The applicability of ultrahigh-performance supercritical fluid chromatography coupled with mass spectrometry (UHPSFC/MS) for the qualitative analysis of metabolites with a wide polarity range (log P: −3.89–18.95) was evaluated using a representative set of 78 standards belonging to nucleosides, biogenic amines, carbohydrates, amino acids, and lipids.
Trang 1journalhomepage:www.elsevier.com/locate/chroma
Department of Analytical Chemistry, Faculty of Chemical Technology, University of Pardubice, Studentská 573, Pardubice 53210, Czech Republic
a r t i c l e i n f o
Article history:
Received 9 December 2021
Revised 13 January 2022
Accepted 13 January 2022
Available online 15 January 2022
Keywords:
Supercritical fluid chromatography
Metabolites
Amino acids
Human plasma
Mass spectrometry
a b s t r a c t
Theapplicability ofultrahigh-performancesupercriticalfluidchromatographycoupledwithmass spec-trometry (UHPSFC/MS) for the qualitative analysis of metabolites with a wide polarityrange (log P:
−3.89–18.95)wasevaluatedusingarepresentativesetof78standardsbelongingtonucleosides,biogenic amines,carbohydrates,aminoacids,andlipids.Theeffectsofthegradientshapeandthepercentageof water(1,2,and5%)wereinvestigatedontheViridisBEHcolumn.Thescreeningofeightstationaryphases wasperformedforcolumnswithdifferentinteractionsites,suchashydrogenbonding,hydrophobic,π
-π,oranionicexchangetypeinteractions.Thehighestnumberofcompounds(67)ofthesetstudiedwas detectedontheTorusDiolcolumn,whichprovidedaresolutionparameterof39.TheDEAcolumnhad thesecondbestperformancewith58detectedstandardsandtheresolutionparameterof54.The over-allperformanceofotherparameters,suchasselectivity,peakheight,peakarea,retentiontimestability, asymmetryfactor,andmassaccuracy,ledtotheselectionoftheDiolcolumnforthefinalmethod.The comparisonofadditivesshowedthatammoniumacetategaveasuperiorsensitivityoverammonium for-mate.Moreover,theinfluenceoftheionsource ontheionizationefficiency wasstudiedbyemploying atmosphericpressurechemicalionization(APCI)andelectrosprayionization(ESI).Theresultsprovedthe complementarityofbothionizationtechniques,butalsothesuperiorionizationcapacityoftheESIsource
inthenegativeionmode,forwhich53%oftheanalytesweredetectedcomparedtoonly7%fortheAPCI source.Finally,optimized analyticalconditionswereappliedtotheanalysisofapooledhumanplasma sample.44compoundsfromthepreselectedsetweredetectedinhumanplasmausingESI-UHPSFC/MS
inMSEmodeconsideringbothionizationmodes
© 2022TheAuthors.PublishedbyElsevierB.V ThisisanopenaccessarticleundertheCCBY-NC-NDlicense (http://creativecommons.org/licenses/by-nc-nd/4.0/)
1 Introduction
Metabolomicscanbe describedasthecomprehensivestudyof
smallmolecules,calledmetabolites,intheorganismandthe
asso-ciationofthose withpathophysiologicalstates [1,2].Metabolomic
analysisrequirestheuseofhighlypowerfulanalytical techniques,
such asmass spectrometry (MS) hyphenated to chromatographic
separation techniques, such as liquid chromatography (LC) or
gas chromatography, asa means to simultaneously analyze
com-plex mixtures of metabolites [3] The most widespread
separa-tion modes used for metabolomicsare reversed-phase
ultrahigh-performance liquid chromatography (RP-UHPLC) and hydrophilic
interactionliquidchromatography(HILIC-UHPLC)coupledto
high-∗ Corresponding author
E-mail address: denise.wolrab@upce.cz (D Wolrab)
resolution (HR) MS instruments, such as quadrupole-time-of-flight (QTOF) or Orbitrap [4] The comprehensive analysis of the metabolomeischallengingduetothehighchemicalandstructural diversityofmetabolites InRP-UHPLC,intermolecularhydrophobic interactionsbetweenanalytes,stationaryphase,andmobile phase allow analysis ofa large part of the metabolome of complex bi-ological samples such as urine, plasma, and tissue extracts [5– 7].However, polarand/or ionicspeciesarepoorly retainedin RP-UHPLC[8].TheretentionmechanismintheHILICmodeisbasedon the interactionof polaranalyteswith thepolar stationaryphase, whichallowstheseparationoftheanalytes Therefore,HILIC pro-videscomplementarychromatographicseparationcomparedto RP-UHPLC/MS [9,10] Nonpolar compounds may elute in or close to the voidvolume inHILIC mode.A comprehensivetechniquethat allowsthe separationofpolarandnon-polarmetabolites, suchas lipids,aminoacids,andnucleosides,isdesired
https://doi.org/10.1016/j.chroma.2022.462832
0021-9673/© 2022 The Authors Published by Elsevier B.V This is an open access article under the CC BY-NC-ND license ( http://creativecommons.org/licenses/by-nc-nd/4.0/ )
Trang 2instruments allow stableandreproducibleresults aswell as
rou-tine hyphenationto massspectrometry[11,12].Mainly, the
back-pressureregulator,injector,andcolumntechnologywereimproved
for thenew generationinstruments, leadingto better acceptance
ofUHPSFC/MS forabroadapplicationrange.Generally,theuseof
supercritical fluid chromatography(SFC) wasfirstdescribedmore
than50yearsago[13],butitsapplicationforlipidomic[14,15]and
metabolomic analysis [12,16,17] represents a rather recent trend
UHPSFC may represent a potential alternative to RP- and
HILIC-UHPLC forthe comprehensiveanalysisof themetabolome by
re-ducing the costs and analysis time Nowadays, UHPSFC mainly
usessupercriticalCO2mixedwithorganicmodifiersasthemobile
phase.The additionoforganicsolvents,typically 2–40%,broadens
the range of analytes that can be separated withUHPSFC Polar
solvents, such as methanol,ethanol, or acetonitrile,are the most
commonlyusedandfacilitatetheelutionofpolarcompounds.The
additionofsmallpercentagesofwater,salts,bases,and/oracid
ad-ditivestothemodifiercanfurtherimprovethepeakshapeandthe
elution of polar andionic compounds [18–20] The low viscosity
and highdiffusionof themobile phase in UHPSFCallow the use
of highflow rateswithout losingseparationefficiencyand
there-fore allow high-throughput analyzes [21,22] Furthermore,almost
all stationary phases used for UHPLC can also be used for
UH-PSFC including modern stationary phasespacked with sub-3 μm
core shell andfully porous particles [23–26] Recently, dedicated
UHPSFC stationary phases were also introduced to the market,
such as the Torus column series from Waters These stationary
phasesare basedon silicamodifiedwithdifferentselectors, such
as propanediol,1-aminoanthracene, diethylamine, 2-picolylamine,
or ethylene-bridged silica, which are potentially suitable for the
separationofmedium topolarmetabolites[18,21,27].Itshouldbe
mentionedthattheanalysisofverypolarcompoundsstillremains
challenging for UHPSFC/MS employingcommon chromatographic
conditions The increase inthe percentageof modifierin CO2 up
to100%duringthegradientincreasestheelutionstrengthfor
po-lar compounds and extends the polarity range of analytes
suit-able for theseparation withUHPSFC/MS instruments These
spe-cialmodesusingincreasedmodifier andCO2 asthemobilephase
arecalledunifiedchromatographyorenhancedfluid
chromatogra-phy[20,28].However,suchmobilephaseconditionsandtheuseof
sub-2 μmparticlecolumnscanlead toexceedingsystempressure
forcing adjustmentofparameters, suchasbackpressure,
tempera-ture,orflowrate.Consequently,theanalyzesmaybeperformedby
increasingtheorganicsolventinthemobilephasegradientand,at
thesametime, decreasingtheflowrates.Followingthisapproach
in metabolomics and using a polar stationary phase, the elution
rangesfromnonpolartopolarcompounds[12]
Inrecentyears,moreUHPSFC/MSapplicationshavebeen
inves-tigated, accompanied by unconventionaland innovative
develop-ments regardingtheapplied conditionsandinstrumentalsettings
Analysis ofnaturalproducts, biological samples[29–32],
pharma-ceuticals, nutrients, and environmental samplesare examples for
the application areasof UHPSFC[33].However, onlya few
stud-iesinvestigatedmetabolomicsbyUHPSFC/MS.Thepotentialof
UH-PSFC/MSfortheanalysisofpolarurinarymetaboliteswas
investi-gatedbySenetal.,whoevaluated12differentcolumns,3column
temperatures, and9 differentadditives inmethanol, forthe
sep-aration of60 polar metabolites (log P −7 to 2) [18] Desfontaine
et al described the application of UHPSFC coupled to a
triple-quadrupole MSfortheanalysisofnucleosides,smallbases,lipids,
small organic acids, andsulfated metabolites [12] The analytical
methodwasoptimizedbyinvestigatingseveralparameterssuchas
the kinetic performance, the percentageofcosolvent, the type of
stationary phase, and the composition of the mobile phase
Ad-ditionally, the mixture of 57 compounds was also analyzed by
unifiedchromatographycoupledwithMS Losacco etal.analyzed
49metabolitesinplasmaandurine usingUHPSFC/QTOF-MSwith theevaluationoftheimpact ofthebiologicalmatrix.Mostof se-lectedcompounds werenotaffectedbymatrixinterference(63%), whereby 16% ofcompounds showeda matrix effectinurine and plasmasamples[16].TheUHPSFC/MS analysisoffreeaminoacids wasinvestigatedbyRaimbaultetal.[20].Theseparationof18 na-tive proteinogenic amino acids wasachieved by applying unified chromatographicconditions,startingfrom90%CO2 to100% modi-fier
The aim of this work was to evaluate the suitability of UH-PSFC/MSfortheanalysisof78metabolitesselectedfromthe Hu-manMetabolome Database (HMDB)database based on their rel-evance incancer research Toachieve the elution ofall analytes, enhanced fluid chromatography wasapplied because the analyte set covers a wide polarity range (log P: −3.89 – 18.95) The in-fluence of the percentage of waterin the modifier, the gradient shape,andthe type ofstationaryphase fortheseparation ofthe analyteset wasevaluatedusingUHPSFC/QTOF-MS Theionization efficiencyoftheselected metabolitesemployingelectrospray ion-ization(ESI)andatmosphericpressure chemical ionization(APCI) wascompared.TheMSEmode wasappliedfortheanalysisofthe standardmetabolitemixtureandplasmasamples
2 Materials and methods
2.1 Chemicals
Methanol (CH3OH), acetonitrile (ACN), 2-isopropanol, hexane (all LC/MS gradient grade), water (H2O; LC/MS Ultra, UHPLC/MS grade),andchloroform (LCgrade,stabilizedwith0.5–1% ethanol) were purchased from Honeywell (Charlotte, North Caroline, US) Ammonium acetate, ammonium formate (LC/MS, gradient grade), andformicacid(98–100%,Suprapur)werepurchasedfrom Sigma-AldrichorMerck(St.Louis,MO,U.S.A),respectively.Carbondioxide (CO2)4.5grade(99.995%)wasobtainedfromMesserGroupGmbH (BadSoden,Germany)
2.2 Standards
The standards were purchased from Sigma-Aldrich (see Table1).Standardstocksolutionswerepreparedbydissolvingeach compoundintheappropriatesolventorsolventmixture(TableS1)
toobtainthefinalconcentration of10mgmL−1.Afterwards,nine standardmixtureswerepreparedaccordingtotheanalytecategory anddilutedinMeOH,thatis,mixturesofnucleosides(N°1;100ng
μL−1), biogenic amines (N°2; 20–100 ng μL−1), sugars together withotherorganiccompounds(N°3;100–10ngμL−1),aminoacids (N°4–8;2–100 ngμL−1) andlipids (N°9;10ngμL−1) Analytesin eachmixtureare reportedinTableS1.Standardconcentrationsin the final mixture were established by investigatingthe efficiency and sensitivity of ionization using direct infusion MS for 10 ng
μL−1 and100ngμL−1 aswellastwodifferentsamplesolvent so-lutions,MeOH andMeOH/ACN(50:50,v/v) Theoptimized source parameters,standardconcentration,andsamplesolventwereused forfurtherexperiments.Thesestandardmixtures(100μLofeach) andguanine (10 μL) were mixed, andthe final standard solution wasdilutedwithacetonitriletoobtainafinalsolventcomposition
ofCH3OH/ACN(50:50,v/v)forUHPSFC/MSanalyzes(TableS1)
2.3 Stationary phases
The final standard solution was analyzed using eight differ-ent columns,reported inTable 2 The stationaryphasesdiffer in their column chemistry, providing different interaction sites and
Trang 3Table 1
List of standard compounds
Nucleosides
Uridine
C 9 H 12 N 2 O 6 244.0695 −2.28
Biogenic amines
N-acetyl-5-hydroxytryptamine C 12 H 14 N 2 O 2 218.1055 0.44
Amino acids
L-Histidine monohydrochloride C 6 H 9 N 3 O 2 155.0695 −1.67
Acetyl- l -carnitine hydrochloride C 9 H 17 NO 4 203.1157 −0.66 D,L-Octanoylcarnitine chloride C 15 H 29 NO 4 287.2096 1.68 D,L-Decanoylcarnitine chloride C 17 H 33 NO 4 315.2409 2.46
D,L-Myristoylcarnitine chloride C 21 H 41 NO 4 371.3035 4.02
D,L-Palmitoylcarnitine chloride
Sugars
Other organic compounds
Polar lipids
( continued on next page )
Trang 4Table 1 ( continued )
Mono-Sulfo-GalCer(d18:1/24:1) C 48 H 91 NO 11 S 889.6313 13.94
Nonpolar lipids
Table 2
Columns screened in this study
ACQUITY UPC 2 Torus Diol Fully porous hybrid silica Propanediol 100 × 3.0 1.7
ACQUITY UPC 2 Torus 2-PIC Fully porous hybrid silica 2-Picolyl-amine 100 × 3.0 1.7
ACQUITY UPC 2 Torus 1AA Fully porous hybrid silica 1-Amino-anthracene 100 × 3.0 1.7
ACQUITY UPC 2 Torus DEA Fully porous hybrid silica Diethylamine 100 × 3.0 1.7
ACQUITY UPC 2 HSS C18 SB Fully porous silica Octadecyl, nonendcapped 100 × 3.0 1.8
therefore may show different selectivities The Viridis BEH
col-umn(100× 3.0mmI.D,1.7μm)wasselectedforthepreliminary
study In addition, the Torus columns, namely, Diol, 2-PIC, 1-AA,
andDEA(100× 3.0mmI.D,1.7μm),wereevaluated.Thedioland
BEH columns representthe most polar stationary phases among
the selected ones, characterized by propandiol bonded silica
support andfree silanols, respectively.Furthermore, twocolumns
packedwithmodifiedC18silicawereincludedforcolumn
screen-ing,namely,HSSC18SBcolumn(100× 3.0mm,1.8μm)andHSS
T3(100× 2.1I.D;1.8μm).Additionally,thechiralstationaryphase
Trefoil CEL1(150 × 3.0mm I.D;2.5μm) wastested Allcolumns
werepurchasedfromWaters(Milford,MA,USA)
2.4 UHPSFC/MS/MS instrumentation
UHPSFC/MS/MS analysis was performed on the Acquity UPC2
(Waters, Milford, MA, USA) hyphenated with the Synapt G2-Si
(Waters) QTOF mass spectrometer The UHPSFC instrument was
coupled to the MS using the commercial interface kit (Waters)
Gradient mode was used for screening the stationary phase
se-lected forthe separation of the metabolite mixture Supercritical
carbondioxide(scCO2)wasusedasthemobilephaseA,andMeOH
with30 mmolL−1 ammonium acetate and1,2,or 5% ofH2Oor
MeOHwith30mMammoniumformateand2%ofH2Owas
inves-tigatedasmobilephaseB(modifier).Thegradientstartedwith5%
ofB,wasincreasedto70%Bin8.5min,thento100%Bin10.5min,
kept constant for2.5min,andfinally returnedto theinitial
con-dition within 1 min and re-equilibrated for 1 min, with a total
runtimeof15min.Furthermore,aflowgradientwasemployedto
avoidinstrumentoverpressureat100%ofthemodifier;thestarting
flow wasset to2.0mLmin−1,decreasedto0.8mLmin−1 within
14 min,and backto the initial flowin 1min The ABPRwas set
at 1800 psi and the column temperatureat 60 °C The injection
volumewas1μL,andtheinjectionneedlewaswashedaftereach
injectionwithhexane/2-isopropanol/H2O(2:2:1,v/v/v).MeOHwith
0.1% of formic acid and5% of H2Owas used as a make-up
sol-vent to improve the ionization Furthermore, a flow gradient for
themake-up solventwasused: 0min– 0.6mLmin−1,8.5min–
0.2mL min−1,13 min – 0.2mL min−1, 14min – 0.6 mLmin−1,
15min– 0.6mLmin−1 Thefollowing settingswere usedforQTOFmeasurements:HR mode,amassrangeofm/z50–950,andthecontinuummodewith
ascantimeof0.1 wereapplied.Leucineenkephalinwasusedas thelock mass,inwhich thelock masswasacquired withascan timeof0.1 in10 intervals,butnoautomaticlockmass correc-tionwasapplied.ESIandAPCIinpositiveandnegativeionmodes wereinvestigated.ThefollowingparameterswereusedfortheESI mode:capillaryvoltage2.50kV,samplingcone20V,sourceoffset
90V,sourcetemperature150°C,desolvation temperature350°C, conegasflow50L/h,desolvationgasflow1000L/handnebulizer gasflow3.5bar.ThefollowingparameterswereusedfortheAPCI mode: corona current1.0μA, samplingcone 10V, conegas flow
50L/h,nebulizergasflow3.5bar,sourceoffset50V,source tem-perature 150°C, probe temperature600 °C andlockspray capil-lary voltage 3.0kV Column screeningwas performedin positive andnegativeionmodeusingESIandfullscanspectraacquisition Furthermore,theMSEmodewasappliedtodetecttheMSspectra andthecorrespondingfragmentspectraofeachcompoundinone run The MSE methodis characterized by two stages In stage 1, allions aretransmittedfromtheionsourcethroughthe collision cell,whereinlowcollisionenergyisappliedsothatno fragmenta-tioncan beobservedinthemass analyzer,andions arerecorded
astheprecursorions.Instage2,allionsaretransmittedfromthe ionsourcethrough thecollision cell,andacollision energyramp
is applied togenerate andrecord fragment ionsin the mass an-alyzer Hence,thesoftware isabletogeneratetwo spectraatthe sametime; thefirst one showsthe precursorions withno colli-sion energy,and the second one generates fragment ionsdue to theappliedcollisionenergy.Thetrapandtransfercollisionenergy
ofthe lowenergyfunction waskept off andtheramp trap colli-sionenergyofthehighenergyfunctionwassetfrom5to30V.– Theobtainedfragmentswerecomparedwithonlinedatabases,i.e., HMDBandMzCloudforfurtherconfirmation
Trang 52.5 Biological sample preparation
Humanplasmacollectedfromdifferenthealthyvolunteerswas
pooled, worked up, and analyzed with UHPSFC/MS under
opti-mized conditions All subjects signed an informed consent An
additional step, namely, limiteddigestion with proteinase K was
added.Beforeproteinprecipitation,2μLofProteinaseKand2μL
of250mM CaCl2 were addedto100 μLofplasmasampleto
ob-tain a final concentration of 5 mM and sonicated for15 min at
40°C.Forproteinprecipitation,1mLofCH3OH/EtOH(1:1,v/v)was
addedtothepooledplasmasamplesonicatedfor15minatroom
temperature(25°C)andstoredfor1hat−20°C.Thesamplewas
centrifuged for15minat10,000rpm,thesupernatantwas
trans-ferred to a glassvialandevaporated undernitrogen The residue
wasdissolvedwith35μLofACN/CH3OH/H2O(4:4:2,v/v/v)+0.1%
formicacidanddiluted1:10withthesamesolventmixture.1μL
wasinjectedforthesubsequentUHPSFC/MSanalysis
2.6 Data processing
Data were acquiredwiththeMassLynxsoftware(Waters).The
WatersCompressionToolwasusedfornoisereduction,whichalso
minimized the raw data file size facilitating data handling The
Accurate MassMeasure tool in MassLynxwasused to apply lock
masscorrectionforbettermassaccuracyandfortheconversionof
datafromcontinuumtocentroidmode,whichfurtherreducedthe
file size Finally, TargetLynx was used to extract retention times,
peak areas,peak height, peak widths(Pkwidth), andasymmetry
factors by providing exact masses and expected retention times
ofall compounds.The resultingtableswere exportedas.csv files
and further processed with Microsoft Excel, i.e., to calculate the
numberofidentifiedstandardsandtherelativestandarddeviation
(RSD%)oftheretentiontime,peakareaandpeakheight
MarkerL-ynxwasusedtogeneratefeaturelistsofm/zwiththe
correspond-ing retentiontime,whichallowed thecalculationofthemass
ac-curacy The complete summary tablesfromMarkerLynx were
ex-portedas.csvfilesandmanually checkedforexperimentalm/z of
eachstandardortargetanalyte
Furthermore,MZmine2.53software[34] softwarewasusedto
assess the influence of the data processing procedure The
fol-lowing settings were applied: the targeted peak detection
mod-ule was used to search the list of compounds with a precursor
mass tolerance of 0.002 m/z or5 ppm and a retention time
tol-eranceof0.1min.Peakintegrationwascheckedandmanually
cor-rected when needed Retention times, peak areas, peak heights,
peak widths (FWHM), asymmetry factors, and experimental m/z
were exported as.csv filesand furtherprocessed with Microsoft
Excel For thechromatographic evaluation, theresolution and
se-lectivity (α)were calculated.Theresultsoftwodifferentsoftware
solutionswerecompared(Tables3andS11)
MSDIAL ver 4.70, was used for metabolite identification in
realhumanplasmausingthedatabasesMSMS-Public-pos-VS15for
positive and MSMS-Public-Neg-VS15 for negative ion mode The
databasesarecomposedofmetabolitesinplasma,whichwere
de-tected bytheMSDIALcommunity,whereby13,303entriesare
in-cluded for positiveion mode and12,879entries for negativeion
mode
3 Results and discussion
3.1 Selection of standard compounds
Theselectedanalytes(Tables1andS1)wereselectedfromthe
Human MetabolomeDatabase based ontheir biological relevance
with a special focus on the metabolites involved in cancer
pro-gression [12,16] The final metabolite mixture varied in
molecu-Fig 1 Relation of partition coefficients (log 10 P) and molecular weights for the in- vestigated analytes and the number of metabolites categorized by compound class: nucleosides (red), biogenic amine (yellow), amino acids (green), sugars (orange), others (dark green), polar lipids (light blue), and nonpolar lipids (blue)
lar weights from 89 to 900 Da and log10P values from −3.89
to 18.95 (Fig 1) The substantial variety in the structural com-position andchemical characteristics ofselected compounds, i.e., polar amino acids to hydrophobic lipids such as triacylglycerols, leadstohighlydiverseretentiontime behavioranddifferent ion-ization efficiencies, altering sensitivity The list includes metabo-litesinvolved invariousbiological pathways,e.g.,metabolites de-rived from the tryptophan pathway Tryptophan is an essential aminoacid,abuildingblockforproteinbiosynthesisandfunctions
asaprecursor forthe conversiontoseveralother metabolites in-cludedinourlist,i.e.,5-hydroxytryptophan,tryptamine,serotonin, melatonin,N-acetyl-5-hydroxytryptamine, l-kynurenine,l-alanine, andglutamic acid Furthermore,clinical studies have shownthat tryptophanmetabolismpromotestumorprogressionthrough mul-tiple mechanisms [35], andits metabolic derivative l-kynurenine
is involved in Alzheimer’s disease and the early stages of Hunt-ington’s disease The catecholamines dopamine, adrenaline, and noradrenaline are derived from the tyrosine pathway [36] with
an implication inthe treatment ofdopamine-responsive dystonia and Parkinson’s disease In general, biogenic amines and amino acids were chosen for their importance in several types of can-cer,namely,ovarian,breast,pancreatic,colon,andoralcancers,and neurodegenerativediseases.Similarly,sugarswereincludedinthis optimizationdue totheir large consumption by tumor cells[37] Other two important biological classes of compounds are nucle-osides andlipids, forwhich evident changes havebeen observed
in cancer patients [38,39] The involvement of lipids in different typesof tumorssuch aspancreatic,gastric, liver,lung, colorectal, andthyroidcancer wasshown[40] Aschematicoverviewofthe biosynthesisreactionsispresentedinthesupplementary informa-tion(Figs.S1andS2), clearlyillustratinghowthevariousanalytes areinterrelated.Intotal64fromthe78analytesshown,the miss-ing14analytesmainlyincludemetabolites,whichare uptakenby dietarymeanssuchasessentialaminoacids(6),caffeineandfolic acid,andconsequentlynobiosynthesiscanbeshown.The remain-inganalyteswereincludedintheanalytesetformechanistic ques-tions,i.e.,isomers.Theimportanceandconnectionofaminoacids forthe biosynthesisof biogenic amines,glucose aswell as lipids canbe seen.Further,itiscommonlyassumedthat the dysregula-tionorabsenceofsomemetabolitesmayharmthewell-being
Tofacilitatethe elutionofnon-polar, polar,andionic metabo-lites,the addition ofmethanol, includingadditives,to scCO2 was necessary.Asmallamountofwaterwasaddedtothemobilephase
to improve the peak shape and solvation ofpolar analytes As a consequenceof thelimitedmaximum upperpressure ofthe UH-PSFC system,it wasnot possibletomaintainhighflow ratesand reach 100% of the modifier for the elution of polar compounds
Trang 6Table 3
Summary tables of mass accuracy, selectivity, resolution, and the repeatability of the retention time, signal area, and signal height calculated by the average RSD% for ammonium acetate and ammonium formate in positive and negative ionization mode using TargetLynx and MZmine as data processing software
Repeatability (RSD%)
MZmine
Repeatability (RSD%)
at thesame time Therefore,a decreasing flow rategradient was
applied simultaneouslywiththeorganicmodifiergradient,
allow-ingustoreach100%ofthemodifier.Thisapproach,calledunified
chromatography, wasintroduced by Chester [28] Theadjustment
oftheeluentstrengthofthemodifierupto100%enabledtowiden
thepolarityrangeoftheanalytesetsuitableforUHPSFC/MS
mea-surements
3.2 Screening of water percentage and gradient evaluation with BEH
column
The chromatographic performance of the Viridis BEH column
was evaluated forthree differentpercentages (1%, 2% and5%) of
waterinthemethanolic modifier,whilethe concentrationof
am-moniumacetatewaskeptconstantat30mmolL−1.Theadditionof
2–7%ofwatertothemodifier[12,16,20,31]tofacilitatetheelution
of polarcompounds andtoimprovepeak shape,probablycaused
bytheimprovedsolubility,iscommonlyreportedintherecent
lit-erature Six consecutive injections of the standard mixture were
performedinpositiveandnegativeionmodestotesttheinfluence
of1,2,and5%H2Ointhemodifieronthechromatographic
perfor-mance Retentiontimesof78selectedmetabolitesarereportedin
Table S2.Fig.2showstheoverlayofchromatogramsobtainedfor
threedifferentamountsofwaterinthemodifierforvarious
com-pounds Generally, the retention time increased with increasing
amountofwaterinthemodifier.Forsomeanalytes(Fig.2A,C),the
peak shape worsened,andpeak tailing occurredusing5% of
wa-ter inthemodifier.Furthermore,agradualincrease inthesystem
pressure was observed using 5% of waterin the modifier, which
regularly causedoverpressure of the instrument The experiment
was repeatedafter severalweeksto ensure that this observation
wasnotan artefact.Thesametrendofthegradualincreaseofthe
systempressurewasobserved,whichcouldbecausedbysolubility
issuesoftheadditiveinthemobilephaseleadingtoprecipitation
and columnblockage.A goodcompromise was obtainedwith2%
or1% ofwaterinthemodifier; all compoundswere eluted
with-out anyimpairmentoftheGaussianpeakshape(Fig.2).However,
nonpolar triacylglycerols and diacylglycerols were eluted closeto
thevoidvolumewith1%ofwaterinthemodifier.Consequently,2%
ofwaterinthemodifierwasassessedasthemostsuitableamount
of waterin the modifier to achieve the bestbalance in terms of
retentiontimeandpeakshape
The nextstep ofthestudywasto improvethegradientshape
toobtainagoodseparationoftheentirestandardmixtureaswell
as good peak shapes Two gradients were evaluated, gradient A
(usedfortheevaluationofthepercentageofwaterinthemodifier)
startingfrom5% modifierto75% in10min,andgradientB start-ingfrom5% modifierto 70%modifier in8.5min.Analyteseluted faster andshowed better peak shapeswith gradient B (Fig S3B) compared to gradient A (Fig S3A) Consequently, gradient B was furtherusedforcolumnscreening
3.3 Column screening and performance evaluation
The choice ofstationaryphase chemistry, columndimensions, and mobile phase composition is crucial for successful separa-tion Subsequently, various stationary phases were evaluated for the separation performance of the standard metabolite mixture, such asDiol, 2-PIC, 1-AA, DEA, BEH,CEL 1,HSSC18 SBandHSS T3(Table2).AllcolumnsareclassifiedasUHPSFCcolumns(except HSST3), areproduced by the samemanufacturer (Waters, Torus, andViridiscolumns)andmosthadthesamesub-2μmparticle di-mensionsaswellascolumnlength anddiameter,forbetter com-parability (100 × 3.0 mm I.D, 1.7μm, fully porous hybridsilica) [18,21,31]
The eight stationary phases screened were dedicated UHPSFC columnsfromthesamemanufacturerwithsub-2μmparticles.The majorityofthestationaryphasesarecomposedofthesame back-bone (bridged ethylene hybrid particles) ensuring that the non-selective interactions are comparable and the different selectors bonded on the silica support cause differencesin the chromato-graphic performance The different selectorsbonded to the silica supportallowdifferentselectivitiesfortheseparationofthe stud-iedmetabolitesasaresultofthedifferentinteractioncapabilities
of the analyte andthe stationary phase The simplest stationary phaseregardingtheselectorstructureistheBEHcolumnwithfree silanolsonthesurface,allowingH-bonding andhydrophilic inter-actions.FortheDiolcolumn, propandiolislinkedtothemodified silicasupport.Consequently,thehydroxylgroupsallowH-bonding andhydrophilic interactions, but the hydrocarbon chains provide hydrophobicinteractionsaswell.ThesilicaparticlesofTorus2-PIC, Torus1-AA,andTorusDEA aremodifiedwith2-picolyl-amine, 1-amino-anthracene,anddiethylamine,respectively.Thesestructures allowmultipleinteractionsoftheselectorwiththeanalyte,suchas stericinteractions, hydrogenbonding, Vander Waalsinteractions, dipole-dipoleinteractions,anionic exchangetype, orπ-π interac-tions.ThecolumnsHSSC18SB(100× 3mmI.D;1.8μm)andHSS T3(100× 2.1mmI.D;1.8μm)columnsarepackedwithsilica par-ticlesmodifiedwithoctadecylbondedligandson thesurface, en-ablinghydrophobicinteractions.Thecolumnsdifferintheir resid-ualactivityofthesilanolgroup,asHSSC18T3isend-capped com-paredtoHSSC18 SB.Thefree residualsilanolgroupsadditionally
Trang 7Fig 2 Effect of water percentage (1% - green, 2% - red, and 5% - blue) in the mobile phase on the retention behavior of selected metabolites: A) l -tryptophan, B) N-acetyl-
5-OH-tryptamine (I °) and serotonin (II °), C) d -glucose (I °) and myo-inositol (II °), and D) LPC (18:1) (I °) and PC (36:2) (II °) Analytical conditions: BEH (100 × 3.0 mm, 1.7 μm) column; 60 °C; 1800 psi (ABPR); mobile phase: CH 3 OH + 30 mmol L −1 ammonium acetate, and 1%, 2%, and 5%; composition of the make-up solvent: CH 3 OH + 0.1% formic acid and 5% of H 2 O
allowhydrophilicinteractionsinthecaseoftheHSSC18SB,which
maybeadvantageous fortheanalysisofpolarcompounds.Trefoil
CEL1 (150× 3.0 mm I.D; 2.5μm)is astationaryphase basedon
polysaccharides, inwhichthesilica gelismodifiedwithcellulose
tris-(3,5-dimethylphenylcarbamate), allowingmultiple interactions
suchasstericinteractions,hydrogenbonding,π-πinteractions
Different chromatographic parameters were evaluated to
de-termine the best column for the separation of the analyte
mix-ture, such as the number of compounds not detected and the
peak asymmetry factor (As), which is calculated as the ratio of
the peak widthinthe backhalf andthe peakwidthin thefront
halfat10% ofthepeak height.Forbettercomparability,the same
gradient was applied forthe separation of78 metabolites forall
stationary phases investigated (Fig S3B) The retention times of
each standard for each tested column are reported in Table S2
Fig 3A shows the chromatogramsof guanine and guanosine
de-pending on the employed stationary phase The highest number
of compounds detected, depending on the stationary phase, was
as follows: Diol > BEH > 2-PIC > HSS C18 SB > DEA > Cel
1 > 1-AA > HSS T3 (Fig 3B) This indicates that with
increas-ing hydrophobicity and bulkiness of the stationary phase
selec-tor, lessanalytesare detected However, some hydrophobic
inter-actions favor separationand detectionincomparison to only
hy-drophilicinteractions, asreflected forthe DiolandBEH columns
Eachcompoundwasinjectedseparatelyaswellasinamixturefor
each column Therefore, thecompounds not detected inthe
ana-lytemixturearebelowthedetectionsensitivitybecausetheywere
identifiedwheninjectedseparately,someathigherconcentrations
Thisproves that theanalytesareelutedforeach column,but
be-causeofthebroadpeakshapeandhighasymmetry,theywere
be-low thedetectionsensitivity,notallowingtheir identification.The
mostdifficultcompoundstoidentifyformostofthecolumnswere
metabolites with primary amines in the structure The primary
amines may undergo ionic interactions with the free silanols of
thestationaryphase.Asionicinteractionsaregenerallyslow,broad
peakshapescanbeobserved,whichmayleadtosensitivityissues
Theenhancementofthecationconcentrationinthemobilephase
could lead to an improvement of the peak shape and sensitivity
sincecations functionasdisplacers.Onthe otherhand,increased baseconcentrationsinthemobilephasemayleadtoion suppres-sion The detection sensitivity was diminished especially for the metabolites PS, LPS, PG, LPG, PA and LPA, putrescine, spermine, spermidine,anddopamine,and5-methoxyindole,5-hydroxyindole, D,L-dylurenine,andfolicacid
The performance of the columnwas generally considered ac-ceptablewhentheAsvaluewasintherangeof0.9–1.5.Therefore, foreach column, the percentageof compounds not detected, the analyteswithAs below0.9andwithAsgreaterthan1.5were cal-culated(Fig.S4).Thehighestpercentageofsymmetricalpeakswas observed forBEH> DEA > Diol> Cel 1> 1-AA > 2-PIC > HSS C18>HSST3inpositiveionmodeandBEH>DEA>2-PIC>HSS T3 > Cel 1 > Diol > 1-AA > HSS C18 in negative ion mode (Table S3) Furthermore, the asymmetry values of each detected compoundandtheirtotalaverage fortheeight screenedcolumns are reported in Table S3 Broad chromatographic peaksand tail-ingwerefound forthelipidclassesPS,LPS,PG,LPG,PA,andLPA,
as also known from the literature For biogenic amines, namely spermine,spermidine,andputrescine,abroadanddistortedpeak shapewasobserved.TheresultssuggestthattheDiolcolumn rep-resentsagoodcompromisebetweenthenumberofdetectedpeaks andtheasymmetryvaluescomparedtotheothersevencolumns The performance of the method was further investigated
by determining the mass accuracy, selectivity, resolution, peak area, peak height, retention time stability, and number of total compounds detected in both polarity modes (Tables S4–S9) L tryptophan wasselected as the reference compound forthe cal-culationofresolutionandselectivity,asoneofthelasteluting an-alytes The average time of the first peak in the run from three consecutive blanks (solution of CH3OH/CHCl3, 1:1 v/v) was con-sideredasthe voidtime needed tocalculate the capacityfactors (TableS5) Themedianofselectivityandtheaverageofall deter-mined resolutionvalues foreach columnwere investigated.They were determined from the average values of six consecutive in-jectionsofthestandard mixtureforeach columnbyapplyingthe optimized conditions The median and the average of the over-all mass accuracy, selectivity, and resolution are reported in
Trang 8Ta-Fig 3 (A) Selected chromatograms for guanine (I °) and guanosine (II °) on the various stationary phases (red: 1-AA, green: BEH, light blue: Diol, violet: 2-PIC, dark blue:
DEA) (B) Bar chart for the number of detected compounds (blue) and non-detected compounds (red) for individual screened columns (C) Median of the selectivity values and (D) Average of the resolution values for all metabolites on the various stationary phases Analytical conditions: mobile phase: CH 3 OH + 30 mmol L −1 ammonium acetate and 2% of water; mobile phase of the make-up pump: CH 3 OH + 0.1% formic acid and 5% of H 2 O; 60 °C, 1800 psi (ABPR), ESI ( + ) and ESI ( −)
bles S4–S6, as well as the average and RSD% of the peak area,
peak height, andretentiontime inTables S7–S9.The median
ob-tainedforallanalyteswasusedtocomparetheoverallselectivity
of the stationary phases, since the overall average may be
influ-enced by analytes eluted close to the void volume The highest
median selectivity was observed for DEA > Cel 1 > 2-PIC =
1-AA > BEH > HSS T3> Diol > HSSC18 in positive ion (Fig 3C)
mode and BEH > 1-AA > 2-PIC = HSS T3 = HSS C18 > Cel
1>Diol>DEAinnegativeionmode(TableS5).Itshouldbe
men-tionedthatmuchfewercompoundsweredetectedanddifferences
in themedianselectivity are negligiblein thenegativeionmode
compared tothepositive ionmode(TablesS2 andS5).The
high-est average resolutionwasobserved forDEA > Diol> BEH >
2-PIC>1-AA>Cel1>HSSC18>HSST3inpositive(Fig.3D)and
DEA > Diol > BEH > 2-PIC > 1-AA > HSSC18 > Cel 1 > HSS
T3 in negative ion mode (Table S6) A small shift in mass
pre-cision wasobserved, whichcorresponds to the retentiontime of
the analyte (Table S4) The ionization efficiency depends on the
gradient shape ofthechromatographic runand, consequently, on
the retention times ofthe analytes Furthermore, the type of
in-teractionsoftheanalyteswiththestationaryphaseinfluencesthe
peak shape, because ionic interactions are slow andmaylead to
broaderpeaksincomparisontofasterinteractionssuchas
interac-tions basedon partitionorsolubility.The peakarea, peakheight,
and averageretention time together withthe retentiontime
sta-bility were investigated The highest average peak area was
ob-served forHSST3>HSSC18>Diol>BEH>Cel1>1-AA>
2-PIC > DEA in positive ion mode and HSS T3 > HSS C18 > Cel
1> > BEH>2-PIC >Diol>1-AA> DEAinnegativeionmode
(Table S7).Thehighestaveragepeakheight wasobservedforHSS
T3 > HSSC18 > 2-PIC > Diol> 1-AA> BEH > Cel1 > DEA in
positive ionmode andDiol> DEA >2-PIC > BEH> 1-AA>Cel
1 > HSSC18> HSST3in negativeionmode (Table S8) The
av-erageretentiontimesonthedifferentstationaryphasesshowthe
distribution of analyteswithin the chromatographic run andthe
extent and type of interactions of the analyte with the selector
The highestaverage retentiontime was observedfor BEH> HSS C18 > Diol > 1-AA > DEA > 2-PIC > Cel 1 > HSS T3 in pos-itive ion mode and BEH > DEA > Diol > 1-AA > 2-PIC > HSS C18>Cel1>HSST3innegativeionmode.Therelativestandard deviationof the retention timesof the analyte for6 consecutive injectionsofthemetabolitemixtureoneachstationaryphasewas investigated, describing the reproducibility of the retention time ThehigheststabilityofretentiontimewasobservedforDiol>Cel
1>DEA> 1-AA> 2-PIC> BEH> HSSC18 >HSST3inpositive ionmodeandDEA >Diol>1-AA>2-PIC> BEH>Cel1> HSS C18>HSST3innegativeionmode
TheDiolcolumndidnotprovidethebestresultsforeach eval-uated parameter,butthe comparisonof thechromatographic pa-rametersmentionedforeachcompound andstationaryphase, in-cluding the total numberof detected peaksin positive and neg-ative ion modes, reveals that the overall best performance was achievedwiththeDiolcolumn,asalsopreviouslyreportedfor uri-narymetabolites[18].67ofthe78structuralandchemicalhighly diversecompounds inthemixtureweredetectedontheDiol col-umn In order to investigate the reason for detected and non-detected compounds dependingon theanalyte structure, the an-alyte set was classified accordingto their functional groups (Ta-ble S14) However, no general trend depending on the presence
offunctionalgroupswasobserved.TheDiolcolumnwasusedfor furtherevaluation withinthis study.The putative explanationfor whichDiolworkedwellfortheseparationofmostoftheanalytes
isthe possiblepolarandhydrophobic interactions of the station-ary phase withthe polar andhydrophobic parts of the analytes Further,thesmallselectorstructureoftheDiolcolumnmayfavor theaccessibilityoftheanalytestointeractwiththeselectorofthe stationaryphase,incontrast tothebiggerselectorstested, which maybeleadingtosterichindrance
Thereliable separationofisomeric and/or isobaricmetabolites
incomplexbiologicalsamplesisimportantinmetabolomics stud-ies Examples of isomeric metabolites included in our selected standardmixtureareleucine,isoleucine,andnorleucine,some
Trang 9as well asN-acetyl-serotonin and1-methyl-tryptophan.Examples
of isobaric metabolitesinvestigated are asparagine andornithine,
asparticacidand5–hydroxy-indole,glutamine,andlysine,aswell
as glutamic acid and 5–methoxy-indole Only the BEH and the
Diolcolumnyieldedapartial separationofleucineandisoleucine
with respect to norleucine, while their coelution with all other
columns was detected Dopamine and octopamine are also
im-portant examples of isomers The first was below its detection
sensitivity, which did not allow detection in most cases; on the
other hand,octopamine waseasily detected However,each
stan-dard wasinjected individually,allowing goodseparation between
thesetwocompounds.Infact,octopaminewaselutedoneach
col-umnbetween5 and7min,while dopamine waseluted between
8 and 9 min on various stationary phases Finally, the isomers
d-glucose and myo-inositol,as well as N-acetyl-serotonin and
1-methyl-tryptophan, were always separated independently of the
stationary phase All detected isobaric metabolite pairs were as
well separatedonallstationaryphasesinvestigated.Theseresults
have shownthatthe optimizedmethodprovides goodseparation
ofnotonlyverydiversemetabolitesbutalsosomeoftheirisomers
andisobars
18 of the 78 compounds investigated in our study were also
included in the analyte set investigated by Losacco et al of 49
compounds[16]andDesfontaineetal.of57compounds[12],such
assome aminoacids,biogenic amines,nucleosides andlipids.For
comparison reasons, special focus was placed on the evaluation
of theseparation performance ofthose compounds Agood peak
shape and peakasymmetry havebeenreached applyingthe Diol
column, i.e., foradenosine,leucine,andsphingomyelin (1.61,1.45,
and 1.31,respectively; TableS3) Additionally,the% RSD of
reten-tiontimestabilitywasdeterminedforeachcolumn(TableS9).The
overall stabilityofthe retentiontime was0.31% RSDforthe Diol
columnand0.06,0.41and0.07%foradenosine,leucineand
sphin-gomyelin,respectively.Inconclusion,thereportedmethodyielded
comparableresultsfortheDiolcolumnincomparisontothedata
shownbyLosaccoetal.usingthePoroshellHILICcolumn[16].Itis
importanttoemphasize,thatthecolumnscreeningwasperformed
forUHPSFC/MSdedicatedsub2-μmcolumnsfromthesame
man-ufacturer and the larger set ofanalytes in terms of polarity and
mass rangeinthepresentwork.The metabolitesinvestigatedare
mainly interrelated (Figs.S1 andS2), besideschosen analytes
in-cludedtostudymechanisticaspects.Thisenabledacompleteand
exhaustiveanalysisofchromatographicandmassspectrometric
pa-rameters
3.4 Evaluation of ammonium acetate versus ammonium formate as
an additive in the modifier
The influence of the type of additive in the mobile phase
on retention time, peak area, peak height, mass accuracy,
selec-tivity, resolution, and peak asymmetry for the standard mixture
was investigated using the Diol column Six consecutive
injec-tionswereperformedusing30mmolL−1ofammoniumacetatein
CH3OH/H2O(98:2,v/v)or30mmolL−1 ofammoniumformate in
CH3OH/H2O(98:2,v/v)asa modifier.The peakareasandheights
of each detected compound were normalized to the average
in-tensity ofthe lockmass todiminish theinfluence ofthe drift of
the instrumental response over time (Table S10) The processed
dataofthenormalizedareaandnormalizedheightwerecompared
using bar graphs for the positive (Figs S5 andS6) and negative
(Figs S7 andS8) ionization mode Signal responses forall
com-pounds were higher for ammonium acetate compared to
ammo-nium formate (Table S10) As a result, a highernumber of
com-poundsweredetectedwithammoniumacetate(67)thanwith
am-monium formate (65) Table 3 shows a summary of the average
Fig 4 Base peak intensity chromatograms of the standard set of metabolites ob-
tained under optimized conditions (black) and reconstructed ion current chro- matograms for selected compounds: caffeine (red), MG (0:0/18:1/0:0) (blue), Cer (d36:2) (olive), sphinganine (d18:0) (orange), melatonin (wine), adenine (magenta), adenosine (violet), N-acetyl-5-OH-tryptamine (royal), LPC (18:1/0:0) (cyan), palmi- toylcarnitine (dark yellow), acetylcarnitine (dark cyan), taurine (pink), l -tyrosine (light magenta), l -tryptophan (dark gray), 5-OH- l -tryptophan (light orange), and l - arginine (light blue) Analytical conditions: Diol (100 × 3.0 mm; 1.7 μm); mobile phase: CH 3 OH + 30 mmol L −1 ammonium acetate and 2% of water; composition of the make-up solvent: CH 3 OH + 0.1% formic acid and 5% of H 2 O; 60 °C, 1800 psi (ABPR), ESI ( + )
mass accuracy, selectivity, resolution and RSD of the peak area, peak height, andretentiontime Furthermore,detailed valuesfor each compound, depending on the additive applied on the Diol column,arereportedinTablesS3–S10 forammoniumacetateand Tables S10–S11 for ammonium formate The average mass accu-racy,selectivity,andresolutionwasslightlyhigherforammonium formate than ammoniumacetate (Table3) No generaltrendwas observedforthesignal andretentiontimestabilitydependingon theadditive.However,ammoniumacetatewasselectedasthe ad-ditive of choice in the mobile phase for the investigated analyte setbecauseoftheslightlyhighernumberofdetected compounds and the higher signal response Data processing was performed independently with TargetLynx, which was used by default, and MZmine, to assess whether the data processing software has an impactonthe results(TableS11) Thesamechromatographicand method parameters were investigated with MZmine aswith Tar-getLynx,and comparedto each other Data were comparablebut notthesame,whichshowsthatthedataprocessingsoftware em-ployedmayhaveanimpactontheresults
Finally,theDiolcolumn(100× 3mmI.D;1.7μm),themodifier
ofMeOHwith30mmolL−1 ammoniumacetate and2%ofwater, themake-upsolventofMeOHwith0.1%formicacidand5%of wa-ter(seeFig.4)wereevaluatedasthebestchoicefortheseparation
oftheinvestigatedanalyteset
3.5 Comparison of ESI and APCI ionization techniques
Theionization efficiencymaychangedependingonthe chem-icalpropertiesandchemicalstructureoftheanalyzedcompounds andtheappliedionsource.ESIisthemostwidelyusedionsource Sensitivitystronglydependsontheflowratesemployed,sinceESI representsaconcentration-dependentionizationtechnique.APCIis
amassflowdependentionizationprocessmoresuitableforhigher flow rates UHPSFC/MS methods generally use flow rates higher thanthoseofUHPLC/MSmethods;therefore,theevaluationofthe ionsource on the ionization efficiencyof target compounds may
be ofinterest.However, themajorityof UHPSFC/MSmethods use
Trang 10a splitter,whichreduces theflow intothemass spectrometer
fa-voring ESI.A systematic investigation wasconducted to evaluate
the influenceofthe ionizationsourceonthe numberandtype of
detected analytes Thestandard mixturewasanalyzedby ESIand
APCIinbothpolaritymodes.Theoptimizedchromatographic
con-ditionsandoptimizedionsourceparameterswereapplied.The
re-sults showed that ESI in general led to a higher ionization
effi-ciency compared to APCI(Figs S9–S12).However, forsome
ana-lytes,thepeakarea andpeak height(normalizedtothesumarea
and sumheight considering the total compounds in the positive
andnegativeionization mode fortheESIandAPCIsources)were
higher forthe APCIsource than forthe ESIsource, showingthat
ESI andAPCI can be complementary (Table S12) The sensitivity
was higherforseveralamino acids,such asl-tyrosine,ornithine,
phenylalanine, taurine, as well as l-tryptophan, l-arginine, and
l-lysine and two nucleosides (adenine and guanine) using APCI,
and dopamine wasonly detected using the APCIsource On the
other hand, the areas and heights of the l-carnitine derivatives,
LPC(18:0),andmelatoninwereenhancedwithESI.l-carnitineand
acetyl-l-carnitine were only detected using ESI To illustrate the
comparisonofthenormalizedareaandheightforsomeidentified
standardsinpositiveionmodeforbothionsources,bargraphsare
showninFigs.S9–S12.Thecorrespondingvaluesofthenormalized
peak areas,heights,andretentiontimesofeach standardforboth
ionsourcesarereportedinTableS12.Thetotalnumberofdetected
compoundswas67and48forESIandAPCI,respectively,showing
thewider applicationrangeofESIfortheinvestigatedanalyteset
[32].Inthenegativeionizationmode,mostanalyteswere not
de-tectedusingAPCI(Figs.S11andS12).Furthermore,inthenegative
ionization mode, thesignal responsewassignificantlylower than
inthepositiveionization mode,regardlessoftheionsourcetype
ESI provided theoverall best ionization efficiencyforthe analyte
setinbothionmodes
3.6 Application to human plasma
Optimized chromatographic and MS conditions were applied fortheanalysisofpooledhumanplasmasamplestoevaluate the applicability of the method to real samples The protocol used for sample preparation was based on the application of an ad-ditionalstep prior to protein precipitationbased on the addition
ofproteinase K.Thisprocedure allowed the releaseof associated metabolitesthrough relaxationof the tertiary structure of native proteins andconsequently a higherpossibility oftheir identifica-tion [41] In addition,plasma samples obtained by the following protein precipitation were injected and analyzed in MSE mode
MSE modeallows theuntargeted scanningof theMS andMS/MS levels by applyinglow andhigh collisionenergy within one run Thisincreasesthe identificationconfidence ofmetabolites, asthe characteristicfragmentsofmetabolitesprovideadditional informa-tion.First,thestandardmixturewasanalyzedtoobtainclean frag-mentionspectraasreferencewithoutinterferencescausedbythe complexmatrixofarealhumanplasmasampleusingESIandAPCI
No differencesinfragmentationbehavior were observedbetween ESIandAPCI(TableS13).Thedilutedhumanplasmasample(1:10) wasanalyzedusingESIandAPCI.Fig.5AshowstheTICofhuman plasmaobtainedwithESIandAPCI Itcanbeseenthatthe sensi-tivityishigherforESIthanAPCI,alsoforrealhumanplasma sam-ples The extractedionchromatograms (XIC)of selected metabo-litesdetectedinhumanplasmaarepresentedinFig.5BusingESI Thetargeteddata analysisrevealed that44 and5 compounds in-cludedintheanalytesetwerealsodetectedinthedilutedplasma sampleusing ESI andAPCI, respectively, in positive andnegative ionmode(Table S15).The reductionofthesamplecomplexityby optimizingthesamplepreparationprotocol,i.e.,usingsolid phase extraction,mayhelpincreasing thesensitivitytodetectthewhole analyte set in human plasma The untargeted MSE approach al-lowedthe use ofmetabolomicsdatabases to link m/z features to
Fig 5 (A) Impact of the type of ion source on sensitivity Base peak intensity chromatogram of human plasma using red) ESI and blue) APCI (B) Selected extracted ion
chromatograms of human plasma using ESI (green: ornithine, blue: glucose, red: serotonine, black: adenosine) Pie charts of the untargeted m/z feature analysis in human plasma for (C) positive and (D) negative ion mode using MS DIAL 5823 m/z features were detected in positive and 2769 m/z features in negative ion mode The m/z features were categorized according to the compound class: red) analyte set of the study, green) nucleoside and derivatives, orange) amino acids and derivatives, violet) polyphenols, light blue) lipids and dark blue) other metabolites Analytical conditions: Diol (100 × 3.0 mm; 1.7 μm); mobile phase: CH 3 OH + 30 mmol L −1 ammonium acetate and 2% of water; composition of the make-up solvent: CH 3 OH + 0.1% formic acid and 5% of H 2 O; 60 °C, 1800 psi (ABPR), ESI ( + )