BRITISH STANDARD Flow-metering devices for connection to terminal units of medical gas pipeline systems The European Standard EN 13220:1998 has the status of a British Standard res 11
Trang 1BRITISH STANDARD
Flow-metering devices for
connection to terminal units of
medical gas pipeline systems
The European Standard EN 13220:1998 has the status of a
British Standard
res 11.040.10; 17.120.10
NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHr LAW
BSEN
13220:1999
Trang 2BS EN 13220:1999
This British Standard, having
been prepared under the
direction of the Health and
Envirorunent Sector Committee,
was published under the
authorily of the Standards
Committee and comes into effect
on 15 April 1999
© BS( 04·1999
ISBN 0580 30686 0
National foreword
This British Standard is the English JanllPJal~e version of EN 13220: 1998
The UK participation in its preparation was entrusted by Technical Conunittee
CHl44, Anaesthetic machines, breathing ati:aclunents, medical gas systems
- aid enquirers to understand the pre~eJlt to the responsible committee any enquiries on the interpretation, or proposals for change, and the UK interests infonned.;
- monitor related international and European developments and promulgate them in the UK
A list of organizations represented on this subcommittee can be obtained on request
to its secretaIy
Cross·references
The British Standards which implement international or European publications referred to in this document may be found in the BSI Standards Catalogue under the section entitled "International Standards Correspondence Index", or by the
"Find~ of the BSI Standards Electronic Catalogue
A British Standard does not purport to include all the necessary of a contract Users of British Standards are responsible for their correct application
COlm))lliLn(:e with a British Standard does not of itself confer fmm unity
obligations
Summary of pages
This document comprises a front cover, an inside front cover, the EN title page, pages 2 to an inside back cover and a back cover
Amendments issued since publication
Trang 3EUROPEAN STANDARD
NORME EURopEENNE
EUROPAIscHE NORM
ICS 11.040.10; 17.120.10
EN October 1998
Descriptors: medical equipment, anaesthetic equipment, artificial breathing apparatus, junctions, gas pipelines, medical gases,
flow measurements, measuring instruments, defirutions
DisJXJS,itifs a debilmetre
murales Durch:flu&neBeiruichtungen zum Anschl1 1S an
Entnahmestellen von Rohrleitungssytemen fUr medizinil§che Gase
CEN members are bound to comply with the CEN/CENELEC Internal Re~~tiOI'IS which stipulate the conditions for giving this Standard the status of a national standard without any alteration lists and
references such national standards may be obtained on application to
This StandaId exists in one official vemion in accordance with Resolution
by translation under the responsibility of a CEN member into its own language and notified to the Central Secretariat has the same status as the official versions
CEN members are the national standards bodies of Austria, Belgium, Czech
United Kingdom
CEN European Committee for Standardization Comit.e Europeen de Normalisation Europaisches Komitee ffir N,.",rnnln<r
Central Secretariat: rue de Stassart 8-1050 Brussels
©1998 CEN All rights of eXl)loiltatiion in any fonn and by allY means reserved worldwide for CEN national Members
Ref No EN 13220: 1998 E
Trang 4Page 2
Foreword
This European Standard has been by
anaesthetic the Secretariat of which is held
by BSI
This European Standard shall be the status of a
text or by endorsement, at the latest Aprillggg, and
conflicting national standards shall be withdrawn at
the latest by April 1999
This European Standard has been under a
mandate given to by the European CoJrnIllLission
supports essential requirements of EU llWective(s)
For relationship with ED Directive(s), see informative
annex ZA, which is an integrnl part of this standard
For special national conditions for 5.4.1, 7.2.1
and Thble 2, see annex D
Annex D forms a normative part of this I!:W:'ODIOOII
Standard Annexes C and ZA are given for
information only
According to the CENICENELEC Internal Re~DllatiOIIS,
the national standards organizations of the UJJJll,nllU':
countries are bound to implement this "~n'il)')j_n
Standard: Austria, Belgiwn, Czech Republic,
France, Germany, Greece, Iceland, Ireland,
Luxembourg, Netherlands, Norway, Portug:ll,
Sweden, Switzerland and the United KiJ1tgctl)m
Contents
Foreword Introduction
I
2 Normative references
3 Definitions
4
5 General requirements
6 Thst methods
8 Information to be supplied manufacturer
Annex A (informative) 1YPical of flow-metering systems
Annex B (informative) Bib,lioJuaJJQV Annex C (in1foffllatilve) Rationale Annex D (normative) Special national conditions Annex ZA (informative) Clauses of this European Standard addressing essential requirements or other provisions of EU Directives
2
3
3
3
3
4
4
6
7
8
12
15
15
15
16
BSI 04-1999
Trang 5Introduction
J:<'low-rne1JeruljO{ devices are widely used in the delivery
of medical gases supplied by a medical gas supply
system to a patient It is essential that these
devices deliver accurate flows under varying conditions
of temperature and inlet It is therefore
irnllOrtant that the characteristics be
This standard pays particular attention to:
sui1tability of materials;
(mechanical safe relief of excess
pressure and resistance to ignition);
- gas specificity;
- cleanliness;
-accuracy;
rruorrnationsupplied
Clauses and subclauses marked with B after their
in annexC
1 Scope
1.1 This European Standard applies to:
- flow-metering devices which are connected and
disconnected the operator at tenninal units of a
medical pipeline system for measurement and
medical gases; they can be connected
or by means of flexible 'n"'''ti.'C1
- flow-metering devices which are connected and
disconnected the operator at ~asesPleciflc
regulators
no'/V-nleUenfll! devices for the oxygen;
- nitrous oxide;
- air for breathing;
- carbon dioxide;
xenon;
specified mixtures of the gases listed
O]GrgeJVnlUOIlS oxide mixture (50 %/ 50 % VIV)
1.3 Electrical or electronic flow-metering devices are
excluded from the scope of this standard
© SST 04-1999
2 Nonnative references
Page 3
EN 13220:1998
undated reference, provisions from publications These normative references are cited at the
appropriate in the text and the publications are listed hereafter For dated subsequent anlendrnents to or revision of any of these apply to this European Standard onlY when inc1orpora.ted in it by anlendrnent or revision For undated references, the latest edition of the publication referred to atJIJU""'
campressed medical gases and VllC'Uum
ES 739, Low-pressure hose assemblies for use with
medical gases
EN 1441, Meaic.al devices - Risk (rm111J"~i_~
E~ 12218, Rail systems for su.pporting medical equipment
identification of content
3 Definitions
nu."""" " of this European Standard the defiinitions apply
flow gauge
gauge which measures pressure differential and which
is calibrated in units of flow NOTE The flow
upstream of a 3.2
flowmeter
indicates flow by measuring the pressure
device that measures and indicates the flow of a specific gas
3.3
n(1.w-m'\".4li1l'l~Ul device
device fitted with an inlet and an outlet connector and which incOlporates one of the following:
a) a flowmeter and a flow control b) a flow and a fixed orifice with a flow control
mUttlplC fixed orifices with a means of sel;ectIDg
3.4
gas-specific
having characteristics which prevent interchangeability and allow assignment to one gas or vacuum service only
Trang 64
EN 18220:1998
3.5
flWiI-sDe,clric connection point
that part of the socket which is the receptor for a
3.6
medical gas
any gas or mixture of gases intended to be
administered to patients for therapeutic, diagnostic or
prophylactic pruposes, or for surgical tool 3pI)llca:ticlR
3.7
medical gas pll)elltne
pipeline disltriblllti(J.n and terminal units at the
points where medical gases or anaesthetic gas
~:we'ruIime: can be n:><n11.C'f>l'I
3.8
medical gu supply ",i-
either:
a) a medical gas system; or
b) any other installation having no permanent
pipeline system but employing a medical gas supply
3.9
NIST connectors
non-interchangeable screw-threaded connectors
different a left or right-hand screw
gas
3.10
probe
non-interchangeable male component designed for
acceptance by and retention in the socket
3.11
rated inlet pressure,
rated maximum u~mam pressure for which the
3.12
single-fault condition
a hazard in is aelet-"tlVe or a single
external abnonnal condition is ~~£"'£""
3.13
socket
that part of a terminal unit which is either int.,,-.,,[ or
3.14
terminal unit
outlet assembly (inlet for vacUUI10) in a medical gas
connections and disconnections NOTE Terminal units and gas-specific connection points can also
be connected to pressure regulators or low-pressure hose
assemblies
4 Thrminology Typical examples of tlo'w-llnetJeliIU!: S1llStE~~'l are shown
in annexA
5 General requirements
5.1 Safety
.t.U~l<U.I,t:u, operated in normal use according to the instructions of the manUltact;un~, cause no safety hazard which could be foreseen using
risk procedures in accordance with EN 1441 and which is connected with their intended in
nonnal condition and in single-fault condition
5.2R Alternative construction
t+lOtw-:meterinj."{ devices and or
materials or having forms of
.l!iUrDJJean Standard (except for dimensions and allocation of connectors and probes used as inlet c01IDe:ctclm) shall be accepted if it can be
demonstrated that an equivalent degree of is
obtained
Such evidence shall be provided by the manufacturer
See annex D for special national conditions
5.3 Materials
5.3.1 The materials in contact with the shall be
cOl1np:JltibJle with oxygen, and with the medical gases and their in the teIlnPE!rature range specified in 5.3.2
NOTE 1 Corrosion resistance includes resistance against moisture and surrolmding materials
NOTE 2 with oxygen involves both combustibility and ease of which burn in air will burn violently
in pure oxygen Many materials which do not burn In air will do
so in pure under pressurR Similarly, materials which can in air require lower ignition energies in
oxygen materials can be ignited by friction at a valve
pressure is rapidly Into a system initially at low pressure
NOTE 3 13159 GompatWitity of medical equipment u>i.th
5.3.2 The m.ateri.als shall pemrit the flow-metering
of -20
© BSI 04-1999
Trang 75.3.3 Flow-metering devices shall be capable, while
environmental conditions as stated by the
manufacturer
5.8.4 Springs, highly-strained components and
liable to wear, which corne into contact with .>' 1
gases, shall not be plated
NOTE Plating could come off
5.3.5R Evidence of coruol:mity with the
of is.3.2, is.3.3 and shall be provided by the
manufacturer
5.4 Design requirements
5.4.1 Inlet connector
5.4.1.1 The inlet connector shall be one of the
following:
comp1ying with
A2a);
737-6:1996
b) a NIST nut and nipple complying with EN 739
A2b);
c) an consisting of a hose insert, a length
of hose and a probe complying with prEN 737.0:1996
FIgure A2c);
d) an assembly consisting of a hose insert, a length
of hose and a NIST nut and complying
with EN 739 Figure
5.4.1.2 These assemblies shall comply with EN 739,
except that the outlet connector is replaced by the
hose insert specified in EN 739
See annex D for special national conditions
COlnpUian<:e shall be checked visual inspection
5.4.2 Outlet connector
The outlet connector shall be one of the following:
a) a permanently connected hose insert;
b) a proprietary fitting, with or without a hose insert
Compliance shall be checked by visual inspection
is.4.3R Filtration
A filter shall be provided which:
a) is located of the now-metering device;
b) is reptac~~aI)le;
c) has not exceeding 100 IJ.IIl or equivalent
mesh
Evidence shall be provided by the manufacturer
5.4.4 Scales and indicators
5.4.4.1 All tlmv-metel:1ng
units of litres per minute
NOTE For flows of less than lllrnin the flow-metering devices
can be graduated in millUitres per minute (mI/min)
5.4.4.2 Unit graduations in flow-metering device scale
increments shall be not less than the stated accuracy
at a given flow
Page 5
13220:1998
5.4.4.3 The indicator of a flow-metering device shall
be visible to the user at all now rates, including zero flow
5.4.4.4 The scale of the now-metering device shall be
necessary, of 1, or standing 1 m the
at an illuminance of215lx 5.4.4.5 Compliance with the requirements of 5.4.4.1
to is.4.4.4 is checked by visual inspection
5.4.5 Meclumical strength
The no'w-nleOenrlg containing for 10
of 1 OOOkPa
C8J)ab.le of either HO"'''''''U'', a pressure The test for mechanical strength is given in 6.2
The accuracy of the flow at any flow graduation of a
flow-metering device shaH be within flO % of the indicated value for flows between 10% and 100% of
full scale or ±o,5 whichever is For flows below 10 % of full the accuracy be disclosed the manufacturer
For flow-metering devices for measuring flows less
than or equal to ll/min, the accuracy of the flow at
any flow graduation shall be within ilO % of the full
scale
The accuracy shall be measured throughout the range
of inlet pressure by the manufacturer; when the flow is discharged into ambient atmosphere, after the test for mechanical has been carned out The test for accuracy of flow is given in 6.3
NOTE To enhance accuracy and to reduce the hazard of static discharge, means should be to minimize the build-up of electrostatic charges both and outside the flowmeter tube and its housing
5.4.7 Flow control "aive
5.4.7.1 The control knob and the valve
be designed so that cannot be ruseng:ag~~ the use of a tool
The flow control valve shall be designed so that the flow increases when the knob is turned
counter-clockwise
5.4.7.2 Compliance with the requirements of5.4.7.1 shall be checked by visual IIlSpe(:tion
5.4.8 Leakage
The total internal with the flow control valve closed with a torque N·m, shall not exceed 0,3 ml/min (0,030 3 at after the tests for mechanical strength and accuracy have been carned out
The total external leakage (to the atmosphere) shall not exceed 0,5 mllmin (0,0506 kPa·llmin) at after tlle tests for mechanical and accuracy have been earned out
The test for leakage is in 6.4
Trang 8Page 6
EN 13220:1998
5.4.9B Resistance to ignition
For flow-metering devices for all gases, the
components in contact with the gas, including the
sealing materials and lubricants (if used), shall not be
lower than 160 °e
Evidence of with tllis requirement shall be
provided by the manufacturer
The test for the detennination of the auto-il!nitlon
temperature is in 6.5
oennitb!d operating temperature of tested material
safety margin is np'c:'~!il.C;;aTV because it covers an unforeseen increase of the "neraUl""
temperature and the fact that the auto-ignition temperature not
a constant In this connection, it should be emphasized that values
of the auto-ignition always depend on l.he lest method
used, which does not simulate all possible operating
conditions
5.4.10 Connection to rail systems
If a flow-metering fitted with an inlet connector
complying with 5.f.l.lc or 5.4.1.1d, is intended to be
"" ,.nn,r-t.oorl by a rail complying with EN it
shalJ be provided with one of the following devices
comp)ying with EN 12218:
a rail clamp;
- an equipment mount;
- an equipment mount pin
5.5 Constructional reqrull'emlenlts
5.5.1B Cleanlng
The components of tlow-lrneter:IIUZ devices for all
shall be supplied clean and free oil, grease
jJ ,\.-I , "" matter
Evidence shall be provided by the manufacturer
which effectively removes all surface dirt and hydrocarbons, and
which leaves no residue Itself Chemlcal cleaning methods will
normally require a subsequent washing and drying process to
remove residues
NOTE 2 of cleaning procedures will be described
13159 Clnll.patimlity 11Jith oxygen,
pre):larn.tion by
S.5.2R Lubricants
If lubricants are used, they shall be compatible with
oxygen, and with the other medical and their
m.i.xtures, in the temperature range specified in 5.3.2
up to the test pressure of 1 000 kPa
Evidence shall be provided by the manufacturer
S.5.3 Loosening torque
from flow-metering device shall be greater than or
to20N·m
The test for loosening torques is in 6.6
5.5.4 Flow setting torque
required at the maximum radil.L."i of the device
to change from the "off" position and from one setting
to another shall be not less than 5 N and not more than 50 N, and the control shall self-centre at each
setting
The test for torques is given in 6.6
5.5.liB Inadvertent disassembly
Means shall be to prevent inadvertent disassembly of the flow-metering device
Evidence shall be provided by the manufacturer
6 Test methods 6.1 Gene:ml
6.1.1 Ambient conditions
where otherwise out at (23 ± 2) °e
6.1.2 'lest gas
tests shall be carried
Tests shall be carried out with clean, oil-free dry
nitrogen or the gas, with a maximum moisture content of 50 tLg/g (50 corresponding to a dew
point of -48 °e at pressure
6.1.3 R~e,.ence condition
Flows shall be corrected to 23 °e and 101,3 kPa
If a flow-metering device is tested with a gas other
than the specific gas, flows shall be converted
6.2 Test method for mechanical strength
Compliance shall be detennined by using the lU1JlUVYll''''
ordered steps:
open the flow control the flow-metering device to a test pressure
('rn1t!lilnjTI~ the pressure for at
return the flow-metering device to and carry out the tests for accuracy and lealaij~e given
in 6.3 and 6.4
6.3 Test method for accuracy of flow
Test the accuracy of flow, the shown
in 1, after the test for mechaniclll strength ha.<;
been carried out
This test shall be carried out at the maximum and
© BSI 04-1999
Trang 96.4 Test method for leakage
6.4.1 Internal leakage
The device shall be tested with pressure Pl to
the inlet connector, with the flow control valve closed
with a torque of 0,4 N·m, after the tests for mechanical
strength and accuracy have been carried out
6.4.2 External leakage
The device shall be tested with pressure to
6.7 Test method for colour coding
Page 7
EN 13220:1998
of markings and
Rub markings and colour hand, without undue pressure, first for 15 s with a cloth rag soaked with distilled water, then for 15 s with a cloth rag soaked with methylated and then for 15 s with a cloth
rag soaked with isopropyl alcohol Carry out this test
at ambient tenlPe~ratm:e
and the outlet sealed or plugged, after the tests for
out
6.6 Test method for determining the
auto-ignition temperature of non-metallic
materials
maJterials, including the sealing materials
as follows, using the apparatus shown in
Place finelY divided test material into a stainless-steel
tube with a chromium-nickel steel cladding, in
quantities of about to g 1b obtain large
fibrous ceramic materiaL Fill the gas-.tight tube,
containing the with oxygen at a specified
pressure (see note 2) and then inductively heat it by a
low-frequency heater in an approximately linear
manner at 120°CIrnin Monitor the temperature of the
sample as a function of time by the use of a
thermocouple, and monitor the pressure by a pressure
transducer Record both pressure and temperature by a
dual-channel recorder The point at which spontaneous
ignition occurs is denoted a sudden rise in
temperature and pressure
temperature and the COITesPOIUlliU{
pressure can be seen from
NOTE 1 Auto-ignition tenlpe~rattlIes
generally be reproduced
to 200 °C Variations of about ·C and in some cases even
higher, are known to occur in the range from 200 °C to 500 0c
Usually five tests at the same pressure are performed on each
sample
materials depend upon the test and there are differences
in the values obtained by different test The
measurement of the temperature of non-metallic
materials is typically out at a of 4 000 kPa, and
data is unavailable for a pressure of kPa The
rel~,tiOJ1Shiip is that the auto-ignition temperature non-metallic
materials with increasing oxygen pressure to an
app,roximately constant value above of 4 000 kPa It is
known that this is not followed by
"' ,'"'''''''' the properties of new non-metallic materials which
might used for oxygen service
6.6 Test method for loosening and operating
torques
Mt'asure the torques using an apjClropri:ate torque
spanner
© BSI 04-1999
7.1.1 Flow-metering devices and their gas-specific components shall be durablY and marked with the symbol of the relevant in accordance with Table 1 In addition to the the name of the gas
may be used The test for durability of is
in 6.7
The indicated value of the flow shall be legible to an
visual corrected if necessary,
1, seated or standing I m the flow-metering with an illuminance of 215 Ix
In addition, flow-metering devices and their gas-specific components shall be marked with the following:
a) the name and/or the trademark of the manufacturer or distributor;
or serial number or year of the rated inlet pressure, Pl;
d) point of reference for the flow-indicating
of inlet pressure for which the tlow-lne1t:eIilIlg device is calibrated;
1) the words "USE NO OIL", or the symbol shown
NOTE National Janguages can be used for "l:SE NO OIL" 7.1.2 Flow control valves shall be
marked with the direction for increasing 7.1.3 Compliance with 7.1.1 and 7.1.2 shall be checked visual insl)e<:t:io,n
7.2 Colour <.:V<L1.U1lS 7.2.1 If colour accordance with See annex D for
is used, it shall be in
l
national conditions
7.2.2 Colour coding shall be durable
The test for dw-ability of colour coding is giwn in 6.7 7.3 Packaging
sealed to protect particulate contamination and packaged to nre1ient damage during storage and transportation
7.3.2 sha1l n "nl'IP a means of identification
of the contents
Trang 10PageS
EN 13220:1998
Table 1 Medical gases, mArkina and colonl'
Nitrous oxide
Carbon dioxide
Mixtures of the above gases 3)
l) According to ISO 32
2) An example of light brown is NCS 303O-Y30R in SS 0191 02
3) According to the components
NOTE National languages may be used for "Air"
S Infonnation to be supplied by
manufacturer
8.1 Flow-metering devices shall be acc:omparrled
docwnents conta.in.i.ng at least a technical
instructions for use and an address to which the user
can refer The accompanying documents shall be
<'''' ~'''iI as a component of flow-metering
8.2 Instructions for use shall contain all information
ne<;esl!!alY to operate the flow-metering device in
ac_c:ordaIlce with its and shall include
eXl~laJllation of the function of controls, connection and
disconnection of detachable parts and accessories
Instructions for use shall include indications on
rec:ogJrIizE~ accessories and detachable parts if the use
of accessories and parts can degrade the
minimwn safety Instructions for use shall detailed
instructions for the safe of cleaning,
indicate the of such activities A
list of recommended spare shall be provided The
"h,~ ,."",nn.,." on the flow-metering device shall be
eJq:)laiJled in the instructions for use
White Blue Black-white Brown Grey Light ulUw!1 2)
3)
8.3 Particular attention shall be satetv'-rellat€!G items:
a) the of fire or arising from the use of l"h,ri "ITIi'<: not recommended by the manufacturer;
b) the of changing the connectors on flow-metering devices;
c) the of disassembling the flow-metering device while it is under pressure;
the etIect on the accuracy of flow which can
result from inlet pressure;
e) the etIect on the accuracy of flow which can
result from varying outlet ret:~isbf.lnc:e;
f) the etIect on the accuracy of flow due to variations in ambient from 0 <>C to +40
g) the effect on mechanical accuracy and leakage of the Dow-metering device to trnnsport and storage outside the .range of temperatures 8p€Cified by the manufacturer
© BSI 04-1999