1. Trang chủ
  2. » Tất cả

Astm d 323 15a

11 2 0

Đang tải... (xem toàn văn)

Tài liệu hạn chế xem trước, để xem đầy đủ mời bạn chọn Tải xuống

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Standard Test Method For Vapor Pressure Of Petroleum Products (Reid Method)
Thể loại Standard test method
Năm xuất bản 2015
Định dạng
Số trang 11
Dung lượng 256,53 KB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

Designation D323 − 15a Standard Test Method for Vapor Pressure of Petroleum Products (Reid Method)1 This standard is issued under the fixed designation D323; the number immediately following the desig[.]

Trang 1

Designation: D32315a

Standard Test Method for

This standard is issued under the fixed designation D323; the number immediately following the designation indicates the year of

original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A

superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

This standard has been approved for use by agencies of the U.S Department of Defense.

1 Scope*

1.1 This test method covers procedures for the

determina-tion of vapor pressure (seeNote 1) of gasoline, volatile crude

oil, and other volatile petroleum products

1.2 Procedure A is applicable to gasoline and other

petro-leum products with a vapor pressure of less than 180 kPa

(26 psi)

1.3 Procedure B may also be applicable to these other

materials, but only gasoline was included in the interlaboratory

test program to determine the precision of this test method

1.4 Procedure C is for materials with a vapor pressure of

greater than 180 kPa (26 psi)

1.5 Procedure D for aviation gasoline with a vapor pressure

of approximately 50 kPa (7 psi)

NOTE 1—Because the external atmospheric pressure is counteracted by

the atmospheric pressure initially present in the vapor chamber, the Reid

vapor pressure is an absolute pressure at 37.8 °C (100 °F) in kilopascals

(pounds-force per square inch) The Reid vapor pressure differs from the

true vapor pressure of the sample due to some small sample vaporization

and the presence of water vapor and air in the confined space.

1.6 This test method is not applicable to liquefied petroleum

gases or fuels containing oxygenated compounds other than

methyl t-butyl ether (MTBE) For determination of the vapor

pressure of liquefied petroleum gases, refer to Test Method

D1267or Test MethodD6897 For determination of the vapor

pressure of gasoline-oxygenate blends, refer to Test Method

D4953 The precision for crude oil has not been determined

since the early 1950s (see Note 3) Test Method D6377 has

been approved as a method for determination of vapor pressure

of crude oil IP 481 is a test method for determination of the

air-saturated vapor pressure of crude oil

1.7 The values stated in SI units are to be regarded as the

standard The values given in parentheses are for information

only

1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use It is the responsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use Specific warning

statements are given in Sections 7 and18, and 12.5.3, 15.5, 21.2,A1.1.2,A1.1.6, andA2.3

2 Referenced Documents

2.1 ASTM Standards:2

D1267Test Method for Gage Vapor Pressure of Liquefied Petroleum (LP) Gases (LP-Gas Method)

D4057Practice for Manual Sampling of Petroleum and Petroleum Products

D4175Terminology Relating to Petroleum, Petroleum Products, and Lubricants

D4953Test Method for Vapor Pressure of Gasoline and Gasoline-Oxygenate Blends (Dry Method)

D6377Test Method for Determination of Vapor Pressure of Crude Oil: VPCRx(Expansion Method)

D6897Test Method for Vapor Pressure of Liquefied Petro-leum Gases (LPG) (Expansion Method)

E1Specification for ASTM Liquid-in-Glass Thermometers E2251Specification for Liquid-in-Glass ASTM Thermom-eters with Low-Hazard Precision Liquids

2.2 Energy Institute Standards:

IP 481Test Method for Determination of the Air Saturated Vapour Pressure (ASVP) of Crude Oil3

3 Terminology

3.1 Definitions:

3.1.1 Bourdon spring gauge, n—pressure measuring device

that employs a Bourdon tube connected to an indicator

3.1.2 Bourdon tube, n—flattened metal tube bent to a curve

that straightens under internal pressure

1 This test method is under the jurisdiction of ASTM Committee D02 on

Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of

Subcommittee D02.08 on Volatility.

Current edition approved June 1, 2015 Published June 2015 Originally

approved in 1930 Last previous edition approved in 2015 as D323 – 15 DOI:

10.1520/D0323-15A.

2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or

contact ASTM Customer Service at service@astm.org For Annual Book of ASTM Standards volume information, refer to the standard’s Document Summary page on

the ASTM website.

3 Available from Energy Institute, 61 New Cavendish St., London, WIG 7AR, U.K., http://www.energyinst.org.uk.

*A Summary of Changes section appears at the end of this standard

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959 United States

Trang 2

3.1.3 oxygenate, n—oxygen-containing ashless organic

compound, such as an alcohol or ether, which may be used as

3.1.4 Reid vapor pressure (RVP), n—resultant total pressure

reading, corrected for measuring error, of a specific empirical

test method (Test Method D323) for measuring the vapor

pressure of gasoline and other volatile products

3.1.5 vapor pressure, n—pressure exerted by the vapor of a

liquid when in equilibrium with the liquid D4175

3.2 Abbreviations:

3.2.1 ASVP, n—air saturated vapor pressure.

3.2.2 LPG, n—liquefied petroleum gases.

3.2.3 MTBE, n—methyl t-butyl ether.

3.2.4 RVP, n—Reid Vapor Pressure.

4 Summary of Test Method

4.1 The liquid chamber of the vapor pressure apparatus is

filled with the chilled sample and connected to the vapor

chamber that has been heated to 37.8 °C (100 °F) in a bath The

assembled apparatus is immersed in a bath at 37.8 °C (100 °F)

until a constant pressure is observed The reading, suitably

corrected, is reported as the Reid vapor pressure

4.2 All four procedures utilize liquid and vapor chambers of

the same internal volume Procedure B utilizes a

semiauto-matic apparatus immersed in a horizontal bath and rotated

while attaining equilibrium Either a Bourdon gauge or

pres-sure transducer may be used with this procedure Procedure C

utilizes a liquid chamber with two valved openings Procedure

D requires more stringent limits on the ratio of the liquid and

vapor chambers

5 Significance and Use

5.1 Vapor pressure is an important physical property of

volatile liquids This test method is used to determine the vapor

pressure at 37.8 °C (100 °F) of petroleum products and crude

oils with initial boiling point above 0 °C (32 °F)

5.2 Vapor pressure is critically important for both

automo-tive and aviation gasolines, affecting starting, warm-up, and

tendency to vapor lock with high operating temperatures or

high altitudes Maximum vapor pressure limits for gasoline are

legally mandated in some areas as a measure of air pollution

control

5.3 Vapor pressure of crude oils is of importance to the

crude producer and the refiner for general handling and initial

refinery treatment

5.4 Vapor pressure is also used as an indirect measure of the

evaporation rate of volatile petroleum solvents

6 Apparatus

6.1 The required apparatus for Procedures A, C, and D is

described inAnnex A1 Apparatus for Procedure B is described

inAnnex A2

7 Hazards

7.1 Gross errors can be obtained in vapor pressure

measure-ments if the prescribed procedure is not followed carefully The

following list emphasizes the importance of strict adherence to the precautions given in the procedure:

7.1.1 Checking the Pressure Gauge—Check all gauges

against a pressure measuring device (seeA1.6) after each test

to ensure higher precision of results (see12.4) Read the gauge while in a vertical position and after tapping it lightly

7.1.2 Checking for Leaks—Check all apparatus before and

during each test for both liquid and vapor leaks (seeNote 5)

7.1.3 Sampling—Because initial sampling and the handling

of samples will greatly affect the final results, employ the utmost precaution and the most meticulous care to avoid losses through evaporation and even slight changes in composition (see Section8and12.1) In no case shall any part of the Reid apparatus itself be used as the sample container prior to actually conducting the test

7.1.4 Purging the Apparatus—Thoroughly purge the

pres-sure gauge, the liquid chamber, and the vapor chamber to be sure that they are free of residual sample This is most conveniently done at the end of the test in preparation for the next test (see12.5and15.5)

7.1.5 Coupling the Apparatus—Carefully observe the

re-quirements of12.2

7.1.6 Shaking the Apparatus—Shake the apparatus

vigor-ously as directed to ensure equilibrium

8 Sampling

8.1 The extreme sensitivity of vapor pressure measurements

to losses through evaporation and the resulting changes in composition is such as to require the utmost precaution and the most meticulous care in the handling of samples The provi-sions of this section shall apply to all samples for vapor pressure determinations, except as specifically excluded for samples having vapor pressures above 180 kPa (26 psi); see Section19

8.2 Sampling shall be done in accordance with Practice D4057

8.3 Sample Container Size—The size of the sample

con-tainer from which the vapor pressure sample is taken shall be

1 L (1 qt) It shall be 70 % to 80 % filled with sample 8.3.1 The present precision statement has been derived using samples in 1 L (1 qt) containers However, samples taken

in containers of other sizes as prescribed in PracticeD4057can

be used if it is recognized that the precision could be affected

In the case of referee testing, the 1 L (1 qt) sample container shall be mandatory

8.4 The Reid vapor pressure determination shall be per-formed on the first test specimen withdrawn from the sample container The remaining sample in the container cannot be used for a second vapor pressure determination If necessary, obtain a new sample

8.4.1 Protect samples from excessive heat prior to testing 8.4.2 Do not test samples in leaky containers They should

be discarded and new samples obtained

8.5 Sampling Handling Temperature—In all cases, cool the

sample container and contents to 0 °C to 1 °C (32 °F to 34 °F) before the container is opened Sufficient time to reach this temperature shall be ensured by direct measurement of the

Trang 3

temperature of a similar liquid in a like container placed in the

cooling bath at the same time as the sample

9 Report

9.1 Report the result observed in12.4or15.4, after

correct-ing for any difference between the gauge and the pressure

measuring device (seeA1.6), to the nearest 0.25 kPa (0.05 psi)

as the Reid vapor pressure

10 Precison and Bias

10.1 The following criteria are to be used for judging the

acceptability of results (95 % confidence):

10.1.1 Repeatability—The difference between successive

test results obtained by the same operator with the same

apparatus under constant operating conditions on identical test

material would, in the long run, in the normal and correct

operation of the test method, exceed the following value only

in one case in twenty

Range Repeatability

D Aviation

10.1.2 Reproducibility—The difference between two, single

and independent results, obtained by different operators

work-ing in different laboratories on identical test material would, in

the long run, in the normal and correct operation of the test

method, exceed the following value only in one case in twenty

Range Reproducibility

D Aviation

NOTE 2—These precision values are derived from a 1987 cooperative

program 4 and the current Committee D02 Statistical Method

RR:D02-1007.

NOTE 3—These precision values were developed in the early 1950’s,

prior to the current statistical evaluation method.

10.2 Bias:

10.2.1 Absolute Bias—Since there is no accepted reference

material suitable for determining the bias for this test method,

bias cannot be determined The amount of bias between this

test vapor pressure and true vapor pressure is unknown

10.2.2 Relative Bias—There is no statistically significant

bias between Procedures A and B for gasolines as determined

in the last cooperative test program

PROCEDURE A FOR PETROLEUM PRODUCTS HAVING REID VAPOR PRESSURES BELOW 180 kPa (26 psi)

11 Preparation for Test

11.1 Verification of Sample Container Filling—With the

sample at a temperature of 0 °C to 1°C, take the container from the cooling bath or refrigerator and wipe dry with absorbent material If the container is not transparent, unseal it, and using

a suitable gauge, confirm that the sample volume equals 70 %

to 80 % of the container capacity (seeNote 4) If the sample is contained in a transparent glass container, verify that the container is 70 % to 80 % full by suitable means (seeNote 4)

NOTE 4—For nontransparent containers, one way to confirm that the sample volume equals 70 % to 80 % of the container capacity is to use a dipstick that has been pre-marked to indicate the 70 % and 80 % container capacities The dipstick should be of such material that it shows wetting after being immersed and withdrawn from the sample To confirm the sample volume, insert the dipstick into the sample container so that it touches the bottom of the container at a perpendicular angle, before removing the dipstick For transparent containers, using a marked ruler or

by comparing the sample container to a like container that has the 70 % and 80 % levels clearly marked, has been found suitable.

11.1.1 Discard the sample if its volume is less than 70 % of the container capacity

11.1.2 If the container is more than 80 % full, pour out enough sample to bring the container contents within the 70 %

to 80 % range Under no circumstances shall any sample poured out be returned to the container

11.1.3 Reseal the container, if necessary, and return the sample container to the cooling bath

11.2 Air Saturation of Sample in Sample Container: 11.2.1 Nontransparent Containers—With the sample again

at a temperature between 0 °C and 1 °C, take the container from the cooling bath, wipe it dry with an absorbent material, remove the cap momentarily taking care that no water enters, reseal, and shake vigorously Return it to the cooling bath for

a minimum of 2 min

11.2.2 Transparent Containers—Since11.1does not require that the sample container be opened to verify the sample capacity, it is necessary to unseal the cap momentarily before resealing it so that samples in transparent containers are treated the same as samples in nontransparent containers After per-forming this task, proceed with11.2.1

11.2.3 Repeat11.2.1twice more Return the sample to the bath until the beginning of the procedure

11.3 Preparation of Liquid Chamber—Completely immerse

the open liquid chamber in an upright position and the sample transfer connection (seeFig A1.2) in a bath at a temperature between 0 °C and 1 °C (32 °F and 34 °F) for at least 10 min

11.4 Preparation of Vapor Chamber—After purging and

rinsing the vapor chamber and pressure gauge in accordance with 12.5, connect the gauge to the vapor chamber Immerse

4 Supporting data have been filed at ASTM International Headquarters and may

be obtained by requesting Research Report RR:D02-1245.

Trang 4

the vapor chamber to at least 25.4 mm (1 in.) above its top in

the water bath maintained at 37.8 °C 6 0.1 °C (100 °F 6

0.2 °F) for not less than 10 min just prior to coupling it to the

liquid chamber Do not remove the vapor chamber from the

bath until the liquid chamber has been filled with sample, as

described in12.1

12 Procedure

12.1 Sample Transfer—Remove the sample from the

cool-ing bath, uncap, and insert the chilled transfer tube (seeFig 1)

Remove the liquid chamber from the cooling bath, and place

the chamber in an inverted position over the top of the transfer

tube Invert the entire system rapidly so that the liquid chamber

is upright with the end of the transfer tube approximately 6 mm

(0.25 in.) from the bottom of the liquid chamber Fill the

chamber to overflowing (in addition to other precautions, make

provision for suitable containment and disposal of the

over-flowing sample to avoid fire hazard) Withdraw the transfer

tube from the liquid chamber while allowing the sample to

continue flowing up to complete withdrawal

12.2 Assembly of Apparatus—Immediately remove the

va-por chamber from the water bath and couple the filled liquid

chamber to the vapor chamber as quickly as possible without

spillage When the vapor chamber is removed from the water

bath, connect it to the liquid chamber without undue movement

that could promote exchange of room temperature air with the

37.8 °C (100 °F) air in the chamber Not more than 10 s shall

elapse between removing the vapor chamber from the water

bath and completion of the coupling of the two chambers

12.3 Introduction of the Apparatus into Bath—Turn the

assembled apparatus upside down and allow all the sample in

the liquid chamber to drain into the vapor chamber With the

apparatus still inverted, shake it vigorously eight times up and

down With the gauge end up, immerse the assembled

appara-tus in the bath, maintained at 37.8 °C 6 0.1 °C (100 °F 6

0.2 °F), in an inclined position so that the connection of the

liquid and vapor chambers is below the water level and

carefully examine for leaks (see Note 5) If no leaks are

observed, further immerse the apparatus to at least 25 mm

(1 in.) above the top of the vapor chamber Observe the

apparatus for leaks throughout the test and discard the test at

anytime a leak is detected

NOTE 5—Liquid leaks are more difficult to detect than vapor leaks Pay particular attention to the coupling between the chambers, which is normally in the liquid section of the apparatus.

12.4 Measurement of Vapor Pressure—After the assembled

apparatus has been in the water bath for at least 5 min, tap the pressure gauge lightly and observe the reading Withdraw the apparatus from the bath and repeat the instructions of12.3 At intervals of not less than 2 min, tap the gauge, observe the reading, and repeat 12.3 until a total of not less than five shakings and gauge readings have been made Continue this procedure, as necessary, until the last two consecutive gauge readings are the same, indicating that equilibrium has been attained Read the final gauge pressure to the nearest 0.25 kPa (0.05 psi) and record this value as the uncorrected vapor pressure of the sample Without undue delay, remove the pressure gauge from the apparatus (see Note 6) without attempting to remove any liquid that may be trapped in the gauge, check its reading against that of the pressure measuring device (seeA1.6) while both are subjected to a common steady pressure that is within 1.0 kPa (0.2 psi) of the recorded uncorrected vapor pressure If a difference is observed between the pressure measuring device and the pressure gauge readings, the difference is added to the uncorrected vapor pressure when the pressure measuring device reading is higher, or subtracted from the uncorrected vapor pressure when the pressure mea-suring device reading is lower, and the resulting value recorded

as the Reid vapor pressure of the sample

NOTE 6—Cooling the assembly prior to disconnecting the gauge will facilitate disassembly and reduce the amount of hydrocarbon vapors released into the room.

12.5 Preparation of Apparatus for Next Test:

12.5.1 Thoroughly purge the vapor chamber of residual sample by filling it with warm water above 32 °C (90 °F) and allowing it to drain Repeat this purging at least five times Purge the liquid chamber in the same manner Rinse both chambers and the transfer tube several times with petroleum naphtha, then several times with acetone, then blow dry using dried air Place the liquid chamber in the cooling bath or refrigerator in preparation for the next test

12.5.2 If the purging of the vapor chamber is done in a bath,

be sure to avoid small films of floating sample by keeping the bottom and top openings of the chamber closed as they pass through the water surface

12.5.3 Preparation of Gauge—Disconnect the gauge from

its manifold connection with the pressure measuring device and remove trapped liquid in the Bourdon tube of the gauge by repeated centrifugal thrusts This is accomplished in the following manner: hold the gauge between the palms of the hands with the right palm on the face of the gauge and the threaded connection of the gauge forward Extend the arms forward and upward at an angle of 45° Swing the arms rapidly downward through an arc of about 135° so that centrifugal force aids gravity in removing trapped liquid Repeat this operation at least three times or until all liquid has been expelled from the gauge Connect the gauge to the vapor chamber with the liquid connection closed and place in the 37.8 °C (100 °F) bath to condition for the next test

(Warning—Do not leave the vapor chamber with the gauge

FIG 1 Simplified Sketches Outlining Method Transferring

Sample to Liquid Chamber from Open-Type Containers

Trang 5

attached in the water bath for a longer period of time than

necessary to condition for the next test Water vapor can

condense in the Bourdon tube and lead to erroneous results.)

PROCEDURE B FOR PETROLEUM PRODUCTS HAVING REID

VAPOR PRESSURES BELOW 180 kPa (26 psi),

(HORIZONTAL BATH)

13 Sampling

13.1 Refer to Section8

14 Preparation for Test

14.1 Refer to Section11

15 Procedure

15.1 Sample Transfer— Remove the sample from the

cool-ing bath, uncap, and insert the chilled transfer tube (seeFig 1)

Remove the liquid chamber from the cooling bath, and place

the chamber in an inverted position over the top of the transfer

tube Invert the entire system rapidly so that the liquid chamber

is upright with the end of the transfer tube approximately 6 mm

(0.25 in.) from the bottom of the liquid chamber Fill the

chamber to overflowing (in addition to other precautions, make

provision for suitable containment and disposal of the

over-flowing sample to avoid fire hazard) Withdraw the transfer

tube from the liquid chamber while allowing the sample to

continue flowing up to complete withdrawal

15.2 Assembly of Apparatus—Immediately remove the

va-por chamber from the water bath Disconnect the spiral tubing

at the quick action disconnect Couple the filled liquid chamber

to the vapor chamber as quickly as possible without spillage or

movement that could promote exchange of room temperature

air with the 37.8 °C (100 °F) air in the vapor chamber Not

more than 10 s shall elapse between removing the vapor

chamber from the water bath and completion of the coupling of

the two chambers

15.3 Introduction of the Apparatus into the Bath—While

holding the apparatus vertically, immediately reconnect the

spiral tubing at the quick action disconnect Tilt the apparatus

between 20° and 30° downward for 4 s or 5 s to allow the

sample to flow into the vapor chamber without getting into the

tube extending into the vapor chamber from the gauge, or

pressure transducer Place the assembled apparatus into the

water bath maintained at 37.8 °C 6 0.1 °C (100 °F 6 0.2 °F)

in such a way that the bottom of the liquid chamber engages the

drive coupling and the other end of the apparatus rests on the

support bearing Turn on the switch to begin the rotation of the

assembled liquid-vapor chambers Observe the apparatus for

leakage throughout the test (see Note 5) Discard the test at

anytime a leak is detected

15.4 Measurement of Vapor Pressure—After the assembled

apparatus has been in the bath for at least 5 min, tap the

pressure gauge lightly and observe the reading Repeat the

tapping and reading at intervals of not less than 2 min, until

two consecutive readings are the same (Tapping is not

necessary with the transducer model but the reading intervals

should be the same.) Read the final gauge or transducer pressure to the nearest 0.25 kPa (0.05 psi) and record this value

as the uncorrected vapor pressure Without undue delay, disconnect the gauge from the apparatus Connect the gauge or pressure transducer to a pressure measuring device Check its reading against that of the pressure measuring device while both are subjected to a common steady pressure that is within 1.0 kPa (0.2 psi) of the recorded uncorrected vapor pressure If

a difference is observed between the pressure measuring device and gauge or transducer readings, the difference is added to the uncorrected vapor pressure when the pressure measuring device reading is higher, or subtracted from the uncorrected vapor pressure when the pressure measuring device reading is lower, and the resulting value recorded as the Reid vapor pressure of the sample

15.5 Preparation of Apparatus for Next Test:

15.5.1 Thoroughly purge the vapor chamber of residual sample by filling it with warm water above 32 °C (90 °F) and allowing it to drain Repeat this purging at least five times Purge the liquid chamber in the same manner Rinse both chambers and the transfer tube several times with petroleum naphtha, then several times with acetone, then blow dry using dried air Place the liquid chamber in the cooling bath or

refrigerator in preparation for the next test (Warning—Do not

leave the vapor chamber with the gauge attached in the water bath for a longer period of time than necessary to condition for the next test Water vapor can condense in the Bourdon tube and lead to erroneous results.)

15.5.2 If the purging of the vapor chamber is done in a bath,

be sure to avoid small films of floating sample by keeping the bottom and top openings of the chamber closed as they pass through the water surface

15.5.3 Preparation of Gauge or Transducer—In the correct

operation of this procedure, liquid should not reach the gauge

or transducer If it is observed or suspected that liquid has reached the gauge, purge the gauge as described in12.5.3 The transducer has no cavity to trap liquid Ensure that no liquid is present in the T handle fitting or spiral tubing by forcing a stream of dry air through the tubing Connect the gauge or transducer to the vapor chamber with the liquid connection closed and place in the 37.8 °C (100 °F) bath to condition for the next test

PROCEDURE C FOR PETROLEUM PRODUCTS HAVING REID VAPOR PRESSURES ABOVE 180 kPa (26 psi)

16 Introduction

16.1 With products having vapor pressure over 180 kPa (26 psi) (seeNote 7), the procedure described in Sections8 –

12 is hazardous and inaccurate Consequently, the following sections define changes in apparatus and procedure for the determinations of vapor pressures above 180 kPa Except as specifically stated, all the requirements of Sections1 – 12shall apply

NOTE 7—If necessary, either Procedure A or B can be used to determine

if the vapor pressure of a product is above 180 kPa.

Trang 6

17 Apparatus

17.1 Apparatus as described inAnnex A1using the liquid

chamber with two openings

17.2 Pressure Gauge Calibration—A dead weight tester

(see A1.7) may be used as a pressure measuring device (see

A1.6) for checking gauge readings above 180 kPa (26 psi) In

7.1.1,9.1,12.4, and12.5.3where the words pressure

measur-ing device and pressure measurmeasur-ing device readmeasur-ing appear,

include as an alternative dead weight tester and calibrated

gauge reading, respectively.

18 Hazards

18.1 The precaution in7.1.6shall not apply

19 Sampling

19.1 Paragraphs8.3,8.3.1,8.4, and8.5shall not apply

19.2 Sample Container Size—The size of the sample

con-tainer from which the vapor pressure sample is taken shall not

be less than 0.5 L (1 pt) liquid capacity

20 Preparation for Test

20.1 Paragraphs11.1and11.2shall not apply

20.2 Any safe method of displacement of the test sample

from the sample container that ensures filling the liquid

chamber with a chilled, unweathered sample may be employed

Paragraphs 20.3 – 20.5describe displacement by self-induced

pressure

20.3 Maintain the sample container at a temperature

suffi-ciently high to maintain superatmospheric pressure but not

substantially over 37.8 °C (100 °F)

20.4 Completely immerse the liquid chamber, with both

valves open, in the water cooling bath for a sufficient length of

time to allow it to reach the bath temperature of 0 °C to 4.5 °C

(32 °F to 40 °F)

20.5 Connect a suitable ice-cooled coil to the outlet valve of

the sample container (see Note 8)

NOTE 8—A suitable ice-cooled coil can be prepared by immersing a

spiral of approximately 8 m (25 ft) of 6.35 mm (0.25 in.) copper tubing in

a bucket of ice water.

21 Procedure

21.1 Paragraphs12.1and12.2 shall not apply

21.2 Connect the 6.35 mm (0.25 in.) valve of the chilled

liquid chamber to the ice-cooled coil With the 12.7 mm

(0.5 in.) valve of the liquid chamber closed, open the outlet

valve of the sample container and the 6.35 mm (0.25 in.) valve

of the liquid chamber Open the liquid chamber 12.7 mm

(0.5 in.) valve slightly and allow the liquid chamber to fill

slowly Allow the sample to overflow until the overflow

volume is 200 mL or more Control this operation so that no

appreciable drop in pressure occurs at the liquid chamber

6.35 mm (0.25 in.) valve In the order named, close the liquid

chamber 12.7 mm (0.5 in.) and 6.35 mm (0.25 in.) valves; and

then close all other valves in the sample system Disconnect the

liquid chamber and the cooling coil (Warning—Combustible.

Keep away from heat, sparks, and open flame Keep container closed Use only with adequate ventilation Avoid prolonged breathing of vapor or spray mist Avoid prolonged, repeated

contact with skin.) (Warning—In addition to other

precautions, provide a safe means of disposal of liquid and vapor escaping during this whole operation.)

21.2.1 To avoid rupture because of the liquid-full condition

of the liquid chamber, the liquid chamber must be quickly attached to the vapor chamber and the 12.7 mm (0.5 in.) valve opened

21.3 Immediately attach the liquid chamber to the vapor chamber and open the liquid chamber 12.7 mm (0.5 in.) valve Not more than 25 s shall pass in completing the assembly of the apparatus after filling the liquid chamber, using the follow-ing sequence of operations:

21.3.1 Remove the vapor chamber from the water bath 21.3.2 Connect the vapor chamber to the liquid chamber 21.3.3 Open the liquid chamber 12.7 mm (0.5 in.) valve 21.4 If a dead weight tester is used as a pressure measuring device (see 17.2), apply the calibration factor in kilopascals (pounds-force per square inch) established for the pressure gauge to the uncorrected vapor pressure Record this value as the calibrated gauge reading and use in Section9in place of the pressure measuring device reading

PROCEDURE D FOR AVIATION GASOLINES APPROXIMATELY 50 kPa (7 psi) REID VAPOR

PRESSURE

22 Introduction

22.1 The following sections define changes in apparatus and procedure for the determination of the vapor pressure of aviation gasoline Except as specifically stated herein, all the requirements set forth in Sections 1 – 12shall apply

23 Apparatus

23.1 Ratio of Vapor and Liquid Chambers—The ratio of the

volume of the vapor chamber to the volume of the liquid chamber shall be between the limits of 3.95 and 4.05 (see A1.1.4)

24 Sampling

24.1 Refer to Section8

25 Preparation for Test

25.1 Checking the Pressure Gauge or Pressure Transducer—The gauge shall be checked at 50 kPa (7 psi)

against a calibrated pressure measuring device (see A1.6, A1.6.1, andA1.7) before each vapor pressure measurement to ensure that it conforms to the requirements of A1.2 This preliminary check shall be made in addition to the final gauge comparison specified in12.4or 15.4

26 Procedure

26.1 Refer to Section12

Trang 7

27 Keywords

27.1 crude oils; gasoline; Reid vapor pressure;

spark-ignition engine fuel; vapor pressure; volatility

ANNEXES (Mandatory Information) A1 APPARATUS FOR VAPOR PRESSURE TEST PROCEDURE A

A1.1 Reid Vapor Pressure Apparatus, consisting of two

chambers, a vapor chamber (upper section) and a liquid

chamber (lower section), shall conform to the following

requirements:

A1.1.1 Vapor Chamber— The upper section or chamber, as

shown in Fig A1.1, shall be a cylindrical vessel having the

inside dimensions of 51 mm 6 3 mm (2 in 6 1⁄8in.) in

diameter and 254 mm 6 3 mm (10 in.61⁄8in.) in length, with

the inner surfaces of the ends slightly sloped to provide

complete drainage from either end when held in a vertical

position On one end of the vapor chamber, a suitable gauge

coupling with an internal diameter of not less than 4.7 mm

(3⁄16in.) shall be provided to receive the 6.35 mm (1⁄4in.) gauge

connection In the other end of the vapor chamber, an opening

approximately 12.7 mm (1⁄2in.) in diameter shall be provided

for coupling with the liquid chamber Care shall be taken that the connections to the openings do not prevent the chamber from draining completely

A1.1.2 Liquid Chamber—One Opening—The lower section

or liquid chamber, as shown inFig A1.1, shall be a cylindrical vessel of the same inside diameter as the vapor chamber and of such a volume that the ratio of the volume of the vapor chamber to the volume of the liquid chamber shall be between 3.8 and 4.2 (seeA1.1.3) In one end of the liquid chamber an opening of approximately 12.7 mm (1⁄2in.) in diameter shall be provided for coupling with the vapor chamber The inner surface of the coupling end shall be sloped to provide complete drainage when inverted The other end of the chamber shall be

completely closed (Warning—To maintain the correct

vol-ume ratio between the vapor chamber and the liquid chamber,

DIMENSIONS OF VAPOR PRESSURE APPARATUS

A Vapor chamber, length 254 ± 3 10 ± 1 ⁄ 8

B, C, D Vapor and gasoline chambers,

Liquid ID

51 ± 3 2 ± 1 ⁄ 8

FIG A1.1 Vapor Pressure Apparatus

Trang 8

paired chambers shall not be interchanged without

recalibra-tion to ascertain that the volume ratio is within the required

limits.)

A1.1.3 The ratio of paired vapor and liquid chambers to be

used for aviation gasoline testing shall be between 3.95 and

4.05

A1.1.4 Liquid Chamber—Two Openings—For sampling

from closed vessels, the liquid section of liquid chamber, as

shown inFig A1.1, shall be essentially the same as the liquid

chamber described inA1.1.2, except that a 6.35 mm (0.25 in.)

valve shall be attached near the bottom of the liquid chamber

and a 12.7 mm (0.5 in.) straight-through, full-opening valve

shall be introduced in the coupling between the chambers The

volume of the liquid chamber, including only the capacity

enclosed by the valves, shall fulfill the volume ratio

require-ments as set forth inA1.1.2

A1.1.5 In determining the capacities for the two-opening

liquid chamber (Fig A1.1), the capacity of the liquid chamber

shall be considered as that below the 12.7 mm (0.5 in.) valve

closure The volume above the 12.7 mm (0.5 in.) valve closure,

including the portion of the coupling permanently attached to

the liquid chamber, shall be considered as a part of the vapor

chamber capacity

A1.1.6 Method of Coupling Vapor and Liquid Chambers—

Any method of coupling the vapor and liquid chambers can be

employed, provided that no sample is lost from the liquid

chamber during the coupling operation, that no compression

effect is caused by the act of coupling, and that the assembly is

free of leaks under the conditions of the tests To avoid

displacement of sample during assembly, the male fitting of the

coupling must be on the liquid chamber To avoid compression

of air during assembly, a vent hole must be present to ensure

atmospheric pressure in the vapor chamber at the instant of

sealing (Warning—Some commercially available apparatus

do not make adequate provision for avoiding air compression

effects Before employing any apparatus, it shall be established

that the act of coupling the two chambers does not compress air

in the vapor chamber This can be accomplished by tightly

stoppering the liquid chamber and coupling the apparatus in the

normal manner, utilizing a 0 kPa to 35 kPa (0 psi to 5 psi)

gauge Any observable pressure increase on the gauge is an

indication that the apparatus does not adequately meet the

specifications of this test method If this problem is encountered, consult the manufacturer for a remedy.)

A1.1.7 Volumetric Capacity of Vapor and Liquid Chambers—To ascertain if the volume ratio of the chambers is

within the specified limits of 3.8 to5.2(seeA1.1.3), carefully measure a quantity of water greater than will be required to fill the two chambers (A dispensing buret is a convenient vessel for this operation.) Without spillage fill the liquid chamber completely The difference between the original volume and the remaining volume of the measured water quantity is the volume of the liquid chamber Without spillage couple the liquid and vapor chambers and fill the vapor chamber to the seat of the gauge connection with more of the measured water The difference between the final volume of the measured water quantity and the intermediate volume measured after ascertain-ing the liquid chamber volume is the volume of the vapor chamber

A1.2 Pressure Gauge—The pressure gauge shall be a

Bour-don type spring gauge of test gauge quality 100 mm to 150 mm (4.5 in to 6.5 in.) in diameter provided with a nominal 6.35 mm (0.25 in.) male thread connection with a passageway not less than 4.7 mm (3⁄16in.) in diameter from the Bourdon tube to the atmosphere The range and graduations of the pressure gauge shall be governed by the vapor pressure of the sample being tested, in accordance with Table A1.1 Only accurate gauges shall be continued in use When the gauge reading differs from the pressure measuring device reading, or dead-weight tester reading when testing gauges above 180 kPa (26 psi), by more than 1 % of the scale range of the gauge, the gauge shall be considered inaccurate For example, the cali-bration correction shall not be greater than 0.3 kPa (0.15 psi) for a 0 kPa to 30 kPa (0 psi to 15 psi) gauge or 0.9 kPa (0.3 psi) for a 0 kPa to 90 kPa (0 psi to 30 psi) gauge

NOTE A1.1—Gauges 90 mm (3.5 in.) in diameter can be used in the 0 kPa to 35 kPa (0 psi to 5 psi) range.

A1.3 Cooling Bath—A cooling bath shall be provided of

such dimensions that the sample containers and the liquid chambers can be completely immersed Means for maintaining the bath at a temperature between 0 °C and 1 °C (32 °F and 34

°F) must be provided Do not use solid carbon dioxide to cool

TABLE A1.1 Pressure Gauge Range and Graduations

Reid Vapor Pressure

Gauge to be Used Scale Range

Maximum Numbered Intervals

Maximum Intermediate Graduations

Trang 9

samples in storage or in the preparation of the air saturation

step Carbon dioxide is appreciably soluble in gasoline and its

use has been found to be the cause of erroneous vapor pressure

data

A1.4 Water Bath—The water bath shall be of such

dimen-sions that the vapor pressure apparatus can be immersed to at

least 25.4 mm (1 in.) above the top of the vapor chamber

Means for maintaining the bath at a constant temperature of

37.8 °C 6 0.1 °C (100 °F 6 0.2 °F) shall be provided In order

to check this temperature, the bath thermometer shall be

immersed to the 37 °C (98 °F) mark throughout the vapor

pressure determination

A1.5 Thermometer —An ASTM Reid Vapor Pressure

Ther-mometer 18C (18F) having a range from 34 °C to 42 °C (94 °F

to 108 °F) and conforming to the requirements in Specification

E1 Alternative non-mercury-containing liquid-in-glass

ther-mometers such as thermometer S18C in Specification E2251

conforming to the temperature range with equal or better

accuracy may be used

A1.6 Pressure Measuring Device—A pressure measuring

device having a range suitable for checking the pressure gauge

employed shall be used The pressure measuring device shall

have a minimum accuracy of 0.5 kPa (0.07 psi) with

incre-ments no larger than 0.5 kPa (0.07 psi)

A1.6.1 When a mercury manometer is not used as the

pressure measuring device, the calibration of the pressure

measuring device employed shall be periodically checked

(with traceability to a nationally recognized standard) to ensure that the device remains within the required accuracy specified

inA1.6

A1.7 Dead-Weight Tester—A dead weight tester may be

used as the pressure measuring device (A1.6) for checking gauge readings above 180 kPa (26 psi)

A1.8 Sample Transfer Connection—This is a device for

removing liquid from the sample container without interfering with the vapor space The device consists of two tubes inserted into a two-holed stopper of appropriate dimensions to fit the opening of the sample container One of the tubes is short for the delivery of the sample, and the other is long enough to reach the bottom corner of the sample container Fig A1.2 shows a suitable arrangement

A2 APPARATUS FOR VAPOR PRESSURE TEST PROCEDURE B

A2.1 Vapor Pressure Apparatus—Refer toA1.1.1 through

A1.1.7

A2.2 Pressure Gauge—The pressure measuring system

shall be a Bourdon type spring gauge, as described inA1.2or

a suitable pressure transducer and digital readout The pressure

measuring system shall be remotely mounted from the vapor

pressure apparatus and terminations provided for use of a quick

connection type fitting

A2.3 Cooling Bath—(Warning—To maintain the correct

volume ratio between the vapor chamber and the liquid

chamber, paired chambers shall not be interchanged without

recalibration to ascertain that the volume ratio is within the

required limits.)

A2.4 Water Bath—The water bath shall be of such

dimen-sions that the vapor pressure apparatus can be immersed in a

horizontal position Provision shall be made to rotate the

apparatus on its axis 350° in one direction and then 350° in the

opposite direction in repetitive fashion Means for maintaining

the bath at a constant temperature of 37.8 °C 6 0.1 °C (100 °F

60.2 °F) shall be provided In order to check this temperature, the bath thermometer shall be immersed to the 37 °C (98 °F) mark throughout the vapor pressure determination A suitable bath is shown in Fig A2.1and is available commercially

A2.5 Thermometers—Refer toA1.5

A2.6 Pressure Measuring Device—Refer toA1.6

A2.7 Flexible Coupler—A suitable flexible coupling shall

be provided for connection of the rotating vapor pressure apparatus to the pressure measuring device

A2.8 Vapor Chamber Tube—The vapor chamber tube of

inner diameter 3 mm (1⁄8 in.) and length of 114 mm (4.5 in.) shall be inserted into the pressure measuring end of the vapor chamber to prevent liquid from entering the vapor pressure measuring connections (seeFig A2.2)

A2.9 Sample Transfer Connection—Refer toA1.8

FIG A1.2 Sample Transfer Connection

Trang 10

SUMMARY OF CHANGES

Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue (D323 – 15) that may impact the use of this standard (Approved June 1, 2015.)

(1) Deleted former 3.1.3, definition for gasoline-oxygenate

blend.

Subcommittee D02.08 has identified the location of selected changes to this standard since the last issue (D323 – 08 (2014)) that may impact the use of this standard (Approved April 1, 2015.)

(1) Revised Referenced Documents to add Specification

E2251

(2) Revised A1.5

FIG A2.1 Apparatus for Vapor Pressure, Procedure B

FIG A2.2 Vapor Chamber Tube Inserted in Vapor Chamber

Ngày đăng: 03/04/2023, 16:00

TỪ KHÓA LIÊN QUAN

w