1. Trang chủ
  2. » Luận Văn - Báo Cáo

The applications and values of the eos imaging system

48 0 0
Tài liệu đã được kiểm tra trùng lặp

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề The Applications And Values Of The EOS Imaging System
Tác giả Dr. Cam Van Pham Thi
Trường học Medic Medical Center
Chuyên ngành X-ray Radiology
Thể loại thesis
Thành phố Vietnam
Định dạng
Số trang 48
Dung lượng 9,7 MB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

How can X-ray images be used more and more effectively like a full-body X-ray?... The conventional X-ray machine The process of lower limb stitching... Plain Stitching stitching room DR

Trang 1

THE APPLICATIONS AND VALUES

OF THE EOS IMAGING SYSTEM

-MEDIC MEDICAL CENTER- VIETNAM

Dr Cam Van Pham Thi X-ray Radiologist, MEDIC polyclinic, VietNam

Trang 2

I Overview of full-body skeletal imaging

II Beneficial comparison

1 Radiation dose

2 Quality of images 2.1 Length of view

2.2 Diagnostic imaging

2.3 Measurement

2.4 3D imaging III Pitfall comparison

1 Motion Artifact

2 Artifact from metal bodies

IV Cases of EOS Applications in Viet Nam

CONTENTS:

Trang 3

What do you think when you see this image?

Trang 4

Methods for quantitative evaluation of axial vertebral rotation [21] Cobb (A), Nash and Moe (B), Stokes (C), and

Perdriolle (D)

Axial balance Frontal and sagittal balance

Trang 5

How can X-ray images be used more and more effectively like a full-body

X-ray?

Trang 6

The conventional

X-ray machine

The process of lower limb stitching

Trang 7

STITCHING ROOM DR

Full-leg / full-spine image stitching, a new and accurate CR-based imaging technique P Dewaele, P Vuylsteke, S Van de Velde, E Schoeters

Trang 8

I.2 FULL-SPINE FROM THE LONGTAIL DR

LENGTH OF 42 X 17 INCH =108 X43 CM

Trang 9

Plain Stitching <–> stitching room DR

or Longtail DR Comparison

The Longtail DR lower limb images

Trang 11

I.3 THE EOS IMAGING

Trang 12

GEORGES CHARPAK’S XENON MULTIWIRE PROPORTIONAL CHAMBER

EOS provides a true-size image

Trang 13

THE EOS IMAGING MACHINE

Trang 14

STER EOS SOFTWARE

Fig 3

Trang 15

The EOS radiolucent chair was used for patients who cannot stand

Trang 16

II BENEFICIAL COMPARISON

II.1 RADIATION DOSE

https://www.radiologyinfo.org/en/info/safety-xray

Trang 18

“as low as reasonably achievable“

Trang 19

II.2 LENGTH OF VIEW

Max length

of view

spine or lower limbs

175cm

Trang 20

II.3 THE WEIGHT-BEARING STANDING POSITION

Lying CT scan

dislocation appears only in the standing position

Standing EOSStanding cervical X-ray Cervical spine CT

Trang 23

Knee plain X-ray

Trang 24

II.4 Accuracy Measurement

Trang 25

II.5.1 MSCT 3D MODELS OF FULL SPINE

Presurgery full spine 3D CT reconstructions

the thoracolumbar spine on a CT scan

(3D reconstruction),

Trang 26

II.5.2 EOS 3D MODELS OF THE LOWER LIMBS

Trang 30

III PITFALL COMPARISON

III.1 Motion Artifacts

III.1.1 X-ray Motion Artifact

Unshape images

Trang 31

III.1.2 CT Motion Artifact

https://radiopaedia.org/articles/motion-artifact-2

Step artifact in the sternal body due to breathing artifact.

CT brain with severe motion artifact

Trang 32

III.1.3 EOS Motion Artifact The 1st doing The 2nd doing

Trang 33

EOS Motion Artifact

pseudoCobb angle

Poking ribs

No shifting end plates

Trang 34

EOS Motion Artifact Plain X-ray

pseudotumor

Trang 35

III.2 The radiopaque object Artifacts

III.2.1 The Conventional X-ray Radiopaque Object

Trang 36

III.2.2 The CT radiopaque

Object

Trang 37

Halo line around metal

bodies

III.2.3 The EOS radiopaque Object

Without “IMPLANT”

function

Trang 39

Radiograph of the right hip in a 75-year-old

woman with a painful total hip prosthesis painful total hip prosthesis a 67-year-old man with a

DIFFERENT DIAGNOSTICS

Trang 40

Stitching Longtail DR CT scan EOS

Nonmoving 3 seconds x No parts 3 seconds 5 -10 seconds 10-20 seconds

4 Magnification exaggeration exaggeration Scale 1:1 Scale 1:1

Trang 41

Worked in Viet Nam since 2014, providing care for over 40,000 patients, including:

25,000 full-body cases,

35,000 full spine cases,

10,000 full lower limb cases,

10 full upper limb cases

and 10 standing pelvimetry cases

IV Some Cases of EOS Applications in Viet Nam

Trang 42

A female 69 yo with spinal degeneration

the Bending test and full spinal EOS

Trang 43

A female 12yo with thoracolumbar Scoliosis

Trang 44

Sitting spinal EOS images of a female 27 yo, thoracolumbar Scoliosis before and after surgery

Trang 45

Pelvimetry assessment of the 22-year-old pregnant woman

Trang 46

A Female, 24 years old

Full upper limb EOS was checked

The left upper limb axis is varus

The functional lengths of the upper limbs are almost the same.

Trang 47

5 Jump up to:2.0 2.1 2.2 2.3"Pelvimetry by Three-Dimensional Computed Tomography in Non-Pregnant

Multiparous Women Who Delivered Vaginally" Polish Journal of Radiology

6 https://www.radiologyinfo.org/en/info/safety-xray

7 a 67-year-old man with painful total hip prosthesis, Catherine Cyteval

8 Concurrent Scoliosis and Dentofacial Anomaly: A Case Report

9 IMAGES IN FORENSICS Lethal mesenteric perforation by osteophytes after bluntabdominal traumaArnaud Le Gallo1&Christophe Loire1&Maisy Lossois2&Pierre-Antoine Peyron2

10 https://radiopaedia.org/articles/motion-artifact-2

Trang 48

THANK YOU FOR YOUR ATTENTION

Ngày đăng: 22/01/2025, 09:30

Nguồn tham khảo

Tài liệu tham khảo Loại Chi tiết
5. Jump up to:2.0 2.1 2.2 2.3 "Pelvimetry by Three-Dimensional Computed Tomography in Non-Pregnant Multiparous Women Who Delivered Vaginally ". Polish Journal of Radiology Sách, tạp chí
Tiêu đề: Pelvimetry by Three-Dimensional Computed Tomography in Non-Pregnant Multiparous Women Who Delivered Vaginally
1. Charpak G. Prospects for the use in medicine of new detectors of ionizing radiation. Bull Acad Natl Med 1996;180:161–8 Khác
2. Deschenes S, Charron G, Beaudoin G, et al. Diagnostic imaging of spinal deformities: reducing patients’ radiation dose with a new slot-scanning X-ray imager. Spine 2010;35:989–94 Khác
3. The third dimension of scoliosis: The forgotten axial plane Tamás S. Illés a,b,c,∗, Francois Lavaste d, Jean F. Dubousset c Khác
4. A retrospective review of performance and utility of routine clinical pelvimetry ". Family Medicine 36 (7): 505–7. 2004. PMID 15243832 Khác
7. a 67-year-old man with painful total hip prosthesis, Catherine Cyteval 8. Concurrent Scoliosis and Dentofacial Anomaly: A Case Report Khác
9. IMAGES IN FORENSICS Lethal mesenteric perforation by osteophytes after bluntabdominal traumaArnaud Le Gallo1&amp;Christophe Loire1&amp;Maisy Lossois2&amp;Pierre-Antoine Peyron2 Khác

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN

w