With combined aortic stenosis and coarctation, the diameters of the aortic valve annulus and the coarctation with the appropriate adjacent aortic diameters are meas-ured from a left vent
Trang 1completely The sheath is withdrawn and pressure
applied over the vessel manually As soon as all the
sheaths are out, the drapes over the patient are removed
completely so that the lower abdomen and upper thigh
areas adjacent to the entire inguinal area are visible While
holding pressure over the arterial puncture site with the
fingers of one hand, a pulse distal to the puncture site
(dorsalis pedis or posterior tibial) is palpated with the
fingers of the other hand The amount of pressure applied
over the arterial puncture site is varied or “titrated” so
that just enough pressure is applied to prevent bleeding
(or a subcutaneous hematoma formation) while at the
same time allowing the palpation of a peripheral pulse
continually In larger or heavier patients, it is better
to apply pressure over the arterial site with a firm “roll” of
“4 × 4s”, since the exact site of arterial puncture deep
within thick subcutaneous tissues does not correspond at
all to the site of the skin puncture and subsequent
pres-sure The pressure from the “roll” of bandage covers a
wider area deep within the tissues and is more likely to
control deep bleeding from the artery The same
tech-nique for monitoring the peripheral pulse is used while
pressure is applied with the “roll” of bandage
In addition to the care of the local puncture sites, these
patients are monitored systemically very closely in a
recovery area for at least six hours During this
observa-tion period, they should have a secure intravenous line
and receive a high maintenance infusion of 1/4 normal
saline or Ringer’s lactate The intravenous fluids are
con-tinued for twelve hours A patient following dilation of a
coarctation of the aorta usually has diuresed during the
procedure from both the contrast agents used and from
the increased renal blood flow as a consequence of the
relief of the coarctation Often these patients start out “dry”
from being NPO for a prolonged period of time before the
procedure begins The combination of these factors results
in a very significant volume depletion, which, in turn,
aggravates vagal or other vascular responses caused by
the changes in the distribution of arterial blood flow
Surprisingly, post-dilation patients rarely suffer from
the “post-coarctectomy” syndrome that is seen commonly
following surgery Dilation patients rarely have any
aggravation of their upper extremity blood pressure,
although the systemic pressures may not drop to normal
immediately13 Post-dilation patients essentially never have
abdominal discomfort and usually resume oral intake
within 6–12 hours after dilation There have been reports
of serious post-procedure complications following the
dilation of coarctations, so no matter how smoothly the
procedure went, these patients are observed overnight
Although they may exhibit pain during the actual balloon
inflation in the coarctation site, once the balloon is deflated,
the pain subsides Any persistent or recurrent pain should
be taken seriously and investigated thoroughly
Dilation of coarctation neonates and young infants
Coarctation of the aorta in the neonatal patient presentssome unique features These infants frequently present tothe cardiologist catastrophically ill with heart failure, inacidosis and in shock They frequently have additionaldefects, some of which complicate the catheter manage-ment and some of which actually help the catheter man-agement The identification of any additional defects ismade from the clinical examination and the echocardio-gram The echo cannot consistently provide the detailsabout the coarctation size, anatomy and severity, and fre-quently underestimates the size of the adjacent aortic seg-ments Decisions about therapeutic intervention for the
neonatal coarctation of the aorta should not be made
entirely on the basis of the echo
A vigorous, but brief attempt is made at stabilizinginfants who are less than three to four weeks old with vent-ilation, inotropics, volume support and prostaglandin.Stabilization attempts are continued only as long as the
infant improves If there is going to be any improvement in
the clinical condition, it is noticeable within one to twohours With, or without, the stabilization after that dura-tion of time, the infant is taken to the catheterization lab-oratory If the infant is not responding noticeably to theresuscitative efforts, more time only allows further deteri-oration If the infant is responding, the improvement willcontinue on the way to, and in, the catheterization labor-atory The exact procedure performed depends upon thepresence or absence of associated lesions and, when pres-ent, which associated lesions are present
Besides their general hemodynamic instability, the verysmall femoral arteries with the associated very weak orabsent femoral pulses represent the greatest challenge forballoon dilation of coarctation of the aorta in the verysmall infant In each individual case, the most expeditioustechnique possible is utilized to approach and treat thecoarctation Once the coarctation area has been reachedwith a catheter, the gradient is measured, the selectiveaortogram(s) performed and accurate measurements ofthe coarctation and adjacent vessels are made in order tochoose the proper balloon
When the infant responds to the administration ofprostaglandins with opening of the ductus arteriosus, thestabilization of the infant is usually rapid and very effect-ive The presence of the patent ductus indirectly facili-tates access to the coarctation by improving perfusion
to the lower extremities significantly, and increases theamplitude of the femoral pulses, which facilitates percuta-neous arterial access The presence of a patent ductus,however, compromises the results of balloon dilation of
a neonatal coarctation Without the patent ductus, the
Trang 2balloon is constrained within the aorta and the “path of
least resistance” is to crush the abnormal ridge of tissue
within the aortic lumen A wide open ductus arteriosus,
on the other hand, allows the dilation balloon to move
away from the coarctation “ridge” into the “ampulla” of
the ductus rather than crushing or compressing the ridge
In addition, when the ductus does constrict during its
nor-mal closure, it probably also constricts some of the
adjac-ent aorta, recreating the coarctation
Most of the newborns and small infants with
coarcta-tion do have at least a potential interatrial communicacoarcta-tion
In all of these infants, venous access is obtained and an
angiographic catheter is introduced into at least the left
ventricle from the venous approach At the same time, in a
small sick infant, no prolonged effort should be made at
advancing this catheter from the left ventricle into the
aorta, as is utilized in coarctation dilation in the older
patient The chambers and vessels are very small and the
tissues “softer” and prone to puncture The catheter in the
left ventricle provides continual systemic pressure
mon-itoring before, during and immediately after the dilation
procedure The pressure in the left ventricle reflects the
severity of the coarctation unless there is associated aortic
stenosis or the infant is in terrible heart failure A left
ven-tricular angiocardiogram is performed which provides
some, and possibly, very good information about the
coarctation and the overall anatomy
Infants who do not respond to prostaglandins and who
have no associated lesions, and those who have associated
aortic stenosis, are the most difficult for coarctation dilation
The echocardiogram should diagnose associated aortic
stenosis and give an estimate of its severity, although
either one of the two lesions can mask the significance of
the other The prograde left ventricular catheter is useful
in these patients for monitoring pressures and for the
administration of fluid and medication, and is possibly
helpful in determining the relative severity of
com-bined lesions
Percutaneous entry into the artery is accomplished with
meticulous attention to detail, a very delicate single wall
vessel puncture technique, and patience (as described in
Chapter 4) The area around the expected arterial
punc-ture site is infiltrated with local anesthesia, being as
care-ful as possible not to puncture the artery with the needle
during the infiltration If the artery is punctured
inadvert-ently, pressure is held over the site for at least 2–4 minutes
before beginning the purposeful puncture of the artery
Special 21-gauge percutaneous needles and extra-floppy
tipped 0.018″ or 0.014″ wires are essential for the
percuta-neous entry into the artery in these patients Once the
artery has been entered with the guide wire, the area
sur-rounding the puncture site is re-infiltrated liberally with
local anesthesia A 3- or 4-French “fine tipped”
sheath/dila-tor is introduced into the artery The preliminary diagnostic
information about the coarctation is obtained with the trograde catheter If the infant is still very unstable, an endand side hole multipurpose catheter is used This type ofcatheter allows pressure measurements and quality aor-togram(s), and at the same time can be used to position theguide wire for the dilation without the necessity of evenone catheter exchange The second catheter previously
re-positioned in the left ventricle helps to confirm the
pres-ence of associated aortic stenosis However, with poor leftventricular function and an associated coarctation of theaorta, even very severe aortic stenosis can be masked
Once the coarctation has been identified as the only
significant lesion, it is measured accurately and the dilation
of the coarctation is carried out as expediently as possibleusing the single-catheter, retrograde technique It ispreferable to stabilize the distal end of the guide wire inone of the subclavian arteries, although an excessivelylong time or effort should not be taken to achieve this loca-tion The balloon for the coarctation dilation is chosen toequal the size of the smallest segment of the aorta adjacent
to the coarctation Once the dilation has been completedand the balloon removed over the wire, the end-holecatheter is re-advanced over the wire and very gently pastthe coarctation to the aorta proximal to the coarctation.The wire is removed and pressures are recorded simulta-neously through the catheter from the ascending aortaand through the side arm of the sheath from the femoralartery A repeat aortogram is recorded, injecting throughthe end-hole catheter proximal to the coarctation site.The catheter should not be withdrawn across the coarc-tation site until all post-dilation studies or any re-dilationhas been accomplished If the dilation was unsatisfactoryand a re-dilation is necessary, the wire is replaced throughthe catheter which is already across and well beyond thelesion in order that no catheter manipulation back acrossthe freshly dilated area is necessary When the catheterhas been withdrawn across the area, it should not bemanipulated back across the area3
Co-existent critical aortic stenosis and coarctation in infants
When there is significant aortic stenosis in associationwith the neonatal coarctation, it is preferable to addressthe aortic stenosis first In the presence of the combinedlesions, perfusion of the coronary and cerebral circula-tions is dependent, at least partially, on the increasedafterload in the ascending aortic pressure provided by thecoarctation Removing this afterload before opening theaortic valve could compromise the coronary and cerebralcirculations even further In addition, if the coarctation isdilated first, all of the manipulations (sometimes extens-ive) required for crossing the aortic valve, and the aorticvalve dilation, will have to be through the freshly dilated
Trang 3coarctation site with the potential for traumatizing the
already damaged aortic intima in the coarctation site even
further
With combined aortic stenosis and coarctation, the
diameters of the aortic valve annulus and the coarctation
with the appropriate adjacent aortic diameters are
meas-ured from a left ventricular angiocardiogram or an aortic
root injection before an attempt is made to pass a catheter
across the stenotic valve A left ventricular
angiocardio-gram is obtained with injection through a prograde left
ventricular catheter, while the aortic root injection is
obtained with the retrograde multipurpose or
angio-graphic catheter that has been manipulated past the
coarc-tation and around the arch to the aortic root Once the
valve and coarctation measurements are obtained, the
appropriate dilation balloon for the aortic valve dilation is
prepared using the “minimal prep” technique but with
a prolonged attempt at removing all air No attempt is
made to cross the aortic valve until the balloon for the
dila-tion is prepared and “poised” for introducdila-tion These
infants are often so precarious that even a tiny 4-French
catheter crossing the stenotic orifice is enough to cause a
rapid decompensation in their hemodynamics
To cross these valves, a 0.018″ or 0.014″, very floppy
tipped, exchange length, torque-controlled, coronary
artery, guide wire is advanced through a multipurpose or
selective right coronary catheter which is already
posi-tioned in the aortic root The wire is advanced out of the
catheter and multiple, rapidly repeated, short probes are made
toward the aortic valve area with the very soft wire tip
Because even a very soft tipped wire exiting the tip of the
catheter can be very stiff for the first few millimeters out of
the tip, the tip of the catheter is kept a centimeter away
from the valve annulus during the probes with the tip of
the wire The tip of the wire is redirected within the aortic
root by simultaneously rotating the catheter (to change
the anterior to posterior direction) and moving the
catheter to and fro (to change the right to left side angle)
Unless another wire or catheter is already passing
through the valve, the exact location of the orifice really is
not known The more “probes” that are made with the
wire along with multiple changes in the angle of the wire,
the more likely is the chance of the wire passing through
the orifice of the “invisible” valve
If the wire does not cross the valve after trying for
several minutes using the original multipurpose catheter,
the multipurpose catheter is replaced with a preformed,
either right or left coronary catheter and the “probing” at
the valve repeated with similar changes in direction of the
catheter tip The tighter, preformed curves at the tips of
coronary catheters allow a greater ability to change the
angle of approach toward the small valve orifice
Once the wire crosses the valve, it is advanced as far as
possible into the ventricle, hopefully even looping the soft
tip within the left ventricle apex With the wire passed asfar as possible into the ventricle, the catheter is advancedover the wire into the ventricle If the infant’s hemody-
namics do not remain stable with the catheter across the valve, the wire is fixed in the ventricle and the catheter is immediately removed and replaced rapidly over the wire
with the previously prepared balloon dilation catheter.The dilation of the valve is carried out with as rapid aninflation and deflation as possible and while recordingangiographically or on “stored fluoroscopy” After theinflation/deflation, the balloon is immediately with-drawn out of the valve over the wire into at least theascending aorta
If the infant remains stable with the end-hole catheterpassed through the valve, the wire is removed A “pig-tail” is formed on the tip of the long floppy tipped, stiffer,exchange guide wire and a 180° curve is formed on thetransition zone between the long floppy tip and the stiffshaft of the wire This individually formed wire is re-advanced through the catheter into the left ventricle Thepreformed “pig-tail” on the wire keeps the wire from digging into the myocardium as it is manipulated in the
ventricle The 180° curve at the transition area of the wire
directs the tip of the wire back toward the left ventricular
outflow tract and allows the stiff portion of the shaft of the
wire to be positioned further across the valve With thewire maintained in position, the catheter is removed andreplaced with the already prepared balloon catheter Theballoon is positioned across the valve and the inflationperformed while recording the inflation angiographically
or on “stored fluoroscopy” The inflation/deflation is formed as rapidly as possible During the inflation the leftventricular pressure increases and the infant developsbradycardia As soon as the deflation of the balloon iscomplete, the balloon is withdrawn over the wire and out
per-of the aortic annulus The infant’s heart rate should returnand the left ventricular pressure will drop to normal levels.The return of a good heart rate and a good, but lower,left ventricular pressure are immediate indications of thesuccess of the dilation Ideally, there will be a lower leftventricular pressure, but in the presence of the associatedcoarctation, the gradient may be “moved downstream” tothe coarctation site with little lowering of the left ventricu-lar pressure The radiographic recording of the inflation/deflation is reviewed If a “waist” appeared on the balloonand then disappeared during the inflation, some true dilation of the valve orifice is assumed The balloon iswithdrawn back to the area of the coarctation If the dia-meter of this balloon is smaller, or, at least, no more than amillimeter larger, than the measurement of the smallestdiameter of the aorta adjacent to the coarctation, the coarctation site is dilated with the same balloon If the bal-loon is two or more millimeters larger than the adjacentaorta in the area of the coarctation, the balloon is replaced
Trang 4with an appropriate diameter balloon and the
coarcta-tion dilated
After both the aortic valve and the coarctation have
been dilated the balloon is removed and replaced with a
catheter to re-evaluate the hemodynamics If there is a left
ventricular catheter in place, the “net” results of the
com-bined dilation are determined by measuring the left
vent-ricular and femoral artery pressures simultaneously using
the side arm of the sheath for the femoral artery If the net
left ventricular to femoral artery gradient is low, it is
assumed that both procedures were successful and the
procedure can be concluded A left ventricular
angiocar-diogram through the prograde catheter provides
visual-ization of both the aortic valve and the coarctation area
An end-hole catheter is passed over the wire to the left
ventricle and the retrograde wire removed Pressures are
recorded on withdrawal of the retrograde catheter from
the left ventricle, to the aorta and across the coarctation
site to quantitate the residual gradients at each area
If the net gradient is still significant, the culprit lesion(s)
is/are identified from the pressures and angiograms If
there is no prograde catheter in the left ventricle, the
results of the dilations must still be determined A
Multi-Track™ catheter provides the most “secure” way of
iden-tifying the major residual problem without having to
remove the wire from the left ventricle Pressures are
recorded and angiograms are performed through the
Multi-Track™ catheter anywhere along the course from
the left ventricle to the descending aorta and all without
having to remove the wire, which can be maintained
posi-tioned in the left ventricle A small, stiff wire is placed in
the true catheter lumen of the Multi-Track™ catheter to
stiffen and support the catheter shaft as Multi-Track™ is
being advanced over the exchange wire Once the
Multi-Track™ catheter is in the left ventricle, the wire within the
true lumen is removed Angiograms are performed where
appropriate and pressures are recorded as the
Multi-Track™ is withdrawn over the exchange wire When a
significant area of residual obstruction is identified, the
decision is made whether this should or can be treated
from the pressures, the review of the balloon dilations and
the current angiograms performed through the
Track™ catheter If a lesion is to be re-dilated, the
Multi-Track™ is removed over the wire and the appropriate
balloon dilation catheter reintroduced to the culprit lesion
If a Multi-Track™ catheter is not available or cannot be
used with the particular system, then an end-hole,
multi-purpose catheter is advanced over the wire to the left
ventricle and the exchange wire removed Pressures
are recorded and sequential angiograms are obtained
through this catheter as it is withdrawn from the left
vent-ricle to the ascending aorta and from the ascending aorta
to the descending aorta This, of course, removes the
pre-viously secure wire access back into the ventricle! If the
significant or predominant gradient is at the aortic valve,the catheter withdrawal is stopped in the aortic root and adecision is made as to whether further catheter therapy ispossible If the aortic valve is to be re-dilated, the softfloppy tipped wire is reintroduced through the catheterand manipulated across the aortic valve into the left ven-tricle The dilation procedure is repeated with a moreappropriate diameter balloon
If there is a significant net left ventricle to femoral arterygradient, but little or no gradient at the aortic valve, thecoarctation is still the culprit The previous coarcta-tion dilation is reviewed If a larger balloon can be used,the wire is reinserted into the catheter and positioned
in the aortic root before the catheter is withdrawn across the coarctation The end-hole catheter is removed over thewire, the appropriate diameter balloon is passed over the wire and the coarctation re-dilated Reassessment ofthe result depends upon which catheters are available,
as described above
Whenever an infant is found to have coarctation of theaorta with aortic stenosis, associated mitral stenosis ofsome type and degree should be suspected The Shone’scomplex of multiple left heart obstructive lesions includesvarious types of mitral valve stenosis along with thecoarctation and aortic stenosis, and frequently results
in an even sicker infant Any one, or all, of the levels ofobstructive lesions can be severe and require intervention,however usually the coarctation and aortic lesions are themost pressing and are the only ones that can be addressedreasonably in the newborn period Congenital mitralstenosis can be treated by balloon dilation in the slightlyolder child, and is addressed separately in Chapter 20.The most favorable associated lesions in an infant withcoarctation that need dilation are a ventricular septaldefect (VSD) and transposition of the great arteries AVSD provides access to the aorta from the right heart and
in turn from the venous access Although access to the
coarctation through the VSD is anticipated, these infantsshould have an indwelling arterial line for systemic pres-sure monitoring In these patients, a curved tipped, end-hole or multipurpose catheter introduced through a shortsheath from a percutaneous venous introduction is usedfor the “right heart” procedure In infants, when manipu-lating a venous catheter from the right ventricle towardthe outflow tract, as the catheter is torqued clockwise,
it frequently and, even preferentially, passes dorsallythrough the ventricular defect and, from there, into theascending aorta Otherwise, with purposeful manipula-tion of the catheter dorsally and cephalad from the rightventricle, entering the aorta almost always is accom-plished in infants with a significant VSD From there,approaching the coarctation around the arch is a straight-forward procedure with a wire and curved tip catheter
A selective aortogram is performed and the lesion and
Trang 5adjacent vessels measured accurately The coarctation
is crossed with a prograde end-hole catheter and the
catheter is exchanged for the exchange length, stiff, guide
wire The balloon is introduced from the vein and
advanced through the right ventricle, through the VSD
and to the coarctation The dilation is accomplished all
prograde from the venous entry site without the need
for even a 3-French sheath in a femoral artery The
post-procedure measurements are carried out with an end-hole
catheter replacing the balloon catheter in the aorta before
the “prograde” guide wire is removed
The same ability to perform a dilation of a coarctation
from a venous access holds true for infants with
coarcta-tion of the aorta in associacoarcta-tion with transposicoarcta-tion of the
great arteries (with or without a ventricular septal defect
or aortic or pulmonary override) This is a particularly
common association in patients with the so-called Taussig–
Bing complex of transposition of the great arteries,
ven-tricular septal defect and pulmonary artery override A
standard end-hole or a balloon “wedge” catheter passes
easily from the right ventricle and out into the aorta With
minimal manipulation the catheter is maneuvered into
the descending aorta to the coarctation The diagnosis and
dilation are performed similarly to that in the infant with
an associated VSD
Intravascular stents in coarctation of the aorta
The use of intravascular stents for the treatment of
coarc-tation of the aorta has become the primary approach for
coarctation in the larger adolescent and adult The use and
advantages of intravascular stents in both native and
re/residual coarctation are covered in detail in Chapter
25 Their use in conjunction with dilation of coarctation of
the aorta has changed the approach to these lesions
dra-matically Intravascular stents not only support the
ves-sels at the maximal dilated diameter of the balloon, but
also, in doing so, eliminate the need for any over-dilation
of the vessel
However, the basic rules for the implant of
intravascu-lar stents in pediatric and congenital lesions apply even
more stringently to their use in coarctations In particular,
no stent should be implanted in the aorta if it cannot be
dilated up to the ultimate adult diameter of the aorta in that
particular patient This “rule” so far has precluded the
rea-sonable and/or sensible use of stents in coarctation of the
aorta in infants and small children The implant of a stent
that cannot be dilated to the diameter of the adult aorta,
creates a new, very fixed diameter coarctation as the
patient grows A coarctation with a stent in place, which is
too small for the aorta and which cannot be dilated
fur-ther, must be managed surgically There is a much higher
risk for surgery on a coarctation with a stent in place than
for the surgical repair of native coarctation, even in an
infant The narrow segment of aorta that contains thestent, at the very least, must be “filleted” open and thenpatched, if not totally resected or bypassed with a pros-thetic graft At the same time, this same patient whorequires such extensive surgery on the aorta, no longerhas the extensive collaterals to protect the spinal cord andthe lower half of his body Hopefully, the further develop-ment of “open ring” stents or some type of resorbablestent, will eventually make it possible to use stents in theinitial treatment of coarctation in infants and children
In the interim, balloon dilation of coarctation is veryeffective in infants and children Even if the obstruction ofthe coarctation is not eliminated totally with an initialdilation, or even with several balloon dilations, if the aorta
is not over-dilated and torn or ruptured, a successful loon dilation with a stent implant can be accomplishedwhen the patient reaches adolescence or adulthood Forthe infant and child, a “conservative” dilation of bothnative and re/residual coarctation should be the primaryapproach to these lesions
bal-Complications specific to coarctation dilation
The majority of the early dilations reported in theValvuloplasty and Angioplasty of Congenital Anomalies(VACA) registry were for re-coarctation of the aorta and,
as a consequence, the majority of the acute complicationswere in these re-coarctation dilations In the VACA series
of re-coarctation dilations there were five deaths (twovagal, one aortic rupture, one CNS death and one deathdue to shock)1
A better understanding of the effect of the post-dilationvasodilation and the exaggerated physiologic vasovagalresponse to the pressure drop associated with dilationshas helped to eliminate the extreme vasovagal response
to the procedure The patients are maintained on a highmaintenance volume of intravenous fluids during thedilation and this fluid infusion is maintained for six totwelve hours after the dilation The intravenous access
is left in place until after the patient is ambulated Withthese precautions, the vasovagal type reactions have been eliminated
Significant central nervous system (CNS) injuries and atleast one death attributed to CNS injury have beenreported following dilation of coarctation of the aorta The
exact etiology of all of the CNS injuries has never been
determined unequivocally Fortunately, with more tion to the position of the distal end of the guide wire forsupport of the dilation balloon and extremely stringentprecautions concerning emboli from the materials and/orthe procedure, the central nervous system complica-tions essentially have been eliminated It is critical that the
Trang 6atten-distal end of the support wire is not positioned in a carotid
or vertebral artery or possibly even in the ascending aorta
Catheters passed over wires are maintained on a
contin-ual flush in order to eliminate any accumulation of blood
and clot around the wire within the catheter or at the
wire–catheter interface Flushing over the wire is
accom-plished by introducing the wire through a wire
back-bleed valve with a flush side port, which is attached to the
hub of the catheter The side port of the wire back-bleed
valve is maintained on continual flush from the
pres-sure/flush system In spite of the theoretical potential for
an aortic tear, patients undergoing coarctation dilation are
heparinized fully with the hope of eliminating clot
forma-tion on the wires or catheters
Although possible, acute, through and through tears in
the aortic wall from the dilation of coarctations have been
exceedingly rare One report early in the history of
dila-tion of coarctadila-tion of the aorta was in a patient with
re-coarctation of the aorta who had remarkably little or
no reaction or scar formation from the prior surgery and
apparently little or no reaction to the acute local injury
The patient had a through and through tear in the aortic
wall and eventually succumbed to the lesion Possibly,
aortic tears cannot be avoided unequivocally, however it
appears likely that aortic tears can be prevented by less
aggressive initial dilation of aortic coarctations and not
over-dilating any segment of the aorta adjacent to the area
of the coarctation by measuring that area of the aorta and
using that measured diameter to determine the diameter
of the balloon for the dilation Dilation of a typical “native”
coarctation presumably tears a “membrane” that lies
across the lumen of the aorta and does not extend the tear
significantly into the wall of the aorta Dilation of an
equally tight “re or residual” coarctation probably involves
dilation of the actual wall of the aorta which has
con-stricted down in the area of the previous therapy and, as a
consequence, these should be dilated more conservatively
Most patients exhibit pain acutely during the process of
inflating a balloon in the aorta, however the pain subsides
when the balloon is deflated If a patient has persistent pain
after acute dilation of the coarctation, the area of the
coarc-tation should be investigated carefully with selective
aortography or intravascular ultrasound in the area of the
dilation Even without identifying a tear, the patient
should be observed in the catheterization laboratory until
the pain subsides or for one or more hours if the pain
persists If a through and through tear with progressive
extravasation is detected, the balloon is reinflated at a low
pressure in the area of the tear The inflation is just enough
to “tamponade” the vessel and allow the patient to be
taken to the operating room for a repair Such a surgical
aortic repair certainly will require relatively prolonged
cross clamping of the aorta and in turn, may require
femoral vein to femoral artery bypass
There has been one report of paraplegia following a balloon dilation of a re/residual coarctation of the aorta in
a small infant with associated complex congenital heartdisease14 The prior surgery on the aorta had involved anuncomplicated end-to-end anastomosis There was no pro-longed ischemia, no evidence of aortic tear nor any evid-ence for embolic phenomena to explain the paraplegia
In the early reported series of coarctation dilation, therewas an 8.5% incidence of significant enough injury to thearterial introductory site to require intervention or havepermanent sequelae at the entrance site into the vessel Inthose early years of balloon dilation, the balloon dilationcatheters were very large and the balloons themselveswere large and grotesque by today’s standards All of theballoons greater than 6 mm in diameter were on 9-Frenchcatheter shafts, the balloons were relatively thick walledand they did not fold well around the catheter so that theballoon “mass” had a very large profile In order to beintroduced through a sheath these balloons required an11- or 12-French sheath! As a consequence to “reduce thesize” of the entry hole into the vessel most of these earlyballoons were introduced into the vessels directly over awire without a sheath in the vessel This in fact probablydid not reduce the diameter of the “hole” in the vessel
wall and certainly contributed to significantly more local
trauma to the vessel walls Considering the ballooncatheters used at that time, the incidence of arterial injuryactually was remarkably low!
Many refinements in the balloons with marked tion in the size of the catheter shafts and decrease in theballoon profiles have not only allowed a smaller “hole” for introduction into the vessel, but have also allowed dilation balloons routinely to be introduced throughindwelling sheaths With the combination of smaller balloon profiles, routine use of indwelling sheaths, verymeticulous care of the introductory sites into the arteries,liberal, repeated use of local anesthesia, heparinization(perhaps?), and personal attention to the site duringhemostasis after the balloon/sheath is removed, local vessel injury is now very rare
reduc-When an acute arterial injury does occur, it is treatedaggressively at the time of the catheterization With anabsent pulse, the patient is continued on heparin therapyfor one or two hours or until the pulse returns If there is
no return of a pulse within 1–2 hours the patient is eitherbegun on thrombolytic therapy with continued heparin
or is taken back to the catheterization laboratory for amechanical recanalization of the vessel, as described inChapter 34 on purposeful perforations
With only a 20-year history of balloon dilation of tion of the aorta, any long-term adverse sequelae from the procedure are as yet unknown The “longer-term”adverse sequelae that have occurred so far, seem to occurmore with the dilation of “native” coarctation of the aorta
Trang 7coarcta-where the aorta is not “protected” by a surrounding area
of scar tissue, as is found around the re-coarctations The
most bothersome finding after coarctation dilation is the
creation of an “aneurysm” at the site of the coarctation
dilation The incidence of these aneurysms has varied
markedly in different series, but in general they are rare
The “aneurysms” are usually a small out-pouching in
the aortic wall in the area of the balloon dilation This
out-pouching is usually apparent immediately after the
dilation In most cases the out-pouching does not enlarge
and in fact seems to remodel into the aortic wall along with
the overall remodeling of the aorta The out-pouching of
the aorta is considered an aneurysm when the area
“remodels” into a persistent, discrete out-pouching There
are no reported long-term adverse consequences of the
aneurysms, however, in at least one series, several of
the patients with an aneurysm did undergo surgery for
it The indication for the surgery was not because of
any clinical problem due to the aneurysm, but because of
the fear of what might happen to the aneurysm over time
The pathology of those aneurysms that were operated
upon showed tears through the intima and media of the
vessel wall Some aneurysms have been followed for
as long as 18 years with no increase in their size and
no sequelae
Aneurysms of the aorta following surgical repair are not
uncommon These aneurysms are more frequent when
there is a persistent narrowing of the aorta proximal to the
coarctation repair site or when the surgical repair was
carried out with a patch angioplasty These post-surgical
aneurysms are large saccular dilations of the entire area
of the coarctation/aorta distal to the more proximal
“obstruction” Most of these aneurysms continue to
grow with time and most have been referred for surgical
revision
The total incidence of aneurysms following
coarcta-tion dilacoarcta-tion initially was small and actually seems to be
decreasing and, possibly, not occurring at all in more
recent series In the earlier days of coarctation dilation, the
measurements of the coarctation and the adjacent vessels
were less accurate, and less attention was paid to the
narrow adjacent structures Often, marked over-dilation
of these areas was carried out with balloons significantly
larger in diameter than the adjacent vessel, either because
of the inaccuracies in measurement or purposefully in an
attempt to achieve a more lasting result Perhaps the more
conservative dilations done with the knowledge that the
aorta can be dilated further later or can be held open
with an intravascular stent without the need for any
over-dilation will totally eliminate the aneurysms associated
with coarctation dilations When a discrete aneurysm
does occur, it should be followed closely, probably at
least every one or two years with repeated CT, MRI, or
angiographic imaging
A discrete aneurysm can be excluded by a stent graft(large covered stent) or covered with a standard intra-vascular stent to support the aortic wall in the area andthen to have coils packed into the aneurysm behind theintravascular stent
Conclusion
Balloon dilation is an accepted standard treatment forboth native and re/residual coarctation of the aorta inpatients of all ages in many major cardiovascular centersdealing with congenital heart disease The safety of theacute procedure now appears to be very good, perhapsbecause of a better understanding of the lesions and amore conservative approach to the dilation procedure
A patient who undergoes a conservative dilation of tation of the aorta and who has a less than optimal result,but at the same time has no complications, can always havethe dilation of the coarctation repeated with, or without,the implant of an intravascular stent to “complete” theprocedure
coarc-The success of dilation of coarctation of the aorta overthe years and the reduction in the complications and the difficulties with repeat surgery on patients withre/residual coarctation have made dilation of native andre/residual coarctation the procedure of choice in mostcenters Most centers stipulate that all patients who havedilation of any type of coarctation of the aorta should befollowed indefinitely In patients who are, or are at leastnear to, full-grown many centers now regularly performprimary intravascular stent implants along with dilation
of coarctation of the aorta The use of stents in coarctation
of the aorta is covered in detail in Chapter 25
References
1 Hellenbrand WE et al Balloon angioplasty for aortic
re-coarctation: Results of the Valvuloplasty and Angioplasty
of Congenital Anomalies Registry Am J Cardiol 1990; 65:
793–797.
2 Tynan M et al Balloon angioplasty for the treatment of native
coarctation: Results of the Valvuloplasty and Angioplasty
of Congenital Anomalies Registry Am J Cardiol 1990; 65:
790–792.
3 Finley JP et al Balloon catheter dilatation of coarctation of the
aorta in young infants Br Heart J 1983; 50: 411–415.
4 Lock JE et al Balloon dilation angioplasty of aortic
coarcta-tions in infants and children Circulation 1983; 68(1): 109–116.
5 Kappetein AP et al More than thirty-five years of coarctation repair An unexpected high relapse rate J Thorac Cardiovasc
Trang 87 Kalita J et al Evoked potential changes in ischaemic
myelo-pathy Electromyogr Clin Neurophysiol 2003; 43(4): 211–215.
8 John CN et al Report of four cases of aneurysm complicating
patch aortoplasty for repair of coarctation of the aorta Aust
NZ J Surg 1989; 59(9): 748–750.
9 Fletcher SE et al Balloon angioplasty of native coarctation of
the aorta: midterm follow-up and prognostic factors J Am
Coll Cardiol 1995; 25(3): 730–734.
10 Anjos R et al Determinants of hemodynamic results of
bal-loon dilation of aortic recoarctation Am J Cardiol 1992; 69(6):
665–671.
11 Bonhoeffer P et al The multi-track angiography catheter: a
new tool for complex catheterisation in congenital heart
13 Choy M et al Paradoxical hypertension after repair of
coarc-tation of the aorta in children: balloon angioplasty versus
surgical repair Circulation 1987; 75(6): 1186–1191.
14 Ussia GP, Marasini M, and Pongiglione G Paraplegia ing percutaneous balloon angioplasty of aortic coarctation: a
follow-case report Catheter Cardiovasc Interv 2001; 54(4): 510–513.
Trang 9In most major pediatric centers balloon dilation of the
aor-tic valve in the cardiac catheterization laboratory is the
accepted standard for the primary treatment of aortic
valve stenosis Balloon dilation of aortic valve stenosis to
treat valvar aortic stenosis was first published in 19841
Balloon dilation of the aortic valve provides palliation that
is comparable to the palliation for similar aortic valve
stenosis achieved by a surgical aortic valvotomy, but
without the risks and morbidity of surgery2,3 Significant
improvements in the dilation balloons, guide wires and
techniques over the past 15 years have improved the
suc-cess rate and decreased, but not eliminated, the morbidity
and mortality of the aortic dilation procedure for infants,
children and adolescents The indications for dilation of
the aortic valve are similar to the indications for surgical
aortic valvotomy and, as with the indications for surgery,
the indications for balloon dilation vary with the age of
the patient
In the newborn the diagnosis of aortic valve stenosis
comprises a very heterogeneous spectrum of anatomy,
including everything from a nearly atretic aortic valve
with a small aortic annulus and/or an associated very
small, hypoplastic left ventricle, to an equally stenotic
valve, but with a large, dilated poorly functioning left
ventricle The exact anatomy and the resultant left
ventric-ular function determine the indications for balloon
dila-tion in this group The echocardiographic demonstradila-tion
of aortic valve stenosis with associated poor left
ventricu-lar function or low cardiac output, particuventricu-larly with an
otherwise “normal” sized left ventricle, is a major
indi-cation for intervention in a newborn regardless of the
measured gradient by either echo or catheterization4
After the clinical evaluation along with a quality
echocar-diogram, the definitive diagnosis of valvular aortic
steno-sis is established by the hemodynamics obtained in the
catheterization laboratory In the older infant and young
child with clinical findings of aortic valve stenosis, theindication for valvotomy is determined from the peak topeak hemodynamic gradient measured across the valve inthe catheterization laboratory In the absence of any signs
of “poor ventricular function”, aortic valve dilation is formed arbitrarily in very young children for a peak topeak gradient greater than 65 mmHg across the valve.Very young children do not participate in organized,severely strenuous, or sustained physical activities and donot create much additional gradient with their level of activ-ity In adolescent and adult patients, who are more likely
per-to participate in severely strenuous or sustained physicalactivity, symptoms referable to the heart or a measuredpeak to peak gradient across the valve of over 50 mmHg isthe arbitrary indication for valve dilation These criteria
were established for a surgical aortic valvotomy on the
basis of “natural history” studies, and probably are tooconservative for balloon dilation of the aortic valve
General anesthesia with intubation and controlled lation is used in newborns, in critically ill patients withaortic stenosis, and in any patient in whom the carotidartery approach is being used The same general proced-ure is used for the “diagnostic” cardiac catheterization of
venti-the patient who is not critically ill but who is undergoing
aortic dilation, as is used for the catheterization of anyother patient The catheterization is performed underdeep sedation and local xylocaine anesthesia with supple-mental sedation given intravenously periodically through-out the procedure A secure peripheral intravenous and
an arterial line are established and an indwelling “Foley”catheter placed in the bladder in all patients past infancywho are undergoing a balloon dilation of the aortic valve.There are several different approaches and techniquesutilized to accomplish balloon dilation of a stenotic aorticvalve The specific technique used depends upon the
19
Trang 10particular circumstances of each individual patient and
also on the individual preferences of the operator and/or
the catheterization laboratory The aortic valve can be
approached retrograde from the femoral, brachial,
um-bilical, or carotid arteries The aortic valve can also be
approached prograde from the right heart after passing
into the left heart through either a patent foramen ovale
or by crossing through the intact atrial septum with a
transseptal puncture The atrial transseptal approach to
the left atrium is the preferred technique to acquire the left
heart hemodynamics and angiography when the atrial
septum is intact, while the retrograde approach through
the femoral arteries is the most commonly used approach
for the actual balloon dilation of the aortic valve A
double-balloon dilation of the valve using a retrograde approach
from both femoral arteries is preferred for most aortic
valve dilations5
Technique
For the combined prograde and retrograde approach to
the aortic valve dilation procedure, a short venous sheath
is introduced into one femoral vein and two very small
indwelling arterial cannulae are introduced into both the
right and left femoral arteries The right heart
catheteriza-tion is performed using an angiographic “marker” catheter
introduced through the short venous sheath When a
transseptal procedure is to be performed and there are
any concerns about or a peculiarity of any part of the
anatomy of the left heart, an angiocardiogram is
per-formed with injection into the pulmonary artery before
the transseptal puncture The recirculation of the contrast
through the left atrium and left ventricle clearly
demon-strates the exact positions and any peculiarities of the left
heart anatomy
The left heart hemodynamics are obtained by means of
a prograde left heart catheterization either through a
pre-existing interatrial communication or through a
trans-septal atrial puncture Using the prograde approach to the
left heart, all of the hemodynamics, as well as quality,
select-ive left ventricular or aortic angiograms, are obtained
before any “time” is incurred in the arteries with the larger
indwelling arterial sheaths When the necessary right
heart information has been obtained, the prograde venous
catheter is advanced into the left atrium and from there
into the left ventricle Pressures are recorded from the left
ventricle and the femoral arteries, and the left ventricular
angiography is obtained to determine the severity and
type of the aortic stenosis
In the absence of a pre-existing atrial communication,
the right heart catheter and short venous sheath are
replaced with a transseptal set of the largest French size
the patient can accommodate comfortably and safely The
long sheath of the transseptal set should have an attachedback-bleed valve with a side arm/flush port A trans-septal left atrial puncture is performed using the longsheath/dilator transseptal set as described in detail inChapter 8 Once the long sheath is positioned in the leftatrium and the sheath and its back-bleed valve apparatusare cleared meticulously of all air and clots, the side port
of the sheath is attached to the pressure/flush system Atthis time in the procedure, the patient is given 100 mg/kg
of heparin through the long sheath An angiographic
catheter one French size smaller than the sheath is advanced
through the sheath and manipulated from the left atriuminto the left ventricle Simultaneous pressures are re-corded from the left atrium, left ventricle and a femoralartery With the knowledge that the femoral artery pres-sure can be as much as 20 millimeters higher than the aor-tic root pressure as a result of the elastic recoil of thesystemic vasculature, this measured difference in pres-sures between the left ventricle and femoral artery gives
an estimate, but does not give an accurate measurement of
the true transvalvular aortic gradient
Following the transseptal puncture and when the initialpressures have been recorded, a selective biplane angio-cardiogram is performed with an injection into the leftventricle The angiocardiogram defines the preciseanatomy of the aortic root and the aortic valve stenosis,and demonstrates any associated left ventricular outflowtrack abnormalities At least one view of the angiocardio-grams should be as close to perpendicular to the valveannulus as possible The lateral (LAT) X-ray tube is placed
in a 60° left anterior oblique position with between 30° and60° cranial angulation The amount of cranial angulation
is determined by how horizontally the heart is situated
in the chestathe more horizontally the heart lies in the
thorax, the greater the cranial angulation The posterior–anterior (PA) X-ray tube is placed in a 30°, right anterioroblique position with 30+° of caudal angulation The PA
tube should be almost perpendicular to the LAT tube inboth planes If the valve is not cut precisely on edge withthe initial picture, the X-ray tubes are rotated appropri-ately and the angiocardiogram repeated Very accurate
angiographic measurements are made of the diameter of the
valve annulus at the base of the aortic sinuses where thevalve leaflets attach or “hinge” in the annulus The meas-urements are obtained from a systolic frame where theleaflets are open and the annulus is at its largest diameterduring the cardiac cycle The measurements must be cali-brated against a valid reference measurement system.Depending upon the measurement system in the particu-
lar laboratory, an exact determination of the actual
dia-meter of the valve annulus is made or calculated utilizing an accurate reference system for calibration of the measure-
ments As discussed earlier in Chapter 11, the use of thediameter of a catheter as the reference measurement is not
Trang 11satisfactory, and, in fact, creates dangerous errors when
measuring large structures such as valves and large
ves-sels A calibrated “marker” catheter, which is positioned
exactly in the angiographic field, is frequently used and
represents a very accurate reference system for
calibra-tion The measuring “bands” on the calibrated catheter are
placed in the plane of the valve and aligned exactly
per-pendicular to the left ventricular outflow tract during the
injection for the valve measurement The catheter tip can
be either in the aortic root, left ventricular outflow tract, or
even in the left ventricular apex Often the catheter moves
during a pressure injection and as a consequence of this
movement, the “marker bands” on the catheter will
usu-ally align precisely on edge for the calibration
measure-ments during at least several frames of the angiogram If
not, the catheter is repositioned to align the marks on edge
before either the X-ray tubes, the table or the patient are
moved and a very brief “cine” of the exactly aligned marks
is recorded with the new catheter position
As an alternative, a “marker” catheter can be positioned
in the superior vena cava immediately adjacent to the
aor-tic root, and in this position it serves as a precise
calibra-tion reference in the same reference plane as the aortic
valve The separate marker catheter in the superior vena
cava, simultaneously with the use of the transseptal
tech-nique, necessitates the use of an additional venous line for
the marker catheter Similarly, a marker catheter can be
introduced into the esophagus with the “marks”
posi-tioned in a location immediately behind the cardiac
silhouette and adjacent to the aortic valve and, in that
position, used as an accurate reference system for
calibra-tion Some X-ray systems have calibration marks
embed-ded in the image intensifier screen The changes in the
distance of these marks to the “isocenter” of the image are
computed constantly and accurately by the sophisticated
computer system, which allows the use of these marks for
a very accurate calibration reference regardless of the
position and distance of the intensifier When such a
built-in calibration reference system is available and it is
verified against a grid or marker catheter reference
sys-tem, it is the most convenient and accurate calibration
system available
An external grid or a metal “sphere” of a known, precise
diameter placed in the exact plane of the valve as described in
Chapter 11 (Angiography) can also be used as the accurate
calibration reference The external grid or sphere is a very
accurate but very inconvenient reference system
Once the measurements of the valve annulus have been
completed, the X-ray tubes are placed in the positions that
will be utilized during the valve dilation Biplane records
of the best images of the valve from both planes of the
biplane angiograms in this same angulation are stored in a
“freeze frame” to be used as a “road map” during the
valve dilation If there is not a good “freeze frame” replay
capability available, it is useful to place tiny lead markers
on the chest wall exactly over the area of the aortic valveannulus on the fluoroscopic image Fluoroscopy is used toalign the “lead” marks on the chest wall according to thelocation of the valve as seen on the previous angiocardio-grams in the same X-ray views
The long transseptal sheath is advanced over thecatheter and into the left ventricle An “active deflectorwire” is introduced through a wire back-bleed/flushvalve into the original angiographic catheter and thecatheter is deflected 180° toward the aortic valve Whileholding the curve on the deflector wire and fixing the wire
in place, the catheter is advanced off the wire, through the
left ventricular outflow tract, and into the aorta This isaccomplished readily when an end-hole or angiographic
woven dacron catheter is used These catheters become
very soft and flexible after only a few minutes in the lation at body temperature, and as a consequence are relat-ively atraumatic and easily deflected
circu-An alternative technique for advancing a catheter fromthe left ventricle into the aorta is to use a floating ballooncatheter through the long sheath, which is positioned inthe left ventricle The original angiographic catheter isremoved from the long sheath while the long sheath ismaintained in position in the left ventricle The longsheath is cleared thoroughly of all air and/or clot The dis-tal end of a floating balloon angiographic catheter, which
is at least one French size smaller than the sheath, is “pulled”
or softened in heat and then formed into at least a 180°curve at the distal end The balloon angiographic catheter
is introduced into the long sheath and advanced throughthe sheath and into the left ventricle During the introduc-tion and the entire time the balloon catheter is beingadvanced within the sheath, both the sheath and the
catheter are maintained on a constant flush Once the
bal-loon tip is beyond the sheath in the left ventricle, the balloon catheter and sheath are switched to pressure monitoring, the balloon is inflated with CO2, and usually,using minimal manipulations, the balloon catheter floatsout of the left ventricle, across the stenotic valve, and intothe ascending aorta In the presence of a very tight steno-sis, the balloon must often be deflated partially, or evencompletely, just beneath the valve in order for the catheter
to pass through the stenotic orifice Occasionally the
“active deflector wire” is used along with the floating loon catheter in order to redirect the catheter tip 180° awayfrom the apex and toward the valve
bal-Once a catheter has passed into the aorta, simultaneousleft ventricular, aortic root, and femoral artery pressuresare recorded, with the left ventricle pressure recordedfrom the side arm of the sheath, the ascending aortic pres-sure from the catheter, and the femoral pressure from the femoral line These pressures provide the true peak-
to-peak hemodynamic gradient across the valve and a
Trang 12comparison with the original left ventricle to femoral
artery pressures In order to obtain even better
visualiza-tion of the valve leaflets, an aortic root aortogram is
recorded through this prograde catheter The soft catheter
alone passing through the valve does not produce
significant artifactual aortic regurgitation When the
aor-tic root angiogram has been recorded, the prograde
catheter is advanced further out of the ventricle and the
tip advanced around the arch and into the descending
aorta (at least beyond the brachiocephalic trunk!) When
the hemodynamics have been recorded, both the catheter
and the transseptal sheath are placed on a slow
continu-ous flush The transseptal sheath remains in the left
vent-ricle and the catheter in the descending aorta before, during
and after the dilation procedure This allows
simultane-ous aortic and left ventricular pressure monitoring and
recording during and immediately after the dilation
proced-ure In addition, both the catheter in the aorta and the
transseptal sheath in the left ventricle serve as routes
for rapid infusions of intracardiac fluids or medications
Of equal importance, the shaft of the catheter that is
pass-ing through the stenotic orifice of the aortic valve serves
as a visible and definitive “guide” for the subsequent
introduction of the retrograde wires/catheters across the
valve
With the prograde catheters in place and the
measure-ments of the valve completed, the balloons for the dilation
are chosen and prepared Extra long balloons are used for
balloon dilation of the aortic valve For smaller patients
(and usually smaller 10 and 12 mm diameter balloons), a
4 cm balloon length is used, and for larger patients (and
usually larger balloon diameters), 6–8 cm long balloons
are used when available The combination of the
double-balloon technique, the use of super stiff wires, and the use
of longer balloons, virtually eliminates the problem of the
balloons squirting or bouncing in and out of the valve and
further damaging the valve during the balloon inflations
The combined inflated diameters of the two balloons for a
double-balloon dilation is equal to a maximum of 1.2 to 1.3
times the maximum accurately measured diameter of the
aortic annulus If a single-balloon technique is used, the
single balloon diameter to start with is no more than 0.9 to
1 times the diameter of the accurately measured annulus
For the usual balloon dilation of the aortic valve, the
balloons undergo a standard preparation in order to clear
them completely of any air before introduction into the
sys-temic circulation A “minimal balloon preparation”
simi-lar to the balloon preparation in very small and/or
critically ill infants is used when there is concern that the
standard balloon prep would increase the deflated
bal-loon profile enough to necessitate a larger arterial sheath
The “minimal balloon preparation” is intended to clear a
balloon of air completely, but at the same time not to
“unfold” it from its “factory wrap” around the catheter
With the prograde aortic catheter in place across thevalve and the dilation balloon(s) prepared, the inguinalareas around the arterial puncture sites are re-infiltratedliberally with 2% xylocaine The indwelling arterial can-nulae are replaced with sheath/dilator sets that willaccommodate the balloons chosen for the dilation Anend-hole catheter that accommodates a 0.035″ guide wire
is passed retrograde around the arch and into the aorticroot from one of the arterial sheaths
Crossing the stenotic aortic valve with the wire(s) andthen positioning the catheter(s)/wire(s) securely andsafely in the left ventricle is often the most difficult part ofthe aortic valve dilation procedure The aortic root is oftenvery dilated and distorted, with the stenotic orifice of thevalve located very eccentrically
The course of the prograde catheter passing through thestenotic valve from the left ventricle into the aorta pro-vides the most reliable technique for identifying the exactposition and “direction” of the stenotic orifice, and is avaluable asset for crossing the stenotic aortic valve fromthe retrograde approach The course of the catheter pass-ing through the valve orifice provides a visible “guide” tothe actual course from the left ventricle, through the nar-row and otherwise invisible orifice, and into the aorta
In the absence of a prograde catheter through the valve,the valve leaflets themselves and the valve orifice are
“invisible” on the fluoroscopy and are only identified bycomparing the fluoroscopic image to the “jet” of contrastpassing through the orifice on the previous angiographicrecording The “freeze frame” of that recording is used as
a guide in “finding” the valve orifice during the grade approach
retro-A torque-control catheter with a preformed, slightlycurved tip, together with a steerable wire with a slightlycurved soft tip, are used to maneuver across the valve Apreformed right or left coronary catheter is most useful fordirecting the tip of the wire toward the aortic valve orifice.Occasionally the preformed retrograde catheter itself can
be directed purposefully and precisely along the course ofthe prograde catheter and through the valve orifice.Usually crossing the stenotic orifice requires the use of thecombination of several catheters and wires, even with aprograde catheter serving as a guide Rotating the shaft
of a catheter clockwise or counterclockwise with the preformed tip positioned in the aortic root, moves the
catheter tip (and wire) anteriorly or posteriorly Moving
the shaft of the catheter forward and backward moves the curved tip of the catheter (and, in turn, the wire tip)
from side to side in the aortic root.
The prograde catheter passing through the valve orificeprovides the exact “route” through the valve orifice,which is visualized simultaneously as the wire is manipu-lated through the retrograde catheter The combination
of the known location of the orifice, which is delineated
Trang 13precisely by the course of the prograde catheter through
the orifice, and the precise control over the direction of the
wire tip, allows the operator to maneuver the tip of the
wire purposefully, precisely along and immediately
adja-cent to the prograde catheter In contrast to the rapid,
repeated maneuvers used during a “blind” probing with a
wire at a stenotic aortic valve, the maneuvers with the
catheter and the wire when “following along the prograde
catheter” are carried out slowly and purposefully The tip
of the wire is observed frequently on both PA and lateral
fluoroscopy as the wire is “tracked along” the prograde
catheter This maneuvering of the wire into the ventricle is
still not “automatic” and may take several attempts, each
with a readjustment of the angle of approach of the wire to
the valve Occasionally, when the angle of the valve orifice
is very distorted, it is necessary to change to a catheter
with a completely different preformed curve at the tip in
order to angle the wire properly at the orifice A very soft
tipped wire can create a problem with this technique
With a very tight stenosis, the high velocity of the jet of
blood through the orifice, literally, blows the tip of a very
soft tipped wire away from the orifice When this problem
occurs repeatedly, a wire with a slightly stiffer tip is used
to accomplish the retrograde crossing of the valve The
technique of tracking along the “adjunct” prograde
catheter to guide the retrograde wire through the valve
represents the most definitive technique available for
cross-ing the stenotic aortic valve from a retrograde approach
Some operators prefer not to go through the extra
maneuvering for the adjunct prograde catheter technique,
and occasionally the prograde catheter from the left
ventricle into the aorta cannot be used or accomplished,
especially in very small, very sick infants Various other
“tricks” are described for crossing the stenotic aortic
valves from the retrograde approach without the adjunct
prograde catheter Most of the techniques first involve the
positioning of a precurved, end-hole catheter retrograde
into the aortic root An attempt is often made using the
retrograde catheter itself to probe at the “invisible” orifice
A soft tipped catheter is “bounced” gently, rapidly and
repeatedly off the valve and occasionally even “backs”
through a tight orifice with a 180° loop formed on the
catheter A catheter with a stiff tip should never be used
with this technique A stiff catheter tip repeatedly
bounced off of a stenotic valve can easily create an orifice by
perforating a leaflet!
When there is no prograde catheter through the valve, a
very soft tip, small diameter, spring guide wire is
maneu-vered toward the valve through a precurved, end-hole,
retrograde catheter, somewhat similar to the catheter/
wire maneuvers just described The main difference is that
the wire is advanced rapidly and repeatedly, in and out of
the catheter tip while the angle of the catheter tip is
changed repeatedly This, in turn, bounces, or backs the
wire tip in and out of the aortic sinuses until the tip or aloop of the wire eventually falls (somewhat accidentally)through the valve opening Since the valve leaflets and the
orifice are not visible at all, this requires rapid, but patient,
repetitions rather than any particular skill The more
repetitive “probes” at the valveaeach with a change in the
direction of the wire by changing the catheter tip position
athe more likely is the chance of crossing the valve This
is the opposite of tracking along a known course of a prograde catheter through the valve and is an instancewhere slow, meticulous catheter and/or wire manipula-
tions definitely are not indicated Slow maneuvers in this
situation increase the radiation used and, at the sametime, decrease the “chances” of hitting the valve orificeper unit time of fluoroscopy used If one particularwire/catheter combination does not accomplish crossingthe valve after several minutes of rapid “probing”, thewire and/or the catheter is/are exchanged A similar
“probing” at the valve with the new catheter/wire bination is used
com-Although a very soft tipped wire is unlikely to perforate
a valve cusp, there are potential problems associated with
the rapid probing technique With the rapid in and outmovements of the wire, occasionally the wire dropsthrough the valve, but is pulled back and out of the ventri-cle before the position is recognized because of the rapid-ity of the maneuvering This is even more likely in thepresence of a markedly distorted aortic root/valve or anunusual course into the ventricle A more serious problem
is the inadvertent and unrecognized cannulation of acoronary artery with the wire/catheter As the aortic root
is probed, the wire passes from the coronary sinus into
a coronary artery often more easily than through thestenotic valve In the presence of a markedly distorted aor-tic root, the abnormal wire position in the coronary arterycan go unnoticed When the soft tipped wire only isadvanced into the coronary artery, it usually does not create a problem, but if the wire is advanced far into thecoronary, particularly if the catheter is advanced over thewire deep into the coronary thinking it is in the ventricle,the coronary artery can be damaged The treatment forthis potential problem is awareness and prevention.Perforation of an aortic valve leaflet during the retro-grade probing of an aortic valve is a constant potential
problem When the tip of the retrograde catheter is deep into, or actually becomes buried in the aortic sinus and
then a wire is advanced out of the catheter tip, the wiredoes not have room to “buckle” or bend and the wire isthen more likely to perforate directly through the valve
tissue in front of itaparticularly the tissue of a thin valve
leaflet If a valve leaflet is perforated by the wire and then
a dilation balloon is passed over the wire through the foration, the valve will be destroyed rather than dilated!The stiffer the wire and/or the catheter that is used for the
Trang 14per-retrograde probing of the aortic valve, the less ability the
wire and/or catheter has of “buckling away” from the
aortic sinuses and the greater the potential for this type of
leaflet perforation Even the “soft” tip of a “standard”
0.025″ spring guide wire is very stiff for the first few
mil-limeters as it extends out of the tip of a catheter
A Terumo™ (or Glide™) wire creates an equal or
prob-ably greater problem of perforation of the aortic leaflets
when it is used to pass retrograde across an aortic valve
Terumo™ wires become much stiffer and much sharper
when they have no room to “buckle” between the tip of
the catheter and the structure/tissue it is attempting
to cross The Terumo™ wire also “glides” through the
tissues easily once it does perforate, and as a consequence
an abnormal course of the wire within tissues can go
unrecognized
The “blind” retrograde probing technique for crossing
the valve usually and eventually is successful, however, it
has obvious drawbacks Crossing the valve depends more
on chance and multiple repeated attempts than on any
skill The use of biplane imaging to help “identify” the
location of the orifice in “three dimensions” is essential to
the success of this as well as all of the other techniques for
crossing the aortic valve Even when the valve, and
particu-larly the orifice of the valve, are not visible, the second
plane of fluoroscopy allows the operator to know when
the catheter/wire is repeatedly probing in a totally
inap-propriate or non-productive area of a valve sinus
Once the left ventricle has been entered with the
retro-grade catheter and/or wire using whatever combination
of retrograde wires, special catheters or special curves
happen to work most effectively in the particular patient,
a soft, curved tipped, spring guide wire is advanced
retro-grade as far as possible into the ventricle When this wire
has been secured in the ventricle, an angled, end-hole
catheter is advanced over the wire into the ventricle With
a combination of pushing and backing maneuvers of both
the wire and the catheter, the tip of the retrograde catheter
in the ventricle is deflected and directed 180° back toward
the aortic valve When the angled end-hole catheter
can-not be directed fairly expeditiously back toward the aortic
valve, the original angled catheter is exchanged for a
pig-tail catheter over the original wire positioned in the left
ventricle Once the pig-tail catheter passes over the wire
and the pig-tail curve is positioned securely in the left
ven-tricular apex, the original wire is removed and replaced
with the larger, long floppy tipped, preformed, stiff,
exchange length wire, which should almost automatically
point toward the left ventricular outflow tract when it is
advanced out of the catheter tip
The largest diameter exchange wire that the balloon
catheters chosen for the dilation will accommodate is
introduced at this time In larger patients, the exchange
wire is a 0.035″, Super Stiff™ exchange length wire with
a long floppy tip This wire should have a “J” or, even, a
> 360° pig-tail curve formed manually at its distal floppy tipand a second more proximal, smooth, but short 180° curveformed at the “transition area” between the rigid part andthe floppy portion of the wire The preformed 180° curve
at the transition area of the wire will allow the stiff portion
of the wire to be completely across the valve when the180° curve at the transition area is seated in the apex of theventricle Either the original retrograde, angled, end-holecatheter looping 180° toward the outflow tract or the pig-tail catheter in the apex of the left ventricle directs the tip
of the Super Stiff™ wire back toward the left ventricularoutflow tract and allows the 180° curve on the stiff portion
of the wire to “seat” in the apex of the left ventricle The tip
of the wire is then directed back toward the LV outflowtract or even back across the aortic valve
When the double-balloon aortic dilation technique isused, the catheter is left in position over the first wire and a slow flush is maintained around the wire while thesecond wire/catheter is positioned5 Once the first stiffwire is in a stable position, the opposite groin is infiltratedliberally with xylocaine A second sheath/dilator set isintroduced into the opposite femoral artery An angled,end-hole catheter, identical to the catheter which was ini-tially used to cross the valve, is introduced into the secondarterial sheath and passed retrograde around the arch.The second catheter is maneuvered retrograde throughthe orifice and into the left ventricle “following” the firstcatheter/wire precisely while using the course of the firstwire/catheter through the valve orifice to serve as a vis-ible guide for introducing the second wire even if anadjunct prograde catheter was not used Once the secondwire has passed into the ventricle, the angled end-holecatheter is advanced into the ventricle and curved backtoward the outflow tract similar to the positioning of thefirst catheter When an 180° curve cannot be formed in theventricle with the angled end-hole retrograde catheter, it
is replaced with a pig-tail catheter, which is positioned inthe left ventricular apex adjacent to the first wire A sec-ond teflon-coated stiff exchange wire is pre-shaped, ident-ical to the first exchange wire, passed through the secondcatheter, and positioned in the left ventricle adjacent toand in a similar position to the first wire
Once the two stiff wires are in a satisfactory, stable position within the ventricle, the end-hole catheters areremoved over the wires and the previously prepared bal-loons are introduced over the wires It is important thatthe wires are observed continuously and maintained pre-cisely and securely in their positions during the introduc-tion of the balloons through the sheath and during theprocess of advancing the balloon catheters retrogradearound the arch The portions of the wires that are outside
of the body are maintained straight and are fixed against afirm structure on the table (or against the patient’s leg)
Trang 15while the wires and the balloon catheters within the body
are observed frequently with fluoroscopy The wires
out-side of the body are adjusted in or out slightly in order to
keep the left ventricular wire loops from being withdrawn
from, or pushed further into (through!) the ventricle
dur-ing the various manipulations of the balloon catheters
This is important particularly when the balloon catheters
are being maneuvered next to each other The balloons are
maneuvered over the wires to a position just above the
aortic valve The two wires are advanced toward the apex
of the ventricle until all “slack” is out of the wires and the
wires are pushed against the outer circumference of the
course from the descending aorta, around the aortic arch,
and into the ascending aorta
Once all of the preparations are ready for the controlled
pressure inflation of the two balloons, the two wires are
fixed in position and the balloons are advanced one at a
time over the wires and across the aortic valve The
posi-tion of the balloons in the aortic valve is compared with
the “freeze frame” image of the valve With the balloons in
their proper position, the area of the valve leaflets should
be exactly in the center of the two markers at the end of
each balloon This positions the center of the parallel
sur-faces of the balloons at the narrowest portion of the valve
If the balloons are unstable in the initial position across the
valve, they are advanced further over the wires toward
the apex of the ventricle while maintaining the parallel
walls of the balloons in the narrow portion of the valve
and keeping the curved wires in the apex of the left
vent-ricle This readjustment is easier and safer when using
longer dilation balloons If a balloon is too far within the
ventricle, the wire is advanced gently in order to push the
balloon back rather than withdrawing the balloon This
maneuver maintains the wires at the “outer
circumfer-ence” of the course around the arch and maintains better
control over the wires
In very severe valve obstruction, the passage of the first
deflated balloon across the narrow orifice can cause a
sud-den and dramatic deterioration in the patient’s
hemody-namics In this circumstance, the first balloon is positioned
rapidly in the valve, inflated, deflated and withdrawn
back into the aortic root The patient is allowed to stabilize
(or is resuscitated) before attempting to position the
sec-ond balloon The subsequent positioning of the two
bal-loons is accomplished as rapidly as possible and followed
immediately by simultaneous rapid inflation/deflation of
the two balloons All inflations/deflations of the balloons
during a dilation procedure are recorded on either biplane
angiography or “stored fluoroscopy”
During the dilation, the balloons are inflated until the
indentations or “waists” in the balloons caused by the
stenotic valve disappear or until the balloons reach their
maximum advertised pressure, whichever comes first As
soon as the balloons reach this full inflation, they are
deflated as rapidly as possible During the balloon inflation
in the aortic valve, almost total obstruction of all cardiacoutput is created by the fully inflated balloon(s) The heartrate slows, the systemic arterial pressure (monitoredthrough the prograde arterial catheter still passingthrough the valve or through the side arm of one of thearterial sheaths) drops, and the left ventricular pressure(monitored through the transseptal sheath still positioned
in the ventricle) increases markedly With a successful
dila-tion of the valve, these hemodynamic parameters rapidly
return to “normal” once the balloons are deflated, ever, the balloons should be withdrawn rapidly over thewires and back into the aortic root to be sure of, and tofacilitate, this recovery
how-The inflation/deflation of the balloons is repeated eral times, changing the positions of the balloons beforeeach subsequent inflation by moving them in and out ofthe valve slightly and, if possible, changing their relativeanterior/posterior/lateral relationships in the process.With the pressures available through the long sheath inthe left ventricle and through the prograde catheter posi-tioned in the aorta, the hemodynamic results of the dila-tion are known as soon as the balloons have been deflatedand without moving the balloon catheters The end point
sev-of the dilation is the absence sev-of any “waist” on the
bal-loons during the initial phase of subsequent inflations and
a reduction of the gradient across the valve to less than
25 mmHg The left ventricular pressure should approachnormal or near normal as monitored by the indwellingtransseptal sheath(s)
Occasionally, the heart rate and blood pressures do notreturn to normal very rapidly after the inflation/deflation.When this return of heart rate/blood pressure is slow, orappears non-existent, if not already accomplished, the bal-loons are withdrawn over the wires, out of, and awayfrom, the valve The patient is given atropine through theleft ventricle sheath and if necessary given several exter-nal manual compressions to stimulate cardiac function.With even partial relief of the aortic obstruction the heartrate and pressure do return
Before making the final post-dilation hemodynamic
measurements, both balloons are withdrawn back overthe wires into at least the descending aorta so that the only
“equipment” passing through the valve is the two grade wires and the prograde balloon catheter, which ispositioned in the aorta The residual gradient across thevalve is recorded from the pressure in the left ventricle(through the sheath) and the pressure in the aorta(through the prograde catheter) and/or one of the femoralartery pressures through the side arm of the femoralartery sheath The pressure gradient and the contour ofthe pulse curves give an excellent immediate indication ofthe results of the procedure before the removal of anycatheters or wires With a very successful relief of the
Trang 16retro-obstruction, the left ventricular systolic pressure
approx-imates the aortic systolic pressure, the left ventricular
end-diastolic pressure is less than 15 mmHg, the gradient
across the aortic valve is abolished or reduced to less than
20 mmHg and there is little or no aortic valve regurgitation
When mild, or no aortic regurgitation is produced, the
pulse pressure is 25–30 mmHg even with the prograde
catheter and the wires still across the valve An aortic root
angiocardiogram to assess aortic regurgitation is recorded
with injection through the prograde catheter This
injec-tion helps to assess the aortic regurgitainjec-tion, particularly if
there is a wide pulse pressure If there is significant aortic
regurgitation but good relief of the obstruction, nothing
further therapeutic is performed in the catheterization
laboratory at this time If there is a significant residual
gradi-ent and significant aortic regurgitation as demonstrated by
the post-dilation pulse pressure or the aortogram, the
wires are removed from across the aortic valve in order
to assess the aortic regurgitation more accurately
Occa-sionally, a stiff wire is forced against one side of the
annu-lus and “holds the valve open”, creating false aortic valve
regurgitation Once the wires across the annulus have
been withdrawn, the aortic regurgitation is reassessed
with repeat pressure measurements and a repeat aortic
root aortogram The prograde catheter passing through
the valve alone does not, or very rarely, produce
signi-ficant aortic regurgitation However, if signisigni-ficant aortic
valve regurgitation is demonstrated from the injection
through the prograde catheter, the prograde catheter is
withdrawn into the ventricle and a repeat aortic root
angiocardiogram performed with an injection in the aortic
root through a retrograde catheter If there is still
signific-ant aortic regurgitation, regardless of any residual gradient,
no further dilation is performed
If dilation with the original balloons did not produce
sufficient relief of the obstruction, but also did not produce
significant aortic valve regurgitation, the angiographic
images of the balloon inflations and the post-dilation
angiograms are reviewed The diameters of the inflated
balloon(s) and valve annulus are re-measured accurately
on these angiograms of the inflations If angiographically
the balloon diameters are the same or smaller than the
diameter of the annulus, the combined balloon diameters
are increased up to 1.2 times the annulus diameter and the
dilation repeated If the balloon diameters are 1.2 times (or
more) larger than the aortic annulus on the angiographic
images of the balloon inflation, no further dilation is
attempted If a single balloon is used for the dilation and
the balloon is exactly the same diameter as the annulus on
the inflation images, the balloon diameter is increased by
at most 10% and the dilation is repeated
When the transseptal sheath and prograde catheter
were not used during the procedure, there are several
alternative techniques to record the postdilation pressures
before the wires are removed A separate, new transseptal
procedure can be performed to enter the left heart and leftventricle The left ventricular pressure through thetransseptal catheter is compared with the femoral arterypressure measured through the side arm of one of thefemoral artery sheaths Another alternative technique is toremove one balloon dilation catheter over the wire Thewire is maintained in the left ventricle and the balloondilation catheter is replaced over this wire with a Multi-Track™ catheter The Multi-Track™ catheter is advancedretrograde over the wire all the way to the left ventricle.Pressures from the Multi-Track™ catheter can berecorded from any location along the course of the wireduring either the introduction or the withdrawal of theMulti-Track™ catheter and with the wire remaining inplace Angiograms can also be recorded at any locationthrough an angiographic Multi-Track™ catheter By usingthe Multi-Track™ catheter, the retrograde wire is still inplace in the ventricle without any further manipulations
of the wire if a new balloon dilation catheter must be duced for a repeat dilation
intro-The third, and least attractive alternative to either ofthese previous techniques for reassessment of the hemo-dynamics post-dilation is to withdraw one of the retro-grade balloon dilation catheters over the wire andcompletely out of the body while leaving the wire in theleft ventricle The balloon catheter is replaced with an end-hole diagnostic catheter, which is passed over the wire,retrograde, all of the way to the left ventricle The wire isthen removed completely The left ventricular pressure ismeasured through this catheter and compared with thefemoral arterial pressure measured through the side arm
of the other arterial sheath If the simultaneous pressuresfollowing the dilation demonstrate an unsatisfactoryrelief of the obstruction, an aortic root aortogram to assess
aortic regurgitation cannot be obtained without
with-drawing this catheter out of its position in the ventricle
Once the catheter is withdrawn and if the dilation does
need to be repeated, one retrograde wire still is passingthrough the valve and into the ventricle This wire serves
as a “guide” along which the wire/catheter is directedduring the reintroduction of the retrograde catheter backinto the ventricle Once the catheter is positioned back
in the ventricle, the stiff exchange wire is re-advanced
through this catheter and repositioned in the ventricle for
introduction of the new balloons
Once it is determined that a satisfactory relief of the aortic obstruction has been obtained, the balloon catheters are withdrawn slowly over the wires and out of the vessels As they are withdrawn, the balloon cathetersare rotated in the direction of the balloon folds and
the negative pressure to the balloons is released during
the withdrawal through the aorta and out through thesheaths
Trang 17Occasionally the retrograde wires, with their
pre-formed loops and curves, become entangled in the
ven-tricular structures Once the balloon catheters have been
withdrawn out of the sheaths, end-hole catheters are
advanced over the wires and into the ventricle While
observing on fluoroscopy, the wires are removed through
these catheters Withdrawing them through catheters
pre-vents the wires from damaging or disrupting chordae or
papillary muscles around which they may have been
entrapped
There is an additional technique for the guaranteed
introduction of a balloon dilation catheter retrograde
across even the most stenotic aortic valve This technique
is used only when the aortic valve is particularly difficult
or impossible to cross with any of the previous retrograde
catheter/wire techniques This technique is most suitable
for a single-balloon dilation technique but can be used for
a dilation using double balloons The technique is a
varia-tion of the use of the prograde catheter along with the
ret-rograde technique, but does add some additional complex
maneuvering to the procedure An end-hole, floating
bal-loon catheter is introduced prograde into the left ventricle
through a long transseptal sheath The end-hole balloon
catheter is advanced (floated) prograde across the aortic
valve and into the descending aorta An extra long (400 cm),
exchange length, spring guide wire is passed through
the prograde catheter and into the descending aorta A
snare catheter is introduced retrograde from a femoral
artery and the distal end of the prograde wire, which was
advanced through the prograde catheter, is snared with
the retrograde snare and withdrawn out through the
femoral artery sheath This produces a “through and
through” wire from the femoral vein to the right atrium,
left atrium, left ventricle, through the aortic valve, into the
aorta, and out through the femoral artery The desired
bal-loon dilation catheter is introduced over the arterial end of
the wire, through an arterial sheath, advanced retrograde
over the through and through wire and through the aortic
valve For a double-balloon technique this procedure is
repeated from the other femoral artery and a second
femoral vein The position(s) of the balloon(s) in the aortic
valve is/are adjusted with the catheters and wires and the
dilation of the valve carried out similarly to any other
bal-loon dilation of the aortic valve
This technique guarantees that the balloon dilation
catheters cross the aortic valve orifice and provides the
maximum control over the position and movement of the
balloons during the inflation/deflation At the same time
it does add some complexity to the procedure The
femoral artery and left ventricular pressures can be
mea-sured simultaneously before, during and immediately
after the dilation through a side arm of the transseptal
sheath and a side arm of the arterial introductory sheath
since, in order to accommodate the passage of even the
deflated dilation balloon, the sheath lumen is always
larger than the shaft of the balloon dilation catheter.
With the through and through wires in place, the loon dilation catheters are easily exchanged or removed.Once the dilation of the aortic valve is completed success-fully, the balloon dilation catheters are withdrawn overthe wires and replaced with end-hole catheters, which areadvanced retrograde over the wires and into the left ven-
bal-tricular apex Once the catheters are in place over the wires
the wires are withdrawn through the long sheaths fromthe venous end The retrograde catheters prevent the stiffwires from cutting intracardiac structures when they
“tighten” as they are withdrawn
When the catheters and wires have been withdrawn out
of the sheaths, more local anesthesia is administered ally around the sheaths at the arterial puncture sites, andthe sheaths are withdrawn Direct manual pressure isapplied over the arterial puncture sites to stop local bleed-ing Either the dorsalis pedis or the posterior tibial pulse inthe same extremity as the puncture is palpated simultane-ously while holding pressure on the artery, being sure thatthe peripheral pulse is not obliterated by the pressureapplied to obtain the hemostasis Pressure is maintainedbalancing the control of bleeding vs the peripheral pulseuntil hemostasis is achieved The systemic heparin is notusually reversed, so achieving hemostasis safely after theremoval of a large sheath can take up to an hour or more
liber-After hemostasis is achieved, a very light bandage just to
cover the puncture site is applied The bleeding should becontrolled before the bandage is applied A tight compres-
sion pressure bandage is not used over the artery The
bandage is left in place for 4–6 hours The patient isobserved in a monitored recovery bed for a minimum offour hours and until they are fully awake Particular atten-tion is paid to the puncture site and the pulses in the
extremity peripheral to the puncture site The “Foley”
catheter is left in place until males are awake and untilfemales are moved out of the recovery ward Patients whoundergo aortic valve dilation are observed in the hospitalovernight On the assumption that all of the adrenergicstress and any volume load from the catheterization aredissipated by the following morning, these patients usu-ally undergo an echocardiogram before discharge on theday following the catheterization This echo informationserves as the baseline, non-invasive value for subsequentfollow-up evaluations The patient is discharged with no
imposed limitations because of the catheterization.
Infant and/or critical aortic stenosis dilation
The diagnosis of aortic stenosis is made from the clinicalfindings and is based more on the appearance of the valve
by echocardiogram and the associated poor left lar function rather than on a “gradient”4 Dilation of aortic
Trang 18ventricu-stenosis in the newborn or small infant is one of, if not
the most difficult and dangerous procedures performed
by pediatric interventional cardiologists The dilation of
infant aortic stenosis is not just a “small version” of any
other aortic valve dilation The newborn or small infant
requiring treatment of aortic stenosis is usually very sick,
often is on prostaglandin and requires inotropic support
before and during the valve dilation procedure The
vas-cular access is small, partivas-cularly in proportion to the
sheaths, catheters, wires and balloons that are available
for small infants
These patients usually do have a patent interatrial
com-munication allowing the introduction of a prograde
venous catheter into the left ventricle A left ventricular
catheter remaining in place throughout the procedure is
invaluable for pressure monitoring and, occasionally, for
a left ventricular angiocardiogram In the very sick and
small infant, once the left ventricle is entered with the
pro-grade catheter, this catheter is not advanced further into
the aorta at this time unless a prograde approach for the
dilation is anticipated When there is no naturally
occur-ring atrial communication, the decision is made at the
onset of the procedure as to whether acquiring the
hemo-dynamic information from the left ventricle before it is
entered retrograde is worth the extra effort and slightly
increased risk of a transseptal puncture in a very small
infant
With meticulous attention to detail, patience, very
delic-ate technique, and some special instrumentation,
percu-taneous entry into a femoral artery essentially always is
accomplished, even in small infants Small, sick infants are
the patients in whom the special 21-gauge percutaneous
needles and the extra floppy tipped 0.018″ wires are
invaluable for entering the artery (Chapter 4) A
“single-wall” puncture technique is always attempted Once the
artery is cannulated with the guide wire, the surrounding
area to the puncture site is re-infiltrated liberally with
local anesthesia A 3- or 4-French (depending upon the
balloon to be used), fine or “feathered” tipped sheath/
dilator set is introduced into the artery Once the sheath is
secured in the artery and cleared of any air, an end-hole,
pre-curved right coronary catheter is introduced through
the sheath and advanced retrograde to the aortic root
The aortic pressures and aortic root angiocardiograms
are performed with the retrograde catheter This catheter
allows baseline pressure measurements, and satisfactory
aortic root aortogram(s) can be obtained with end-hole
catheters in these small infants At the same time, an
angled, end-hole catheter is the most useful for
manipula-tion of the guide wire across the stenotic valve for the
sub-sequent dilation, all without an unnecessary exchange of
catheters Only one arterial catheter is necessary since a
single-balloon, retrograde approach is used for dilation of
the aortic valve in small infants However, if the other
femoral artery is entered inadvertently during the ture of the vessels, it is cannulated with a 20-Gauge, teflonQuick-Cath™ for continual arterial monitoring through-out the procedure
punc-The “venous” catheter previously positioned in the leftventricle along with the arterial line allows recording ofsimultaneous left ventricular and aortic pressures This
helps to confirm the severity of the aortic stenosis,
how-ever, in these infants with poor left ventricular function,the gradient, particularly in very severe aortic stenosis, isoften minimal A left ventriculogram is not usually neces-sary and potentially is dangerous in a critically ill infant If
a ventriculogram is desired, it is recorded with injection inthe left ventricle through the prograde catheter When theleft ventriculogram is accomplished before the valve is
crossed, it helps to localize the area of the valve orifice.
The diameter of the aortic valve annulus is measuredvery accurately from the aortic root injection or the leftventricular angiocardiogram Once the valve measure-ments are obtained, the appropriate dilation balloon for
the aortic valve dilation is chosen and prepared before an
attempt is made at crossing the valve The balloon chosen
is equal in diameter to the aortic annulus measured at thevalve hinge points The preferred balloons for very smallinfant and neonatal aortic valve dilations currently are thevery small Tyshak Mini™ (NuMED Inc., Hopkinton, NY)balloons The Tyshak Mini™ balloons up to 8 mm indiameter pass through a 3-French sheath while the 9 and
10 mm Mini™ balloons require a 4-French sheath TheMini™ balloons, however, accept only a 0.014″ wire TheHi-Torque Iron Man™ wire (Guidant Corp, Santa Clara,CA) is as effective a wire as any available for supportingthese balloons The dilation balloon is prepared using a
“minimal prep” technique The stenotic aortic valve iscrossed only after the balloon has been prepared and isready to be introduced These infants are so precariousthat the 3- or 4-French catheter alone crossing the stenoticorifice is enough to occlude the orifice and cause the infants
to decompensate acutely and occasionally irreversibly
A 3- or 4-French, pre-curved right or left Judkins™coronary catheter or a 3- or 4-French curve-tipped multi-purpose catheter is advanced retrograde into the aorticroot from the femoral artery To cross the aortic valves in aneonate, a 0.014″ or a 0.018″ very floppy tipped exchange
length, torque-controlled guide wire is used The wireused depends upon what the balloon chosen for the dila-tion will accommodate When a 0.018″ wire can be used, aPlatinum Plus™ (Boston Scientific, Natick, MA) or a V-18Control™ (Boston Scientific, Natick, MA) wire is veryuseful both for crossing the valve and for supporting thedilation balloon across the valve A very slight curve isformed at the tip of the wire to allow changes in directionwhen torquing the wire The wire is advanced through thepre-curved catheter positioned in the aortic root
Trang 19Crossing the aortic valve in the neonate is similar to
the “blind” crossing in an older patient except that all
structures are much smaller and, in turn, all catheter
manipulations are much finer Occasionally, the
pre-curved catheter will pass through the stenotic orifice
when maneuvered directly at the valve Crossing with the
catheter is attempted several times while simultaneously
rotating and moving the catheter in and out in the aortic
root When the catheter does not cross the valve, the
catheter tip is positioned at least one centimeter above the
valve and the very soft tip of a fine guide wire is
manipu-lated out of the catheter The wire is directed toward the
aortic valve orifice by almost continuously rotating the
catheter while simultaneously moving both the catheter
and the wire to and fro Multiple, rapidly repeated, short
probes are made with the wire toward the aortic valve
orifice rather than any slow, methodical maneuvers The
exact location of this orifice really is not known unless
another wire or catheter is already passing through the
valve The more “probes” that are made with the wire
during the simultaneous, multiple changes in the angle/
direction of the wire, the more likely is the tip of the wire
to cross the valve If the wire does not cross the valve after
several minutes of trying with the first angled catheter
and wire, the catheter and/or the wire are replaced with
either the right or left coronary catheter or the
multipur-pose catheterawhichever was not used initiallyaand the
“probing” with the wire is repeated during continual
changes in the direction of the catheter tip Eventually,
more by chance than skill, the wire crosses the valve into
the ventricle
Once the wire crosses the valve, it is advanced as far as
possible into the ventricle, hopefully even looping the soft
tip of the wire within the apex of the left ventricle With
the wire advanced as far as possible into the ventricle, the
catheter is advanced over the wire to the ventricular apex
If the infant remains stable with the end-hole catheter
passed through the valve, the catheter is fixed in the
ven-tricle and the wire is removed Preferably, the original
fine, soft tipped wire is replaced through the catheter
posi-tioned in the left ventricle with a stiffer, exchange length
wire of the largest diameter that the prepared balloon will
accommodate The stiffer wire should be pre-shaped
before introduction into the catheter/ventricle A
“pig-tail” curve is formed on the long floppy tip of the new
guide wire and a 180° curve, which conforms to the cavity
size at the ventricular apex, is formed at the transition
zone of the wire between the long floppy tip and the
stiff shaft of the wire This specifically formed wire is
advanced through the catheter into the left ventricle The
pre-formed “pig-tail” keeps the wire from digging into
the myocardium as it is manipulated within the ventricle
The 180° curve at the transition area of the wire prevents the
stiffer wire from perforating the left ventricle and, at the
same time, directs the tip of the wire 180° back toward the left ventricular outflow tract This position allows the
stiff portion of the wire proximal to the curve at the
trans-ition area to be postrans-itioned completely across the valve and deeper within the ventricle With the wire maintained
in position, the catheter is removed and replaced with the already prepared balloon catheter The balloon is positioned precisely across the valve and an inflation/deflation is performed as rapidly as possible The inflation/deflation is recorded on biplane angiography or onbiplane “stored fluoroscopy”
During the inflation of the balloon, the left ventricularpressure increases, the systemic arterial pressure drops,and the infant develops bradycardia As soon as thedeflation of the balloon is complete, the balloon is with-drawn over the wire and out of the aortic valve On relief
of the obstruction, the infant’s heart rate returns and theleft ventricular pressure decreases toward normal Therapidity of the stabilization of the infant is a partial indica-tion of the success of the dilation The infant’s heart rateshould return to a rate faster than the baseline rate and theleft ventricular pressure should be lower than predila-tion A very sick infant may require some medical ormechanical assistance for return of the cardiac output
When the infant does not remain stable with even the small end-hole catheter across the valve, the original wire
used to cross the valve is stabilized rapidly and the
catheter is immediately withdrawn over the wire and out of
the valve orifice while leaving the wire in place The
end-hole catheter is rapidly replaced with the previously
pre-pared balloon dilation catheter, which is passed over the
wire and positioned across the valve as expeditiously, but
at the same time as accurately, as possible Dilation of thevalve is carried out with as rapid an inflation/deflation ofthe balloon as is possible The balloon inflation/deflation
is recorded on biplane, stored fluoroscopy or ically After the inflation/deflation, the balloon is immedi-ately withdrawn out of the valve, over the wire and into atleast the ascending aorta to allow the infant to stabilize.The wire is kept securely in place in the ventricle when-ever possible
angiograph-Once vascular access is obtained and the aortic valve iscrossed with a retrograde catheter, aortic valve dilationfrom the femoral route usually is very successful.However, femoral arterial access is the most difficult andthe most hazardous part of the procedure in the neonateand small infant The vessels are very small relative toeven the smallest catheters/dilating balloons available foraortic valve dilation As a consequence, arterial complica-tions are relatively common in this group of patients.Although extreme occlusive problems with necroticischemia of a limb are extremely rare, cool extremitieswith poor perfusion of the extremity immediately after
a retrograde catheterization are common and usually
Trang 20represent compromise, if not total loss of the deep femoral
artery of the involved limb These patients usually acutely
regain perfusion and warmth of the extremity but often on
subsequent attempts at access, the deep femoral artery is
totally occluded or occasionally the patient has a relative
growth failure of that extremity As a consequence of the
femoral access problems several other routes for aortic
valve dilation have evolved
Carotid artery introduction for the retrograde
approach
Transapical balloon dilation of the aortic valve in neonates
set a precedent for direct collaboration between the
car-diovascular surgeon and the interventional cardiologist
for performing aortic valvotomies in critically ill infants
Because of the difficulties with arterial access and then
entering the ventricle from the femoral approach,
particu-larly with the earlier balloons, several centers performed
balloon dilation of the neonatal aortic valve in the
operat-ing room through a thoracotomy and a controlled, apical,
left ventricular puncture, but without the necessity of
car-diopulmonary bypass Although this approach avoided
cardiopulmonary bypass, it did not avoid deep general
anesthesia, the thoracotomy, the ventricular perforation,
and all of the inherent problems of these particular
proced-ures The improvement in balloons and balloon
tech-niques soon put this particular collaborative technique
to rest
An alternative, unique and, initially, seemingly radical
collaborative approach for aortic valve dilation was
devel-oped as a consequence of the ongoing technical problems
with balloon dilation of the infant aortic valve and, at the
same time, other vascular technology which had been
developing concurrently The continuing difficulties with
the arterial access site when using the femoral artery
approach in neonates and the persistent difficulties with
the catheter manipulation around the arch and across the
aortic valve from both the femoral artery and the
umbili-cal artery approaches made these approaches less than
ideal At the same time the expanded use of
extracorpo-real membrane oxygenation (ECMO) led to almost
“rou-tine” cut-downs and repairs of the carotid artery by pediatric
vascular surgeons
The combination of the persistent problems with the
femoral/umbilical catheter approaches and the proficiency
of the surgeons with the carotid cut-downs resulted in the
consideration of a carotid artery approach to the balloon
dilation of the aortic valve6 The carotid artery itself has
considerable appeal as an approach to the aortic valve Of
most importance the approach to the aortic valve from the
entrance into the carotid artery is a very short, very
straight shot! With this “straight line” from the right
carotid artery to the left ventricle, minimal catheter
manipulation is required to cross even a severely stenoticaortic valve and to enter the left ventricle with a wire orcatheter The carotid artery is approached and entered
by a cut-down procedure performed and repaired by avascular/cardiovascular surgeon The carotid approachcombined with essentially no vascular or central nervoussystem sequelae when a skilled vascular surgeon exposesand repairs the artery, makes the carotid approach seemlike the ultimate solution to a major problem
Patients undergoing a carotid cut-down approach foraortic valve dilations are anesthetized with general anes-thesia to ensure that they are maintained absolutely stillthroughout the procedure They undergo endotrachealintubation for control of their respiration and to keep thehead and face out of the “operating” field in the neck Thecarotid approach for the dilation is usually used in con-junction with a prograde catheter from a systemic veinand a separate indwelling femoral arterial monitoringline The majority of the hemodynamic and anatomicinformation is obtained through these lines before thecarotid cut-down is initiated The venous catheter isadvanced through the patent foramen ovale (PFO) andinto the left ventricle in almost all of these infants Thevenous catheter and femoral arterial line are introducedsimultaneously while the surgeon is introducing thesheath into the carotid artery and usually the progradecatheterization and femoral lines do not add any over-all time to the procedure The prograde catheter andindwelling arterial line actually simplify and add to thesafety of the procedure The venous prograde catheter inthe left ventricle and the separate indwelling femoral arte-rial line provide continuous left ventricular and systemicarterial pressure monitoring before, during and after thedilation without having to cross the aortic valve repeat-edly with a retrograde catheter
A vascular surgeon performs a cut-down on the neckover the mid portion of the right carotid artery and iso-lates the right common carotid artery A floppy tippedwire that will accommodate the balloon dilation catheter
is introduced into the artery either through a small sion performed by the surgeon or through a needle intro-duced by the pediatric cardiologist
inci-The surgeon introduces a short 4-French sheath/dilatorwith an attached back-bleed valve/flush port into thecarotid artery over the soft tipped wire, and advances
it just to the base of the carotid artery as observed onfluoroscopy All further introduction and positioning ofthe sheath tip are performed by the pediatric cardiologistand are visualized continuously on fluoroscopy If even ashort sheath is introduced to its hub in the carotid artery in
a small infant, the tip of the sheath extends to (or past) thearea of the aortic valve If the tip of the sheath does not
pass through the orifice of the valve, it potentially
pro-duces catastrophic damage to the aortic valve Rarely, the
Trang 21sheath/dilator/wire passes into the descending aorta
from the right carotid artery cut-down and has to be
with-drawn and maneuvered specifically back into the aortic
root Alternatively, the wire may pass directly into the left
ventricle during its initial introduction When the sheath
is in the proper position in the ascending aorta, the dilator
and wire are removed and the sheath is cleared
meticu-lously of any air or clots before flushing The sheath is
secured in the artery by means of a purse string and
sub-cutaneous sutures so that the tip of the sheath is fixed at
least 1–1.5 cm above the aortic valve An aortic root
aor-togram is performed with an injection directly through
the sheath or through a catheter introduced into the
sheath Accurate measurements of the valve annulus are
carried out using one of the reference calibration
tech-niques described in Chapter 11, “Angiography” The
appropriate balloon for the valve dilation is chosen from
these measurements exactly as it is chosen for the femoral
approach The balloon is prepared with a “minimal prep”
in order to eliminate all air from the balloon
A soft tip exchange wire, with a very slight curve at
the tip, which will accommodate the prepared balloon
dilation catheter is introduced through the sheath The
wire is introduced either directly through the sheath or
through an end-hole catheter introduced into the sheath
first A catheter within the sheath provides more control
over the tip of the wire and allows this catheter to be
advanced over the wire into the ventricle as soon as the
wire enters the ventricle At the same time, the catheter
adds considerable “length” outside of the introductory
site into the carotid artery, and it extends well above the
infant’s head for the manipulations The wire is
manipu-lated through the aortic valve into the left ventricle This is
usually accomplished with only a few “probes” and
mini-mal readjustment in the direction of the wire tip Once
through the valve, the wire is advanced deep into the
ventricle until the stiff portion of the guide wire is entirely
across the valve The wire is fixed in position in the
ven-tricle while the catheter is removed over the wire from
the sheath
The balloon catheter is advanced over the wire, into the
sheath, and to the area of the valve The balloon is
cen-tered precisely across the valve annulus as previously
identified by angiography or as visualized on a
trans-esophageal, or even a transthoracic, echocardiogram, and
a rapid balloon inflation/deflation is performed The
inflation/deflation is recorded on a biplane angiogram or
on “stored fluoroscopy” The balloon is withdrawn over
the wire and back into the sheath immediately after it is
deflated As with any other aortic valve dilation, the
in-fant’s hemodynamics deteriorate during the balloon
infla-tion in the valve, but usually return rapidly to “normal”
(or better) with the deflation of the balloon and its
with-drawal from the valve When the prograde left ventricular
and indwelling femoral arterial lines are in place, thehemodynamic results of the dilation are available imme-diately The patient is allowed to stabilize while therecording of the inflation/deflation is reviewed particu-larly for the balloon position and the appearance/dis-appearance of a waist on the balloon When the infant’shemodynamics have stabilized, the balloon is reintro-duced across the valve and the inflation/deflation isrepeated at least one more time Before each reinflation,the balloon is repositioned forward or backward in thevalve slightly The balloon is observed for any persistent
“waist” on it during subsequent inflations, particularlyduring the very initial phase of the inflation After a successful dilation, no residual waist should appear onthe balloon, even during the early phases of subsequentinflations
When satisfied with the appearance of the inflationsand the resultant hemodynamics, the balloon is with-drawn over the wire and out of the sheath, and the sheath
is passively cleared very carefully of any air or clot The
balloon dilation catheter is replaced with an end-holecatheter, which is advanced into the ventricle over thewire that is still positioned in the left ventricle The wire isremoved slowly through the catheter, the catheter cleared
of air/clot and the simultaneous left ventricle and femoralartery pressures are recorded If a femoral line was not inplace or there was no prograde catheter in the ventricle,the pressures are obtained following the dilation by either
a pull-back of the end-hole catheter across the valve or,preferentially, with the left ventricular pressures obtainedthrough the catheter still positioned in the ventricle andthe arterial pressure obtained simultaneously from theside port of the carotid artery sheath In this case, thesheath which is in the carotid artery/aorta must be at least one French size larger than the catheter When the
stenosis is not relieved by the initial dilation and aortic
insufficiency is not significant, a wire is repositioned inthe ventricle (either through the catheter still in the ven-tricle or by manipulating a wire/catheter back through avalve), and the dilation repeated with a larger balloon orwith repositioning of the balloon If a very wide pulsepressure is recorded or massive aortic regurgitation isvisualized by echocardiogram, either the procedure isconcluded or the catheter is withdrawn into the aorta and
an aortic root angiogram is performed to verify the sence and amount of aortic regurgitation
pre-When the prograde catheter is already in the ventricle,the pressures are recorded simultaneously from thiscatheter, the femoral arterial line, and the sheath in theaortic root When there is no prograde catheter in the ven-tricle, an end-hole catheter is advanced over the wire intothe left ventricle, the wire is removed, and the pressurefrom the left ventricle is recorded through the catheterwith a simultaneous femoral artery pressure from the
Trang 22femoral arterial line If the resultant pressure in the left
ventricle appears satisfactory compared to the femoral
arterial pressure, a “pull-back” pressure is recorded as the
catheter is withdrawn from the left ventricle into the aorta
A follow-up aortic root aortogram is recorded with
injec-tion through the “retrograde” carotid catheter The
deci-sion for any further ballooning of the valve is made
exactly as for other aortic valve dilations When satisfied
with the results of the dilation, or if significant aortic
regurgitation was created regardless of the residual
gradi-ent, the sheath is withdrawn from the artery and the artery
repaired meticulously by the vascular surgeon
There are growing data indicating that the retrograde
approach to the aortic valve from the carotid artery may
be the safest and, in turn, the preferred approach for
bal-loon dilation of the aortic valve in very small infants Once
the sheath is secured in the carotid artery, the carotid
approach represents the most expedient way of crossing
and dilating a severely stenotic aortic valve in an infant
The results of the dilation are comparable to other dilation
techniques and there are no reported acute
complica-tions from the properly performed carotid approach The
repaired carotid arteries have good Doppler flow
immedi-ately following the dilation and on short-term follow-up
of the carotid repairs Whether this technique is used
rou-tinely in any particular center depends upon the
availabil-ity and co-operation of the surgeon, the surgeon’s skill at
vascular access and repair, and the working relationship
between the surgeon and the interventional cardiologist
When all of these “elements” fall into place, this is the
pre-ferred approach to the dilation of critical aortic valve
stenosis in the newborn and small infant
Umbilical artery introduction for the retrograde
approach
In the newborn infant, at least one of the umbilical arteries
is potentially patent for up to a week (or more) after birth
In a newborn infant with severe aortic stenosis who
requires aortic valve dilation, an umbilical artery
ap-proach for the retrograde catheter provides a potential
arterial access without compromise of a femoral or the
carotid artery7,8 The newborn with severe aortic stenosis
should have a venous catheterization with a prograde left
heart cardiac catheterization After the first few days of
life, the umbilical vein often is not accessible, in which
case the usual femoral vein approach is used for the
venous catheterization
The infant with severe aortic stenosis usually is critically
ill and frequently already has an end-hole polyethylene
“umbilical artery” catheter positioned by the
neonato-logist in the abdominal aorta from one of the umbilical
arteries! In that situation, a fine, very small diameter, but,
preferably, relatively stiff, exchange length, teflon-coated
wire is passed through the umbilical catheter and as far
as possible into the thoracic aorta, ascending aorta
and, with the ultimate luck, even into the left ventricle.
Unless the wire ends up in the left ventricle, the ethylene umbilical artery catheter is replaced over this wire with a 4-French angled tipped (right coronary)catheter
poly-When there is no pre-existing umbilical artery line, the
umbilical cord stump is scrubbed very thoroughly and then
“amputated” parallel with and close to the abdominalwall This exposes the stumps of the two umbilical arteriesand the umbilical vein The two arteries are smaller indiameter, rounder and thicker walled than the single,larger diameter and irregular vein Usually, all three ofthese umbilical vessels are obliterated with thrombi, however, patency of the arteries can often be restored
by “probing” the arterial lumen with a small diameter,smooth and blunt tipped metal probe One edge of theexposed wall of the end of one of the arteries is graspedwith a small forceps and the occluded lumen of the artery
is probed with a very small, blunt, metal, vessel probe afew millimeters at a time until the probe passes severalcentimeters into the lumen The probe usually stayswithin the walls of the vessel as it dissects through thethrombus With the stump of the vessel opened in thismanner, an end-hole “umbilical artery catheter” is intro-duced into the channel that was created with the probe,and advanced into the newly opened vessel lumen whileplacing some “counter tension” on the exposed edge ofthe vessel with the forceps As the catheter is introduced
into the lumen, it is directed posteriorly and caudally
toward the posterior pelvis Once the catheter hasadvanced several centimeters into the artery, it is visual-ized on fluoroscopy to determine whether the catheter isbeing directed to the right or the left inguinal area Withthat information, the catheter is pushed in that directiontoward that groin and the “counter traction” on the
“stump” is pulled away from that direction
Once the catheter has advanced one to two centimeterswithin the artery, the vessel often seems to “open up” andallow the catheter to move more freely through the umbil-ical artery Simultaneously, there may be blood returninto the catheter Usually resistance is encountered inmaneuvering the catheter around the sharp 90–130° curve
at the junction of the umbilical artery and the iliac artery.Occasionally, advancing the umbilical catheter is facilit-ated by advancing a fine, floppy tipped, torque wire or a
fine, curved Terumo™ wire through, and slightly in
advance of, the umbilical catheter Once the umbilicalcatheter has advanced into the descending aorta, anexchange length wire is advanced further into the thoracicaorta, the ascending aorta and, as before, with consider-able luck, into the left ventricle If the wire does notadvance all of the way to the ventricle, the soft umbilical
Trang 23catheter is removed over the wire and replaced with a
4-French right Judkins™ coronary catheter
From the descending aorta, the combination of the wire
and right coronary catheter is advanced/manipulated
around the aortic arch into the aortic root This maneuver
is often very difficult and requires several exchanges of
different wires and/or catheters Once the catheter has
reached the aortic root, the wire is removed, the aortic
pressure recorded, and a biplane aortogram is recorded
with the X-ray tubes angled to “cut the valve on edge”
(some degree of LAO–Cranial and RAO–Caudal) with
the two planes Accurate measurements of the valve
annulus are made using one of the various accurate X-ray
calibration techniques A “marker catheter” can be placed
in the superior vena cava adjacent to the aorta or even the
left ventricle when there is prograde venous access If
there is no venous line, a calibration marker catheter
can be inserted gently into the infant’s esophagus and
advanced to the area behind the cardiac silhouette as the
reference for the angiographic calibration system When
positioned just behind the center of the heart shadow, the
calibration marks on the catheter in the esophagus are in
the plane of, and very close to, the aortic valve
Once the measurements are complete, a very soft tipped,
fine torque wire is inserted into the catheter and the
stenotic aortic valve is probed with the catheter/wire
combination The technique is exactly as with any other
isolated retrograde approach to the aortic valve with the
exception that it is far more difficult from the umbilical
artery By the time the catheter/wire has been advanced
into the aortic root, the combination catheter/wire has
made two fairly acute and nearly 180° curves (passing
from the umbilical to the iliac artery and from the
des-cending to the asdes-cending aorta) As a consequence, torque
control and to-and-fro control over the catheter are
re-stricted markedly In addition, the “push” on the proximal
shaft of the catheter entering the umbilical artery must be
toward the groin and, counter-intuitively, “away” from the
direction of the aortic root Finally, there frequently is
significant arterial spasm along the course of the catheter,
which restricts catheter manipulation even further
If the wire can be manipulated across the valve and well
into the ventricle, and depending on which wire is used to
cross the valve, an attempt is made at passing the catheter
into the ventricle If the shaft of the original wire is
relat-ively soft, and depending upon the lumen diameter of the
dilation balloon catheter that is to be used, the initial wire
usually must be replaced with a wire which the dilation
balloon catheter can accommodate and which is stiffer in
order to support the delivery of a balloon/catheter
through the circuitous course Once the proper wire is in
place, the catheter that is over the wire is placed on a
con-tinuous flush through a wire back-bleed valve and the
appropriate balloon is prepared
When the dilation balloon is ready, the original catheter
is withdrawn over the wire leaving the wire positionedsecurely in the left ventricle The balloon dilation catheter
is advanced over the wire into the umbilical artery andthrough the circuitous course to the aortic valve This fre-quently cannot be accomplished or is only accomplishedwith one or more exchanges of wires When the bal-loon is positioned accurately across the valve, a rapidinflation/deflation is performed while recording the pro-cedure on biplane angiography or stored fluoroscopy.When the infant stabilizes after the balloon deflation,
if possible, the balloon is left in place across the valve until the angiograms of the inflation are reviewed Theinflation/deflation is repeated at least one more time Thereinflation/deflation verifies that the “waist” on the bal-loon does not reappear with the initial reinflation of theballoon If the infant does not stabilize rapidly after thedeflation with the balloon still across the valve, the bal-loon is withdrawn into the ascending (or descending!)aorta while simultaneously advancing the wire to main-
tain the wire’s position in the ventricleaif at all possible.
Once stabilized, an attempt is made at re-advancing thedilation balloon across the valve for at least one moreinflation/deflation
When the dilations have been completed, the balloon iswithdrawn over the wire, still leaving the wire in the ven-tricle The balloon catheter is replaced with an end-holecatheter, which is advanced over the wire into the ven-tricle If there is no prograde catheter in the left ventricle,the wire is withdrawn from the catheter, pressures arerecorded from the ventricle, and a biplane left ventricularangiogram is performed through this catheter Eventhough the retrograde umbilical catheter is an end-holecatheter, satisfactory left ventricular angiograms can beobtained through these catheters in a neonate A with-drawal pressure tracing from the left ventricle to theascending aorta is recorded, following which an aor-togram is performed in the aortic root to assess the move-ment of the valve leaflets and the degree of aorticregurgitation The catheter is removed and the umbilicalartery stump is oversewn When there is a progradevenous catheter positioned in the left ventricle during theprocedure, the pre- and post-left ventricular pressuresand any left ventricular angiograms are obtained throughthis catheter The prograde catheter obviates the necessity
of multiple recrossing of the valve with the umbilical/retrograde catheter
Because of the difficulties in maneuvering the catheters,wires and balloons from the umbilical artery approachand now with the much smaller balloon dilation cathetersthat can be introduced into the femoral arteries throughvery small (3- or 4-French) sheaths, the umbilical arteryapproach is no longer attempted in our center, although it
is still used as the preferred approach in some centers
Trang 24Prograde approach to aortic valve dilation
There still are difficulties, particularly in smaller patients,
in using any arterial approach for aortic valve dilation
The various potential problems from the different arterial
approaches led to the development of techniques for the
venous approach for the prograde delivery of the dilation
balloon(s) to the aortic valve for aortic valve dilation The
usual prograde approach to the left heart is with catheters
introduced from the femoral veins, although the
umbil-ical veins have been used in neonates and the jugular/
brachial veins have been used in the presence of a
pre-existing interatrial communication These approaches
obviate virtually all of the particular problems with the
arteries that occur with the arterial approaches, but they
do have some significant difficulties of their own9,10
Prograde catheterization of the left ventricle and the
aorta through a pre-existing intracardiac communication
or by means of a transseptal atrial puncture are techniques
that are used very commonly The use of a diagnostic
catheter which is advanced prograde into the aorta after it
has been introduced into the left heart through a
trans-septal atrial puncture is part of the routine procedure for
dilation of the aortic valve in the older patient as described
earlier in this chapter When a prograde dilation of the
aortic valve is anticipated, two additional measures are
necessary First, when the transseptal puncture is
per-formed, the transseptal sheath (set) that is used must
be large enough in diameter to accommodate the largest
dilation balloon catheter that is to be used for the prograde
aortic valve dilation Secondly, the passage of the catheter
from the left atrium to the left ventricle initially is
accom-plished using a “floating” balloon catheter The inflated
balloon of a floating balloon catheter is more likely to float
preferentially through the true, central orifice of the mitral
valve, cleanly between the papillary muscles, and away
from any chordae This is essential in order to avoid the
eventual passage or expansion of the relatively large
dila-tion balloons through narrow channels in, or entangled
with, these valve structures
The hemodynamic data are recorded and angiography
performed in both the left ventricle and the aorta through
the prograde catheters as described earlier in this chapter
Once the hemodynamics and angiography are completed
and it is established that aortic valve dilation is indicated,
a long transseptal sheath is advanced over the floating
bal-loon catheter and positioned securely in the left ventricle as
near to the apex as possible If a double-balloon technique
is to be used, a second, appropriately sized, long
trans-septal sheath is introduced into the left atrium and the left
ventricle in a similar fashion to the first long sheath Either
an end-hole, floating balloon or torque-controlled catheter
is advanced through each long transseptal sheath As the
catheter tip reaches the tip of the sheath, the sheath is
withdrawn very slightly to allow the tip of the catheter
to exit the sheath without digging into the ventricularmyocardium The catheter(s) is(are) manipulated fromthe left ventricle, 180° into the aorta, around the arch and well into the descending aorta using the techniquesdescribed earlier The end-hole catheter used for this can
be either a soft, torque-controlled catheter or an end-hole,floating balloon Swan™ type catheter A floating ballooncatheter is usually easier and safer to use for this manipu-lation, although the smaller floating balloon catheters donot accommodate a 0.035″ guide wire
When the end-hole catheter(s) has/have been advancedwell into the descending aorta, an exchange length guide
wire is introduced into each catheter through a wire back
bleed valve With the catheter on a slow continuous flush,
the wire is advanced through the catheter and positionedsecurely in the descending aorta The exact wires useddepend a great deal on the size of the patient and the typeand size of balloons to be used Whenever possible, a relat-ively stiff wire is desirable in order to provide support forthe passage of a balloon catheter around the two 180°curves en route to the aortic valve and to hold the bal-loon(s) in place during the dilation At the same time, thewire cannot be so stiff that it holds the mitral valve openand that it “straightens” the normal 360° course from theright atrium to the aorta into a straight line! Occasionallythe stiff wire will not traverse around the curves throughthe original prograde catheter to the aorta or, even if thewire can be positioned in the aorta, it is so stiff that it holdsthe mitral and/or aortic valves open or “splints” the ven-tricular walls apart Any of these difficulties with the wireinterferes with the ventricular function to such a degreethat it cannot be left in place for even a few minutes Thewires in their course from the left atrium, through the ven-tricle, to the aorta must have a long loop, deep into the
apex of the left ventricle and must not pass across the
ven-tricular cavity The wires must be maintained in that tion throughout the dilation procedure in order to preventdamage to the mitral valve
posi-Once the appropriate wires are in satisfactory, stablepositions and the patient still remains stable, the catheter(s)remain(s) over the wire(s) while all of the catheter(s)and/or the long sheath(s) in the left ventricle are main-tained on a slow continuous flush while the dilation balloon(s) are prepared The dilation balloon for the pro-grade approach should be shorter than those used for aretrograde aortic valve dilation The shorter balloons facil-itate the delivery of the relatively stiff “balloon segment”
of the balloon dilation catheter through the tight curves
in the circuitous prograde course to a position across theaortic valve The shorter balloons also help to ensure that
the proximal ends of the balloons are positioned entirely in
the left ventricular outflow tract and well away from themitral apparatus when they are inflated After a balloon is
Trang 25prepared, the original catheter is removed over the wire
while maintaining the wire in its position well into the
descending aorta and through the long sheath and, at the
same time, the tip of the sheath is maintained near the left
ventricular apex
Each balloon dilation catheter is introduced over the
wire into the long sheath in the femoral vein and
advanced to the aortic valve, all of the time maintaining
the sheath and the catheter lumen on a continuous flush
This circuitous route through two 180° curves within the
heart usually takes considerable and meticulous
manipu-lation of both the catheters and wires while maintaining
the sheath in position securely within the left ventricle
Care must be taken that the wire loops neither tighten nor
elongate excessively on themselves When the wires begin
to tighten, they pull against, and cut into, the valvular
structures and begin to pull the tips of the wires back out
of the aorta When the loops of wire elongate excessively
they “stent” the valvular structures open or cut into the
valves along the “outer circumference” of the loops The
long sheath maintained in position deep within the left
ventricular apex makes the balloon passage from the
left ventricle to the aorta slightly more difficult, but this
position of the sheath is essential to help prevent mitral
valve damage
When a wire repeatedly pulls back from the aorta
and cannot be maintained in the descending aorta as
the balloon dilation catheter is advanced over it, a snare
introduced retrograde is used to hold the wire in place A
Micro Vena™ snare catheter is introduced into a femoral
artery through a 4-French sheath The distal end of the
prograde wire in the descending aorta is grasped with a
10 mm Micro Vena™ snare introduced retrograde The
wire is held securely with the snare in this position,
which secures the distal end of the prograde wire in the
descending aorta Alternatively, the distal end of the
pro-grade, exchange length wire is withdrawn through the
descending aorta and “exteriorized” through the femoral
arterial sheath Either the snare catheter holding the
prograde wire or the exteriorized distal end of the wire is
secured on the table outside of the artery When a
pro-grade, double-balloon dilation is being performed, the
second wire also is grasped with a retrograde snare This
can be performed through the same femoral artery with
the same snare grasping the two wires simultaneously or
with a second snare through the opposite femoral artery
When a single snare catheter is used, unless the first wire
or both wires is/are “exteriorized”, the exact tension on
the separate wires cannot be controlled separately
If a retrograde sheath/catheter cannot be introduced
into a femoral artery for the snare catheter, or the
intro-duction is contraindicated, an attempt is made at
manipu-lating the prograde wire all the way around the arch, to
the descending aorta and into one of the femoral arteries,
advancing the tip of the wire to a position well beyond theinguinal ligament The prograde wire can often be fixed
in this position by firm, manual, finger pressure directlyover the artery, compressing the artery and, in turn, secur-ing the wire in the inguinal area This fixes the wire butdoes not allow any “counter traction” or other change inposition or tension on the wire from the distal end
The through and through control on the wire with theretrograde snare keeps the wire from being pulled out ofthe aorta and back into the left ventricle as the balloon isbeing advanced At the same time, the loops in the wirepassing through the heart with traction at both ends of thewire, can “tighten” the loops dangerously about the struc-tures in the heart For many reasons, significant tightening
of the wires must be avoided If the bare wires tighten,
they cut into the structures “within” the loopsain
particu-lar the medial leaflet of the mitral valve, the mitral dae, the left ventricular outflow tract and/or the aorticvalve itself As the loops of the wires tighten, the wires
chor-pull across the sub-valve apparatus of the mitral valve
If the wires remain across the sub-valve apparatus of the
valve as the balloon is advanced, the larger diameter,rougher surfaced, balloon also passes through or isexpanded in the sub-valve mitral apparatus causing dis-ruption of the apparatus The long sheath across the mitralvalve helps to protect the mitral valve but does not guar-antee that the mitral valve cannot be severely damaged.Even when the tightening loops do not cause injury, the
“tightening” and, therefore, narrower loops in the wire
create more resistance to the movement of the balloon
catheter over the entire course of the wire The tighter thecourse of the wire loops, the more difficult the passage ofthe balloon becomes The most extreme and dangerousdegree of “tightening” of the loops of the wire results inthe wire starting from its previous 360° course while pass-ing from the inferior vena cava, to the right atrium, leftatrium, left ventricle and out into the aorta, actually
straightening into an entirely straight course through the
same structures! This occurs with a sudden “flip” of thewire and unequivocally will damage intracardiac struc-tures This very dangerous phenomenon is prevented bythe continual observation of the course of the wires andnever allowing the “loops” in the wires to begin to tighten.The significant difficulties with the passage of a balloondilation catheter over the wire require an exchange of theoriginal wire for a smaller diameter wire or a wire withentirely different characteristics The prograde delivering
of the balloon on the balloon dilation catheter to the valve
is the most difficult part of this procedure
Once the balloon is positioned across the aortic valve,any wire loops or curves that were tightened excessively,are “loosened” by advancing the wire very carefully The
proximal end of the prograde balloon across the aortic
valve must be positioned cephalad in the left ventricular
Trang 26outflow tract and away from the mitral valve apparatus A
long smooth loop of the wire is formed proximal to the
balloon in the left ventricular apex so that the wire will be
positioned well away from the mitral valve Initially the
balloon is inflated very slowly and only partially to be sure
that there are no unexpected or unwanted, additional
“waists” appearing from entrapment in part of the mitral
valve apparatus When the partial inflation appears
“clear” of abnormally located indentations, the balloon is
inflated and deflated rapidly as with any other aortic
valve dilation
In addition to the difficulty in delivering the balloon to
the aortic valve, there are some additional disadvantages
to the prograde technique for the dilation of the aortic
valve The smaller the patient or the smaller the heart, the
tighter the 180° curves or loops of wire/catheter become
and the more difficult it is to advance the balloons around
the multiple curves When the patient is
hemodynamic-ally unstable to begin with, all of the balloon/wire
manip-ulation is not tolerated These problems are compounded
in critically ill infants and particularly in small infants
with a relatively small left ventricular cavity where it is most
appealing to avoid the arterial approach If the patient
does not tolerate the balloon inflation or the balloon across
the valve per se, it is more difficult to withdraw the balloon
quickly from the valve and into a stable location than it is
from the retrograde approach In addition to the risk of
damage of the mitral valve and/or valve apparatus from
the wire/balloon catheter alone, and during the balloon
inflation, there also is a higher incidence of left bundle
branch block or complete heart block created during the
prograde procedure
Considering both the technical challenges and the
added risks of the prograde approach, the benefits of the
prograde approach for the dilation of the aortic valve must
be weighed heavily against the disadvantages before
embarking on this approach for aortic valve dilation
Complications of aortic valve dilation
The inability to complete an adequate dilation of the aortic
valve is not a “complication”, but is a failure of the
pro-cedure The failure of the procedure is not, in itself, an
adverse event for the patient unless there is an additional
separate complication from the procedure The
underly-ing disease and not the failure of the procedure itself,
cre-ates the necessity for further intervention At the same
time, complications specifically related to aortic valve
dilation are more common than with most, if not all, of the
other therapeutic catheter procedures
One of the most common complications specifically
associated with aortic valve dilation is injury to a
periph-eral artery Although the incidence of arterial injury has
decreased as a result of improvements in both the niques and in the available equipment for the procedure,arterial spasm, clots, tears and disruptions of the arterystill occur Prevention using meticulous technique and theuse of the least traumatic equipment possible preventsmost arterial injuries (Chapter 4) Early recognition of
tech-an arterial problem with early intervention ctech-an preventpermanent sequelae The availability and use of specificcatheter interventions and of more effective thrombolyticsfor arterial thrombi are discussed in Chapters 2 & 35 It
is now extremely rare for a patient to require a surgicalarterial repair or have any permanent sequelae as a consequence of a retrograde arterial procedure even in
a small infant
The most serious complications from aortic valve tions are injuries to the central nervous system As withany procedure on the “systemic side” of the circulation,the potential for embolic events to systemic organs, par-ticularly the head, is always present during aortic valve dilations There are innumerable opportunities for theintroduction of air and/or clots into the circulation prox-imal to the head vessels during the multiple manipula-tions in the aortic root, left ventricle, and left atrium Airembolization absolutely should be preventable by the use
dila-of meticulous technique
Solid particle embolization is a different story Thrombieasily form on guide wires, on and within catheters and sheaths and specifically on balloon dilation catheterswithin the folds of collapsed balloons Patients undergo-ing these procedures are always anticoagulated with hep-arin during the procedure Unfortunately, even this is notsufficient to prevent thrombus formation unequivocally.The best that can be done to prevent thrombi is to keep thepatient’s ACT level above 300 seconds, to be as expedient
as possible with manipulations in the systemic circulation,and to keep all catheters and wires through catheters on
a continuous flush “Parking” catheters and, particularly,balloon catheters in the descending aorta as opposed tothe ascending aorta when they are not being used forrecordings/procedures specifically in the aortic root,reduces the likelihood of a “head” event
The most unpredictable, frustrating and probably, themost common complication of aortic valve dilation is
the creation of significant aortic valve regurgitation It is
the most frustrating complication because the specificcause of the regurgitation in any one patient is still not
known although it is a potential risk, as a consequence of
every balloon dilation of the aortic valve Significant aortic
regurgitation occurs in approximately 10% of all balloondilations of the aortic valve This number is comparable
to its occurrence following surgical aortic valvotomy, but whether it is a problem in the same patients whetherthey undergo surgery or catheterization is not known Itappears that over-sizing of the balloon for the aortic valve
Trang 27annulus is one predictor for the creation of aortic
regurgi-tation, however aortic regurgitation certainly is created
even when using a precisely measured balloon to annulus
ratio of less than one Unquestionably, the wire and then
the balloon passage through a valve leaflet rather than
through the commissures/orifice of the valve results in
significant regurgitation This problem may not be
recog-nized until after the fact This particular cause of
regurgi-tation is avoided by gentle retrograde probing of the valve
with very soft, floppy tipped wires and by directing the
retrograde wire along the precise course of a previously
positioned catheter that is passing prograde through the
valve orifice.
Once even a moderate amount of aortic regurgitation
is created during a balloon dilation procedure, no further
dilation of the aortic valve is attempted regardless of
the residual obstruction Fortunately, patients with very
significant aortic stenosis and, particularly, those with
significant chronic aortic stenosis tolerate even moderate
to severe degrees of acutely created aortic regurgitation
without the immediate need for valve repair or
replace-ment Massive aortic regurgitation with acute left
ventric-ular decompensation does occur, particventric-ularly in patients
who have only moderate aortic stenosis with no
ventricu-lar hypertrophy before the dilation, and whose ventricles
are “unprepared” for the sudden volume load This is the
major contraindication to dilation of mild or moderately
severe aortic valve stenosis Facilities must be available
for urgent surgery and acute valve replacement in the case
of such an occurrence in any patient undergoing aortic
valve dilation
Mitral valve damage during the prograde approach
for aortic valve dilation has been discussed previously
Avoiding the prograde approach for dilation of the aortic
valve is the best method of preventing this complication
However, the mitral valve apparatus can be disrupted
even during a retrograde dilation of the aortic valve It is
possible for the retrograde wire and then the balloon to
pass through and become entrapped in the mitral valve
chordae If this is not recognized, the mitral valve can be
damaged An indicator of this potential problem is a very
posterior and “less mobile” position of the retrograde wire/
balloon in the ventricle A left ventricular
angiocardio-gram or a transesophageal echocardioangiocardio-gram demonstrates
the errant wire/balloon position in the mitral apparatus
Perforation of the left ventricle is possible during
dila-tion of the aortic valve A straight wire, particularly with
a shorter, straight “floppy” tip is more likely to perforate
the myocardium as it is extruded out of the tip of a
catheter, although even a wire with a long floppy tip can
perforate if enough forward force is placed on the wire
when the tip of the catheter is “buried” in the myocardium
and cannot bow or bend away from the tip of the catheter
If there is a very sharp kink or angle on the stiff portion of
the wire when a “curve” on the stiff or transition zone ofthe wire is positioned in the apex of the left ventricle
and the balloon “milks” into the ventricle during inflation,
the balloon pushes the acute, sharp angle on the wire
fur-ther into the ventricle A straight wire with no curve on the
transition zone can kink acutely and be driven throughthe ventricular wall by a rapid forward force on the wirecaused by a “squirting” balloon This is true particularlywith small sick infants
A problem that is more likely to occur in, but is nottotally limited to, sicker patients is the inability of theheart to recover and the patient to stabilize after the totalobstruction of cardiac output by the balloon inflation/deflation in the aortic valve Usually, the relief of theobstruction by the dilation is sufficient to allow the rapidreturn of the heart rate, blood pressure and, in turn, a bet-ter cardiac output with rapid stabilization of the patientimmediately after the balloon(s) is/are deflated On rareoccasions, the myocardium initially is so damaged and is
“stunned” even further by the acute and total obstruction
of coronary flow during the balloon inflation that whenthe balloon is deflated, there is no return of cardiac func-tion in spite of all resuscitative efforts This may not beentirely preventable It is less likely to happen if the patientwas pretreated adequately and heart failure, shock andacidosis were corrected before the dilation procedure The double-balloon technique allows some forward flowthrough the valve when the balloons are at full inflation,which reduces the adverse heart rate and blood pressureeffects
The so-called “hooded coronary” orifice can result in acatastrophic outcome in association with balloon dilation
of the aortic valve With this lesion the orifice of a coronaryartery lies low or deep in the coronary sinus While thevalve is stenotic, the “tethering” or doming of the leafletsholds the leaflets away from the abnormally locatedorifice, but when the valve is dilated successfully, particu-larly if one leaflet becomes partially flail, the leaflet thenfolds completely against the wall of the sinus and, in turn,over the orifice of the coronary artery, acutely obstructingflow into the coronary The best indicator of this abnor-mality is a very low position of the coronary orifice andthe closeness of the aortic leaflet to the coronary orificebefore the valve undergoes dilation Recognition andimmediate surgical intervention are the only managementfor this rare abnormality
Rupture of the aortic valve annulus is another trophic event that can occur during aortic valve dilation,but probably is preventable This complication is unlikely
catas-to occur unless the balloon(s) is/are oversized markedlyfor the annulus size or the patient has underlying aortic wall disease such as the medial necrosis as seen inMarfan’s syndrome As a consequence, this problem theoretically should be avoidable by meticulous attention
Trang 28to the measurements of the annulus, using valid reference
calibration measurements, and by thorough knowledge of
the patient’s history
In spite of the numerous potentially serious
complica-tions of balloon dilation of the aortic valve, the benefits of
dilation of the aortic valve still outweigh the risks of the
catheter procedure and the risks of the alternative
man-agement/therapy as long as meticulous attention is given
to the details of the performance of the procedure
References
1 Lababidi Z, Wu JR, and Walls TJ Percutaneous balloon aortic
valvuloplasty results in 23 patients Am J Cardiol 1984; 53:
194–197.
2 Rocchini AP et al Balloon aortic valvuloplasty: Results of the
Valvuloplasty and Angioplasty of Congenital Anomalies
Registry Am J Cardiol 1990; 65: 784–789.
3 Justo RN et al Aortic valve regurgitation after surgical versus
percutaneous balloon valvotomy for congenital aortic valve
stenosis Am J Cardiol 1996; 77(15): 1332–1338.
4 Zeevi B et al Neonatal critical valvar aortic stenosis A parison of surgical and balloon dilation therapy Circulation
com-1989; 80(4): 831–839.
5 Mullins CE et al Double balloon technique for dilation of
valvular or vessel stenosis in congenital and acquired heart
disease J Am Coll Cardiol 1987; 10(1): 107–114.
6 Fischer DR et al Carotid artery approach for balloon dilation
of aortic valve stenosis in the neonate: a preliminary report
9 Magee AG et al Balloon dilation of severe aortic stenosis
in the neonate: comparison of anterograde and retrograde
catheter approaches J Am Coll Cardiol 1997; 30(4): 1061–
Trang 29In the more developed countries of the world, mitral
stenosis in children is predominately congenital in origin
and, fortunately, is a relatively rare abnormality1 The
stenosis in the congenitally malformed mitral valve has
multiple variations and involves all parts of the mitral
valve apparatus The “characteristic” congenital mitral
stenosis has a single papillary muscle, the valve leaflets
are fused into a funnel-like or “parachute” deformity, and
there are no real commissures of the valve The leaflets
alone can be fused without the fusion of the papillary
muscles or vice versa There also often is a “supra-valve”
obstructive membrane within the left atrium just above
the mitral valve annulus with or without the other
abnor-malities of the mitral valve apparatus In spite of the
com-plexity of these lesions, some of these valves are amenable
to balloon dilation2,3
The success of either balloon dilation or reconstructive
surgery for congenital mitral valve stenosis depends a
great deal upon the initial anatomy of the stenosis A
suc-cessful balloon or surgical dilation opens the stenosis and
relieves the obstruction, but still only represents
palliat-ive therapy This palliation hopefully delays the almost
inevitable mitral valve replacement Dilation of
congeni-tal mitral valve stenosis by any technique is likely to
pro-duce some mitral regurgitation Fortunately, the degree of
insufficiency usually is not great, is tolerated moderately
well and certainly is tolerated better than a very tight
stenosis Because of the relative difficulty and the risks
and the lack of predictability of the results of dilation
of congenital mitral stenosis, the criteria for performing
balloon dilation of congenital mitral stenosis are more
stringent than for most other therapeutic procedures in
the congenital cardiac catheterization laboratory Even
moderate mitral stenosis symptomatically is tolerated
quite well and the presence of the stenosis per se is not
an indication for intervention Patients with congenital
mitral stenosis are considered for intervention only whensymptoms become significant or when secondary pul-monary hypertension develops When balloon dilation
of a congenital mitral valve stenosis is anticipated, thedefinite possibility of creating significant mitral regurgita-tion, which will require a mitral valve replacement, mustalways be included in the decision for and the discussionsabout the balloon valvotomy
In the developing countries of the world where acuterheumatic fever is still rampant, rheumatic mitral valvestenosis is very common and does occur in young children.Because the rheumatic stenosis represents the fusion of whatwere previously normal, bicuspid, mitral valve leaflets,the dilation of the acquired fusion is far more successfulthan the dilation of the congenitally deformed valves As aconsequence, the indications for catheter intervention forrheumatic mitral stenosis are much more lenient than forthe congenital mitral lesions The presence of rheumaticmitral stenosis with any symptoms or with even the sug-gestion of an increase in right ventricular and pulmonaryartery pressures is an indication for intervention for rheu-matic mitral stenosis in children, adolescents and adults4.Balloon dilation is considered the standard primarytherapy of rheumatic mitral stenosis at virtually any age,however, in the older patient with mitral stenosis, the deci-sion for valve dilation is weighed against the amount ofacquired degenerative changes in the valve5 In the child
or adolescent with rheumatic mitral stenosis, the ative changes in the valve that produce the contraindications
degener-to balloon therapy are usually not present In the youngerage group the mitral valve stenosis must be distinguishedfrom congenital mitral stenosis, which may represent acontraindication to balloon dilation of the valve
The applicability of a stenotic rheumatic mitral valve for
a transcatheter mitral valvuloplasty is determined on thebasis of a combined echocardiographic and X-ray gradingsystem The thickness of the leaflets, the restriction in the valve motion, the sub-valvular thickening due to thefusion of the chordae or the papillary muscles and the
20
Trang 30amount of calcium in the valve on X-ray are all
determin-ates in the “mitral valve score” Each of these factors is
graded 1 through 4 according to the worsening degree
of severity, and when the numerical grades are added
together, give the “score” The higher the score, the less
likely the procedure is to succeed and the greater the
chances of complications6 Thrombi or masses within
the left atrium are relative contraindications to balloon
dilation of the valve regardless of the other findings
The balloon types and sizes used for mitral valve
dila-tion vary according to the size of the mitral valve annulus,
whether the dilation is with a single “standard” balloon, a
double-balloon technique or with the Inoue™ balloon
The dilation technique and type of balloon(s) used are
mostly chosen according to the preference of the
particu-lar operator and the availability of the particuparticu-lar
neces-sary equipment in the catheterization laboratory There is
a great variability in how the mitral annulus diameter is
measured and, in turn, how the balloon sizes are
deter-mined for balloon dilation of the mitral valve The mitral
annulus is not actually measured in some adult series, but,
instead, the balloon diameter is chosen arbitrarily
accord-ing to the patient’s body size Other series utilize a
for-mula or a graph according to the patient’s measured
height and/or body surface area to “calculate” the balloon
size to be used
The size of the Inoue™ balloon is frequently
deter-mined according to the patient’s height using the formula:
Inoue™ balloon size = patient’s height (in cms) ÷ 10 + 10
This diameter is the maximum diameter of the
particu-lar balloon With the variability in the diameter of the
Inoue™ balloon possible during the dilation, this allows
the operator to begin the dilation using a diameter 4 mm
smaller than the maximum for that balloon and then
sequentially increasing the diameter up to the calculated
end or maximal diameter of that balloon, all without
hav-ing to exchange balloons
For the double-balloon techniques it is better to
meas-ure the various diameters of the mitral valve annulus
angiographically using several angled views or by
echo-cardiography in several different planes When actually
measured, the dimension of the mitral annulus in the
lat-eral projection usually is 25–30% longer than the posterior–
anterior dimension The longest annulus diameter is used
in choosing the two balloons The combined diameter
of the two balloons used is usually equal to, or up to, 1.3
times the longest measured diameter of the mitral
annu-lus Occasionally, when there is very gross malformation
of a congenitally stenotic mitral valve, the balloon dilation
is started with smaller diameter balloons If the annulus
diameter is not adequately measured by either modality,
then a graph of normal valve diameters according to body
surface area is used to determine the balloon sizes7
Techniques for balloon dilation of the mitral valve
With the exception of one of several retrograde proaches for balloon dilation of the mitral valve and theextremely rare incidence of either transhepatic or trans-
ap-jugular access to the atrial septum, most balloon dilation
procedures of the mitral valve are performed from thefemoral venous approach The usual, as well as the alter-native approaches, to transcatheter dilation of the mitralvalve are discussed in this chapter
The mitral valve can be dilated using a single, large,
“standard” angioplasty balloon, with one of several ferent double-balloon techniques or, in larger patients,dilation with the Inoue Balloon™, which is designed spe-cially for mitral valve dilation All of these procedureshave advantages and disadvantages Each procedure hasits advocates The double-puncture, double-balloon tech-nique and the Inoue Balloon™ procedure are discussed
dif-in detail The sdif-ingle-balloon technique usdif-ing a large
“standard” balloon and the double-balloon using a singletransseptal puncture are mentioned only in comparison tothe other two, more preferred techniques
The safest technique (and the standard of care in oped countries) is to use biplane fluoroscopic guidance forcontrol of the transseptal procedure(s) as well as the actualmitral valve dilation procedure The actual dilation of thevalve also is supported with echocardiographic guidancecontrol Transthoracic echo (TTE) can be used particularly
devel-in the smaller patient, but now the trend is to use esophageal echo (TEE) or intracardiac echo (ICE) for guid-ance of the procedure In addition to the more reliableimages of the valve, both TEE and ICE have the advantagethat the transducer and the echocardiographer’s handsare out of the fluoroscopy field during the dilation pro-cedure The TEE procedure has the disadvantage that itrequires general anesthesia for most patients to toleratethe TEE probe, and at the same time to hold still for theduration of the procedure The experience with ICE is still
trans-limited, and it still requires the presence of an additional
large venous sheath
The catheterization and dilation of the mitral valve can
be performed with deep sedation and local anesthesia
when transesophageal echocardiography is not used to
guide the procedure If transesophageal graphic monitoring is used during the dilation procedure,then general anesthesia is used for the whole procedure
echocardio-Double-balloon, double transseptal puncture technique
A separate venous sheath is introduced percutaneouslyinto both the right and left femoral veins, and a small
Trang 31indwelling arterial cannula placed percutaneously into
one femoral artery As mentioned in the description of
the transseptal technique in Chapter 8, it is imperative that
the catheter course from the femoral puncture passes
through the true femoral vein and not through pelvic
col-lateral veins, which pass deep into the posterior pelvic
and perirectal venous system Right heart pressures and
cardiac output determinations are obtained with a right
heart catheterization Pulmonary artery and pulmonary
arterial wedge pressures obtained during this part of the
procedure help to substantiate the severity of the mitral
stenosis Following the acquisition of the “right heart” data,
the transseptal left heart catheterization is performed
When the double-balloon, double transseptal technique is
used, separate transseptal punctures are performed with
an approach from both inguinal areas The exact severity
of the stenosis and the anatomy of the valve and valve
apparatus are determined after the left heart is entered
transseptally
A double-balloon technique through two separate
transseptal punctures introduced from separate groins
is preferred for children as well as for smaller adults
This allows the introduction of two smaller balloons
into smaller peripheral veins and the passage across
the intra-atrial septum through two smaller and widely
separated holes in the atrial septum With the current
bal-loon technology, both balbal-loon catheters are introduced
through long sheaths The two punctures in the atrial
sep-tum are separated from each other by at least a centimeter
to prevent the two openings from coalescing and creating
one large and permanent hole in the septum! In a large
patient with a large mitral annulus, the double-balloon
technique allows the use of two balloons of adequate
combined diameter without the necessity of using one
very large, very bulky, and irregular, single, mitral
“angio-plasty” balloon with its very rough and traumatic deflated
profile
The two transseptal punctures are performed from
catheters introduced from the separate right and left
femoral veins (The transseptal technique is described in
detail in Chapter 8.) In the adult and in the moderate, or
large sized child or adolescent, both transseptal sheath/
dilator sets can be introduced through separate punctures
into the same femoral vein in either inguinal area when
there is an access problem from one side When using
bilateral femoral veins, it is preferable to perform the first
transseptal puncture with the catheter that is introduced
into the left groin The transseptal procedure from the left
groin usually requires bending the patient’s thorax to the
right in order for the needle to align more perpendicular
to the septum and to engage or impinge on the septum
during the actual puncture Performing the first
punc-ture from the left femoral vein allows freer bending
and movement of the patient during the first transseptal
puncture without fear of dislodging a previously tioned transseptal sheath by the movement of the patientduring the second puncture
posi-Although certainly not essential (or maybe even able!), the use of TEE guidance is popular for needle guid-ance during the transseptal puncture It is complementary
desir-to the fluoroscopy, but should not replace the use of biplane
fluoroscopy as the primary means of visualizing the tures Once the tip of the needle and the transseptal set isimpinged on the septum, TEE, then, is helpful for localiz-ing the site of the puncture on the septal surface and helps
punc-to ensure that the two puncture sites are separated fromeach other by at least a centimeter The TEE image of theneedle on the septum also provides a security check inavoiding the aortic root and the atrial septum–posteriorwall junction during the puncture, particularly whenthere is a small left atrium
When the right heart catheterization is performed fromthe left inguinal area or if sheaths are in both femoral
veins, the left venous catheter and sheath are replaced
over a wire with a 7- or 8-French transseptal set Thetransseptal set has a sheath with a back-bleed valve with aflush port and a radio-opaque marker band at the distaltip of the sheath The transseptal set is advanced over thewire into the superior vena cava and, preferably, into the left innominate vein The wire is withdrawn from thetransseptal set, the dilator cleared of air and flushed, the transseptal needle is introduced into the dilator,
advanced to just within the tip of the dilator and then
attached to the pressure/flush system The patient’s rax is angled (bent, not twisted!) to the patient’s right Thisbending creates a more perpendicular angle between thetip of the transseptal needle and the atrial septum once thetip of the needle has been withdrawn into the right atrium.With the pressure from the needle displayed on a 20 or
tho-40 mmHg scale, the sheath/dilator/needle is withdrawn/rotated from the superior vena cava and along the rightatrial surface of the interatrial septum, as described indetail in Chapter 8 (“Transseptal Technique”)
The transseptal puncture is performed as low on theatrial septum as is possible For these low positions for
puncture through the septum, TEE is useful to ensure that
the needle tip and, in turn, the puncture is not originatingfrom within the coronary sinus After the needle tip passesthrough the septum into the left atrium, visualization of
the posterior wall of the left atrium by TEE helps, but usually
is unnecessary, in avoiding puncture through the ior wall into the pericardium While observing pressurecontinuously through the needle, the needle/dilator/sheath are advanced through the septum and into the left
poster-atrium until the tip of the sheath is well within the poster-atrium.
With any loss of pressure, advancing the needle/longsheath/dilator is stopped, the tip of the needle is rotatedhorizontally slightly and the location of the needle tip is
Trang 32determined precisely with a small injection of contrast
through the needle
Once the tip of the sheath is well into the left atrium, the
needle first and then, separately, the dilator are removed
slowly from the sheath The dilator is allowed to bleed
back and drip fluid/blood out of the end of its hub as it is
withdrawn The sheath is then cleared meticulously of
any air and clots by allowing passive free flow of blood out
of the side arm of the back-bleed device while tapping and
rotating the back-bleed valve chamber The long sheath
in the left atrium along with the wire/dilator/catheter
exchanges through the sheath positioned in the left heart
represent the greatest potential hazards of the procedure
The wire/catheter manipulations through the long sheath
with the sheath tip positioned within the left atrium create
a huge potential for the introduction of air or clot into the
left heart Meticulous attention must be taken in clearing
the sheath to avoid this Once the sheath, including the
hub/chamber of its back-bleed valve, unequivocally is
cleared of any air or clot, the side arm of the sheath is
attached to the pressure/flush system At this point in the
procedure the patient is given 100 mg/kg of heparin
intra-venously through the long sheath When pressures are not
being recorded through the sheath, the sheath is
main-tained on a slow flush of heparinized flush solution
Once the first transseptal sheath has been secured in the
left atrium, the hemodynamics and anatomy of the mitral
stenosis are confirmed An accurate left atrial pressure is
recorded through the side arm of the sheath or through
a catheter introduced through the sheath into the left
atrium When the catheter is at least one French size
smaller than the sheath, pressures can be recorded
through both the catheter and the sheath simultaneously
The two simultaneous pressures from the left atrium
serve as an extremely reliable method of documenting the
accuracy of the two transducers (Chapter 10) The two
pressure tracings should produce a single line on the
mon-itoring screen Once the pressure in the left atrium is
recorded, angiocardiograms can be performed in the left
atrium to define the valve anatomically and to localize
various left heart structures An angiographic catheter is
advanced into the left atrium through the long sheath In
order to obtain the most useful anatomic information
about the valve, the X-ray tubes are positioned at angles
that are as perpendicular to the annulus of the valve as
possible The mitral valve usually is “cut on edge” with a
right anterior oblique/caudal (RAO/caudal) angulation of
the posterior–anterior (PA) X-ray tube A very steep left
anterior oblique/cranial (LAO/cranial or “four
cham-bered”) angulation of the lateral tube provides the closest
to perpendicular view of the valve from the lateral X-ray
plane However, it often is impossible to obtain a steep
enough cranial angulation with the lateral X-ray system to
cut the valve absolutely “on edge” from the LAO/cranial
angle Even though the entire valve may not be “on edge”with the LAO/cranial view, this view does give a seconddimension for measurement of the diameter of the valve
annulus and the sub-valve apparatus Since no two
patients and valves are the same, frequently several leftatrial angiocardiograms with changes in the X-ray tubeangulations are needed to position the mitral valve optim-ally “on edge”
From the left atrial angiocardiograms, the exact size,anatomy and position of the orifice of the mitral valve aredetermined During the angiocardiograms, either a cali-brated marker catheter, some type of external referencegrid, or large diameter calibration system built into the X-
ray system is used in each plane of the angiocardiograms
for accurate measurements of the valve annulus The urements should be recorded in as many views as pos-sible including specifically where the valve attachmentsare visualized within the ventricle The maximum meas-ured diameter of the valve annulus is used to determine the combined balloon diameters for the dilation TEE/ICE
meas-is used along with the angiocardiogram to define thevalvular anatomy and for a corroborative measurement ofthe annulus
A frame of the angiocardiogram from both planes of theX-ray is placed in a “freeze frame” image to be used as a
“road map” during the dilation If quality “freeze frame”
or image storage capabilities are not available, an externalradio-opaque marker can be placed on the skin surfacecorresponding to the location of the valve as visualized onthe angiogram This marker is used to identify the level ofthe valve orifice during the dilation A lead shot (#5!) or alead number 1 from a radiographic film “labeling set”,taped onto the chest wall in both planes, serves as anexcellent reference marker
With the left atrial pressure recorded and the mitralanatomy defined and measured accurately angiographi-cally, an angiographic catheter at least one French sizesmaller than the long transseptal sheath is manipulatedfrom the left atrium into the left ventricle With the angio-graphic catheter positioned in a stable location in the leftventricle and the tip of the larger French sized sheath still
within the left atrium, very accurate, simultaneous
pres-sure tracings of the left atrial and left ventricular prespres-suresare recorded from the side arm of the sheath and thecatheter, respectively These pressures provide the exactgradient across the stenotic mitral valve The valve areacan be calculated using this gradient along with the car-diac output determinations obtained during the rightheart catheterization
Once accurate and simultaneous left atrial and left ricular pressures are recorded, the X-ray tubes again arepositioned into the angles that are the most perpendicular
vent-to the valve annulus and a biplane left ventricular diogram is performed Often the negative shadow of the
Trang 33angiocar-mitral valve in a densely opacified left ventricular image
gives more information about the exact location of the
opening, size and excursion of the mitral valve leaflets
than the left atrial angiocardiogram If so, a frame of the
left ventricular angiocardiogram in the exact position
that will be utilized during the valve dilation procedure
is stored on the “freeze frame” to serve as a reference for
positioning the dilating balloons once they are exactly
within the lesion
After accurate measurements have been made of the
mitral annulus, the balloons are chosen for the dilation
The combined balloon diameters should be equal to or
1.2–1.3 times greater than the measured largest diameter
of the mitral valve annulus Relatively long balloons
are used, but, at the same time, balloons with short
“shoulders” and short tips are necessary Some balloons
are available with a “pig-tail” curve at the tip rather than
a short stubby tip The “pig-tail” is designed to keep the
balloon tip from driving into or through the ventricular
wall if the balloon is displaced into the ventricle during
inflation The longer balloons are necessary to span the
entire “depth” of the mitral valve and the valve apparatus,
and to help keep the dilating surface of the balloons from
moving in and out of the valve during the inflation
Once the angiocardiograms are completed, the
meas-urements obtained, and the size of the balloons chosen,
the angiographic catheter is withdrawn from the left
ven-tricle and from the body through the sheath Preparations
are made for the second transseptal puncture The two
balloons will be delivered through the septum through
two separate long transseptal sheaths If a larger sized
sheath than the one already in the left atrium is necessary
to accommodate one or both of the balloons that are to
be used for the dilation, the initial transseptal sheath is
exchanged for the appropriate larger sized sheath/dilator
at this time The exchange is made over a wire passed
through the first sheath into the left atrium or
pulmon-ary vein Once the new, larger sheath is secure in the left
atrium, the second transseptal puncture from the opposite
femoral vein is carried out using a second larger
trans-septal sheath/dilator set that will accommodate the
dilation balloon
The exact location on the septum of the second puncture
will depend upon where on the septum the earlier
punc-ture was made The second puncpunc-ture should be at least a
centimeter away from the first puncture and can be above,
below, in front or behind the first punctureawherever
there is the most room and distance from the first
punc-ture The location of the second puncture on the septum in
relation to the first puncture site and the rest of the septum
is identified from the earlier left atrial angiograms and by
TEE/ICE interrogation The second transseptal puncture
is carried out identically to the first transseptal procedure
except that when approaching from the right inguinal
area, the patient’s body usually does not have to be bent toalign the septum better
After both transseptal sheaths are secured in the leftatrium and cleared of air/clot, the side arms of bothsheaths are attached to the pressure/flush system andplaced on a slow continuous flush When the two longsheaths are placed on their separate pressure recordings,the two simultaneous pressures from the left atrium againserve to document the accuracy of the two transducersand, in turn, the reliability of the previous hemodynamics
The two pressure tracings should produce a single line on
the screen
In order to position the guide wires across the mitralvalve for the dilation, an end-hole catheter is advancedthrough one of the sheaths that are positioned in the leftatrium, and maneuvered through the mitral valve andinto the left ventricle It is safer, usually easier and alwaysbetter to “float” an end-hole, floating balloon catheterthrough the mitral valve than to maneuver a standardend-hole catheter from the left atrium to the left ventricle.The inflated balloon has a better chance of passingthrough the “center” and/or the largest orifice of themitral valve apparatus and not through a narrow slit in aleaflet and/or fused chordae of the valve TEE/ICE, aswell, is very helpful in determining the exact course of thecatheter through the valve apparatus
When the opening in the mitral orifice tolerates it, thesheath is advanced over the catheter and into the left ven-tricle Once the sheath and the end-hole catheter are in theleft ventricle and well past the mitral valve apparatus, thetip of the catheter is advanced out of the sheath andmaneuvered toward the left ventricular outflow tract Theexact mechanism of this maneuver depends upon the sta-bility of the patient, the type of catheter in the left ventricleand the preference of the operator The sheath over thecatheter and through the mitral valve into the left vent-ricle helps to support the more proximal shaft of a floatingballoon catheter in this maneuver As the balloon catheter
is pushed forward through the sheath, the sheath holdsthe shaft of the catheter in the left ventricle and keeps
it from backing up while the tip of the balloon cathetergenerally is forced to turn or deflect cephalad more or less,toward the LV outflow tract
When the long sheath cannot be maintained across themitral valve or the tip of the balloon catheter does notdeflect toward the outflow tract on its own in the small leftventricular cavity, a controllable deflector wire is used todeflect the catheter tip toward the outflow tract The activedeflector wire also provides an effective technique for deflecting a standard, “non-floating” end-hole catheter or
a floating balloon catheter that is not passing through asheath, toward the outflow tract The deflector wire, orany other wire introduced through the catheters in the
left heart, are introduced into the catheters through a wire
Trang 34back-bleed valve and the catheters are maintained on a slow
continuous flush around the wire The tip of the catheter
in the apex of the left ventricle is deflected into a 180°
curve with the deflector wire The catheter is advanced off
the wire toward, and well into, the left ventricular outflow
tract (LVOT) or even out through the aortic valve If the
catheter tip deflects in the direction of the aortic valve, but
the catheter cannot be advanced into the outflow tract,
the deflector wire is removed and replaced with a
torque-controlled, exchange length, guide wire with a slightly
curved, very floppy tip While keeping the catheter tip
directed toward the outflow tract, the curved floppy tipped
wire is passed through the catheter and maneuvered
out of the catheter and into the ascending aorta using
the torque mechanism on the wire to direct the tip Once
the wire has been secured in the aorta, the catheter is
advanced over the wire
In a small patient or in a patient with a small left
ventricu-lar cavity, the catheter passage into the left ventricuventricu-lar
outflow tract/aorta often cannot be accomplished with
a catheter large enough to accommodate the larger Super
Stiff™ wire (Boston Scientific, Natick, MA) that is to
be used for the balloon dilation of the mitral valve In that
circumstance, the deflection into the outflow tract is
per-formed with a smaller catheter, using a smaller deflector
wire, and then is exchanged for a smaller diameter
exchange length wire The smaller exchange length wire,
however, will not support the balloon dilation catheters
adequately for the dilation procedure The smaller gauge
wire is advanced through the smaller catheter and into the
aorta Once the smaller exchange wire is well into the
aorta, the original catheter is withdrawn over the smaller
exchange wire and replaced with a 6- or 7-French,
end-hole, either woven dacron, extruded polyethylene, or
even a Toray™ “Glide” catheter These catheters are large
enough to accommodate a 0.035″ wire, but at the same
time are malleable enough to follow the smaller wire into
the aorta These larger catheters positioned in the aorta
are rigid enough to guide a larger, pre-curved, Super
Stiff™ wire through the circuitous route from the venous
entry site to the apex of the left ventricle Once the larger
catheter has been advanced over the smaller gauge,
exchange length wire and into the ascending aorta, the
smaller wire is withdrawn, the catheter cleared of all
debris and placed on a slow continuous flush through a
wire back-bleed/flush port
Alternatively, a 6- or 7-French “pig-tail” catheter can be
used to position the larger Super Stiff™ wire in the left
ventricle The “pig-tail” catheter is introduced into the
left atrium directly through the long sheath and then
deflected from the left atrium into the left ventricle The
loop of the “pig-tail” passing through the mitral apparatus
has the same effect as a floating balloon catheter, as the
diameter of the “loop” of the “pig-tail” “backing” through
the valve obligates the catheter to pass through the largest
orifice of the stenotic mitral valve If a small exchange wire
already had been introduced into the aorta or left lar outflow tract, the pig-tail catheter is advanced over thesmaller wire, through the long sheath and into the leftatrium The pig-tail catheter is advanced through themitral valve and the curved tip or “loop” is pushed into
ventricu-the apex of ventricu-the left ventricle When ventricu-the pig-tail is passed
over the small wire, once the curve of the pig-tail is fixed
in the apex, the smaller wire is withdrawn completelyfrom the catheter The pig-tail curve seated in the left ventricular apex will usually deflect the long tip of a SuperStiff™ wire 180° when it is advanced into, and through,the loop and, in turn, directs the floppy tip of the wiretoward the left ventricular outflow tract This allows thetransition zone and curved stiffer portion of the wire to
“seat” in the apex of the ventricle
Once the first larger end-hole catheter is positionedsecurely across the mitral valve with the tip directed intothe LVOT, the procedure is repeated through the secondlong transseptal sheath The second long transseptalsheath already is positioned in the left atrium The mitralvalve is crossed with a second floating balloon catheterpassed through the second long sheath The catheter ismaneuvered into the LVOT or aorta using a techniquesimilar to that used for the first catheter as described
above The patient usually tolerates the two catheters
across the mitral valve and into the LVOT better than evenone of the Super Stiff™ wires within those catheters acrossthe valve
With the prograde catheters in place and ready for theinsertion of the Super Stiff™ wires, the indwelling femoralartery line is replaced with a small arterial sheath Anangiographic catheter is introduced through the femoralartery sheath and advanced retrograde into the left vent-ricle This retrograde catheter in the left ventricle was not
an absolutely essential part of the procedure up to thispoint, but during the rest of the procedure, the retrogradecatheter facilitates the assessment of the results of the balloon dilation immediately after the mitral dilationwithout the need for removing or exchanging either balloons or wires
Once all of the catheters are in place, two Super Stiff™ exchange length wires with long floppy tips (7 cm)
(Boston Scientific, Natick, MA) are pre-shaped specifically
for the mitral dilation and specifically for the particularpatient The wires used must be able to pass through theballoon dilation catheters that have been chosen for thedilation An exchange length, teflon-coated, 0.035″ Super
Stiff™ wire with a long floppy tip is used to support each of
the balloons during a double-balloon mitral valve tion Preferably, the long floppy tip should have a tight “J”
dila-or “pig-tail” curve fdila-ormed at the very end of the floppytip If only straight, long, floppy tipped Super Stiff™ wires
Trang 35are available, a J or tight pig-tail curve is formed on the
entire floppy portion of the wire The J or pig-tail curve at
the tip of the wire keeps the wire from digging into the
myocardium during the manipulations within the left
ventricle An additional fairly tight but smooth 180° curve
is formed just at the “transition” zone between the stiff and
floppy portions of the Super Stiff™ wire Regardless of the
other curves in the wire, this curve at the transition
between the end of the rigid portion of the wire and the
beginning of the “floppy” portion is essential to keep the
wire from kinking and becoming a “perforating” weapon
when “seated” in the apex of the left ventricle outside of a
catheter The diameter of the curve formed at the
“transi-tion zone” should correspond to the transverse diameter
of the particular patient’s left ventricular cavity distal to
the mitral valve During the dilation, this curve in the
transition/stiff area of the catheter will “seat” in the apex
of the left ventricle The 180° smooth curve helps to fix the
wire in the left ventricular apex while at the same time
preventing the wire (and the balloon tip, which is over the
wire) from perforating the left ventricle This 180° curve is
particularly important if a balloon “milks” forward
dur-ing the inflation of the balloons The straight portion of the
wire, which is proximal to the transition curve, should
direct the dilation balloon toward the apex rather than
across the ventricle and/or into the outflow tract of the
left ventricle The floppy portion of the wire distal to the
transition curve is directed 180° away from the inflow
of the ventricle and toward the left ventricular outflow
tract/aorta
An alternative curve and position for the support wire
is to form a 360+° “circular” loop or coil on the entire long
floppy tip of the wire, beginning with the transition zone
of the wire The wire still should have the 180° curve at the
transition zone just proximal to this 360+° looping of the
long floppy portion The “coil” of floppy wire remains
looped in the body of the ventricle as opposed to
extend-ing toward/into the left ventricular outflow tract This
positioning of the wire is simpler than directing the wire
specifically into the outflow tract, and gives just as much
stability, control and safety for the wires during the
inflation It is slightly harder to keep track of the separate
wires/balloons when two wires are looped around in the
ventricular apex
The first pre-curved Super Stiff™ wire is advanced
through one of the pre-positioned catheters and fixed with
the “transition zone” curve of the stiff wire seated in the left
ventricular apex and with the tip of the wire heading
toward the LVOT or aorta When the pig-tail catheter is
used to place the Super Stiff™ exchange wire, the pig-tail
loop of the catheter first is seated securely into the apex of
the left ventricle The Super Stiff™ wire is introduced into
the pig-tail catheter and advanced into the left ventricle
As the wire tip passes through the catheter, ideally, the
stiff “pig-tail curve” at the tip of the catheter “uncoils”
par-tially and deflects the soft tip of the wire 180° and towardthe direction of the LVOT as the wire passes through theloop of the catheter The Super Stiff™ wire is advanced
further into the pig-tail catheter until the preformed curve
at the transition zone of the wire reaches the pig-tail curve at the distal end of the catheter This results in the transition
curve of the stiff wire “seating” in the apex of the left tricle still within the pig-tail catheter
ven-It is imperative to have the straight and stiff portion ofthe Super Stiff™ wire extending entirely through the
mitral valve and to the apex of the ventricle At the same
time, the distal “end” of this stiff portion of the wire must
be protected by the bend or curve at the transition area ofthe wire, which directs the stiff straight wire away fromthe apex The placement and positioning of this wire maytake a variety of maneuvers with the sheath and/orcatheters and often is time consuming but, like most otherwire positioning for balloon dilation procedures, this par-ticular position of the wire is critical for the successful andsafe balloon dilation of the mitral valve Although it is notabsolutely necessary to have the tip of the wire in theaorta, it is desirable to have the soft portion of the wire,which is distal to the “transition” curve, heading at leasttoward the left ventricular outflow tract This positionhelps to stabilize the balloons in the correct area of thevalve and reduces ventricular ectopy due to the wires.The first catheter remains in place over the wire thatwas positioned first and maintained on a slow flushthrough a wire back-bleed valve, while the second SuperStiff™ wire is maneuvered into position in the apex of the left ventricle through the second catheter, which isalready positioned across the mitral valve and into theapex of the left ventricle The wires passing through thevalve can be visualized with TEE and/or ICE TEE/ICEwill show the position of the wires relative to the valveorifice and verify that they are passing through the center
of the orifice of the valve, and ensure that they are notpassing through some unwanted structure in the valve
or valve apparatus Once the second wire is in positionthrough the catheter, the patient is ready for the introduc-tion of the balloons and for the dilation of the valve
If not already across the mitral valve, the two longsheaths are advanced over the catheters and wires andinto the left ventricle The patient’s blood pressures andheart rate are observed closely for several minutes for anycompromise due to the increased diameters of the sheathscrossing the valve When the hemodynamic status of thepatient tolerates the two sheaths across the valve, thesheaths are maintained across the valve for the delivery
of the dilation balloons If the patient does not tolerate the two sheaths positioned across the mitral valve, the
sheaths are withdrawn back into the left atrium and tained in the left atrium The catheters that were used for
Trang 36main-the introduction of main-the wires are withdrawn over main-the
wires, through the sheaths and out of the body from each
groin while the wires are maintained fixed in place and
visualized continuously on fluoroscopy Extra care must
be taken to maintain the wires in their exact positions with
the 180° “transition” curve seated in the left ventricular apex
during the withdrawals of the catheters off the wires
With both wires positioned securely across the septum
and preferably, if tolerated, through the long sheaths
passing through the mitral valve, with the “transition”
curves of the wires in the left ventricular apex and the tips
directed into the left ventricular outflow tract, the
intro-duction of the balloons is accomplished
A double-balloon mitral dilation requires at least three
experienced personnel to be scrubbed while all of the
per-sonnel in the catheterization laboratory assisting with the
dilation procedure prepare for the balloon inflations
Recording of the monitored pressures from the left
ven-tricle and left atrium are started on a continuously running,
slow-speed, permanent recording The first balloon is
passed over the wire, through the long sheath and across
the area of the mitral valve In very severe mitral stenosis,
particularly if the long sheaths could not be maintained
across the mitral valve and were withdrawn into the left
atrium, the first balloon is “parked” over the wire in the
left atrium just proximal to the mitral valve The second
balloon catheter is introduced into the second long sheath
and advanced to the left atrium or the mitral valve
depending upon where the sheaths are positioned When
the sheaths are across the mitral valve, both balloons are
advanced across the valve still within the sheaths The two
balloons are positioned side by side exactly within the
valve as compared to the earlier “freeze frame” of the left
atrial or left ventricular angiograms of the valve and as
visualized on TEE and/or ICE The sheaths are
with-drawn off the balloons and back into the left atrium The
positions of the balloons in the valve again are compared
to the earlier “freeze frame” image and/or to a hand
injec-tion angiocardiogram through one of the long sheaths
and/or a left ventriculogram through the retrograde
catheter TEE or ICE gives a good image of the balloons
across the valve and helps verify their proper position in
the valve
If the sheaths had to be withdrawn into the left atrium
from the mitral orifice before the balloons were
intro-duced, the two balloons are advanced one at a time from
the left atrium, over the two separate wires, into the mitral
valve orifice The first balloon is advanced into position
across the mitral valve If the patient tolerates this balloon
across the valve at all, the second balloon is advanced
quickly over the other wire and immediately adjacent to
the first balloon The introduction of the two balloons
across the valve is performed smoothly, but as rapidly as
possible since as soon as even the deflated balloons reach
their appropriate positions across the mitral valve, thedeflated balloons themselves produce significant obstruc-tion The centers of each balloon (between the marks at theends of the balloon) are positioned at the stenotic area ofthe valve orifice As described above, the positioning can
be verified with TEE and/or ICE when utilized The loon positions can be documented further by a hand injec-tion through one of the two long sheaths in the left atriumand compared to the earlier “freeze frame” angiogram ofthe valve, or the positions of the two balloons relative to
bal-the valve and bal-the structures within bal-the left ventricle are
visualized on a left ventriculogram performed by tion through the retrograde catheter This angiocardio-gram is also important for assessing the amount of mitralregurgitation that is present with the stiff wires (and bal-
injec-loons) across the valve prior to dilation of the valve with the
balloons
Not only should the balloons be across the mitral valve, but the proximal ends of the balloons must be wellaway from the atrial septum while the opposite ends aredirected toward the apex of the left ventricle Neither balloon should be bent or turning across the ventricularcavity either posteriorly toward the mitral apparatus oranteriorly, “across” the ventricle and toward the outflowtract If the balloons are inflated in either of these abnor-mal positions, disruption of the mitral apparatus is likely
to occur
When both balloons are in their proper positions acrossthe valve, the two balloons are inflated simultaneouslyusing pressure-monitored and controlled indeflators.Unless the patient is very unstable with the deflated bal-loons across the valve, the balloon inflations are relat-ively slow and very controlled The balloon inflations are
visualized continuously fluoroscopically and recorded on
either slow-frame-rate biplane angiograms or biplane
“stored fluoroscopy” The inflation of the two balloons iscontinued until the recommended maximum inflationpressure of each balloon is reached or the “waist” in the
balloons created by the stenosis disappearsawhichever
comes first The balloons should not move in their tions within the valve during the inflations and a single,discrete “waist” should appear around the two balloons
posi-at the level of the stenotic valve The appearance of the
“waist” on the balloons and then the opening of the valve with the disappearance of the waist are also visibleand can be observed on TEE or ICE performed during theballoon inflation If either of the balloons begins to moveeither further into, or out of, the ventricle or if an unusual
or second “waist” appears on either balloon, especially deepwithin the ventricle, the inflation is stopped instantly, theballoons are deflated immediately and rapidly and, oncethey are fully deflated, withdrawn out of the valve
A very forceful displacement of the balloons into the
ventricle during the pressure inflation could continue to
Trang 37push the balloon tip, along with the stiff wire, even with
the curve at its transition zone, into, and actually through
the ventricular wall!! If a balloon that is displaced into the
ventricle does not push the wire with it, but follows the
stiff curve on the “transition” zone of the wire, it “turns”
toward the outflow tract and tracks perpendicularly across
the mitral apparatus Further inflation in this position
could disrupt the mitral apparatus An “abnormal waist”
on a balloon is created by a balloon that is positioned
abnormally through a chorda or in a fused papillary
muscle in the left ventricle The presence of any abnormal
waist should result in the immediate stopping of the
inflation of the balloon Further inflation of the balloon in
such abnormal positions very likely will disrupt the mitral
apparatus Displacement of the balloons backwards into
the left atrium pushes the balloons actually into the
intera-trial septum Continued inflation with the balloon pushed
backwards dilates a large and possibly permanent
open-ing in the atrial septum
If the inflation is interrupted, the inflation angiogram
and the TEE/ICE during the inflation are reviewed and
the balloons and/or the wires are repositioned to more
stable locations When comfortable with the new
posi-tions of the balloons and wires, the inflation is repeated
The inflation again is slow, is recorded on biplane
angio-graphy and observed on TEE/ICE following the same
precautions concerning abnormal positioning and/or
movement during each subsequent inflation of the
bal-loons Once the maximum pressures of the balloons are
reached or the “waists” on the balloons disappear, the
bal-loons are deflated as rapidly as possible and the patient’s
hemodynamics (vital signs) are allowed to stabilize
During the inflation of the balloons, the systemic blood
pressure drops dramatically as a result of the almost
com-plete obstruction of the forward flow of blood through the
mitral valve even when using the double-balloon
tech-nique If the systemic pressure does not begin to return
immediately with just the deflation of the balloons, the
bal-loons are withdrawn over the wires, which remain fixed
in place, and out of the valve orifice in order to allow freer
prograde flow through the valve Once the patient has
stabilized either with the balloons still across the valve
or after reinserting the balloons across the valve, the
inflation/deflation cycle is repeated two to four more
times By either advancing the balloons further into the
ventricle over the wire or withdrawing them slightly
back toward the left atrium by pushing the wires forward,
the position of the balloons is changed slightly during
each repeat inflation of the balloons This maneuver is
to ensure that the maximum parallel surfaces of the
balloons are inflated simultaneously and exactly in the
narrowest portion of the stenotic valve The positioning
is verified with TEE or ICE during each change of the
balloons’ position
In very severe mitral valve stenosis when even onesheath positioned across the valve cannot be tolerated oreven the first deflated balloon crossing the valve causes
a significant drop in systemic pressure, preparations are
made for an initial, rapid, single-balloon dilation in order
to open the valve partially before the double-balloon dilation The partial opening of the valve created by thesingle-balloon inflation allows the patient to tolerate thetwo deflated balloons across the valve for a subsequent,double-balloon dilation The wire(s) is (are) already inplace across the valve With the previous “freeze frame”images aligned with the bony landmarks of the thorax andone of the two balloons ready for immediate inflation, theballoon is advanced across the mitral valve and a rapid,but controlled inflation with biplane angiographic and
TEE or ICE recording is started Exactly as with the
con-trolled double-balloon inflation, if there is any
displace-ment of the balloon or the “waist” does not appear in its proper (expected) location as seen on angiography orTEE/ICE, the inflation is stopped and the balloon isdeflated rapidly and withdrawn out of the valve Thepatient is allowed to stabilize and the procedure repeatedwhen the balloon/wire position and inflation appears correct and stable Very rarely, a single-balloon dilation
must be accomplished before even the second wire can be
positioned for the double-balloon dilation When a “goodwaist” appears on the single balloon and the “waist” issuccessfully abolished with the balloon inflation, usuallythe procedure can be continued with a double-balloondilation as described above
How the results of the dilation are assessed ately after the dilation depends upon the placement of the
immedi-various catheters prior to the actual dilation procedure and
on whether TEE or ICE imaging is utilized With the grade catheter already positioned in the left ventricle,simultaneous left ventricular and left atrial pressures areobtained immediately through the retrograde catheterand the side arm of one of the long sheaths that are in theleft atrium The estimate of any residual gradient and
retro-of the diameter retro-of the valve orifice is also available fromthe TEE or ICE If the gradient across the mitral valve
is abolished or minimized, then the dilation procedure has accomplished the opening of the valve The variouscatheters are removed regardless of the other findings
If there still is a significant gradient across the valve, thedecision must be made whether further dilation of thevalve can and should be accomplished An estimate of the degree of mitral regurgitation is made with the TEE
or ICE imaging A left ventricular angiocardiogram is performed to “quantitate” any mitral valve regurgitationvery roughly If significant mitral regurgitation was pro-duced by the dilation, regardless of any residual gradient,
no further dilation is performed If there is no, or minimal,mitral regurgitation and still a residual gradient, the
Trang 38angiograms of the balloon inflations and the TEE
informa-tion are analyzed to determine if further dilainforma-tion is
pos-sible The diameter of the valve annulus and the combined
diameters of the fully inflated balloons in the annulus are
remeasured accurately and together in the same
angio-graphic sequence and by TEE or ICE If the combined
diameter of the two balloons is less than 1.6 times the
diameter of the mitral annulus in its largest diameter and
the dilation did not produce significant mitral regurgitation,
then larger diameter balloons are used for further dilation
The larger diameter balloons are introduced with the
same technique used for the initial balloons Occasionally
when a larger diameter balloon is used, the diameter of
the long sheath will also have to be increased In that
situ-ation the original sheath is withdrawn over the
support-ing exchange wire, which is maintained in place across the
mitral valve and in its position in the left ventricular apex
The new, larger French sized sheath with its dilator is
advanced over the wire, into the left atrium and, if
toler-ated, across the mitral valve This is repeated with the
other sheath if two larger balloons are to be used The
larger diameter balloons are introduced through the new
larger sheaths The dilation procedure is carried out with
the larger balloons exactly as with the initial dilation
balloons Usually when the valve is being dilated further
during the same catheterization procedure, the patient
remains more stable with the sheaths across the valve and
during the subsequent balloon inflations
Reassessment of the results of the dilation is repeated
exactly as after the initial dilation Once the gradient has
been eliminated or the waists on the larger balloons no
longer reappear during early balloon inflation, the
dila-tion procedure is completed and the diladila-tion balloons are
removed When there is a residual gradient or there is the
possibility of further dilation of the valve, the presence of
mitral regurgitation is assessed with a repeat left
ventricu-lar angiocardiogram and TEE/ICE before the wires and
balloons are removed from the valve TEE or ICE is more
sensitive at detecting tears and other disruption of the
valve leaflets and valve apparatus than angiograms If
there is significant mitral regurgitation and no valve
dis-ruption, but with the wires and balloons still across the
mitral valve, the balloons are withdrawn back into the
sheaths in the left atrium The TEE and the left ventricular
angiocardiogram are repeated If there still is significant
mitral regurgitation, the dilation balloons are removed
from the sheaths and replaced with end-hole catheters,
which are advanced into the left ventricle over the wires,
and the wires are withdrawn carefully through these
catheters The wires with their preformed curves readily
catch on the mitral valve apparatus, and must be
with-drawn through catheters that are fixed in the ventricle.
Once the wires are removed, the catheters are withdrawn
back into the left atrium/sheaths TEE/ICE and the
angiocardiogram are repeated once all balloons, wiresand catheters are out of the mitral valve
TEE or ICE interrogation of the valve and the left
ven-tricular angiocardiogram are repeated with nothing across
the mitral valve The presence of mitral regurgitation with
nothing crossing the valve represents real mitral valveinsufficiency! No further dilation of the valve is con-sidered in the presence of significant mitral regurgitation
If, however, the mitral regurgitation disappears, or even isreduced markedly when the wires are removed from thevalve, then the decision is made on the basis of the resid-ual gradient and the valve anatomy seen on TEE or ICEwhether the dilation needs to be repeated If the dilation
is to be repeated, it means almost starting from scratch,except that the transseptal sheaths are still in place acrossthe atrial septum The catheters and then the wires arereplaced across the mitral valve, larger balloons inserted,and the dilations repeated as described above
Once the final dilation is completed, pressure ments are recorded, a left ventricular angiocardiogram,preferably in the same view as utilized for the original leftventriculogram, and a repeat TEE or ICE analysis of thevalve are performed ICE imaging of the valve can nowprovide a direct image of the valve orifice facing it fromthe left atrial side through one of the large long sheathsthat are still in the left atrium A repeat left atrial angiocar-diogram, again in comparable views to the pre-dilationleft atrial angiogram, is recorded to demonstrate the change
measure-in the diameter of the valve orifice, the change measure-in valvemotion, changes in the emptying of the left atrium, andthe size and amount of leak across the interatrial septum
as a result of the transseptal approach
Occasionally and although it is not advocated, when verylarge balloons and, in turn, large sheaths must be used toaccommodate the balloons, some operators utilize a double-
balloon technique for dilation of the mitral valve without
the long sheaths remaining in the left atrium In that cumstance, the two venous punctures and two separatetransseptal punctures are performed with 7- or 8-Frenchtransseptal sets It still is important that the two transseptal
cir-punctures are at least one centimeter away from each other
so that the two holes do not “coalesce” into one Once thetransseptal puncture is completed, the end-hole cathetersand wires are positioned exactly as for a double-balloondilation through long sheaths Once the wires are in posi-tion, the original transseptal sheaths are removed over the wires and the dilation balloons introduced directlythrough the skin over the wires When very large “profile”balloons are necessary for the dilation, the holes in theatrial septum often first must be dilated with a smaller,6–8 mm diameter angioplasty balloon before the deflated,large dilating balloons will pass through the septum Oncethrough the septum, the dilation is carried out exactly aswith two balloons delivered through two long sheaths
Trang 39The only “advantages” of not introducing the two
bal-loons through long sheaths are of not having the long
sheaths in the “left heart” for as long and that the “factory”
folded balloons introduced directly over the wires initially
are smaller in outside diameter than the outside diameter
of the sheaths which they otherwise would be passing
through However, once the balloons are inflated and
deflated in the body their deflated “profiles” and
dia-meters are much larger than the comparable sheaths
There are multiple disadvantages of not using long
sheaths, no matter how much larger they might be The
large dilation balloons have horrible deflated profiles, and
when withdrawn through the atrial septum they make
excellent “septostomy” devices, which are very likely to tear
and extend the openings in the septumaresulting in very
large communications or tearing the two puncture sites so
that they coalesce into a single very large opening in the
septum The deflated large balloons are equally traumatic
to the smaller more peripheral venous structures and
often disrupt the introductory veins as the balloons are
withdrawn, resulting in later total occlusion of the vein
When the balloons are delivered directly over the wires,
there is no way of measuring left atrial pressure
immedi-ately after the dilation for assessment of the results of
the dilation while still keeping the two wires in position
across the mitral valve and while still keeping the two
balloons in the left atrium Either one balloon must be
removed over a wire, replaced with an end-hole catheter,
and then the wire completely removed, or a third
trans-septal puncture must be performed to introduce a third
catheter into the left atrium for the measurements The
third transseptal is possible with an additional long
transseptal sheath/dilator introduced “piggy-back”
adja-cent to one of the other femoral venous punctures
This third transseptal system can be as small as a 6-French,
since it will only be used for pressures and possibly
angiography
With the major disadvantages and much greater
poten-tial for long-term complications when the balloons are
not delivered through long sheaths, it is now standard
practice to use the long transseptal sheath technique for
double-balloon dilation of the mitral valve Regardless
of how large the diameters of the long sheaths are, they
always are less traumatic to the atrial septum and venous
structures than “bare” deflated balloons
Single standard balloon dilation of mitral valve
The single “standard” balloon dilation of the mitral valve
is still advocated by some operators for mitral dilations
The single standard balloon technique certainly is useful
for the very rare mitral valve dilation when the approach
is not from the femoral veins, and particularly the hepatic
and jugular vein approaches, which involve significantly
different transseptal and dilation procedures A single
balloon also is used in very small infants where the mitral
annulus is less than 10 mm in diameter The technique forthe transseptal puncture, the acquisition of the hemody-namics, performance of the angiography and the position-ing of a single stiff support wire for the procedure, all arethe same for a single balloon dilation approached from the femoral vein as described for the first wire of the double-balloon technique
For a mitral valve dilation using a single, standardangioplasty balloon, the diameter of the balloon usedshould be the same as the diameter as the mitral annulus.Because of the relatively large diameter of the mitral
annulus, this of course necessitates a very large balloon
re-lative to the patient’s body size Because of the necessarylong “taper” of the ends of these very large balloons, they
are also extremely long compared to the usable, parallel
walled sections of the balloons When the usable length
of these large balloons is “centered” in the valve annulusfor the inflation, the two ends of the balloon often
extend from the apex of the left ventricle back into the
atrial septum!
Because of the very large diameter of the balloon that isnecessary in the larger child or adult for a single-balloondilation of the mitral valve, these balloons have very large
deflated profiles, and there often is not a large enough long
sheath available to accommodate the larger balloon evenwith the initial “factory” fold on the catheter As a con-sequence, the balloon dilation catheter is usually passed
directly over the wire and not through a long transseptal
sheath or even a short venous sheath A separate dilation with a separate smaller balloon of the introduc-tory site into the subcutaneous tissues, into the vein andacross the atrial septum, often is necessary before the verylarge sized mitral dilation balloon can be advanced to themitral valve The pre-dilation of the subcutaneous tissuesand the vein is accomplished with a separate dilator ordilation balloon that is larger than the deflated diameter ofthe mitral dilation balloon The atrial septal opening isdilated with a 6 or 8 mm, low-profile, dilation balloon,which is advanced to the septum over the same wire Afterthe opening in the septum is enlarged, the large single mitraldilation balloon is advanced over the pre-positioned wire,through the septum and across the mitral valve
pre-In the very small infant, the single balloon for mitralvalve dilation is delivered through a long transseptalsheath In that specific circumstance, the long sheathremains in the left atrium and is available for immediatehemodynamic measurements after the dilation The longsheath also protects the intracardiac and venous struc-tures from the rough profile of the deflated balloon duringits withdrawal
Similar techniques and precautions are observed ing the inflation of the single balloon as are used in the
Trang 40dur-double-balloon technique The single balloon has to be
inflated and deflated as rapidly as possible and carefully
but rapidly withdrawn from the valve back into the left
atrium over the wire immediately after its deflation to
allow the return of the systemic cardiac output The large
single balloon inflates and deflates much slower than two
separate balloons used for the same dilation so that the
obstruction of cardiac output is much longer during the
dilation procedure As a consequence there usually is a
very significant drop in heart rate and blood pressure, and
a more prolonged recovery period or even a transient
“resuscitation” of the patient is required after the single
balloon is deflated
The single, standard balloon technique superficially
appears simpler than any of the other mitral valve dilation
techniques It does avoid the second venous puncture, the
second transseptal puncture, and the positioning of two
wires across the valve and in the left ventricular outflow
tract This appears as an advantage to very inexperienced
operators, however, to gain this small advantage, there
are many and significant disadvantages, which have
already been described The slower inflation/deflation of
the larger balloon has already been mentioned In larger
patients, the very large diameter single dilation balloon
catheters cannot be delivered or positioned through a long
sheath Reassessment of the hemodynamics immediately
after the single-balloon dilation can be performed only
by echocardiography or by placing a second catheter in
the left atrium from a second venous puncture and second
transseptal puncture! Without the second catheter in the left
atrium, the balloon catheter must be removed completely
out of the body and replaced with a catheter through
which pressures can be recorded By the use of a
Multi-Track™ catheter for this post-dilation assessment, at least
the wire does not have to be removed as well! If the results
indicate that further dilation is required, an even larger
balloon would have to be introduced
A much greater problem with the single standard
bal-loon dilation technique for the mitral valve is that a much
larger hole is purposefully made in both the femoral vein
and the atrial septum to introduce the large diameter,
large-profile balloon Even worse, when these very large
dilation balloons are deflated, they develop an even
larger, horribly irregular and rough profile This profile is
guaranteed to tear the septum further as the deflated
bal-loon is withdrawn directly through the septum, and can
be very traumatic to the iliofemoral venous system when
withdrawn out of the vein Because of these
disadvant-ages, this technique is not recommended
Bifoil balloon dilation of the mitral valve
For a while, with the idea of creating a more linear dilation
profile as occurs with a double-balloon dilation, but at the
same time, still being able to use a single puncture and single wire system, “bifoil” balloon catheters were de-veloped and used for valve dilations These bifoil ballooncatheters had two balloons side by side attached to thesame single catheter shaft and central lumen The bifoilballoon catheter was passed over a single wire similar to
a single standard balloon technique, however once the
“balloon” was in position in the orifice of the valve, thereactually were two balloons side by side, which were fixedsecurely in their side-by-side relationship
Unfortunately, all the advantages of a single punctureand single wire are obviated by the multiple disadvant-ages of the bifoil balloons The bifoil balloons could not
be introduced through any sheaths of any length Because
of the large and irregular profiles of the bifoil balloonseven before they are inflated, it was necessary to pre-dilate
the introductory venous tract, the atrial septum, and even the mitral valve itself with a 10 mm angioplasty balloon before
the bifoil system could be introduced The bifoil balloonshad problems with unequal inflation and the inability todeflate one or both balloons After the dilation, there is nomeans of measuring left atrial pressure without removingthe balloons completely Even when deflated properly,bifoil balloons had an even worse and more traumatic,deflated profile than even one large, single standard balloon As a consequence bifoil balloons were almostguaranteed to leave a permanent atrial septal defect andpermanently traumatize the introductory vein Because ofthe real disadvantages compared to any possible advant-
ages, these balloons faded from use and production and
fortunately are no longer available
Double-balloon, double-wire, through single
vein and single transseptal puncture: (Block™)
technique of mitral dilation
The double-balloon dilation over two wires introduced
through a single venous and single transseptal puncture
was developed to “simplify” and expedite the procedure
by eliminating one venous and one transseptal puncture.The special double-lumen Block™ catheter was developedspecifically for this procedure to carry the two wiresthrough the single puncture A single catheter was intro-duced percutaneously into one femoral vein The rightheart catheterization and cardiac outputs were obtained
A single transseptal procedure using a Mullins transseptalset was performed The hemodynamics and angiocardio-grams were recorded from the left heart using the catheterthrough the transseptal sheath as described previously
in this chapter
Once the anatomy and hemodynamics were obtained,
an end-hole catheter was maneuvered through the mitralvalve, into the left ventricle and to the left ventricularoutflow tract A stiff exchange length guide wire was