These include instruments for measurement of voltage, current and power; instrument transformers; oscilloscopes and transducers.. The other flux is created by the moving coil, called the
Trang 1MARKS DISTRIBUTION FOR GATE QUESTIONS
54
TOPIC DISTRIBUTION FOR GATE QUESTIONS
2011 Wattmeter, AC bridge, Error analysis
2010 Wattmeter, Ammeter, AC bridge
2009 Dynamometer, Oscilloscope, Wattmeter
Trang 3ELECTRICAL AND ELECTRONIC MEASUREMENTS
Measurement techniques have played a significant role
from the starting from fair exchange of goods in early
civilizations to regulation of trade in industrialised
societ-ies Better measurement and instrumentation techniques
evolved as production of goods became industrialised
and advent of computers saw their enormous
applica-tion to measurement, process control and monitoring In
this chapter, we will discuss the instruments used
com-monly for electrical and electronic measurements and
about their error analysis These include instruments for
measurement of voltage, current and power; instrument
transformers; oscilloscopes and transducers
7.1 CLASSIFICATION OF MEASURING
INSTRUMENTS
Instruments can be classified based on their mode of
operation, manner of energy conversion, measuring
tech-niques and kind of output signal The main instrument
types are discussed as follows
1. Primary (or absolute) and secondary type:
Instruments that measure the absolute physical quantity directly in terms of the constants of the instrument and the deflection are called primary or absolute instruments (e.g., tangent galvanometer)
If the actual value of the quantity being measured
is proportional to some other absolute value of the quantity, the instrument is called secondary The instrument is pre-calibrated using the absolute instrument (e.g., voltmeter, ECG recorder, etc.)
2. Active and passive type: Instruments that can
be directly used for the quantity being measured are known as the active-type If the quantity being measured simply modulates the magnitude of some external power source the instrument is known as passive type
3. Deflection and null type: In a deflection-type instrument, the physical effect generated by the quantity being measured, produces an equivalent opposing effect in some other part of the instru-ment, which in turn causes deflection (or mechanical
Trang 4displacement) which is a measure of the quantity
In the null-type instrument the physical effect
gen-erated by the physical quantity under
measure-ment is nullified by either a manual or automatic
balancing device The equivalent null causing effect
is the measure of the quantity
4. Analog and digital type: In analog instrument,
the physical quantities under measurement show
continuous (step-less) variation with time In
digi-tal instruments, the physical quantities are discrete
and vary in steps with time
Based on the mode of operation, the secondary
instru-ments are further classified into three types:
1. Indicating type: In this category the
measur-ing instrument indicates the quantity bemeasur-ing
mea-sured through a pointer or some type of indicator
Majority of the measuring instruments fall under
this category, for example, voltmeter, ammeter, etc
2. Integrating type: In this category of measuring
instruments, the measurement is done with the
help of integrating device or arrangement over a
period of time For example, in the case of energy
meter the rotation of disc over a period of time
gives the reading of the energy consumed
3. Recording type: In this category, the
measur-ing instrument is used to record certain quantities
to be used for analysis For example, plot chart
recorder, ECG, EEG, etc
7.1.1 Indicating Type Instruments
The different types of torques that function in any
indi-cating measuring instruments are:
1. Deflecting torque: This torque, also called
oper-ating torque, is developed by the magnetic,
electro-static, chemical or thermal effects produced by the
quantity to be measured
T
d ∝ Operating quanity
2. Controlling torque: The instruments are so
designed that the controlling torque acts on its
moving part It can (i) control/stop the
move-ment of the pointer beyond the desired reading
and (ii) bring the pointer back to its zero position,
when the operating quantity is removed The
gr is the gravity constant
Note: In instruments with spring control ated scales are used and cramped scales are used for gravity control instruments
gradu- 3. Damping torque: In an instrument, the combined effect of deflecting and controlling torques on the movement of the pointer on the scale, causes it to oscillate when indicating the final reading These oscillations are prevented by using a damping mech-anism, either by generating air or fluid friction or by action of eddy currents This torque is proportional
to the angular velocity of the moving system and hence operated only when the system is in motion
ddtdamp = damp q
The effect of damping on deflection (q ) is depicted
in Fig (7.1), where graphs I, II and III represent under-damped, critically damped and over-damped instruments, respectively
Figure 7.1 | Effect of damping on deflection
7.2 TYPES OF INDICATING INSTRUMENTS
The basic components of all indicating instruments include:
1. Support for moving system: This can be achieved either by pivoting or suspension
2. Permanent magnets: These should have constant strength over a period of time
3. Pointers and scales: The pointers should be light
in weight with low constant of inertia A strip of mirror is mounted on the scale beneath the pointer and reading is taken after removing the parallax error between the pointer and its mirror image
4. Cases: These are the outer covering of the instrument and should be made-up of non-magnetic material
Trang 5proportional to the measured quantity, that is, age or current This electromagnetic torque is counter balanced by the mechanical torque of control springs (bronze hair springs) attached to the movable coil The coil is wound on an aluminium former which moves in the magnetic field of the permanent magnet to provide eddy current damping When the torques are balanced, the pointer attached to the moving coil will stop and its angular deflection will represent the amount of electri-cal current to be measured against a fixed reference or scale The light weight pointer is carried by the spin-dle and it moves over this graduated scale The scales
volt-of the PMMC instruments are usually linearly spaced
as the deflecting torque and hence the pointer deflections are directly proportional to the current passing through the coil The scale and pointer on the pivot are depicted
in the top view of PMMC in Fig 7.3
Scale
Pointer
Controlspring
Pivot
Balanceweight
Figure 7.3 | Top view of PMMC instrument
The electromagnetic torque is equal to the multiplication
of force with distance to the point of suspension The total deflection torque is given by
NBAk
Depending on the mode of operation of the permanent
magnet, the indicating type instruments can be classified
into the following types:
1. Moving coil-type instruments: This is further
categorised into:
(i) Permanent magnet moving coil: This can
be used for direct current and voltage measurements
(ii) Dynamometer type: This can be used either
directly or through alternating current and voltage measurements
2. Moving iron-type instruments: This can be
used for AC/DC current and voltage measurement
These instruments are discussed in detail in the
follow-ing sections
7.2.1 Permanent Magnet Moving Coil Instruments
Figure 7.2 shows the construction of a permanent
magnet moving coil instrument The instrument has a
moving coil of fine wire (circular or rectangular), with
N-turns suspended in the uniform, horizontal and radial
magnetic field of a permanent magnet in the shape of a
horse-shoe It is free to turn about its vertical axis The
coil with is placed around an iron core, which is
spheri-cal if the coil is circular, and is cylindrispheri-cal if the coil is
rectangular Since the coil is moving and the magnet is
permanent, the instrument is called permanent magnet
moving coil or a PMMC instrument
ControlspringScale
Pointer
N
S
Permanentmagnet
Rotating coil of N turnsStationary iron core
Figure 7.2 | Permanent magnet moving coil
instrument
When the current is passed through the coil it
pro-duces another magnetic field and the interaction of this
field with the magnetic field of the permanent magnet
produces an electromagnetic torque The amount of
force experienced by the coil is proportional to the
current passing through the coil which again becomes
Trang 6and move towards the coil The spindle is rigidly nected to the pointer, controlling weight, moving iron and to the piston The repulsion-type MI instrument consists of two cylindrical soft iron vanes mounted within a fixed current carrying coil One iron vane is kept fixed to the coil frame and other, attached to the pointer shaft, is free to rotate Two irons lie in the mag-netic field produced by the coil and current in the coil makes both vanes to become magnetised with the same polarity The repulsion between the similarly magne-tised vanes produces a proportional deflection of the pointer In MI type instruments, rotation is opposed
con-by a hairspring that produces the restoring torque The damping is achieved by air or fluid friction damping
For an excitation current I carried by the stationary coil, the torque produced that causes the iron disc to move inside the coil is given by
T
I dMd
d =2
The advantages of MI-type of instruments are as follows:
1. Suitable for AC and DC circuits
2. Simple construction and low cost
3. Measures voltage in the range of 0−30 V but a series resistance can be inserted in the circuit to measure higher voltages
4. Accuracy is high
5. Frictional error is less as torque/weight ratio is high
7.2.3 Electrodynamic Type Meters
Electrodynamic type meters, also known as eter type meters, can measure both DC as well as AC sig-nals up to a frequency of 2 kHz The schematic diagram
dynamom-The advantages of PMMC instruments are listed as follows:
1. Uniform scale
2. Accurate and reliable
3. High sensitivity
4. Free from hysteresis error and not affected by
external (stray) magnetic fields
5. Simple and effective damping mechanism
6. Low power consumption
7. Extension into multirange instruments possible
7.2.2 Moving Iron Type Instruments
The moving iron (MI) type instruments can measure
AC signals at frequencies up to 125 Hz, in addition to
DC signals In these instruments, the signal (current)
to be measured is allowed to flow through a
station-ary coil, which produces a magnetic field proportional
to the quantity to be measured The moving iron piece
(made of soft iron) is fixed with the moving system
(a spindle and pointer), gets attracted/repelled
pro-portionately and gives reading on the calibrated scale
These instruments are accordingly classified as
attrac-tion or repulsion type Figure 7.4(a) shows the schematic
of attraction type moving iron type instrument and that
for the repulsion type is shown in Fig 7.4(b)
Fixed coil
Cramped scale
Pointer
Spring
Moving iron disc
(a)
PointerSpringMoving iron piece (vane)(b)
Figure 7.4 | Moving-iron meter: (a) Attraction type
and (b) repulsion type
The attraction-type moving-iron meter, the moving iron
disc (vane) placed near the coil is free to get attracted
Trang 7for dynamometer type instrument is shown in Fig 7.5
The instrument has a moving circular coil which is placed
in the magnetic field produced by two circular stationary
coils which are wound separately and connected in series
Pointer
Scale
Fixed coilsMoving coil
Figure 7.5 | Schematic representation of electrodynamic
meter
The deflection torque in this type of wattmeter is
pro-duced by the interaction of two magnetic fluxes One of
the fluxes is produced by a fixed coil, called current coil,
which carries a current proportional to the load current
The other flux is created by the moving coil, called the
voltage or potential coil, which carries a current
pro-portional to the load voltage The deflecting torque is
dependent upon the mutual inductance between the two
coils and can be given by
T I I
dMd
the angular displacement between the coils The torque
is thus proportional to square of the current If the
mea-sured current is alternating, the meter is unable to follow
the alternating torque values and instead displays the
mean value of square of the current The squared or rms
value of the measured current (or any other quantity)
can be obtained by suitable modification of the scale
The advantages of electrodynamic type meters are as follows:
1. More accurate than moving-coil and moving-iron
instruments but expensive
2. Voltage, current and power can all be measured by
suitable connections of fixed and moving coils
3. Used to measure voltages in the range of 0−30 V
but can be modified by placing a series resistance
to measure higher voltages
7.2.4 Measurement of High-Frequency Signals
In the instruments discussed in the sections above, the
maximum frequency limit is of the order of 2 kHz for the
dynamometer type meters and only 100 Hz in the case of
the moving-iron type instrument The limitation of low permissible frequency can be overcome, to an extent, by rectifying the voltage signal and then applying it to a moving-coil meter, as shown in Fig 7.6
Bridge rectifier
Moving-coilmeter
Figure 7.6 | Measurement of high-frequency
voltage signals
The circuit with bridged rectifier extends the upper limit
of measurable frequency to 20 kHz but makes the surement more sensitive to change in temperature of the environment and resulting non-linear behavioursignifi-cantly impacts measurement accuracy for voltages An alternative method to overcome the low frequency limit
mea-is provided by the thermocouple meter
7.3 BRIDGES AND POTENTIOMETERS
The AC bridge networks are used for the measurement
of inductance and capacitance in the circuits These are modified form of Wheatstone bridge; consisting similarly
of four arms, an excitation source and balance detector
DC bridges along with potentiometers are used for the measurement of resistance This is depicted in the flow chart shown in Fig 7.7
Bridges
AC Bridges
Inductance measurement
Capacitance measurement
DC Bridges
Wheatstone bridge Kelvin double bridge
(Resistance measurement)
De Sauty’s bridge Schering bridge Wein’s bridge
Maxwell bridge Hay’s bridge Owen’s bridge Anderson’s bridge
Figure 7.7 | Bridges for measurement of inductance,
capacitance and resistance
Trang 8Figure 7.9 | Maxwell’s bridge (a) Circuit diagram
(b) Phasor diagram
The construction of Maxwell’s bridge shows:
1. One arm consisting of a capacitor C
1 in lel with a resistor R
paral-2 Both these variables have adjustable values
2. The opposite arm consisting of an inductor L
1 in series with a resistor R
4 Both these variables are unknown values and need to be measured
3. The other two arms consist of resistors R
1 and R
3., for which the values are known
The Maxwell bridge measures inductance after ment of C
adjust-1 and R
2 such that the current through the bridge between points A and B becomes zero, which occurs when the voltages at points A and B are equal
This is known as balancing of circuit
When the Maxwell bridge is balanced, the impedances can be written as
ZRRZ1 1 3 2
where Z
1 is the impedance of resistor R
2 in parallel with capacitor C
2, and Z
2 is the impedance of inductor L
1 in series with resistor R
4 Thus, from Eq (7.3), the relation can be mathematically represented as:
=+
The general form of an AC bridge under balance
condi-tion is shown in Fig 7.8, where all four arms are
consid-ered as impedance (frequency dependent components) If
=+
Figure 7.8 | AC bridge under balance condition
7.3.1 Measurement of Inductance
Commonly used bridges for measurement of inductance
are Maxwell’s bridge, Maxwell—Wien bridge and Hay’s
bridge; other’s include Owen’s bridge and Anderson bridge
7.3.1.1 Maxwell’s Inductance-Capacitance
Bridge
The Maxwell bridge is used to measure unknown
induc-tance in terms of calibrated resisinduc-tance and capaciinduc-tance
The circuit arrangement for Maxwell’s bridge is shown
in Fig 7.9(a) and the corresponding phasor diagram is
Trang 9When the Hay’s bridge is balanced, the impedances can
be written as
ZRRZ1 2 3 4
where, Z
4 is the impedance of the arm containing C
4 and R
2 3
1 4
1 4
2 3 1 4
1 4
1 4
1 4
2 3 4
4 4 21
=+ (w )and
R
C R R R
R C1
4 2 3 4
4 4 22
1
=+
w w
2
2. Second arm consists of an inductor L
1 in series with
a resistor R
1 These are the values to be determined
3. Third arm contains a known capacitor C
1 3 2
From the value of R
4 determined by Eq (7.5), L
1 can be determined using Eq (7.4)
7.3.1.2 Hay’s Bridge
A Hay’s bridge is another AC bridge circuit, which is a
modification of Maxwell’s bridge Figure 7.10 shows the
circuit diagram of the Hay’s Bridge It can be used for
measuring an unknown inductance by balancing its four
arms, one of which contains the unknown inductance
One of the arms of a Hay’s Bridge has a capacitor of
known value, which is the principal component that is
used to determine the unknown inductance value
The construction of Hay’s bridge is as follows:
1. One arm of the bridge consists of a capacitor C
4 in series with a resistor R
4 The resistor R
4 and C
4 are both adjustable
2. The second arm consists of an inductor L
1 in series with a resistor R
1 These are the unknown values
3. The other two arms contain known resistors R
2 and R
3
Trang 10Equating real and imaginary parts, we have
L R R C
1 = 2 3 4and
R
C RC1
4 3 2
I3
I2D
r
C
I1D
1 and R
1, and Z
4 is the impedance of the arm containing C
4 Then
=
w w
1 = 1 + w 1When the bridge is balanced,
114
3
//
Trang 11e1 = e2
e3,e4
e(b)
Figure 7.13 | De Sauty’s bridge (a) Circuit diagram
(b) Phasor diagram
It measures an unknown capacitance by comparing it with a known standard capacitance Two ratio arms of this bridge consist of non-inductive resistors (R
1 and R
2) and two consist of capacitors (C
1 and C
2) where one is of unknown value and another is standard capacitor If C
1
is the capacitor whose capacitance is to be measured and C
2 is the standard (known) capacitance, then the circuit
is balanced is by varying either R
1 or R
2 In the balanced circuit, B and D are at the same potential Then
I R I R
CIjCI
2 2
Dividing Eqs (7.13) and (7.14) we get
RRCC
RR1
2 1 2
1 2
The bridge has maximum sensitivity when C
1 = C
2 The method is simple in construction and use but perfect balance
is difficult to achieve if the capacitors show dielectric loss
7.3.2.2 Schering Bridge
The Schering bridge is used to measure unknown cal capacitance and its dissipation factor The dissipa-tion factor of a capacitor is the ratio of its resistance to its capacitive reactance The circuit for a Schering bridge
I j CR R R j CR R I R
1( w 3 + 3+ w 3 4)= 2 4 (7.12)From Eqs (7.11) and (7.12), we have
I R j L j CR R
I
R RR
j CR R RR
j CR R
1
2 3 4
2 3 4
RR
R R R R R1
2 3 4
1
3 4
7.3.2 Measurement of Capacitance
The AC bridges commonly used for measurement of
capacitance are De Sauty’s bridge and Schering bridge
These are also based on the principle of Wheatstone
bridge and have two arms; one of which has the unknown
capacitance to be determined
7.3.2.1 De Sauty’s Bridge
The De Sauty’s bridge is a modified form of the
Wheatstone bridge with the DC source replaced by an
AC source Figure 7.13 shows the circuit for De Sauty’s
(a)
Trang 127.3.2.3 Wien’s Bridge
Wien’s bridge is used for measurement of unknown capacitance and also frequency The circuit for the bridge
is shown in Fig 7.15 The construction is as follows:
1. One arm consists of a capacitor C
2 in series with a resistor R
2 These are the quantities to be determined
2. The second arm consists of a capacitor C
1 in allel to a resistor R
par-1 Both these quantities are adjustable
3. The other two arms consist of one known resistors each R
3 are present on the other two arms
3. The fourth arm has the unknown capacitor C
4 with
a resistor (R
4) connected in series (or parallel)
Under the balanced condition, we have
ZCRZ1 2 3 4
1 11
jRC
R R CC
1 4
4 1
3 2
3 4 4 2
4 3 2
=
C
R CR1 4 3
2
=Note: The balancing of a Schering bridge is independent
of frequency
Trang 13E
CA
Equating real and imaginary parts, we get
R RR
1 3
2 4
=+
⇒
−and
R RR
1 3
2 4
=+
⇒
−
ZZRR1 2 3 4 =
where Z
1 is the impedance of the arm containing C
1 and R
1 Z
2 is the impedance of the arm containing C
2 and R
2 Then,
RR
R
j C RR
jC
3 4
1
1 1 2
1
=+
2
w w
3 2 1 1 2
Thus, the balanced equation involves a factor of
fre-quency, even though individual bridge elements may be
independent of frequency
7.3.3 Measurement of Mutual Inductance
Mutual inductance can exist only in the presence of
self-inductance An important measure of mutual inductance
(M) is its ratio to the geometric mean of the two self-
L L
=
1 2Different AC bridge circuits are used for measurement of
mutual inductance These include:
7.3.3.1 Heaviside Bridge
In Heaviside bridge, the mutual inductance is measured
in terms of self-inductance The circuit (Fig 7.16)
con-sists of four non-inductive resistors R
1, R
2, R
3 and R
4
connected on the four arms of the bridge The mutual
inductor with unknown inductance is connected in series
of this bridge circuit
Trang 147.3.4.1 Kelvin Double Bridge
Kelvin bridge, also known as Kelvin double bridge or Thomson Bridge is widely used to measure an unknown electrical resistance below 1 Ω Its circuit configuration and principle of operation is similar to the Wheatstone bridge except for the presence of additional resistors
The bridge uses a second set of ratio arms and hence the name double bridge The schematic circuit configuration for Kelvin double bridge is shown in Fig 7.18 The first ratio arms is R
1 and R
2 The second set of ratio arms R
1¢ and R
2¢ is used to connect the galvanometer to a point
D at a suitable potential between points M and N such that the effect of connecting lead resistance r between the unknown resistance R
3 and the standard resistance R
4 is eliminated The ratio of resistances in the first arm (R
1/R
2) and second arm (R
1¢/R2¢) are made equal
R2
R′2G
I
Figure 7.18 | Kelvin double bridge circuit
Under balance conditions there is no current through the galvanometer which means that the voltage drop between
A and B, E
AB is equal to voltage drop E
AMD between A and C Then
The modified Heaviside-Campbell bridge is used to
mea-sure the unknown value of self inductor in terms of mutual
inductance A balancing coil L and resistance r is added to
the arm to which mutual inductor is connected in series
(Fig 7.17) A short circuit switch is connected across R
1
=+
−/
Different methods may be used for measurement of
resis-tance, depending on resistance value:
1. Low resistance: If the resistance is low (of the
order of 1 Ω or low); ammeter-voltmeter method,
Kelvin’s double bridge method and potentiometers
are used
2. Medium resistance: If the resistance is medium
(1 to 10,000); Wheatstone bridge, Carey-Foster
bridge or substitution method are used
3. High resistance: If the resistance is high (>10,000 Ω);
Megohm bridge or direct deflection, loss of change
and Megger methods are used
Trang 15is relatively large The voltage drop in the ammeter may lead to error in the reading.
In the circuit shown in Fig 7.19(b), the measured resistance is given by
RVI
V
I I
RII
+
=+
x
x
x1
If I
x>> IV, the unknown resistance is equal to the sured value of the resistance So this connection of volt-meter-ammeter is used when resistance to be measured
mea-is relatively small The current through voltmeter may lead to error in the reading
The method may be used for low and medium level resistances
An ohmmeter, using only one meter, that is voltmeter
or ammeter, is also used for measurement of resistance
Here, one of the parameters (current or voltage) is kept constant A basic series ohmmeter consists of a perma-nent magnet moving coil instrument connected in series with standard resistance
7.3.4.3 Wheatstone Bridge
Wheatstone bridge is most commonly used for ing medium level resistances Fig 7.20 shows the circuit for Wheatstone bridge It consists of three known resis-tances (R
Figure 7.20 | Circuit diagram for Wheatstone bridge
Under balanced condition
AD = AB and I R I R
1 1= 2 2Also,
DC = BC and I R I R
3 3 = 4 4
RRRR
R r
RRRR3
1 2 4
2
1 2 1 2
RRRR1 2 1 2
1 2 4
=
There are some commercial devices available with
accu-racies of 2% for resistance ranges between 0.000001 Ω
to 25 Ω
7.3.4.2 Ammeter-Voltmeter Method
This method is mainly used in the laboratories but not in
practical applications It involves use of two meters and
the accuracy is determined by the accuracy of both the
voltmeter and ammeter Their possible arrangements are
as shown in Figs 7.19 (a) and (b)
VA
VIA
IA
V
(b)
Figure 7.19 | Ammeter-voltmeter method circuits
In the circuit shown in Fig 7.19(a), the measured
resis-tance is given by
RVI
I
RVI
x>> VA, the unknown resistance is equal to the
mea-sured value of the resistance So this connection of
volt-meter-ammeter is used when resistance to be measured
Trang 16Comparing the two balanced conditions, simplifying and solving, we get
2 are the length of slide rule when slide wire
is calibrated using known resistance R
4 Let l1
′ and l2
′ be the length of slide wires at balance points, when known resistance R
resis-r and the unknown resistance R
x To sure R
mea-x, first switch S
1 is put on point 1, switch S
2 is closed and reading of ammeter is noted Then switch S
1
is moved to point 2 and the known variable resistance R
is adjusted until the ammeter gives the same deflection
as in the first case The value of the unknown resistance
is obtained directly from the known variable resistance, producing the same deflection
Figure 7.22 | Circuit for substitution method
7.3.4.6 Measurement of High Resistances
The methods that can be adopted for measuring tances of the order of 0.1 MΩ and higher are:
resis- 1. Direct deflection method
2. Loss of charge method
3. Megohm bridge
4. Megger methodThese are discussed as follows
Direct Deflection MethodFigure 7.23 (a) shows an arrangement for measurement
of high resistance in cables having metallic sheath by
RRRR
R RR1
3 2 4 4
2 3 1
7.3.4.4 Carey-Foster Bridge
The Carey-Foster bridge is a more elaborate
modifica-tion of Wheatstone bridge, particularly useful for
mea-suring or comparing two nearly equal resistances The
circuit for Carey-Foster bridge is shown in Fig 7.21
is added between resistances R
approxi-3/R
4 by sliding contact on slide wire
Let r be the resistance per unit length of the slide wire
Figure 7.21 | Carey-Foster bridge
When the bridge is balanced, let l
1 be the distance of the sliding contact from the left hand end of the slide-wire
Next the resistances R
3 and R
4 are interchanged and the bridge is balanced by moving the slide rule to distance l
2 Then:
For first balance condition
RR
R l r
R l l r1
RR
R l r
R l l r1
Trang 17= (−/ ) ⇒ = (−/ ),The insulation resistance is thus,
Figure 7.25 shows the circuit for the Megohm bridge
The circuit is completely self-contained and includes inbuilt power supplies, amplifiers, bridge members, and
an indicating instrument It has range from 0.1 MΩ to
106 MΩ
G
E
SR
RAG
RBGA
V+
−
Figure 7.25 | Megohm bridge
The accuracy is usually within 3% to possible 10% above
10000 MΩ Sensitivity of balancing at a high resistance
is obtained by usage of adjustable high voltage supplies
of 500 V to 1000 V The use of a sensitive null indicating arrangement such as a high gain amplifier with an elec-tronic voltmeter or a cathode ray oscilloscope can also be used for the purpose
Megger MethodFigure 7.26 shows Megger arrangement for measuring high resistances The current coil is the same as that in
direct deflection method The leakage current I
L is ried by guard wire wound on the insulation and therefore
car-does not flow through the galvanometer as shown
V
Guardwire
IR
IL
Metallicsheath
Conductor Insulating
materialG
(a)
V
−+
IL
IR
wireCable
(b)
Figure 7.23 | Direct deflection method for measurement
of high resistance
Figure 7.23 (b) shows the arrangement for measurement
of high resistance in cables without metal sheaths The
ends of the cable are immersed in water in a tank The
water and the tank then form the return path of the
cur-rent The insulation resistance of the cable is,
R
VI
=RLoss of Charge Method
Figure 7.24 shows the circuit for loss of charge method of
measuring high resistances In this method, the unknown
resistance is connected in parallel with capacitor and
electrostatic voltmeter The capacitor is charged initially
to a voltage V and then allowed to discharge through the
resistance R The voltmeter reading is v
Trang 18the use of these meters should not change the quantity to
be measured For this, ideally, the voltmeter should have
an infinite resistance, so that current component is not altered by the inclusion of the voltmeter and an ammeter should have zero resistance so that the load voltage is not altered by the inclusion of the ammeter in the circuit
However, for practical purposes, voltmeters have very high resistance and ammeters have very low resistance
These meters can eb used in both AC and DC circuits
7.4.1 Shunts and Multipliers
Generally moving iron instruments are used as ammeter and voltmeters The use of these meters for operating a moving coil instrument would be impractical due to bulk and weight of the coil required So, to enhance use of these meters and extend their range, shunts (in case of amme-ters) and multipliers (in case of voltmeters) are used
7.4.1.1 Extension of Range of Ammeters
The circuit for extension of an ammeter is shown in Fig 7.27
I RIsh
m m sh
=
sh = − mThus
R
I R
I Ish
m m m
=
−
II
RRm
m sh
G
Figure 7.26 | Megger method for measuring high
resistance
The voltage coil V
1 is in weak magnetic field when the pointer is at infinity and so this coil exerts lesser torque
The torque exerted by V
1 increases as it moves into a stronger field This torque becomes maximum when it is
under the pole face of the magnet and under this
condi-tion the pointer will be at zero of the resistance scale
In order to modify the torque in the voltage circuit,
2 combined function like a spring
of variable stiffness, such that It is very stiff near zero
when the current in the current coil is very small due
to the presence of unknown resistance R
x which is very large As a consequence, the low resistance portion of the
scale is compressed and high resistance part of the scale
opens up
The voltage range for measurement using a Megger
circuit can be controlled by varying the series resistance
R (to R′or R′′) connected with the current coil The test
voltages can be varied as 500, 1000 or 2500 V and can be
supplied using generator G
7.4 MEASUREMENT OF CURRENT
AND VOLTAGE
An ammeter is used to measure the current in a circuit
which is connected in series with the components
carry-ing the current A voltmeter is required to measure the
voltage across a particular element in the circuit and is
connected in parallel with the component across which
the voltage is to be measured For accurate measurement,
Trang 19The sensitivity of a voltmeter is determined in ohms per volt It is found by the division of the sum of the resistance of the meter (R
m), plus the series resistance (R
s), by the full-scale reading in volts Mathematically sensitivity is expressed as:
Power consumed in DC circuits is measured as a product
of reading of ammeter and voltmeter, that is,
P =VIHowever, in order to obtain correct power consumed by
a load, corrections must be applied for the power loss in the instrumentthat is it should include the power con-sumed by the instrument closer to the load terminal
Therefore, considering power loss in ammeter, power consumed by a load is
V VI I R
L = − 2 aAnd considering power loss in voltmeter, power con-sumed is
V VI
VRL
where cosf is the power factor of the load.
7.5.1 Measurement of Power in AC Circuits
For measurement of AC power wattmeter is used instead
of voltmeter and ammeter The most commonly used
is electrodynamic- or dynamomter-type wattmeter
Induction-type wattmeter gives the integrated measure
of power with respect with time, which is the measure of energy It is hence called induction type energy meter
7.5.1.1 Dynamometer-Type Wattmeter
Figure 7.29 shows the circuit for type wattmeter In this instrument, there are two low-resistance current coils (CC), which are fixed at their positions A high-resistance moving coil called the poten-tial or pressure coil (PC) is placed between the two fixed coils, such that it may cut the magnetic field created by the two coils The spindle, carrying spring S and pointer
electrodynamometer-II
RRm
m sh
= +1
IImm
=
The ratio of current to be measured and the full scale
deflection current is known as the instrument constant or
the multiplying factor For the same instrument with
dif-ferent shunts, the instrument constant will be difdif-ferent
resistance of the shunt is
R
Rm
RIIsh
7.4.1.2 Extension of Range Voltmeters
For extension of range of voltmeters, a series resistor or a
multiplier is required as shown in the circuit in Fig 7.28
s is multiplier tance and V is full range voltage of instrument
resis-Then
V =I R +R
m( s m)R
V I RI
VIRs
m m
s m
= + 1The circuit for multiple range of extension of voltmeter
can be achieved using more number of multiplier
resis-tances in series
7.4.2 Sensitivity of Ammeter and Voltmeter
The amount of current required by the meter coil to
produce full-scale deflection of the pointer is known as
ammeter sensitivity If the amount of current required
to produce the full-scale deflection is low, the sensitivity
of the meter is high
Trang 20Since the deflection of such instruments is proportional
to the average power they almost have uniform scales
So, this type of instruments can be used in both ac and
it Hence the angle between the current in the rent coil and current in the potential coil is less than f If the angle is f′ = −f b, then the watt-meter reading is proportional to VIcos(f - b)′cos f
cur-whereas the true power should be VIcosf So the
correction factor is
coscos( ) cos
3. Due to mutual inductance of coils: These errors resulting from mutual inductance between current and potential coils are more operative at higher frequencies As a consequence of this error, the phase angle for connection of voltage coil on the load side is increased and phase angle when current coil is connected to load is decreased
4. Due to Eddy current: Alternating magnetic field of the current coil, leads to the generation of Eddy currents The magnetic field generated by the Eddy currents modifies the phase and magni-tude of current in the current coil, thus introduc-ing errors
5. Due to connections: The diagram for connection
of a wattmeter in a circuit with small load current
is shown in Fig 7.30 Error is introduced in power measurement due to loss of power in current or voltage coils
P on it, is rigidly connected with the fixed coil The two
current coils are connected in series with each other
The sense of winding in them is such that they produce
magnetic field in the same direction (Fleming’s
right-hand rule)
Fixed coil(CC)
Moving coil(MC)
Load High
resistance (R)Supply
Figure 7.29 | Dynamometer type wattmeter
One of the fluxes is produced by a fixed coil which carries
a current proportional to the load current and, therefore,
is called the current coil The other flux is created by
a coil which carries a current proportional to the load
voltage and thus called the voltage or potential coil A
high non-inductive resistance is connected to the
poten-tial coil so that its current is almost in phase with the
load voltage The deflecting torque is produced by the
magnetic effect of electric current The control torque
is provided by control springs The damping torque is
provided by air friction damping
T
=
m m1
0sinw sin(w ±f)
P =V Irms rmscosf Watt
Trang 21There are two electromagnets in this instrument which comprise the driving system The series magnet is con-nected in series with the load and is energised by the
CC The shunt magnet takes the pressure coil which carries the current proportional to the supply voltage
The magnetic fields produced by the two magnets act upon an aluminium disc, which is free to rotate around a spindle It cuts the fluxes of both the magnets A deflect-ing torque is produced by the flux of each magnet, which tries to rotate the disc This is known as the moving system of the energy meter The magnetic effects of cur-rents through CC and PC will keep acting on a rotating disc to give a cumulative value of power consumed with respect to time (i.e energy)
A permanent magnet known as the brake (or drag) magnet is used to control the movement of the disc
It forms the braking system of the energy system It placed near the edge of aluminium disc Eddy currents are induced due to the rotation of the disc in the field
of braking magnet and the flux produced by the eddy current opposes the main flux This produces a torque proportional to the speed of disc
The spindle of the disc is connected to a counting mechanism, which records the number of revolutions of disc and indicates the energy consumed directly in kWh
The number of revolutions made by the disc for sumption of one kilowatt hour of energy is known as meter constant (K) Therefore,
con-Number of revolutions Energy in kWh,Number of revolut
=
=
KK
×iionskWh
7.5.2 Measurement of Power in a Three-Phase Circuit
Consider the star connected three-phase circuit shown
Figure 7.30 | Connections of wattmeter in a circuit
The voltage across the potential coil is equal to the sum of
voltages across the current coil and load The wattmeter
measures the power consumed by the load and power loss
in the current coil This error can be removed by use of a
compensating coil connected in series with the potential
coil and identical and coincident with the current coil
7.5.1.2 Induction Type Energy Meter
Measurement of energy means integrated measurement
of power with respect to time
E=∫ P dtHence, an energy meter would work to continuously
measure the cumulative effects of currents and
volt-ages in the circuit The energy meter is an instrument
that measures the electrical energy consumed It is also
known as watt-hour or kilowatt-hour meter
In an energy meter, a single pointer is not sufficient to
measure the total energy consumed The time dependent
measurement can be achieved by having a continuous
rotation of a disc rather than a deflection If the number of
revolution at a constant speed are proportional to the time
and the speed is proportional to the power; the energy
consumed can be obtained easily Figure 7.31 shows the
schematic diagram for induction type energy meter
Shuntmagnet
Movingaluminimdisc
Seriesmagnet
Currentcoil(CC)
Load
Figure 7.31 | Schematic of induction type energy meter
Trang 22Various methods of measurement of three phase power use different number of wattmeters:
1. One-wattmeter method: In this method, the voltage coil of the wattmeter is connected across the phase and the current coil is connected in series with the phase Thus, at a time, the wattmeter gives power in one phase only So after the measure-ment in one phase, the wattmeter is disconnected and then reconnected in each of the two remaining phases The three reading are finally added to get total power The method is thus time consuming
2. Two-wattmeter method: In this method two wattmeters are connected in an appropriate manner, such that the sum of the two wattmeter readings gives the total three-phase real power
This is discussed in detail in the next section
3. Three-wattmeter method: In this method, three wattmeters are used; one connected across each phase This gives the total three-phase power by summing up the readings of the three wattmeters
The method is faster but expensive as it requires use of three wattmeters simultaneously
7.5.2.1 Two Wattmeter Method of Power Measurement
Figure 7.34 shows the schematic circuit diagram for wattmeter method of power measurement in a three-phase star-connected system
Figure 7.34 | Schematic circuit diagram for two-wattmeter
method of power measurement in a phase star-connected system
three-The current coils of the wattmeters 1 and 2, are in series with the two phases, R and B The pressure or volt-age coils are connected between RY and BY phases for the two wattmeters The total instantaneous power con-sumed in the load circuit is given by,
W =i v +i v +i v
RN⋅ RN YN⋅ YN BN⋅ BN
Real power= L cos Watts
phVI3
Reactive power= L sin VAR
phVI3
33
×V ×I
Figure 7.33 | Three-phase delta circuit
If the per phase voltage is V
L and magnitude of per phase current is I
L/ 3 then per phase
Apparent power= VA
L LVI3
Real power= Watts
L LVI3cosf
Reactive power= VAR
L LVI3sinf
The real power in an three-phase electrical circuit is
measured using the wattmeter It consists of two coils,
namely the voltage coil (also called the potential coil or
the pressure coil) carrying large number of thin turns
and the current coil carrying less number of thick turns
The voltage coil is connected in parallel and the current
coil is connected in series to the load in which the power
is to be measured The reading given by the wattmeter
is the product of the rms values of the voltage (across
the voltage coil), rms value of the current (in the current
coil) and the cosine of the angle between them
P =VIcosf
Trang 23
The reading of the first wattmeter W
The line voltage, V
RY leads the respective phase voltage, V
RN by 30o, and the phase voltage, leads the phase rent, I
cur-RN by f Therefore, the phase difference between
V
RY and I
RN is 30o+ f which can also be seen from the
phasor diagram From Eqs (7.21 and 7.22), the sum of the two wattmeter readings is given by,
on the power factor of the load is given in Table 7.1
From Fig 7.34, the voltage across the pressure coil in
the wattmeterW
1 is
RY = RN− YNand the current through the current coil is i
RN Then the instantaneous power measured by this wattmeter W
1 is given by,
RN⋅ RY RN( RN− YN)Similarly, the instantaneous power measured by the W
Substituting the value of i
YN from Eq (7.20) in Eq (7.19),
we get
RN⋅ RN BN⋅ BN YN⋅ YN
So, it can be concluded that the sum of the two
watt-meter readings is the total power consumed in the
three-phase circuit
The phasor diagram for a three-phase balanced
star-connected circuit is shown in Fig 7.35 Here V
BY and V
RY are line voltages and V
RN is the phase voltage
Figure 7.35 | Phasor diagram for two-wattmeter
method of power measurement
Trang 24Table 7.1 | Variation of two-wattmeter readings with change in power factor of the load current
Power Factor of Load
Wattmeter Readings (W)
InferenceW
Instrument transformers are used for metering and
pro-tection in a power system Electrical measurements and
relaying decisions in a power system are made based
on the current and voltage obtained from the system
Relays work with smaller magnitudes of these signals
Real life currents and voltages thus have to be scaled
to lower levels This job is done by current and
volt-age transformers also known as instrument transformers
They also electrically isolate the relaying system from
the power equipment and working personnel
7.6.1 Current Transformer
Current transformers (CTs) are extensively used in
power system for power measuring circuits CTs are
gen-erally used in panel boards in substations or grid station
to measure the high valued bus bar currents These are
also used in combination with the relays for protection
purposes Current transformer can measure high
rents in a conductor The conductor carrying high
cur-rent passes through circular and laminated iron core of
the transformer The conductor is the single turn
pri-mary winding The secondary winding will constitute of
large number of turns of wire wound around this circular
core The secondary current (I
S) in turn is reduced to a lower value than the higher valued primary current (I
P)
as the secondary voltage is stepped up The secondary
is connected to an ammeter for current measurement A
typical current transformer is shown in Fig 7.36
Laminatediron coreConductor
carrying
IP
Secondarywindingaround core
A
Figure 7.36 | Current transformer
There may be different kinds of current transformers based on their use in metering and protection circuits
When a current transformer is used for both metering and protection purposes, it has to be of required accu-racy class to suit both accuracy of measurement and protection It has to be precise and sensitive for both small and large values of current
Circuit for measurement of current using a CT is given
in Fig 7.37
Trang 25g
a g d
Figure 7.38 | Phasor diagram for CT
7.6.2 Voltage Transformer
Voltage or Potential transformer (PT) is a type of former used to measure high voltages; they basically function as step down transformers They have smaller number of secondary turns than turns in the primary
trans-Figure 7.39 shows a voltage transformer circuit
VVoltage to be
measured
Voltmeter
Figure 7.39 | Voltage transformer
The voltage to be measured is connected across the mary circuit The low voltage secondary circuit is con-nected to a voltmeter The power rating of this type of transformer is usually lower The circuit for measure-ment of voltage using a PT is given in Fig 7.40
pri-LoadV
Primarywinding
Secondarywinding
ACSupply
Figure 7.40 | Circuit for measurement of voltage
using a PT
Load
Primarywinding
ACSupply
Secondarywinding
Three types of core construction can be employed for
CTs namely, core type, shell type and ring type The
core type construction has an advantage that sufficient
space is available for insulation purposes which makes
it more suitable for high voltage work Shell type gives
better protection for the windings Ring type is the most
common of the core constructions for CT It has very
small leakage reactance as it has no joints in the core
7.6.1.1 Transformer Burden
In CTs, the secondary has very small impedance referred
to as burden, so the CT practically operates on short
circuit conditions The burden for CT is the volt-ampere
(VA) loading which is imposed on the secondary at rated
current The burden can also be expressed as the ratio
between secondary voltage and secondary current
A metering CT has lower VA capacity than a
pro-tection CT A metering CT has to be accurate over its
complete measuring range Such a CT’s magnetising
impedance at low current and hence low flux should be
very high The magnetising impedance is not constant
for a CT’s operating range due to the non-linear
charac-teristics of the B-H curve It cannot give linear response
during large fault currents For protection CT, linear
response is expected for up to 20 times the rated current
It is also expected to give precise performance in the
normal operating currents up to high fault level currents
7.6.1.2 CT Phasor Diagram
Figure 7.38 shows the phasor diagram for the current
transformer Here, flux fm is taken as the reference The
are respectively the induced emf for the primary and
secondary windings lagging behind the flux by 90° The
magnitudes of the emf are proportional to their number
of turns in windings
Trang 267.6.3 Linear Variable Differential Transformer (LVDT)
The linear variable differential transformer (LVDT) is a passive inductive transformer which requires an external source of power It is used to measure linear displace-ment It consists of a primary winding and two second-ary windings These windings are wound over a hollow tube and the primary winding is kept between the two secondaries Figure 7.42 shows the schematic and work-ing of LVDT
Vin
Primarycoil
Primarycoil
Iron core
Secondarycoil
Secondarycoil
Advantages of LVDT: It can produce high output voltage with relatively low change in core position It is also less costly, solid and robust in construction
7.6.4 Errors in Instrument Transformers
There can be two types of errors in instrument formers These are listed as follows:
trans- 1. Ratio error: For CTs, the current transformation ratio should be constant It should also be within the given limits Practically, it can be seen some-times this ratio may vary with power factor This gives an error known as the ratio error
The ratio between actual current transformation and the normal ratio is known as ratio correction factor (RCF) Mathematically,
n
=kk
7.6.2.1 PT Phasor Diagram
Figure 7.41 shows the phasor diagram for a voltage
transformer Starting with the flux reference, E
2 is the induced emf in the secondary winding and V
2 is the minal voltage across the secondary Then
2 = 2− 2 2cosf2− 2 2sinf2The primary induced emf E
q q
q
f f
f
f0
f2
fin
Figure 7.41 | Phasor diagram for PT
7.6.2.2 Differences between a PT and a CT
The main points of difference are listed as follows:
1. The secondary of the CT is under short circuit as the
primary circuit is energised, a PT can operate with
its secondary under open circuit conditions without
any damage to the transformer or to the operator
2. Under normal operating conditions, the line
volt-age of the PT is nearly steady The flux density
and the exciting current of a PT vary between
small ranges On the other hand, the primary
cur-rent of a CT varies over wide ranges under normal
operating conditions
3. The primary current of CT is independent of
sec-ondary winding conditions, while primary current
in a PT depends on the secondary burden
4. For a PT, the primary is connected across full
volt-age In case of CT, the primary is in series with a
line and therefore a very small voltage exists across
the terminals The CT primary on the other hand
carries full line current
Trang 275. Potentiometric type: It is based on used of a calibrated potentiometer The unknown voltage
is measured by comparison with reference voltage whose value is fixed using the potentiometer
The advantages of DVM over conventional voltmeters are:
1. Higher accuracy (+0.5% or better in some cases)
2. Less human error as reading is displayed not read
3. Voltage input range from +1.000 V to +1000 V with the automatic range selection and indication for overload condition
4. Higher resolution as 1 µV reading can be measured
on 1 V range
5. Input impedance is as high as 10 MΩ
6. Small in size and hence portable
A digital multimeter on the other hand is an electronic volt ohm meter with a digital display It is capable of measuring AC and DC voltages and circuits and resis-tances over several ranges The basic circuit of a digital multimeter is generally a DC voltmeter, so any param-eter that is to be measured is first converted into voltage form The analog voltage form is then converted into digital form using analog digital converter and the digi-tal data displayed in decimal or BCD form It has supe-rior accuracy than analog instruments
7.8 CATHODE RAY OSCILLOSCOPE
The cathode-ray oscilloscope (CRO) is a common ratory instrument which can provide accurate time and amplitude measurements of voltage signals over a wide range of frequencies Its reliability, stability and ease of operation make it suitable for usage as a general purpose laboratory instrument for various signal measurements
labo-The block diagram for a CRO is shown in Fig 7.43
Horizontalinput
Horizontalamplifier
Vetricalinput
Vetricalamplifier
CRT
Sweepgenerator
Sweeptrigger
AC linesignal
Externaltrigger
Figure 7.43 | Block diagram for cathode ray
oscilloscope
The most important component of CRO is the cathode ray tube (CRT) (Fig 7.44)
2. Phase angle error: This is the angle by which
the secondary current differs in phase from the
pri-mary current when reversed This error is due to
no-load or exciting current in the transformer
7.6.4.1 Error Minimisation Methods
The magnetising and core loss component of currents
have to be kept at low values The core material should
have high value of permeability, large cross-section and
shorter magnetic path in order to minimise these
cur-rents The materials from which the core can be
con-structed for this purpose are hot rolled silicon, cold
rolled grain oriented silicon steel and nickel iron alloys
Suitable turns ratio can be provided and number of
secondary turns can be minimised by one or two turns
Large current on the secondary should be reduced by
put-ting a suitable valued shunt on either side This process also
reduces phase angle error in the instrument transformers
7.7 DIGITAL VOLTMETERS AND
MULTIMETERS
Digital voltmeters or DVM are analog to digital
convert-ers with a digital display unit These measure voltage
across two points in a circuit and display the voltage in
the form of discrete numerical instead of pointer
deflec-tion They can be used to measure both AC and DC
voltages The most important component of DVM is
analog to digital converter (ADC), which converts any
analog signal into digital For any input analog voltage,
the output is in the form of binary digital values
Digital meters achieve the required measurements
by converting the analog input into digital signal by a
sequence of digital samples spaced uniformly in time The
input signals are processed in discrete time domain with
the measured signal displayed in digital structure Thus
unlike analog instruments whose signals are processed in
continuous time domain, digital instruments have the
sig-nals processed in discrete time domain hence the name
Different types of digital voltmeters are:
1. Ramp type: It uses ramp signals as reference to
convert analog input into digital form
2. Continuous balance type: It uses a number of
test voltages in succession to calibrate the voltmeter
3. Successive approximation type: It uses a sequence
of test voltages to calibrate the voltmeter It is more
rapid in operation that continuous balance type but
more complex in construction and expensive
4. Integrating type: It is based on voltage to
fre-quency conversion and measures the actual average
of the input voltage over a fixed measuring time
Trang 28same time a voltage that increases linearly with time is applied to the horizontal deflection plates This causes the beam to be deflected horizontally at a uniform or constant rate The signal applied to the vertical plates can be thus displayed on the screen as a function of time
The horizontal axis thus serves as a uniform time scale
The linear deflection or sweep of the beam horizontally
is actually accomplished by the use of a sweep generator which is part of oscilloscope control circuitry The volt-age output of such a generator is a saw tooth wave as shown in Fig 7.45
a0
V
t
Figure 7.45 | Voltage difference V between horizontal
plates as a function of time t
Application of one cycle of this voltage to the tal plates causes the beam to deflect across the tube face also linearly with time When the voltage suddenly falls to zero at the end of each sweep (points a, b, c, d
horizon-in Fig 7.45), each of the beam flies back to its horizon-initial position The horizontal deflection of the beam is thus repeated periodically and the frequency of this periodic-ity is adjustable by external controls
The working of CRO involves the following steps:
1. The signal to be displayed is first amplified by the vertical amplifier and then applied to the vertical deflection plates of the CRT
2. A portion of the signal in the vertical amplifier is also applied to the sweep trigger as a triggering signal
Electron gun
+
−
FocusinganodeAcceleratinganode
Electron beam Vacuum
DeflectingsystemHigh
voltagesupply
6V
Figure 7.44 | Schematic of cathode-ray tube
The important components of cathode ray tube their
and functions are listed as follows:
1. Electron gun: It is the total assembly of the
fila-ment, cathode, control (intensity) grid, focus grid
and accelerating anode It generates the electron
beam and control its intensity and focus
(i) Filament gets heated up when current is passed through it and then heats up the cathode
(ii) Cathode (a negative electrode)emits electrons
when heated
(iii) Control (intensity) grid controls the number
of electrons reaching the fluorescent screen
(iv) Accelerating anode accelerates the electrons
towards the fluorescent screen
(v) Focusing gird (anode focuses the beam of
elec-trons on the screen
2. Deflecting plates: These are pair of metal plates
that are oriented in a manner that one set provided
horizontal deflection (Y-plates) and the other set
vertical deflection (X-plates) The combined effect
of these plates help control the deflection of
elec-tron beam to reach any desired point on the
fluo-rescent screen
3. Fluorescent screen: It is a glass screen coated
with fluorescent material, which converts the kinetic
energy of the electrons colliding with the screen into
heat and light So wherever the electron beam hits
the screen, the fluorescent material (phosphor) is
excited and light is emitted from that point
7.8.1 Working of CRO
To study a signal in oscilloscope, it is first amplified and
then applied to the vertical deflection plates and at the
Trang 29(ii) Sweep time/cm variable: Provides ously variable sweep rates The calibrated position is fully clockwise.
continu-(iii) Position: Controls horizontal position of trace
on screen
(iv) Horizontal variable: Controls the attenuation
or reduction of the signal applied to horizontal amplifier through external horizontal connector
4. Trigger: This section selects the timing of the beginning of the horizontal sweep and has the fol-lowing related controls
(i) Slope: Selects whether the triggering occurs
on an increasing (+) or decreasing (−) tion of trigger signal
por-(ii) Coupling: Selects whether triggering occurs at
a specific DC or AC level
(iii) Source: Selects the source of the triggering signal
(iv) Level: Selects the voltage point on the ing signal at which sweep is triggered It also allows automatic (auto) triggering
trigger-7.8.3 Measurements of Voltage
Consider the circuit shown in Fig 7.46(a) that can be used for the measurement of voltage The signal genera-tor is used to produce a 1000 Hz sine wave The AC voltmeter and the leads to the vertical input of the oscil-loscope are connected across the generator’s output By adjusting the horizontal sweep time/cm and trigger, a steady trace of the sine wave may be displayed on the screen as shown in Fig 7.46(b) The trace represents a plot of voltage vs time The vertical deflection of the trace about the line of symmetry (DC) is proportional
to the magnitude of the voltage at any instant of time
Signalgenerator
Voltmeter
VerticalinputAC
(a)
Vm
Vp-pDC
(b)
Figure 7.46 | Measurement of voltage
3. The sweep trigger then generates a pulse which is
coincident with a selected point in the cycle of the
triggering signal This pulse turns on the sweep
generator and thereby initiating the saw-tooth
wave form
4. The saw-tooth wave is then amplified by the
hori-zontal amplifier and applied to the horihori-zontal
deflection plates
5. Some additional provisions for external signals are
usually made for applying an external triggering
signal or utilising the 60 Hz line for triggering
Also, in some cases, the sweep generator may not
be used and an external signal applied directly to
the horizontal amplifier
7.8.2 CRO Controls
The controls present in most oscilloscopes provide a wide
range of operating conditions and thus make the
instru-ment suitable for measuring a wide variety of signals A
brief description of controls that are common to most
oscilloscopes (as depicted in Fig 7.44) are as follows:
1. Cathode-ray tube (CRT): This section
per-forms the following control functions:
(i) Power and scale illumination: Turns the
instrument on and controls illumination of the screen
(ii) Focus: Adjusts the focus the spot or trace on
the screen
(iii) Intensity: Regulates the brightness of the spot
or trace
2. Vertical amplifier section: This section
com-prises of the following control functions:
(i) Position: Controls vertical positioning of
oscil-loscope display
(ii) Sensitivity: Selects the sensitivity of the
verti-cal amplifier in verti-calibrated steps
(iii) Variable sensitivity: Provides a continuous
range of sensitivities between the calibrated steps
(iv) AC-DC-GND: Selects the desired coupling
( AC or DC) for incoming signal applied to the vertical amplifier or grounds (GND)the amplifier input If DC coupling is selected, the input is directly connected to the amplifier
When AC coupling is selected, the signal is first passed through a capacitor to block out any constant or DC component, before enter-ing the amplifier
3. Horizontal sweep section: This section
com-prises of the following control functions:
(i) Sweep time/cm: Selects the desired sweep
rate from calibrated steps or admits external signal to horizontal amplifier
Trang 307.8.4.1 Lissajous Figures
When the inputs to the horizontal and vertical ers are sine-wave signals of different frequencies, a sta-tionary pattern is formed on the CRT When the ratio
amplifi-of the two input frequencies is an integral fraction such
as 12234315, , , , etc., these stationary patterns are known
as Lissajous figures and can be used for comparison and measurement of frequencies (Fig 7.47) Using two oscil-lators of different frequencies, some simple Lissajous figures can be generated, like those shown in Fig 7.48
Figure 7.47 | Lissajous figures for horizontal-to-vertical
frequency ratios of (a) 1:1, (b) 2:1, (c) 1:2 and (d) 3:1
Phase difference x-y
Frequency ratio y/x
180°
Figure 7.48 | Lisajous patterns
To determine the size of the voltage signal appearing at
the output of terminals of the signal generator, an AC
voltmeter is connected in parallel across these terminals
The AC voltmeter is designed to read the effective DC
value of the voltage This effective value is also known as
the rms value of the voltage The peak or maximum
volt-age is V
m volts and is represented by the distance from the symmetry line CD to the maximum deflection The
magnitude of the peak voltage displayed on the
oscil-loscope is related to the effective or rms voltage (V
rms) displayed on AC voltmeter as
V
rms = 0.707 V
m (for a sine or cosine wave)
VV
m = rms
0 707.For a symmetric sinusoidal wave, the value of peak volt-
age V
m can be taken as 1/2 the peak to peak of the
volt-age signal (V
p-p)
Note: The mathematical realtion for rms signals is valid
only for sinusoidal signals
7.8.4 Measurement of Frequency
With the application of horizontal sweep voltage, the
voltage measurements can be obtained from the
ver-tical deflection and the signal can be displayed as a
function of time If the time base (or the, sweep) is
calibrated, such that measurements of pulse duration
or signal period can be made, then frequency of the
signals can then be determined as reciprocal of the
time period
To measure the frequency of a signal, the following
steps are required:
1. Set the oscillator to 1000 Hz and display the signal
on the CRO
2. Then set the horizontal gain so that only one
com-plete wave form is displayed
3. Measure the period of the oscillations using the
horizontal distance between two points such as C
to D as shown in Fig 7.46(b)
4. Then reset the horizontal gain until five waves are
seen on the screen Keep the time base control in a
calibrated position
5. Measure the distance (and hence time scale) for
five complete cycles and calculate the frequency
from this measurement
f
T(Hz)= seconds
1
6. Repeat the measurements for other frequencies
(e.g 150 Hz, 5 kHz and 50 kHz) as set on the
signal generator
Trang 31The quality factor (Q) of a coil is the ratio of reactance
to resistance in a frequency dependent circuit tion The Q factor or quality factor of an inductance is commonly expressed as the ratio of its series reactance to its series resistance For inductor in series or parallel, the ratio of reactance to resistance in a frequency dependent circuit configuration is given by
configura-QXRLR
= s =s s s
w
QRXRL
p p p p
w
The value of Q varies from 5 to 1000
The Q factor of a capacitance is the ratio of its series reactance to its series resistance, although for capaci-tors, generally dissipation factor (D) is used which is the reciprocal of Q For capacitor in parallel or series, the ratio of reactance to resistance in a frequency dependent circuit configuration is given by
DX
C R
= s =s
A signal generator can be used to measure the frequency
of an unknown sinusoidal signal It is connected to the
vertical amplifier (or horizontal) and the calibrated
signal source of frequency is fed to the horizontal
ampli-fier (or vertical) The frequency of the signal generator is
adjusted so that a steady Lissajous pattern is obtained
If f
v and f
h are the frequencies of the signals applied
to vertical and horizontal amplifiers, respectively, then
these are related to the number of tangencies (points
at the edge of arcs) along the vertical and horizontal
iizontaltangencies crossings( )
If f
h is known, the unknown frequency f
v can be lated using the above relation
calcu-Note: It is difficult to maintain the Lissajous figures in a
fixed configuration because the two oscillators are not in
phase and frequency locked Their frequencies and phase
drift slowly causes the two different signals to change
slightly with respect to each other
7.8.5 Measurement of Phase Difference
When both applied waveforms are sinusoidal, the
result-ing Lissajous pattern may take many forms dependresult-ing
upon the frequency ratio and phase difference between
the waveforms If straight line is obtained, the phase
angle difference will either be zero or 180° However if an
ellipse is obtained, the phase difference between the two
signals can be determined from the Lissajous figure The
formation of ellipse is illustrated in Fig 7.49
Let the two sinusoidal voltage signals be given by
x y
=
1 2sinsin
Since deflection is directly proportional to the
ampli-tude of voltage, we have from the figure
sinf = =
YYXX1 2 1 2Note: The ellipse forms can be used to determine only
the phase angle between two sinusoidal voltages It
does not indicate which one is leading and which one is
lagging
Trang 32An active transducer is one which does not require any power source for operation The input of physical quantity generates a proportional electric signal A pas-sive transducer requires external power source for opera-tion The output signal represents variation of electrical parameters (R, C, etc.) and needs to be converted into equivalent current or voltage signal.
7.10.1 Strain Gauge
A strain gauge is an instrument used to measure strain produced on a wire by a force generated by varying the electrical resistance of the wire It is an example of a passive transducer It can effectively measure strain, displacement, weight, pressure or mechanical force A bonded strain gauge is made of fine wire looped from side to side on a mounting plate which is attached to the element which is experiencing the stress Consider the equation of resistance,
RLA
= r
where r is the specific resistance of conductor wire L is
the length of the conductor (in m) while A is the cross sectional area (in m2) When under strain, the length will increase and the area will decrease As a result, the value
of resistance will increase The gauge factor is given as
kRRLL
Vxtxf
=∆ n =∆where, n is the number of detection elements which passes the detector in t seconds f is the frequency of output signal
A Q-meter is used to measure the Q-factor of a coil and
related electrical properties It is based on the principle that
Q-factor of a resonant circuit is equal to its voltage
magni-fication factor and can be expressed as the ratio of voltage
developed across its reactive elements to the voltage injected
in series with the circuit to produce the developed voltage
The basic circuit for a Q-meter is shown in Fig 7.50
The circuit contains terminals for connecting the
induc-tance (L
x) to be measured and this is brought in resonance
by a variable tuning capacitor (C) There are terminals
available for adding capacitance (C
x), if required
When an unknown coil is connected to the test
termi-nals, the circuit is excited by a tunable signal source
This is achieved by setting the oscillator to a given
fre-quency and varying the internal capacitor (C) or adding
a capacitor C
x of the desired value and adjusting the frequency of the oscillator This leads to development of
voltage across a resistor in series with the tuned circuit
The resistance of the resistor should be very small
(frac-tional in ohms) so that it can be neglected in comparison
to the loss resistance of the components to be measured
The AC injection voltage (V
in) across the series resistor and the AC output voltage (V
out) across the terminals of the tuning capacitor are measured using voltmeters The
circuit for output measurement must be a high input
impedance circuit so that loading of the tuned circuit by
the metering circuit is prevented
The Q-factor is measured by adjusting the source
fre-quency and/or the tuning capacitor for a peak output
voltage corresponding to resonance Q-factor is
deter-mined as the ratio of output voltage measured across the
tuned circuit to voltage injected into it When L
x and C
x
are at resonance:
QVV
= outinThe voltage across the variable capacitor is measured
with electronic voltmeter V
2, in which the scale is brated to read Q directly
cali-7.10 TRANSDUCERS
An electrical transducer is a device which can convert a
non-electrical physical quantity into an electrical quantity,
like current or voltage and generates an electrical signal
Unknown coil
Lowresistance
Tunablesignalsource
Highimpedanceinterface(R)
Trang 33Figure 7.51 shows a typical configuration of thermocouple.
Wire A
Wire B
ColdHot
Figure 7.51 | A typical thermocouple
The temperature ranges of thermocouple vary with the metals used for making them Temperature ranges of some common thermocouples are listed below:
Chromel-Alumel −270 to 1370o
CChromel-Constantan −270 to 790o
CIron-Constantan −210 to 1050o
CCopper-Constantan −270 to 400o
CNicros-Nisil −260 to 1300o
Pitot tube is extensively used for velocity measurement
in aircraft It works on the principle that if a blunt object
is placed in the flow channel, the velocity of fluid before
it will be zero; considering the fluid to be incompressible
Thus from Bernoulli’s equation,
p vg
p vg
A typical Pitot tube is shown in Fig 7.52
This is usually performed optically, mechanically,
inductively or capacitively
7.10.2.1 Tachometer
Tachometers are used to measure speeds or rotational
movements, revolutions per minute, or sometimes they
can be used to measure rate of flow A scale factor can be
applied to produce readings of the desired type They may
be the analog or digital types with contact or non-contact
types Tachometers can be energised by either AC or DC
7.10.3 Temperature Sensing Devices
7.10.3.1 Resistance Temperature Detector
(RTD)
RTDs are temperature detectors that are based on the
change in electrical resistance of some materials with
change in temperature The resistance changes linearly
with temperature, within a limited temperature range
The relation is given as,
made of platinum as it has high chemical stability and
highly reproducible electrical properties Although they
are accurate and stable, they have higher initial cost
They are known to be less rugged in vibration locations
7.10.3.2 Thermistors
Thermistors are also temperature sensing devices They
can be negative temperature coefficient type used for
tem-perature sensing or positive temtem-perature coefficient type
used for temperature control They have a typical
tem-perature range of −70 to 300°C (~ −100°F to 600°F)
Thermistors are widely accepted as the most
advanta-geous and less costly sensors for a number of applications
for temperature measurement and control
7.10.3.3 Thermocouples
Thermocouple is a temperature sensing device which
works on the principle of Seebeck effect This effect
causes a voltage generation in a circuit containing two
different metals with their junctions kept at different
temperatures These devices are rugged but sensitive
and widely as they are inexpensive and can be used over
a wide temperature range
Trang 347.10.4.3 Rotameter
Rotameter works as a constant pressure drop and able area meter for measuring fluid flow It is easy to install and simple in construction but has lower accuracy and can only be installed in vertical pipelines
vari-7.10.4.4 Venturimeter
Venturimeter can measure horizontal flow rate and has two pressure taps It has high mechanical strength but has higher cost Figure 7.54 shows a venturimeter
on the diaphragm will cause a change in distance between the diaphragm and the static plate causing a change in capacitance This change in capacitance is measured using a bridge circuit or tank circuit
Dielectric
Static plateDiaphragm
Pressure
Insulatingmaterial
Figure 7.55 | Capacitive transducer
7.10.6 Piezo-Electric Transducer
Piezo-electric transducer is used to measure pressure as
a function of voltage generated Force and acceleration can also be measured using this transducer As the name suggests, it contains a piezo-electric crystal in which an
Figure 7.52 | Pitot tube
7.10.4.2 Orifice Meter
Orifice meter is the most common type of instruments
used to measure the fluid flow Here an orifice plate is
used in a pipeline Figures 7.53(a) and (b) show an
ori-fice meter and the structure of oriori-fice plate respectively
The volumetric flow rate Q in an orifice meter can be
obtained at different diameters d
1 and d
2 of the orifice plate as,
1, A is cross-sectional area g and g are
acceleration due to gravity and specific weight of fluid
Trang 357.10.8 Bellows
Bellows are made of soft material and are pressed to form convolutions One end is fixed where air can go inside using a port while the other end is free to move (Fig 7.58)
A spring is used often to counter the bellow movement
The pressure to be measured can be obtained as,
pkxA
Figure 7.58 | Bellows
7.11 ERROR ANALYSIS
Error analysis is an essential part of measurement Error
in measurement can be defined in as,Error = Instrument reading — true reading
Percentage error
=Instrument reading true reading
True readi
−nng
×100
7.11.1 Types of Errors
There can be three types of errors, namely gross error, systematic error and random error and these are explained as follows
1. Gross error: It arises due to human mistakes ing to wrong reading of values, mistake in record-ing measured data, parallax error, etc
lead- 2. Systematic error: Zero error and instrument bias are source of systematic error There may be con-structional error in instruments They may be due
to error in construction of the instrument scale, pointer, etc This error can lead to error in every reading taken while measuring
3. Random error: It occurs when the exact cause
of error cannot be determined These can never be corrected but can be reduced to a certain extent by averaging or finding the error limits The random errors have a deviation of the readings and it
electric charge is produced on the surface on application
of force Voltage produced will be proportional to the
applied force Usually a quartz crystal is a piezoelectric
crystal An instrument amplifier is used on its output for
measurement purpose Figure 7.56 shows a piezoelectric
transducer
Outputvoltage
Pressure port
Quartzcrystal
Forcesummingmember
Base
Figure 7.56 | Piezo-electric transducer
7.10.7 Bourdon Tube
Bourdon tube is used for pressure measurement It
con-sists of a C-shaped hollow tube whose one end is fixed
and connected to a pressure tapping while the other end
is kept free (Fig 7.57) When a pressure is applied, the
circular tube tries to straighten, which makes the free
end to move A deflecting mechanism attached to an
indicating instrument is attached to the free end that
moves a pointer The materials commonly used for this
tube are brass, phosphor bronze, and beryllium-copper
They can measure very high differential pressures of
700 MPa
p
Free end ofthe tubeDeflecting
mechanism
Pointer
Figure 7.57 | Bourdon tube
Trang 36follows a particular distribution which is generally
normal distribution
7.11.2 Mean Value and Deviation
Mean value is the most probable value for a set of
read-ings, which can be given as,
xnxi
n
=
i11
=
∑
where, n is the total number of readings and x
i is the individual readings value The deviation d of readings
from this value can be derived as,
i =
i−
7.11.3 Precision and Calibration
The precision of measurement is sometimes defined as
this deviation from the mean value The mean square
deviation or variance can now be used to measure the deviation from a set of readings as,
Vn
x xn
i1
1
2 1
2
− i=∑( − ) s
where, s is the standard deviation.
Calibration is the process of comparing the measured value obtained from a particular instru-ment to that of a standard instrument Comparison of actual input values with the output indication of the system will result determination of systematic errors
Errors at these calibrating points are then reduced
by adjusting the components or by using calibration charts
The true values can also be obtained from standard look up tables prepared for this purpose Calibration can
be re-performed from time to time to reduce the effects
of instrument wear over time
IMPORTANT FORMULAS
1 Types of torque in indicating instruments
(a) Deflecting torque
T
d ∝ Operating quanity (b) Controlling torque
2 Permanent magnet moving coil (PMMC) instrument:
Deflecting torque T = NbAI
3 Moving iron type instrument:
Deflecting torque, T
I dMd
=2
2 q
Deflecting angle, q
q
=IkdMd22
4 Electrodynamic instrument:
Deflecting torque, T I I
dMd
4 are the impedance
of AC bridge arms, then at balance point:
I Z I ZI
=+
1 3 2
Trang 37(b) Heaviside-Campbell bridge
L
R R2
1
=+
−/
9 Measurement of resistance (a) Kelvin double bridge
RRRR3
1 2 4
= (b) Ammeter-voltmeter method
RVI
I
RVI
RVI
V
I I
RII
+
=+
x V x1
(c) Wheatstone bridge
R
R RR4
2 3 1
=R (f) Loss of charge method
V ve
Vve
=ln
10 Measurement of current and voltage (a) Extension of ammeter range
II
RRmm
m sh
1
4 2 3 4
4 4
121
=+
=+
and
w w w
=
w w
1= 1+ w 1
C RC
4 3 2
(d) Anderson bridge
R
R RR
RR
R R R R R1
2 3 4
1
3 4
1 2
4
4 41
=+ w
j C
R1
1 11
w
C
R CR1 4 3
2
= (c) Wein’s bridge
RRRRCC4
3 2 1 1 2
f
R R C C
=12
1
1 2 1 2
8 Measurement of mutual inductance
(a) Heaviside bridge
Trang 38Rm
RIIsh
(b) Extension of voltmeter range
V
I R
RR
m m
s m
11 Measurement of power and energy in AC circuits
(a) Dynamometer type wattmeter
P =V Irms rmscosf
(b) Induction type energy meter
Number of revolutions = K×Energy in kWhwhere K is meter constant
(c) In three-phase (star) circuits total power
consumed by the balanced load is equal to (W
1+W
2)
12 Power factor angle f =
+tan−
13 Cathode ray oscilloscope (CRO)
(a) Voltage measurement
VV
m = rms
0 707. (b) Frequency measurement
ffh v
=Number of vertical tangencies crossingsNumber of hor
iizontaltangencies(crossings) (c) Measurement of phase difference
sinf = =
YYXX1 2 1 2
14 Q-Meter
QVV
= outin
15 Transducers (a) Strain gauge
Vxtxf
=∆ =
∆n
(c) Resistance temperature detector
=Instrument reading true reading
True readi
−nng
×100
(a) Mean x
nxn
=
i
i =11
∑
(b) Deviation d x x
i = i− (c) Variance (mean square deviation)
Vn
x xi
n
i1
1
2 1
2
− ∑= ( − ) s
Trang 39SOLVED EXAMPLES
Types of Indicating Instruments
1 The coil of a PMMC instrument has dimensions
15 mm × 12 mm The flux density is given as
4 A 10 A electrodynamic ammeter is controlled by
a spring having a constant of 0.1 × 10−6 Nm/
degree The full scale deflection is 110° Determine
the inductance of the instrument when measuring
a current of 10 A Given that the mutual tance at 0° deflection is 2 µH and the change in the mutual inductance is linear with deflection
induc-(a) 4.42 µH (b) 7.2 µH(c) 2.21 mH (d) 2.21 µHSolution: We have
110 0 1 10100
m
Thus total inductance = 2 + 0.21 µH = 2.21 µH
Ans (d)
Bridges and Potentiometers
5 A Maxwell’s capacitance bridge is used to sure an unknown inductance in comparison with a known capacitance as shown in the figure below
(a) 240 Ω (b) 24 Ω(c) 2.4 kΩ (d) 240 kΩ
Trang 40V
Thus, R
1 =10kΩ20kΩ=6 67 kΩ
R
2 =10kΩ20kΩ=7 14 kΩThus,
100 48 3 51 7100
9 For the given circuit, if the voltmeter reads 60 V, the value of the unknown resistance R will be
100 VDC+
−
R
100kΩ V
(a) 2.2 kΩ (b) 50 kΩ(c) 32.2 Ω (d) 100 ΩSolution: Voltmeter reading = 60 VThus,
100
50 100150
10
403
+
=RR
Solving we get R = 2.2 kΩ
Ans (a)
10 A moving coil instrument of resistance 5 Ω requires
a potential difference of 75 mV to give a full scale deflection The value of shunt resistance needed to give a full scale deflection at 30 A is
2 3 4
600 4001000
240
WAns (a)
7 An Owen’s bridge is used to measure the impedance
of a steel specimen at 2 kHz Given that
R
2 = 834 Ω, C2 = 0.124 µF,R