F 1390 – 97 Designation F 1390 – 97 Standard Test Method for Measuring Warp on Silicon Wafers by Automated Noncontact Scanning 1 This standard is issued under the fixed designation F 1390; the number[.]
Trang 1Standard Test Method for
Measuring Warp on Silicon Wafers by Automated
This standard is issued under the fixed designation F 1390; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1 Scope
1.1 This test method covers a noncontacting, nondestructive
procedure to determine the warp of clean, dry semiconductor
wafers
1.2 The test method is applicable to wafers 50 mm or larger
in diameter, and 100 µm (0.004 in.) approximately and larger in
thickness, independent of thickness variation and surface
finish, and of gravitationally-induced wafer distortion
1.3 This test method is not intended to measure the flatness
of either exposed silicon surface Warp is a measure of the
distortion of the median surface of the wafer
1.4 This test method measures warp of a wafer corrected for
all mechanical forces applied during the test Therefore, the
procedure described gives the unconstrained value of warp
This test method includes a means of canceling
gravity-induced deflection which could otherwise alter the shape of the
wafer.2 The resulting parameter is described by Warp(2) in
Appendix X2 Shape Decision Tree in SEMI Specification M 1
(See Annex A1.)
N OTE 1—Test Method F 657 measures median surface warp using a
three-point back-surface reference plane The back-surface reference
results in thickness variation being included in the recorded warp value.
The use (in this test method) of a median surface reference plane
eliminates this effect The use (in this test method) of a least-squares fit
reference plane reduces the variability introduced in three-point plane
calculations by choice of reference point location The use (in this test
method) of special calibration or compensating techniques minimizes the
effects of gravity-induced distortion of the wafer.
1.5 The values stated in SI units are to be regarded
sepa-rately as the standard The values given in parentheses are for
information only
1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use It is the
responsibility of the user of this standard to establish
appro-priate safety and health practices and determine the
applica-bility of regulatory limitations prior to use.
2 Referenced Documents
2.1 ASTM Standards:
F 657 Test Method for Measuring Warp and Total Thickness Variation on Silicon Slices and Wafers by Noncontact Scanning3
F 1241 Terminology of Silicon Technology3
2.2 SEMI Standard:
M 1 Specifications for Polished Monocrystalline Silicon Wafers4
3 Terminology
3.1 Definitions:
3.1.1 mechanical signature— of an instrument, that
compo-nent of a measurement that is introduced by the instrument and that is systematic, repeatable, and quantifiable
3.1.2 median surface—of a semiconductor wafer, the locus
of points equidistant from the front and back surfaces
3.1.3 quality area—that portion of a wafer within the
specified parameter is determined
3.1.4 reference plane— of a semiconductor wafer, a plane
from which deviations are measured
3.1.5 reference plane deviation (RPD)—the distance from a
point on a reference plane to the corresponding point on a wafer surface A dome-shaped wafer is considered to have positive RPD at its center; a bowl-shaped wafer is considered
to have negative RPD at its center
3.1.6 thickness—of a semiconductor wafer, the distance
through the wafer between corresponding points on the front and back surfaces
3.1.7 wafer—of a semiconductor, the difference between
the maximum and minimum distances of the median surface of
a free, unclamped wafer from a reference plane
3.1.7.1 Discussion—Although warp may be caused by
un-equal stresses on the two exposed surfaces of the wafer, it cannot be determined from measurements on a single exposed surface The median surface may contain regions with upward
or downward curvature or both; under some conditions the median surface may be flat (see figures in Appendix X1) 3.2 Other definitions relative to silicon material technology can be found in Terminology F 1241
1
This test method is under the jurisdiction of ASTM Committee F-1 on
Electronicsand is the direct responsibility of Subcommittee F01.06 on Silicon
Materials and Process Control.
Current edition approved June 10, 1997 Published August 1997 Originally
published as F 1390–92 Last previous edition F 1390–92{1.
2 Poduje, N., “Eliminating Gravitational Effect in Wafer Shape Measurements,”
NIST/ASTM/SEMI/SEMATECH Technology Conference, Dallas, TX; Technology
for Advanced Materials/Process Characterization, February 1, 1990.
3
Annual Book of ASTM Standards, Vol 10.05.
4 Available from SEMI, 805 East Middlefield Road, Mt View, CA 94043.
1
AMERICAN SOCIETY FOR TESTING AND MATERIALS
100 Barr Harbor Dr., West Conshohocken, PA 19428 Reprinted from the Annual Book of ASTM Standards Copyright ASTM
Trang 24 Summary of Test Method
4.1 A calibration procedure is performed In addition to
setting the instrument’s scale factor and other constants, this
procedure determines the mechanical signature of the
instru-ment and the effect of gravity on the wafer
4.2 The wafer is supported by a small-area chuck and both
external surfaces are simultaneously scanned along a
pre-scribed pattern by both members of an opposed pair of probes
4.3 The paired displacement values are used to construct a
median surface
4.4 The median surface is mathematically corrected for
gravitational effects and for mechanical signature of the
instru-ment
4.5 A least-squares reference plane is constructed from the
corrected median surface
4.6 The reference plane deviation (RPD) is calculated at
each measured point
4.7 Warp is reported as the algebraic difference between the
most positive RPD and the most negative RPD
5 Significance and Use
5.1 Warp can significantly affect the yield of semiconductor
device processing
5.2 Knowledge of this characteristic can help the producer
and consumer determine if the dimensional characteristics of a
specimen wafer satisfy given geometrical requirements
5.3 Changes in wafer warp during processing can adversely
affect subsequent handling and processing steps These
changes can also provide an important process monitoring
function
5.4 The test method is suitable for measuring the warp of
wafers used in semiconductor device processing in the
as-sliced, lapped, etched, polished, epitaxial, or other layer
condition and for monitoring thermal and mechanical effects
on the warp of wafers during device processing
5.5 Until the results of a planned interlaboratory evaluation
of this test method are established, use of this test method for
commercial transactions is not recommended unless the parties
to the test establish the degree of correlation that can be
obtained
6 Interferences
6.1 Any relative motion along the probe measuring axis
between the probes and the wafer holding device during
scanning will produce error in the measurement data Vibration
of the test specimen relative to the probe-measuring axis will
introduce error Such errors are minimized by system signature
analysis and correction algorithms Internal system monitoring
may also be used to correct non-repetitive and repetitive
system mechanical translations, and failure to provide such
corrections may cause errors
6.2 If a measured wafer differs substantially in diameter,
thickness, fiducials or crystal orientation from that used for the
gravitational compensation procedure, the results may be
incorrect Approximate errors for differences in diameter and
thickness are shown in Appendix X2 If the crystal orientation
of the sample to be measured differs from the crystal
orienta-tion of the gravity-compensaorienta-tion wafer, then the measured
warp value may differ from the actual warp value by up to
15 % Error tables for fiducial variation have not been gener-ated
6.3 Different methods for implementing gravitational pensation may give different results Varying levels of com-pleteness of implementing a method may also give different results
N OTE 2—The recommended method for gravitational compensation is representative wafer inversion, 5 since it allows the use of a single wafer
to establish the compensation that is subsequently applied to sample wafers The sample wafer inversion method requires that every wafer be measured twice, once in a normal and once in an inverted position, which increases measurement time and subjects the sample to additional han-dling Theoretical modeling requires only a single measurement per sample, however it does not address machine signature issues, nor is a rigorous theory presently known to exist.
6.4 Mechanical variations in wafer holding devices between systems may introduce measurement differences See 7.1.1 6.5 Most equipment systems capable of this measurement have a definite range of wafer thickness combined with warp (dynamic range) that can be accommodated without readjust-ment If the sample moves outside this dynamic range during either calibration or measurement, results may be in error An overrange signal can be used to alert the operator and mea-surement data examiners to this event
6.6 The quantity of data points and their spacing may affect the measurement results See 7.1.2
7 Apparatus
7.1 Warp-Measuring Equipment, consisting of wafer
hold-ing device, multiple-axis transport mechanism, probe assembly with indicator, and system controller/computer, including data processor and suitable software The system must be equipped with an overrange signal Instrument data reporting resolution shall be 100 nm or smaller
7.1.1 Wafer-Holding Device, for example a chuck whose
face is perpendicular to the measurement axis, and on which the wafer is placed for the measurement scan The diameter of the wafer holding device shall be 22-mm (0.9-in.) diameter, 33-mm (1.3-in.) diameter, or other value as agreed upon between using parties
7.1.2 Multiple-Axis Transport Mechanism, which provides a
means for moving the wafer-holding device, or the probe assembly, perpendicularly to the measurement axis in a con-trolled fashion in several axes This motion must permit data gathering over a prescribed scan pattern the entire quality area Maximum data point spacing to be used shall be 4 mm, or other value as agreed upon between using parties
7.1.3 Probe Assembly With Paired Non-Contacting Displacement-Sensing Probes, Probe Supports, and Indicator Unit—The probes shall be capable of independent
measure-ment of the distance between the probed site on each surface of the sample wafer and the motion plane The probes shall be mounted above and below the wafer in a manner so that the probed site on one surface of the wafer is opposite the probed site on the other The common axis of these probes is the
5 The representative wafer method of gravitational correction is covered by a patent held by ADE Corporation, 77 Rowe Street, Newton, MA 02166 Alternate methods are described in 13.1.2.
2
Trang 3measurement axis (see Fig 1 ) The probe separation D shall be
kept constant during calibration and measurement
Displacement resolution shall be 0.1 µm or better The probe
sensor size shall be 4 by 4 mm, or other value to be agreed
upon between using parties Systems employing either
representative wafer inversion of sample wafer inversion
methods for gravity compensation must provide precise
positioning in both measurement orientations so that
measurements are taken at identical locations for each
orientation of the sample
where:
D 5 the distance from probe b to probe a,
a 5 the distance from the top surface of the wafer to probe
a,
b 5 the distance from probe b to the bottom surface of the
water,
t 5 wafer thickness, (always a positive number) and
z 5 the distance between the wafer medium surface and
the point halfway between the upper and lower
probes
8 Materials
8.1 Set-up Masters, suitable to accomplish calibration and
standardization as recommended by the equipment
manufacturer
8.2 Reference Wafer, the warp value#20 µm with a data set
that is used to determine the level of agreement between the
system under test and the data set (see Annex A1)
8.3 Representative Wafer—If using the representative wafer
inversion method, a wafer identical in nominal diameter,
thickness, fiducials, composition and crystalline orientation to
those being measured is required for the calibration procedure
Its warp need not be known
9 Suitability of Test Equipment
9.1 The suitability of the test equipment shall be determined
with the use of a reference wafer and its associated data set in
accordance with the procedures of Annex A1, or by
performance of a statistically-based instrument repeatability study to ascertain whether the equipment is operating within the manufacturer’s stated specification for repeatability
N OTE 3—For further information on instrument repeatability studies contact Subcommittee F1.95.
9.2 Determination of degree of suitability is currently under investigation
10 Sampling
10.1 This test method is nondestructive and may be used on either 100 % of the wafers in a lot or on a sampling basis 10.1.1 If samples are to be taken, procedures for selecting the sample from each lot of wafers to be tested shall be agreed upon by the parties to the test, as shall the definition of what constitutes a lot
11 Calibration and Standardization
11.1 Calibrate in accordance with the manufacturer’s instructions
12 Procedure
12.1 Prepare the apparatus for measurement of wafers, including selection of diameter, peripheral fiducials, scan area, and data display/output functions
12.2 Introduce the test specimen into the measurement mechanism and initiate the measurement sequence
13 Calculation
13.1 The instrument is assumed to be direct reading with all necessary calculations performed internally and automatically
as follows:
13.1.1 The displacements (distances) between each probe and the nearest surface of the wafer are determined (in pairs) at
intervals along the scan pattern The distance between Probe a
and the nearest surface of the wafer is displacement value a.
The distance between Probe b and the nearest surface of the
wafer is displacement value b The probes are separated by the
distance, D, (see Fig 1) Half the difference of each pair of displacements (0.5[b − a]) yields the position z (along the
measurement axis) of the median surface of the wafer at each point with respect to a plane halfway between the upper and lower probes
13.1.2 Gravitational compensation is applied to the median surface by one of the following methods:
13.1.2.1 Representative Wafer Inversion— A wafer
representative of the lot is measured first in a normal and then
in an inverted position The median surface measurement values are determined at each measurement location in each sample orientation The results are added to obtain 0.5
(z normal + z inverted) This cancels the effect of the representative wafer’s shape while retaining the effect of gravity, resulting in
a value of each measurement point expressed as z gravity The effect of gravity on subsequent measurements on a sample
wafer is cancelled by subtracting z gravity from z normalto produce
z compensatedat each measurement point
N OTE 4—The representative wafer inversion technique deals not only with first-order gravitational effects, but also with other effects that may influence the measured value, such as wafer-periphery effects, some machine-specific signature, etc.
3
Trang 413.1.2.2 Sample Wafer Inversion—Each sample wafer is
measured in a normal position then in an inverted position The
median surface measurement values are determined at each
measurement location in each measurement orientation One
half the difference between the normal and inverted
measurement values at each point yields the sample wafer’s
gravity-compensated shape:
z com5z nor 2 z inv
where:
z com 5 compensated position at each measurement point,
z nor 5 normal position at each measurement point, and
z inv 5 inverted position at each measurement point
13.1.2.3 Theoretical Modeling—Measure each sample
wafer and apply gravitational correction that has been
developed from a theoretical model A rigorous model is not
known to exist although approximate corrections have been
calculated.6
13.1.3 A reference plane is constructed that is a
least-squares fit to the median surface z-position data at all points of
the scan pattern The z-value of the reference plane (z ref) is
subtracted from the measured z-position at each point at all the
points of the scan pattern to yield reference plane deviation
(RPD) at each point:
13.2 The difference between the largest (most positive) and
smallest (most negative) of the reference plane deviations is
taken as the warp:
13.3 Record the calculated warp value
13.4 For referee or other measurements where the wafer is
measured more than once, calculate the maximum, minimum, sample standard deviation, average and range of all measurements on the sample
14 Report
14.1 Report the following information:
14.1.1 Date, time and temperature of test, 14.1.2 Identification of operator,
14.1.3 Identification of measuring instruments, including wafer holding device diameter, data point spacing, sensor size, and gravitational correction method,
14.1.4 Lot identification, including nominal diameter and center point thickness,
14.1.5 Description of sampling plan, and 14.1.6 Warp of each wafer measured
14.2 For referee tests the report shall also include the standard deviation of each set of wafer measurements
15 Precision and Bias
15.1 Interlaboratory evaluation of this test method is planned to verify its suitability and reliability Until the results are established, use of this test method for commercial transactions is not recommended unless the parties to the test establish the degree of correlation that can be obtained 15.2 No standards exist against which the bias of this test method can be evaluated
N OTE 5—For further information on producing related reference materials to certify the wafer artifacts contact Subcommittee F1.95.
N OTE 6—Since no standard reference material exists for the measurement of warp, the measurement analysis shall include the capability to calibrate warp results to standards agreed upon between the participants in the measurement.
16 Keywords
16.1 noncontact measurement; semiconductor; shape; silicon; wafers; warp
ANNEX (Mandatory Information) A1 COMPARING DATA SETS A1.1 Introduction
A1.1.1 In qualifying a measurement system for operation, it
can be useful to compare the values ascribed to an artifact such
as a reference standard against those obtained for that artifact
on a machine under test This annex outlines a way in which
the multiple measurement data points that generate a
single-value quantity of warp can be used to monitor the effects of
Interferences more informatively than by using that
single-value alone
A1.1.2 A data set is that set of data used in computation of
warp It is corrected data, that is, all possible after interferences
have been removed and the data replanarized in accordance
with this test method
A1.1.3 A referee wafer (artifact) is accompanied by its own data set (referee data set (RDS)), in which each data point is the average of a number of values obtained for that point over a number of “passes” (repeat measurements) The artifact is measured on a machine under test and its RDS is compared against the resultant measured sample data set Delta-point, delta-warp and other values are computed from the differences The parameter used to determine agreement between the artifact and the system under test and the acceptable level of this agreement is to be agreed upon between the using parties
A1.2 Summary of Test Method
A1.2.1 Select a referee wafer of appropriate criteria, for which an RDS has been obtained
6 Application Note; Gravitational Sag in Silicon Wafers, ADE Corporation, 77
Rowe Street, Newton, MA 02166.
4
Trang 5A1.2.2 Measure the referee wafer on the machine under test
to obtain a sample data set (SDS)
A1.2.3 Subtract the two to obtain a difference data set
(DDS):
A1.2.4 The DDS represents the differences between the
measurements made on the machine under test and the referee
data set The DDS contains many values The simplest metric
that can be used to determine acceptability is maximum
difference, the largest absolute value in the DDS This
represents the worst-case disagreement between the machine
under test and the referee data
A1.2.5 Accept the machine as suitable for measurement if the maximum difference is less than a value that is agreed upon between the parties to the test
A1.2.6 More complex calculations may also be used, for example, a histogram of the (point-by-point) values of the DDS along with statistical measures (mean, sigma, etc.) may be compared These measures can be compared to application-specific limits or used to provide insight into the nature and source of the difference, or both
APPENDIXES (Nonmandatory Information) X1 VISUALIZATION OF WARP
X1.1 To calculate warp for a given case, it may be
convenient to transform the measurement geometry and to
consider the distance between the upper surface of the wafer
and a reference plane as d1, taken to be positive above the plane
and negative below, and the distance between the lower surface
of the wafer and the reference plane as d2, taken to be positive
below the plane and negative above it, as indicated in the
example in Fig X1.1
RPD5d12 d2
where:
RPD 5 reference plane deviation,
d1 5 distance between upper surface of the wafer, and
d2 5 distance between lower surface of the wafer
X1.2 See Fig X1.2 for examples of warped wafers with
stylized shapes Sample 4 in Fig X1.2 represents the example
Calculations for warp of each of these examples is given in
Table X1.1
FIG X1.1 Visualization of Warp
N OTE 1—T1is two units, T2is four units, and T3is three units; the warp values are calculated from Eq 2 The individual measured distances and the calculated differences are shown in Table X1.1.
FIG X1.2 Visualization of Warp—Stylized Examples
5
Trang 6X2 MEASUREMENT ERRORS DUE TO DIFFERENCES IN DIAMETER AND THICKNESS BETWEEN A REPRESENTATIVE
WAFER AND A WAFER UNDER TEST
X2.1 For small variations about the calibration values, the
relative change of the gravity effect is four times the relative
change of the diameter and minus two times the relative change
of thickness The following gives examples of gravity effect
errors (in micrometres), that is, deflection at the wafer edge
relative to the wafer center point, using a value of 1.603 10
12for Young’s modulus with the wafer supported horizontally
by a point located at the wafe center point Standard thickness and diameter tolerances in accordance with SEMI Specification
M 1 are used for the calculations, given in Fig X2.1 X2.2 The deflection induced in a wafer by gravity is modeled as follows:
TABLE X1.1 Values for Fig X1.2
6
Trang 7The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection with any item mentioned in this standard Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn Your comments are invited either for revision of this standard or for additional standards and should be addressed to ASTM Headquarters Your comments will receive careful consideration at a meeting of the responsible technical committee, which you may attend If you feel that your comments have not received a fair hearing you should make your views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.
Actual Diameter (mm) Typical Deflection (µm) for Nominal Thickness/
* * * * * * *
Actual Diameter (mm)
* * * * * * *
Actual Diameter (mm)
* * * * * * *
Actual Diameter (mm)
* * * * * * *
Actual Diameter (mm)
N OTE 1—the Deflection induced in a wafer by gravity is modeled as follows:
to get relative (9percentage9) errors:
S5kD 4
T2 5 gravity effect ~“sag”!
dS
dD 54
kD3
T2 dS
dD 5 22
kD4
T3 dS
S 54
dD D dS
S 5 22dT T
FIG X2.1 Examples of Gravity Effect Errors
7