Designation C759 − 10 Standard Test Methods for Chemical, Mass Spectrometric, Spectrochemical, Nuclear, and Radiochemical Analysis of Nuclear Grade Plutonium Nitrate Solutions1 This standard is issued[.]
Trang 1Designation: C759−10
Standard Test Methods for
Chemical, Mass Spectrometric, Spectrochemical, Nuclear,
and Radiochemical Analysis of Nuclear-Grade Plutonium
Nitrate Solutions1
This standard is issued under the fixed designation C759; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision A number in parentheses indicates the year of last reapproval A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1 Scope
1.1 These test methods cover procedures for the chemical,
mass spectrometric, spectrochemical, nuclear, and
radiochemi-cal analysis of nuclear-grade plutonium nitrate solutions to
determine compliance with specifications
1.2 The analytical procedures appear in the following order:
Sections
Plutonium by Amperometric Titration with Iron(II) 2
Plutonium by Diode Array Spectrophotometry
Free Acid by Titration in an Oxalate Solution 8 to 15
Free Acid by Iodate Precipitation-Potentiometric Titration
Test Method
16 to 22 Uranium by Arsenazo I Spectrophotometric Test Method 23 to 33
Thorium by Thorin Spectrophotometric Test Method 34 to 42
Iron by 1,10-Phenanthroline Spectrophotometric Test Method 43 to 50
Impurities by ICP-AES
Chloride by Thiocyanate Spectrophotometric Test Method 51 to 58
Fluoride by Distillation-Spectrophotometric Test Method 59 to 66
Sulfate by Barium Sulfate Turbidimetric Test Method 67 to 74
Plutonium —238 Isotopic Abundance by Alpha Spectrometry
Americium-241 by Extraction and Gamma Counting 77 to 85
Gamma-Emitting Fission Products, Uranium, and Thorium by
Gamma-Ray Spectroscopy
94 to 102 Rare Earths by Copper Spark Spectrochemical Test Method 103 to 105
Tungsten, Niobium (Columbium), and Tantalum by Spectro
chemical Test Method
106 to 114 Sample Preparation for Spectrographic Analysis for General
Impurities
115 to 118 1.3 The values stated in SI units are to be regarded as
standard The values given in parentheses are for information
only
1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use It is the
responsibility of the user of this standard to establish
appro-priate safety and health practices and determine the
applica-bility of regulatory limitations prior to use For specific
safeguard and safety hazard statements, see Section 6
2 Referenced Documents
2.1 ASTM Standards:3 C697Test Methods for Chemical, Mass Spectrometric, and Spectrochemical Analysis of Nuclear-Grade Plutonium Dioxide Powders and Pellets
C852Guide for Design Criteria for Plutonium Gloveboxes
C1009Guide for Establishing and Maintaining a Quality Assurance Program for Analytical Laboratories Within the Nuclear Industry
C1068Guide for Qualification of Measurement Methods by
a Laboratory Within the Nuclear Industry
C1108Test Method for Plutonium by Controlled-Potential Coulometry
C1128Guide for Preparation of Working Reference Materi-als for Use in Analysis of Nuclear Fuel Cycle MateriMateri-als
C1156Guide for Establishing Calibration for a Measure-ment Method Used to Analyze Nuclear Fuel Cycle Mate-rials
Controlled-Potential Coulometry in H2SO4at a Platinum Working Electrode
C1206Test Method for Plutonium by Iron (II)/Chromium (VI) Amperometric Titration(Withdrawn 2015)4
C1210Guide for Establishing a Measurement System Qual-ity Control Program for Analytical Chemistry Laborato-ries Within the Nuclear Industry
C1235Test Method for Plutonium by Titanium(III)/ Cerium(IV) Titration(Withdrawn 2005)4
C1268Test Method for Quantitative Determination of
241
Am in Plutonium by Gamma-Ray Spectrometry
C1297Guide for Qualification of Laboratory Analysts for the Analysis of Nuclear Fuel Cycle Materials
1 These test methods are under the jurisdiction of ASTM Committee C26 on
Nuclear Fuel Cycle and are the direct responsibility of Subcommittee C26.05 on
Methods of Test.
Current edition approved June 1, 2010 Published July 2010 Originally approved
in 1973 Last previous edition approved in 2004 as C759 – 04 DOI: 10.1520/
C0759-10.
2 Discontinued as of November 15, 1992.
3 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org For Annual Book of ASTM Standardsvolume information, refer to the standard’s Document Summary page on
the ASTM website
4 The last approved version of this historical standard is referenced on www.astm.org.
Trang 2C1307Test Method for Plutonium Assay by Plutonium (III)
Diode Array Spectrophotometry
C1415Test Method for238Pu Isotopic Abundance By Alpha
Spectrometry
C1432Test Method for Determination of Impurities in
Plutonium: Acid Dissolution, Ion Exchange Matrix
Separation, and Inductively Coupled Plasma-Atomic
Emission Spectroscopic (ICP/AES) Analysis
D1193Specification for Reagent Water
E50Practices for Apparatus, Reagents, and Safety
Consid-erations for Chemical Analysis of Metals, Ores, and
Related Materials
E115Practice for Photographic Processing in Optical
Emis-sion Spectrographic Analysis(Withdrawn 2002)4
E116Practice for Photographic Photometry in
Spectro-chemical Analysis(Withdrawn 2002)4
3 Significance and Use
3.1 These test methods are designed to show whether a
given material meets the purchaser’s specifications
3.1.1 An assay is performed to determine whether the
material has the specified plutonium content
3.1.2 Determination of the isotopic content of the plutonium
in the plutonium-nitrate solution is made to establish whether
the effective fissile content is in compliance with the
purchas-er’s specifications
3.1.3 Impurity content is determined by a variety of
meth-ods to ensure that the maximum concentration limit of
speci-fied impurities is not exceeded Determination of impurities is
also required for calculation of the equivalent boron content
(EBC)
4 Committee C26 Safeguards Statement 5
4.1 The material (plutonium nitrate) to which these test
methods apply is subject to nuclear safeguards regulations
governing its possession and use The following analytical
procedures in these test methods have been designated as
technically acceptable for generating safeguards accountability
measurement data: Plutonium by Controlled-Potential
Cou-lometry; Plutonium by Amperometric Titration with Iron(II);
Plutonium by Diode Array Spectrophotometry and Isotopic
Composition by Mass Spectrometry
4.2 When used in conjunction with appropriate Certified
Reference Materials (CRMs), these procedures can
demon-strate traceability to the national measurement base However,
adherence to these procedures does not automatically
guaran-tee regulatory acceptance of the resulting safeguards
measure-ments It remains the sole responsibility of the user of these test
methods to assure that their application to safeguards has the
approval of the proper regulatory authorities
5 Reagents and Materials
5.1 Purity of Reagents—Reagent grade chemicals shall be
used in all test methods Unless otherwise indicated, it is
intended that all reagents shall conform to the specifications of
the Committee on Analytical Reagents of the American Chemi-cal Society, where such specifications are available.6 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination
5.2 Purity of Water—Unless otherwise indicated, reference
to water shall be understood to mean reagent water conforming
to SpecificationD1193
6 Safety Hazards
6.1 Since plutonium bearing materials are radioactive and toxic, adequate laboratory facilities, gloved boxes, fume hoods, etc., along with safe techniques, must be used in handling samples containing these materials A detailed discussion of all the precautions necessary is beyond the scope of these test methods; however, personnel who handle these materials should be familiar with such safe handling practices as are given in Guide C852and in Refs ( 1 ) through ( 2 ).7
6.2 Adequate laboratory facilities, such as fume hoods and controlled ventilation, along with safe techniques, must be used
in this procedure Extreme care should be exercised in using hydrofluoric and other hot, concentrated acids Use of proper gloves is recommended Refer to the laboratory’s chemical hygiene plan and other applicable guidance for handling chemical and radioactive materials and for the management of radioactive, mixed, and hazardous waste
6.3 Hydrofluoric acid is a highly corrosive acid that can severely burn skin, eyes, and mucous membranes Hydroflu-oric acid is similar to other acids in that the initial extent of a burn depends on the concentration, the temperature, and the duration of contact with the acid Hydrofluoric acid differs from other acids because the fluoride ion readily penetrates the skin, causing destruction of deep tissue layers Unlike other acids that are rapidly neutralized, hydrofluoric acid reactions with tissue may continue for days if left untreated Due to the serious consequence of hydrofluoric acid burns, prevention of exposure or injury of personnel is the primary goal Utilization
of appropriate laboratory controls (hoods) and wearing ad-equate personal protective equipment to protect from skin and eye contact is essential
7 Sampling
7.1 A sample representative of the lot shall be taken from each lot into a container or multiple containers that are of such composition that corrosion, chemical change, radiolytic de-composition products, and method of loading or sealing will not disturb the chemical or physical properties of the sample (A flame-sealed quartz vial that is suitable for accommodating pressure resulting from radiolytic decomposition is generally considered to be an acceptable sample container.)
5 Based upon Committee C26 Safeguards Matrix ( C1009 , C1068 , C1128 , C1156 ,
C1210 , C1297 ).
6Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia and National Formulary, U.S Pharmacopeial Convention, Inc (USPC), Rockville,
MD.
7 The boldface numbers in parentheses refer to the list of references at the end of these test methods.
Trang 37.2 Sample size shall be sufficient to perform the following:
7.2.1 Assay and acceptance tests at the seller’s plant,
7.2.2 Assay and acceptance tests at the purchaser’s plant,
and
7.2.3 Referee tests in the event they become necessary
7.3 All samples shall be identified clearly, including the
seller’s lot number
7.3.1 A lot is defined as any quantity of aqueous plutonium
nitrate solution that is uniform in isotopic, chemical, and
physical characteristics by virtue of having been mixed in such
a manner as to be thoroughly homogeneous
7.3.2 All containers used for a lot shall be identified
positively as containing material from a particular
homoge-neous solution
PLUTONIUM BY CONTROLLED-POTENTIAL
COULOMETRY
(This test method was discontinued in 1992 and replaced by
Test Method C1165.)
PLUTONIUM BY CONTROLLED-POTENTIAL
COULOMETRY
(With appropriate sample preparation, controlled-potential
coulometric measurement as described in Test Method
C1108 may be used for plutonium determination.)
PLUTONIUM BY AMPEROMETRIC TITRATION
WITH IRON(II)
(This test method was discontinued in 1992 and replaced by
Test MethodC1206.)
TEST METHOD FOR PLUTONIUM ASSAY BY
PLUTONIUM(III) DIODE ARRAY
SPECTROPHOTOMETRY
(With appropriate sample preparation, the measurement
de-scribed in Test MethodC1307may be used for plutonium
determination.)
FREE ACID BY TITRATION IN AN OXALATE
SOLUTION
8 Scope
8.1 This test method covers the determination of free acid in
plutonium nitrate solutions ( 3 , 4 ).
9 Summary of Test Method
9.1 Free acid is determined by titrating an aliquot of sample,
which contains an excess of ammonium oxalate added to
complex the plutonium, back to the original pH of the
ammonium oxalate solution with standard sodium hydroxide
solution Micropipets and microburets are required to measure
the small volume of sample and titrant used
10 Interferences
10.1 Any metal ions not complexed by oxalate which form
precipitates at the pH of the end point of the titration will cause
interference in this test method
N OTE 1—A “rule of thumb” is that 1 mL of saturated ammonium
oxalate solution will complex 6.4 mg of plutonium.
11 Apparatus
11.1 Magnetic Stirrer.
11.2 Microburet.
11.3 Micropipets.
11.4 pH Meter.
12 Reagents and Materials
12.1 Ammonium Oxalate Solution, saturated.
12.2 Nitric Acid (3.50 N)—Prepare solution by diluting
concentrated nitric acid (HNO3, sp gr 1.42) with water
Standardize by titrating 0.500-mL aliquots with 0.100 N NaOH
solution
12.3 Sodium Hydroxide Solution (0.100 N)—Prepare and
standardize in accordance with PracticesE50
13 Procedure
13.1 Transfer 1.0 mL of saturated ammonium oxalate solu-tion to a small vial and dilute to about 2 mL with water 13.2 Add a stirring bar and insert the electrodes and start stirrer When the pH value becomes stable, record the value as the pH of reagent
N OTE 2—Normally, the pH value for the saturated solution is approxi-mately 6.4.
13.3 Add 20 µL of sample to the vial, rinse the pipet thoroughly with water, and stir the solution for 1 min
13.4 Titrate with 0.100 N NaOH solution to within one pH
unit of the end point; then, by adding successively smaller increments, titrate to the pH of the ammonium oxalate reagent and record the volume of titrant
N OTE 3—Allow time for the pH reading to stabilize between additions
of titrant as the end point is approached.
13.5 Make a daily check of the system by adding 20 µL of
3.50 N HNO3to a sample that has already been titrated to the
end point and titrate with standard 0.100 N NaOH solution
back to the same pH
14 Calculation
14.1 Calculate the free acid (H+, N) as follows:
H 1, N 5~A 3 N!/V (1)
where:
A = microlitres of standard NaOH solution required to titrate sample,
N = normality of NaOH standard solution, and
V = volume of sample, µL
15 Precision and Bias
15.1 Precision—Of individual results, 65 % at the 95 %
confidence level
15.2 Bias—99.4 %.
FREE ACID BY IODATE PRECIPITATION-POTENTIOMETRIC TITRATION TEST METHOD
16 Scope
16.1 This test method covers the determination of free acid
in strong acid solutions of plutonium nitrate
Trang 417 Summary of Test Method
17.1 Free acid is determined by potentiometric titration with
standard sodium hydroxide solution after precipitation and
subsequent removal of plutonium (up to 50 mg) as plutonium
iodate
18 Interferences
18.1 Any hydrolyzable ions that are not precipitated with
iodate will interfere
19 Reagents and Materials
19.1 Hydrochloric Acid (sp gr 1.19)—Concentrated
hydro-chloric acid (HCl)
19.2 Nitric Acid (1 + 14)—Dilute 14 volumes of water with
1 volume of concentrated nitric acid (HNO3, sp gr 1.42)
19.3 Potassium Iodate (0.3 M)—Dissolve 64.2 g of
potas-sium iodate (KIO3) in 900 mL of water, adjust the pH to 4.3 by
adding HNO3(1 + 14), and dilute to 1 L with water
19.4 Sodium Hydroxide (0.3 M)—Prepare and standardize in
accordance with Practices E50 after making the following
alterations: Use 15 mL of the NaOH solution (50 g/50 mL), and
in step 42.2, transfer 1.200 g of National Institute for Standards
and Technology (NIST) potassium acid phthalate SRM 84 h or
its replacement to a 250-mL Erlenmeyer flask instead of 0.4000
g
20 Procedure
20.1 Pipet 50 mL of KIO3(0.3 M) into a beaker and stir
while adding an aliquot of sample solution containing no
greater than 50 mg of plutonium
20.2 After precipitation is complete, filter the solution
through either a medium porosity glass frit or a fine textured
acid-washed filter paper and collect the filtrate in a beaker
20.3 Wash the precipitate with two 25-mL portions of 0.3 M
KIO3solution, and combine the washings with the filtrate from
step20.2
20.4 Dissolve the precipitate in HNO3(sp gr 1.42) or HCl
(sp gr 1.19) and transfer to a residue bottle
20.5 Transfer the sample from 20.3 to the titration
apparatus, position the electrodes and a magnetic stirring bar in
the solution, and start the stirrer
20.6 Titrate the free acid in the solution by adding the 0.3 M
NaOH solution from a 5-mL buret and plot the titration curve
(pH versus mL NaOH solution).
20.7 Determine the end point of the titration from the
midpoint of the inflection on the titration curve and record the
volume of 0.3 M NaOH solutions by the steps given in 20.5
through20.7of the procedure
21 Calculation
21.1 Calculate the free acid (H+, N) as follows:
H 1, N 5~Vs2 Vb!N/S (2)
where:
N = normality of NaOH solution,
Vs = millilitres of NaOH solution to titrate sample aliquot,
Vb = millilitres of NaOH solution to titrate reagent blank, and
S = millilitres of sample aliquot
22 Precision and Bias
22.1 The relative standard deviation, based on 49 titrations,
is 0.9 % for aliquots of sample containing a minimum of 0.2 milliequivalents of acid
22.2 Between 99 and 100 % of the free acid in standard plutonium (IV) solutions has been measured by this procedure; however, when the plutonium was in the (III) oxidation state,
the results showed a negative bias of as much as 8 % ( 5 ).
URANIUM BY ARSENAZO I SPECTROPHOTOMETRIC TEST METHOD
23 Scope
23.1 This test method covers the determination of uranium
in the range from 300 to 3000 µg/g of plutonium in plutonium nitrate solutions
24 Summary of Test Method
24.1 Plutonium is reduced to Pu(III) in HCl (1 + 1) solution with hydroxylamine hydrochloride The uranium and pluto-nium are then separated by anion exchange, and the urapluto-nium is determined by measuring the absorbance of the U(VI)-Arsenazo I complex in a 1-cm cell at a wavelength of 600 nm
versus a reagent blank.
25 Interference
25.1 Iron at 500 µg/g of plutonium is the most likely interference in this test method
26 Apparatus
26.1 Columns, ion exchange, 1 by 10 cm Columns can be
made by sealing a 1-cm diameter filtering tube with a coarse glass frit to the bottom of a 40-mL centrifuge tube and cutting the tube off diagonally just below the frit
26.2 Spectrophotometer and 1-cm Matched Cells.
27 Reagents and Materials
27.1 Ammonium Hydroxide (1 + 13)—Dilute 1 volume of
concentrated ammonium hydroxide (NH4OH, sp gr 0.90) with
13 volumes of water
27.2 Arsenazo I Reagent (0.500 g/L)—Dissolve 250 mg of
the purified reagent [(3-2-arsonophenylazo)-4,5-dihydroxy-2,7 naphthalenedisulfonic acid, disodium salt] in water and dilute
to 500 mL with water
N OTE 4—Purify reagents as follows: To a saturated aqueous solution of Arsenazo I, add an equal volume of HCl (sp gr 1.19), filter the orange precipitate, wash with acetonitrile, and dry at 100°C for 1 h.
27.3 Hydrochloric Acid (0.1 N)—To prepare, dilute 8.3 mL
of hydrochloric acid (HCl, sp gr 1.19) to 1 L with water
27.4 Hydrochloric Acid (1 + 1)—To prepare, dilute 500 mL
of hydrochloric acid (HCl, sp gr 1.19) to 1 L with water
Trang 527.5 Hydroxylamine Hydrochloride Solution (100 g/L)—
Dissolve 10 g of (NH2OH·HCl) in water and dilute to 100 mL
with water
27.6 Nitric Acid (1 + 2)—Dilute 100 mL of nitric acid
(HNO3, sp gr 1.42) to 300 mL with water
27.7 Phenolphthalein Solution (0.25 g/L)—Dissolve 25 mg
of phenolphthalein in a water-ethanol (1 + 1) solution and
dilute to 100 mL with the water-ethanol solution
27.8 Plutonium Matrix Calibration Solution (7 g/L)—
Dissolve approximately 700 mg of plutonium metal, NIST
SRM 949e or its replacement, or other metal containing less
than 20 ppm of uranium in 5 mL of HCl (1 + 1), and dilute to
100 mL with HCl (1 + 1)
27.9 Sodium Cyanide Solution (50 g/L)—Dissolve 5 g of
sodium cyanide (NaCN) in water and dilute to 100 mL with
water
27.10 Resin, Anion Exchange—Use Dowex 1-X2 anion
exchange resin, chloride form, 100 to 200 mesh, or equivalent
resin
27.11 Stannous Chloride Solution (700 g/L)—Dissolve 7 g
of stannous chloride (SnCl2·2 H2O) in hydrochloric acid (HCl,
sp gr 1.19) and dilute to 10 mL with HCl (sp gr 1.19) Prepare
reagent fresh daily
27.12 Sulfuric Acid (1 + 2)—Dilute 1 volume of sulfuric
acid (H2SO4, sp gr 1.84) with 2 volumes of water
27.13 Sulfuric Acid (1 + 8)—Dilute 1 volume of sulfuric
acid (H2SO4, sp gr 1.84) with 8 volumes of water
27.14 Triethanolamine Buffer-Ethylenediamine-Tetraacetic
Acid Complexing Solution—Dissolve 74.5 g of triethanolamine
and 72 mg of ethylenediamine-tetraacetic acid, disodium salt
(EDTA) in 750 mL of water and 14.0 mL of nitric acid (HNO3,
sp gr 1.42) and dilute to 1 L with water Allow solution to stand
overnight before using
27.15 Uranium Standard Solution (20 mg/L)—Dissolve
23.60 mg of U3O8 (NIST SRM 950b or its replacement), or
uranium oxide of equal purity, in 1 mL of HNO3(1 + 2) and
dilute to 1 L with H2SO4(1 + 8)
28 Preparation of Ion Exchange Columns
28.1 Wash 250 g of the anion exchange resin alternately
with three 350-mL portions of HCl (sp gr 1.19) and three
350-mL portions of water Allow the resin to remain in each
solution for 30 min
28.2 Fill each column to a height of 10 cm with ion
exchange resin and rinse each column with 30 mL of HCl (sp
gr 1.19)
N OTE 5—Immediately before each analysis, rinse each column with 30
mL of HCl (sp gr 1.19) and remove any entrapped air from the column.
29 Calibration and Standardization
29.1 Pipet ten 10-mL aliquots of plutonium matrix
calibra-tion solucalibra-tion (7 g/L) into separate 50-mL beakers and add 2 mL
of H2SO4(1 + 2)
29.2 Add 0.0, 1.0, 4.0, 7.0, and 10.0 mL of uranium standard solution (20 mg/L), respectively, to each of the 5 pairs
of solutions prepared in 29.1and evaporate to dryness 29.3 Add 4.0 mL of HCl (1 + 1) to each beaker and dissolve the residue
29.4 Add 3 mL of hydroxylamine hydrochloride solution (NH2OH·HCl, 100 g/L) to each beaker and warm the solution under infrared lamps until the plutonium is reduced to Pu(III)
as indicated by the blue color If the solution is not blue, add more NH2OH·HCl solution and warm again
N OTE 6—Plutonium is not adsorbed on the resin if it is in the reduced Pu(III) state.
29.5 Cool the solutions to room temperature and add 3 drops of SnCl2·2 H2O solution (700 g/L) to each beaker
N OTE 7—The stannous chloride prevents air oxidation of the Pu(III) during subsequent steps in the procedure.
29.6 Add 13 mL of HCl (sp gr 1.19) to each beaker 29.7 Transfer each solution to a separate ion exchange column using five 1-mL portions of HCl (sp gr 1.19) to wash each beaker
29.8 Wash the Pu(III) from each column with six 5-mL portions of HCl (sp gr 1.19)
29.9 Next, elute the uranium from each column by washing
each column with six 5-mL portions of 0.1 N HCl Collect the
wash solutions from each column in a 50-mL beaker and evaporate to dryness on a hot plate under infrared lamps 29.10 Add 3 drops of HCl (sp gr 1.19) to dissolve each residue and wash the sides of the beaker with water
29.11 Add 4 drops of NaCN solutions (50 g/L) and 2 drops
of phenolphthalein solution to each beaker; then add NH4OH (1 + 13) until the indicator remains slightly pink
29.12 Pipet 5 mL of triethanolamine buffer and 3.0 mL of Arsenazo I solution to each beaker
29.13 Transfer each solution to a 25-mL volumetric flask and dilute to volume with water
29.14 Allow the solutions to stand 1 h for maximum color development, and then measure the absorbance at 600 nm in
1-cm cells versus a reference solution prepared from the
reagents starting at29.11
29.15 Calibration Curve:
29.15.1 Process the results obtained in29.14in accordance with the procedure described in31.1and31.2
29.15.2 Each time samples are analyzed verify the calibra-tion by processing duplicate aliquots of plutonium matrix calibration solutions containing no uranium; also process a set
of duplicates that contain 5 mL each of uranium standard (20 mg/L) added to aliquots of plutonium matrix calibration solution by the procedure given in 29.3through29.14 29.15.3 Process the results obtained in 29.15.2 in accor-dance with the procedure outlined in 31.3 If the individual calibration value disagrees at the 0.05 significance level with
Trang 6the value of the constant obtained from the complete
calibra-tion set, investigate and rectify the cause before proceeding
with further analyses
30 Procedure
30.1 Prepare duplicate reagent blanks starting with30.3
30.2 Transfer a sample aliquot containing approximately 70
mg of plutonium weighed to 60.1 mg into a 50-mL beaker
30.3 Add 5 mL of HCl (sp gr 1.19) to the beaker
30.4 Evaporate the solution to near dryness slowly to avoid
loss of sample
N OTE 8—This eliminates excess nitrate which would prevent reduction
of the plutonium.
30.5 Proceed with the analysis as described in29.3through
29.14
30.6 Calculate the concentration of uranium in micrograms
per gram of plutonium in accordance with instructions in
Section32
31 Calculation of Calibration Factors
31.1 Calculate the corrected absorbance value for each
standard solution as follows:
where:
Y = corrected absorbance value for standard,
r = absorbance value of standard obtained in29.14, and
s = average absorbance value obtained in 29.14 for the
duplicate calibration blanks with no uranium added
31.2 Use the least squares formulas and the data from31.1
to calculate values of A and B in the linear calibration equation:
where:
A, B = constants (B should be approximately zero),
Y = corrected absorbance value from31.1, and
x = micrograms of uranium in the standard calibration
solution
31.3 Calculate the individual calibration value for each
standard solution processed simultaneously with each set of
samples as follows:
where:
A' = individual calibration value for each standard solution,
n = micrograms of uranium in the standard solution, and
m = corrected absorbance of standard = p − q
where
p = absorbance for standard solution, and
q = average absorbance obtained from duplicate blank
solutions
31.4 Each individual value of A' should agree at the 0.05
significance level with the value of A obtained from the
complete calibration set
32 Calibration of Uranium Concentration
32.1 Calculate the uranium concentration in the sample, R,
micrograms per gram Pu, as follows:
where:
R = micrograms U per gram plutonium,
A, B = constants in linear calibration equation,
C = grams Pu per gram plutonium nitrate solution in
sample,
W = weight of sample aliquot, g, and
Y = a − b = corrected absorbance of sample solution where:
a = absorbance of sample solution, and
b = average absorbance of duplicate calibration blanks
33 Precision and Bias
33.1 In the range from 300 to 1100 µg U/g Pu the standard deviation is 6100 µg/g; in the range from 1500 to 3000 µg U/g
Pu it is 650 µg/g
THORIUM BY THORIN SPECTROPHOTOMETRIC
TEST METHOD
34 Scope
34.1 This test method covers the determination of 10 to 150
µg of thorium per gram of plutonium in plutonium nitrate solutions
35 Summary of Test Method
35.1 Lanthanum is added as a carrier and is precipitated along with thorium as insoluble fluoride, while the plutonium remains in solution and is decanted after centrifugation of the sample The thorium and lanthanum fluoride precipitate is dissolved in perchloric acid, and the absorbance of the thorium-thorin complex is measured at a wavelength of 545 nm
versus a reference solution The molar absorptivity of the
colored complex is 15 600 for thorium concentration in the range from 5 to 70 µg Th/10 mL of the solution
36 Interferences
36.1 Cations that form insoluble fluorides and colored complexes with thorin interfere in this test method
37 Apparatus
37.1 Infrared Heat Lamps, 250-W, borosilicate glass 37.2 Aluminum Heating Block—Drill a 150-mm high
alu-minum block to hold 16 12-mL centrifuge tubes and a thermometer In use the block is heated to 220°C
37.3 Platinum Stirring Rod, 1 mm in diameter by 160 mm
long
37.4 Spectrophotometer, with matched cells having 10-mm
light path
37.5 Vacuum Transfer Device, approximately 150 mm long
with a10⁄18standard-taper ground-glass joint that fits a 10-mL volumetric flask
Trang 738 Reagents and Materials
38.1 Ammonium Peroxydisulfate ((NH4)2S2O8)
38.2 Hydrochloric Acid (sp gr 1.19)—Concentrated
hydro-chloric acid (HCl)
38.3 Hydrofluoric Acid (1 + 24)—Dilute 1 volume of
con-centrated hydrofluoric acid (HF, sp gr 1.15) with 24 volumes of
water and store in a polyethylene wash bottle
38.4 Hydrogen Peroxide (30 %)—Concentrated hydrogen
peroxide (H2O2)
38.5 Hydroxylamine Hydrochloride Solution (250 g/L)—
Dissolve 25 g of hydroxylamine hydrochloride (NH2OH·HCl)
in water and dilute to 100 mL with water
38.6 Lanthanum Nitrate Solution (10 g La/L)—Dissolve
3.12 g of lanthanum nitrate (La(NO3)3·6 H2O) in water and
dilute to 100 mL with water
38.7 Nitric Acid (sp gr 1.42)—Concentrated nitric acid
(HNO3)
38.8 Perchloric Acid (70 to 72 %)—Concentrated perchloric
acid (HClO4)
38.9 Silver Nitrate Solution (2.5 g/L)—Dissolve 250 mg of
silver nitrate (AgNO3) in water and dilute to 100 mL with
water Store solution in an amber bottle
38.10 Sulfuric Acid (1 + 35)—Add 1 volume of
concen-trated sulfuric acid (H2SO4, sp gr 1.84) to 35 volumes of water
38.11 Thorin Solution (1 g/L)—Dissolve 1 g of thorin
o-(2-hydroxy-3,6-disulfo-1-naphthylazo) benzenearsonic acid
disodium salt in water and dilute to 1 L
38.12 Thorium Standard Solution (20.00 mg/L)—Dissolve
20.00 mg of high-purity thorium as the metal, oxide, or nitrate
in HCl (sp gr 1.19) and H2O2(30 %) Add 83 mL of HClO4(70
to 72 %) and dilute to 1 L with water
39 Calibration and Standardization
39.1 Reference Standards and Blanks:
39.1.1 Pipet 1.00 mL of thorium standard (20 mg/L) into
each of two 20-mL beakers, 2.00 mL into each of 2 more
beakers and 3.00 mL into each of a third pair of beakers
39.1.2 To two additional 20-mL beakers and to each of the
solutions from 39.1.1, add 1 mL of HNO3(sp gr 1.42) and 2
mL of HClO4(70 to 72 %)
39.1.3 Evaporate each solution to approximately 2 mL on a
steam bath; then continue the evaporation to dryness under
infrared lamps on a hot plate
39.1.4 Remove the beakers from the hot plate, and dissolve
each residue in approximately 2 mL of H2SO4(1 + 35),
dis-pensed from a polyethylene wash bottle
39.1.5 Transfer each solution to a 12-mL centrifuge tubing
using three 2-mL rinses of H2SO4(1 + 35)
39.1.6 Add 5 drops of La(NO3)3·6 H2O (10 g/L) solution
and 0.1 mL of AgNO3solution (2.5 g/L) and approximately 1
g of (NH4)2S2O8to each centrifuge tube
39.1.7 Heat the tubes in a steam bath for 15 min
39.1.8 Remove the tubes from the steam bath, cool, and add
1 mL of HF (sp gr 1.15) Stir the mixture with a platinum
stirring rod; rinse the rod with HF (1 + 24) after each stirring After 5 min, centrifuge the tubes for 5 min
39.1.9 Withdraw the supernatant plutonium-containing so-lution by means of vacuum and transfer to a plutonium residue bottle Invert the tubes onto a tissue for 1 to 2 min; then draw off to the residue bottle any liquid that has drained down the inner wall of the tubes
39.1.10 Wash the precipitate by adding 3 mL of HF (1 + 24) and mixing with the platinum rod Rinse the platinum rod with
HF (1 + 24), wait 5 min, and centrifuge for 5 min Repeat the procedure in 39.1.9and proceed to step39.1.11
39.1.11 Add 1 mL of HClO4(70 to 72 %) to each tube and place the tubes in the heated aluminum heating apparatus for
30 min
39.1.12 Remove the tubes, cool, and add HClO4 (70 to
72 %) to adjust the volume in each tube to 0.5 mL
39.1.13 Transfer each solution to a 10-mL volumetric flask using the vacuum transfer device with three 2-mL water rinses; then add 0.5 mL of NH2OH·HCl solution to each flask 39.1.14 Prepare a reference solution by adding 0.5 mL of HClO4(70 to 72 %), 0.5 mL of NH2 OH·HCl solution (250 g/L), and 6 mL of water to a 10-mL volumetric flask 39.1.15 Place the flasks on a steam bath for 30 min 39.1.16 Remove the flasks from the steam bath, cool, and add 1 mL of Thorin solution to each flask Dilute to volume with water, stopper, and shake
39.1.17 Measure the absorbance of the solutions in 10-mm
cells versus the reference solution at a wavelength of 545 nm.
39.1.18 Process the data obtained in39.1.17in accordance with the procedure described in41.1
39.2 Checking Calibration Curve:
39.2.1 Each time a set of samples is analyzed verify the procedure and calibration factor by processing two 2-mL thorium standards and two blank solutions (with no thorium added) in accordance with the instructions in 39.1.2 through
39.1.17 39.2.2 Process the data obtained in 39.2.1 in accordance with the procedures described in39.2 If an individual calibra-tion value disagrees at the 0.05 significance level with the value of the constant obtained from the complete calibration set, investigate and rectify the cause of the difficulty before proceeding with further analyses
40 Procedure
40.1 Transfer a weighed aliquot of sample containing from
5 to 70 µg of thorium and no greater than 500 mg of plutonium into a 20-mL beaker and proceed with the analysis as described
in39.1.2through39.1.17 40.2 Calculate the thorium concentration in accordance with the procedure described in 41.3
41 Calculation
41.1 Equation for Calibration Data:
41.1.1 Calculate the corrected absorbance value for each standard calibration solution as follows:
where:
Trang 8Y = corrected absorbance value for the standard calibration
solution,
r = absorbance value obtained in 39.1.17 for the standard
calibration solution, and
s = average absorbance value obtained in 39.1.17 for the
duplicate calibration blanks
41.1.2 Use least squares formulas and the data from41.1.1
to calculate values of A and B in the linear calibration equation:
where:
A, B = constants (B should be approximately zero),
Y = corrected absorbance from41.1.1, and
x = micrograms thorium in the standard calibration
solu-tion
41.2 Individual Calibration Values:
41.2.1 Calculate the individual calibration value for each
standard solution processed with the samples as follows:
where:
A' = individual calibration value for each standard solution,
n = micrograms of thorium in standard solution,
m = corrected absorbance value for standard solution p − q
where:
p = absorbance value from standard solution, and
q = average absorbance of duplicate blank solutions from
39.1.17
41.2.2 Each individual A' should agree at the 0.05
signifi-cance level with the value of A obtained from the complete
calibration set
41.3 Determine the thorium concentration of the sample as
follows:
Th, µg/g Pu 5 R 5~Y 2 B!/AWC (10)
where:
A and B = constants in the linear calibration equation,
W = sample mass, g,
C = grams Pu per gram of sample, and
Y = a − b = corrected absorbance of sample solution
where:
a = absorbance value for sample solution, and
b = average absorbance value from the duplicate reagent
blanks described in39.2.1
42 Precision and Bias
42.1 The relative standard deviation is less than 2 % at
thorium concentrations between 66 and 144 µg/g Pu, 4 % at a
concentration of 34 µg/g Pu, and 11 % at a concentration of 10
µg/g Pu
42.2 The average value for thorium found in 91
measure-ments of 5 to 70 µg of thorium was 99 6 1 %
IRON BY 1,10-PHENANTHROLINE SPECTROPHOTOMETRIC TEST METHOD
43 Scope
43.1 This test method covers the determination of micro-gram quantities of iron in plutonium nitrate solutions
44 Summary of Test Method
44.1 Ferric ion is reduced to ferrous ion with hydroxylamine hydrochloride Solutions of 1,10-phenanthroline and acetate buffer are added and the pH adjusted to 3.5 to 4.5 The absorbance of the red-orange complex [(C12H8N2)3Fe]+2 is read at 508 nm against a sample blank containing all of the
reagents except the 1,10-phenanthroline ( 6 ).
45 Interferences
45.1 Plutonium must be reduced to Pu(III) to avoid causing interference
45.2 Silver and bismuth form precipitates
45.3 Tolerance limits for 2 µg/mL Fe for elements that
interfere in this determination are as follows ( 7 ):
46 Apparatus
46.1 Spectrophotometer, visible range with matched 10-mm
cells
47 Reagents and Materials
47.1 Acetate Buffer Solution—Dissolve 410 g of sodium
acetate, (Na2C2H3O2) in water, add 287 mL of glacial acetic acid, and dilute to 1 L with water
47.2 Ammonium Hydroxide (1 + 9)—Dilute 1 volume of
NH4OH (sp gr 0.9) with 9 volumes of water
47.3 Hydrochloric Acid (1 + 9)—Dilute 1 volume of HCl
(sp gr 1.19) with 9 volumes of water
47.4 Hydroxylamine Hydrochloride Solution (104 g/L)—
Dissolve 104 g of hydroxylamine hydrochloride (NH2OH·HCl)
in water and dilute to 1 L with water
47.5 Iron Standard (100 µg Fe/mL)—Carefully dissolve 100
mg of high-purity iron wire in 165 mL of HCl (1 + 1), cool, and dilute to 1 L with water
47.6 1,10-Phenanthroline Solution (0.2 weight/volume %)—
Dissolve 2 g of 1,10-phenanthroline in water and dilute to 1 L with water
48 Procedure
48.1 Transfer an aliquot of sample that contains 5 to 75 µg
of iron to a 30-mL beaker and add 10 mL acetate buffer
Trang 9solution and 1 mL of hydroxylamine hydrochloride solution.
Let solution stand for 10 min
48.2 Add 1 mL of 1,10-phenanthroline solution and adjust
the pH of the solution to the range from 3.5 to 4.5 with HCl
(1 + 9) or NH4OH
48.3 Transfer the solution to a 25-mL volumetric flask Use
water to wash the beaker and to dilute to volume Stopper the
flask and mix thoroughly
48.4 After 10 min, measure the absorbance of the sample
aliquot against a sample blank that contains all of the reagents,
except 1,10-phenanthroline, at a wavelength of 508 nm
N OTE 9—For sample aliquots that contain iron in the range of 5 µg, cells
of 5-cm length or longer should be used.
48.5 Prepare a calibration curve by adding to separate
30-mL beakers, containing 10 mL of acetate buffer solution and
1 mL of hydroxylamine hydrochloride solution, the following
amounts of iron standard: 0, 50, 100, 250, and 500 µL of iron
standard solution (100 µg Fe/mL) Follow the steps given in
48.2 through48.4 of the procedure; then plot the absorbance
versus the micrograms of iron per 25 mL final volume of the
solution
49 Calculation
49.1 Calculate the iron in micrograms per gram of
pluto-nium as follows:
Fe, µg/g Pu 5 C/PW (11)
where:
C = micrograms of Fe from calibration curve,
W = weight of sample, g, and
P = Pu, g/g of sample
50 Precision and Bias
50.1 The relative standard deviation is 6 %
IMPURITIES BY ICP-AES
(Cationic impurities may be determined using Test Method
C1432(Impurities by ICP-AES) with appropriate sample
preparation and instrumentation
CHLORIDE BY THIOCYANATE
SPECTROPHOTOMETRIC TEST METHOD
51 Scope
51.1 This test method ( 8 ) is used to determine chloride in
plutonium nitrate solution
52 Summary of Test Method
52.1 After the sample aliquot is mixed with a solution
containing ferrous ammonium sulfate, sulfamic acid,
phos-phoric acid, and sulfuric acid, the chloride is steam distilled at
a temperature of 140°C (Note 10) An aliquot of the distillate
is mixed with ferric ammonium sulfate and mercuric
thiocya-nate solutions Thiocyathiocya-nate ion is released in direct proportion
to the chloride ion concentration The absorbance of the
resulting red-brown ferric thiocyanate complex is read at 460
nm against a reagent blank
N OTE 10—Save a portion of the distillate to use for the fluoride determination.
53 Interferences
53.1 Iodide, bromide, cyanide, and thiocyanate ions inter-fere Nitrite interference is eliminated by use of sulfamic acid
54 Apparatus
54.1 Steam Distillation Apparatus, including a steam
gen-erator and heating mantles
54.2 Spectrophotometer and Matched 10-mm Cells.
55 Reagents and Materials
55.1 Chloride Standard Solution (5 µg Cl/mL)—Prepare a
stock solution, A, Cl = 500 µg/mL, by dissolving 824.4 mg of dried NaCl in water and diluting to 1 L Prepare chloride standard, 5 µg Cl/mL, by diluting 10 mL of stock solution A to
1 L with water
55.2 Ferric Ammonium Sulfate Solution (0.25 M)—
Dissolve 12 g FeNH4(SO4)2·12 H2O in H2SO4 (5 + 95) and dilute to 100 mL with H2SO4(5 + 95)
55.3 Ferrous Ammonium Sulfate (0.2 M) Sulfamic Acid (0.5 M) Solution—Dissolve 78.4 g of Fe(NH4)2(SO4)2·6 H2O and 48.6 g of NH2SO3H in H2SO4(5 + 95) and dilute to 1 L with
H2SO4(5 + 95)
55.4 Mercuric Thiocyanate Solution (saturated)—Add
Hg-(SCN)2to 90 % ethyl alcohol until the solution is saturated
56 Procedure
56.1 Transfer 25 mL of acid mixture consisting of 0.2 M ferrous ammonium sulfate-0.5 M sulfamic acid solution,
phos-phoric acid, and sulfuric acid mixed in the ratio 1 + 1 + 2.5, to
a steam distillation flask and steam distill at 140°C until 100
mL of distillate is collected Retain this solution for use as a reagent blank
56.2 Transfer an accurately weighed aliquot of plutonium nitrate solution that contains approximately 500 mg of pluto-nium to a steam distillation flask and add 25 mL of acid mixture as described in56.1 Steam distill at 140°C until 100
mL of distillate is collected
56.3 Transfer up to 6 mL of sample distillate, and 6 mL of reagent blank distillate, to separate 10-mL volumetric flasks
To each flask, add 2 mL of 0.25 M ferric ammonium sulfate
solution, 2 mL of saturated mercuric thiocyanate solution, and dilute to volume with water solution and mix
56.4 After 10 min, transfer the solutions to 1-cm cells and
measure the absorbance of the sample versus the reagent blank
at a wavelength of 460 nm
56.5 Prepare a calibration curve by adding 0, 0.5, 1, 2.5, and
4 mL of the chloride standard (5 µg Cl/mL) to 10-mL volumetric flasks and dilute to about 5 mL with water solution
Add 2 mL of 0.25 M ferric ammonium sulfate solution and 2
mL of mercuric thiocyanate solution, mix, and dilute to volume with water solution Mix solutions again and after 10 min transfer the solution to 1-cm absorption cells and read the
Trang 10absorbance versus a reagent blank at a wavelength of 460 nm.
Plot the micrograms of Cl per 10 mL of solution versus the
absorbance reading
57 Calculation
57.1 Calculate the micrograms of Cl per gram of plutonium
as follows:
Cl, µg/g Pu 5 CD/WP (12)
where:
C = micrograms Cl from calibration curve,
D = dilution factor = distillate, mL/aliquot of distillate,
mL,
W = weight of sample, g, and
P = Pu, g/g of sample
58 Precision and Bias
58.1 The precision and bias of this test method is 1006 5 %
with a sodium chloride matrix No plutonium standard is
available
FLUORIDE BY
DISTILLATION-SPECTROPHOTOMETRIC TEST METHOD
59 Scope
59.1 This test method covers the determination of
micro-quantities of fluoride in plutonium nitrate solutions
60 Summary of Test Method
60.1 Fluoride is separated from the plutonium nitrate
dis-solved in a mixture of phosphoric and sulfuric acid by steam
distillation at 140 6 5°C (Note 11) The fluoride, in an aliquot
of the distillate, is complexed with Amadac F and the
absor-bance of the blue-colored complex is read in 1-cm cells versus
a reagent blank at a wavelength of 620 nm
N OTE 11—An aliquot of the distillate from the test method for the
determination of chloride (see 52.1 ) can also be used to determine
fluoride.
61 Interferences
61.1 Sulfate or phosphate ions, which may be carried over
by bumping or steam distillation at too high a temperature,
cause low absorbance reading by bleaching the colored
com-plex The formation of the colored complex is sensitive to pH
range and high salt concentration
62 Apparatus
62.1 See Section54
63 Reagents and Materials
63.1 Amadac F Solution (0.1 g/mL)—Dissolve 10 g of
Amadac F reagent in 60 % isopropyl alcohol in a 100-mL
volumetric flask and dilute to volume with 60 % isopropyl
alcohol
63.2 Fluoride Standard Solution (10 µg F/mL)—Prepare a
stock solution, 1.000 mg F/mL, by dissolving 2.210 g of dry
NaF in water and diluting to 1 L Pipet 10 mL of stock solution,
1.000 mg F/mL, into a 1-L volumetric flask and dilute to volume with water to prepare the fluoride standard, 10 µg F/mL
64 Procedure
64.1 Transfer 20-mL aliquots of the sample and of the reagent blank distillates, which were prepared during the determination of chloride, to 50-mL beakers and adjust the pH
to 5.0 to 5.2 by the addition of fluoride-free NaOH solution or HCl Dilute these solutions to 25 mL
64.2 Transfer 8-mL aliquots of the solutions prepared in
64.1to 10-mL flasks and add 2 mL of Amadac F reagent (0.1 g/L) to each solution and mix
64.3 Allow the solutions to stand 1 h; then measure the
absorbance of the blue-colored complex in the sample versus
the reagent blank solution at a wavelength of 620 nm in 1-cm cells
64.4 To prepare a calibration curve, pipet 0, 50, 100, 200,
500, and 1000-µL aliquots of the fluoride standard solution (10
µg F/mL) into separate 10-mL volumetric flasks Add 2 mL of Amadac F solution (0.1 g/L) to each flask and dilute to volume Mix, allow solutions to stand in the dark for 1 h, and measure
the absorbance of each in 1-cm cells versus a reagent blank at
a wavelength of 620 nm Plot the micrograms of fluoride in the
10-mL volume of solution versus the absorbance.
65 Calculation
65.1 Calculate the fluoride in micrograms per gram of plutonium as follows:
F, µg/g Pu 5 CD/WP (13)
where:
C = micrograms of F from calibration curve,
D = dilution factor = V1V2/A1A2 where:
V1 = volume of distillate,
A1 = aliquot from V1, mL,
V2 = volume to which A1is diluted, mL, and
A2 = aliquot of V2taken for analysis, mL,
W = weight or original sample aliquot, g, and
P = Pu, g/g of sample
66 Precision and Bias
66.1 Precision and bias of the analysis is 100 6 5 % with a sodium fluoride matrix No plutonium matrix standard is available
SULFATE BY BARIUM SULFATE TURBIDIMETRIC
TEST METHOD
67 Scope
67.1 This test procedure covers the determination of the sulfate concentration in plutonium nitrate solutions in the range from 50 to 700 µg/g of plutonium
68 Summary of Test Method
68.1 In plutonium nitrate solutions plutonium is removed by extraction with tributylphosphate, TBP The sulfate in the