How can X-ray images be used more and more effectively like a full-body X-ray?... The conventional X-ray machine The process of lower limb stitching... Plain Stitching stitching room DR
Trang 1THE APPLICATIONS AND VALUES
OF THE EOS IMAGING SYSTEM
-MEDIC MEDICAL CENTER- VIETNAM
Dr Cam Van Pham Thi X-ray Radiologist, MEDIC polyclinic, VietNam
Trang 2I Overview of full-body skeletal imaging
II Beneficial comparison
1 Radiation dose
2 Quality of images 2.1 Length of view
2.2 Diagnostic imaging
2.3 Measurement
2.4 3D imaging III Pitfall comparison
1 Motion Artifact
2 Artifact from metal bodies
IV Cases of EOS Applications in Viet Nam
CONTENTS:
Trang 3What do you think when you see this image?
Trang 4Methods for quantitative evaluation of axial vertebral rotation [21] Cobb (A), Nash and Moe (B), Stokes (C), and
Perdriolle (D)
Axial balance Frontal and sagittal balance
Trang 5How can X-ray images be used more and more effectively like a full-body
X-ray?
Trang 6The conventional
X-ray machine
The process of lower limb stitching
Trang 7STITCHING ROOM DR
Full-leg / full-spine image stitching, a new and accurate CR-based imaging technique P Dewaele, P Vuylsteke, S Van de Velde, E Schoeters
Trang 8I.2 FULL-SPINE FROM THE LONGTAIL DR
LENGTH OF 42 X 17 INCH =108 X43 CM
Trang 9Plain Stitching <–> stitching room DR
or Longtail DR Comparison
The Longtail DR lower limb images
Trang 11I.3 THE EOS IMAGING
Trang 12GEORGES CHARPAK’S XENON MULTIWIRE PROPORTIONAL CHAMBER
EOS provides a true-size image
Trang 13THE EOS IMAGING MACHINE
Trang 14STER EOS SOFTWARE
Fig 3
Trang 15The EOS radiolucent chair was used for patients who cannot stand
Trang 16II BENEFICIAL COMPARISON
II.1 RADIATION DOSE
https://www.radiologyinfo.org/en/info/safety-xray
Trang 18“as low as reasonably achievable“
Trang 19II.2 LENGTH OF VIEW
Max length
of view
spine or lower limbs
175cm
Trang 20II.3 THE WEIGHT-BEARING STANDING POSITION
Lying CT scan
dislocation appears only in the standing position
Standing EOSStanding cervical X-ray Cervical spine CT
Trang 23Knee plain X-ray
Trang 24II.4 Accuracy Measurement
Trang 25II.5.1 MSCT 3D MODELS OF FULL SPINE
Presurgery full spine 3D CT reconstructions
the thoracolumbar spine on a CT scan
(3D reconstruction),
Trang 26II.5.2 EOS 3D MODELS OF THE LOWER LIMBS
Trang 30III PITFALL COMPARISON
III.1 Motion Artifacts
III.1.1 X-ray Motion Artifact
Unshape images
Trang 31III.1.2 CT Motion Artifact
https://radiopaedia.org/articles/motion-artifact-2
Step artifact in the sternal body due to breathing artifact.
CT brain with severe motion artifact
Trang 32III.1.3 EOS Motion Artifact The 1st doing The 2nd doing
Trang 33EOS Motion Artifact
pseudoCobb angle
Poking ribs
No shifting end plates
Trang 34EOS Motion Artifact Plain X-ray
pseudotumor
Trang 35III.2 The radiopaque object Artifacts
III.2.1 The Conventional X-ray Radiopaque Object
Trang 36III.2.2 The CT radiopaque
Object
Trang 37Halo line around metal
bodies
III.2.3 The EOS radiopaque Object
Without “IMPLANT”
function
Trang 39Radiograph of the right hip in a 75-year-old
woman with a painful total hip prosthesis painful total hip prosthesis a 67-year-old man with a
DIFFERENT DIAGNOSTICS
Trang 40Stitching Longtail DR CT scan EOS
Nonmoving 3 seconds x No parts 3 seconds 5 -10 seconds 10-20 seconds
4 Magnification exaggeration exaggeration Scale 1:1 Scale 1:1
Trang 41Worked in Viet Nam since 2014, providing care for over 40,000 patients, including:
• 25,000 full-body cases,
• 35,000 full spine cases,
• 10,000 full lower limb cases,
• 10 full upper limb cases
• and 10 standing pelvimetry cases
IV Some Cases of EOS Applications in Viet Nam
Trang 42A female 69 yo with spinal degeneration
the Bending test and full spinal EOS
Trang 43A female 12yo with thoracolumbar Scoliosis
Trang 44Sitting spinal EOS images of a female 27 yo, thoracolumbar Scoliosis before and after surgery
Trang 45Pelvimetry assessment of the 22-year-old pregnant woman
Trang 46A Female, 24 years old
Full upper limb EOS was checked
The left upper limb axis is varus
The functional lengths of the upper limbs are almost the same.
Trang 475 Jump up to:2.0 2.1 2.2 2.3"Pelvimetry by Three-Dimensional Computed Tomography in Non-Pregnant
Multiparous Women Who Delivered Vaginally" Polish Journal of Radiology
6 https://www.radiologyinfo.org/en/info/safety-xray
7 a 67-year-old man with painful total hip prosthesis, Catherine Cyteval
8 Concurrent Scoliosis and Dentofacial Anomaly: A Case Report
9 IMAGES IN FORENSICS Lethal mesenteric perforation by osteophytes after bluntabdominal traumaArnaud Le Gallo1&Christophe Loire1&Maisy Lossois2&Pierre-Antoine Peyron2
10 https://radiopaedia.org/articles/motion-artifact-2
Trang 48THANK YOU FOR YOUR ATTENTION