1. Trang chủ
  2. » Luận Văn - Báo Cáo

Astm D 6438 - 05 (2015).Pdf

4 3 0

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Standard Test Method For Acetone, Methyl Acetate, And Parachlorobenzotrifluoride Content Of Paints, And Coatings By Solid Phase Microextraction-Gas Chromatography
Thể loại standard test method
Năm xuất bản 2015
Thành phố West Conshohocken
Định dạng
Số trang 4
Dung lượng 89,83 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

Designation D6438 − 05 (Reapproved 2015) Standard Test Method for Acetone, Methyl Acetate, and Parachlorobenzotrifluoride Content of Paints, and Coatings by Solid Phase Microextraction Gas Chromatogra[.]

Trang 1

Designation: D643805 (Reapproved 2015)

Standard Test Method for

Acetone, Methyl Acetate, and Parachlorobenzotrifluoride

Content of Paints, and Coatings by Solid Phase

This standard is issued under the fixed designation D6438; the number immediately following the designation indicates the year of

original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A

superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1 Scope

1.1 This test method is for the determination of acetone,

methyl acetate, or parachlorobenzotrifluoride (PCBTF), or

combination of any of the three, in paints and coatings, by solid

phase microextraction (SPME) headspace sampling, and

sub-sequent injection into a gas chromatograph It has been

evaluated for cellulose nitrate, acrylic, and urethane

solvent-borne systems The established working range of this test

method is: acetone, 28 to 90 %; methyl acetate, 12 to 22 %;

parachlorobenzotrifluoride, 10 to 17 % There is no reason to

believe that it will not work outside these ranges A minor

modification of this test method would make it suitable for the

analysis of the same analytes in water-borne coatings (seeNote

1)

N OTE 1—Water-borne paints are internally standardized and diluted

with water followed by addition of solid sodium chloride.

1.2 The values stated in SI units are to be regarded as

standard No other units of measurement are included in this

standard

1.3 This standard does not purport to address all of the

safety concerns, if any, associated with its use It is the

responsibility of the user of this standard to establish

appro-priate safety and health practices and determine the

applica-bility of regulatory limitations prior to use.

2 Referenced Documents

2.1 ASTM Standards:2

D3925Practice for Sampling Liquid Paints and Related

Pigmented Coatings

D6133Test Method for Acetone, p-Chlorobenzotrifluoride,

Methyl Acetate or t-Butyl Acetate Content of

Solvent-borne and WaterSolvent-borne Paints, Coatings, Resins, and Raw Materials by Direct Injection Into a Gas Chromatograph

E180Practice for Determining the Precision of ASTM Methods for Analysis and Testing of Industrial and Spe-cialty Chemicals(Withdrawn 2009)3

3 Terminology

3.1 Abbreviations:

3.1.1 PCBTF—parachlorobenzotrifluoride

(4-chlorobenzotrifluoride)

3.1.2 MCBTF—metachlorobenzotrifluoride

(3-chlorobenzotrifluoride)

3.1.3 SPME—solid phase microextraction 3.1.4 VOC—volatile organic compound 3.1.5 PEG/DVB—polyethylene glycol/divinylbenzene 3.1.6 FID—flame ionization detector

3.1.7 MS—mass selective or mass spectral 3.1.8 SIM—selected ion monitoring 3.1.9 GC—gas chromatograph 3.1.10 Sr—repeatability standard deviations 3.1.11 SR—reproducibility standard deviations 3.1.12 r—repeatability, 95 % confidence limit 3.1.13 R—reproducibility, 95 % confidence limit

4 Summary of Test Method

4.1 A suitable aliquot of whole solvent-borne paint is internally standardized and diluted with dioctyl phthalate The headspace of this solution is sampled with an SPME fiber, which is then thermally desorbed in the injection port of a gas chromatograph onto a suitable capillary column Either a flame ionization or mass specific detector may be used to measure peak areas of analytes and internal standards

1 This test method is under the jurisdiction of ASTM Committee D01 on Paint

and Related Coatings, Materials, and Applications and is the direct responsibility of

Subcommittee D01.21 on Chemical Analysis of Paints and Paint Materials.

Current edition approved June 1, 2015 Published June 2015 Originally

approved in 1999 Last previous edition approved in 2010 as D6438 – 05 (2010).

DOI: 10.1520/D6438-05R15.

2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or

contact ASTM Customer Service at service@astm.org For Annual Book of ASTM

Standards volume information, refer to the standard’s Document Summary page on

the ASTM website.

3 The last approved version of this historical standard is referenced on www.astm.org.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959 United States

Trang 2

5 Significance and Use

5.1 In order to calculate the volatile organic content (VOC)

of paints containing EPA exempt solvents, it is necessary to

know the acetone, methyl acetate, or

parachlorobenzotrifluo-ride content This gas chromatographic test method provides a

simple and direct way for measuring these solvents Each

analyte is measured with respect to a unique internal standard

For acetone, the internal standard used is acetone-d6, for

methyl acetate it is methyl acetate-d3, and for PCBTF it is

metachlorobenzotrifluoride (MCBTF) These unique analyte/

internal standard pairs behave very nearly as single solvents

with respect to evaporation rate and adsorption rate onto a

coated silica fiber (SPME) but are separable on a gas

chro-matograph (GC) capillary column The only critical analytical

technique required for successfully performing this test method

is the ability of an analyst to weigh a paint sample and internal

standard, corresponding to the analyte of interest, into a septum

capped vial After weighing, solvent evaporation has no effect

on the final value of the determination Since whole paint is not

injected into the gas chromatograph, the analytical system is

never compromised

6 Apparatus

6.1 Manual SPME Holder,4fitted with partially crosslinked

polyethylene glycol/divinylbenzene (PEG/DVB) fiber

assembly, 65–µm film thickness

6.2 Gas Chromatograph, FID Detection—Any capillary gas

chromatograph equipped with a flame ionization detector may

be used Temperature programming capability is desirable, but

isothermal operations may be utilized

6.2.1 For FID instrument conditions, seeTable 1

6.2.2 Inlet Liner, 0.75-mm should be placed in the injection

port.4

6.2.3 Integrator—Any electronic integrator that can

accu-rately quantify a gas chromatographic peak area is acceptable

6.3 Gas Chromatograph, Mass Selective (MS) Detection—

Any capillary gas chromatograph equipped with a mass selec-tive detector may be used The detector must be capable of measuring in the selected ion monitoring (SIM) mode at dwell times of 100 milliseconds or less

6.3.1 For MS instrument conditions, seeTable 2

6.3.2 Inlet liner, 0.75-mm, should be placed in the injection

port.4 6.3.3 The instrument should have a software data system to allow extraction and integration of the SIM ions

7 Column and Fiber Conditioning

7.1 Either or both capillary columns should be conditioned according to the manufacturer’s recommendation The col-umns may then be used indefinitely without further condition-ing

7.2 The SPME fiber should be conditioned and used accord-ing to the manufacturer’s recommendation

7.3 The SPME fiber should be inserted into a 260°C injection port for 30 s prior to daily use

8 Reagents and Materials

8.1 Purity of Reagents—Reagent grade chemicals shall be

used in all tests Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society, where such specifications are available.5Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination

8.2 Carrier Gas, Helium of 99.995 % or higher purity 8.3 Acetone -d6 , 99.9 % isotopic purity.

4 The sole source of supply of the manual holder and fibers known to the

committee at this time is Supelco Company, Supelco Park, Bellefonte, PA

16823-0048 If you are aware of alternative suppliers, please provide this

informa-tion to ASTM Internainforma-tional Headquarters Your comments will receive careful

consideration at a meeting of the responsible technical committee, 1

which you may attend.

5Reagent Chemicals, American Chemical Society Specifications, American

Chemical Society, Washington, DC For suggestions on the testing of reagents not

listed by the American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopia and National Formulary, U.S Pharmacopeial Convention, Inc (USPC), Rockville, MD.

TABLE 1 FID Instrument Conditions

Detector: Flame ionization

Column: 60 m × 0.25 mm 100 % polyethylene glycol, 0.5-µm

film thickness

Flow Rage: 1.0 mL per minute (20 cm per second)

Split Ratio: 200 to 1

Fiber desorption time: 5 to 6 sA

Temperature, ° C

Rate 1 30° per minute to 100°, hold 10 min

Rate 2 30° per minute to 240°, hold 2 min

AIf the fiber desorption is carried out longer than 5 or 6 s, the acetone peaks

exhibit tailing A 5 to 6–s desorption time transfers 98 to 99 % of the analytes to the

capillary column The fiber may be cleaned by inserting it into the GC inlet for 15

s after analytes of interest have eluted.

TABLE 2 MS Instrument Conditions

Detector: Electron ionization or mass selective Detection Mode: Selected ion monitoring (SIM) of ions m/e 58, 64,

74, 77, and 180 Dwell Time: 100 milliseconds or less Solvent Delay: 0.0 min

Column: 25 m × 0.20 mm 5 % phenyl/95 % methyl

polydimethylsiloxane

Flow Rate: 1.0 mL per minute (20 cm per second)

Fiber desorption time: 5 to 6 sA

Temperatures, ° C

Rate 1 10° per minute to 90°, hold 1 min Rate 2 40° per minute to 240°, hold 1 min

AIf the fiber desorption is carried out longer than 5 or 6 s, the acetone peaks exhibit tailing A 5 to 6–s desorption time transfers 98 to 99 % of the analytes to the capillary column The fiber may be cleaned by inserting it into the GC inlet for 15

s after analytes of interest have eluted.

Trang 3

8.4 Methyl Acetate -d3, 99.9 % isotopic purity (seeNote 2).

8.5 Parachlorobenzotrifluoride, technical grade, 99+ % (see

Note 3)

8.6 Metachlorobenzotrifluoride, 97 %, (seeNote 4)

8.7 Dedicated Syringes, 250-µL syringes, and Disposal

Syringes, 1.0 mL.

8.8 Septum Vials, 22 or 40 mL, with fluorocarbon-faced

silicone septa

N OTE 2—Methyl acetate-d3 was prepared by the acid-catalyzed

reac-tion of methanol with acetic-d3 acid, 99.9 % isotopic purity.

N OTE 3—Parachlorobenzotrifluoride contained 0.5 % of the meta

iso-mer and 0.1 % of the ortho isoiso-mer.

N OTE 4—The metachlorobenzotrifluoride standard was

chromato-graphically analyzed on a 60–meter 100 % polyethylene glycol column

(see 6.2 ) and exhibited only a single peak.

9 Preparation of Apparatus

9.1 Place 6-mL dioctyl phthalate into a 22 or 40-mL vial and

seal with a septum cap

9.2 Using dedicated 250-µL syringes, weigh approximately

100 mg each of parachlorobenzotrifluoride and

metachloroben-zotrifluoride to 0.1mg into the vial Excess solvent should be

wiped from the syringe needle prior to piercing the septum

Any solvent adhering to the outside of the septum cap after

making the transfer should be wiped off with tissue Weigh the

amounts of PCBTF and MCBTF to 0.1 mg Repeat the

procedure by adding approximately 150 mg each of acetone,

acetone-d6, methyl acetate, and methyl acetate-d3, and weigh

each component to 0.1 mg

9.3 Manually shake the vial for 15 s to mix the contents and

then let the vial stand at room temperature for 30 min

9.4 Remove the cap and wipe the septum to remove solvent

from the inner surface and reseal immediately, or replace the

septum with a new one

9.5 Sample the headspace with the SPME fiber for 60 s

9.6 Desorb the SPME fiber for 5 to 6 s in the injection port

of the gas chromatograph using either the conditions described

in6.2or6.3

9.7 Clean the fiber by placing it in the GC injection port for

15 s after the analytes of interest have passed through the

capillary column

9.8 If using the GC/FID procedure measure the peak areas

of the three analytes and three standards by integration (See

Note 5) Curve fitting software for measuring peak areas is

advantageous in some instances

N OTE 5—The approximate retention times in minutes are as follows:

acetone-d6, 10.35; acetone, 10.56; methyl acetate-d3, 11.06; methyl

acetate, 11.22: MCBTF, 22.33; PCBTF, 23.53.

9.9 If using the GC/MS procedure, measure the peak areas

for the extracted ions indicated inTable 3

9.10 Calculate response factors (RF) for acetone, methyl

acetate, and PCBTF according to the equation given in 10.4

The numerical values should agree within 1 % of each other

10 Procedure

10.1 Using a disposable 1-mL syringe, weigh 0.5 to 1.0 g of well mixed paint (see Note 6 and Practice D3925) into a septum-capped vial containing 6 mL of dioctyl phthalate The paint should drop directly into the dioctyl phthalate Carry out the same procedures as described in 9.3 – 9.7

N OTE 6—Since acetone and methyl acetate are extremely volatile, it is advisable to split the initial paint sample to be tested into two or more samples in suitable secondary containers.

10.2 Determine which, if any, of the three exempt solvents are present in the paint sample and if interfering peaks are present at the retention times corresponding to the retention times of the internal standards (see Note 7)

N OTE 7—The area of an interfering peak, if any, found in 10.2 , may be subtracted from the area of the standard peak when carrying out 10.5

10.3 Into a septum-capped vial containing 6 mL dioctyl phthalate, weigh 0.5 to 1.0 g of paint sample to 0.1 mg Add

100 mg of each of the internal standards, corresponding to the analytes found in 10.1 Carry out the same procedure as described in9.3 – 9.9

10.4 Calculate analyte concentrations according to the fol-lowing equations:

RF 5 AA 3 MI

and

% Analyte 5AA 3 MI 3 100

AI 3 RF 3 MC (2)

where:

RF = response factor,

AA = area of analyte,

MI = mass of internal standard,

AI = area of internal standard,

MA = mass of analyte, and

MC = mass of coating

10.5 Repeat the procedure given in 10.3 using paint and internal standard amounts such that the relative peak areas of the analyte and its internal standard chromatographic peak areas are approximately the same size A duplicate injection of the same sample should give relative peak areas that agree within 1 % of each other

10.6 Samples may also be analyzed for acetone content using Test Method D6133

11 Precision and Bias

11.1 Precision—A round robin was conducted involving

five laboratories All of the laboratories used the SPME

TABLE 3 Approximate Retention Times and Masses of Analytes

and Internal Standards

Analyte/Internal Standard m/e Retention Time, min

Trang 4

sampling technique Three of the laboratories used flame

ionization (FID) detection to measure analyte amounts and two

of the laboratories used mass spectral (MS) detection to

measure analyte amounts

11.2 Six commercial solvent-borne coatings were analyzed

Four of the coatings contained acetone and two of the same

four coatings contained both acetone and

parachlorobenzotri-fluoride Coatings containing methyl acetate were not yet

available in the marketplace and were therefore prepared by

adding known amounts of methyl acetate to different cellulose

nitrate lacquers

11.3 Precision data was calculated separately for the three

laboratories using FID detection and the two laboratories using

MS detection and also for the combination of FID and MS

detection Average values, repeatability standard deviations

(Sr), and reproducibility standard deviations (SR) are given for

each coating and exempt solvent inTable 4 andTable 5

11.4 The results were analyzed for precision using Practice

E180

11.5 Precision (Within-Laboratory, Between-Days

Variability, Formerly Called Repeatability)—The standard

de-viation of results (each the average of duplicates), obtained by

the same analyst on different days, have been estimated to be

0.42 % absolute at 39 df for acetone, 0.35 % absolute at 19 df

for parachlorobenzotrifluoride, and 0.16 % absolute at 19 df for

methyl acetate The 95 % confidence limit for the difference

between two such averages is 1.18 % absolute for acetone, 0.97 % absolute for parachlorobenzotrifluoride, and 0.46 % absolute for methyl acetate

11.6 Reproducibility (Multilaboratory)—The standard

de-viation of results (each of the average of duplicates), has been estimated to be 0.69 % absolute at 39 df for acetone, 0.53 % absolute at 19 df for parachlorobenzotrifluoride, and 0.25 % absolute at 19 df for methyl acetate The 95 % confidence limit for the difference between two such averages is 1.94 % absolute for acetone, 1.47 % absolute for parachlorobenzotrifluoride, and 0.70 % absolute for methyl acetate

11.7 Bias—Bias cannot be determined for this method

because there are no accepted standards for acetone, methyl acetate and parachlorobenzoptrifluoride in paints and coatings

12 Keywords

12.1 acetone; acetone content by GC; gas chromatography (GC); head space sampling; methyl acetate; methyl acetate content by GC; parachlorobenzotrifluoride (PCBTF); PCBTF content by gas chromatography; solid phase microextraction (SPME); VOC

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned

in this standard Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk

of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and

if not revised, either reapproved or withdrawn Your comments are invited either for revision of this standard or for additional standards

and should be addressed to ASTM International Headquarters Your comments will receive careful consideration at a meeting of the

responsible technical committee, which you may attend If you feel that your comments have not received a fair hearing you should

make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,

United States Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above

address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website

(www.astm.org) Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222

Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

TABLE 4 Precision Data for Acetone

TABLE 5 Precision Data for Parachlorobenzotrifluoride and

Methyl Acetate

trifluoride

Methyl AcetateA

ACellulose nitrate lacquers obtained from different manufacturers were diluted under carefully controlled laboratory condition to give samples E and F The calculated values for methyl acetate in these two samples were: Sample E, 21.2 %; Sample F, 12.0 %.

Ngày đăng: 03/04/2023, 21:40

TỪ KHÓA LIÊN QUAN

TÀI LIỆU CÙNG NGƯỜI DÙNG

  • Đang cập nhật ...

TÀI LIỆU LIÊN QUAN