the left atrium When the device is deployed and flowthrough the tract is decreased/stopped, the contrast willreflux “backward” into the tract slightly
The device is delivered into the tract of the perivalvularmitral leak and positioned in the tract similarly to thepositioning in perivalvular aortic leaks and according tothe type of device being used Once opened in the tract,the positioning, degree of occlusion and distance awayfrom the leaflets of the prosthetic valve are examined
by repeat biplane angiography, injecting through the
Trang 38retrograde catheter or through the long sheath over the
delivery catheter, and by TEE/Doppler™ interrogation
When satisfied with the occlusion and the position, the
device is released A large volume left ventricular
angio-cardiogram is performed through the retrograde catheter
to visualize the final occlusion of the perivalvular leak and
the function of the prosthetic mitral valve, and to exclude
any additional perivalvular leaks If there is persistent
flow through the original perivalvular communication
and/or if there are any additional perivalvular leaks
detected, these are addressed during this same
catheter-ization for the same reasons as for residual perivalvular
aortic leaks
Complications of vascular
occlusions
As with all other therapeutic catheterization procedures,
there are potential complications related to the
catheter-ization itself and those more specifically related to the
occlusion procedure Since a majority of the abnormal
ves-sels and/or vascular channels requiring occlusion arise
off a systemic artery, most transcatheter occlusion
proced-ures are performed entirely or, at least, partially from an
arterial approach This, alone, increases the likelihood
of local arterial injury and of systemic embolic
complica-tions These complications and the steps taken to prevent
them are discussed in detail in Chapter 4 (“Vascular
Access”) and Chapter 9 (“Retrograde Technique”)
Embolization of the occlusion device and/or
inadvert-ent occlusion of the blood supply to a vital structure or
organ are the major complications unique to intravascular
occlusion of systemic vessels, fistulae and perivalvular
leaks Occlusion of unwanted vessels/structures is
pre-vented by test occlusion with a balloon catheter before
depositing an occlusion device and by the use of
controlled-release/retrievable devices whenever there is
any question about compromise of a vital adjacent vessel/
structure This is particularly important in the occlusion of
coronary-cameral fistulae, where the extension of even a
small part of the occlusion device into a vital coronary
artery can obstruct and/or thrombose an essential
cor-onary blood supply
Remote embolization of a vascular occlusion device
is always possible, but almost always is preventable
Certainly, the use of the controllable and retrievable
devices adds to the safety and secure positioning of
occlu-sion devices for peripheral vascular occluocclu-sions When
“controllable” release devices are not available or cannot
be used, extra care is taken in the sizing of the device for
the lesion and in choosing the safest location for the
implant of the occlusion device When a vascular
occlu-sion device does embolize to a different unwanted and/or
dangerous area or structure, it becomes a foreign bodythat requires removal It is imperative that the catheteriza-tion laboratory performing vascular occlusion procedureshas the equipment and expertise for the removal of alltypes of foreign bodies
The migration of a coil or other small occlusion devicecompletely through a large pulmonary arteriovenousfistulous communication into the pulmonary vein and, inturn, embolization into the systemic arterial circulationalways is a possibility and potentially, results in very seri-ous consequences This type of embolization is prevented
by the use of a much larger device and/or an “irregular”device, like an umbrella, which becomes trapped securely
in the afferent vessel of the fistula
Whenever there is incomplete occlusion of a vessel orother lesion with a resultant high-velocity residual “jet”leak through or adjacent to an implanted device, any ofthe vascular occlusion devices can produce hemolysis.Again, treatment is prevention by occluding the lesioncompletely during the initial catheterization for the occlu-sion When a high-velocity lesion does remain and doesresult in hemolysis, the patient initially is treated medic-ally with extra volume infusions and blood replacement
as necessary for symptoms Often the hemolysis is limited after 2–7 days, but when it is unrelenting and/orprogressive, re-intervention to eliminate the high-velocitylesion is necessary Usually additional transcatheter deliv-ered occlusion devices to stop the flow will solve the prob-lem, but rarely surgical removal of the device and closure
self-of the vessel/leak will be necessary
Complications of catheter occlusions of aorta to left ventricular tunnels and peri-prosthetic valvular leaks
Like all other therapeutic catheterization procedures, andparticularly those within the systemic arterial system, all
of the basic hazards of the cardiac catheterization ure are present with these procedures The occlusion ofaortic to left ventricular tunnels and perivalvular leaks ofboth the aortic and mitral valves with catheter-delivereddevices involves very extensive manipulations proximal
proced-to the head vessels in the systemic arterial system withwires, catheters and delivery systems, which, in turn, cre-ates a very high potential for cerebrovascular embolic
problems There must be constant vigilance to prevent the
introduction of air into the system and to keep all wirespassing through catheters, catheters and delivery systems
on a continuous flush to prevent thrombi from forming.The potential for embolization of the occlusion deviceitself from the defect to the systemic circulation is alwayspresent The equipment must be available for the immedi-ate capture and retrieval of any embolized device before itresults in permanent damage to a vital organ
Trang 39The occlusion of perivalvular leaks around mechanical
prosthetic valves involves extensive
catheter/wire/deliv-ery system manipulations adjacent to the valve orifice
and the mechanical leaflets of the involved prosthetic
valves with the potential for entrapment in the valve
and significant interference with the valve’s function The
manipulations in the area of these valves must be directed
very precisely and cautiously, and observed continually
on biplane fluoroscopy to avoid the valves.
Any of the occlusion devices placed in perivalvular
leaks can interfere with the function of a mechanical
pros-thetic valve Prevention is the optimal management by the
use of retrievable devices and the very careful positioning
and testing of each device as it is delivered If the function
of a prosthetic valve is compromised significantly, the
device must be removed either by retrieval with a catheter
technique or surgically
Summary
Unwanted shunts and/or vascular leaks in virtually all
cases can be occluded with a catheter-delivered device
Refinements in the imaging systems, the variety of
avail-able devices, and improvements in delivery have made
these procedures more effective and far safer In 2005,
trans catheter occlusion of residual and/or unwanted
vascular communications and/or leaks should be the
primary approach for their management
References
1 Bookstein JJ et al Transcatheter hemostasis of
gastrointest-inal bleeding using modified autogenous clot Radiology 1974;
113(2): 277–285.
2 Gold RE and Grace DM Gelfoam embolization of the left
gastric artery for bleeding ulcer: experimental
considera-tions Radiology 1975; 116(3): 575–580.
3 Kaufman SL et al Transcatheter embolization with
micro-fibrillar collagen in swine Invest Radiol 1978; 13(3): 200–
204.
4 Zuberbuhler JR et al Tissue adhesive closure of
aortic-pulmonary communications Am Heart J 1974; 88(1): 41–46.
5 Serbinenko FA [Balloon occlusion of saccular aneurysms of
the cerebral arteries] Vopr Neirokhir 1974; 4: 8–15.
6 Gianturco C, Anderson JH, and Wallace S Mechanical
devices for arterial occlusion Am J Roentgenol Radium Ther
Nucl Med 1975; 124(3): 428–435.
7 Yamamoto S et al Transcatheter embolization of bronchial
collateral arteries prior to intracardiac operation for tetralogy
of Fallot J Thorac Cardiovasc Surg 1979; 78(5): 739–743.
8 Grifka MR and Jones TK Transcatheter closure of large PDA using 0.052″ gianturco coils: controlled delivery using a biop-
tome catheter through a 4 French sheath Catheter Cardiovasc
Interv 2000; 49(3): 301–306.
9 Kuhn MA and Latson LA Transcatheter embolization coil
closure of patent ductus arteriosusamodified delivery for enhanced control during coil positioning Cathet Cardiovasc
Diagn 1995; 36(3): 288–290.
10 Uzun O et al Transcatheter occlusion of the arterial duct with
Cook detachable coils: early experience Heart 1996; 76(3):
269–273.
11 Grifka RG et al New Gianturco-Grifka vascular occlusion device Initial studies in a canine model Circulation 1995;
91(6): 1840–1846.
12 White RI Jr et al Therapeutic embolization with detachable
balloons Physical factors influencing permanent occlusion.
Radiology 1978; 126(2): 521–523.
13 Goto K et al Permanent inflation of detachable balloons with
a low-viscosity, hydrophilic polymerizing system Radiology
1988; 169(3): 787–790.
14 Moore JW and Murphy JD Use of a bow tie stent occluder for
transcatheter closure of a large anomalous vein Catheter
Cardiovasc Interv 2000; 49(4): 437–440.
15 Khan MS and Moore JW Treatment of abdominal aortic pseudoaneurysm with covered stents in a pediatric patient.
Catheter Cardiovasc Interv 2000; 50(4): 445–448.
16 Tulloh RM and Rigby ML Transcatheter umbrella closure of
aorto-pulmonary window Heart 1997; 77(5): 479–480.
17 Jureidini SB, Spadaro JJ, and Rao PS Successful transcatheter closure with the buttoned device of aortopulmonary win-
dow in an adult Am J Cardiol 1998; 81(3): 371–372.
18 Richens T and Wilson N Amplatzer device closure of a
resid-ual aortopulmonary window Catheter Cardiovasc Interv 2000;
50(4): 431–433.
Trang 40Transcatheter closure of patent ductus arteriosus (PDA)
with one or more of several (many) devices/techniques
now represents the established, standard approach for the
correction of a PDA throughout the entire world
includ-ing even the United States (US) There now is one device
approved even by the US Food and Drug Administration
(FDA) specifically for closure of the PDA
The patent ductus arteriosus presents in an infinite
vari-ety of shapes and each shape appears in an equally wide
range of sizes1 The most common shape is conical with a
large aortic end (ampulla) tapering to a narrower
pul-monary end The persistent ductus however, can be long
and “finger-like” with or without an additional tapering
at the pulmonary end; it can have several areas of
narrow-ing or even be shaped like a long tube with no tapernarrow-ing at
any area The long tubular ductus usually is relatively
straight but can be tortuous At the other extreme the PDA
can be a very short and flat communication or “window”,
between the aorta and the pulmonary artery Each shape
of the PDA can vary in diameter at its narrowest portion
from less than 1 mm to greater than 10 mm The flow
through the ductus and the differential in pressure
be-tween the two ends of the ductus depend upon the size of
the ductus and the relative resistances of the pulmonary
arterial and systemic arterial circulations Obviously, no
single device can be optimal or even applicable for all
of the varieties in the size and shape of the PDA, and
fortunately over the past three plus decades multiple
devices and the procedures for delivering them have
been developed for transcatheter closure of the patent
ductus arteriosus
Transcatheter occlusion of the PDA was first introduced
by Porstmann in 1967 et al.2, however, PDA occlusion in
the catheterization laboratory was not popularized until
the early 1980s, when the more practical Rashkind™ PDA
occluder was introduced into clinical trials in the US and
the rest of the world Since the introduction of theRashkind™ PDA device, many different devices and tech-niques for transcatheter occlusion of the PDA have beenintroduced and used clinically for almost 25 years with
minimal morbidity and no mortality from the catheter cedures The success and complications of the procedures
pro-for transcatheter occlusion of the PDA have been variableand are related both to the type and size of the ductus, theparticular device used, the delivery procedure utilized,and to the experience and skill of the operators perform-ing the occlusion
Depending upon where one resides in the world, in
2004 there were between three and ten devices and/or
techniques available for PDA occlusion in the
catheteriza-tion laboratory The availability of devices in any lar area or country depends upon the developmentalstage of new devices and the approval for human use
particu-by the regulatory agencies of the various countries.Ironically, most of the viable devices and techniques forocclusion of the PDA in the catheterization laboratorywere designed, developed and clinically validated in the
US Unfortunately for the pediatric and adult patientswith congenital heart lesions in the US, the FDA has beenultra conservative toward children in their review andapproval of specific devices As a consequence, the UnitedStates has the fewest devices available and the mostrestricted “approval” for the use of those devices that areavailable for specific congenital heart lesions throughoutthe rest of the world
There are several devices that are approved by the FDA
in the US for the occlusion of “vascular lesions” that havebeen adapted for transcatheter PDA occlusion and areused for PDA occlusions in the catheterization laboratory
in the US These devices and their special delivery niques have been demonstrated to be extremely safe and
tech-effective and are accepted as the standard therapy for PDA occlusion by all knowledgeable medical professionals
throughout the world, including in the US However, in
2004, the Amplatzer™ PDA device was the only device