1. Trang chủ
  2. » Kỹ Thuật - Công Nghệ

Performance of a radiophotoluminescence (RPL) system in environmental and area monitoring

8 0 0

Đang tải... (xem toàn văn)

THÔNG TIN TÀI LIỆU

Thông tin cơ bản

Tiêu đề Performance of a radiophotoluminescence (RPL) system in environmental and area monitoring
Tác giả E.G. Yukihara, F. Assenmacher
Trường học Paul Scherrer Institute
Chuyên ngành Radiation Safety and Security
Thể loại journal article
Năm xuất bản 2021
Thành phố Villigen
Định dạng
Số trang 8
Dung lượng 0,91 MB

Các công cụ chuyển đổi và chỉnh sửa cho tài liệu này

Nội dung

The objective of this work was to perform an extensive side-by-side comparison between a radiophotoluminescence (RPL) dosimetry system and thermoluminescent dosimeters (TLDs) or EDIS-1™ Environmental Direct Ion Storage Dosimeters during environmental and area monitoring.

Trang 1

Available online 29 December 2020

1350-4487/© 2021 The Authors Published by Elsevier Ltd This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

Performance of a radiophotoluminescence (RPL) system in environmental

and area monitoring

E.G Yukihara*, F Assenmacher

Department of Radiation Safety and Security, Paul Scherrer Institute, 5232, Villigen, PSI, Switzerland

A B S T R A C T

The objective of this work was to perform an extensive side-by-side comparison between a radiophotoluminescence (RPL) dosimetry system and thermoluminescent dosimeters (TLDs) or EDIS-1™ Environmental Direct Ion Storage Dosimeters during environmental and area monitoring The measurement locations include points around and within the perimeter of the Paul Scherrer Institute (PSI) and nearby facilities These data are complemented by a study on the RPL detection limit, uncertainty and dose linearity, as well as two intercomparisons of environmental dosimeters, a Swiss intercomparison organized by PSI and an international

intercomparison organized by the German National Metrology Institute (Physikalisch-Technische Bundesanstalt, PTB, Germany) The laboratory irradiations show that the detection limit of the RPL dosimeters is < 50 μSv if the time between pre-dose measurement and readout is < 200 days, and the response is linear up to at least

100 mSv with less than 15% deviation from linearity, satisfying the requirements of the Swiss dosimetry ordinance The RPL doses were more consistent than the TLD doses over time The RPL system shows slightly lower doses (12–14%) in comparison with EDIS-1 dosimeters The intercomparisons for passive environmental

dosimeters in terms of H*(10) showed a good agreement between the RPL dose values and the conventional true values Altogether, the results demonstrate the

equivalence between RPL and the other dosimetry systems, providing support for the RPL adoption for environmental dosimetry

1 Introduction

Radiophotoluminescence (RPL) dosimeters based on Ag+-doped

phosphate glasses are now commercially used for dosimetry worldwide

(Miyamoto et al., 2011), having been adopted in Europe by the Institute

for Radiological Protection and Nuclear Safety (Institut de radioprotection

et de Sûret´e Nucl´eaire, IRSN, France) and the International Atomic Energy

Agency (IAEA) The system is used for individual monitoring since 2016

by the accredited Dosimetry Laboratory of the Paul Scherrer Institute

(PSI) Assenmacher et al (2017) presented the commissioning data of

the system in terms of the operational personal dose equivalent Hp(10),

showing that it satisfies the requirements of the Swiss Ordinance of the

Federal Department of Home Affairs (EDI) on Personal and

Environ-mental Dosimetry (hereafter called “Swiss dosimetry ordinance”) (Swiss

Federal Council, 2017)

The RPL is based on the UV-induced photoluminescence signal due to

luminescence centers (Ag0 and Ag++) created in Ag+-doped phosphate

glass upon exposure to ionizing radiation (Miyamoto et al., 2011) The

badge design used at PSI, IRSN and IAEA (GBFJ-01) consists of a holder

containing five different filter types, which defines five areas in the glass

detector with different photon energy responses read out by an

auto-mated reader (Hocine et al., 2011; Hocine, 2012; Assenmacher et al.,

2017) These five signals are combined using a proprietary linear

algorithm to calculate the operational quantities Hp(10), Hp(0.07), and

H*(10)

For environmental dosimetry, however, the GBFJ-01 badge design is not necessarily appropriate because of the non-isotropic response

Assenmacher et al (2017) showed that the Hp(10) angle dependence for S–Cs (Cs-137) and N-80 radiation qualities (ISO, 1999) does not deviate

by more than 20% for angles up to 60◦ For environmental dosimetry, however, the angle dependence needs to be verified for the operational

quantity ambient dose equivalent H*(10) Furthermore, environmental

dosimeters are required to be tested up to larger angles than personal dosimeters, both according to the Swiss dosimetry ordinance and the international standard IEC 62387 (IEC, 2020)

Limited information is available on the applicability of the RPL system for environmental and area monitoring A previous study on a different environmental RPL dosimetry system reports on characteristics such as batch homogeneity, reproducibility, linearity, detection limit, energy dependence and UV sensitivity (Ranogajec-Komor et al., 2008), but the applicability of those results is limited: firstly, it does not necessarily apply to the dosimetry system used at PSI, IRSN and IAEA;

secondly, the results are not presented in terms of H*(10) and the angle

dependence was not investigated

To fill this knowledge gap, the energy and angle dependence of the

H*(10) for the dosimetry system consisting of the glass type FD-7, GBFJ-

* Corresponding author Department of Radiation Safety and Security, Paul Scherrer Institute, 111 Forschungsstrasse, 5232, Villigen, Switzerland

E-mail address: eduardo.yukihara@psi.ch (E.G Yukihara)

Contents lists available at ScienceDirect Radiation Measurements

journal homepage: http://www.elsevier.com/locate/radmeas

https://doi.org/10.1016/j.radmeas.2020.106514

Received 25 September 2020; Received in revised form 23 November 2020; Accepted 21 December 2020

Trang 2

01 badge, FGD-660 reader and CDEC-Easy software (Chiyoda Technol

Corp.) was investigated (Assenmacher et al., 2020) Although the energy

dependence was shown to satisfy both the Swiss dosimetry ordinance

and the IEC 62387, the response for angles close to 90◦exceed the limits

of the Swiss dosimetry ordinance (±20%) for the N-80 radiation quality

Regarding the IEC 62387 standard, the response at 90◦shall be

deter-mined by rotating the dosimeter 360◦around the reference direction of

the dosimeter or by eight irradiations rotated by 45◦each In this “full

rotation” scenario, the response of the RPL dosimeters was within the

limits of the IEC 62387 At 75◦irradiation for the N-80, however, the

response was still outside the requirements of the IEC 62387

Nevertheless, the environmental radiation is typically non-

directional and characterized by a broad photon energy spectrum

(Tereda et al., 1980) Therefore, the angle dependence investigated in

laboratory with directional irradiation at low energies may not be

relevant for practical environmental and area monitoring, except in the

case of an unexpected low energy irradiation at angles close to 90◦ For

this reason, it is important to carry out a side-by-side comparison

be-tween the RPL dosimetry system and other dosimetry systems used for

environmental and area monitoring, to demonstrate the equivalence of

RPL in realistic conditions

PSI’s Dosimetry Laboratory performs routine environmental and

area monitoring within and outside PSI’s perimeter and in the vicinity of

various other facilities, including at the fence of the Beznau nuclear

power plant and the central interim storage facility Zwilag (Zwilag

Zwischenlager Würenlingen AG) Because of the diversity of dose rates

and energy spectra in the various measurement points, this monitoring

network offers a perfect opportunity to perform this comparison

be-tween RPL and other techniques Outside PSI’s perimeter the radiation

field is dominated by cosmic and terrestrial natural radiation (typically

~0.6 mSv/year) Inside PSI’s perimeter, the radiation field can be

increased due to radiation from the various accelerator facilities,

generated by bremsstrahlung with energies up to hundreds of MeV or a

few GeV, depending on the accelerator, as well as prompt gamma

ra-diation due to nuclear reactions and activation of the shielding

com-ponents In monitored points within sealed off areas the photon dose can

reach values >10 mSv/year

To demonstrate the feasibility of using RPL for environmental

dosimetry for the direct applications at PSI, since 2017 we have carried

out a side-by-side comparison between the RPL dosimetry system and

other dosimetry systems, including Al2O3:C, TLDs and EDIS-1™

Envi-ronmental Direct Ion Storage Dosimeters In addition, we participated in

two intercomparison for environmental dosimeters, a Swiss

intercom-parison organized by PSI, and an international intercomintercom-parison

orga-nized by the Physikalisch-Technische Bundesanstalt (PTB, Germany)

(Dombrowski, 2019)

In this paper we report representative results of this extensive side-

by-side comparison between the RPL system and the TLD or EDIS-1

dosimeters during environmental and area monitoring These data are

complemented by a study on the detection limit (ISO, 2019),

un-certainties and dose linearity, as well as the two intercomparisons for

passive environmental dosimeters organized by PSI and PTB

2 Materials and methods

2.1 Dosimetry systems

The RPL dosimetry system used in this study is described in Table 1

Before use, the RPL glass detectors were regenerated and the pre-dose

(PD) signal of the detector (intrinsic background signal) was read and

automatically stored in the database The RPL glass detectors were then

assembled in the badges and stored in the laboratory until deployment

After deployment, the RPL dosimeters were disassembled and the RPL

glass detectors were pre-heated for 1 h at 100 ◦C to accelerate the build-

up process of the RPL signal and then read out Two identical RPL

readers were available for these studies

The TLD dosimetry system used in this study is part of the ISO/IEC

17025 (ISO/IEC, 2017) accredited system from the Paul Scherrer Insti-tute (accreditation number STS 0491, Swiss Accreditation Service SAS) and is described in Table 1 Because of their high sensitivity, Al2O3:C detectors are used in the environmental dosimeters and in area dosim-eters outside controlled areas, whereas 7LiF:Mg,Ti detectors are used inside controlled areas, where the dose limits are higher The TLDs were regenerated approximately 14 days before deployment For each data point, the same TLD is used in alternating quarters (e.g., 2018Q01, 2018Q03, 2019Q01, 2019Q3, etc.) The glow curves and integral values are stored in a database to be processed later Auxiliary TLDs (AUSD) were used to estimate the dose accumulated while in storage after the reset and before deployment The TLD readout is performed typically the day after return of the dosimeters

The EDIS-1™ Environmental Direct Ion Storage Dosimeters are commercialized by Mirion Technologies, Inc

2.2 Dose calculation algorithm

The RPL dose calculation was performed using one of the dose calculation algorithms included in the software CDEC-EASY The

selected algorithm calculates H*(10) without background subtraction,

only subtraction of the pre-dose (PD) value read after regeneration The dose is, therefore, the total dose since the PD measurement

For the environmental and area monitoring, the ambient dose equivalent rate during storage in the laboratory, estimated as 1.9 μSv/ day at PSI based on historical data and confirmed here (see Section 3.1), was subtracted from the measured values using:

day× (T dd) (1)

where M is the final (background subtracted) estimate for the dose equivalent H*(10), H_10X is the RPL indicated value in terms of H*(10) calculated by CDEC-EASY, T d is the total number of days between the

pre-dose measurement and the detector readout, d is the number of days

the detector was deployed For the intercomparisons no subtraction was done, since the dose transport and storage doses were evaluated using transport detectors

For the study of the detection limit, uncertainties and dose response,

the detectors were evaluated in terms of Hp(10) without subtraction of

the background This study was performed in terms of Hp(10) because it was part of a characterization of the RPL system for personal dosimetry

Nevertheless, the difference between the Hp(10) and H*(10) calculation

algorithm is only ~12% and, therefore, the results are approximately

Table 1

Luminescence dosimetry systems used in this study

Parameter RPL TLD (Al 2 O 3 :C) TLD ( 7 LiF:Mg,Ti) Detector FG-7 Ag-doped

phosphate glass detectors (35 mm ×

7 mm × 1.5 mm)

Al 2 O 3 :C single crystal (~5 mm diameter × 0.9

mm thickness)

7 LiF:Mg,Ti pellet (~4.5 mm diameter × 0.9 mm thickness) Dosimeter GBFJ-01 ALNOR-type with

2 detectors ALNOR-type with 3 detectors Annealing 370 ◦ C/10 min for

regeneration (complete cycle: 2 h);

100 ◦ C/1 h before readout

310 ◦ C for 15 s in reader 305

◦ C for 10 s in reader

Readout – 265 ◦ C for 12 s 305 ◦ C for 10 s Readers FGD-660 (two

identical readers) ALNOR (DOSACUS) ALNOR (DOSACUS) Dose

calculation software/

method

CDEC-Easy Average of two

detectors, corrected for individual detector sensitivity

Average of three detectors, corrected for individual detector sensitivity

Trang 3

valid for H*(10) as well

The TLD doses are calculated as the average dose of two or three

detector elements (see Table 1), corrected by the individual element

sensitivity The calibration factor is determined using calibration

de-tectors irradiated using Cs-137 with H*(10) = 1.25 mSv in case of Al2O3:

C TLDs and H*(10) = 2.5 mSv in case of 7LiF:Mg,Ti TLDs The dose due

to the period between the dosimeter’s regeneration and start of the

deployment is subtracted using the AUSD dosimeters, prepared together

with the routine dosimeters, but read when the deployment begins

Here the indicated values of the systems are also referred to as “RPL

doses”, “TLD doses” or “EDIS-1 doses”

2.3 Calibration and laboratory irradiations

The RPL and TLD systems used in this study are calibrated at PSI’s

Calibration Laboratory using a Cs-137 source The source calibration is

traceable to the primary standards at the PTB (Germany) and has a

relative uncertainty (coverage factor k = 2) of the irradiated dose values

in the range from 3.4% to 3.9% The EDIS-1 detectors are calibrated by

the manufacturer (Mirion Technologies) In the case of the RPL, the

entire system is calibrated every quarter using calibration dosimeters In

the case of TLDs, calibration detectors are irradiated in the middle of the

monitoring interval to account for the possible fading

Additional laboratory irradiations were carried out at PSI’s

Cali-bration Laboratory to characterize the dose response, uncertainties

involved, and detection limits Various RPL glasses were used for each

dose as specified in the text and the dosimeters were read various times

after irradiation As mentioned in Section 2.2, in these studies the

do-simeters were irradiated on phantom with Cs-137 in terms of Hp(10) (on

phantom) with 2 mm PMMA build-up and with doses values ranging

from 0.010 mSv to 100 mSv Unirradiated detectors were also used The

readouts (after pre-heating) were carried out at various times after the

irradiation

2.4 Environmental and area monitoring network

The environmental monitoring network of PSI includes ~35

mea-surement points outside PSI’s perimeter, up to almost 10 km distant

from PSI These are complemented by ~42 points immediate outside

PSI’s perimeter or at the fence, as well as ~22 points at the fence of

nearby facilities (Beznau nuclear power plant and Zwilag) These were

measured using Al2O3:C TLDs and RPL dosimeters

Area dosimeters (inside PSI’s perimeter) outside controlled areas

consisted of Al2O3:C TLDs and RPL dosimeters Because of the low doses

in these points, the results are similar to the environmental monitoring

network described above and will not be presented Area dosimeters

inside controlled areas were carried out using 7LiF:Mg,Ti TLDs and RPL

dosimeters installed in wooden boxes with a transparent plastic cover

affixed to the wall inside the buildings The reference levels are 80 mSv/

year for permanent workplaces, 200 mSv/year for temporary

work-places; locked-off areas have no reference levels These reference levels

were established considering the occupancy of the areas and the

occu-pational dose limits for radiation workers of 20 mSv/year

In addition to the environmental and area monitoring of PSI,

mea-surements were also carried out using Al2O3:C TLDs from PSI in parallel

with the EDIS-1 dosimeters used routinely by the Leibstadt nuclear

power plant (Kernkraftwerk Leibstadt, KKL) in their environmental

monitoring program

2.5 Intercomparisons

2.5.1 Swiss intercomparison 2016

The Swiss intercomparison was organized by PSI’s Calibration

Lab-oratory under commission of the Swiss Federal Nuclear Safety

Inspec-torate (ENSI) in agreement with the Swiss Federal Office of Public

Health (FOPH) PSI’s Dosimetry Laboratory was an independent

participant in the intercomparison and was not involved in its organi-zation The intercomparison included passive detectors (TLD, RPL) from three institutions, as well as active detectors and spectrometers (ioni-zation chambers, Geiger-Müller counters and high-purity germanium detectors) from a total of nine institutions

The passive dosimeters were exposed in a reference location at PSI for a period of six months in 2016 Irradiations were also carried out in laboratory with Cs-137 at two dose levels, one typical for environmental dosimeters (0.3 mSv) and a higher dose to check the calibration of the dosimeters (1.7 mSv) Transport detectors were also used and stored in a lead shielding for the period the other dosimeters were exposed in the field The dose in the lead shielding was provided by the organizers and subtracted from the measured values

The dose for the field irradiation was estimated using a pressurized ionization chamber RSDetection (model RS-131-S131-200) from the PSI’s Calibration Laboratory, calibrated in the PTB reference fields for

the ambient dose equivalent rate dH*(10)/dt for photon energies

be-tween 65 keV and 6700 keV

2.5.2 IC2017prep intercomparison

The objective of the IC2017prep intercomparison, which took place

between October 2017 and April 2018, was to evaluate passive H*(10)

dosimeters that could be used in the aftermath of a radiological or nu-clear event The dosimeters were exposed in two reference sites of the PTB: a free field (terrestrial and secondary cosmic radiation) and a free- floating platform (secondary cosmic radiation) In addition, irradiations were performed with Cs-137 at 0◦ and 90◦ Transport detectors were stored in a lead shielding at an underground facility, where the accu-mulated dose for a 6-month period is ~0.5 μSv only For a complete detail of the intercomparison, please see Dombrowski (2019)

3 Results

3.1 Dose response and detection limit

To demonstrate the ability of the RPL detectors to measure the low doses involved in environmental dosimetry and their uncertainties, we first analyzed twenty RPL detectors that were annealed and stored in the laboratory for a period of up to ~270 days The same detectors were read repeatedly at different periods (pre-heated only the first time), in some cases with two identical readers, here called reader 1 and reader 2,

in the same day

Fig 1a shows the indicated value of the dosimeters as a function of the time interval since the PD measurement The indicated value follows essentially the line corresponding to a dose rate of 1.9 μSv/day, which corresponds to the estimated background radiation level at PSI A broadening of the response with time within the dosimeters was observed, but the reason for this broadening is unclear Previous data has shown that the coefficient of variation for glasses is below 2.0% for

doses >0.5 mSv (Assenmacher et al., 2017), whereas the coefficient of variation in Fig 1a is ~4.0% for the highest doses Therefore, this broadening can only be partially explained by difference in glass sensitivities

Subtracting the fixed dose of 1.9 μSv/day from the data, we obtain the background subtracted doses shown in Fig 1b, where the black data points are for the measurements obtained using reader 1 and the red data points are for the measurements obtained using reader 2 The readouts with two identical RPL readers were performed to demonstrate the equivalence of the readers For short periods the intra-day variation (detector-to-detector variability) is small, but there is a large inter-day variation This suggests that for short intervals the uncertainties are mostly due to reader sensitivity fluctuations As the time interval in-creases, however, the detector-to-detector variability starts to dominate One can also see that the increase in the indicated value is not linear, but the background-subtracted dose deviates by not more than ±10 μSv up

to 200 days since the PD measurement

Trang 4

The background-subtracted doses in Fig 1b can then be used to es-timate the detection limit of the system The detection limit (DL) was defined in a first approximation as DL ≅ 3.3 ×σ0, where σ0 is the standard deviation of the background (standard deviation of the net indicated value), assuming the standard deviation of the signal as being approximately the same as the standard deviation of the background (Currie, 1968; ISO, 2019) If the detection limit is plotted for each time interval, we obtain the black (reader 1) and red lines (reader 2) shown in

Fig 1c The detection limit increases with time, mainly because of the increase in the detector-to-detector variability with time

For short periods of time, however, the detection limit is under- estimated when using only the readouts in a single day, because the uncertainty is dominated by the day-to-day reader variation Therefore,

to improve this detection limit estimate we took all the detector readouts

up to 90 days and obtained the mean net indicated value of (2.9 ± 5.0)

μSv, which leads to the detection limit of ~16.5 μSv indicated in Fig 1c This value is well below the lower dose measurement range of 50 μSv required by the Swiss dosimetry ordinance

For completeness, Fig 2 presents the dose response for the set of dosimeters irradiated with doses from 0.010 mSv to 100 mSv The de-viation from linearity remains within 15% for the entire range, also satisfying the Swiss dosimetry ordinance

3.2 Comparison between RPL and TL dosimeters 3.2.1 Environmental monitoring

Fig 3 shows the RPL doses plotted versus the TLD (Al2O3:C) doses for all environmental dosimeters around PSI The solid line indicates the 1:1 relationship and the dashed lines indicate the ±10% deviation The data in Fig 3 shows an overall good agreement between the RPL and TL dosimetry systems, with an average daily dose of approximately 1.9 μSv Deviations outside the ±10% are observed in a few quarters (e

g 2019Q01, 2019Q03) Nevertheless, this is more likely an underesti-mation from the TLDs As mentioned before (Section 2.1), the same TLDs are used in alternating quarters, e.g in 2019Q01, 2019Q03, which in-dicates that the same sets of TLDs are affected

Fig 4 shows a box-and-whisker plot of the average RPL and TLD values for all quarters The box indicate approximately the first and third quartiles, the black dot in the center represents the median, the whiskers represent the largest and smallest data points within 1.5 times the size of the box from the quartiles, and the points outside are considered as

Fig 1 (a) Indicated values from unirradiated detectors as a function of time

since pre-dose (PD) measurement, (b) background subtracted dose for the same

detectors and (c) detection limit based on the standard deviation of the

de-tectors for reader 1 (black) and reader 2 (red), as well as the detection limit

based on all measurements within the first 90 days (16.5 μSv line) The red

dashed line in (a) represents the expected dose assuming a dose of 1.9 μSv/day

The error bars represent the standard deviation of the data, i.e., the range of

variation from detector-to-detector and, therefore, provide an estimate of the

uncertainty of each detector See text for details (For interpretation of the

references to color in this figure legend, the reader is referred to the Web

version of this article.)

Fig 2 Dose response of detectors irradiated with various doses read out 16–17

days after the PD measurement The solid line represents the 1:1 relationship and the dashed lines the ±15% deviations The error bars are barely visible in this graph

Trang 5

Fig 3 Comparison between doses per day estimated using RPL dosimeters and Al2O3:C TLDs for environmental measurements around PSI The solid line indicates the 1:1 relation, whereas the dashed lines indicate a ±10% deviation

Fig 4 Comparison between the average dose per day estimated using RPL and TLD dosimeters in the environmental monitoring network of PSI’s Dosimetry

Laboratory, presented as box-and-whisker plots (see text for details)

Trang 6

“outliers” and are used to indicate when the distributions are skewed

(Dalgaard, 2008) In this graph it is possible to see that the median RPL

value stays constant over time, whereas the TLD system occasionally

shows lower values than the RPL system (e.g 2019Q01)

The actual mean dose in these measured points is not known and,

therefore, one cannot affirm that one system is more correct than the

other Nevertheless, one can state that the RPL system showed a more

constant mean dose per day over the period investigated

3.2.2 Area monitoring at PSI inside controlled areas

Fig 5 compares the doses measured using RPL with those measured

using 7LiF:Mg,Ti TLDs inside controlled areas at PSI The doses are again

below the reference levels per quarter (80 mSv/year or 20 mSv/

quarter), even though the data points include locked-off areas

Furthermore, the RPL doses correlate with the TLD doses, with a few

exceptions

3.3 Comparison between RPL and EDIS-1 dosimeters

Fig 6 compares the RPL data with the EDIS-1 dosimeters exposed

during two quarters at various measurement points at KKL (see Section

2.4) The figure shows that the points with higher dose rate are correctly

tracked by both RPL and EDIS-1 dosimeters The data again shows an

overall good agreement between the two dosimetry systems, with the

EDIS-1 showing a higher dose than the RPL dosimeters, ~14% in

2019Q04 and ~12% in 2020Q01 This difference is not surprising, given

the different construction of the RPL dosimeters and the EDIS-1

do-simeters and their expected different photon energy response

3.4 Intercomparisons

3.4.1 Swiss intercomparison

The results of the Swiss intercomparison are presented in Table 2 for

both Al2O3:C TLDs and RPL dosimeters Because of the small number of

detectors used, all values are presented and only the mean of the relative

response was calculated

The results show an overall agreement of the TLDs and RPL

dosim-eters in reference conditions (Cs-137) for both doses, with deviations <

5% from the reference value For the field irradiation, both TLD and RPL showed an under-response of ~8% in comparison with the dose esti-mated using the RSDetection pressurized ionization chamber, but both TLD and RPL systems showed the same dose

3.4.2 IC2019prep intercomparison

IC2019prep results for the RPL dosimetry system are shown in

Table 3 The initial objective was to compare the TL and the RPL dosimetry systems, but the light-sensitive Al2O3:C TLDs were exposed to light by mistake and could not be evaluated

The results for the reference irradiation with Cs-137 at 0◦showed a good agreement with the conventional true value The irradiation with Cs-137 at 90◦ showed an under-response, as also reported by Assen-macher et al (2020) The results for both free field exposure (terrestrial plus secondary cosmic radiation) and floating platform exposure (sec-ondary cosmic radiation) were in good agreement with the conventional true values, with an over-response of the RPLs between 8 and 14%

4 Conclusions

The results reported here provide further supporting data for the application of RPL in environmental and area dosimetry

The laboratory irradiations show that the detection limit of the RPL dosimeters is below the 50 μSv up to ~200 total days of use (time be-tween pre-dose measurement and readout) Moreover, the response is linear up to 100 mSv, with less than 15% deviation Both characteristics satisfy the requirements of the Swiss dosimetry ordinance for passive environmental dosimeters

The comprehensive comparison between the RPL, TLD and EDIS-1 dosimetry systems for the environmental and area measurement points in various locations (outside the PSI perimeter, inside PSI perimeter inside controlled areas, environmental measurements at KKL) showed a good agreement between the techniques In general the environmental doses obtained using the RPL system were more constant

Fig 5 RPL versus 7LiF:Mg,Ti TLD doses measured inside controlled areas at PSI The solid line indicates the 1:1 relation, whereas the dashed lines indicate a

±10% deviation

Trang 7

over time, the dose per day remaining around 1.9 μSv, whereas the TLD

system shows larger deviations from this value In locations where

higher dose rates are expected, the RPLs also showed reliable

perfor-mance, tracking the TLD or EDIS-1 doses

In the case of the comparison with EDIS-1 dosimeters, slightly lower

doses (12–14%) were observed using RPL Because of the very different

construction of the EDIS-1 dosimeters, a large difference in the energy

dependence between the two types of dosimeters is expected and,

therefore, such discrepancies are not unusual

The intercomparisons for passive dosimeters in terms of H*(10)

showed a good agreement between the RPL dose values and the

conventional true values In the Swiss intercomparison, both TLD and RPL results under-estimated the conventional true value by ~8%, whereas in the IC2019prep intercomparison the RPL over-estimated the conventional true value by ~8–14%, depending on the radiation field These discrepancies are not related to the calibration of the dosimeters, which was independently checked using Cs-137 irradiations

The results presented here, combined with the characterization of the RPL dosimetry system for personal and environmental applications (Assenmacher et al., 2017, 2020), support the adoption of the RPL dosimetry system for environmental dosimetry

Declaration of competing interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper

Acknowledgements

We thank the dosimetry service of the Leibstadt Nuclear Power Plant (Kernkraftwerk Leibstadt, KKL) for providing the data on the EDIS-1 detectors and Sabine Mayer (PSI) for comments on the manuscript This work was funded by the Swiss Federal Nuclear Safety Inspectorate ENSI, contracts no CTR00491

Fig 6 Comparison between EDIS-1 and RPL dosimeters at KKL for various measurement points and two quarters

Table 2

Conventional true value Ht and indicated value Hm for Al2O3:C TLDs and the RPL dosimetry systems for the 2016 Swiss intercomparison for environmental dosimeters

The uncertainties correspond to the standard deviation of the mean multiplied by the corresponding tp(ν) value for a 95.45% level of confidence (k = 2) from the t-

distribution for degrees of freedom ν, as in Table G.2 from the ISO/IEC Guide 99–3:2008 (ISO/IEC, 2008)

Radiation field Ht [mSv] Hm [mSv] (TLD) H m /H t(TLD) Hm [mSv] (RPL) H m /H t(RPL)

0.294 0.97 ± 0.05 0.311 0.295 1.0 ± 0.4

1.694 0.996 ± 0.008 1.636 1.635 0.962 ± 0.004 Combined terrestrial/secondary cosmic 0.182 0.164

0.175 0.164 0.163

0.92 ± 0.05 0.170

0.176 0.162 0.164

0.92 ± 0.06

Table 3

Conventional true value and indicated value for the RPL system for the

IC2019prep intercomparison The uncertainties correspond to the standard

de-viation of the mean multiplied by the corresponding tp(ν) value for a 95.45%

level of confidence (k = 2) from the t-distribution for degrees of freedom ν, as in

Table G.2 from the ISO/IEC Guide 99–3:2008 (ISO/IEC, 2008)

Radiation field H t [mSv] Hm [mSv] Hm/Ht

Cs-137, 0 ◦ 30.0 30.37 ± 0.23 1.012 ± 0.008

Cs-137, 90 ◦ 30.0 19 ± 3 0.64 ± 0.10

Free field 0.294 0.318 ± 0.018 1.08 ± 0.06

Floating platform 0.137 0.156 ± 0.005 1.14 ± 0.03

Trang 8

References

Assenmacher, F., Boschung, M., Hohmann, E., Mayer, S., 2017 Dosimetric properties of a

personal dosimetry system based on radiophotoluminescence of silver doped

phosphate glass Radiat Meas 106, 235–241

Assenmacher, F., Musto, E., Yukihara, E.G., 2020 Dosimetric properties of an

environmental H*(10) dosemeter based on radiophotoluminescence of silver doped

phosphate glass Radiat Meas 138, 106468

Currie, L.A., 1968 Limits for qualitative detection and quantitative determination -

application to radiochemistry Anal Chem 40, 586–593

Dalgaard, P., 2008 Introductory Statistics with R Springer, New York

Dombrowski, H., 2019 Preparedness intercomparison of passive H*(10) area photon

dosemeters in 2017/2018 (IC2017prep) J Instrum 14, P10008

Hocine, N., Donadille, L., Huet, C., Iti´e, C., Clairand, I., 2011 Personal monitor glass

badge: theoretical dosimeter response calculated with the Monte Carlo transport

code MCNPX Radiat Protect Dosim 144, 231–233

Hocine, N., 2012 Calculated angular responses of an RPL dosimeter to photon and beta

radiation Radiat Protect Dosim 151, 374–378

IEC, 2020 International Standard IEC 62387:2020-01: radiation protection

instrumentation – dosimetry systems with integrating passive detectors for

individual, workplace and environmental monitoring of photon and beta radiation

In: International Electrotechnical Commission, Geneva

ISO, 1999 International Standard ISO 4037-3: X and Gamma Reference Radiation for

Calibrating Dosemeters and Doserate Meters and for Determining Their Response as

a Function of Photon Energy - Part 3: Calibration of Area and Personal Dosemeters and the Measurement of Their Response as a Function of Energy and Angle of Incidence International Organization for Standardization, Geneva

ISO, 2019 International Standard ISO 11929:2019: determination of the characteristic limits (decision threshold, detection limit and limits of the confidence interval) for measurements of ionizing radiation – fundamentals and application - Part 1: elementary applications In: International Organisation for Standardisation

ISO/IEC, 2008 ISO/IEC Guide 98-3: Guide to the Expression of Uncertainty in Measurement (GUM:1995) International Organization for Standardization, Geneva

ISO/IEC, 2017 ISO/IEC 17025:2017: General Requirements for the Competence of Testing and Calibration Laboratories International Organization for Standardization, Geneva

Miyamoto, Y., Takei, Y., Nanto, H., Kurobori, T., Konnai, A., Yanagida, T., Yoshikawa, A., Shimotsuma, Y., Sakakura, M., Miura, K., Hirao, K., Nagashima, Y., Yamamoto, T.,

2011 Radiophotoluminescence from silver-doped phosphate glass Radiat Meas 46, 1480–1483

Ranogajec-Komor, M., Knezevic, Z., Miljanic, S., Vekic, B., 2008 Characterisation of radiophotoluminescent dosimeters for environmental monitoring Radiat Meas 43, 392–396

Swiss Federal Council, 2017 Verordnung des EDI über die Personen- und Umgebungsdosimetrie (Dosimetrieverordnung)

Tereda, H., Sakai, E., Katagiri, M., 1980 Environmental gamma-ray exposure rates measured by in-situ Ge(Li) spectrometer J Nucl Sci Technol 17, 281–290

Ngày đăng: 24/12/2022, 02:10

TÀI LIỆU CÙNG NGƯỜI DÙNG

TÀI LIỆU LIÊN QUAN

🧩 Sản phẩm bạn có thể quan tâm